A Review on MR Image Intensity Inhomogeneity Correction
Open accessThe DOI matches its Crossref record; 84 of the 84 references it cites were found in a registry or an index.
Abstract
Intensity inhomogeneity (IIH) is often encountered in MR imaging, and a number of techniques have been devised to correct this artifact. This paper attempts to review some of the recent developments in the mathematical modeling of IIH field. Low‐frequency models are widely used, but they tend to corrupt the low‐frequency components of the tissue. Hypersurface models and statistical models can be adaptive to the image and generally more stable, but they are also generally more complex and consume more computer memory and CPU time. They are often formulated together with image segmentation within one framework and the overall performance is highly dependent on the segmentation process. Beside these three popular models, this paper also summarizes other techniques based on different principles. In addition, the issue of quantitative evaluation and comparative study are discussed.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Stochastic relaxation, gibbs distributions, and the bayesian restoration of images.
Geman, S., & Geman, D. (1984) · IEEE Transactions on Pattern Analysis and Machine Intelligence · doi:10.1109/tpami.1984.4767596 · Cited by 17,949 · cited 1×
“A useful tool for spatial modeling is the Markov random field (MRF), which is first employed by Geman and Geman [72] for image segmentation.”
MethodVoxel-Based Morphometry—The Methods
Ashburner, J., & Friston, K. J. (2000) · NeuroImage · doi:10.1006/nimg.2000.0582 · Cited by 8,676 · cited 1×
“In this way, the two different problems, IIH correction and image segmentation, are formulated in one framework and solved simultaneously.”
Fischl, B., Salat, D. H., Busa, E., Albert, M., Dieterich, M., Haselgrove, C., van der Kouwe, A., Killiany, R., Kennedy, D., Klaveness, S., et al. (2002) · Neuron · doi:10.1016/s0896-6273(02)00569-x · Cited by 8,610
Ashburner, J., & Friston, K. J. (2005) · NeuroImage · doi:10.1016/j.neuroimage.2005.02.018 · Cited by 7,507 · cited 1×
“In addition, there even exist attempts to address these three processes, IIH correction, segmentation, and registration together [93, 94].”
Zhang, Y., Brady, M., & Smith, S. (2001) · IEEE Transactions on Medical Imaging · doi:10.1109/42.906424 · Cited by 6,982 · cited 1×
“A large category of surface fitting approaches search the parameters by fitting with respect to a set of tissue points encoding information about the IIH.”
BackgroundSpatial Interaction and the Statistical Analysis of Lattice Systems
Besag, J. (1974) · Journal of the Royal Statistical Society Series B: Statistical Methodology · doi:10.1111/j.2517-6161.1974.tb00999.x · Cited by 6,067
A nonparametric method for automatic correction of intensity nonuniformity in MRI data
Sled, J. G., Zijdenbos, A. P., & Evans, A. C. (1998) · IEEE Transactions on Medical Imaging · doi:10.1109/42.668698 · Cited by 4,731
Non-Uniform Random Variate Generation
Devroye, L. (1986) · Springer eBooks · doi:10.1007/978-1-4613-8643-8 · Cited by 4,041
Spatial registration and normalization of images
Friston, K. J., Ashburner, J., Frith, C. D., Poline, J. ‐., Heather, J. D., & Frackowiak, R. S. J. (1995) · Human Brain Mapping · doi:10.1002/hbm.460030303 · Cited by 4,028 · cited 1×
“Image registration has also been utilized to aid the IIH correction [89, 90].”
MethodRoemer, P. B., Edelstein, W. A., Hayes, C. E., Souza, S. P., & Mueller, O. M. (1990) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910160203 · Cited by 2,322
Cited by
N4ITK: Improved N3 Bias Correction
Tustison, N. J., Avants, B. B., Cook, P. A., Zheng, Y., Egan, A., Yushkevich, P. A., & Gee, J. C. (2010) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2010.2046908 · Cited by 6,073
Minimization of Region-Scalable Fitting Energy for Image Segmentation
Li, C., Kao, C.-Y., Gore, J. C., & Ding, Z. (2008) · IEEE Transactions on Image Processing · doi:10.1109/tip.2008.2002304 · Cited by 1,721
State of the art survey on MRI brain tumor segmentation
Gordillo, N., Montseny, E., & Sobrevilla, P. (2013) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2013.05.002 · Cited by 697
Lesion identification using unified segmentation-normalisation models and fuzzy clustering
Seghier, M. L., Ramlackhansingh, A., Crinion, J., Leff, A. P., & Price, C. J. (2008) · NeuroImage · doi:10.1016/j.neuroimage.2008.03.028 · Cited by 317
What are normal relaxation times of tissues at 3 T?
Bojorquez, J. Z., Bricq, S., Acquitter, C., Brunotte, F., Walker, P. M., & Lalande, A. (2017) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2016.08.021 · Cited by 286
Segmentation of multiple sclerosis lesions in brain MRI: A review of automated approaches
Lladó, X., Oliver, A., Cabezas, M., Freixenet, J., Vilanova, J. C., Quiles, A., Valls, L., Ramió-Torrentà, L., & Rovira, À. (2012) · Information Sciences · doi:10.1016/j.ins.2011.10.011 · Cited by 240
“Two different studies have recently reviewed different ways to overcome this preprocessing step problem [52,105].”
BackgroundSynthMorph: Learning Contrast-Invariant Registration Without Acquired Images
Hoffmann, M., Billot, B., Greve, D. N., Iglesias, J. E., Fischl, B., & Dalca, A. V. (2022) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2021.3116879 · Cited by 201
“We further corrupt the image with a spatially varying intensity-bias field B [74], [75].”
BackgroundKraus, M. F., Liu, J. J., Schottenhamml, J., Chen, C.-L., Budai, A., Branchini, L., Ko, T., Ishikawa, H., Wollstein, G., Schuman, J., et al. (2014) · Biomedical Optics Express · doi:10.1364/boe.5.002591 · Cited by 169
“For each input volume I, illumination effects are estimated from the fundus projection , 1 ( , , ) i j k x y z Based on the assumption that illumination effects have low spatial frequency, each fundus projection is low pass filtered using a 2D Gaussian filter with a 10 pixel standard deviation [18].”
MethodLCC-Demons: A robust and accurate symmetric diffeomorphic registration algorithm
Lorenzi, M., Ayache, N., Frisoni, G. B., & Pennec, X. (2013) · NeuroImage · doi:10.1016/j.neuroimage.2013.04.114 · Cited by 162
“Bias correction is often performed prior to image registration in order to remove global intensity inhomogeneity (Hou, 2006; Sled et al., 1998; Tustison et al., 2010), but this is often not sufficient to remove local changes in the bias field.”
BackgroundSudre, C. H., Cardoso, M. J., Bouvy, W. H., Biessels, G. J., Barnes, J., & Ourselin, S. (2015) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2015.2419072 · Cited by 162
“1) Intensity Inhomogeneity Correction (IIH): Intensity inhomogeneities, also called bias field, may appear in MR images due to spatial inhomogeneity of the scanner main magnetic field during the acquisition of the image [43].”
BackgroundFulton, L., Dolezel, D., Harrop, J., Yan, Y., & Fulton, C. (2019) · Brain Sciences · doi:10.3390/brainsci9090212 · Cited by 158
“noise) not related to the patient’s anatomical image which include patient movement, nearby static fields, radio-frequency variation, or other non-patient source [23].”
BackgroundMethodSegmentation and quantification of adipose tissue by magnetic resonance imaging
Hu, H. H., Chen, J., & Shen, W. (2016) · Magnetic Resonance Materials in Physics, Biology and Medicine · doi:10.1007/s10334-015-0498-z · Cited by 90
A review of Methods for Bias Correction in Medical Images
Song, S., Zheng, Y., & He, Y. (2017) · Biomedical Engineering Review · doi:10.18103/bme.v3i1.1550 · Cited by 62
Metadata
- Authors
- Hou, Z.
- Venue
- International Journal of Biomedical Imaging
- Publisher
- Wiley
- Published
- 2006-01
- Volume
- 2006
- Issue
- 1
- Pages
- 49515-49515
- ISSN
- 1687-4188 (ISSN-L)
- 1687-4196
- OpenAlex
- W2008101963
- Cited by
- 200
- References
- 84
- FWCI
- 10.19
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.