CoChat

Sirolimus-Eluting Stent for the Treatment of In-Stent Restenosis: A Quantitative Coronary Angiography and Three-Dimensional Intravascular Ultrasound Study

Open access
Sousa, J. Eduardo, Costa, Marco A., Abizaid, Alexandre, Sousa, Amanda G.M.R., Feres, Fausto, Mattos, Luiz A., Centemero, Marinella, Maldonado, Galo, Abizaid, Andrea S., Pinto, Ibraim, Falotico, Robert, Jaeger, Judith, Popma, Jeffrey J., Serruys, Patrick W. · 2003 · Circulation
Cited by 203

The DOI matches its Crossref record; 10 of the 10 references it cites were found in a registry or an index.

Abstract

Background— We have previously reported the safety and effectiveness of sirolimus-eluting stents for the treatment of de novo coronary lesions. The present investigation explored the potential of this technology to treat in-stent restenosis. Methods and Results— Twenty-five patients with in-stent restenosis were successfully treated with the implantation of 1 or 2 sirolimus-eluting Bx VELOCITY stents in São Paulo, Brazil. Nine patients received 2 stents (1.4 stents per lesion). Angiographic and volumetric intravascular ultrasound (IVUS) images were obtained after the procedure and at 4 and 12 months. All vessels were patent at the time of 12-month angiography. Angiographic late loss averaged 0.07±0.2 mm in-stent and −0.05±0.3 mm in-lesion at 4 months, and 0.36±0.46 mm in-stent and 0.16±0.42 mm in-lesion after 12 months. No patient had in-stent or stent margin restenosis at 4 months, and only one patient developed in-stent restenosis at 1-year follow-up. Intimal hyperplasia by 3-dimensional IVUS was 0.92±1.9 mm 3 at 4 months and 2.55±4.9 mm 3 after 1 year. Percent volume obstruction was 0.81±1.7% and 1.76±3.4% at the 4- and 12-month follow-up, respectively. There was no evidence of stent malapposition either acutely or in the follow-up IVUS images, and there were no deaths, stent thromboses, or repeat revascularizations. Conclusion— This study demonstrates the safety and the potential utility of sirolimus-eluting Bx VELOCITY stents for the treatment of in-stent restenosis.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1161/01.cir.0000047063.22006.41

Cites

10 of 10 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 10 references

Cited by

Metadata

Authors
Sousa, J. E., Costa, M. A., Abizaid, A., Sousa, A. G. M. R., Feres, F., Mattos, L. A., Centemero, M., Maldonado, G., Abizaid, A. S., Pinto, I., Falotico, R., Jaeger, J., Popma, J. J., & Serruys, P. W.
Venue
Circulation
Published
2003-01-07
Volume
107
Issue
1
Pages
24-27
ISSN
  • 0009-7322 (ISSN-L)
  • 1524-4539
OpenAlex
W2167351939
Cited by
203
References
10
FWCI
44.19
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).