Comparison of Automated Office Blood Pressure With Office and Out-Off-Office Measurement Techniques: A Systematic Review and Meta-Analysis
Open accessThe DOI matches its Crossref record; 9 of the 9 references it cites were found in a registry or an index.
Abstract
Automated office blood pressure (AOBP) has emerged as a valuable tool to assess patient’s BP status, but the lack of strong evidence to establish a threshold value for hypertension diagnosis limits its use in clinical practice. We aimed at synthesizing the published literature through a meta-analysis of studies comparing AOBP with other BP measurement techniques and at analyzing the differences between AOBP and physician’s office BP, nonphysician’s office BP, daytime ambulatory BP monitoring, and home BP monitoring. We searched PubMed database for articles published up to April 2018; eligible studies compared AOBP with office and out-of-office measurement techniques and reported the BP differences or BP values obtained. Twenty-six studies, for a total of 7116 patients, were included in the analysis. AOBP values were lower than physician (systolic BP, −10.48 mm Hg; 95% CI, −13.15 to −7.81/diastolic BP, −4.44 mm Hg; 95% CI, −6.07 to −2.80) and nonphysician office ones (systolic BP, −6.89 mm Hg; 95% CI, −8.75 to −5.04/diastolic BP −3.82 mm Hg; 95% CI, −4.86 to −2.78). No significant differences were found between AOBP and daytime ambulatory BP monitoring (systolic BP, −1.85; 95% CI, −4.50 to 0.79/diastolic BP, 0.12; 95% CI, −1.42 to 1.66) and home BP monitoring (systolic BP, −2.65; 95% CI, −8.42 to 3.12/diastolic BP, −1.67; 95% CI, −4.20 to 0.87). AOBP readings did not differ significantly from out-of-office blood pressure, still remaining an office technique; it may improve hypertension diagnosis by overcoming some of office BP limitations, including the white coat effect.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Meta-analyses show that standardized automated office BP provides BP measurements that closely approximate daytime ambulatory BP monitoring (ABPM) and home BP monitoring (HBPM). 23 , 24”
From Hypertension Canada guideline for the diagnosis and treatment of hypertension in adults in primary care (2025) · cites this paper as [23]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
2018 ESC/ESH Guidelines for the management of arterial hypertension
Williams, B., Mancia, G., Spiering, W., Agabiti Rosei, E., Azizi, M., Burnier, M., Clement, D. L., Coca, A., de Simone, G., Dominiczak, A., et al. (2018) · European Heart Journal · doi:10.1093/eurheartj/ehy339 · Cited by 10,286
Whelton, P. K., Carey, R. M., Aronow, W. S., Casey, D. E., Collins, K. J., Dennison Himmelfarb, C., DePalma, S. M., Gidding, S., Jamerson, K. A., Jones, D. W., et al. (2018) · Hypertension · doi:10.1161/hyp.0000000000000065 · Cited by 7,391
Whelton, P. K., Carey, R. M., Aronow, W. S., Casey, D. E., Collins, K. J., Dennison Himmelfarb, C., DePalma, S. M., Gidding, S., Jamerson, K. A., Jones, D. W., et al. (2018) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2017.11.006 · Cited by 6,616
A Randomized Trial of Intensive versus Standard Blood-Pressure Control
The SPRINT Research Group (2015) · New England Journal of Medicine · doi:10.1056/nejmoa1511939 · Cited by 6,325
Closing the Gap between Methodologists and End-Users: R as a Computational Back-End
Wallace, B. C., Dahabreh, I. J., Trikalinos, T. A., Lau, J., Trow, P., & Schmid, C. H. (2012) · Journal of Statistical Software · doi:10.18637/jss.v049.i05 · Cited by 1,125
Nerenberg, K. A., Zarnke, K. B., Leung, A. A., Dasgupta, K., Butalia, S., McBrien, K., Harris, K. C., Nakhla, M., Cloutier, L., Gelfer, M., et al. (2018) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2018.02.022 · Cited by 573
Lovibond, K., Jowett, S., Barton, P., Caulfield, M., Heneghan, C., Hobbs, F. R., Hodgkinson, J., Mant, J., Martin, U., Williams, B., et al. (2011) · The Lancet · doi:10.1016/s0140-6736(11)61184-7 · Cited by 311
Measurement of Blood Pressure in the Office: Recognizing the Problem and Proposing the Solution
Myers, M. G., Godwin, M., Dawes, M., Kiss, A., Tobe, S. W., & Kaczorowski, J. (2010) · Hypertension · doi:10.1161/hypertensionaha.109.141879 · Cited by 260
Comparability of two commonly used automated office blood pressure devices in the severely obese
Padwal, R. S., & Majumdar, S. R. (2016) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000196 · Cited by 9
Research works printed but not linked: 48
- 10.1016/S0895-7061(03)00058-X · DOI
- 10.1097/HJH.0b013e3283577b05 · DOI · Paper page
- 10.1186/1471-2261-5-18 · DOI · Paper page
- 10.1097/HJH.0b013e32831b9e6b · DOI · Paper page
- 10.1097/HJH.0b013e328356abd3 · DOI
- 10.1097/MBP.0b013e328340d128 · DOI · Paper page
- 10.1097/HJH.0000000000000319 · DOI
- 10.3109/08037051.2015.1134086 · DOI · Paper page
- 10.1097/HJH.0000000000001076 · DOI
- 10.1097/01.mbp.0000189794.36230.a7 · DOI
- 10.1097/MBP.0b013e32832c5167 · DOI · Paper page
- 10.1097/MBP.0b013e328346e0db · DOI
- 10.1093/fampra/cmq067 · DOI · Paper page
- 10.1136/bmj.d286 · DOI · Paper page
- 10.1016/j.cjca.2011.05.001 · DOI
- 10.1097/MBP.0b013e3283540785 · DOI · Paper page
- 10.1038/ajh.2012.76 · DOI
- 10.1016/j.jash.2013.07.002 · DOI
- 10.3747/pdi.2012.00027 · DOI
- 10.1097/MBP.0000000000000133 · DOI · Paper page
- 10.1016/j.jash.2014.11.004 · DOI
- 10.1016/j.cjca.2017.09.011 · DOI
- 10.1016/j.jash.2017.01.009 · DOI
- 10.1016/j.cjca.2017.01.021 · DOI
- 10.1161/HYPERTENSIONAHA.117.09464 · DOI
- 10.1161/HYPERTENSIONAHA.117.10026 · DOI · Paper page
- 10.1080/08037051.2018.1425606 · DOI
- 10.1159/000487900 · DOI
- 10.1016/j.cjca.2017.09.022 · DOI
- 10.1161/JAHA.118.008994 · DOI
- Glantz, SA. Statistica per Discipline Biomediche. 6th edition. McGraw-Hill Education; 2007. https://www.mheducation.com.
- 10.1038/sj.jhh.1001626 · DOI
- 10.1016/j.jash.2011.08.007 · DOI
- 10.1016/j.jacc.2016.08.035 · DOI
- 10.3399/bjgp14X677851 · DOI · Paper page
- 10.1161/01.hyp.9.2.209 · DOI · Paper page
- 10.1097/HJH.0b013e3283605c71 · DOI
- 10.1093/ajh/hpx180 · DOI
- 10.1093/ajh/hpy053 · DOI
- 10.1097/HJH.0000000000001775 · DOI
- 10.1038/sj.jhh.1002190 · DOI
- 10.1038/ajh.2011.38 · DOI
- Myers, MG. Self-measurement of blood pressure at home: the potential for reporting bias. Blood Press Monit. 1998;3(suppl 1):S19–S22.
- 10.1016/S0895-7061(98)00241-6 · DOI · Paper page
- 10.1097/MBP.0b013e32831e314f · DOI
- 10.1016/j.cjca.2017.01.020 · DOI
- Omron Healthcare Inc. IntelliSense Blood Pressure Monitor Model HEM 907XL. Omron Instruction Manual.
- BpTRU Medical Devices. Operator’s Manual.
Cited by
(2022) · PubMed Central · doi:10.6515/acs.202205_38(3).20220321a · Cited by 117
Goupil, R., Tsuyuki, R. T., Santesso, N., Terenzi, K. A., Habert, J., Cheng, G., Gysel, S. C., Bruneau, J., Leung, A. A., Campbell, N. R. C., et al. (2025) · Canadian Medical Association Journal · doi:10.1503/cmaj.241770 · Cited by 60
Karnjanapiboonwong, A., Anothaisintawee, T., Chaikledkaew, U., Dejthevaporn, C., Attia, J., & Thakkinstian, A. (2020) · BMC Cardiovascular Disorders · doi:10.1186/s12872-020-01736-2 · Cited by 27
Metadata
- Authors
- Pappaccogli, M., Di Monaco, S., Perlo, E., Burrello, J., D’Ascenzo, F., Veglio, F., Monticone, S., & Rabbia, F.
- Venue
- Hypertension
- Published
- 2019-02
- Volume
- 73
- Issue
- 2
- Pages
- 481-490
- ISSN
- 0194-911X (ISSN-L)
- 1524-4563
- OpenAlex
- W2906108284
- Cited by
- 74
- References
- 9
- FWCI
- 7.17
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.