In Silico Preclinical Trials: A Proof of Concept in Closed-Loop Control of Type 1 Diabetes
Open accessThe DOI matches its Crossref record; 49 of the 49 references it cites were found in a registry or an index.
Abstract
Arguably, a minimally invasive system using subcutaneous (s.c.) continuous glucose monitoring (CGM) and s.c. insulin delivery via insulin pump would be a most feasible step to closed-loop control in type 1 diabetes mellitus (T1DM). Consequently, diabetes technology is focusing on developing an artificial pancreas using control algorithms to link CGM with s.c. insulin delivery. The future development of the artificial pancreas will be greatly accelerated by employing mathematical modeling and computer simulation. Realistic computer simulation is capable of providing invaluable information about the safety and the limitations of closed-loop control algorithms, guiding clinical studies, and out-ruling ineffective control scenarios in a cost-effective manner. Thus computer simulation testing of closed-loop control algorithms is regarded as a prerequisite to clinical trials of the artificial pancreas. In this paper, we present a system for in silico testing of control algorithms that has three principal components: (1) a large cohort of n = 300 simulated “subjects” ( n = 100 adults, 100 adolescents, and 100 children) based on real individuals' data and spanning the observed variability of key metabolic parameters in the general population of people with T1DM; (2) a simulator of CGM sensor errors representative of Freestyle Navigator™, Guardian RT, or Dexcom™ STS™, 7-day sensor; and (3) a simulator of discrete s.c. insulin delivery via OmniPod Insulin Management System or Deltec Cozmo® insulin pump. The system has been shown to represent adequate glucose fluctuations in T1DM observed during meal challenges, and has been accepted by the Food and Drug Administration as a substitute to animal trials in the preclinical testing of closed-loop control strategies.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“The characteristics of these”
From Developing the UVA/Padova Type 1 Diabetes Simulator: Modeling, Validation, Refinements, and Utility (2023) · cites this paper as [66]
In the citing paper
“This UVA-PADOVA Type 1 Diabetes Simulator (S2008) emulated meal challenges and included a population of 300 in silico subjects (100 adults, 100 adolescents, 100 children).”
From The UVA/PADOVA Type 1 Diabetes Simulator: New Features (2014) · cites this paper as [4]
In the citing paper
“However, the extensive use of computer simulation started in 2008 when a paradigm change occurred in the field of type 1 diabetes mellitus (T1DM): for the first time a computer model has been accepted by FDA as a substitute for preclinical trials of certain insulin treatments, including closed-loop algorithms. 4”
From The UVA/PADOVA Type 1 Diabetes Simulator: New Features (2014) · cites this paper as [4]
In the citing paper
“Thus, the creation and validation of the in silico population involved proving that the parameters of the in silico population cover well key parameter distributions observed in vivo and provide comprehensive analysis of control performance.66,67”
From Developing the UVA/Padova Type 1 Diabetes Simulator: Modeling, Validation, Refinements, and Utility (2023) · cites this paper as [66]
In the citing paper
“The tracer-based”
From Developing the UVA/Padova Type 1 Diabetes Simulator: Modeling, Validation, Refinements, and Utility (2023) · cites this paper as [66]
5 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Continuous glucose monitoring and intensive treatment of type 1 diabetes
(2008) · New England Journal of Medicine · doi:10.1056/nejmoa0805017 · Cited by 1,559
Nonlinear model predictive control of glucose concentration in subjects with type 1 diabetes
Hovorka, R., Canonico, V., Chassin, L. J., Haueter, U., Massi-Benedetti, M., Federici, M. O., Pieber, T. R., Schaller, H. C., Schaupp, L., Vering, T., et al. (2004) · Physiological Measurement · doi:10.1088/0967-3334/25/4/010 · Cited by 1,274
Meal Simulation Model of the Glucose-Insulin System
Dalla Man, C., Rizza, R. A., & Cobelli, C. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.893506 · Cited by 955 · cited 1×
“[20,30] mg/dl 3.”
Self-citationKlonoff, D. C. (2005) · Diabetes Care · doi:10.2337/diacare.28.5.1231 · Cited by 724
Weinzimer, S. A., Steil, G. M., Swan, K. L., Dziura, J., Kurtz, N., & Tamborlane, W. V. (2008) · Diabetes Care · doi:10.2337/dc07-1967 · Cited by 567
Deiss, D., Bolinder, J., Riveline, J.-P., Battelino, T., Bosi, E., Tubiana-Rufi, N., Kerr, D., & Phillip, M. (2006) · Diabetes Care · doi:10.2337/dc06-1134 · Cited by 539
Feasibility of Automating Insulin Delivery for the Treatment of Type 1 Diabetes
Steil, G. M., Rebrin, K., Darwin, C., Hariri, F., & Saad, M. F. (2006) · Diabetes · doi:10.2337/db06-0419 · Cited by 473
Improvement in Glycemic Excursions With a Transcutaneous, Real-Time Continuous Glucose Sensor
Garg, S., Zisser, H., Schwartz, S., Bailey, T., Kaplan, R., Ellis, S., & Jovanovic, L. (2006) · Diabetes Care · doi:10.2337/diacare.29.01.06.dc05-1686 · Cited by 433
(1985) · DSpace@MIT (Massachusetts Institute of Technology) · Cited by 417
An Artificial Endocrine Pancreas
Albisser, A. M., Leibel, B. S., Ewart, T. G., Davidovac, Z., Botz, C. K., & Zingg, W. (1974) · Diabetes · doi:10.2337/diab.23.5.389 · Cited by 379
Research works printed but not linked: 4
- Clemens AH, 1977, Horm Metab Res., 23
- Fischer U, 1978, Endokrinologie., 71, 65
- Clemens AH, 1979, Med Prog Technol., 6, 91
- Zisser H, Shwartz S, Ratner R, Wise J, Bailey T. Accuracy of a seven-day continuous glucose sensor compared to YSI blood glucose values. Proc 27th Workshop of the AIDPIT Study Group. Innsbruck, Austria. AIDPIT; 2008. p. S03. http://www.aidpit.org.
Cited by
Items 501–600 of 870
Rad, F. S., & Li, J. (2025) · JMIR Diabetes · doi:10.2196/72874 · Cited by 7
Piratla, V. S. A., Saxena, S., Bhatia, S., & Kumar, N. (2025) · Intelligent Cybersecurity Conference (ICSC) · doi:10.1109/icsc65596.2025.11140338 · Cited by 7
“…algorithms have been developed and implemented, each with its own strengths and limitations. i. Proportional-Integral-Derivative Control: PID controllers are a classic control algorithm widely used in industrial applications and have been adapted in systems [2, 4] where k = 0 , . . . , N − 1 .”
InfluentialSaday, A., Yildirim, Y. C., & Tekin, C. (2023) · neural information processing systems · doi:10.52202/075280-3032 · Cited by 7
In silico Models of Alcohol Dependence and Treatment
Kovatchev, B. P., Breton, M. D., & Johnson, B. A. (2012) · Frontiers in Psychiatry · doi:10.3389/fpsyt.2012.00004 · Cited by 7
“However, while population averages are sufficient to simulate silico experiments at an individual level (Kovatchev et al., 2009).”
BackgroundMethodSelf-citationModeling the acute effects of exercise on glucose dynamics in healthy nondiabetic subjects
Frank, S., Jbaily, A., Hinshaw, L., Basu, R., Basu, A., & Szeri, A. J. (2021) · Journal of Pharmacokinetics and Pharmacodynamics · doi:10.1007/s10928-020-09726-9 · Cited by 7
On the centenary of the insulin discovery
Dedov, I. I., & Shestakova, M. V. (2021) · Diabetes Mellitus · doi:10.14341/dm12733 · Cited by 7
Reiterer, F., Reiter, M., del Re, L., Christensen, M. B., & Nørgaard, K. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818770694 · Cited by 7
Method to generate a large cohort in-silico for type 1 diabetes
Orozco-López, O., Rodríguez-Herrero, A., Castaneda, C. E., García-Sáez, G., & Hernando, M. E. (2020) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2020.105523 · Cited by 7
Ríos, Y. Y., García-Rodríguez, J. A., Sanchez, E. N., Alanís, A. Y., Ruiz-Velazquez, E., & García, A. P. (2021) · ISA Transactions · doi:10.1016/j.isatra.2021.07.045 · Cited by 7
Automated Adaptation of Insulin Treatment in Type 1 Diabetes
Fabris, C., Gautier, T., & Breton, M. D. (2021) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc46164.2021.9630191 · Cited by 7
“Here, we present results obtained by testing the optimization algorithm in an FDAaccepted simulation environment known as the University of Virginia/Padova T1D Simulator [21]-[23].”
MethodSelf-citationA software interface for in silico testing of type 2 diabetes treatments
VISENTIN, R., Cobelli, C., & Man, C. D. (2022) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2022.106973 · Cited by 7
Self-citationVHDL implementation of the Biostator II glucose control algorithm for critical care
Pagkalos, I., Herrero, P., El-Sharkawy, M., Pesl, P., Oliver, N. S., & Georgiou, P. (2011) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2011.6107735 · Cited by 7
The mathematician's control toolbox for management of type 1 diabetes
CSETE, M. E., & Doyle, J. C. (2014) · Interface Focus · doi:10.1098/rsfs.2014.0042 · Cited by 7
“A major advance in modelling was FDA approval of a T1DM patient simulator for pre-clinical development of a clinical product [22].”
Metaheuristic Patient Estimation Based Patient-Specific Fuzzy Aggregated Artificial Pancreas Design
Kirubakaran, V., Radhakrishnan, T. K., & Sivakumaran, N. (2014) · Industrial & Engineering Chemistry Research · doi:10.1021/ie5009647 · Cited by 7
Propuesta de algoritmos de predicción de glucosa en pacientes diabéticos
Gandía, C. P. (2014) · UPM Digital Archive (Technical University of Madrid) · doi:10.20868/upm.thesis.25604 · Cited by 7
Closed-loop insulin delivery in adults with type 1 diabetes
Kumareswaran, K. (2012) · Apollo (University of Cambridge) · doi:10.17863/cam.16644 · Cited by 7
“In 2008, the USA Food and Drug Administration approved an in silico simulation environment as an alternative to animal trials for pre-clinical closed-loop studies.[125]”
Modeling Intraperitoneal Insulin Absorption in Patients with Type 1 Diabetes
Schiavon, M., Cobelli, C., & Dalla Man, C. (2021) · Metabolites · doi:10.3390/metabo11090600 · Cited by 6
Self-citationMoscoso-Vásquez, M., Fabris, C., & Breton, M. D. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231206798 · Cited by 6
Self-citationLong-term blood glucose prediction using deep learning-based noise reduction
Kim, S., Moon, J. S., & Jung, S. (2025) · ICT Express · doi:10.1016/j.icte.2025.05.009 · Cited by 7
The Impact of Exogenous Insulin Input on Calculating Hepatic Clearance Parameters
McHugh, A. D., Chase, J. G., Knopp, J. L., Ormsbee, J. J., Kulawiec, D. G., Merry, T. L., Murphy, R., Shepherd, P. R., Burden, H. J., & Docherty, P. D. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820986878 · Cited by 6
Optimal Periodic Impulsive Strategies in Glycemic Control
Borri, A., Cacace, F., Gaetano, A. D., Pompa, M., & Panunzi, S. (2024) · IEEE Transactions on Control Systems Technology · doi:10.1109/tcst.2024.3398288 · Cited by 6
“To this end, we consider the model in [33] (more recently extended in [45]), for which equations and parameter sets are freely available, and which constitutes the basis of the first version of the FDA-accepted Uva/Padova simulator [34].”
MethodEstimation and personalization of clinical insulin therapy parameters
Palma, R. C. (2013) · Texas ScholarWorks (Texas Digital Library) · Cited by 6
Pharmacokinetics of insulin lispro in type 2 diabetes during closed-loop insulin delivery
Ruan, Y., Thabit, H., Kumareswaran, K., & Hovorka, R. (2014) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2014.07.004 · Cited by 6
Nonlinear Models of Glucose Concentration
Georga, E. I., Fotiadis, D. I., & Tigas, S. K. (2017) · Elsevier eBooks · doi:10.1016/b978-0-12-804831-3.00006-6 · Cited by 6
Getting Daily Blood Glucose Tracks Using Clinical Protocols of the DirecNet Database
Rudenko, P. A., Bazaev, N. A., Витольдович, П. К., Litinskaia, E. L., Гринвальд, В. М., & Chekasin, A. I. (2018) · Biomedical Engineering · doi:10.1007/s10527-018-9745-8 · Cited by 6
Fox, I. (2020) · Deep Blue (University of Michigan) · Cited by 6
Dynamic of Glucose Homeostasis in Virtual Patients: A Comparison between Different Behaviors
Alonso‐Bastida, A., Adam‐Medina, M., Posada, R., Salazar-Piña, D. A., Osorio-Gordillo, G. L., & Valdés, L. G. V. (2022) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph19020716 · Cited by 6
ESCADA: Efficient Safety and Context Aware Dose Allocation for Precision Medicine
Demirel, İ., Celik, A. A., & Tekin, C. (2022) · neural information processing systems · doi:10.52202/068431-1990 · Cited by 6
Subcutaneous neural inverse optimal control for an Artificial Pancreas
Leon, B. S., Alanís, A. Y., Sanchez, E. N., Tellez, F. O., & Ruiz-Velazquez, E. (2013) · International Joint Conference on Neural Networks (IJCNN) · doi:10.1109/ijcnn.2013.6707090 · Cited by 6
“In this paper, the attention is drawn to identi cation for unknown nonlinear dynamic of the blood glucose level in response to meal and continuous insulin infusion for a T1DM patient.”
BackgroundUsing Modeling and Simulation in the Design of Closed-Loop Insulin Delivery System
Hoadley, D., & Ananthan, A. (2013) · Journal of Medical Devices · doi:10.1115/1.4024340 · Cited by 6
Naumova, V., Pereverzev, S. V., & Sivananthan, S. (2012) · Journal of Complexity · doi:10.1016/j.jco.2012.06.001 · Cited by 6
“To illustrate this we use data set of 100 virtual subjects which are obtained from Padova/University of Virginia simulator [12].”
MethodControl by pulses under MPC schemes, with applications to artificial pancreas
Abuin, P., Ferramosca, A., Rivadeneira, P. S., Godoy, J. L., & González, A. H. (2019) · Workshop on Information Processing and Control · doi:10.1109/rpic.2019.8882137 · Cited by 6
Asare, P., Dickerson, R. F., Wu, X., Lach, J., & Stankovic, J. A. (2013) · International Conference on Body Area Networks · doi:10.4108/icst.bodynets.2013.253694 · Cited by 6
Bequette, B. W., Cameron, F. M., Baysal, N., Howsmon, D. P., Buckingham, B. A., Maahs, D. M., & Levy, C. J. (2016) · Processes · doi:10.3390/pr4040039 · Cited by 6
“, [36]), they do not fully mimic the variability observed in humans, thus testing algorithms in closed-loop with human subjects is critical.”
BackgroundCarbMetSim: A discrete-event simulator for carbohydrate metabolism in humans
Goyal, M., Aydas, B., Ghazaleh, H., & Rajasekharan, S. (2020) · PLoS ONE · doi:10.1371/journal.pone.0209725 · Cited by 6
Hypoglycemia Prevention in Closed-Loop Artificial Pancreas for Patients with Type 1 Diabetes
Abu-Rmileh, A., & García-Gabín, W. (2011) · InTech eBooks · doi:10.5772/22647 · Cited by 6
“Some of these models have been implemented in simulation environments designed to support the development of the closed-loop artificial pancreas (Kovatchev et al. (2009); Wilinska et al. (2010)).”
BackgroundAutomated optimal glycaemic control using a physiology based pharmacokinetic, pharmacodynamic model
Schaller, S., Schuppert, A., & Mitsos, A. (2015) · RWTH Publications (RWTH Aachen) · Cited by 6
Ongoing Debate About Models for Artificial Pancreas Systems and In Silico Studies
Forlenza, G. P. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0038 · Cited by 6
“31,32 Serving as a substitute for timely and costly animal studies, the simulator allows for preclinical assessment of certain insulin treatment strategies and closed-loop algorithms.”
BackgroundFaruqui, S. H. A., Alaeddini, A., Du, Y. N., Li, S., Sharma, K., & Wang, J. (2024) · arXiv (Cornell University) · doi:10.48550/arxiv.2401.02661 · Cited by 6
“Most of these studies focus on controlling insulin infusion devices 12,13 to control blood glucose levels.”
BackgroundZarkogianni, K. (2011) · DSpace - NTUA (National Technical University of Athens) · doi:10.26240/heal.ntua.305 · Cited by 6
Metabolic Models, in Silico Trials, and Algorithms
Çınar, A., Basu, A., Bequette, B. W., Breton, M. D., Buckingham, B. A., Cengiz, E., Cobelli, C., Dassau, E., Doyle, F. J., Fabris, C., et al. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2025.18800.ac · Cited by 5
Self-citationOptimizing Blood Glucose Control through Reward Shaping in Reinforcement Learning
Rad, F. S., & Li, J. (2023) · IEEE International Conference on e-Health Networking, Applications and Services (Healthcom) · doi:10.1109/healthcom56612.2023.10472341 · Cited by 5
Wang, Q., Xie, J., Molenaar, P. C. M., & Ulbrecht, J. S. (2015) · Proceedings of the ... American Control Conference/Proceedings of the American Control Conference · doi:10.1109/acc.2015.7172245 · Cited by 5
Lunze, K., Walter, M., & Leonhardt, S. K. (2014) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2014.05.004 · Cited by 5
“The response behaviour of the nonlinear minipig model to a ubcutaneously injected insulin impulse of sc(t) = 1 IU Insulin · ı(t) (4) is used as reference with IU being the abbreviation of internaional units.”
MethodUse of intraperitoneal insulin delivery for artificial pancreas
Renard, É., Farret, A., & Place, J. (2019) · Elsevier eBooks · doi:10.1016/b978-0-12-815655-1.00014-4 · Cited by 5
An OGI model for personalized estimation of glucose and insulin concentration in plasma
Wang, W., Wang, S., Geng, Y., Qiao, Y., & Wu, T. (2021) · Mathematical Biosciences & Engineering · doi:10.3934/mbe.2021420 · Cited by 5
Determining Losses in Jet Injection Subcutaneous Insulin Delivery: A Model-Based Approach
McHugh, A. D., Chase, J. G., Knopp, J. L., Zhou, T., & Holder-Pearson, L. R. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221085032 · Cited by 5
Evaluation of Stochastic Adjustment for Glucose Sensor Bias During Closed-Loop Insulin Delivery
Kollman, C., Calhoun, P. M., Lum, J. W., Sauer, W., & Beck, R. W. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0133 · Cited by 5
“Results from the UVA-Padova T1DM simulator 14 are shown in Figure 4.”
Evaluation and Enhancement of an Intraoperative Insulin Infusion Protocol via In-Silico Simulation
Kohl, B., Chen, S., Mullen‐Fortino, M., & Lee, I. (2013) · IEEE International Conference on Healthcare Informatics (ICHI) · doi:10.1109/ichi.2013.43 · Cited by 5
“The virtual population in the software was identified based on real data collected from a group of individuals who participated in a triple-tracer meal study [52]: the meals were marked with isotope tracers so that the glucose/insulin fluxes in the body can be directly measured.”
Fravolini, M. L., & Fabietti, P. G. (2013) · Computer Methods in Biomechanics & Biomedical Engineering · doi:10.1080/10255842.2012.753064 · Cited by 5
“The above model has been recently approved by the Food and Drug Administration (FDA) as a substitute for animals trials in preclinical testing of closed loop control algorithm (Kovatchev et al. 2009); for this reason, this model is deemed particularly appropriate for our study.”
BackgroundMethodInfluentialCameron, F., Bequette, B. W., Buckingham, B. A., Wilson, D. M., Lee, H., & Niemeyer, G. (2011) · IFAC Proceedings Volumes · doi:10.3182/20110828-6-it-1002.03141 · Cited by 5
“…control Using a control algorithm as described in (Cameron 2010), a prediction algorithm as described in (Cameron & Niemeyer 2010), and the meal priors as described here we simulated using the 10 adult patients in the UVa/Padova Metabolic simulator for Type 1 Diabetes (B.P. Kovatchev et al. 2009).”
MethodArtificial pancreas: the bridge to a cure for type 1 diabetes
Thabit, H., Hovorka, R., & Evans, M. L. (2012) · European Diabetes Nursing · doi:10.1002/edn.207 · Cited by 6
Personalized glucose-insulin model based on signal analysis
Goede, S. L., de Galan, B. E., & Leow, M. K. (2016) · Journal of Theoretical Biology · doi:10.1016/j.jtbi.2016.12.018 · Cited by 5
Parvinian, B., Bighamian, R., Scully, C. G., Hahn, J., & Pathmanathan, P. (2021) · Frontiers in Physiology · doi:10.3389/fphys.2021.705222 · Cited by 5
“A precursor to clinical applications of subject-specific mathematical models is their use in pre-clinical safety evidence generation where they can be used to build a cohort of in silico subjects to complement and/or replace traditional animal studies (Kovatchev et al., 2009; Jean-Quartier et al., 2018; Mirinejad et a…”
BackgroundModel-Based Sensor-Augmented Pump Therapy
Grosman, B., Voskanyan, G. R., Loutseiko, M., Roy, A., Mehta, A., Kurtz, N., Parikh, N., KAUFMAN, F. R., Mastrototaro, J. J., & Keenan, D. B. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700224 · Cited by 5
In-Silico Trials for Glucose Control in Hospitalized Patients with Type 2 Diabetes
Choi, K., Oh, T. J., Lee, J. C., Kim, M.-J., Kim, H. C., Cho, Y. M., & Kim, S. (2016) · Journal of Korean Medical Science · doi:10.3346/jkms.2016.31.2.231 · Cited by 5
“To resolve this, in-silico simulation methods with virtual patients could provide an effective evaluation environment for glucose control prior to clinical studies, thus reducing the risks and costs of clinical evaluation (10-16).”
BackgroundMethodDistributed aspects of the artificial pancreas
Patek, S. D., Chen, S., Keith-Hynes, P., & Lee, I. (2013) · Allerton Conference on Communication, Control, and Computing · doi:10.1109/allerton.2013.6736572 · Cited by 5
Can technological solutions for diabetes replace islet cell function?
Gregory, J. M., & Moore, D. J. (2011) · Organogenesis · doi:10.4161/org.7.1.14028 · Cited by 5
Developing a mobile case-based reasoning application to assist type 1 diabetes management
Brown, D., Bayley, I., Harrison, R., & Martin, C. E. (2013) · IEEE International Conference on e-Health Networking, Applications and Services (Healthcom) · doi:10.1109/healthcom.2013.6720782 · Cited by 5
“To aid the case study for diabetes management the UVa/Padova T1DM Simulator to simulate diabetes will be used [10].”
MethodInsulin Dependent Diabetes Mellitus Control
Colmegna, P. H., & Peña, R. S. S. (2013) · CONICET Digital (CONICET) · Cited by 5
Miller, B., Vahid, F., & Givargis, T. (2012) · Design, Automation, and Test in Europe · doi:10.1109/date.2012.6176585 · Cited by 5
“Lee presents a closed-loop artificial pancreas design [8], which is validated through the use of a human diabetic subject simulator [7].”
BackgroundClosing the Loop: Another Step Forward
Weinstock, R. S. (2011) · Diabetes Care · doi:10.2337/dc11-1217 · Cited by 5
“Food and Drug Administration (FDA) of computer-based simulation studies (in silico testing) (instead of animal trials) has helped to more rapidly advance control algorithm development over the past several years (4,5).”
BackgroundThe diabetic patient software agent.
Wang, Z. (2012) · oURspace (University of Regina) · doi:10.82465/5800 · Cited by 5
Best Use of Models to Advance the Artificial Pancreas
Steil, G. M. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0420 · Cited by 5
“Nonetheless, a model developed by the UVa/Padua team 15 was accepted by the FDA as a replacement for pre-clinical animal studies— becoming commercially available in July 2011.”
Dynamic Calibration of Nonlinear Sensors with Time-Drifts and Delays by Bayesian InferencePreprint
Talukder, S., Kundu, S., & Kumar, R. (2022) · arXiv (Cornell University) · doi:10.48550/arxiv.2208.13819 · Cited by 5
“Figure 1: BGL profiles corresponding to the respective 20 virtual patients FDA-approved simulation models of type-I diabetic patients [18], is used to synthetically generate BGL values for 20 virtual patients of diverse age-groups subject to varied food intake and insulin injection profiles over a period of 19 hours (…”
MethodFera, N., Procopio, A., Zaffino, P., Cutruzzolà, A., Irace, C., & Cosentino, C. (2026) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2026.111601 · Cited by 4
Nonparametric Identification of Glucose-Insulin Process in IDDM Patient with Multi-meal Disturbance
Bhattacharjee, A., & Sutradhar, A. (2012) · Journal of The Institution of Engineers (India) Series B · doi:10.1007/s40031-013-0037-0 · Cited by 4
Dassau, E., Hennings, T., Fazio, J. C., Atlas, E., & Phillip, M. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.1504 · Cited by 4
Personalized hybrid artificial pancreas using unidirectional sliding-modes control algorithm
Orozco-López, O., Castaneda, C. E., García-Sáez, G., Hernando, M. E., & Rodríguez-Herrero, A. (2022) · Journal of Applied Biomedicine · doi:10.1016/j.bbe.2022.10.003 · Cited by 4
Robust Satisficing Gaussian Process Bandits Under Adversarial Attacks
Saday, A., Yıldırım, Y. C., & Tekin, C. (2025) · neural information processing systems · doi:10.52202/085713-1680 · Cited by 4
Subcutaneous blood glucose neural inverse optimal control for type 1 diabetes mellitus patients
Leon, B. S., Alanís, A. Y., Sanchez, E. N., Tellez, F. O., & Ruiz-Velazquez, E. (2012) · World Automation Congress (WAC) · Cited by 4
Neural inverse optimal control applied to type 1 diabetes mellitus patients
Leon, B. S., Alanís, A. Y., Sanchez, E. N., Tellez, F. O., & Ruiz-Velazquez, E. (2012) · IEEE Latin American Symposium on Circuits and Systems (LASCAS) · doi:10.1109/lascas.2012.6180310 · Cited by 4
T1DM subject as a time-delay system: Modeling and adaptive control
Ludwig, T., Ottinger, I., Tárník, M., & Miklovičová, E. (2013) · International Conference on Process Control · doi:10.1109/pc.2013.6581388 · Cited by 4
“Closing the glucose control loop with a fully automated device will dramatically improve the management of diabetes and the quality of life of the patients [1], [2].”
MethodArtificial Blood Glucose Control using a DDE Modelling Approach
Penet, M., Guéguen, H., & Belmiloudi, A. (2014) · IFAC Proceedings Volumes · doi:10.3182/20140824-6-za-1003.01233 · Cited by 4
“In order to test the interest of the proposed approach, simulation has been realized on a virtual testing platform (Uva/Padova T1DM metabolic simulator the distributed version [Kovatchev et al., 2009]).”
MethodInfluentialHuman endocrine system modeling based on ontologies
Viale, P., Bora, J. J., Benegui, M., & BASUALDO, M. S. (2016) · Knowledge-Based Systems · doi:10.1016/j.knosys.2016.08.002 · Cited by 4
Robust glucose control via μ-synthesis in type 1 diabetes mellitus
Morales-Contreras, J., Ruiz-Velazquez, E., & García-Rodríguez, J. A. (2017) · IEEE International Autumn Meeting on Power, Electronics and Computing (ROPEC) · doi:10.1109/ropec.2017.8261671 · Cited by 4
“As such, it has been accepted by the Food and Drug Administration (FDA) as a substitute for preclinical animal studies for various scenarios of the glucose control system [14].”
BackgroundMethodMei, Y., Huynh, T., Khor, R., & Rollins, D. K. (2019) · Journal of Dynamic Systems Measurement and Control · doi:10.1115/1.4043335 · Cited by 4
Franco, R., Ríos, H., de Loza, A. F., Cassany, L., Gucik‐Derigny, D., Cieslak, J., & Henry, D. (2022) · International Journal of Robust and Nonlinear Control · doi:10.1002/rnc.6030 · Cited by 4
HypoTreat: Reducing Hypoglycemia in Artificial Pancreas Simulation
Chlumsky-Harttmann, M., Ayad, A., & Schmeink, A. (2024) · IEEE Conference on Evolving and Adaptive Intelligent Systems (EAIS) · doi:10.1109/eais58494.2024.10569096 · Cited by 4
“The publicly available Simglucose is based on the UVA/Padova Simulator’s oldest version 2008 [14] and comes with a cohort of 30 in silico T1DM patients (10 adults, adolescents, and children each) with patient-specific model parameters.”
BackgroundMethodCampetelli, G., Lupo, M., Fina, B. L., Zumoffen, D., BASUALDO, M. S., & Rigalli, A. (2010) · IFAC Proceedings Volumes · doi:10.3182/20100707-3-be-2012.0097 · Cited by 4
Li, P., Yu, L., Guo, L., Dong, J., Hu, J., & Fang, Q. (2012) · World Congress on Intelligent Control and Automation · doi:10.1109/wcica.2012.6359435 · Cited by 4
“Though the development of CGM and CSII has been advanced greatly in recent years [4-6], to develop a reliable insulin-glucose metabolic model for APS that can be widely accepted is still a not fully resolved research question.”
BackgroundAdjustment of basal insulin infusion rate in T1DM by hybrid PSO
Lou, Z., Liu, B., Xie, H., & Wang, Y. (2014) · Soft Computing · doi:10.1007/s00500-014-1378-6 · Cited by 4
“Nowadays the metabolism model for DM has been accepted by Food and Drug Administration as a substitute to animal trials in the preclinical testing of control strategies (Kovatchev et al. 2009).”
BackgroundMethodRios, Y., García-Rodríguez, J. A., Sanchez, E. N., & Acevedo, C. M. D. (2018) · REVISTA COLOMBIANA DE TECNOLOGIAS DE AVANZADA · doi:10.24054/16927257.v32.n32.2018.3023 · Cited by 4
Gessa, F., Asare, P., Bray, A., Clipp, R. B., & Poler, S. M. (2019) · ACM SIGBED Review · doi:10.1145/3357495.3357499 · Cited by 4
“Two of the well-known frameworks for evaluation of closed-loop MCPS are the simulation platform for the artificial pancreas [17] and the that for pacemakers [24].”
MethodInfluentialNon-linear Continuous Action Spaces for Reinforcement Learning in Type 1 Diabetes
Hettiarachchi, C., Malagutti, N., Nolan, C. J., Suominen, H., & Daskalaki, E. (2022) · Lecture notes in computer science · doi:10.1007/978-3-031-22695-3_39 · Cited by 4
Liu, W., Chen, T., Liang, B., Wang, Y., & Jin, H. (2023) · Frontiers in Endocrinology · doi:10.3389/fendo.2023.1115436 · Cited by 4
High-Confidence Medical Device Software Development
Jiang, Z., & Mangharam, R. (2015) · now publishers, Inc. eBooks · doi:10.1561/9781680830699 · Cited by 4
Challenges and Opportunities in Design of Control Algorithm for Artificial Pancreas
Ghorbani, M., & Bogdan, P. (2014) · DROPS (Schloss Dagstuhl – Leibniz Center for Informatics) · doi:10.4230/oasics.mcps.2014.49 · Cited by 4
A new model for closed-loop control in type 1 diabetes
Abbes, I. B., Lefebvre, M.-A., Cormerais, H., & Richard, P.-Y. (2011) · IFAC Proceedings Volumes · doi:10.3182/20110828-6-it-1002.02154 · Cited by 4
“This model has been implemented as a simulation toolbox developed by Kovatchev et al. [2009] in the framework of Matlab-Simulink.”
MethodModel based control of artificial pancreas under meal disturbances
Ata, S., & Khan, Z. H. (2017) · International Symposium on Recent Advances in Electrical Engineering (RAEE) · doi:10.1109/raee.2017.8246033 · Cited by 4
“Furthermore, the embedded controller actuates the insulin pump via Bluetooth to release the appropriate dose of insulin into the patient [17], [18].”
MethodPredictive tools for designing new insulins and treatment regimens
Klim, S. (2009) · Cited by 4
The Digital Mouse: why computational modelling of mouse models of disease can improve translationPreprint
van Riel, N. A. W., Müller, R., & Dall’Ara, E. (2020) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2020.05.04.075812 · Cited by 4
Abbes, I. B. (2013) · theses.fr (ABES) · doi:10.70675/97253ab3zc724z418ez8f5bzf76529aaf961 · Cited by 4
Diabète de type 1 du modèle ... à la boucle fermée
Magdelaine, N. (2017) · theses.fr (ABES) · doi:10.70675/abaad835z8c10z4e88z8e21z1c1b6b3604e1 · Cited by 4
Dorosan, M., Chen, Y.-L., Zhuang, Q., & Lam, S. W. (2025) · JMIR Research Protocols · doi:10.2196/63875 · Cited by 3
Wang, J., Faruqui, S. H. A., Alaeddini, A., Du, Y. N., Li, S., Yang, Y., & Sharma, K. (2026) · npj Health Systems · doi:10.1038/s44401-026-00118-8 · Cited by 3
A rat–human scale-up procedure for the endocrine system
Campetelli, G., Lombarte, M., Biset, H., Rigalli, A., & BASUALDO, M. S. (2014) · Computers & Chemical Engineering · doi:10.1016/j.compchemeng.2014.09.015 · Cited by 3
An algorithm mimicking pancreas pulsatile behavior improves artificial pancreas performance
Cocha, G., Tedesco, V., D’Attellis, C. E., & Amorena, C. E. (2021) · The International Journal of Artificial Organs · doi:10.1177/03913988211027176 · Cited by 3
Forlenza, G. P., Tabatabai, I., & Lewis, D. M. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0073 · Cited by 3
Intraperitoneal insulin infusion: on the way to the artificial pancreas
Александрович, К. В., Fedorova, E. A., Philippov, Y. I., Mayorov, A. Y., & Shestakova, M. V. (2015) · Diabetes Mellitus · doi:10.14341/dm2015332-45 · Cited by 3
Metadata
- Authors
- Kovatchev, B. P., Breton, M., Dalla Man, C., & Cobelli, C.
- Affiliations
- University of Virginia (US)
- Department of Information Engineering, University of Padova, Padova, Italy
- University of Padua (IT)
- Corresponding author
- Kovatchev, Boris P.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2009-01
- Volume
- 3
- Issue
- 1
- Pages
- 44-55
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC2681269
- OpenAlex
- W2167160009
- Semantic Scholar
- 98207f5f90ddab161fa26eb0ed21911b4960fc86
- Cited by
- 760
- References
- 49
- Citations by year
- 2026: 26 · 2025: 23 · 2024: 29 · 2023: 31 · 2022: 30 · 2021: 42 · 2020: 50 · 2019: 55 · 2018: 54 · 2017: 51 · 2016: 48 · 2015: 48 · 2014: 64 · 2013: 66 · 2012: 49
- Type
- conference abstract
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Institutes of Health (5r01dk051562-23)
- National Institute of Diabetes and Digestive and Kidney Diseases (R01DK 51562, R01-DK-051562)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Pancreatic function and diabetes
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Keywords
- type 1 diabetes, feedback control, artificial pancreas, continuous glucose monitoring, subcutaneous insulin administration, insulin pump, control algorithm simulation, meal challenge, glycemic variability
- MeSH terms
- Adolescent, Adult, Algorithms, Blood Glucose, Child, Computer Simulation, Diabetes Mellitus, Type 1, Drug Evaluation, Preclinical, Glucose, Humans, Insulin Infusion Systems, Blood Glucose Self-Monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.