The Surveillance Error Grid
Open accessThe DOI matches its Crossref record; 16 of the 16 references it cites were found in a registry or an index.
Abstract
Introduction: Currently used error grids for assessing clinical accuracy of blood glucose monitors are based on out-of-date medical practices. Error grids have not been widely embraced by regulatory agencies for clearance of monitors, but this type of tool could be useful for surveillance of the performance of cleared products. Diabetes Technology Society together with representatives from the Food and Drug Administration, the American Diabetes Association, the Endocrine Society, and the Association for the Advancement of Medical Instrumentation, and representatives of academia, industry, and government, have developed a new error grid, called the surveillance error grid (SEG) as a tool to assess the degree of clinical risk from inaccurate blood glucose (BG) monitors. Methods: A total of 206 diabetes clinicians were surveyed about the clinical risk of errors of measured BG levels by a monitor. The impact of such errors on 4 patient scenarios was surveyed. Each monitor/reference data pair was scored and color-coded on a graph per its average risk rating. Using modeled data representative of the accuracy of contemporary meters, the relationships between clinical risk and monitor error were calculated for the Clarke error grid (CEG), Parkes error grid (PEG), and SEG. Results: SEG action boundaries were consistent across scenarios, regardless of whether the patient was type 1 or type 2 or using insulin or not. No significant differences were noted between responses of adult/pediatric or 4 types of clinicians. Although small specific differences in risk boundaries between US and non-US clinicians were noted, the panel felt they did not justify separate grids for these 2 types of clinicians. The data points of the SEG were classified in 15 zones according to their assigned level of risk, which allowed for comparisons with the classic CEG and PEG. Modeled glucose monitor data with realistic self-monitoring of blood glucose errors derived from meter testing experiments plotted on the SEG when compared to the data plotted on the CEG and PEG produced risk estimates that were more granular and reflective of a continuously increasing risk scale. Discussion: The SEG is a modern metric for clinical risk assessments of BG monitor errors that assigns a unique risk score to each monitor data point when compared to a reference value. The SEG allows the clinical accuracy of a BG monitor to be portrayed in many ways, including as the percentages of data points falling into custom-defined risk zones. For modeled data the SEG, compared with the CEG and PEG, allows greater precision for quantifying risk, especially when the risks are low. This tool will be useful to allow regulators and manufacturers to monitor and evaluate glucose monitor performance in their surveillance programs.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435
Self-citationSeaquist, E. R., Anderson, J., Childs, B., Cryer, P., Dagogo-Jack, S., Fish, L., Heller, S. R., Rodriguez, H., Rosenzweig, J., & Vigersky, R. (2013) · Diabetes Care · doi:10.2337/dc12-2480 · Cited by 1,389
Self-citationParkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H. (2000) · Diabetes Care · doi:10.2337/diacare.23.8.1143 · Cited by 620
Self-citationVigersky, R. A., Fonda, S. J., Chellappa, M., Walker, M. S., & Ehrhardt, N. M. (2012) · Diabetes Care · doi:10.2337/dc11-1438 · Cited by 301
Self-citationFreckmann, G., Schmid, C., Baumstark, A., Pleus, S., Link, M., & Haug, C. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600510 · Cited by 237
Johnson, S. R., Cooper, M. N., Jones, T. W., & Davis, E. A. (2013) · Diabetologia · doi:10.1007/s00125-013-3007-9 · Cited by 178
Accuracy of Perceiving Blood Glucose in IDDM
Cox, D. J., Clarke, W. L., Gonder-Frederick, L., Pohl, S., Hoover, C., Snyder, A., Zimbelman, L., Carter, W. R., Bobbitt, S., & Pennebaker, J. (1985) · Diabetes Care · doi:10.2337/diacare.8.6.529 · Cited by 169
Self-citationSelf‐reported non‐severe hypoglycaemic events in Europe
Östenson, C. G., Geelhoed‐Duijvestijn, P., Lahtela, J., Weitgasser, R., Markert Jensen, M., & Pedersen‐Bjergaard, U. (2014) · Diabetic Medicine · doi:10.1111/dme.12261 · Cited by 168
Technical Aspects of the Parkes Error Grid
Pfützner, A., Klonoff, D. C., Pardo, S., & Parkes, J. L. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700517 · Cited by 122
Self-citation
Cited by
Continuous glucose monitoring and metrics for clinical trials: an international consensus statement
Battelino, T., Alexander, C. M., Amiel, S. A., Arreaza-Rubin, G., Beck, R. W., Bergenstal, R. M., Buckingham, B. A., Carroll, J., Ceriello, A., Chow, E., et al. (2023) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(22)00319-9 · Cited by 831
Self-citationVillena Gonzales, W., Mobashsher, A., & Abbosh, A. (2019) · Sensors · doi:10.3390/s19040800 · Cited by 772
Lee, I., Probst, D., Klonoff, D., & Sode, K. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113054 · Cited by 288
Self-citationAssessing Sensor Accuracy for Non-Adjunct Use of Continuous Glucose Monitoring
Kovatchev, B. P., Patek, S. D., Ortiz, E. A., & Breton, M. D. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0272 · Cited by 234
Self-citationMeasures of Accuracy for Continuous Glucose Monitoring and Blood Glucose Monitoring Devices
Freckmann, G., Pleus, S., Grady, M., Setford, S., & Levy, B. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818812062 · Cited by 202
Shah, V. N., Laffel, L. M., Wadwa, R. P., & Garg, S. K. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0143 · Cited by 197
Galindo, R. J., Migdal, A. L., Davis, G. M., Urrutia, M. A., Albury, B., Zambrano, C., Vellanki, P., Pasquel, F. J., Fayfman, M., Peng, L., et al. (2020) · Diabetes Care · doi:10.2337/dc19-2073 · Cited by 181
Blood Glucose Prediction with Variance Estimation Using Recurrent Neural Networks
Martinsson, J., Schliep, A., Eliasson, B., & Mogren, O. (2020) · Journal of Healthcare Informatics Research · doi:10.1007/s41666-019-00059-y · Cited by 168
Investigation of the Accuracy of 18 Marketed Blood Glucose Monitors
Klonoff, D. C., Parkes, J. L., Kovatchev, B. P., Kerr, D., Bevier, W. C., Brazg, R. L., Christiansen, M., Bailey, T. S., Nichols, J. H., & Kohn, M. A. (2018) · Diabetes Care · doi:10.2337/dc17-1960 · Cited by 144
Self-citationJacobs, P. G., Herrero, P., Facchinetti, A., Vehi, J., Kovatchev, B., Breton, M. D., Cinar, A., Nikita, K. S., Doyle, F. J., Bondia, J., et al. (2024) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2023.3331297 · Cited by 113
Self-citationFreckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
“In panel (a), the four types of error grids are the Clarke error grid, introduced by Clarke et al, 172 the consensus error grid, introduced by Parkes et al, 173 the CG-EGA, introduced by Kovatchev et al, 175 and the surveillance error grid, introduced by Klonoff et al. 174 In panel (b), the thresholds for switching be…”
BackgroundSelf-citationBaker, M., Musselman, M. E., Rogers, R., & Hellman, R. (2022) · American Journal of Health-System Pharmacy · doi:10.1093/ajhp/zxab456 · Cited by 27
Performance of a new real‐time continuous glucose monitoring system: A multicenter pilot study
Zhou, J., Zhang, S., Li, L., Wang, Y., Lu, W., Sheng, C., Li, Y., Bao, Y., & Jia, W. (2018) · Journal of Diabetes Investigation · doi:10.1111/jdi.12699 · Cited by 22
Tian, T., Aaron, R. E., Kohn, M. A., & Klonoff, D. C. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231214281 · Cited by 16
“3 After the CEG and PEG were published, new short-acting analog insulins with faster absorption and shorter duration of action were developed, with the first, Lispro, being approved in 1996.”
BackgroundSelf-citationFreckmann, G., Baumstark, A., & Pleus, S. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817713220 · Cited by 13
Giovannetti, E. R., Lee, R. O., Thomas, R. L., Wolinsky, T., Talbot, A. V., Ali, R. S., Cavaiola, T. S., Kulasa, K., & Boeder, S. C. (2025) · Endocrine Practice · doi:10.1016/j.eprac.2025.05.751 · Cited by 7
Metadata
- Authors
- Klonoff, D. C., Lias, C., Vigersky, R., Clarke, W., Parkes, J. L., Sacks, D. B., Kirkman, M. S., Kovatchev, B., Panel, T. E. G., Aurand, L., Benson, C., Bernhardt, P., Blonde, L., Breton, M., Buckingham, B., Cariski, A., Colburn, S., Flacke, F., Harrison, B., Herman, C., Hinzmann, R., Johnson-Lyles, D., Karinka, S. A., Landree, L., Lebinger, T., Lee, S., Mastrototaro, J., Peyser, T., Phillips, S., Pinkos, A., Price, D., Scott, S., Serrano, K., & Simmons, D.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2014-06-13
- Volume
- 8
- Issue
- 4
- Pages
- 658-672
- ISSN
- 1932-2968 (ISSN-L)
- OpenAlex
- W2114558541
- Cited by
- 174
- References
- 16
- FWCI
- 10.77
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.