Real-Time Detection of Infusion Site Failures in a Closed-Loop Artificial Pancreas
Open accessAbstract
Background: As evidence emerges that artificial pancreas systems improve clinical outcomes for patients with type 1 diabetes, the burden of this disease will hopefully begin to be alleviated for many patients and caregivers. However, reliance on automated insulin delivery potentially means patients will be slower to act when devices stop functioning appropriately. One such scenario involves an insulin infusion site failure, where the insulin that is recorded as delivered fails to affect the patient’s glucose as expected. Alerting patients to these events in real time would potentially reduce hyperglycemia and ketosis associated with infusion site failures. Methods: An infusion site failure detection algorithm was deployed in a randomized crossover study with artificial pancreas and sensor-augmented pump arms in an outpatient setting. Each arm lasted two weeks. Nineteen participants wore infusion sets for up to 7 days. Clinicians contacted patients to confirm infusion site failures detected by the algorithm and instructed on set replacement if failure was confirmed. Results: In real time and under zone model predictive control, the infusion site failure detection algorithm achieved a sensitivity of 88.0% (n = 25) while issuing only 0.22 false positives per day, compared with a sensitivity of 73.3% (n = 15) and 0.27 false positives per day in the SAP arm (as indicated by retrospective analysis). No association between intervention strategy and duration of infusion sets was observed ( P = .58). Conclusions: As patient burden is reduced by each generation of advanced diabetes technology, fault detection algorithms will help ensure that patients are alerted when they need to manually intervene. Clinical Trial Identifier: www.clinicaltrials.gov,NCT02773875
AI snapshot
Written from the paper’s full text.
TL;DR
In a randomized crossover outpatient trial, a real-time infusion site failure detection algorithm under Zone-MPC closed-loop achieved 88.0% sensitivity with 0.22 false positives/day, compared with retrospective SAP performance of 73.3% sensitivity and 0.27 false positives/day. The algorithm detected failures earlier than patients in many cases, but additional studies are needed to characterize its impact on hyperglycemia.
Key findings
- Real-time closed-loop Zone-MPC: sensitivity 88.0% (n = 25), 0.22 false positives/day (one FP alarm every 4.55 days), mean glucose at detection 260 mg/dL, and three false negatives.
- Retrospective SAP: sensitivity 73.3% (n = 15), 0.27 false positives/day; if alerts had been acted on in real time, there would have been 33% less time spent in hyperglycemia in the four hours prior to SF detection compared with patient detection.
- No association was observed between intervention strategy and infusion set duration (P = .58), nor with the number of sets lasting the full seven days (P = .64); there were 25 and 16 patient-confirmed SFs under Zone-MPC + SF detection and SAP, respectively.
- Hyperglycemia duration (>250 mg/dL) in the four hours preceding an SF had median reductions from 92 to 63 minutes with retrospective SAP detection (P = .085) and from 63 to 30 minutes with real-time Zone-MPC detection (P = .060); excluding subject 13, reductions were 99 to 67 minutes (P = .086) and 67 to 23 minutes (P = .087).
Methods
- Six-week outpatient randomized crossover controlled trial at Barbara Davis Center and Stanford University; 20 subjects enrolled, 19 completed (10 at Stanford, 9 at Barbara Davis), with 1:1 randomization. Group A started with Zone-MPC + SF detection and Group B with sensor-augmented pump (SAP); there was one week of washout, and each block consisted of one week of sensor run-in followed by two weeks of prolonged infusion set and sensor wear up to seven days.
- Zone-MPC was implemented on the DiAs platform with Dexcom G4 PLATINUM CGM; the Roche Accu-Chek Spirit Combo pump was used in the Zone-MPC + SF detection arm, while each patient’s personal pump was used in the SAP arm. The SF detection algorithm used adaptive patient-specific glucose fault metric (GFM), insulin fault metric (IFM), and glucose slope thresholds; alarms were confirmed or refuted by patients and clinical staff.
- Analysis assessed sensitivity and false positives per day on an alarm-event basis, compared infusion set survival using Kaplan-Meier estimates and log-rank tests, compared censored sets with chi-square tests, and analyzed relationships between hyperglycemia and carbohydrate intake via F-tests for regression.
Limitations
- The paper states that additional studies are needed to fully characterize the effects of real-time SF detection and various intervention strategies on SF-associated hyperglycemia, and describes the work as a pilot study.
- The paper notes three false negatives: one was detected very early by a diligent patient, and two were attributed to prolonged SF dynamics from gradual increases in insulin feedback under closed-loop control, slower than the open-loop cases on which the algorithm was trained.
- Evident from the text: the sample was small (19 completers), each arm lasted only two weeks, the algorithm used a previously identified fixed parameter set, and false positives may alert patients to other inadequate-control scenarios such as unannounced meals because meal information was not provided to the algorithm.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
7 of 30 cited sources checked · 6 supported · 0 flagged · accuracy not yet measured
- Premisecitation§ Conclusion
Problems with infusion sets are a reason for patients discontinuing insulin pump therapy.
“and a reason for patients discontinuing insulin pump therapy. 21,37-39”
Quote in the paperSubstantial claim3 of 4 sources checked- cites [21] 21. Wong JC, Boyle C, DiMeglio LA, et al. Evaluation of pump discontinuation and associated factors in the T1D exchange clinical registry. J Diabetes Sci Technol . 2017;11(2):224-232. source
- cites [37] 37. Wood JR, Moreland EC, Volkening LK, Svoren BM, Butler DA, Laffel LM. Durability of insulin pump use in pediatric patients with type 1 diabetes. Diabetes Care . 2006;29(11):2355-2360. source
- cites [38] 38. de Vries L, Grushka Y, Lebenthal Y, Shalitin S, Phillip M. Factors associated with increased risk of insulin pump discontinuation in pediatric patients with type 1 diabetes. Pediatr Diabetes . 2011;12(5):506-512. source
- cites [39] 39. Lombardo F, Scaramuzza AE, Iafusco D. Failure of glycated hemoglobin drop after continuous subcutaneous insulin infusion initiation may indicate patients who discontinue: a 4-year follow-up study in children and adolescents with type 1 diabetes. Acta Diabetol . 2012;49(1):99-105. source
- Findingown data§ Impact of SF Detection on SF-Induced Hyperglycemia
Retrospective analysis of the SAP arm indicates a median reduction in hyperglycemia duration from 92 to 63 minutes (P = 0.085) if the SF detection algorithm had been active.
“Retrospective analysis of the SAP arm indicates a median reduction in hyperglycemia duration from 92 to 63 minutes ( P = 0.085) if the SF detection algorithm had been active.”
Quote in the paperSubstantial claim - Findingown data§ Characteristics of SFs
In 76 weeks of prolonged infusion set wear, there were 25 patient-confirmed SFs under Zone-MPC + SF detection and 16 under SAP.
“In 76 weeks of prolonged infusion set wear, there were 25 and 16 patientconfirmed SFs under Zone-MPC + SF detection and SAP conditions, respectively (Table 4, Figure 2).”
Quote in the paperNot a substantial claim - Findingown data§ Results
Nineteen patients (10 at Stanford and 9 at Barbara Davis) successfully completed the six-week study.
“Nineteen patients (10 at Stanford and 9 at Barbara Davis) successfully completed the six-week study”
Quote in the paperSubstantial claim - Findingown data§ Characteristics of SFs
No association was found between the intervention strategy and the number of infusion sets that lasted the full seven days (P = 0.64).
“No association was found between the intervention strategy and the number of infusion sets that lasted the full seven days ( P = 0.64).”
Quote in the paperNot a substantial claim
- Findingown data§ Characteristics of SFs
No association was found between infusion set duration and the intervention strategy.
“Statistical analysis of observed SFs reveals no association between either infusion set duration or the total number of SFs with the intervention strategy.”
Quote in the paperNot a substantial claim - Findingown data§ Performance of Real-Time SF Detection Under Zone-MPC
Mean glucose at detection was 260 mg/dL.
“Mean glucose at detection was 260 mg/dL.”
Quote in the paperNot a substantial claim - Findingown data§ Impact of SF Detection on SF-Induced Hyperglycemia
Removing subject 13 from the analysis produces a reduction in median hyperglycemia duration from 99 to 67 minutes (P = 0.086) when SF detection is applied retrospectively to the SAP portion.
“Removing subject 13 from the analysis produces a reduction in median hyperglycemia duration from 99 to 67 minutes ( P = 0.086) when the SF detection is applied retrospectively to the SAP portion”
Quote in the paperNot a substantial claim - Findingown data§ Performance of Real-Time SF Detection Under Zone-MPC
There were three false negatives during the closed-loop portion of the study.
“There were three FNs during the closed-loop portion of this study.”
Quote in the paperNot a substantial claim - Findingown data§ Abstract
No association between intervention strategy and duration of infusion sets was observed (P = .58).
“No association between intervention strategy and duration of infusion sets was observed ( P = .58).”
Quote in the paperNot a substantial claim - Findingown data§ Abstract
In the SAP arm, retrospective analysis showed the SF detection algorithm issued 0.27 false positives per day.
“and 0.27 false positives per day in the SAP arm (as indicated by retrospective analysis)”
- Also stated in: § Retrospective Analysis of SF Detection Under SAP · plain
“while only issuing 0.27 FPs/day (Table 3)”
Quote in the paperNot a substantial claim - Also stated in: § Retrospective Analysis of SF Detection Under SAP · plain
- Findingown data§ Abstract
In the SAP arm, retrospective analysis showed the SF detection algorithm achieved a sensitivity of 73.3% (n = 15).
“compared with a sensitivity of 73.3% (n = 15) and 0.27 false positives per day in the SAP arm (as indicated by retrospective analysis)”
- Also stated in: § Retrospective Analysis of SF Detection Under SAP · plain
“Of the 15 remaining failures, the SF detection algorithm correctly detected 11 SF events (sensitivity of 73.3%)”
Quote in the paperSubstantial claim - Also stated in: § Retrospective Analysis of SF Detection Under SAP · plain
- Findingown data§ Discussion
Three SFs were not detected by the SF detection algorithm in real time.
“Although three SFs were not detected by the SF detection algorithm in real time (Figure 5, discussed below)”
Quote in the paperSubstantial claim - Findingown data§ Characteristics of SFs
The most common cause of set failures was unexplained hyperglycemia that failed to respond to a correction dose.
“The most common cause of set failures was unexplained hyperglycemia that failed to respond to a correction dose.”
Quote in the paperNot a substantial claim - Findingreasoning§ Performance of Real-Time SF Detection Under Zone-MPC
The 0.22 false positives per day translates to one false positive alarm every 4.55 days.
“This translates to one FP alarm every 4.55 days.”
Quote in the paperNot a substantial claim - Findingown data§ Abstract
In real time under Zone-MPC, the infusion site failure detection algorithm issued 0.22 false positives per day on average.
“while issuing only 0.22 false positives per day”
- Also stated in: § Performance of Real-Time SF Detection Under Zone-MPC · plain
“and issued only 0.22 FPs per day, on average (Table 3)”
Quote in the paperSubstantial claim - Also stated in: § Performance of Real-Time SF Detection Under Zone-MPC · plain
- Findingown data§ Abstract
In real time under Zone-MPC, the infusion site failure detection algorithm achieved a sensitivity of 88.0% (n = 25).
“the infusion site failure detection algorithm achieved a sensitivity of 88.0% (n = 25) while issuing only 0.22 false positives per day”
Quote in the paperSubstantial claim - Findingown data§ Discussion
Of the three false negatives, one undetected SF was detected very early by a patient who noticed insulin was leaking at the insertion site.
“Of the three FNs, one undetected SF was detected very early by a patient who noticed insulin was leaking at the insertion site (Figure 5A).”
Quote in the paperNot a substantial claim - Findingown data§ Retrospective Analysis of SF Detection Under SAP
Of the 15 remaining failures in the SAP arm, the SF detection algorithm correctly detected 11 SF events.
“Of the 15 remaining failures, the SF detection algorithm correctly detected 11 SF events”
Quote in the paperSubstantial claim - Findingown data§ Characteristics of SFs
No association was found between the total number of SFs and the intervention strategy.
“Statistical analysis of observed SFs reveals no association between either infusion set duration or the total number of SFs with the intervention strategy.”
Quote in the paperNot a substantial claim - Findingown data§ Characteristics of SFs
No association between infusion set survival and intervention strategy was observed (P = 0.58).
“Analysis of the KaplanMeier estimates of infusion set survival (Figure 3) produces a P value of 0.58, indicating that no association between infusion set survival and intervention strategy was observed.”
Quote in the paperNot a substantial claim - Findingown data§ Impact of SF Detection on SF-Induced Hyperglycemia
Real-time SF detection in the Zone-MPC arm further reduced hyperglycemia duration from 63 to 30 minutes (P = 0.060).
“Furthermore, real-time SF detection in the Zone-MPC arm further reduced this hyperglycemia duration from 63 to 30 minutes ( P = 0.060).”
Quote in the paperSubstantial claim - Findingown data§ Characteristics of SFs
There was only 1 occlusion alarm generated by the pump over all subjects and in both arms.
“There was only 1 occlusion alarm generated by the pump over all subjects and in both arms.”
Quote in the paperNot a substantial claim - Findingown data§ Impact of SF Detection on SF-Induced Hyperglycemia
Removing subject 13 from the analysis produces a reduction in median hyperglycemia duration from 67 to 23 minutes (P = 0.087) when SF detection is applied in SAP versus ZoneMPC.
“and from 67 to 23 minutes ( P = 0.087) when SF detection is applied in SAP versus ZoneMPC”
Quote in the paperNot a substantial claim - Premisecitation§ Corresponding Author:
Previous studies in the literature have retrospectively evaluated SF detection in both open-loop and closed-loop scenarios, primarily in simulation.
“Previous studies in the literature have retrospectively evaluated SF detection in both open-loop and closed-loop scenarios, primarily in simulation.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [22] 22. Howsmon DP, Cameron F, Baysal N, et al. Continuous glucose monitoring enables the detection of Losses in Infusion Set Actuation (LISAs). Sensors . 2017;17(1):161. source
Checked before — its earlier verdict is held back until a new check runs.
- Interpretationcitation§ Discussion
The similarity in the results from the real-time analysis presented herein and the retrospective analysis conducted previously helps to mitigate concerns surrounding the retrospective determination of the start of SFs.
“The similarity in the results from the real-time analysis presented herein and the retrospective analysis conducted previously 22 helps to mitigate concerns surrounding the retrospective determination of the start of SFs.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [22] 22. Howsmon DP, Cameron F, Baysal N, et al. Continuous glucose monitoring enables the detection of Losses in Infusion Set Actuation (LISAs). Sensors . 2017;17(1):161. source
Checked before — its earlier verdict is held back until a new check runs.
- Interpretationown data§ Discussion
The other two false negatives were attributed to prolonged SF dynamics resulting from gradual increases in insulin feedback directed by the artificial pancreas.
“and the other two FNs were attributed to prolonged SF dynamics resulting from gradual increases in insulin feedback directed by the artificial pancreas.”
Quote in the paperNot a substantial claim - Interpretationown data§ Retrospective Analysis of SF Detection Under SAP
When the algorithm correctly detected an SF, if an alert had been responded to in real time, there would have been 33% less time spent in hyperglycemia in the four hours prior to SF detection, compared to when the patient detected the SF.
“Furthermore, when the algorithm correctly detected an SF, if an alert had been responded to in real time, there would have been 33% less time spent in hyperglycemia in the four hours prior to SF detection, compared to when the patient detected the SF.”
Quote in the paperSubstantial claim- The authors state this as an assumption: An alert would have been responded to in real time.
- Study design and conductown data§ Discussion
The validation strategy used to guard against overfitting the algorithm parameters served its intended purpose.
“indicating that the validation strategy used to guard against overfitting the algorithm parameters served its intended purpose.”
Quote in the paperNot a substantial claim - Premisecitation§ Keywords
Incidence of DKA is associated with higher rates of CSII pump discontinuation.
“incidence of DKA is associated with higher rates of CSII pump discontinuation.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [21] 21. Wong JC, Boyle C, DiMeglio LA, et al. Evaluation of pump discontinuation and associated factors in the T1D exchange clinical registry. J Diabetes Sci Technol . 2017;11(2):224-232. source
Checked before — its earlier verdict is held back until a new check runs.
- Interpretationcitation§ Keywords
Alerting patients to SFs causing hyperglycemia would provide them with an alert they currently do not have.
“Alerting patients to SFs causing hyperglycemia would provide them with an alert they currently do not have”
Quote in the paperSubstantial claim - Interpretationreasoning§ Corresponding Author:
This report marks the first study that evaluates the performance of real-time SF detection in advanced diabetes technology.
“This report marks the first study that evaluates the performance of real-time SF detection in advanced diabetes technology.”
Quote in the paperSubstantial claim- Rests on an assumption: There has been no study to date that investigates real-time SF detection, where patients and clinical personnel can immediately give feedback on alarm performance and missed SF events.
- Premisecitation§ Keywords
Inability to recognize SFs can lead to diabetic ketoacidosis (DKA), a high-cost acute complication that may be fatal.
“Inability to recognize SFs can lead to diabetic ketoacidosis (DKA), a high-cost acute complication that may be fatal.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [18] 18. Maahs DM, Hermann JM, Holman N, et al. Rates of diabetic ketoacidosis: international comparison with 49,859 pediatric patients with type 1 diabetes from England, Wales, the U.S., Austria, and Germany. Diabetes Care . 2015;38(10):1876-1882. source
- Interpretationreasoning§ Discussion
The SF detection algorithm may be used in conjunction with a fully closed-loop artificial pancreas that also does not require meal announcement.
“However, since it does not rely on user-provided meal announcement, the SF detection algorithm may be used in conjunction with a fully closedloop artificial pancreas that also does not require meal announcement.”
Quote in the paperSubstantial claim - Interpretationown data§ Discussion
Retrospective analysis on the Zone-MPC arm suggests that one false negative was due to a diligent patient response.
“retrospective analysis on the Zone-MPC arm suggests that one FN was due to a diligent patient response”
Quote in the paperNot a substantial claim - Interpretationown data§ Impact of SF Detection on SF-Induced Hyperglycemia
Removing subject 13 alleviates concerns that this subject skews the statistical significance of the results.
“alleviating concerns that this subject skews the statistical significance of the results.”
Quote in the paperNot a substantial claim - Interpretationown data§ Impact of SF Detection on SF-Induced Hyperglycemia
Subject 13 may have skewed the distribution due to the total number of failures recorded for this subject.
“It was observed that subject 13 may have skewed the distribution due to the total number of failures recorded for this subject.”
Quote in the paperNot a substantial claim - Interpretationown data§ Discussion
The SF detection algorithm performed well in real-time, closed-loop conditions.
“The SF detection algorithm performed well in real-time, closed-loop conditions.”
Quote in the paperSubstantial claim - Interpretationreasoning§ Keywords
When there is a persistent mismatch of expected insulin action and sustained hyperglycemia, an SF is likely.
“When there is a persistent mismatch of expected insulin action and sustained hyperglycemia, an SF is likely.”
Quote in the paperSubstantial claim - Premisecitation§ Keywords
Most SF episodes fail to trigger a pump occlusion alarm.
“since most SF episodes fail to trigger a pump occlusion alarm.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [14] 14. Gibney M, Xue Z, Swinney M, Bialonczyk D, Hirsch L. Reduced silent occlusions with a novel catheter infusion set (BD FlowSmart): results from two open-label comparative studies. Diabetes Technol Ther . 2015;18(3):136-143. source
Not checked in the free run.
- Premisecitation§ Keywords
Problems with the pump (eg, SFs) are the largest contributor to development of DKA.
“problems with the pump (eg, SFs) are the largest contributor to development of DKA”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [19] 19. Hanas R, Lindgren F, Lindblad B. A 2-yr national population study of pediatric ketoacidosis in Sweden: predisposing conditions and insulin pump use. Pediatr Diabetes . 2009;10(1):33-37. source
- cites [20] 20. Dogan ADA, Jørgensen UL, Gjessing HJ. Diabetic ketoacidosis among patients treated with continuous subcutaneous insulin infusion. J Diabetes Sci Technol . 2016;11(3):631-632.
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationnone given§ Discussion
If a patient receives an alert for hyperglycemia, they do not know if the hyperglycemia is due to insufficient insulin administration or due to an SF.
“However, if a patient receives an alert for hyperglycemia, they do not know if the hyperglycemia is due to insufficient insulin administration (eg, due to an underbolused meal) or due to an SF.”
Quote in the paperSubstantial claim- Rests on an assumption: Current alarm systems for advanced diabetes technology focus on alerting patients of current or impending hyper- and hypoglycemia.
- Premisecitation§ Conclusion
Patient surveys indicate considerable variation in the actual frequency with which patients exchange infusion sets, with mean change intervals ranging from 2.7 to 4.8 days and a range from 1 to 30 days.
“but patient surveys indicate that there is considerable variation in the actual frequency with which patients exchange infusion sets, with mean change intervals ranging from 2.7 to 4.8 days with a range from 1 to 30 days. 33”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [33] 33. Heinemann L, Fleming GA, Petrie JR, Holl RW, Bergenstal RM, Peters AL. Insulin pump risks and benefits: a clinical appraisal of pump safety standards, adverse event reporting, and research needs. Diabetes Care . 2015;38(4):716-722. source
- Interpretationown data§ Discussion
The expected dynamics of the TP and the slower, unexpected dynamics of the FN experienced by the same patient suggest that the time scale of the failure cannot be explained by differences in patient demographics.
“Furthermore, the expected dynamics of the TP and the slower, unexpected dynamics of the FN experienced by the same patient (Figure 5B) suggest that the time scale of the failure cannot be explained by differences in patient demographics.”
Quote in the paperNot a substantial claim - Premisecitation§ Conclusion
Problems with infusion sets continue to be a major contributor to diabetes-related adverse events.
“Problems with infusion sets continue to be a major contributor to diabetes-related adverse events 34-36”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [34] 34. Heinemann L, Krinelke L. Insulin infusion set: the Achilles heel of continuous subcutaneous insulin infusion. J Diabetes Sci Technol . 2012;6(4):954-964. source
- cites [35] 35. Wheeler BJ, Donaghue KC, Heels K, Ambler GR. Family perceptions of insulin pump adverse events in children and adolescents. Diabetes Technol Ther . 2013;16(4):204-207. source
- cites [36] 36. Freckmann G, Arndt S, Fiebelmann A, et al. Randomized cross-over study comparing two infusion sets for CSII in daily life. J Diabetes Sci Technol . 2017;11(2):253-259.
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationreasoning§ Discussion
Fault detection strategies can augment a safety layer for artificial pancreas applications.
“Fault detection strategies can augment a safety layer for artificial pancreas applications.”
Quote in the paperSubstantial claim - Interpretationnone given§ Conclusion
The short duration of infusion set life continues to be a hindrance for artificial pancreas applications.
“The short duration of infusion set life continues to be a hindrance for artificial pancreas applications.”
Quote in the paperSubstantial claim
- Classified as probably not a substantial claim. 28
- May combine two claims. 17
Probably not a substantial claim: may not pass S3 (load-bearing).
Interpretationcitation§ Characteristics of SFsThe observed infusion set duration is comparable to infusion set durations observed in previous studies.
“The observed infusion set duration is comparable to infusion set durations observed in previous studies. 15,29”
Quote in the paperNot a substantial claim1 of 2 sources checked- cites [15] 15. Karlin AW, Ly TT, Pyle L, et al. Duration of infusion set survival in lipohypertrophy versus nonlipohypertrophied tissue in patients with type 1 diabetes. Diabetes Technol Ther . 2016;18(7):429-435. source
- cites [29] 29. Patel PJ, Benasi K, Ferrari G, et al. Randomized trial of infusion set function: steel versus Teflon. Diabetes Technol Ther . 2013;16(1):15-19. source
May combine two claims.
Interpretationcitation§ DiscussionThe retrospective SAP sensitivity of 73.3% with 0.27 FPs/day agrees closely with previous results.
“agreeing closely with previous results 22 (Table 3)”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [22] 22. Howsmon DP, Cameron F, Baysal N, et al. Continuous glucose monitoring enables the detection of Losses in Infusion Set Actuation (LISAs). Sensors . 2017;17(1):161. source
Classified as probably not a substantial claim. May combine two claims.
Premisecitation§ KeywordsSFs may be caused by catheter kinking, insulin precipitation, insulin leakage, site inflammation, set detachment, or otherwise unknown reasons and lead to hyperglycemia and ketonemia if left uncorrected.
“SFs may be caused by catheter kinking, insulin precipitation, insulin leakage, site inflammation, set detachment, or otherwise unknown reasons and lead to hyperglycemia and ketonemia if left uncorrected.”
Quote in the paperSubstantial claim1 of 2 sources checked- cites [15] 15. Karlin AW, Ly TT, Pyle L, et al. Duration of infusion set survival in lipohypertrophy versus nonlipohypertrophied tissue in patients with type 1 diabetes. Diabetes Technol Ther . 2016;18(7):429-435. source
- cites [16] 16. Deiss D, Adolfsson P, Alkemade-van Zomeren M, et al. Insulin infusion set use: European perspectives and recommendations. Diabetes Technol Ther . 2016;18(9):517-524. source
Probably not a substantial claim: may not pass S2 (checkable).
Interpretationcitation§ DiscussionInformed alarms dictating the appropriate patient-directed intervention potentially increase safety in advanced diabetes technology by improving the interactions between patients and technology.
“Informed alarms dictating the appropriate patient-directed intervention potentially increase safety in advanced diabetes technology by improving the interactions between patients and technology. 29”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [29] 29. Patel PJ, Benasi K, Ferrari G, et al. Randomized trial of infusion set function: steel versus Teflon. Diabetes Technol Ther . 2013;16(1):15-19. source
The Introduction (pp. 15–19; passage page unspecified) describes patients checking glucose and changing infusion sets when glucose did not improve. A search for alarms and alerts found no passage addressing whether alarms that specify patient interventions improve safety or patient–technology interactions. (LLM, conf=0.73)
Experimental — accuracy not yet measured
- Rests on an assumption: The root causes of hyperglycemia (insufficient insulin administration versus SF) require very different interventions of giving an additional insulin bolus or replacing the infusion set.
May combine two claims.
Findingown data§ DiscussionSFs were detected at a higher sensitivity of 88.0% and while only issuing 0.22 FPs/day under real-time, closed-loop conditions.
“Furthermore, SFs were detected at a higher sensitivity of 88.0% and while only issuing 0.22 FPs/day under real-time, closed-loop conditions (Table 3).”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown data§ Characteristics of SFsThe median first-weekly infusion set duration coincided with the censoring time of seven days.
“Furthermore, the median first-weekly infusion set duration coincided with the censoring time of seven days (Figure 2).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown data§ ResultsThe study led to 38 patient-weeks of prolonged infusion set wear under both SAP and Zone-MPC + SF detection scenarios, for a total of 76 patient-weeks.
“leading to 38 patient-weeks of prolonged infusion set wear under both SAP and Zone-MPC + SF detection scenarios for a total of 76 patient-weeks.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingnone given§ DiscussionThe glucose remained elevated overnight and into the next day despite frequent boluses.
“However, the glucose remained elevated overnight and into the next day despite frequent boluses.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingnone given§ DiscussionThe patient experienced a slight trend upward in glucose values during the previous day, with normal glucose excursion due to manual bolusing and meals.
“The patient experienced a slight trend upward in their glucose values during the previous day, with normal glucose excursion due to manual bolusing and meals.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingreasoning§ DiscussionThe infusion set was finally deemed unresponsive to correction boluses and was replaced.
“The infusion set was finally deemed unresponsive to correction boluses and was replaced.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingreasoning§ DiscussionAn FP was issued after the glucose level reached 300 mg/dL, but the set was not deemed to have failed at this point because a correction bolus had its intended effect.
“An FP was issued after the glucose level reached 300 mg/dL, but the set was not deemed to have failed at this point because a correction bolus had its intended effect.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingreasoning§ DiscussionThe SF detection algorithm issued an FP, indicating that it had detected the unusual trends in glucose and insulin; however, its failure to continue issuing alarms resulted in the FN.
“The SF detection algorithm issued an FP, indicating that it had detected the unusual trends in glucose and insulin; however, its failure to continue issuing alarms resulted in the FN.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown data§ DiscussionThe other two undetected SFs presented in a similar way.
“The other two undetected SFs presented in a similar way (Figure 5B), thus only one will be discussed in detail.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Interpretationreasoning§ DiscussionSince the SF detection algorithm requires glucose to be elevated relative to each patient's specific baseline, FPs may alert patients to other scenarios of inadequate control, such as unannounced meals.
“Since the SF detection algorithm requires the glucose to be elevated relative to each patient's specific baseline, FPs may alert patients to other scenarios of inadequate control, such as unannounced meals.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductreasoning§ Retrospective Analysis of SF Detection Under Open-LoopSince specificity cannot be determined on a per event basis, the FP rate is reported as the average number of FPs per day.
“Since specificity cannot be determined on a per event basis, the FP rate is reported as the average number of FPs per day.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationreasoning§ KeywordsEarly SF detection promises to decrease both exposure to hyperglycemia and incidence of DKA.
“Early SF detection promises to decrease both exposure to hyperglycemia and incidence of DKA.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim. May combine two claims.
Study design and conductnone given§ SF DetectionSliding window lengths and thresholds were selected to simultaneously maximize the sensitivity and reduce the number of false positives per day.
“Sliding window lengths and thresholds were selected to simultaneously maximize the sensitivity and reduce the number of false positives per day.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Premisenone given§ KeywordsAlthough SFs are often classified into different failure modes for diagnostic purposes, they result in the same action from the patient, that is, inserting a new infusion set in a different location.
“Although SFs are often classified into different failure modes for diagnostic purposes, they result in the same action from the patient, that is, inserting a new infusion set in a different location.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductreasoning§ Statistical AnalysisThe major difference between the real-time and retrospective analysis is that SF start times do not need to be calculated as the infusion set status is determined by the patient and clinical staff for each alarm.
“The major difference between the real-time and retrospective analysis is that SF start times do not need to be calculated as the infusion set status is determined by the patient and clinical staff for each alarm.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Premisecitation§ KeywordsAdvanced device technology, including continuous glucose monitors (CGMs) and continuous subcutaneous insulin infusion (CSII) pumps, improves care for patients living with type 1 diabetes and has recently moved to closed-loop artificial pancreas (AP) systems.
“Advanced device technology, including continuous glucose monitors (CGMs) and continuous subcutaneous insulin infusion (CSII) pumps, improves care for patients living with type 1 diabetes and has recently moved to closed-loop artificial pancreas (AP) systems.”
Quote in the paperNot a substantial claim0 of 4 sources checked- cites [1] 1. Bequette BW. Challenges and recent progress in the development of a closed-loop artificial pancreas. Annu Rev Control . 2012;36(2):255-266.
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [2] 2. Peyser T, Dassau E, Breton M, Skyler JS. The artificial pancreas: current status and future prospects in the management of diabetes. Ann N Y Acad Sci . 2014;1311(1):102-123. source
- cites [3] 3. Thabit H, Hovorka R. Coming of age: the artificial pancreas for type 1 diabetes. Diabetologia . 2016;59(9):1795-1705. source
Queued — the background check reaches it, usually within a day.
- cites [4] 4. Garg SK, Weinzimer SA, Tamborlane WV, et al. Glucose outcomes with the in-home use of a hybrid closed-loop insulin delivery system in adolescents and adults with type 1 diabetes. Diabetes Technol Ther . 2017;19(3):155-163. source
Queued — the background check reaches it, usually within a day.
May combine two claims.
Interpretationown data§ DiscussionThe SF detection algorithm is applicable to both open-loop and closed-loop scenarios, with arguably better performance in closed-loop.
“Overall, these results suggest that the SF detection algorithm is applicable to both open-loop and closed-loop scenarios, with arguably better performance in closed-loop.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S2 (checkable). May combine two claims.
Interpretationreasoning§ KeywordsDetecting SFs would increase trust in an AP, unless there was a high incidence of false alarms.
“Detecting SFs would increase trust in an AP, unless there was a high incidence of false alarms.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Premisenone given§ KeywordsOnly manual intervention from the patient in replacing the infusion set can restore satisfactory glycemic control and prevent eventual development of diabetic ketoacidosis.
“Only manual intervention from the patient in replacing the infusion set can restore satisfactory glycemic control and prevent eventual development of diabetic ketoacidosis.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
Study design and conductnone given§ Study DesignEach SF event was confirmed or refuted (and thus classified as true positive or false positive) following immediate communication with the patient, and patients were instructed to change infusion sets and infusion set locations for confirmed SFs.
“Each SF event was confirmed or refuted (and thus classified as true positive or false positive events) following immediate communication with the patient and patients were instructed to change infusion sets and infusion set locations for confirmed SFs.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone given§ Study DesignRandomization occurred in a 1:1 ratio at each clinical site.
“Randomization occurred in a 1:1 ratio at each clinical site.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown data§ Retrospective Analysis of SF Detection Under SAPOne SF was removed from analysis due to a CGM dropout that started three hours before and continued until two hours after the documented SF.
“One SF was removed from analysis due to a CGM dropout that started three hours before and continued until two hours after the documented SF.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing).
Interpretationnone given§ DiscussionReal-time analysis can better assess the performance of these algorithms than simulated scenarios with assumed SF dynamics.
“Moreover, real-time analysis can better assess the performance of these algorithms than simulated scenarios with assumed SF dynamics.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Interpretationown data§ DiscussionThis SF occurred on the eleventh day of the closed-loop portion of the trial and the patient had already experienced five SFs in this study arm, potentially increasing this patient's attention to their infusion set.
“This SF occurred on the eleventh day of the closed-loop portion of the trial and the patient had already experienced five SFs in this study arm, potentially increasing this patient's attention to their infusion set.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone given§ Study DesignThe study was approved by the US Food and Drug Administration and the institutional review boards at Stanford University and the University of Colorado Denver.
“The study (www.clinicaltrials.gov,NCT02773875) was approved by the US Food and Drug Administration and the institutional review boards at Stanford University and the University of Colorado Denver.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown data§ ResultsOne participant was excluded midstudy due to failure to comply with the study protocol.
“One participant was excluded midstudy due to failure to comply with the study protocol.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ KeywordsThe incidence of DKA is not increased by CSII pump use over multiple daily injections.
“Although the incidence of DKA is not increased by CSII pump use over multiple daily injections”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [17]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [18] 18. Maahs DM, Hermann JM, Holman N, et al. Rates of diabetic ketoacidosis: international comparison with 49,859 pediatric patients with type 1 diabetes from England, Wales, the U.S., Austria, and Germany. Diabetes Care . 2015;38(10):1876-1882. source
Classified as probably not a substantial claim.
Premisecitation§ ConclusionIt is currently recommended that insulin infusion sets are changed every two to three days.
“It is currently recommended that insulin infusion sets are changed every two to three days, 16,28”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [16] 16. Deiss D, Adolfsson P, Alkemade-van Zomeren M, et al. Insulin infusion set use: European perspectives and recommendations. Diabetes Technol Ther . 2016;18(9):517-524. source
Not checked in the free run.
- cites [28] 28. Pfützner A, Sachsenheimer D, Grenningloh M, et al. Using insulin infusion sets in CSII for longer than the recommended usage time leads to a high risk for adverse events: results from a prospective randomized crossover study. J Diabetes Sci Technol . 2015;9(6):1292-1298. source
Not checked in the free run.
Probably not a substantial claim: may not pass S2 (checkable).
Interpretationreasoning§ ConclusionMore reliable infusion sets and SF detection algorithms may potentially reduce the diabetes technology discontinuation rate by reducing patient burnout with changing infusion sets and relieving concerns over complications related to SFs.
“More reliable infusion sets and SF detection algorithms may potentially reduce the diabetes technology discontinuation rate by reducing patient burnout with changing infusion sets and relieving concerns over complications related to SFs.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationreasoning§ DiscussionThe failure to continuously issue an alarm is attributed to the increased insulin delivery due to the feedback control in the artificial pancreas, enabling the fault to present with slower dynamics than in the open-loop cases on which the algorithm was trained.
“The failure to continuously issue an alarm is attributed to the increased insulin delivery due to the feedback control in the artificial pancreas, enabling the fault to present with slower dynamics than in the open-loop cases on which the algorithm was trained.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S2 (checkable), S4 (based).
Interpretationreasoning§ ConclusionAlarm systems that alert patients to specific manual interventions can potentially increase market penetration of artificial pancreas systems by enhancing human/device interactions and ameliorating concerns over troubleshooting these devices.
“Furthermore, alarm systems that alert patients to specific manual interventions, such as providing a correction dose, changing an infusion set, or taking a specific amount of carbohydrates for treating hypoglycemia, can potentially increase market penetration of artificial pancreas systems by enhancing human/device interactions and ameliorating concerns over troubleshooting these devices.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationreasoning§ DiscussionEach fault metric in the SF detection algorithm was either increasing or had already exceeded its threshold, indicating that the SF detection algorithm may have detected this event, albeit at a later time than the patient.
“It should be noted that each fault metric in the SF detection algorithm was either increasing or had already exceeded its threshold, indicating that the SF detection algorithm may have detected this event, albeit at a later time than the patient.”
Quote in the paperNot a substantial claim
The paper relies on this without support.
There has been no study to date that investigates real-time SF detection, where patients and clinical personnel can immediately give feedback on alarm performance and missed SF events.
“However, there has been no study to date that investigates real-time SF detection, where patients and clinical personnel can immediately give feedback on alarm performance (ie, true/false positives) and missed SF events.”
The paper relies on this without support.
SFs are the most common and most problematic issue with infusion pump therapy and will continue to occur with artificial pancreas systems.
“SFs are the most common and most problematic issue with infusion pump therapy and will continue to occur with artificial pancreas systems.”
The authors state this as an assumption.
An alert would have been responded to in real time.
“if an alert had been responded to in real time”
The paper relies on this without support.
Current alarm systems for advanced diabetes technology focus on alerting patients of current or impending hyper- and hypoglycemia.
“Current alarm systems for advanced diabetes technology focus on alerting patients of current or impending hyper- and hypoglycemia.”
The paper relies on this without support.
The root causes of hyperglycemia (insufficient insulin administration versus SF) require very different interventions of giving an additional insulin bolus or replacing the infusion set.
“These root causes of hyperglycemia require very different interventions of either giving an additional insulin bolus or replacing the infusion set, respectively.”
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Zone-MPC failure total differs between text and Table 4
MajorThe text reports 25 patient-confirmed failures under Zone-MPC, but Table 4 totals 24. The table's failure-mode counts also sum to 24.
“there were 25 and 16 patientconfirmed SFs under Zone-MPC + SF detection and SAP conditions, respectively (Table 4, Figure 2).”
“| Total SFs | 16 | 24 |”
V2 sensitivity is incompatible with its failure count
MinorTable 3 lists 10 SFs for V2. With 10 failures, an integer number detected cannot yield the reported sensitivity of 71.4%; seven detected would yield 70.0%.
“| Number of SFs (n) | 23 | 15 | 10 | 15 | 25 |”
“| Sensitivity (%) | 71.8 | 73.3 | 71.4 | 73.3 | 88.0 |”
Cites
Harris, P. A., Taylor, R., Thielke, R., Payne, J., Gonzalez, N., & Conde, J. G. (2009) · Journal of Biomedical Informatics · doi:10.1016/j.jbi.2008.08.010 · Cited by 51,384 · see citing works
Ultrasonic locating devices for central venous cannulation: meta-analysis
Hind, D., Calvert, N., McWilliams, R., Davidson, A., Paisley, S., Beverley, C., & Thomas, S. (2003) · BMJ · doi:10.1136/bmj.327.7411.361 · Cited by 1,083 · see citing works
The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System
Bailey, T., Bode, B. W., Christiansen, M. P., Klaff, L. J., & Alva, S. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0378 · Cited by 685 · see citing works
Garg, S. K., Weinzimer, S. A., Tamborlane, W. V., Buckingham, B. A., Bode, B. W., Bailey, T. S., Brazg, R. L., Ilany, J., Slover, R. H., Anderson, S. M., et al. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0421 · Cited by 585 · see citing works
Self-citationWEINZIMER, S. A., Steil, G. M., Swan, K. L., Dziura, J., Kurtz, N., & Tamborlane, W. V. (2008) · Diabetes Care · doi:10.2337/dc07-1967 · Cited by 567 · see citing works
Maahs, D. M., Hermann, J. M., Holman, N., Foster, N. C., Kapellen, T. M., Allgrove, J., Schatz, D. A., Hofer, S. E., Campbell, F., Steigleder-Schweiger, C., et al. (2015) · Diabetes Care · doi:10.2337/dc15-0780 · Cited by 224 · see citing works
Self-citationComing of age: the artificial pancreas for type 1 diabetes
Thabit, H., & Hovorka, R. (2016) · Diabetologia · doi:10.1007/s00125-016-4022-4 · Cited by 205 · see citing works
Zone Model Predictive Control: A Strategy to Minimize Hyper- and Hypoglycemic Events
Grosman, B., Dassau, E., Zisser, H., Jovanovic, L. G., & Doyle, F. J. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400428 · Cited by 204 · see citing works
Self-citationChallenges and recent progress in the development of a closed-loop artificial pancreas
Bequette, B. W. (2012) · Annual Reviews in Control · doi:10.1016/j.arcontrol.2012.09.007 · Cited by 182 · see citing works
Self-citationFeasibility of Outpatient Fully Integrated Closed-Loop Control
Kovatchev, B. P., Renard, É., Cobelli, C., Zisser, H., Keith-Hynes, P., Anderson, S. M., Brown, S. A., Cherñavvsky, D. R., Breton, M. D., Farret, A., et al. (2013) · Diabetes Care · doi:10.2337/dc12-1965 · Cited by 176 · see citing works
Self-citationBarnard, K. D., Wysocki, T., Allen, J. M., Elleri, D., Thabit, H., Leelarathna, L., Gulati, A., Nodale, M., Dunger, D. B., Tinati, T., et al. (2014) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2014-000025 · Cited by 159 · see citing works
Gondhalekar, R., Dassau, E., & Doyle, F. J. (2016) · Automatica · doi:10.1016/j.automatica.2016.04.015 · Cited by 159 · see citing works
Self-citationInsulin Infusion Set: The Achilles Heel of Continuous Subcutaneous Insulin Infusion
Heinemann, L., & Krinelke, L. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600429 · Cited by 148 · see citing works
The artificial pancreas: current status and future prospects in the management of diabetes
Peyser, T., Dassau, E., Breton, M., & Skyler, J. S. (2014) · Annals of the New York Academy of Sciences · doi:10.1111/nyas.12431 · Cited by 144 · see citing works
Self-citationHeinemann, L., Fleming, G. A., Petrie, J. R., Holl, R. W., Bergenstal, R. M., & Peters, A. L. (2015) · Diabetes Care · doi:10.2337/dc15-0168 · Cited by 143 · see citing works
Stevenson, J. E., Nilsson, G. C., Petersson, G., & Johansson, P. E. (2010) · Health Informatics Journal · doi:10.1177/1460458209345901 · Cited by 133 · see citing works
Closed-Loop Control Without Meal Announcement in Type 1 Diabetes
Cameron, F. M., Ly, T. T., Buckingham, B. A., Maahs, D. M., Forlenza, G. P., Levy, C. J., Lam, D. W.-H., Clinton, P. K., Messer, L. H., Westfall, E., et al. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0078 · Cited by 105 · see citing works
Self-citationDurability of Insulin Pump Use in Pediatric Patients With Type 1 Diabetes
Wood, J. R., Moreland, E. C., Volkening, L. K., Svoren, B. M., Butler, D. A., & Laffel, L. M. B. (2006) · Diabetes Care · doi:10.2337/dc06-1141 · Cited by 98 · see citing works
Hanas, R., Lindgren, F., & Lindblad, B. (2009) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2008.00441.x · Cited by 93 · see citing works
Randomized Trial of Infusion Set Function: Steel Versus Teflon
Patel, P. J., Benasi, K., Ferrari, G., Evans, M. G., Shanmugham, S., Wilson, D. M., & Buckingham, B. A. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0119 · Cited by 90 · see citing works
Self-citationA Closed-Loop Artificial Pancreas Based on Risk Management
Cameron, F., Bequette, B. W., Wilson, D. M., Buckingham, B. A., Lee, H., & Niemeyer, G. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500226 · Cited by 85 · see citing works
Self-citationControl to Range for Diabetes: Functionality and Modular Architecture
Kovatchev, B. P., Patek, S. D., Dassau, E., Doyle, F. J., Magni, L., Nicolao, G. D., & Cobelli, C. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300509 · Cited by 83 · see citing works
Self-citationvan Bon, A. C., Luijf, Y. M., Koebrugge, R., Koops, R., Hoekstra, J. B. L., & DeVries, J. H. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0166 · Cited by 76 · see citing works
Morbidity and Mortality of Diabetic Ketoacidosis With and Without Insulin Pump Care
Realsen, J., Goettle, H., & Chase, H. P. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0161 · Cited by 74 · see citing works
Huyett, L. M., Ly, T. T., Forlenza, G. P., Reuschel-DiVirgilio, S., Messer, L. H., Wadwa, R. P., Gondhalekar, R., Doyle, F. J., Pinsker, J. E., Maahs, D. M., et al. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0399 · Cited by 64 · see citing works
Self-citationEvaluation of Pump Discontinuation and Associated Factors in the T1D Exchange Clinic Registry
Wong, J. C., Boyle, C., DiMeglio, L. A., Mastrandrea, L. D., Abel, K.-L., Cengiz, E., Cemeroglu, P. A., Aleppo, G., Largay, J. F., Foster, N. C., et al. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816663963 · Cited by 63 · see citing works
Insulin Infusion Set Use: European Perspectives and Recommendations
Deiss, D., Adolfsson, P., Alkemade-van Zomeren, M., Bolli, G. B., Charpentier, G., Cobelli, C., Danne, T., Girelli, A., Mueller, H., Verderese, C. A., et al. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.07281.sf · Cited by 62 · see citing works
Harvey, R. A., Dassau, E., Bevier, W. C., Seborg, D. E., Jovanovic, L. G., Doyle, F. J., & Zisser, H. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0231 · Cited by 61 · see citing works
Self-citationCameron, F., Niemeyer, G., Wilson, D. M., Bequette, B. W., Benassi, K., Clinton, P. K., & Buckingham, B. A. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0093 · Cited by 58 · see citing works
Self-citationForlenza, G. P., Deshpande, S., Ly, T. T., Howsmon, D. P., Cameron, F. M., Baysal, N., Mauritzen, E., Marcal, T., Towers, L., Bequette, B. W., et al. (2017) · Diabetes Care · doi:10.2337/dc17-0500 · Cited by 58 · see citing works
Self-citationPfützner, A., Sachsenheimer, D., Grenningloh, M., Heschel, M., Walther-Johannesen, L., Gharabli, R., & Klonoff, D. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815604438 · Cited by 53 · see citing works
de Vries, L., Grushka, Y., Lebenthal, Y., Shalitin, S., & Phillip, M. (2011) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2010.00701.x · Cited by 50 · see citing works
Gibney, M., Xue, Z., Swinney, M., Bialonczyk, D., & Hirsch, L. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0342 · Cited by 32 · see citing works
Karlin, A. W., Ly, T. T., Pyle, L., Forlenza, G. P., Messer, L., Wadwa, R. P., DeSalvo, D. J., Payne, S. L., Hanes, S., Clinton, P., et al. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0432 · Cited by 32 · see citing works
Self-citationFamily Perceptions of Insulin Pump Adverse Events in Children and Adolescents
Wheeler, B. J., Donaghue, K. C., Heels, K., & Ambler, G. R. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0315 · Cited by 27 · see citing works
Continuous Glucose Monitoring Enables the Detection of Losses in Infusion Set Actuation (LISAs)
Howsmon, D., Cameron, F., Baysal, N., Ly, T., Forlenza, G., Maahs, D., Buckingham, B., Hahn, J., & Bequette, B. (2017) · Sensors · doi:10.3390/s17010161 · Cited by 26 · see citing works
Self-citationDiabetic Ketoacidosis Among Patients Treated With Continuous Subcutaneous Insulin Infusion
Dogan, A. D. A., Jørgensen, U. L., & Gjessing, H. J. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816668375 · Cited by 15 · see citing works
Randomized Cross-Over Study Comparing Two Infusion Sets for CSII in Daily Life
Freckmann, G., Arndt, S., Fießelmann, A., Klausmann, G., Pralle, K., Künsting, T., & Petersen, B. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816667510 · Cited by 14 · see citing works
Lombardo, F., Scaramuzza, A. E., & Iafusco, D. (2012) · Acta Diabetologica · doi:10.1007/s00592-011-0344-3 · Cited by 13 · see citing works
Forlenza, G. P., Deshpande, S., Ly, T. T., et al. (2017) · Diabetes Care · doi:10.2337/dc17-er11a · Cited by 1 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Jacobs, P. G., Herrero, P., Facchinetti, A., Vehi, J., Kovatchev, B., Breton, M. D., Cinar, A., Nikita, K. S., Doyle, F. J., Bondia, J., et al. (2024) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2023.3331297 · Cited by 113 · see citing works
Self-citationRealizing a Closed-Loop (Artificial Pancreas) System for the Treatment of Type 1 Diabetes
Lal, R., Ekhlaspour, L., Hood, K. K., & Buckingham, B. A. (2019) · Endocrine Reviews · doi:10.1210/er.2018-00174 · Cited by 97 · see citing works
Self-citationEvaluation of Pump Discontinuation and Associated Factors in the T1D Exchange Clinic Registry
Wong, J. C., Boyle, C., DiMeglio, L. A., Mastrandrea, L. D., Abel, K.-L., Cengiz, E., Cemeroglu, P. A., Aleppo, G., Largay, J. F., Foster, N. C., et al. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816663963 · Cited by 63 · see citing works
Deshpande, S., Pinsker, J. E., Zavitsanou, S., Shi, D., Tompot, R., Church, M. M., ANDRE, C. C., Doyle, F. J., & Dassau, E. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0278 · Cited by 55 · see citing works
Self-citationPractical implementation of automated closed-loop insulin delivery: A French position statement
Tubiana‐Rufi, N., Schaepelynck, P., Franc, S., Chaillous, L., Joubert, M., Renard, É., Reznik, Y., Abettan, C., Bismuth, É., Beltrand, J., et al. (2020) · Diabetes & Metabolism · doi:10.1016/j.diabet.2020.10.004 · Cited by 42 · see citing works
Meneghetti, L., Facchinetti, A., & Favero, S. D. (2020) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2020.3004270 · Cited by 36 · see citing works
Meneghetti, L., Susto, G. A., & Favero, S. D. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819881452 · Cited by 28 · see citing works
Hughes, M. S., Douvas, J. L., Layfield-Bryan, M., Blanco, L. E., Gray, J. C., Zapotoczny, G., Espinoza, J., Wilcox, J. H., & Lal, R. A. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0005 · Cited by 16 · see citing works
Hughes, M. S., Pasquel, F. J., Davis, G. M., Lal, R., Buckingham, B. A., Boughton, C. K., Brown, S. A., & Bally, L. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0343 · Cited by 13 · see citing works
Self-citationRisk analysis for the design of a safe artificial pancreas control system
Kölle, K., Fougner, A. L., Lundteigen, M. A., Carlsen, S. M., Ellingsen, R., & Stavdahl, Ø. (2018) · Health and Technology · doi:10.1007/s12553-018-0269-8 · Cited by 13 · see citing works
Mise en place de l’insulinothérapie automatisée en boucle fermée : position d’experts français
Franc, S., Schaepelynck, P., Tubiana‐Rufi, N., Chaillous, L., Joubert, M., Renard, É., Reznik, Y., Abettan, C., Bismuth, É., Beltrand, J., et al. (2020) · Médecine des Maladies Métaboliques · doi:10.1016/s1957-2557(20)30003-1 · Cited by 12 · see citing works
Human-in-the-Loop Insulin Dosing
Bequette, B. W. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819891177 · Cited by 12 · see citing works
Self-citationMeneghetti, L., Dassau, E., Doyle, F. J., & Favero, S. D. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/1932296821997854 · Cited by 11 · see citing works
Self-citationSecuring automated insulin delivery systems: A review of security threats and protective strategies
Niu, Y., & Lam, S.-K. (2025) · Computers & Security · doi:10.1016/j.cose.2025.104733 · Cited by 10 · see citing works
Metabolic Models, in Silico Trials, and Algorithms
Çınar, A., Basu, A., Bequette, B. W., Breton, M. D., Buckingham, B. A., Cengiz, E., Cobelli, C., Dassau, E., Doyle, F. J., Fabris, C., et al. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2025.18800.ac · Cited by 5 · see citing works
Self-citationOgunnaike, B. A. (2019) · Industrial & Engineering Chemistry Research · doi:10.1021/acs.iecr.9b04228 · Cited by 5 · see citing works
Use of intraperitoneal insulin delivery for artificial pancreas
Renard, É., Farret, A., & Place, J. (2019) · Elsevier eBooks · doi:10.1016/b978-0-12-815655-1.00014-4 · Cited by 4 · see citing works
Metabolic Models, in Silico Trials, and Algorithms
Çınar, A., Basu, A., Bequette, B. W., Breton, M. D., Buckingham, B. A., Cengiz, E., Cobelli, C., Dassau, E., Doyle, F. J., Fabris, C., et al. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251338300 · Cited by 4 · see citing works
Self-citationEmpfehlungen zur Diabetes-Behandlung mit automatischen Insulin-Dosierungssystemen
Berge, T. V. D., Biester, S., Biester, T., Buchmann, A.-K., Datz, N., Grosser, U., Kapitzke, K., Klusmeier, B., Remus, K., Reschke, F., et al. (2021) · Diabetologie und Stoffwechsel · doi:10.1055/a-1652-9011 · Cited by 4 · see citing works
Early Stages of Automated Insulin Delivery
Zisser, H., Hovorka, R., Basu, A., Battelino, T., Boughton, C. K., Breton, M. D., Buckingham, B. A., Brown, S. A., Cherñavvsky, D. R., Dassau, E., et al. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2025.18802.hz · Cited by 3 · see citing works
Self-citation100 Years of insulin: A chemical engineering perspective
Bequette, B. W. (2023) · Korean Journal of Chemical Engineering · doi:10.1007/s11814-022-1308-8 · Cited by 3 · see citing works
Self-citationDevelopment of a Real-time Force-based Algorithm for Infusion Failure Detection
Blanco, L. E., Wilcox, J. H., Hughes, M. S., & Lal, R. A. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241247530 · Cited by 2 · see citing works
Ibrahim, M., Beneyto, A., Contreras, I., & Vehı́, J. (2023) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2023.10.441 · Cited by 1 · see citing works
Factors determining the prognosis of people with type 1 diabetes – current perspective
Araszkiewicz, A., Trzaska, K., Soukharev, A., Spławski, J., Wolbach, J., Marek, P., Dubel, K., Wojak, W., & Zozulińska‐Ziółkiewicz, D. (2025) · Journal of Medical Science · doi:10.20883/medical.e1122 · Cited by 1 · see citing works
Early Stages of Automated Insulin Delivery
Zisser, H., Hovorka, R., Basu, A., Battelino, T., Boughton, C. K., Breton, M. D., Buckingham, B. A., Brown, S. A., Cherñavvsky, D. R., Dassau, E., et al. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251335685 · Cited by 1 · see citing works
Self-citationAutomated Insulin Dosing for Type 1 Diabetes
Bequette, B. W. (2021) · Springer eBooks · doi:10.1007/978-3-030-44184-5_100131 · Cited by 1 · see citing works
Self-citationRenard, É. (2019) · Springer eBooks · doi:10.1007/978-3-319-98119-2_6
Automated Insulin Dosing for Type 1 Diabetes
Bequette, B. W. (2019) · Springer eBooks · doi:10.1007/978-1-4471-5102-9_100131-1
Self-citationEmpfehlungen zur Diabetes-Behandlung mit automatischen Insulin-Dosierungssystemen
Berge, T. V. D., Biester, S., Biester, T., Buchmann, A.-K., Datz, N., Grosser, U., Kapitzke, K., Klusmeier, B., Remus, K., Reschke, F., et al. (2022) · Diabetes aktuell · doi:10.1055/a-1851-2883
Implementation of Multi-Functional Pump Therapy
Brindha, R., Pongiannan, R. K., Franklin, J. A., Bharath, A., Sanjeevi, V. K. S. M., & Monachan, C. (2023) · International Conference on System, Computation, Automation and Networking (ICSCAN) · doi:10.1109/icscan58655.2023.10395851
Enhancing Insulin Delivery in Type 1 Diabetes with the Detection of Insulin Flow Interruption
Hamdi, R., & Cescon, M. (2025) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2025.12.204
Martin, C., Robert, J., Bernard, S., & Girard, A. (2024) · Journal of Applied Pharmaceutical Technologies and Systems · doi:10.68159/z846227873
- Authors
- Howsmon, D. P., Baysal, N., Buckingham, B. A., Forlenza, G. P., Ly, T. T., Maahs, D. M., Marcal, T., Towers, L., Mauritzen, E., Deshpande, S., Huyett, L. M., Pinsker, J. E., Gondhalekar, R., Doyle, F. J., Dassau, E., Hahn, J., & Bequette, B. W.
- Affiliations
- Rensselaer Polytechnic Institute (US)
- Stanford Medicine (US)
- University of Colorado Denver (US)
- University of California San Diego (US)
- Harvard University (US)
- University of California, Santa Barbara (US)
- Sansum Diabetes Research Institute (US)
- Corresponding author
- Bequette, B. Wayne
- ORCID
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2018-02-01
- Volume
- 12
- Issue
- 3
- Pages
- 599-607
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC6154252
- OpenAlex
- W2787658827
- Cited by
- 32
- References
- 39
- Citations by year
- 2025: 7 · 2024: 3 · 2023: 5 · 2022: 1 · 2021: 3 · 2020: 3 · 2019: 7 · 2018: 2 · 2016: 1
- FWCI
- 2.89
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Juvenile Diabetes Research Foundation (17-2013-471)
- National Institutes of Health (DP3DK094331, DP3DK104057, UL1 TR001085)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Engineering › Biomedical Engineering › Intravenous Infusion Technology and Safety
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Keywords
- type 1 diabetes, fault detection algorithm, real-time alerts, sensor-augmented pump, artificial pancreas, hyperglycemia prevention, ketosis prevention
- MeSH terms
- Adult, Algorithms, Diabetes Mellitus, Type 1, Equipment Failure, Female, Humans, Hypoglycemic Agents, Insulin, Insulin Infusion Systems, Male, Middle Aged, Diabetic Ketoacidosis, Cross-Over Studies, Pancreas, Artificial
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-11-06.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-11-06.
Cited sources are checked by gpt-6-sol.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review