Continuous Glucose Monitoring Enables the Detection of Losses in Infusion Set Actuation (LISAs)
Abstract
Reliable continuous glucose monitoring (CGM) enables a variety of advanced technology for the treatment of type 1 diabetes. In addition to artificial pancreas algorithms that use CGM to automate continuous subcutaneous insulin infusion (CSII), CGM can also inform fault detection algorithms that alert patients to problems in CGM or CSII. Losses in infusion set actuation (LISAs) can adversely affect clinical outcomes, resulting in hyperglycemia due to impaired insulin delivery. Prolonged hyperglycemia may lead to diabetic ketoacidosis—a serious metabolic complication in type 1 diabetes. Therefore, an algorithm for the detection of LISAs based on CGM and CSII signals was developed to improve patient safety. The LISA detection algorithm is trained retrospectively on data from 62 infusion set insertions from 20 patients. The algorithm collects glucose and insulin data, and computes relevant fault metrics over two different sliding windows; an alarm sounds when these fault metrics are exceeded. With the chosen algorithm parameters, the LISA detection strategy achieved a sensitivity of 71.8% and issued 0.28 false positives per day on the training data. Validation on two independent data sets confirmed that similar performance is seen on data that was not used for training. The developed algorithm is able to effectively alert patients to possible infusion set failures in open-loop scenarios, with limited evidence of its extension to closed-loop scenarios.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The retrospective SAP sensitivity of 73.3% with 0.27 FPs/day agrees closely with previous results.
“agreeing closely with previous results 22 (Table 3)”
From Real-Time Detection of Infusion Site Failures in a Closed-Loop Artificial Pancreas (2018) · cites this paper as [22]
Previous studies in the literature have retrospectively evaluated SF detection in both open-loop and closed-loop scenarios, primarily in simulation.
“Previous studies in the literature have retrospectively evaluated SF detection in both open-loop and closed-loop scenarios, primarily in simulation.”
From Real-Time Detection of Infusion Site Failures in a Closed-Loop Artificial Pancreas (2018) · cites this paper as [22]
The similarity in the results from the real-time analysis presented herein and the retrospective analysis conducted previously helps to mitigate concerns surrounding the retrospective determination of the start of SFs.
“The similarity in the results from the real-time analysis presented herein and the retrospective analysis conducted previously 22 helps to mitigate concerns surrounding the retrospective determination of the start of SFs.”
From Real-Time Detection of Infusion Site Failures in a Closed-Loop Artificial Pancreas (2018) · cites this paper as [22]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Randomized Trial of Infusion Set Function: Steel Versus Teflon
Patel, P. J., Benasi, K., Ferrari, G., Evans, M. G., Shanmugham, S., Wilson, D. M., & Buckingham, B. A. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0119 · Cited by 90 · see citing works
Self-citationA Hazard Analysis for a Generic Insulin Infusion Pump
Zhang, Y., Jones, P. L., & Jetley, R. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400207 · Cited by 61 · see citing works
Early Detection of Infusion Set Failure During Insulin Pump Therapy in Type 1 Diabetes
Cescon, M., DeSalvo, D. J., Ly, T. T., Maahs, D. M., Messer, L. H., Buckingham, B. A., Doyle, F. J., & Dassau, E. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816663962 · Cited by 51 · cited 1× · see citing works
“[17] also evaluated infusion set failure algorithms on clinical data as opposed to simulated faults, but tried to anticipate—as opposed to detect—failures.”
BackgroundSelf-citation
Flagged first, then by citations
Research works printed but not linked: 22
- [1]Bequette B.W. Challenges and recent progress in the development of a closed-loop artificial pancreas Annu. Rev. Control 2012 36 255 266 10.1016/j.arcontrol.2012.09.007 23175620 PMC3501007 · DOI
- [2]Peyser T. Dassau E. Breton M. Skyler J.S. The artificial pancreas: Current status and future prospects in the management of diabetes Ann. N. Y. Acad. Sci. 2014 1311 102 123 10.1111/nyas.12431 24725149 · DOI · Paper page
- [3]Guilhem I. Leguerrier A. Lecordier F. Poirier J. Maugendre D. Technical risks with subcutaneous insulin infusion Diabetes Metab. 2006 32 279 284 10.1016/S1262-3636(07)70281-1 16799407 · DOI
- [5]Baysal N. Cameron F. Buckingham B.A. Wilson D.M. Chase H.P. Maahs D.M. Bequette B.W. A Novel Method to Detect Pressure-Induced Sensor Attenuations (PISA) in an Artificial Pancreas J. Diabetes Sci. Technol. 2014 8 1091 1096 10.1177/1932296814553267 25316716 PMC4455457 · DOI
- [6]Dai X. Gao Z. From Model, Signal to Knowledge: A Data-Driven Perspective of Fault Detection and Diagnosis IEEE Trans. Ind. Inform. 2013 9 2226 2238 10.1109/TII.2013.2243743 · DOI
- [7]Kovács L. Kulcsár B. Bokor J. Benyó Z. LPV fault detection of glucose-insulin system Proceedings of the 14th Mediterranean Conference on Control and Automation Ancona, Italy 28–30 June 2006
- [8]Vega-Hernandez O. Campos-Cornejo F. Campos-Delgado D.U. Espinoza-Trejo D. Increasing security in an artificial pancreas: Diagnosis of actuator faults Proceedings of the 2009 Pan American Health Care Exchanges Mexico City, Mexico 16–20 March 2009 137 142
- [9]Vega-Hernandez O. Campos-Delgado D. Espinoza-Trejo D. Actuator fault tolerant control for an artificial pancreas Proceedings of the 6th International Conference on Electrical Engineering, Computing Science and Automatic Control Toluca, Mexico 10–13 November 2009
- [10]Eberle C. Ament C. The Unscented Kalman Filter estimates the plasma insulin from glucose measurement Biosystems 2011 103 67 72 10.1016/j.biosystems.2010.09.012 20934485 · DOI
- [11]Rojas R. Garcia-Gabin W. Bequette B.W. Mean glucose slope—Principal component analysis classification to detect insulin infusion set failure Proceedings of the 18th IFAC World Congress Milan, Italy 29 August–3 September 2011
- [12]Rojas R. Garcia-Gabin W. Bequette B. Multivariate statistical analysis to detect insulin infusion set failure Proceedings of the American Control Conference (ACC) San Francisco, CA, USA 29 June–1 July 2011
- [13]Herrero P. Calm R. Vehi J. Armengol J. Georgiou P. Oliver N. Tomazou C. Robust Fault Detection System for Insulin Pump Therapy Using Continuous Glucose Monitoring J. Diabetes Sci. Technol. 2012 6 1131 1141 10.1177/193229681200600518 23063040 PMC3570848 · DOI · Paper page
- [14]Facchinetti A. Favero S. Sparacino G. Cobelli C. An Online Failure Detection Method of the Glucose Sensor-Insulin Pump System: Improved Overnight Safety of Type-1 Diabetic Subjects IEEE Trans. Biomed. Eng. 2013 60 406 416 10.1109/TBME.2012.2227256 23193300 · DOI · Paper page
- [15]Del Favero S. Monaro M. Facchinetti A. Tagliavini A. Sparacino G. Cobelli C. Real-time detection of glucose sensor and insulin pump faults in an artificial pancreas Proceedings of the International Federation of Automatic Control World Congress Cape Town, South Africa 24–29 August 2014
- [16]Baysal N. Cameron F. Buckingham B.A. Wilson D.M. Bequette B.W. Detecting sensor and insulin infusion set anomalies in an artificial pancreas Proceedings of the American Control Conference (ACC) Washington, DC, USA 17–19 June 2013
- [18]Cameron F. Bequette B.W. Wilson D.M. Buckingham B.A. Lee H. Niemeyer G. A closed-loop artificial pancreas based on risk management J. Diabetes Sci. Technol. 2011 5 368 379 10.1177/193229681100500226 21527108 PMC3125931 · DOI
- [19]Cameron F. Niemeyer G. Wilson D.M. Bequette B.W. Benassi K.S. Clinton P. Buckingham B.A. Inpatient Trial of an Artificial Pancreas Based on Multiple Model Probabilistic Predictive Control with Repeated Large Unannounced Meals Diabetes Technol. Ther. 2014 16 728 734 10.1089/dia.2014.0093 25259939 PMC4201242 · DOI
- [20]Bequette B.W. Glucose Clamp Algorithms and Insulin Time-Action Profiles J. Diabetes Sci. Technol. 2009 3 1005 1013 10.1177/193229680900300503 20144413 PMC2769898 · DOI · Paper page
- [21]Cameron F. Niemeyer G. Bequette B.W. Extended multiple model prediction with application to blood glucose regulation J. Process Control 2012 22 1422 1432 10.1016/j.jprocont.2012.05.014 · DOI
- [22]Steil G.M. Palerm C.C. Kurtz N. Voskanyan G. Roy A. Paz S. Kandeel F.R. The Effect of Insulin Feedback on Closed Loop Glucose Control J. Clin. Endocrinol. Metab. 2011 96 1402 1408 10.1210/jc.2010-2578 21367930 PMC3085208 · DOI
- [23]Ly T.T. Maahs D.M. Messer L. DeSalvo D. Payne S. Clinton P. Block J. Buckingham B.A. Duration of Infusion Set Survival in Lipohypertrophy versus Nonlipohypertrophied Tissue in Patients with Type 1 Diabetes Diabetes Technology and Therapeutics Paris, France 2015 PMC4931738 27227290
- [26]Ly T.T. Chernavvsky D. Satin-Smith M. Desalvo D.J. Shanmugham S. Keith-Hynes P. Breton M.D. Buckingham B.A. Closed-Loop Control with DIAS vs. Sensor-Augmented Pump Therapy in Adolescents and Young Adults with Type 1 Diabetes at Camp (99-L) Ned. Tijdschr. Natuurkd. 2013 13 55 56 10.1007/s12467-015-0022-0 · DOI
Cited by
Sorted by citations
Real-Time Detection of Infusion Site Failures in a Closed-Loop Artificial Pancreas
Howsmon, D. P., Baysal, N., Buckingham, B. A., Forlenza, G. P., Ly, T. T., Maahs, D. M., Marcal, T., Towers, L., Mauritzen, E., Deshpande, S., et al. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818755173 · Cited by 32 · see citing works
Self-citation
- Authors
- Howsmon, D., Cameron, F., Baysal, N., Ly, T., Forlenza, G., Maahs, D., Buckingham, B., Hahn, J., & Bequette, B.
- Affiliations
- Department of Chemical & Biological Engineering, Rensselaer Polytechnic Institute, Troy, NY 12180, USA
- Stanford University School of Medicine, Stanford, CA 94305, USA
- Barbara Davis Center for Childhood Diabetes, University of Colorado Denver, Aurora, CO 80045, USA
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2017-01-15
- Volume
- 17
- Issue
- 1
- Pages
- 161
- ISSN
- 1424-8220
- PMC
- PMC5298734
- Cited by
- 26
- References
- 3
- Language
- en
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review