Continuous Glucose Monitoring in Critically Ill Patients With COVID-19: Results of an Emergent Pilot Study
Open accessThe DOI matches its Crossref record; 29 of the 29 references it cites were found in a registry or an index.
Abstract
Background: Amidst the coronavirus disease 2019 (COVID-19) pandemic, continuous glucose monitoring (CGM) has emerged as an alternative for inpatient point-of-care blood glucose (POC-BG) monitoring. We performed a feasibility pilot study using CGM in critically ill patients with COVID-19 in the intensive care unit (ICU). Methods: Single-center, retrospective study of glucose monitoring in critically ill patients with COVID-19 on insulin therapy using Medtronic Guardian Connect and Dexcom G6 CGM systems. Primary outcomes were feasibility and accuracy for trending POC-BG. Secondary outcomes included reliability and nurse acceptance. Sensor glucose (SG) was used for trends between POC-BG with nursing guidance to reduce POC-BG frequency from one to two hours to four hours when the SG was in the target range. Mean absolute relative difference (MARD), Clarke error grids analysis (EGA), and Bland-Altman (B&A) plots were calculated for accuracy of paired SG and POC-BG measurements. Results: CGM devices were placed on 11 patients: Medtronic ( n = 6) and Dexcom G6 ( n = 5). Both systems were feasible and reliable with good nurse acceptance. To determine accuracy, 437 paired SG and POC-BG readings were analyzed. For Medtronic, the MARD was 13.1% with 100% of readings in zones A and B on Clarke EGA. For Dexcom, MARD was 11.1% with 98% of readings in zones A and B. B&A plots had a mean bias of −17.76 mg/dL (Medtronic) and −1.94 mg/dL (Dexcom), with wide 95% limits of agreement. Conclusions: During the COVID-19 pandemic, CGM is feasible in critically ill patients and has acceptable accuracy to identify trends and guide intermittent blood glucose monitoring with insulin therapy.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“The addition of noninsulin agents, such as DPP-4 inhibitors or GLP-1 RAs, before admission or during the perioperative period, has not reduced rates of stress hyperglycemia and may increase nausea and vomiting rates1316–1324 among critically ill persons.”
From American Association of Clinical Endocrinology Clinical Practice Guideline: Developing a Diabetes Mellitus Comprehensive Care Plan-2022 Update. (2022) · cites this paper as [1324]
In the citing paper
“While the use of these agents is safe and may result in lower glucose levels and insulin doses, we do not recommend addition of a DPP-4 inhibitor or GLP-1 RA to IV insulin therapy until efficacy is demonstrated.”
From American Association of Clinical Endocrinology Clinical Practice Guideline: Developing a Diabetes Mellitus Comprehensive Care Plan-2022 Update. (2022) · cites this paper as [1324]
In the citing paper
“This approach has been taken to reduce the use of personal protective equipment and more closely monitor patients so that health care personnel do not have to go into a patient room solely to measure a glucose level (283-286).”
From 7. Diabetes Technology: Standards of Care in Diabetes—2025 (2025) · cites this paper as [284]
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Lu, R., Zhao, X., Li, J., Niu, P., Yang, B., Wu, H., Wang, W., Song, H., Huang, B., Zhu, N., et al. (2020) · The Lancet · doi:10.1016/s0140-6736(20)30251-8 · Cited by 11,907
A concordance correlation coefficient to evaluate reproducibility.
Lin, L. I.-K. (1989) · Biometrics · doi:10.2307/2532051 · Cited by 8,785
Serial Evaluation of the SOFA Score to Predict Outcome in Critically Ill Patients
Ferreira, F. L. (2001) · JAMA · doi:10.1001/jama.286.14.1754 · Cited by 2,862
Zhu, L., She, Z.-G., Cheng, X., Qin, J.-J., Zhang, X.-J., Cai, J., Lei, F., Wang, H., Xie, J., Wang, W., et al. (2020) · Cell Metabolism · doi:10.1016/j.cmet.2020.04.021 · Cited by 1,602 · cited 5×
“14(6) 1065 –1073 © 2020 Diabetes Technology Society Article reuse guidelines: sagepub.”
BackgroundInfluentialCoronavirus disease 2019 case surveillance—United States, January 22–May 30, 2020.
Stokes, E. K., Zambrano, L. D., Anderson, K. N., Marder, E. P., Raz, K. M., El Burai Felix, S., Tie, Y., & Fullerton, K. E. (2020) · MMWR Morbidity and Mortality Weekly Report · doi:10.15585/mmwr.mm6924e2 · Cited by 1,538
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435
Diabetes in COVID-19: Prevalence, pathophysiology, prognosis and practical considerations
Singh, A. K., Gupta, R., Ghosh, A., & Misra, A. (2020) · Diabetes & Metabolic Syndrome: Clinical Research & Reviews · doi:10.1016/j.dsx.2020.04.004 · Cited by 893
Is diabetes mellitus associated with mortality and severity of COVID-19? A meta-analysis
Kumar, A., Arora, A., Sharma, P., Anikhindi, S. A., Bansal, N., Singla, V., Khare, S., & Srivastava, A. (2020) · Diabetes & Metabolic Syndrome: Clinical Research & Reviews · doi:10.1016/j.dsx.2020.04.044 · Cited by 776
Bode, B., Garrett, V., Messler, J., McFarland, R., Crowe, J., Booth, R., & Klonoff, D. C. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820924469 · Cited by 766
Research works printed but not linked: 1
- 10.1186/s13054-014-0453-9 · doi:10.1186/s13054-014-0453-9 · Paper page
Cited by
Blonde, L., Umpierrez, G. E., Reddy, S. S., McGill, J. B., Berga, S. L., Bush, M., Chandrasekaran, S., DeFronzo, R. A., Einhorn, D., Galindo, R. J., et al. (2022) · Endocrine Practice · doi:10.1016/j.eprac.2022.08.002 · Cited by 507
Self-citation16. Diabetes Care in the Hospital: Standards of Medical Care in Diabetes—2022
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s016 · Cited by 440
7. Diabetes Technology: Standards of Medical Care in Diabetes—2022
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s007 · Cited by 358
7. Diabetes Technology: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s007 · Cited by 348
Management of diabetes and hyperglycaemia in the hospital
Pasquel, F. J., Lansang, M. C., Dhatariya, K., & Umpierrez, G. E. (2021) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(20)30381-8 · Cited by 329
7. Diabetes Technology: Standards of Care in Diabetes — 2023
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s007 · Cited by 301
7. Diabetes Technology: Standards of Care in Diabetes—2024
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Hilliard, M. E., Johnson, E. L., Khunti, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s007 · Cited by 281
16. Diabetes Care in the Hospital: Standards of Care in Diabetes—2023
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s016 · Cited by 221
Type 2 Diabetes and the Use of Real-Time Continuous Glucose Monitoring
Jackson, M. A., Ahmann, A., & Shah, V. N. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0007 · Cited by 106
Longo, R. R., Elias, H., Khan, M., & Seley, J. J. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211008446 · Cited by 89
A Consensus Statement for Continuous Glucose Monitoring Metrics for Inpatient Clinical Trials
Spanakis, E. K., Cook, C. B., Kulasa, K. M., Aloi, J., Bally, L., Davis, G. M., Dungan, K. M., Galindo, R. J., Mendez, C. E., Pasquel, F. J., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231191104 · Cited by 93
Faulds, E. R., Boutsicaris, A., Sumner, L., Jones, L., McNett, M., Smetana, K. S., May, C. C., Buschur, E., Exline, M. C., Ringel, M. D., et al. (2021) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgab409 · Cited by 71
“Other reports among COVID-19 ICU patients reported similar or slightly better accuracy (32, 33).”
BackgroundMethodResultInfluentialShang, T., Zhang, J. Y., Thomas, A., Arnold, M. A., Vetter, B. N., Heinemann, L., & Klonoff, D. C. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211007212 · Cited by 71
Continuous Glucose Monitoring in the Hospital
Perez‐Guzman, M. C., Shang, T., Zhang, J. Y., Jornsay, D. L., & Klonoff, D. C. (2021) · Endocrinology and Metabolism · doi:10.3803/enm.2021.201 · Cited by 77
“[29] 2020 USA ICU (n=11) Guardian Connect Dexcom G6 Medtronic MiniMed Dexcom Accuracy and feasibility Capillary, venous, and arterial by POC”
BackgroundMethodResultRecent Updates to Clinical Practice Guidelines for Diabetes Mellitus
Yu, J., Lee, S.-H., & Kim, M. K. (2022) · Endocrinology and Metabolism · doi:10.3803/enm.2022.105 · Cited by 65
“Because of the coronavirus disease 2019 pandemic, CGM devices have been widely used to minimize contact between health care providers and patients [46-52].”
BackgroundShaw, J. L. V., Bannuru, R. R., Beach, L. A., ElSayed, N. A., Freckmann, G., Füzéry, A. K., Fung, A. W. S., Gilbert, J. D., Yun, H., Korpi-Steiner, N. L., et al. (2024) · Diabetes Care · doi:10.2337/dci24-0073 · Cited by 78
Diagnostic Modalities in Critical Care: Point-of-Care Approach
Rajšić, S., Breitkopf, R., Bachler, M., & Treml, B. (2021) · Diagnostics · doi:10.3390/diagnostics11122202 · Cited by 66
Continuous glucose monitoring in the hospital: an update in the era of COVID-19
Gothong, C., Singh, L. G., Satyarengga, M., & Spanakis, E. K. (2021) · Current Opinion in Endocrinology Diabetes and Obesity · doi:10.1097/med.0000000000000693 · Cited by 63
“There were only two studies [35,40] that utilized the Guardian Connect (Enlite) CGM with only one study reporting the overall MARD of 13.”
MethodAccuracy and Potential Interferences of Continuous Glucose Monitoring Sensors in the Hospital
Bellido, V., Freckman, G., Pérez, A., & Galindo, R. J. (2023) · Endocrine Practice · doi:10.1016/j.eprac.2023.06.007 · Cited by 57
Finn, E., Schlichting, L., Grau, L., Douglas, I. S., & Pereira, R. I. (2023) · Diabetes Care · doi:10.2337/dc23-0089 · Cited by 45
Review—Glucose Monitoring Sensors: History, Principle, and Challenges
Jing, Y., Chang, S. J., Chen, C.-J., & Liu, J. (2022) · Journal of The Electrochemical Society · doi:10.1149/1945-7111/ac6980 · Cited by 50
CGM in the Hospital: Is It Ready for Prime Time?
Buschur, E., Faulds, E. R., & Dungan, K. M. (2022) · Current Diabetes Reports · doi:10.1007/s11892-022-01484-x · Cited by 50
Dillmann, C., Amoura, L., Fall-Mostaine, F., Coste, A., Bounyar, L., & Kessler, L. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/1932296821994586 · Cited by 44
Hagerf, B. V., Protuš, M., Németová, L., Mráz, M., Kieslichová, E., Uchytilová, E., Indrova, V., Lelito, J., Girman, P., Haluzík, M., et al. (2024) · Diabetes Care · doi:10.2337/dc23-1663 · Cited by 49
Accuracy of continuous glucose monitoring systems in intensive care unit patients: a scoping review
Nielsen, C. G., Grigonyte‐Daraskeviciene, M., Olsen, M. T., Møller, M. H., Nørgaard, K., Perner, A., Mårtensson, J., Pedersen‐Bjergaard, U., Kristensen, P. L., & Bestle, M. H. (2024) · Intensive Care Medicine · doi:10.1007/s00134-024-07663-6 · Cited by 47
“Following the temporary FDA approval of the use of CGM systems during the COVID-19 pandemic, several clinical studies using a hybrid protocol approach were conducted [19, 44, 48, 52, 125, 126].”
InfluentialLindner, N., Kuwabara, A., & Holt, T. (2021) · Systematic Reviews · doi:10.1186/s13643-021-01644-2 · Cited by 47
Continuous Glucose Monitoring in the Intensive Care Unit
ARROYO, L. I. G., Faulds, E. R., Perez‐Guzman, M. C., Davis, G. M., Dungan, K. M., & Pasquel, F. J. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231169522 · Cited by 36
Marks, B. E., Williams, K. M., Sherwood, J. S., & Putman, M. S. (2021) · Journal of Clinical & Translational Endocrinology · doi:10.1016/j.jcte.2021.100282 · Cited by 41
“[56] Sadhu AR, Serrano IA, Xu J, Nisar T, Lucier J, Pandya AR, et al.”
BackgroundMethodInfluentialBOEDER, S. C., Kobayashi, E., Ramesh, G., Serences, B., Kulasa, K. M., & Majithia, A. R. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221113865 · Cited by 31
Baker, M., Musselman, M. E., Rogers, R., & Hellman, R. (2022) · American Journal of Health-System Pharmacy · doi:10.1093/ajhp/zxab456 · Cited by 27
Glycemic targets in critically ill adults: A mini-review
See, K. C. (2021) · World Journal of Diabetes · doi:10.4239/wjd.v12.i10.1719 · Cited by 28
Implementation of Continuous Glucose Monitoring in Critical Care: A Scoping Review
Faulds, E. R., Dungan, K. M., & McNett, M. (2023) · Current Diabetes Reports · doi:10.1007/s11892-023-01503-5 · Cited by 30
Hybrid closed-loop insulin pump technology can be safely used in the inpatient setting
Pelkey, M. N., Boyle, M. E., Long, A., Castro, J. C., Cook, C. B., & Thompson, B. (2023) · Endocrine Practice · doi:10.1016/j.eprac.2022.11.006 · Cited by 25
Nursing Perspectives on the Use of Continuous Glucose Monitoring in the Intensive Care Unit
Faulds, E. R., Dungan, K. M., McNett, M., JONES, L. G., Poindexter, N., Exline, M. C., Pattison, J., & Pasquel, F. J. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231170616 · Cited by 28
Friman, O., Soltani, N., Lind, M., Zetterqvist, P., Balintescu, A., Perner, A., Oldner, A., Rooyackers, O., & Mårtensson, J. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0035 · Cited by 25
“7,9,24,25 Notably, existing studies, including ours, have insuf fi cient power to provide robust analysis of accuracy during hypo-glycemic episodes.”
Bann, S. A., Hercus, J. C., Atkins, P., Alkhairy, A., Loyal, J. P., Sekhon, M., & Thompson, D. J. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0074 · Cited by 21
Dumitrascu, A. G., Perry, M. F., Boone, R. J., Guzman, M. P., Chirilă, R. M., McNally, A. W., Colibaseanu, D., Meek, S., Ball, C. T., White, L. J., et al. (2022) · Endocrine Practice · doi:10.1016/j.eprac.2022.12.012 · Cited by 15
“The median 155 duration of hospitalization was 11 (4-61) days.”
BackgroundContinuous glucose monitoring (CGM) in a non-Icu hospital setting: The patient's journey
Irace, C., Coluzzi, S., Cianni, G. D., Forte, E., Landi, F., Rizzo, M. R., Sesti, G., Succurro, E., & Consoli, A. (2023) · Nutrition Metabolism and Cardiovascular Diseases · doi:10.1016/j.numecd.2023.06.021 · Cited by 19
Longo, R. R., & Joshi, R. (2022) · Diabetes Spectrum · doi:10.2337/dsi22-0009 · Cited by 20
A Year of Critical Care: The Changing Face of the ICU During COVID-19
Dhala, A. F., Gotur, D. B., Hsu, S. H.-L., Uppalapati, A., Hernandez, M., Alegria, J., & Masud, F. N. (2021) · Methodist DeBakey Cardiovascular Journal · doi:10.14797/mdcvj.1041 · Cited by 18
Review Article – Diabetes Technology in the Hospital: An Update
Visser, M. M., Vangoitsenhoven, R., Gillard, P., & Mathieu, C. (2024) · Current Diabetes Reports · doi:10.1007/s11892-024-01545-3 · Cited by 17
“Several (case) reports described the use of CGM in remote diabetes care in ICU [36–38] and non-ICU patients [39–41].”
Edward, R., & Priefer, R. (2023) · Primary care diabetes · doi:10.1016/j.pcd.2023.10.009 · Cited by 20
Automated Insulin Delivery Technology in the Hospital: Update on Safety and Efficacy Data
Thompson, B., Boyle, M. E., Castro, J. C., Dodoo, C., & Cook, C. B. (2024) · Endocrine Practice · doi:10.1016/j.eprac.2024.07.012 · Cited by 14
Ang, L., Lin, Y. K., Schroeder, L. F., Huang, Y., DeGeorge, C. A., Arnold, P., Akanbi, F., Knotts, S., DuBois, E., Desbrough, N., et al. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241241005 · Cited by 11
“18,26 Our pilot’s MARD and median ARD of 13.2 and 9.8% values are consistent with those presented by Perez-Guzman et al 18 and Sadhu et al. 26 An earlier, relatively smaller study by Ancona et al investigated the accuracy of Freestyle Libre in critically ill patients, demonstrating a similar MARD…”
BackgroundResultInfluentialMårtensson, J., Cutuli, S. L., Yanase, F., Ancona, P., Toh, L., Osawa, E., & Bellomo, R. (2022) · Acta Anaesthesiologica Scandinavica · doi:10.1111/aas.14159 · Cited by 9
“9% and a proportion of values within Clarke zone A and B of 98%.(10,14) Our study showing an overall MARD of 11.”
BackgroundMethodStrongin, L. G., Анатольевна, Н. Т., Викторовна, Б. Д., Георгиевна, К. К., & Петров, А. В. (2022) · Problems of Endocrinology · doi:10.14341/probl12840 · Cited by 9
Izquierdo, V., Afonso-Bouza, N., Montoto-Méndez, E. M., Silva, G. G., Pazos-Couselo, M., & Rodríguez‐Núñez, A. (2025) · European Journal of Pediatrics · doi:10.1007/s00431-025-06368-2 · Cited by 12
“Evidence in the adult ICU population using CGM has demonstrated positive CGM accuracy [17, 27, 28].”
Real-time Continuous Glucose Monitoring in the Intensive Care Unit: The Fifth Vital Sign
Patham, B., Kansara, A., & Sadhu, A. (2024) · Diabetes Care · doi:10.2337/dci24-0020 · Cited by 9
Self-citationDexcom G7 Accuracy and Reproducibility in the Intensive Care Unit
Baran, J. D., Spanbauer, C., Bais, R., Chiang, H.-H., Chao, J. H., Khakpour, D., Panfil, P., Pasquel, F. J., Ullal, J., Gligorijevic, N., et al. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2025.0184 · Cited by 10
Faulds, E. R., Hester, J. C., Badakhshi, Y., Miller, J. D., Basil, R. C., Chandra, S., Chang, A. S., Garcia, M., JONES, L. G., Pintor-Jimenez, K. A., et al. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251343108 · Cited by 11
“3,6,10,15 While POC frequency was reduced significantly from standard hourly testing at all institutions, it is interesting to note that at hospitals A and B, which required Q6 ongoing POC testing and validation, the mean frequency of POC testing was 6.7 ± 3.1 and 5.9 ± 3.5, respectively.”
InfluentialUllal, J., Spanbauer, C., Baran, J. D., Bais, R., Chiang, H.-H., Chao, J. H., Khakpour, D., Panfil, P., Pasquel, F. J., Gligorijevic, N., et al. (2025) · Diabetes Technology & Therapeutics · doi:10.1177/15209156251368933 · Cited by 8
“In previous studies in the ICU, 6,8,9,12,22 – 25 the G6 sensor (not the G6 Pro) has demonstrated minimal bias when reported and a mean RAD generally of 11% – 14%, substantially lower than what we observed with the G6 Pro (Table 5).”
Diabetes Technology Meeting 2020
Shang, T., Zhang, J. Y., Bequette, B. W., Raymond, J. K., Coté, G. L., Sherr, J. L., Castle, J. R., Pickup, J. C., Pavlovic, Y., Espinoza, J., et al. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211016480 · Cited by 9
“37 Furthermore, use of CGM based metrics has helped guide optimization of therapy during telehealth encounters.”
BackgroundTienstra, M., de Boer, J., van Doesum, J. A., Keijzer, K., Morsink, L. M., Hazenberg, C. L. E., Ammatuna, E., Huls, G. A., Choudhary, P., Gans, R. O. B., et al. (2024) · Endocrine Practice · doi:10.1016/j.eprac.2024.09.013 · Cited by 7
Rayan, M. N., Faulds, E. R., Lee, B., McNett, M., Exline, M. C., Lin, C. J., JONES, L. G., Rasul, A., & Dungan, K. M. (2025) · CHEST Critical Care · doi:10.1016/j.chstcc.2025.100193 · Cited by 5
“4–11 Despite these advantages, CGM use in hospitals, particularly in ICU settings, presents challenges because of clinical conditions that affect the accuracy of interstitial glucose measurement, including rapid glucose fluctuations, hypoperfusion, acidosis, and interfering substances.”
Monitoring Devices for glucose in critical patients during the COVID-19 pandemic
Xue, X., Wang, Y., Wang, J., Zhao, J., Tang, Y., Song, X., & Xu, C. (2022) · International Healthcare Review (online) · doi:10.56226/ihr.v1i2.29 · Cited by 5
Innovation, Technology, and Telemedicine in Critical Care
Shaikh, A., Patel, D., & Sakhuja, A. (2025) · Critical Care Clinics · doi:10.1016/j.ccc.2025.11.003 · Cited by 4
Continuous Glucose Monitoring System After Coronary Artery Bypass Graft Surgery: A Feasibility Study
Torre, T., Schlotterbeck, H., Ferraro, F. R., Klersy, C., Surace, G. G., Toto, F., Pozzoli, A., Ferrari, E., & Demertzis, S. (2024) · Asian Cardiovascular and Thoracic Annals · doi:10.1177/02184923241240035 · Cited by 4
Shimizu, N., Hayashi, A., Ito, S., Suzuki, A., Fujishima, R., Matoba, K., Wada, T., Takano, K., Katagiri, M., & Shichiri, M. (2021) · Endocrine Journal · doi:10.1507/endocrj.ej21-0540 · Cited by 4
“Libre system has the advantage of being cheaper than previously reported remote realtime CGMs [5-11].”
BackgroundResultHinoue, T., Yatabe, T., Fujiwara, H., & Nishida, O. (2021) · Journal of Anesthesia · doi:10.1007/s00540-021-02965-1 · Cited by 4
“ble and reliable in reducing the frequency of point-of-care glucose testing, patient and healthcare worker exposure, and personal protective equipment (PPE) use during the pandemic [3, 4].”
BackgroundSakjirapapong, C., Sirinvaravong, S., Preechasuk, L., & Thongtang, N. (2025) · Journal of Diabetes Research · doi:10.1155/jdr/7538573 · Cited by 3
“According to the previous study [9], overall MARD was 13.1% and the standard deviation (SD) of MARD was 10%; a sample size of 15 would provide a precision error of 5% and a 95% con fi dence interval.”
InfluentialAccuracy of continuous glucose monitoring in critically ill patients
Santamaria, J. D., Selers, E., & Reid, D. (2025) · Critical Care and Resuscitation · doi:10.1016/j.ccrj.2025.100118 · Cited by 5
Hospital Diabetes Meeting 2022
Huang, J. K., Yeung, A. M., Nguyen, K. T., Xu, N. Y., Preiser, J., Rushakoff, R. J., Seley, J. J., Umpierrez, G. E., Wallia, A., Drincic, A., et al. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221110878 · Cited by 4
Pañero‐Moreno, M., Muxella-Roson, M., Dubè-Llobet, G., Cabrerizo-Jimenez, M., Guix‐Comellas, E. M., Ordozgoiti, A. V., & Marsà-Perez, N. (2025) · Enfermería Intensiva (English ed ) · doi:10.1016/j.enfi.2025.500534 · Cited by 5
Aziz, Q.-U.-A., Batra, K., Fatima, S., Splinter, J., Champion, A. L., Kumar, A. M. V., & Izuora, K. E. (2025) · Endocrine Practice · doi:10.1016/j.eprac.2025.04.022 · Cited by 6
Continuous Glucose Monitoring in Hospital Settings
Faulds, E. R., & Hughes, M. S. (2025) · Medical Clinics of North America · doi:10.1016/j.mcna.2025.11.008 · Cited by 2
Protuš, M., Hagerf, B. V., Jabor, A., Franeková, J., Németová, L., Uchytilová, E., Indrova, V., Beckova, J., Macek, A., Dolečková, M., et al. (2026) · Therapeutic Advances in Endocrinology and Metabolism · doi:10.1177/20420188251405372 · Cited by 2
“…of perioperative CGM use, 25–27 as well as reports evaluating exceptional applications of CGM in intensive care during the COVID-19 pandemic, 19–24 demonstrate that the accuracy of contemporary CGM systems has improved and that they might be suitable for broader use beyond their originally…”
Paoli, L. L., Rizzi, A., Tartaglione, L., D'Amore, D., Vinti, L., Barberio, A., Sciaraffia, G., Viti, L., Leo, M. A. S. D., Pontecorvi, A., et al. (2026) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.70154 · Cited by 2
“Similarly, a study conducted by Sadhu et al., involving 11 adults with COVID‐19 with a history of DM or newly diagnosed hyperglycemia, showed MARD values of 11.1% for Dexcom G6 and 13.1% for Medtronic Guardian Connect, reinforcing the system's stability in high acuity environments [27].”
InfluentialArtificial intelligence in critically ill diabetic patients: current status and future prospects
Juneja, D., Gupta, A., & Singh, O. (2022) · Artificial Intelligence in Gastroenterology · doi:10.35712/aig.v3.i2.66 · Cited by 3
Bogun, M. M., Wang, C., Kurlansky, P. A., Bedeir, N., & Umpierrez, G. E. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251316887 · Cited by 3
“14-17 Here, we report our experience using real-time CGM to manage adult patients with DKA.”
Hospital care: improving outcomes in type 1 diabetes
BOEDER, S. C., & Kulasa, K. M. (2020) · Current Opinion in Endocrinology Diabetes and Obesity · doi:10.1097/med.0000000000000601 · Cited by 3
Logothetis, G., Avramidis, K., Konstantaki, E., Matziou, V. N., & Doupis, J. (2025) · touchREVIEWS in Endocrinology · doi:10.17925/ee.2025.21.1.7 · Cited by 1
“…rtCGM-POC readings were within acceptable zones A and B, while for Medtronic Guardian Connect, 100% of the readings were within zones A and B. 24,25 In contrast, the analysis of Dexcom G6 resulted in 0.2% of pairs in zone C (readings may have the potential to result in inappropriate…”
InfluentialScoping review of subcutaneous glucose monitoring techniques
Hrubá, E., Kubíček, J., & Augustýnek, M. (2025) · Measurement · doi:10.1016/j.measurement.2025.119940 · Cited by 1
Perioperatives Management von Menschen mit Diabetes mellitus: CGM im Krankenhaus
Reger-Tan, S., Lee-Barkey, Y. H., Stratmann, B., Klose, J. M., & Heinemann, L. (2026) · Die Diabetologie · doi:10.1007/s11428-025-01409-8 · Cited by 1
Diabetes in the Critically Ill Patient: DKA, HHS, and Beyond
Singh, B., Chlebek, S., & Krikorian, A. (2023) · Contemporary Endocrinology · doi:10.1007/978-3-031-44648-1_5 · Cited by 2
Pañero‐Moreno, M., Guix‐Comellas, E. M., & Ordozgoiti, A. V. (2024) · Nursing in Critical Care · doi:10.1111/nicc.13198 · Cited by 4
“13,14,19 Effective glycaemic control is indicated by a time in range (TIR) above 50% and a time below range (TBR) of less than 1%.”
Os procedimentos transradiais e suas complicações: um olhar da enfermagem
de Moraes, L. F., dos Santos, T. G., Spíndola, D. G., Maia, J. S., & Figueredo, L. P. (2021) · Revista Recien · doi:10.24276/rrecien2021.11.35.347-354 · Cited by 1
Tecnologias realizadas pela enfermagem em portadores de COVID-19: revisão integrativa
dos Santos, E. S., Luna, A. A., & Szpalher, A. S. (2022) · Revista Recien · doi:10.24276/rrecien2022.12.37.222-232
Management of Critically Ill Persons with COVID-19 and Diabetes
Mathew, J. T., & Lee, H. J. (2023) · Contemporary Endocrinology · doi:10.1007/978-3-031-28536-3_9
Faulds, E. R., Lee, B., Rasul, A., JONES, L. G., McNett, M., Exline, M. C., Shoben, A. B., Lin, C. J., Madan, R., & Dungan, K. M. (2024) · Research Square · doi:10.21203/rs.3.rs-4243392/v1
González-Aguirre, J. E., Medina-Ramirez, H. A., Soto-Monciváis, B., Sanchez-Ruiz, S. R., Garcia-Arroyo, C., Moreno-Hoyos-Abril, J. F., & Náñez-Terreros, H. (2026) · Archives of Medical Research · doi:10.1016/j.arcmed.2026.103382
Pañero‐Moreno, M., Carbonell-Sabate, C., Conesa-González, R., Risco-Patiño, L., Guix‐Comellas, E. M., Ordozgoiti, A. V., & Cordón-Arnau, T. (2026) · Enfermería Intensiva · doi:10.1016/j.enfi.2026.500589
Pañero‐Moreno, M., Carbonell-Sabate, C., Conesa-González, R., Risco-Patiño, L., Guix‐Comellas, E. M., Ordozgoiti, A. V., & Cordón-Arnau, T. (2026) · Enfermería Intensiva (English ed ) · doi:10.1016/j.enfie.2026.500589
Aguilar-Schotborgh, M., Múñoz, Ó., Henao, D. C., Triana, L. C., Mogollon, J. M., & Medina, A. M. G. (2026) · Clinical Medicine Insights Endocrinology and Diabetes · doi:10.1177/11795514261470879
Metadata
- Authors
- Sadhu, A. R., Serrano, I. A., Xu, J., Nisar, T., Lucier, J., Pandya, A. R., & Patham, B.
- Affiliations
- Houston Methodist (US)
- Houston Methodist Research Institute (US)
- Corresponding author
- Sadhu, Archana R.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2020-10-16
- Volume
- 14
- Issue
- 6
- Pages
- 1065-1073
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC7645121
- OpenAlex
- W3093094212
- Cited by
- 84
- References
- 29
- Citations by year
- 2026: 7 · 2025: 15 · 2024: 14 · 2023: 11 · 2022: 16 · 2021: 20 · 2020: 1
- FWCI
- 9.05
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Keywords
- continuous glucose monitoring, COVID-19 pandemic, critically ill patients, intensive care unit, insulin therapy, CGM accuracy, sensor reliability, Dexcom G6, mean absolute relative difference, Clarke error grid, Bland-Altman analysis
- MeSH terms
- Betacoronavirus, COVID-19, SARS-CoV-2, Adult, Aged, Blood Glucose, Diabetes Mellitus, Feasibility Studies, Female, Humans, Hyperglycemia, Insulin, Insulin Infusion Systems, Intensive Care Units, Male, Middle Aged, Monitoring, Physiologic, Pilot Projects, Pneumonia, Viral, Prognosis, Retrospective Studies, Blood Glucose Self-Monitoring, Reproducibility of Results, Critical Illness, Coronavirus Infections, Point-of-Care Systems, Pandemics
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.