A neuroprosthesis for tremor management through the control of muscle co-contraction
Open accessAbstract
BACKGROUND: Pathological tremor is the most prevalent movement disorder. Current treatments do not attain a significant tremor reduction in a large proportion of patients, which makes tremor a major cause of loss of quality of life. For instance, according to some estimates, 65% of those suffering from upper limb tremor report serious difficulties during daily living. Therefore, novel forms for tremor management are required. Since muscles intrinsically behave as a low pass filter, and tremor frequency is above that of volitional movements, the authors envisioned the exploitation of these properties as a means of developing a novel treatment alternative. This treatment would rely on muscle co-contraction for tremor management, similarly to the strategy employed by the intact central nervous system to stabilize a limb during certain tasks. METHODS: We implemented a neuroprosthesis that regulated the level of muscle co-contraction by injecting current at a pair of antagonists through transcutaneous neurostimulation. Co-contraction was adapted to the instantaneous parameters of tremor, which were estimated from the raw recordings of a pair of solid state gyroscopes with a purposely designed adaptive algorithm. For the experimental validation, we enrolled six patients suffering from parkinsonian or essential tremor of different severity, and evaluated the effect of the neuroprosthesis during standard tasks employed for neurological examination. RESULTS: The neuroprosthesis attained significant attenuation of tremor (p<0.001), and reduced its amplitude up to a 52.33±25.48%. Furthermore, it alleviated both essential and parkinsonian tremor in spite of their different etiology and symptomatology. Tremor severity was not a limiting factor on the performance of the neuroprosthesis, although there was a subtle trend towards larger attenuation of more severe tremors. Tremor frequency was not altered during neurostimulation, as expected from the central origin of Parkinson's disease and essential tremor. All patients showed a good tolerance to neurostimulation in terms of comfort and absence of pain, and some spontaneously reported that they felt that tremor was reduced when the neuroprosthesis was activated. CONCLUSIONS: The results presented herein demonstrate that the neuroprosthesis provides systematic attenuation of the two major types of tremor, irrespectively from their severity. This study sets the basis for the validation of the neuroprosthesis as an alternative, non-invasive means for tremor management.
Overview
CoChat Summary
TL;DR
This paper presents a wearable neuroprosthesis that reduces upper-limb tremor by transcutaneous neurostimulation that induces muscle co-contraction at antagonist muscles. In 6 patients with essential tremor or Parkinsonian tremor, it significantly reduced tremor amplitude, with a mean reduction of 52.33±25.48% (p<0.001), without changing tremor frequency.
Key findings
- In 6 patients (1 female; 2 PD, 4 ET; mean age 68.2±13.8 years; tremor severity 3 to 30 on the Fahn-Tolosa-Marin score), the neuroprosthesis reduced tremor amplitude to 52.33±25.48% of baseline on average.
- The reduction was statistically significant (p<0.001) and was observed in 26 out of 30 trials.
- More severe tremors were attenuated to a greater extent (p=0.008), and the trend in attenuation vs. baseline amplitude had a log-log slope of −0.160 with R²=0.243.
- Tremor frequency was not altered by neurostimulation (p=0.831).
Methods
- The authors built a wearable neuroprosthesis that delivered transcutaneous neurostimulation to antagonist muscles to increase co-contraction and thereby manipulate joint stiffness and viscosity.
- Tremor was estimated from solid-state gyroscopes using a two-stage adaptive signal-processing pipeline: a g-h filter to separate voluntary movement, then a WFLC + Kalman filter cascade to track tremor frequency and amplitude.
- They tested the device in 6 patients with either Parkinson’s disease (n=2) or essential tremor (n=4) during standard neurological tasks; analysis compared trials with vs. without active stimulation using Mann-Whitney tests and a power-spectrum-based attenuation ratio (Ratt).
Limitations
- The paper states that controller gains were selected manually, and says better automatic gain selection could improve outcomes and might reduce the 4 trials in which tremor worsened.
- The authors note that muscle accommodation/fatigue during extended use would require adapting controller parameters over time.
- Evident from the text: the sample was very small (n=6) and heterogeneous, so the findings are promising but limited for generalization.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures

![Figure 2 Rationale behind the tremor suppression approach. Frequency response of a human joint, showing how it is modified by co-contraction of the antagonist muscle pair. The data for the human joint without the neuroprosthesis (black trace) corresponds to the model given in (1) fitted to the parameters identified for the wrist in [39]. The remainder traces represent how the response of the joint is modified by a concurrent increase in viscosity and stiffness (defined as a multiple of the parameters D and K), and correspond to: i) equal increments of DNP and KNP (in multiples of 10, from 10 to 100, shown as blue traces), ii) larger increments of DNP (in multiples of 50, from 50 to 500) than of KNP (in multiples of 10, from 10 to 100, shown as red traces), and iii) larger increments of KNP (in multiples of 50, from 50 to 500) than of DNP (in multiples of 10, from 10 to 100, shown as yellow traces). For all cases, larger values of DNP and KNP cause the response of the joint to have smaller magnitude (as indicated by the arrow in gray).](https://figures.cochat.ai/paper-figures/a771dc83b5fc4b439bcb10808eb4b0eb/a-neuroprosthesis-for-tremor-management-through-the-control-of-muscle-co-figure-2.jpg)
![Figure 3 Example of the controller of the NP. The plot shows the estimation of tremor (solid line) and the amplitude of the current applied at the extensors (dashed line) and flexors (dotted line). The instant at which the NP was triggered is signaled with an arrow. Angular velocity is scaled [x 15] for visualization sake; a positive value corresponds to wrist extension.](https://figures.cochat.ai/paper-figures/a8020494205d4b55b9377c7fc0abd43c/a-neuroprosthesis-for-tremor-management-through-the-control-of-muscle-co-figure-3.jpg)




Verification
Claims the paper makes
1 Valid · 2 Not valid · 23 Not checked · 47 Own results and interpretations
The sources with a verdict support this claim; the rest weren’t checked yet, have no full text we could read, or their verdicts aren’t rated yet.
In the paper
“voluntary [24,25] or artificially elicited [26] muscle contractions affect the severity of the tremor, by generating forces on the musculoskeletal system.”
§ A neuroprosthesis for tremor management through the control of muscle co-contraction > Background
What the sources say
- Not checked — no full text of the sourcecites [24] source
- Not checked — no full text of the sourcecites [25] source
- Supportedcites [26] source
For the artificially elicited contractions attributed to this source as [26], the Introduction says FES activates muscles as actuators and that the proposed system will apply biomechanical loads to suppress tremor. It also states that applying force can change tremor amplitude. The voluntary-contraction portion is cited to other sources. (LLM, conf=0.87)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“All patients showed a good tolerance to neurostimulation in terms of comfort and absence of pain”
§ A neuroprosthesis for tremor management through the control of muscle co-contraction > Abstract > Results
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper states — it cites no source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Tremor frequency was not altered during neurostimulation, as expected from the central origin of Parkinson’s disease and essential tremor.”
§ A neuroprosthesis for tremor management through the control of muscle co-contraction > Abstract > Results
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Tremor severity was not a limiting factor on the performance of the neuroprosthesis”
§ A neuroprosthesis for tremor management through the control of muscle co-contraction > Abstract > Results
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“reduced its amplitude up to a 52.33±25.48%”
§ A neuroprosthesis for tremor management through the control of muscle co-contraction > Abstract > Results
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
38 more claims
Create a free account to see all 73 claims and to verify the 9 references still open.
- It may state two separate claims as one. 18
- Probably not a substantial claim. 17
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“Alternatively, some approaches modulate the impedance of the human-machine system [20, 24, 29].”
From Controlling a motorized orthosis to follow elbow volitional movement: tests with individuals with pathological tremor (2019) · cites this paper as [29]
The cited source says what the claim attributes to it.
In the citing paper
“Of the above systems, several were demonstrated with individuals with tremor [17–20, 27–32] resulting in ~ 20–88% tremor attenuation levels”
From Controlling a motorized orthosis to follow elbow volitional movement: tests with individuals with pathological tremor (2019) · cites this paper as [29]
The cited source says what the claim attributes to it.
In the citing paper
“Other studies reported some migration of tremor to nearby joints [29].”
From Controlling a motorized orthosis to follow elbow volitional movement: tests with individuals with pathological tremor (2019) · cites this paper as [29]
The cited source says what the claim attributes to it.
In the citing paper
“This neuroprosthesis was validated within two PD patients and four ET patients, who reported a reduction of 52.3 ± 25.5% in 26 out of 30 trials compared to trials where the neuroprosthesis was not active.”
From A Review on Wearable Technologies for Tremor Suppression (2021) · cites this paper as [69]
Citations
Cites
Estimation and Tracking: Principles, Techniques, and Software
Bar‐Shalom, Y., & Li, X. (1993) · Medical Entomology and Zoology · Cited by 2,203
Fedorov, V. V., Studden, W. J., & Klimko, E. M. (1972) · Cited by 2,092
Consensus Statement of the Movement Disorder Society on Tremor
Deuschl, G., Bain, P., & Brin, M. (2008) · Movement Disorders · doi:10.1002/mds.870131303 · Cited by 1,981 · cited 5×
“Typical exercises for neurological examination were considered: resting the arm on the lap, keeping both arms outstretched, the finger to finger test, and the finger to nose test [10].”
BackgroundMethodInfluentialPathological synchronization in Parkinson's disease: networks, models and treatments
Hammond, C., Bergman, H., & Brown, P. (2007) · Trends in Neurosciences · doi:10.1016/j.tins.2007.05.004 · Cited by 1,629 · cited 1×
“PD is estimated to afflict more than ten million people all over the world [9].”
BackgroundRascol, O., Brooks, D. J., Korczyn, A. D., De Deyn, P. P., Clarke, C. E., & Lang, A. E. (2000) · New England Journal of Medicine · doi:10.1056/nejm200005183422004 · Cited by 1,539
Repetitive control system: a new type servo system for periodic exogenous signals
Hara, S., Yamamoto, Y., Omata, T., & Nakano, M. (1988) · IEEE Transactions on Automatic Control · doi:10.1109/9.1274 · Cited by 1,407 · cited 1×
“Tremor parameters estimated during a (tremor) period were employed to generate the control action in the subsequent one, analogously to repetitive control theory [40].”
MethodThe central nervous system stabilizes unstable dynamics by learning optimal impedance
Burdet, E., Osu, R., Franklin, D. W., Milner, T. E., & Kawato, M. (2001) · Nature · doi:10.1038/35106566 · Cited by 1,097 · cited 2×
“Indeed, co-contraction is exploited by the intact central nervous system to stabilize the limbs during specific tasks [35,36].”
BackgroundResultPerlmutter, J. S., & Mink, J. W. (2006) · Annual Review of Neuroscience · doi:10.1146/annurev.neuro.29.051605.112824 · Cited by 1,073
Adaptive control of mechanical impedance by coactivation of antagonist muscles
Hogan, N. (1984) · IEEE Transactions on Automatic Control · doi:10.1109/tac.1984.1103644 · Cited by 1,041 · cited 2×
“Indeed, co-contraction is exploited by the intact central nervous system to stabilize the limbs during specific tasks [35,36].”
BackgroundResultSubthalamic nucleus deep brain stimulation: Summary and meta-analysis of outcomes
Kleiner-Fisman, G., Herzog, J., Fisman, D. N., Tamma, F., Lyons, K. E., Pahwa, R., Lang, A. E., & Deuschl, G. (2006) · Movement Disorders · doi:10.1002/mds.20962 · Cited by 957
Cited by
Online tremor suppression using electromyography and low-level electrical stimulation
Dosen, S., Muceli, S., Dideriksen, J. L., Romero, J. P., Rocon, E., Pons, J., & Farina, D. (2015) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2014.2328296 · Cited by 120
“In some recent studies [17]–[20] tremor was detected using inertial sensors to measure the kinematics (angle, velocity, and/or acceleration) of the affected joint(s) and extract tremor features (amplitude, frequency).”
BackgroundResultSelf-citationEvancho, A. M., Tyler, W. J., & McGregor, K. (2023) · Frontiers in Human Neuroscience · doi:10.3389/fnhum.2023.1151218 · Cited by 110
“, 2022), yielding an average rate of tremor inhibition of over 50% (Popović Maneski et al., 2011; Zhang et al., 2011; Gallego et al., 2013; Freeman et al., 2015).”
BackgroundElectrical stimulation of afferent pathways for the suppression of pathological tremor
Dideriksen, J. L., Laine, C. M., Dosen, S., Muceli, S., Rocon, E., Pons, J. L., Benito-Leon, J., & Farina, D. (2017) · Frontiers in Neuroscience · doi:10.3389/fnins.2017.00178 · Cited by 67
“…the use of external devices to suppress tremor by mechanical loading (Pledgie et al., 2000; Rocon et al., 2007) or electrical stimulation (Javidan et al., 1992; Prochazka et al., 1992; Gillard et al., 1999; Popović Maneski et al., 2011; Gallego et al., 2013; Bó et al., 2014; Dosen et al., 2015).”
BackgroundMethodInfluentialSelf-citationPeripheral electrical stimulation to reduce pathological tremor: a review
Pascual-Valdunciel, A., Hoo, G. W., Avrillon, S., Barroso, F. O., Goldman, J. G., Hernandez-Pavon, J. C., & Pons, J. L. (2021) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-021-00811-9 · Cited by 66
“In this strategy, simultaneous electrical stimulation applied above the motor threshold in both agonist and antagonist muscles increase the impedance of the joint and therefore minimize the involuntary oscillations [17, 20, 39].”
BackgroundSelf-citationMedical Devices for Tremor Suppression: Current Status and Future Directions
Mo, J., & Priefer, R. (2021) · Biosensors · doi:10.3390/bios11040099 · Cited by 78
“TREMOR neurorobot ET (4); PD tremor (2) 52% tremor suppression [81]”
Background
Metadata
- Authors
- Gallego, J. Á., Rocon, E., Belda-Lois, J. M., & Pons, J. L.
- Affiliations
- Consejo Superior de Investigaciones Científicas (ES)
- Corresponding author
- Gallego, Juan Álvaro
- Venue
- Journal of NeuroEngineering and Rehabilitation
- Published
- 2013
- Volume
- 10
- Issue
- 1
- Pages
- 36
- ISSN
- 1743-0003 (ISSN-L)
- PMC
- PMC3661364
- OpenAlex
- W2163040354
- Cited by
- 81
- References
- 65
- Citations by year
- 2025: 4 · 2024: 5 · 2023: 8 · 2022: 11 · 2021: 7 · 2020: 5 · 2019: 9 · 2018: 2 · 2017: 8 · 2016: 5 · 2015: 7 · 2014: 6 · 2013: 4
- FWCI
- 4.65
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- Locations
- Journal of NeuroEngineering and Rehabilitation (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DIGITAL.CSIC (Spanish National Research Council (CSIC)) (submittedVersion)
- Funders
- European Commission (ICT-2011, 287739, 224051)
- Universitat de València
- Ministerio de Ciencia e Innovación (CSD2009, CSD2009-00067)
- Consejo Superior de Investigaciones Científicas
- Topics
- Health Sciences › Medicine › Clinical Neurology › Deep Brain Stimulation and Movement Disorders
- Life Sciences › Neuroscience › Cognitive Neuroscience › Motor Control and Adaptation
- Health Sciences › Medicine › Clinical Neurology › Botulinum Toxin and Related Neurological Disorders
- Keywords
- pathological tremor, neural prostheses, co-contraction, Parkinson's disease, essential tremor, gyroscope, adaptive algorithms, non-invasive treatment, antagonists
- MeSH terms
- Adult, Aged, Aged, 80 and over, Algorithms, Transcutaneous Electric Nerve Stimulation, Female, Humans, Male, Middle Aged, Muscle Contraction, Parkinson Disease, Muscle, Skeletal, Essential Tremor, Neural Prostheses
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A neuroprosthesis for tremor management through the control of muscle co-contraction. https://app.cochat.ai/papers/10.1186/1743-0003-10-36
- The paper
- Gallego, J. Á., Rocon, E., Belda-Lois, J. M., & Pons, J. L. (2013). A neuroprosthesis for tremor management through the control of muscle co-contraction. Journal of NeuroEngineering and Rehabilitation. https://doi.org/10.1186/1743-0003-10-36
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.