The Accuracy of 6 Inexpensive Pulse Oximeters Not Cleared by the Food and Drug Administration: The Possible Global Public Health Implications
The DOI matches its Crossref record; 9 of the 9 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Universal access to pulse oximetry worldwide is often limited by cost and has substantial public health consequences. Low-cost pulse oximeters have become increasingly available with limited regulatory agency oversight. The accuracy of these devices often has not been validated, raising questions about performance. METHODS: The accuracy of 6 low-cost finger pulse oximeters during stable arterial oxygen saturations (Sa o 2) between 70% and 100% was evaluated in 22 healthy subjects. Oximeters tested were the Contec CMS50DL, Beijing Choice C20, Beijing Choice MD300C23, Starhealth SH-A3, Jumper FPD-500A, and Atlantean SB100 II. Inspired oxygen, nitrogen, and carbon dioxide partial pressures were monitored and adjusted via a partial rebreathing circuit to achieve 10 to 12 stable target Sa o 2 plateaus between 70% and 100% and Pa co 2 values of 35 to 45 mm Hg. Comparisons of pulse oximeter readings (Sp o 2) with arterial Sa o 2 (by Radiometer ABL90 and OSM3) were used to calculate bias (Sp o 2 – Sa o 2) mean, precision (SD of the bias), and root mean square error (A RMS). RESULTS: Pulse oximeter readings corresponding to 536 blood samples were analyzed. Four of the 6 oximeters tested showed large errors (up to −6.30% mean bias, precision 4.30%, 7.53 A RMS) in estimating saturation when Sa o 2 was reduced <80%, and half of the oximeters demonstrated large errors when estimating saturations between 80% and 90%. Two of the pulse oximeters tested (Contec CMS50DL and Beijing Choice C20) demonstrated A RMS of <3% at Sa o 2 between 70% and 100%, thereby meeting International Organization for Standardization (ISO) criteria for accuracy. CONCLUSIONS: Many low-cost pulse oximeters sold to consumers demonstrate highly inaccurate readings. Unexpectedly, the accuracy of some low-cost pulse oximeters tested here performed similarly to more expensive, ISO-cleared units when measuring hypoxia in healthy subjects. None of those tested here met World Federation of Societies of Anaesthesiologists standards, and the ideal testing conditions do not necessarily translate these findings to the clinical setting. Nonetheless, further development of accurate, low-cost oximeters for use in clinical practice is feasible and, if pursued, could improve access to safe care, especially in low-income countries.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
An estimation of the global volume of surgery: a modelling strategy based on available data
Weiser, T. G., Regenbogen, S. E., Thompson, K. D., Haynes, A. B., Lipsitz, S. R., Berry, W. R., & Gawande, A. A. (2008) · The Lancet · doi:10.1016/s0140-6736(08)60878-8 · Cited by 2,511
Agreement Between Methods of Measurement with Multiple Observations Per Individual
Bland, J. M., & Altman, D. G. (2007) · Journal of Biopharmaceutical Statistics · doi:10.1080/10543400701329422 · Cited by 2,065
Effects of skin pigmentation on pulse oximeter accuracy at low saturation
Bickler, P. E., Feiner, J. R., & Severinghaus, J. W. (2005) · Anesthesiology · doi:10.1097/00000542-200504000-00004 · Cited by 427
Self-citationFeiner, J. R., Severinghaus, J. W., & Bickler, P. E. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000285988.35174.d9 · Cited by 400
Self-citationGlobal operating theatre distribution and pulse oximetry supply: an estimation from reported data
Funk, L. M., Weiser, T. G., Berry, W. R., Lipsitz, S. R., Merry, A. F., Enright, A. C., Wilson, I. H., Dziekan, G., & Gawande, A. A. (2010) · The Lancet · doi:10.1016/s0140-6736(10)60392-3 · Cited by 397
Jubran, A., & Tobin, M. J. (1990) · Chest · doi:10.1378/chest.97.6.1420 · Cited by 296
Ries, A. L., Prewitt, L. M., & Johnson, J. J. (1989) · Chest · doi:10.1378/chest.96.2.287 · Cited by 145
Reliability of noninvasive oximetry in black subjects during exercise and hypoxia
Zeballos, R. J., & Weisman, I. M. (1991) · American Review of Respiratory Disease · doi:10.1164/ajrccm/144.6.1240 · Cited by 108
Effect of skin pigmentation on pulse oximetry accuracy in the emergency department
Adler, J. N., Hughes, L. A., Vtvilecchia, R., & Jr., C. A. C. (1998) · Academic Emergency Medicine · doi:10.1111/j.1553-2712.1998.tb02772.x · Cited by 102
Cited by
Luks, A. M., & Swenson, E. R. (2020) · Annals of the American Thoracic Society · doi:10.1513/annalsats.202005-418fr · Cited by 263
The 2023 wearable photoplethysmography roadmap
Charlton, P. H., Allen, J., Bailón, R., Baker, S., Behar, J. A., Chen, F., Clifford, G. D., Clifton, D. A., Davies, H. J., Ding, C., et al. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acead2 · Cited by 187
A review of the effect of skin pigmentation on pulse oximeter accuracy
Al-Halawani, R., Charlton, P. H., Qassem, M., & Kyriacou, P. A. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acd51a · Cited by 119
Home monitoring in interstitial lung diseases
Wijsenbeek, M. S., Moor, C. C., Johannson, K. A., Jackson, P. D., Khor, Y. H., Kondoh, Y., Rajan, S. K., Tabaj, G. C., Varela, B. E., van der Wal, P., et al. (2023) · The Lancet Respiratory Medicine · doi:10.1016/s2213-2600(22)00228-4 · Cited by 98
Pulse Oximeter Performance, Racial Inequity, and the Work Ahead
Okunlola, O. E., Lipnick, M. S., Batchelder, P. B., Bernstein, M., Feiner, J. R., & Bickler, P. E. (2022) · Respiratory Care · doi:10.4187/respcare.09795 · Cited by 96
“Some inexpensive oximeters perform well enough in laboratory studies to meet FDA accuracy requirements for clinical care, but many do not.(17) Poor performance of many inexpensive devices may be especially true under conditions of low perfusion and dark-skin pigmentation among others.”
BackgroundSelf-citation
Metadata
- Authors
- Lipnick, M. S., Feiner, J. R., Au, P., Bernstein, M., & Bickler, P. E.
- Venue
- Anesthesia & Analgesia
- Published
- 2016-08
- Volume
- 123
- Issue
- 2
- Pages
- 338-345
- ISSN
- 0003-2999 (ISSN-L)
- OpenAlex
- W2339693659
- Cited by
- 100
- References
- 9
- FWCI
- 2.28
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.