Digital Photoplethysmography for Assessment of Arterial Stiffness: Repeatability and Comparison with Applanation Tonometry
Open accessAbstract
INTRODUCTION: Arterial stiffness is an independent risk factor for cardiovascular morbidity and can be assessed by applanation tonometry by measuring pulse wave velocity (PWV) and augmentation index (AIX) by pressure pulse wave analysis (PWA). As an inexpensive and operator independent alternative, photoelectric plethysmography (PPG) has been introduced with analysis of the digital volume pulse wave (DPA) and its second derivatives of wave reflections. OBJECTIVE: The objective was to investigate the repeatability of arterial stiffness parameters measured by digital pulse wave analysis (DPA) and the associations to applanation tonometry parameters. METHODS AND RESULTS: 112 pregnant and non-pregnant individuals of different ages and genders were examined with SphygmoCor arterial wall tonometry and Meridian DPA finger photoplethysmography. Coefficients of repeatability, Bland-Altman plots, intraclass correlation coefficients and correlations to heart rate (HR) and body height were calculated for DPA variables, and the DPA variables were compared to tonometry variables left ventricular ejection time (LVET), PWV and AIX. No DPA variable showed any systematic measurement error or excellent repeatability, but dicrotic index (DI), dicrotic dilatation index (DDI), cardiac ejection elasticity index (EEI), aging index (AI) and second derivatives of the crude pulse wave curve, b/a and e/a, showed good repeatability. Overall, the correlations to AIX were better than to PWV, with correlations coefficients >0.70 for EEI, AI and b/a. Considering the level of repeatability and the correlations to tonometry, the overall best DPA parameters were EEI, AI and b/a. The two pansystolic time parameters, ejection time compensated (ETc) by DPA and LVET by tonometry, showed a significant but weak correlation. CONCLUSION: For estimation of the LV function, ETc, EEI and b/a are suitable, for large artery stiffness EEI, and for small arteries DI and DDI. The only global parameter, AI, showed a high repeatability and the overall best correlations with AIX and PWV.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Over 30 features have been proposed for assessing the stiffness of both the large and small arteries from a single PPG”
From Wearable Photoplethysmography for Cardiovascular Monitoring (2022) · cites this paper as [172]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619 · cited 1×
“We described agreement between the two DPA measurements with Bland-Altman plots [36,37].”
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1191/096228099673819272 · Cited by 9,619
Expert consensus document on arterial stiffness: methodological issues and clinical applications
Laurent, S., Cockcroft, J., Van Bortel, L., Boutouyrie, P., Giannattasio, C., Hayoz, D., Pannier, B., Vlachopoulos, C., Wilkinson, I., Struijker-Boudier, H., et al. (2006) · European Heart Journal · doi:10.1093/eurheartj/ehl254 · Cited by 6,077
McDonald's blood flow in arteries; theoretical, experimental and clinical principles
(1992) · General Pharmacology The Vascular System · doi:10.1016/0306-3623(92)90079-y · Cited by 2,395
Aortic Stiffness Is an Independent Predictor of Primary Coronary Events in Hypertensive Patients
Boutouyrie, P., Tropeano, A., Asmar, R. G., Gautier, I., Bénétos, A., Lacolley, P. J., & Laurent, S. (2002) · Hypertension · doi:10.1161/hy0102.099031 · Cited by 1,745 · cited 2×
“Although pressure PW analysis shows benefits in predicting cardiovascular morbidity [1,10], the impact in clinical practice is small because the tonometry method is cumbersome with a need for highly specialized operators.”
McDonald's Blood Flow in Arteries: Theoretical, Experimental and Clinical Principles
Vlachopoulos, C., O'Rourke, M., & Nichols, W. W. (2011) · doi:10.1201/b13568 · Cited by 1,613
Aortic pulse wave velocity as a marker of cardiovascular risk in hypertensive patients
Blacher, J., Asmar, R., Djane, S., London, G. M., & Safar, M. E. (1999) · Hypertension · doi:10.1161/01.hyp.33.5.1111 · Cited by 1,493 · cited 1×
“The PWV is correlated to age, blood pressure (BP) and arteriosclerosis and has a better predictive value for cardiovascular disease than traditional risk markers [3,4].”
On the Analysis of Fingertip Photoplethysmogram Signals
Elgendi, M. (2012) · Current Cardiology Reviews · doi:10.2174/157340312801215782 · Cited by 1,206 · cited 3×
“The e wave represents the dictrotic notch and corresponds to the aortic valve closure [11].”
InfluentialClinical measurement of arterial stiffness obtained from noninvasive pressure waveforms
NICHOLS, W. (2005) · American Journal of Hypertension · doi:10.1016/j.amjhyper.2004.10.009 · Cited by 1,008 · cited 1×
“AIX is influenced by both large artery stiffness and properties in the PW-reflecting peripheral arterial tree [5 – 7] and is associated with cardiovascular risk, although it is less established than the PWV as a risk factor for cardiovascular morbidity [8 – 10].”
Foundations of Clinical Research
Portney, L. G., & Watkins, M. P. (1993) · Cited by 979
Bartlett, J. W., & Frost, C. D. (2008) · Ultrasound in Obstetrics and Gynecology · doi:10.1002/uog.5256 · Cited by 943 · cited 9×
“Agreement quantifies how close measurements made on the same subject are under identical conditions over a short period of time [34].”
InfluentialKawaguchi, M., Hay, I., Fetics, B. J., & Kass, D. A. (2003) · Circulation · doi:10.1161/01.cir.0000048123.22359.a0 · Cited by 929 · cited 1×
“An inverse relation exists between myocardial contractility and arterial compliance [47] and both the b/a ratio and the AIX reflect these entities.”
A critical review of the systolic time intervals.
Lewis, R. P., Rittogers, S. E., Froester, W. F., & Boudoulas, H. (1977) · Circulation · doi:10.1161/01.cir.56.2.146 · Cited by 624
Determination of age-related increases in large artery stiffness by digital pulse contour analysis
Millasseau, S., Kelly, R. P., Ritter, J., & Chowienczyk, P. (2002) · Clinical Science · doi:10.1042/cs1030371 · Cited by 624
Takazawa, K., Tanaka, N., Fujita, M., Matsuoka, O., Saiki, T., Aikawa, M., Tamura, S., & Ibukiyama, C. (1998) · Hypertension · doi:10.1161/01.hyp.32.2.365 · Cited by 589
The Case for Using the Repeatability Coefficient When Calculating Test–Retest Reliability
Vaz, S. M. A., Falkmer, T., Passmore, A. E., Parsons, R. W., & Andreou, P. (2013) · PLoS ONE · doi:10.1371/journal.pone.0073990 · Cited by 476 · cited 2×
“The ICC takes dimensionless values between zero and one [34,35], where the value of (1 –ICC) relates to the errors in the measurement process.”
Augmentation index is associated with cardiovascular risk
Nürnberger, J., Keflioglu-Scheiber, A., Saez, A. M. O., Wenzel, R. R., Philipp, T., & Schäfers, R. F. (2002) · Journal of Hypertension · doi:10.1097/00004872-200212000-00020 · Cited by 441 · cited 1×
“AIX is influenced by both large artery stiffness and properties in the PW-reflecting peripheral arterial tree [5 – 7] and is associated with cardiovascular risk, although it is less established than the PWV as a risk factor for cardiovascular morbidity [8 – 10].”
Weissler, A. M., Peeler, R. G., & Roehll, W. H. (1961) · American Heart Journal · doi:10.1016/0002-8703(61)90403-3 · Cited by 440 · cited 1×
“The ET varies inversely with the HR and directly with stroke volume [31] and ETc is therefore adjusted to a HR of 60 bpm (beats/second = bpm/60) in the Meridian DPA, reading the calculation equation as ETc = (S-C time)/ p (HR/60); this is a non-linear HR correction equation, originally presented by Bazett in 1920 [32].”
Kelly, R. P., Millasseau, S., Ritter, J. M., & Chowienczyk, P. (2001) · Hypertension · doi:10.1161/01.hyp.37.6.1429 · Cited by 390
Heart rate dependency of pulse pressure amplification and arterial stiffness
Wilkinson, I. B., Mohammad, N. H., Tyrrell, S., Hall, I. R., Webb, D. J., Paul, V. E., Levy, T., & Cockcroft, J. (2002) · American Journal of Hypertension · doi:10.1016/s0895-7061(01)02252-x · Cited by 357 · cited 1×
“from the heart and the amplification is positively correlated to the HR [21], but due to an alteration in the relative timing of the reflected pressure PW, the AIX is inversely related to HR [23].”
Contour analysis of the photoplethysmographic pulse measured at the finger
Millasseau, S., Ritter, J. M., Takazawa, K., & Chowienczyk, P. (2006) · Journal of Hypertension · doi:10.1097/01.hjh.0000239277.05068.87 · Cited by 351 · cited 1×
“Millasseau and colleagues [25] have described the physiological background to the contour analysis of the PPG PW measured in a finger.”
Pulse wave analysis and pulse wave velocity
Davies, J. I., & Struthers, A. D. (2003) · Journal of Hypertension · doi:10.1097/00004872-200303000-00004 · Cited by 292 · cited 3×
“Davies & Struther [20] have described the applanation tonometry method with assessment of the pressure PW.”
InfluentialNoninvasive Assessment of the Digital Volume Pulse
Millasseau, S., Guigui, F. G., Kelly, R. P., Prasad, K., Cockcroft, J., Ritter, J. M., & Chowienczyk, P. (2000) · Hypertension · doi:10.1161/01.hyp.36.6.952 · Cited by 292 · cited 2×
“[12] have in parallel recordings showed that there are close correlations between the contours and characteristics of the peripheral arterial pressure PW and the digital volume PW.”
Bortolotto, L. A., Blacher, J., Kondo, T., Takazawa, K., & Safar, M. E. (2000) · American Journal of Hypertension · doi:10.1016/s0895-7061(99)00192-2 · Cited by 272 · cited 2×
“the more negative, the better; AI increases with age and arteriosclerosis [4,14,16].”
Yasmin, Y. (1999) · QJM · doi:10.1093/qjmed/92.10.595 · Cited by 240 · cited 1×
“AIX is influenced by both large artery stiffness and properties in the PW-reflecting peripheral arterial tree [5 – 7] and is associated with cardiovascular risk, although it is less established than the PWV as a risk factor for cardiovascular morbidity [8 – 10].”
The peripheral pulse wave: Information overlooked
Murray, W. B., & Foster, P. A. (1996) · The Journal of Clinical Monitoring · doi:10.1007/bf02077634 · Cited by 231 · cited 2×
“A correct identification of the dicrotic notch is therefore essential to determine the systolic time [27], but due to variations in HR, cardiac output and vascular tone, with drifting in time positioning of the dicrotic notch with distance from the aortic valve, and to interference with the reflected wave, its applica…”
BackgroundPHOTO-ELECTRIC PLETHYSMOGRAPHY AS A MONITORING DEVICE IN ANAESTHESIA
Dorlas, J. C., & Nijboer, J. A. (1985) · British Journal of Anaesthesia · doi:10.1093/bja/57.5.524 · Cited by 160
Assessing Reproducibility by the Within-Subject Coefficient of Variation with Random Effects Models
Quan, H., & Shih, W.-C. (1996) · Biometrics · doi:10.2307/2532835 · Cited by 152
Hashimoto, J., Chonan, K., Aoki, Y., Nishimura, T., Ohkubo, T., Hozawa, A., Suzuki, M., Matsubara, M., Michimata, M., Araki, T., et al. (2002) · Journal of Hypertension · doi:10.1097/00004872-200212000-00021 · Cited by 122 · cited 5×
“Hashimoto and coworkers [16] write that the weak correlations with PWV are not due to poor reproducibility of the PWV and DPA parameters, but occur because the parameters reflect different physiological properties at central and peripheral sites in the arterial vascular system.”
InfluentialImanaga, I., Hara, H., KOYANAGI, S., & TANAKA, K. (1998) · Japanese Heart Journal · doi:10.1536/ihj.39.775 · Cited by 118 · cited 1×
“Studies have demonstrated that the APG parameters are closely associated with aging and risk factors for arteriosclerotic vascular disease [13 – 15], but it is still unclear which one is clinically most informative [11].”
Photoplethysmography and nociception
Korhonen, I., & Yli‐Hankala, A. (2009) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.2009.02026.x · Cited by 105 · cited 2×
“A correct identification of the dicrotic notch is therefore essential to determine the systolic time [27], but due to variations in HR, cardiac output and vascular tone, with drifting in time positioning of the dicrotic notch with distance from the aortic valve, and to interference with the reflected wave, its applica…”
Otsuka, T., Kawada, T., Katsumata, M., & Ibuki, C. (2006) · Circulation Journal · doi:10.1253/circj.70.304 · Cited by 103 · cited 5×
“The d / a ratio correlates negatively to the Framingham score [17].”
InfluentialSalvi, V., Karnad, D. R., Panicker, G. K., & Kothari, S. (2009) · British Journal of Pharmacology · doi:10.1111/j.1476-5381.2009.00427.x · Cited by 102 · cited 2×
“The Bazett equation has been criticized for over-correcting at high HRs and under-correcting at low HRs [51].”
Detection of a and b waves in the acceleration photoplethysmogram
Elgendi, M., Norton, I., Brearley, M. B., Abbott, D., & Schuurmans, D. (2014) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-13-139 · Cited by 100
Patvardhan, E. A., Heffernan, K. S., Ruan, J., Hession, M. J., Warner, P. J., Karas, R. H., & Kuvin, J. T. (2011) · Cardiology Research and Practice · doi:10.4061/2011/253758 · Cited by 80 · cited 1×
“AIX is influenced by both large artery stiffness and properties in the PW-reflecting peripheral arterial tree [5–7] and is associated with cardiovascular risk, although it is less established than the PWV as a risk factor for cardiovascular morbidity [8–10].”
Detection of c, d, and e waves in the acceleration photoplethysmogram
Elgendi, M. (2014) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2014.08.001 · Cited by 73
Second Derivative of Photoplethysmography for Estimating Vascular Aging
Baek, H. J., Kim, J. S., Kim, Y. S., Lee, H. B., & Park, K. S. (2007) · IEEE International Conference on Information Technology and Applications in Biomedicine · doi:10.1109/itab.2007.4407346 · Cited by 69
Janner, J. H., Godtfredsen, N. S., Ladelund, S., Vestbo, J., & Prescott, E. (2011) · Journal of Human Hypertension · doi:10.1038/jhh.2011.59 · Cited by 62 · cited 1×
“Although pressure PW analysis shows benefits in predicting cardiovascular morbidity [1,10], the impact in clinical practice is small because the tonometry method is cumbersome with a need for highly specialized operators.”
Takada, H., Washino, K., Harrell, J. S., & Iwata, H. (1996) · PubMed · Cited by 56
The finger volume pulse and assessment of arterial properties
Avolio, A. P. (2002) · Journal of Hypertension · doi:10.1097/00004872-200212000-00007 · Cited by 45
Standard Terminologies for Photoplethysmogram Signals
Elgendi, M. (2012) · Current Cardiology Reviews · doi:10.2174/157340312803217184 · Cited by 43
Trends in Citations to Books on Epidemiological and Statistical Methods in the Biomedical Literature
Porta, M. S., Vandenbroucke, J. P., Ioannidis, J. P. A., Sanz, S., Fernández, E., Bhopal, R. S., Morabia, A., Victora, C. G., & López‐Jiménez, T. (2013) · PLoS ONE · doi:10.1371/journal.pone.0061837 · Cited by 38
Measurement of the left ventricular ejection time by digital plethysmography
Chirife, R., Pigott, V. M., & Spodick, D. H. (1971) · American Heart Journal · doi:10.1016/0002-8703(71)90269-9 · Cited by 26 · cited 1×
“[49] Digital Photoplethysmography for Assessment of Arterial Stiffness”
Jang, D.-G., Park, S.-H., & Hahn, M. (2014) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2014.2306679 · Cited by 23 · cited 1×
“On the other hand, differences in age and gender have impacts on the accuracy of the pulse contour analyses [40], indicating that in a heterogeneous population like ours the repeatability of measurements differs from a homogenous population [34,41].”
BackgroundApplication and Interpretation
Stuart D. Summerhayes BSc, M. (2008) · doi:10.1002/9781444302554.ch2 · Cited by 20
Dicrotic notch detection using wavelet transform analysis
Antonelli, L. T., Ohley, W. J., & Khamlach, R. (2002) · Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.1994.415400 · Cited by 20
A morphological approach to calculation of the second derivative of photoplethysmography
Jang, D.-G., Park, J.-H., Park, S.-H., & Hahn, M. (2010) · International Conference on Signal Processing · doi:10.1109/icosp.2010.5656472 · Cited by 8 · cited 1×
“[44].”
Ishikawa, H., Honjo, A., Kagoshima, T., Doi, M., Nonaka, H., & Saishin, M. (1975) · Japanese Circulation Journal-english Edition · doi:10.1253/jcj.39.133 · Cited by 4 · cited 1×
“[50] found, in contrast to Jang et al.”
ResultFeature Extraction from Finger Pulse Profile for High Pitta Detection
Sharma, T., & Singh, M. (2012) · Cited by 2
A comparison of correction algorithms to pulse parameters for estimation of arterial stiffness
Jang, D.-G., Park, S.-H., & Hahn, M. (2010) · International Conference on Signal Processing · doi:10.1109/icosp.2010.5656481 · Cited by 1 · cited 1×
“[39], linear regression equations can be used for adjustments of timerelated PPG parameters.”
Method
Flagged first, then by citations
2 of the 53 references couldn’t be matched to a known paper.
Research works printed but not linked: 3
- R Gonzales, 2008, A computer base photoplethysmographic vascular analyser through derivatives, Comp Cardiol, 35, 177
- T Ozawa, 1978, Pattern of second derivative of volume pulse wave, the relation between non-invasive index of ventricular function and peak acceleration and effect of preloading to peak velocity, Myakuha (Pulse Wave), 8, 22
- HC Bazett, 1920, An analysis of the time-relations of electrocardiograms, Heart, 7, 35
Cited by
Sorted by citations
Self‐Powered Pulse Sensor for Antidiastole of Cardiovascular Disease
Ouyang, H., Tian, J., Sun, G., Zou, Y., Liu, Z., Li, H., Zhao, L., Shi, B., Fan, Y., Fan, Y., et al. (2017) · Advanced Materials · doi:10.1002/adma.201703456 · Cited by 470
Photoplethysmogram Analysis and Applications: An Integrative Review
Park, J., Seok, H. S., Kim, S.-S., & Shin, H. (2022) · Frontiers in Physiology · doi:10.3389/fphys.2021.808451 · Cited by 514
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348
Wearable Photoplethysmography for Cardiovascular Monitoring
Charlton, P. H., Kyriacou, P. A., Mant, J., Marozas, V., Chowienczyk, P., & Alastruey, J. (2022) · Proceedings of the IEEE · doi:10.1109/jproc.2022.3149785 · Cited by 250
An optimal filter for short photoplethysmogram signals
Liang, Y., Elgendi, M., Chen, Z., & Ward, R. (2018) · Scientific Data · doi:10.1038/sdata.2018.76 · Cited by 242
Metadata
- Authors
- von Wowern, E., Östling, G., Nilsson, P. M., & Olofsson, P.
- Affiliations
- Lund University (SE)
- Venue
- PLOS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2015-08-20
- Volume
- 10
- Issue
- 8
- Pages
- e0135659
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC4546304
- OpenAlex
- W1784071036
- Cited by
- 134
- References
- 53
- Citations by year
- 2026: 3 · 2025: 8 · 2024: 9 · 2023: 8 · 2022: 39 · 2021: 19 · 2020: 12 · 2019: 7 · 2018: 13 · 2017: 9 · 2016: 7
- FWCI
- 7.03
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- PLoS ONE (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- RePEc: Research Papers in Economics (submittedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Lund University Publications (Lund University) (submittedVersion, other-oa)
- Funders
- Lunds Universitet
- Region Skåne
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Keywords
- arterial stiffness, digital photoplethysmography, applanation tonometry, pulse wave analysis, pulse wave velocity, augmentation index, repeatability, aging index, left ventricular ejection time, pregnancy, cardiovascular risk
- MeSH terms
- Adult, Aged, Aged, 80 and over, Aorta, Arteries, Blood Pressure, Elasticity, Female, Heart Rate, Humans, Male, Manometry, Middle Aged, Pregnancy, Pulse, Risk Factors, Photoplethysmography, Young Adult, Vascular Stiffness, Pulse Wave Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review