Cooperation experience in a multicentre study to define the upper limits in a normal population for the diagnostic assessment of the functional lupus anticoagulant assays

Pradella, Paola, Azzarini, Gabriella, Santarossa, Liliana, Caberlotto, Livio, Bardin, Cristina, Poz, Alessandra, D’Aurizio, Federica, Giacomello, Roberta·2013·Clinical Chemistry and Laboratory Medicine (CCLM)

Abstract

BACKGROUND: Phospholipid-dependent coagulation tests for lupus anticoagulant (LA) are considered an important step for the diagnosis of anti-phospholipid syndrome; however, LA laboratory detection is difficult because of many variables. Five hospital laboratories, located in a North-Italy area and using the same method for LA testing, cooperated to standardise sample treatment and analytical procedure in order to define the upper values for LA negativity. METHODS: In total, 200 normal subjects (40 for each centre) were studied for six LA functional assays, using the same procedure, reagent lot and analyser type. The first tests done were LA screen and LA confirm assays, based on diluted Russell's Viper Venom Time, with low and high phospholipid content, respectively. The second tests performed were silica clotting time screen and confirm assays, based on activated partial thromboplastin time, with low and high phospholipid content, respectively. Finally, two mixing assays were executed for both screening assays, diluting patient sample with a pool prepared with plasma collected from the study population. RESULTS: Data analysis demonstrated a difference between centres for all assays when results were expressed in seconds; the difference disappeared when results are normalised with the local mean normal value of each centre and are expressed as a normalised ratio. The study population was normally distributed; so the value corresponding to 99th percentile was used as limit value for LA negativity. Values expressed as normalised ratio, for LA and silica clotting time screenings were 1.22 and 1.23, respectively. CONCLUSIONS: The study allowed us to define a uniform approach to LA testing and evaluation for laboratories employing the same methods.

Integrity signals

Registry

Crossref · 10.1515/cclm-2012-0382

Cites

36 of 36 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Pradella, P., Azzarini, G., Santarossa, L., Caberlotto, L., Bardin, C., Poz, A., D’Aurizio, F., & Giacomello, R.
Affiliations
  • Ospedale di Cattinara (IT)
  • Ospedale dell' Angelo (IT)
  • Azienda Ospedaliera Santa Maria Degli Angeli (IT)
  • Ca' Foncello Hospital (IT)
  • University of Udine (IT)
Corresponding author
Pradella, Paola
Venue
Clinical Chemistry and Laboratory Medicine (CCLM)
Published
2012-09-22
Volume
51
Issue
2
Pages
379-385
ISSN
  • 1434-6621 (ISSN-L)
  • 1437-4331
OpenAlex
W1994162452
Cited by
26
References
36
Citations by year
2026: 1 · 2023: 2 · 2022: 3 · 2021: 2 · 2020: 2 · 2019: 1 · 2018: 1 · 2017: 2 · 2016: 2 · 2015: 3 · 2014: 3 · 2013: 3 · 2012: 1
FWCI
2.45
Type
journal article
Language
en
Open access
closed
Locations
  • Clinical Chemistry and Laboratory Medicine (CCLM) (publishedVersion)
  • PubMed (publishedVersion)
  • Institutional Research Information System (University of Udine) (submittedVersion)
Topics
  • Health Sciences › Medicine › Rheumatology › Systemic Lupus Erythematosus Research
  • Health Sciences › Medicine › Hematology › Blood Coagulation and Thrombosis Mechanisms
  • Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
Keywords
lupus anticoagulant, antiphospholipid syndrome, reference limits, healthy population, multicentre collaboration, Russell's viper venom time, activated partial thromboplastin time
MeSH terms
Adolescent, Adult, Aged, Blood Coagulation Tests, Female, Humans, Male, Middle Aged, Reference Values, Multicenter Studies as Topic, Lupus Coagulation Inhibitor, Antiphospholipid Syndrome, Young Adult

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review