Integrated insulin pump therapy with continuous glucose monitoring for improved adherence: technology update
Open accessAbstract
Insulin pump therapy combined with real-time continuous glucose monitoring, known as sensor-augmented pump (SAP) therapy, has been shown to improve metabolic control and to reduce the rate of hypoglycemia in adults with type 1 diabetes compared to multiple daily injections or standard continuous subcutaneous insulin infusion. Glycemic variability is also reduced in patients on SAP therapy. This approach allows patients to monitor their glucose levels being informed of glycemic concentration and trend. Trained diabetic patients, therefore, can appropriately modify insulin infusion and/or carbohydrate intake in order to prevent hypo- or hyperglycemia. For these reasons, SAP therapy is now considered the gold standard for type 1 diabetes treatment. To be clinically effective, however, devices and techniques using advanced technology should not only have the potential to theoretically ameliorate metabolic control, but also be well accepted by patients in terms of satisfaction and health-related quality of life, because these factors will improve treatment adherence and consequently overall outcome. SAP therapy is generally well tolerated by patients; however, many clinical trials have identified significant noncompliance in the use of this device, most notably in the pediatric and adolescent populations. In this review we aim to analyze the main reasons for good or poor adherence to SAP therapy and to provide useful tips in order to fully benefit from this kind of novel therapeutic approach.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Sensor‐augmented pump systems (SAP) enable adaptive insulin delivery, including dual‐wave or square‐wave boluses, which distribute insulin over several hours and better match the slow and prolonged glucose absorption typical of DGP, particularly in high‐fat or high‐protein meals.
“Sensor‑augmented pump systems (SAP) enable adaptive insulin delivery, including dual‑wave or square‑wave boluses, which distribute insulin over several hours and better match the slow and prolonged glucose absorption typical of DGP [54, 112, 132, 148, 149], particularly in high‑fat or high‑protein meals [150–152].”
From Diabetic gastroparesis: pathophysiology and impact on insulin timing choices (2026) · cites this paper as [152]
Split into 2 atoms — 0 confirmed, 1 partially supported, 0 disputed, 0 misattributed, 1 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Continuous glucose monitoring and intensive treatment of type 1 diabetes
(2008) · New England Journal of Medicine · doi:10.1056/nejmoa0805017 · Cited by 1,559 · see citing works
Effectiveness of Sensor-Augmented Insulin-Pump Therapy in Type 1 Diabetes
Bergenstal, R. M., Tamborlane, W. V., Ahmann, A., Buse, J. B., Dailey, G., Davis, S. N., Joyce, C., Peoples, T., Perkins, B. A., Welsh, J. B., et al. (2010) · New England Journal of Medicine · doi:10.1056/nejmoa1002853 · Cited by 894 · cited 5× · see citing works
“In particular, it is necessary to explain to the patient that SAP therapy implies more rather than less attention for appropriate diabetes management.(7,18)”
InfluentialThreshold-based insulin-pump interruption for reduction of hypoglycemia
Bergenstal, R. M., Klonoff, D. C., Garg, S. K., Bode, B. W., Meredith, M., Slover, R. H., Ahmann, A. J., Welsh, J. B., Lee, S. W., & Kaufman, F. R. (2013) · New England Journal of Medicine · doi:10.1056/nejmoa1303576 · Cited by 652 · see citing works
Pickup, J. C., Freeman, S. C., & Sutton, A. J. (2011) · BMJ · doi:10.1136/bmj.d3805 · Cited by 541 · cited 3× · see citing works
“Previous studies, reporting that RT-CGM systems improve clinical outcomes only when they are used consistently, 2,7,9 confirm the critical role of patient acceptance for optimizing the benefits of all new treatments or technologies.”
Deiss, D., Bolinder, J., Riveline, J.-P., Battelino, T., Bosi, E., Tubiana-Rufi, N., Kerr, D., & Phillip, M. (2006) · Diabetes Care · doi:10.2337/dc06-1134 · Cited by 539 · see citing works
the SWITCH Study Group, Battelino, T., Conget, I., Olsen, B., Schütz-Fuhrmann, I., Hommel, E., Hoogma, R., Schierloh, U., Sulli, N., & Bolinder, J. (2012) · Diabetologia · doi:10.1007/s00125-012-2708-9 · Cited by 494 · cited 1× · see citing works
“We must specify that, even if technology could deeply affect the patients’ lifestyle, it would be useless if not supported by a robust and continuous educational approach as demonstrated by the major randomized controlled trials on SAP therapy.(2,6,7) Perhaps the most important characteristic for predicting the best r…”
The effect of continuous glucose monitoring in well-controlled type 1 diabetes
Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group (2009) · Diabetes Care · doi:10.2337/dc09-0108 · Cited by 399 · see citing works
Effectiveness of Continuous Glucose Monitoring in a Clinical Care Environment
Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group (2010) · Diabetes Care · doi:10.2337/dc09-1502 · Cited by 313 · see citing works
Sensor-Augmented Insulin Pump Therapy: Results of the First Randomized Treat-to-Target Study
Hirsch, I. B., Abelseth, J. M., Bode, B. W., Fischer, J. S., KAUFMAN, F. R., Mastrototaro, J. J., Parkin, C. G., Wolpert, H. A., & Buckingham, B. A. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0068 · Cited by 297 · cited 1× · see citing works
“Wearing the device at least 60%–70% of the total time predicted significant HbA 1c reduction in the STAR 1 study.(19) Furthermore, in the JDRF randomized controlled trial, the group with the greatest success (those over 25 years old) also had the greatest frequency of sensor use: 83% of the patients wore the device at…”
Raccah, D., Sulmont, V., Reznik, Y., Guerci, B., Renard, É., Hanaire, H. P., Jeandidier, N., & Nicolino, M. P. (2009) · Diabetes Care · doi:10.2337/dc09-0750 · Cited by 236 · see citing works
Can Interstitial Glucose Assessment Replace Blood Glucose Measurements?
Rebrin, K., & Steil, G. M. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050194332 · Cited by 220 · cited 2× · see citing works
“The delay (3–12 minutes) is due to the time required for glucose diffusion across the capillary endothelial barrier.(22,23) Each patient initiating SAP therapy should be aware of these dynamics because they directly influence the interpretation of RT-CGM.”
Continuous glucose monitoring: an Endocrine Society Clinical Practice Guideline
Klonoff, D. C., Buckingham, B., Christiansen, J. S., Montori, V. M., Tamborlane, W. V., Vigersky, R. A., & Wolpert, H. (2011) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2010-2756 · Cited by 211 · see citing works
O’Connell, M. A., Donath, S. M., O'Neal, D. N., Colman, P. G., Ambler, G. R., Jones, T. W., Davis, E. A., & Cameron, F. J. (2009) · Diabetologia · doi:10.1007/s00125-009-1365-0 · Cited by 210 · see citing works
Mauras, N., Beck, R., Xing, D., Ruedy, K., Buckingham, B., Tansey, M., White, N. H., Weinzimer, S. A., Tamborlane, W., Kollman, C., et al. (2012) · Diabetes Care · doi:10.2337/dc11-1746 · Cited by 204 · see citing works
Hermanides, J., Nørgaard, K., Bruttomesso, D., Mathieu, C., Frid, A. H., Dayan, C. M., Diem, P., Fermon, C., Wentholt, I. M. E., Hoekstra, J. B. L., et al. (2011) · Diabetic Medicine · doi:10.1111/j.1464-5491.2011.03256.x · Cited by 169 · see citing works
Systematic literature review: quality of life associated with insulin pump use in Type 1 diabetes
Barnard, K. D., Lloyd, C. E., & Skinner, T. C. (2007) · Diabetic Medicine · doi:10.1111/j.1464-5491.2007.02120.x · Cited by 136 · see citing works
Hirsch, I. B., Armstrong, D. U., Bergenstal, R. M., Buckingham, B. A., Childs, B. P., Clarke, W. L., Peters, A. L., & Wolpert, H. A. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0016 · Cited by 136 · cited 2× · see citing works
“This phased approach allows the user to become comfortable with the other aspects of the system maintenance, avoiding frequent high and low glucose alerts.(23,30)”
Satisfaction with continuous glucose monitoring in adults and youths with Type 1 diabetes
Tansey, M., Laffel, L. M., Cheng, J., Beck, R. W., Coffey, J., Huang, E. S., Kollman, C., Lawrence, J. M., Lee, J., Ruedy, K. J., et al. (2011) · Diabetic Medicine · doi:10.1111/j.1464-5491.2011.03368.x · Cited by 124 · cited 1× · see citing works
“When bleeding occurs, it is better to blot the blood with a cotton swab before connecting the transmitter.(27,28)”
Use of continuous glucose monitoring in children and adolescents *
Phillip, M., Danne, T., Shalitin, S., Buckingham, B. A., Laffel, L. M., Tamborlane, W. V., Battelino, T., & Participants, F. T. C. F. (2012) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2011.00849.x · Cited by 111 · see citing works
Routine Sensor-Augmented Pump Therapy in Type 1 Diabetes: The INTERPRET Study
Nørgaard, K., Scaramuzza, A. E., Bratina, N., Lalic, N. M., Jarosz-Chobot, P. A., Kocsis, G., Jašinskienė, E., Block, C. E. M. D., Carrette, O., Castañeda, J., et al. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0288 · Cited by 107 · cited 1× · see citing works
“However, surveys from patients on SAP therapy showed that the most common perceived benefits are prevention of hypoglycemia and decrease in hypoglycemia-related anxiety.(13,15)”
Tumminia, A., Crimi, S., Sciacca, L., Buscema, M., Frittitta, L., Squatrito, S., Vigneri, R., & Tomaselli, L. (2015) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.2557 · Cited by 100 · cited 3× · see citing works
“Patients under SAP therapy are required to have a good understanding of the basic principles of insulin therapy for T1D.”
Self-citationBlevins, T. C., Bode, B. W., Garg, S. K., Grunberger, G., Hirsch, I. B., Jovanovič, L., Nardacci, E., Orzeck, E. A., Roberts, V. L., & Tamborlane, W. V. (2010) · Endocrine Practice · doi:10.4158/ep.16.5.730 · Cited by 98 · see citing works
Realistic Expectations and Practical Use of Continuous Glucose Monitoring for the Endocrinologist
Hirsch, I. B. (2009) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2008-2625 · Cited by 93 · cited 1× · see citing works
“One way is to use the rate of glucose change to alter the premeal bolus, with a 1–2 mg/dL/min change indicating the need for a 10% adjustment in bolus dosing (a single arrow up or down), while adjusting the bolus dose by 20% when the CGM system alerts that glucose levels are more rapidly changing (more than 2 mg/dL/mi…”
Danne, T., Tsioli, C., Kordonouri, O., Bläsig, S., Remus, K., Roy, A., Keenan, D. B., Lee, S. W., & KAUFMAN, F. R. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0327 · Cited by 91 · cited 1× · see citing works
“In the PILGRIM study, testing the PLGM algorithm, impending hypoglycemia was prevented 80% of the time, with the average suspension of insulin occurring at 92 mg/dL.(32) These novel features can play a key role in improving adherence to SAP therapy, especially in those patients experiencing frequent hypoglycemic episo…”
Cemeroğlu, A. P., Stone, R., Kleis, L., Racine, M. S., Postellon, D. C., & Wood, M. A. (2009) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2009.00549.x · Cited by 67 · cited 1× · see citing works
“However, surveys from patients on SAP therapy showed that the most common perceived benefits are prevention of hypoglycemia and decrease in hypoglycemia-related anxiety.(13,15)”
Wentholt, I. M. E., Hart, A. A. M., Hoekstra, J. B. L., & DeVries, J. H. (2007) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.0007 · Cited by 54 · cited 1× · see citing works
“The delay (3–12 minutes) is due to the time required for glucose diffusion across the capillary endothelial barrier.(22,23) Each patient initiating SAP therapy should be aware of these dynamics because they directly influence the interpretation of RT-CGM.”
Blood Glucose Monitoring Technology: Translating Data Into Practice
Hirsch, I. B. (2004) · Endocrine Practice · doi:10.4158/ep.10.1.67 · Cited by 47 · cited 2× · see citing works
“Patients under SAP therapy are required to have a good understanding of the basic principles of insulin therapy for T1D.”
Psychological aspects of continuous glucose monitoring in pediatric type 1 diabetes
The Diabetes Research in Children Network (DirecNet) Study Group (2006) · Pediatric Diabetes · doi:10.1111/j.1399-543x.2006.00142.x · Cited by 46 · cited 1× · see citing works
“26 On the basis of these findings, we could speculate that patients who use CGM less frequently may feel annoyed with multiple CGM sensor insertions and the alarms.”
The DAWN Youth WebTalk Study: methods, findings, and implications
Peyrot, M. F., & on behalf of the International DAWN Youth Survey Group (2009) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2009.00612.x · Cited by 46 · see citing works
Continuous Glucose Monitoring and Intensive Treatment of Type 1 Diabetes
Ovalle, F. (2009) · Yearbook of Endocrinology · doi:10.1016/s0084-3741(09)79371-0 · Cited by 45 · see citing works
Raccah, D., Sulmont, V., Reznik, Y., Guerci, B., Renard, É., Hanaire, H. P., Jeandidier, N., & Nicolino, M. P. (2010) · Avances en Diabetología · doi:10.1016/s1134-3230(10)61016-2 · Cited by 36 · see citing works
Frontino, G., Bonfanti, R., Scaramuzza, A. E., Rabbone, I., Meschi, F., Rigamonti, A., Battaglino, R., Favalli, V., Bonura, C., Sicignano, S., et al. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0044 · Cited by 29 · cited 1× · see citing works
“Using these techniques, many very young children can be taught to tolerate CGM well and, in those already on an insulin pump, SAP therapy has been shown to be effective.(29)”
Practical Aspects of Real-Time Continuous Glucose Monitors
Ives, B., Sikes, K. A., Urban, A. D., Stephenson, K., & Tamborlane, W. V. (2009) · The Science of Diabetes Self-Management and Care · doi:10.1177/0145721709352010 · Cited by 24 · cited 3× · see citing works
“facturer’s instructions for frequency and timing of blood glucose calibrations.(28) • Tip 1: quality is more important than quantity for calibration.”
Continuous Glucose Monitoring and Type 1 Diabetes
Montagnana, M., Lippi, G., & Guidi, G. C. (2009) · New England Journal of Medicine · doi:10.1056/nejmc082243 · Cited by 17 · see citing works
The effect of continuous glucose monitoring in well-controlled type 1 diabetes
Vázquez, F. (2010) · Avances en Diabetología · doi:10.1016/s1134-3230(10)64018-5 · Cited by 5 · see citing works
de Solís, M. S. N. (2010) · Avances en Diabetología · doi:10.1016/s1134-3230(10)62014-5 · Cited by 3 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Rodbard, H. W., Peters, A. L., Slee, A., Cao, A., Traina, S. B., & Alba, M. (2017) · Diabetes Care · doi:10.2337/dc16-1353 · Cited by 94 · see citing works
Automated blood glucose control in type 1 diabetes: A review of progress and challenges
Bertachi, A. H., Ramkissoon, C. M., Bondía, J., & Vehı́, J. (2017) · Endocrinología Diabetes y Nutrición · doi:10.1016/j.endinu.2017.10.011 · Cited by 39 · see citing works
Devices for continuous monitoring of glucose: update in technology
Gómez, A. M., Carrillo, D. C. H., & Múñoz, Ó. (2017) · Medical Devices Evidence and Research · doi:10.2147/mder.s110121 · Cited by 29 · see citing works
“Also, higher QoL scores in different populations including pediatric patients were associated with more frequent CGM use, suggesting that compliant patients may perceive more benefits from the use of SAP therapy.(54) In a study conducted in Colombia, using the CORE Diabetes Model, an increase in the average life expec…”
Maternal Diabetes Impairs Insulin and IGF-1 Receptor Expression and Signaling in Human Placenta
Tumminia, A., Scalisi, N. M., Milluzzo, A., Ettore, G., Vigneri, R., & Sciacca, L. (2021) · Frontiers in Endocrinology · doi:10.3389/fendo.2021.621680 · Cited by 26 · see citing works
Self-citationTumminia, A., Milluzzo, A., Festa, C., Fresa, R., Pintaudi, B., Scavini, M., Vitacolonna, E., Napoli, A., & Sciacca, L. (2021) · Nutrition Metabolism and Cardiovascular Diseases · doi:10.1016/j.numecd.2021.03.013 · Cited by 25 · see citing works
Self-citationBombaci, B., Passanisi, S., Alibrandi, A., D’Arrigo, G., Patroniti, S., Averna, S., Salzano, G., & Lombardo, F. (2022) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph191610293 · Cited by 23 · see citing works
“The introduction of sensoraugmented pump (SAP) therapy allowed to integrate glucose monitoring and insulin therapy through the use of decision-making algorithms that adjust insulin delivery in relation to sensor glycemic levels in real-time [9].”
BackgroundAutomated blood glucose control in type 1 diabetes: A review of progress and challenges
Bertachi, A. H., Ramkissoon, C. M., Bondía, J., & Vehı́, J. (2018) · Endocrinología Diabetes y Nutrición (English ed ) · doi:10.1016/j.endien.2018.03.001 · Cited by 22 · see citing works
Múñoz, Ó., Guyatt, G. H., Devji, T., Zhang, Y., Li, S., Alexander, P. E., Henao, D. C., Gomez, A.-M., & Ruiz, Á. J. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0346 · Cited by 14 · see citing works
Physical Activity and Nutrition Interventions for Type 1 Diabetes: A Meta-analysis
Shorey, S., Ng, E. D., Law, E. C. N., Wong, J. C. M., Loke, K. Y., & Tam, W. (2022) · PEDIATRICS · doi:10.1542/peds.2022-056540 · Cited by 10 · see citing works
TELLEZ, S. E., Hornung, L. N., Abu‐El‐Haija, M. A., & Elder, D. A. (2023) · Journal of Clinical Medicine · doi:10.3390/jcm12093319 · Cited by 9 · see citing works
Tumminia, A., Santoro, G. M., Oteri, V., Galeano, F., Baratta, R., Gullo, D., Frittitta, L., Sciacca, L., Moli, R. L., Piticchio, T., et al. (2025) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2025.112173 · Cited by 9 · see citing works
Self-citationKiconco, R., Lumumba, S. A., Bagenda, C. N., Atwine, R., Ndarubweine, J., & Rugera, S. P. (2024) · Therapeutic Advances in Endocrinology and Metabolism · doi:10.1177/20420188241232280 · Cited by 6 · see citing works
“96 Similarly, a study of patients with type 2 diabetes found that 44% of patients receiving basal insulin therapy and 52% of patients receiving basal-bolus insulin therapy experienced hyperglycemia, which is defined as blood glucose levels above 130 mg/dL or 7.2 mmol/L. 97 It is important for…”
BackgroundInfluentialNyholm, D., Adnan, M., & Senek, M. (2020) · Journal of Parkinson s Disease · doi:10.3233/jpd-202114 · Cited by 6 · see citing works
“Developmentofadher-encemonitoringhassofarbeenslow,comparedto forexampleinsulintreatmentindiabetes[22].”
NUEVAS TECNOLOGÿAS EN DIABETES
Apablaza, P., Néstor Soto, I., Román, R., & Ethel Codner, D. (2016) · Revista Médica Clínica Las Condes · doi:10.1016/j.rmclc.2016.04.011 · Cited by 6 · see citing works
Shorey, S., Ng, E. D., Law, E. C., Wong, J. C., Loke, K. Y., & Tam, W. (2021) · Cochrane Database of Systematic Reviews · doi:10.1002/14651858.cd013858 · Cited by 2 · see citing works
RetractedZorena, K., Kowalewska, B., Szmigiero-Kawko, M., Wąż, P., & Myśliwiec, M. (2016) · The Italian Journal of Pediatrics/Italian journal of pediatrics · doi:10.1186/s13052-016-0317-9 · Cited by 1 · see citing works
“Blood glucose levels were measured by continous glicemic monitoring systems (CGMS) using an electrode compatible with “Guardian” device or Medtronic insulin pump (Medtronic Minimed, Northridge, CA).”
BackgroundDiabetic gastroparesis: pathophysiology and impact on insulin timing choices
Galeano, F., Tumminia, A., Oteri, V., Le Moli, R., Piticchio, T., Scilletta, S., Di Marco, M., Di Pino, A., & Frasca, F. (2026) · Endocrine · doi:10.1007/s12020-026-04693-6
“…pump systems (SAP) enable adaptive insulin delivery, including dual-wave or square-wave boluses, which distribute insulin over several hours and better match the slow and prolonged glucose absorption typical of DGP [54, 112, 132, 148, 149], particularly in high-fat or high-protein meals [150–152].”
Self-citationLa gastroparesi diabetica: fisiopatologia e impatto sulla scelta del timing insulinico
Galeano, F., Frasca, F., Oteri, V., Moli, R. L., & Tumminia, A. (2026) · L Endocrinologo · doi:10.1007/s40619-026-01773-7
Self-citationAssociation of resilience with hemoglobin A1c among youth and young adults with youth-onset diabetes
Julceus, E., Mendoza, J., Flory, K., Frongillo, E., Merchant, A., Malik, F. S., Cooper, D. K., Reboussin, B., Sauder, K. A., Bellatorre, A., et al. (2026) · Annals of Epidemiology · doi:10.1016/j.annepidem.2026.110125
Tumminia, A., Papa, G., Cannarella, R., Galeano, F., Oteri, V., Amato, V., Baratta, R., Condorelli, R. A., Palazzolo, R., Gusmano, C., et al. (2026) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2026.113613
Self-citation
- Authors
- Tumminia, A., Sciacca, L., Frittitta, L., Squatrito, S., Vigneri, R., Le Moli, R., & Tomaselli, L.
- Affiliations
- University of Catania (IT)
- Ospedale Garibaldi (IT)
- Corresponding author
- Tumminia, Andrea
- Venue
- Patient Preference and Adherence
- Publisher
- Informa UK Limited
- Published
- 2015-09
- Volume
- 9
- Pages
- 1263
- ISSN
- 1177-889X (ISSN-L)
- PMC
- PMC4567238
- OpenAlex
- W1440504177
- Cited by
- 20
- References
- 36
- Citations by year
- 2026: 4 · 2025: 1 · 2024: 1 · 2023: 1 · 2022: 2 · 2021: 3 · 2020: 1 · 2019: 1 · 2018: 1 · 2017: 2 · 2016: 3
- FWCI
- 0.30
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Patient Preference and Adherence (publishedVersion, cc-by-nc)
- PubMed Central (publishedVersion, cc-by-nc)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Dove Medical Press (Taylor and Francis Group) (submittedVersion, cc-by-nc)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Medicine › Family Practice › Medication Adherence and Compliance
- Keywords
- sensor-augmented pump therapy, insulin pump therapy, continuous glucose monitoring, type 1 diabetes, medication adherence, hypoglycemia, glycemic variability, glycemic control, patient satisfaction, health-related quality of life, device acceptance
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review