Clinical Implications of Real-time and Intermittently Scanned Continuous Glucose Monitoring
Abstract
Two types of continuous glucose monitoring (CGM) systems are now available: real-time CGM (rtCGM) and intermittently scanned (isCGM). Current rtCGM systems automatically transmit a continuous stream of glucose data to the user, provide alerts and active alarms, and transmit glucose data (trend and numerical) in real time to a receiver, smart watch, or smartphone. The current isCGM system provides the same type of glucose data but requires the user to purposely scan the sensor to obtain information, and it does not have alerts and alarms. Both CGM technologies have significant advantages over self-monitoring of blood glucose; however, differences in the features and capabilities of the two approaches must be considered when guiding patient selection of the system that meets their individual needs.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
International consensus on use of continuous glucose monitoring
Danne, T., Nimri, R., Battelino, T., Bergenstal, R. M., Close, K. L., DeVries, J. H., Garg, S., Heinemann, L., Hirsch, I., Amiel, S. A., et al. (2017) · Diabetes Care · doi:10.2337/dc17-1600 · Cited by 2,069
Self-citationBeck, R. W., Riddlesworth, T., Ruedy, K., Ahmann, A., Bergenstal, R., Haller, S., Kollman, C., Kruger, D., McGill, J. B., Polonsky, W., et al. (2017) · JAMA · doi:10.1001/jama.2016.19975 · Cited by 1,185 · cited 1×
“For example, the large, randomized Multiple Daily Injections and Continuous Glucose Monitoring in Diabetes (DIAMOND) trial, which included T1D and T2D participants, showed that use of rtCGM improved HbA1c, reduced time spent in the hypoglycemic (T1D only) and hyperglycemic ranges, and demonstrated reductions in modera…”
BackgroundBolinder, J., Antuna, R., Geelhoed-Duijvestijn, P., Kröger, J., & Weitgasser, R. (2016) · The Lancet · doi:10.1016/s0140-6736(16)31535-5 · Cited by 893
Lind, M., Polonsky, W., Hirsch, I. B., Heise, T., Bolinder, J., Dahlqvist, S., Schwarz, E., Ólafsdóttir, A. F., Frid, A., Wedel, H., et al. (2017) · JAMA · doi:10.1001/jama.2016.19976 · Cited by 705
Self-citationSafety of a hybrid closed-loop insulin delivery system in patients with type 1 diabetes
Bergenstal, R. M., Garg, S., Weinzimer, S. A., Buckingham, B. A., Bode, B. W., Tamborlane, W. V., & Kaufman, F. R. (2016) · JAMA · doi:10.1001/jama.2016.11708 · Cited by 644 · cited 3×
“Recent studies have also demonstrated the benefits of rtCGM as an integrated component of sensor-augmented insulin pumps with predictive low glucose suspend functionality (54,55).”
BackgroundBeck, R. W., Riddlesworth, T. D., Ruedy, K., Ahmann, A., Haller, S., Kruger, D., McGill, J. B., Polonsky, W., Price, D., Aronoff, S., et al. (2017) · Annals of Internal Medicine · doi:10.7326/m16-2855 · Cited by 635
Goldstein, D. E., Little, R. R., Lorenz, R. A., Malone, J. I., Nathan, D., Peterson, C. M., & Sacks, D. B. (2004) · Diabetes Care · doi:10.2337/diacare.27.7.1761 · Cited by 625
Haak, T., Hanaire, H., Ajjan, R., Hermanns, N., Riveline, J.-P., & Rayman, G. (2017) · Diabetes Therapy · doi:10.1007/s13300-016-0223-6 · Cited by 623 · cited 12×
“However, it also has inherent limitations that can impact its utility, especially in insulin-treated individuals with T1D and type 2 diabetes (T2D).”
BackgroundMethodResultInfluentialCavalot, F., Petrelli, A., Traversa, M., Bonomo, K., Fiora, E., Conti, M., Anfossi, G., Costa, G., & Trovati, M. (2006) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2005-1005 · Cited by 569
Pickup, J. C., Freeman, S. C., & Sutton, A. J. (2011) · BMJ · doi:10.1136/bmj.d3805 · Cited by 541 · cited 1×
“Many of the earlier studies focused primarily on use of rtCGM in conjunction with insulin pump therapy (53); however, more recent studies have investigated the potential benefits of rtCGM use in conjunction with multiple daily injections (MDI) therapy (9,10,45,47–50).”
Background
Cited by
Electrochemical glucose sensors in diabetes management: an updated review (2010–2020)
Teymourian, H., Barfidokht, A., & Wang, J. (2020) · Chemical Society Reviews · doi:10.1039/d0cs00304b · Cited by 881
Wearable Electrochemical Glucose Sensors in Diabetes Management: A Comprehensive Review
Saha, T., Del Caño, R., Mahato, K., De la Paz, E., Chen, C., Ding, S., Yin, L., & Wang, J. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00078 · Cited by 442
Maiorino, M. I., Signoriello, S., Maio, A., Chiodini, P., Bellastella, G., Scappaticcio, L., Longo, M., Giugliano, D., & Esposito, K. (2020) · Diabetes Care · doi:10.2337/dc19-1459 · Cited by 364
Grunberger, G., Sherr, J., Allende, M., Blevins, T., Bode, B., Handelsman, Y., Hellman, R., Lajara, R., Roberts, V. L., Rodbard, D., et al. (2021) · Endocrine Practice · doi:10.1016/j.eprac.2021.04.008 · Cited by 288
Lee, I., Probst, D., Klonoff, D., & Sode, K. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113054 · Cited by 288
Nimri, R., Battelino, T., Laffel, L. M., Slover, R. H., Schatz, D., Weinzimer, S. A., Dovc, K., Danne, T., Phillip, M., NextDREAM Consortium, et al. (2020) · Nature Medicine · doi:10.1038/s41591-020-1045-7 · Cited by 242
Yaron, M., Roitman, E., Aharon-Hananel, G., Landau, Z., Ganz, T., Yanuv, I., Rozenberg, A., Karp, M., Ish-Shalom, M., Singer, J., et al. (2019) · Diabetes Care · doi:10.2337/dc18-0166 · Cited by 211
Continuous glucose monitoring in adults with type 2 diabetes: a systematic review and meta-analysis
Jancev, M., Vissers, T. A. C. M., Visseren, F. L. J., van Bon, A. C., Serné, E. H., DeVries, J. H., de Valk, H. W., & van Sloten, T. T. (2024) · Diabetologia · doi:10.1007/s00125-024-06107-6 · Cited by 191
“CGM consists of a subcutaneous sensor that monitors interstitial glucose levels, which approximates blood glucose levels [3].”
BackgroundTeo, E., Hassan, N., Tam, W., & Koh, S. (2022) · Diabetologia · doi:10.1007/s00125-021-05648-4 · Cited by 186
Visser, M. M., Charleer, S., Fieuws, S., De Block, C., Hilbrands, R., Van Huffel, L., Maes, T., Vanhaverbeke, G., Dirinck, E., Myngheer, N., et al. (2021) · The Lancet · doi:10.1016/s0140-6736(21)00789-3 · Cited by 165
Seidu, S., Kunutsor, S. K., Ajjan, R. A., & Choudhary, P. (2024) · Diabetes Care · doi:10.2337/dc23-1520 · Cited by 124
Continuous glucose monitoring for the routine care of type 2 diabetes mellitus
Ajjan, R. A., Battelino, T., Cos, X., Del Prato, S., Philips, J.-C., Meyer, L., Seufert, J., & Seidu, S. (2024) · Nature Reviews Endocrinology · doi:10.1038/s41574-024-00973-1 · Cited by 117
“…values from the glucose sensor to a receiver or smartphone, whereas isCGM systems require users to scan their glucose sensor with the receiver or smartphone to get their up-to-date glucose readings and the associated trend arrows that show the direction and rate of glucose change at any point 53 .”
Ferreira, R. O. M., Trevisan, T., Pasqualotto, E., Chavez, M. P., Marques, B. F., Lamounier, R. N., & van de Sande-Lee, S. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0390 · Cited by 92
Castellana, M., Parisi, C., Di Molfetta, S., Di Gioia, L., Natalicchio, A., Perrini, S., Cignarelli, A., Laviola, L., & Giorgino, F. (2020) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2019-001092 · Cited by 84
Choe, H. J., Rhee, E.-J., Won, J. C., Park, K. S., Lee, W.-Y., & Cho, Y. M. (2022) · Diabetes Care · doi:10.2337/dc22-0764 · Cited by 59
“However, with the advent of CGM, it is possible to monitor glucose patterns every few minutes without repeatedly pricking fingers (6).”
BackgroundAvari, P., Uduku, C., George, D., Herrero, P., Reddy, M., & Oliver, N. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0276 · Cited by 40
Lu, J., Ying, Z., Wang, P., Fu, M., Han, C., & Zhang, M. (2024) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.15328 · Cited by 39
Kong, S.-Y., & Cho, M.-K. (2024) · Healthcare · doi:10.3390/healthcare12050571 · Cited by 36
Urakami, T., Yoshida, K., Kuwabara, R., Mine, Y., Aoki, M., Suzuki, J., & Morioka, I. (2022) · Journal of Diabetes Investigation · doi:10.1111/jdi.13618 · Cited by 26
Zhou, Y., Sardana, D., Kuroko, S., Haszard, J. J., de Block, M. I., Weng, J., Jefferies, C., & Wheeler, B. J. (2024) · Diabetic Medicine · doi:10.1111/dme.15280 · Cited by 25
Hypoglycaemia and its management in primary care setting
Ibrahim, M., Baker, J., Cahn, A., Eckel, R. H., El Sayed, N. A., Fischl, A. H., Gaede, P., Leslie, R. D., Pieralice, S., Tuccinardi, D., et al. (2020) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.3332 · Cited by 21
“CGM should be considered for all individuals with increased risk for hypoglycaemia, impaired hypoglycaemia awareness, frequent nocturnal hypoglycaemia and frequent severe hypoglycaemia.(79)”
Background
Metadata
- Authors
- Edelman, S. V., Argento, N. B., Pettus, J., & Hirsch, I. B.
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2018-10-15
- Volume
- 41
- Issue
- 11
- Pages
- 2265-2274
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2895865306
- Cited by
- 199
- References
- 34
- FWCI
- 11.94
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.