Fastest Thickness Measurements with a Terahertz Time-Domain System Based on Electronically Controlled Optical Sampling

Open access

The DOI matches its Crossref record; 36 of the 36 references it cites were found in a registry or an index; 8 of the 16 sources its claims cite were checked for support.

Abstract

We apply a fast terahertz time-domain spectroscopy (TDS) system based on electronically controlled optical sampling (ECOPS) to contact-free thickness gauging. Our setup achieves a measurement speed of 1600 terahertz pulse traces per second, which—to our knowledge—represents the fastest thickness measurement performed with any terahertz system to-date. Using a silicon wafer as a test sample, we compare data of the ECOPS experiment to results obtained with a conventional terahertz TDS system and a mechanical micrometer gauge. We show that all systems provide consistent results within the measurement accuracy. Moreover, we perform thickness measurements of a rapidly moving sample and characterize the ECOPS setup with respect to time-domain dynamic range, signal-to-noise ratio, and spectral properties.

Overview

CoChat Summary

TL;DR

This paper characterizes an electronically controlled optical sampling (ECOPS) terahertz time-domain system for fast thickness gauging, achieving up to 1600 pulse traces/s at 200 ps scan range and 200 traces/s at 700 ps scan range. It demonstrates non-contact thickness measurements of a high-resistivity silicon wafer that agree with a conventional TDS system and a micrometer gauge, including on a moving sample at ~700 mm/s.

Key findings

  • ECOPS reaches a maximum measurement speed of 1600 pulse traces/s at 200 ps scan range; at 700 ps scan range it maintains 200 traces/s. At maximum speed it is ~40× faster than the state-of-the-art TDS unit with mechanical delay, and at the lowest speed it still exceeds that system by ~5×. It is 160× faster than the TeraFlash pro at 1600 vs. 10 traces/s; after 10-trace averaging it remains 16× faster.
  • Single-shot TD-DR is 61 dB, independent of measurement speed, while the TeraFlash pro achieves 82 dB. Averaging 10, 100, and 1000 traces increases ECOPS TD-DR by approximately 10, 20, and 30 dB, respectively; ECOPS achieves 91 dB in less than one second. Single-shot TD-SNR is 50 dB at 1600 Hz and 55 dB at 200 Hz, versus 74 dB for the TeraFlash pro; a factor of 10 in averaging improves TD-SNR by ~10 dB. Single-shot PDR is 38 dB with signal reaching the noise floor at 3.4 THz; with 1000 averages PDR increases to 68 dB and detectable signal extends to ~4.8 THz within 0.6 s.
  • Thickness measurements of the silicon wafer gave 527 ± 1.5 µm with ECOPS, 527.94 ± 0.04 µm with the TeraFlash pro, and 527.5 ± 2.5 µm with the micrometer gauge; all three agree within a 1-sigma error margin, though TeraFlash pro scatter is ~40× smaller than ECOPS. On a moving sample at ~700 mm/s, ECOPS achieved spatial resolution of ~0.4 mm; thickness varied along the sample by as much as 10 µm, with ECOPS and micrometer-gauge results in good agreement. Preliminary results indicate layers as thin as ~30 µm can still be accurately determined.

Methods

  • The ECOPS setup uses two passively mode-locked femtosecond lasers at ~1560 nm, 80 MHz repetition rate, <80 fs pulse width, and ~25 mW incident power on the terahertz antennas. One laser is fixed (“Master”); the other (“Slave”) has a tunable repetition rate via a piezo-actuated mirror and is phase-locked with a PID controller. Photoconductive switches based on InAlAs/InGaAs are used for terahertz generation and detection. Measurements use a reflection setup at 8° incidence with ~2.5 mm terahertz spot size, a 3.5 MHz transimpedance amplifier with 10^6 V/A gain, and a digital oscilloscope.
  • Thickness is extracted via time-of-flight from the time difference between zero-crossings of pulse echoes from the front and back surfaces of an un-doped high-resistivity silicon wafer (n ~3.4). Results are compared with a TeraFlash pro at 10 traces/s and a manual micrometer gauge; terahertz measurements were repeated 10,000 times and micrometer measurements 50 times. A moving-sample test used a vinyl-record player at 45 rpm, translating to ~700 mm/s lateral velocity.

Limitations

  • Stated: the difference between ECOPS TD-SNR at 1600 Hz and 200 Hz is still subject to further investigation, tentatively attributed to increased timing jitter at higher speed affecting signal amplitude reproducibility. Minimum resolvable layer thickness results are preliminary and not shown; ~30 µm is reported, with thinner layers surmised possible using more sophisticated post-processing. ECOPS is 160× faster than the TeraFlash pro but has ~40× larger scatter; repeatability on the 1 µm level is deemed sufficient for many industrial applications.
  • Evident: thickness validation used a single high-resistivity silicon wafer and a mock-up conveyor (vinyl-record player), not a real production line; the reported ~0.4 mm spatial resolution does not consider the actual focus size.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.3390/app9071283

Cites

36 of 36 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Verification

Claims the paper makes

4 Valid · 1 Not valid · 5 Not checked · 51 Own results and interpretations

37 more claims

Create a free account to see all 61 claims and every source verdict.

See all claims — free account
  • It may state two separate claims as one. 17
  • The quote only partly states this claim. 1
  • Probably not a substantial claim. 4

Create a free account to see these claims.

Citations

Cites

Research works printed but not linked: 1

  1. Naftaly, M., and Dudley, R.A. (2011, January 26–29). Calibration of Terahertz Spectrometers. In Proceeding of the International Society for Optical Engineering, Nanjing, China.

Cited by

See all 63

Metadata

Authors
Yahyapour, M., Jahn, A., Dutzi, K., Puppe, T., Leisching, P., Schmauss, B., Vieweg, N., & Deninger, A.
Affiliations
  • Friedrich-Alexander-Universität Erlangen-Nürnberg (DE)
  • Toptica Photonics (Germany) (DE)
Corresponding author
Yahyapour, Milad
Venue
Applied Sciences
Publisher
MDPI AG
Published
2019-03-27
Volume
9
Issue
7
Pages
1283
ISSN
2076-3417 (ISSN-L)
OpenAlex
W2935454897
Cited by
63
References
33
Citations by year
2026: 2 · 2025: 6 · 2024: 12 · 2023: 9 · 2022: 13 · 2021: 9 · 2020: 6 · 2019: 6
FWCI
3.06
Type
journal article
Language
en
Open access
Locations
  • Applied Sciences (publishedVersion, cc-by)
  • OPUS Repository (Kooperativer Bibliotheksverbund Berlin-Brandenburg) (submittedVersion, cc-by)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion)
Topics
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Terahertz technology and applications
  • Physical Sciences › Engineering › Mechanics of Materials › Thermography and Photoacoustic Techniques
  • Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Laser-Matter Interactions and Applications
Keywords
terahertz time-domain spectroscopy, thickness measurement, measurement speed, silicon wafers, signal-to-noise ratio

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

The claim verdicts this report counts are experimental: their accuracy is not yet measured.

A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Cited sources are checked with a language model’s help (gpt-6-sol).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.