Trends and Challenges in Noninvasive Hemodynamic Monitoring of Neonates Following Cardiac Surgery: A Narrative Review
Open accessThe DOI matches its Crossref record; 87 of the 87 references it cites were found in a registry or an index; 63 of the 113 sources its claims cite were checked for support.
Abstract
Hemodynamic monitoring is essential in the postoperative management of neonates undergoing cardiac surgery, enabling early identification of circulatory failure and its underlying cause, optimization of oxygen delivery to tissues, and evaluation of treatment response. Despite its significant role, there is still no consensus and there remains substantial heterogeneity in bedside hemodynamic monitoring practices. Pediatric intensivists typically rely on macro- and microcirculatory indicators, including arterial blood pressure, urine output, capillary refill time, mixed venous oxygen saturation, lactate concentration, and serial echocardiographic evaluations. However, most of these are indirect hemodynamic indicators and provide only intermittent snapshots of the hemodynamic status, which can be very fluctuating following cardiac surgery. Technological advancements have driven a shift toward continuous, noninvasive monitoring techniques, such as near-infrared spectroscopy (NIRS), electrical biosensing technology, and microcirculatory assessment tools. Real-time, simultaneous tracking of multiple physiological variables through a multimodal hemodynamic monitoring protocol facilitates the understanding of systemic and regional perfusion and oxygenation. This narrative review aims to summarize current techniques and innovations in neonatal hemodynamic monitoring following cardiac surgery, combining clinical evaluation with both intermittent and continuous noninvasive techniques.
Overview
CoChat Summary
TL;DR
This narrative review summarizes current noninvasive and minimally invasive hemodynamic monitoring techniques for neonates after cardiac surgery, emphasizing that no single device meets all ideal requirements. It highlights NIRS, electrical biosensing, echocardiography, and microcirculatory tools, while noting limited validation, lack of standardization, and no consensus on thresholds or treatment targets.
Key findings
- Low cardiac output syndrome (LCOS) is the most common postoperative complication in children with congenital heart disease, with incidence 25–60% and typical onset within 6–18 h after surgery; definitions vary, with metabolic acidosis in 64%, inotropic support in 61%, and clinical manifestations in 59% of pediatric definitions.
- Indirect markers are limited: low cardiac output is clinically recognized in only 25% of pediatric intensive care cases; lactate is used by 99% of pediatric intensivists; postoperative acute kidney injury is reported in approximately 53.8% of neonates undergoing congenital heart surgery.
- Echocardiography remains the most widely used direct method, with neonatal cardiac output values generally 150–300 mL/kg/min, but it has a precision error around 30% versus dilution techniques and intra-/interobserver coefficients of variation from 2.1–22% and 3.1–22%, respectively; targeted neonatal echocardiography changes management in approximately 30–60% of hemodynamically unstable patients.
- NIRS average cerebral oxygen saturation is approximately 65–70% in term and preterm neonates versus 46–57% in cyanotic congenital heart disease; a cerebral oxygen saturation threshold <58% combined with lactate >7.4 mmol/L at 24 h post-neonatal cardiac surgery predicted survival and neurodevelopmental outcomes, but no universally accepted lower intervention threshold exists.
- The ESPR NICOM consensus states thoracic electrical biosensing technology is not currently reliable for precise cardiac output measurement in neonates and should not replace transthoracic echocardiography, though it may be useful for monitoring changes from an individual baseline.
Methods
- Narrative review with a systematic database search in PubMed and Embase for 2015–2025 using terms combining neonate/newborn/infant, congenital heart disease, cardiac surgery/procedure/intervention, and hemodynamic/hemodynamics; additional articles were found by manual search.
- One reviewer screened titles and abstracts for relevance and duplication; full texts in languages other than English were excluded. Evidence was synthesized narratively, with no formal meta-analysis or quantitative pooling described.
Limitations
- The paper states there is no gold standard or consensus in pediatric hemodynamic monitoring, leading to institutional variability; many studies of newer devices are small, single-center, and lack standardized reference values, limiting generalizability.
- The paper states newer devices are not routinely implemented, current expert recommendations support use primarily in research settings, and there are challenges in validation against reference methods, lack of measurement standardization, no consensus for guiding treatment, and limited data on patient outcomes.
- Evident from the text: the search and screening were performed by one reviewer, and no formal risk-of-bias assessment or quantitative synthesis is described, which may limit reproducibility and comprehensiveness.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures









Verification
Claims the paper makes
55 Valid · 15 Not valid · 48 Not checked · 77 Own results and interpretations
At least one cited source doesn’t support this claim or points to a different paper, and none confirmed it — see the verdicts below.
In the paper
“evaluating fluid responsiveness in children with shock [77]”
§ Trends and Challenges in Noninvasive Hemodynamic Monitoring of Neonates Following Cardiac Surgery: A Narrative Review > 4. Direct Hemodynamic Monitoring > 4.5. Thoracic Electrical Biosensing Technology
What the sources say
- Disputedcites [77] Awadhare P. Patel R. McCallin T. Mainali K. Jackson K. Starke H. Bhalala U. Non-invasive Cardiac Output Monitoring and Assessment of Fluid Responsiveness in Children With Shock in the Emergency Department Front. Pediatr. 2022 10 857106 10.3389/fped.2022.857106 35463892 PMC9021702 sourceExperimental — accuracy not yet measured
The Methods report measuring thoracic fluid content (TFC) with an electrical cardiometry device in children with shock. But the Results (Effects of Fluid Therapy and Fluid Responsiveness; page not available) found no significant difference in TFC change between fluid responders and non-responders. The paper identifies other measures—not TFC—as predictors of fluid responsiveness. (LLM, conf=0.88)
Checked by a reasoning model — may be wrong.
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Disputed
At least one cited source doesn’t support this claim or points to a different paper, and none confirmed it — see the verdicts below.
In the paper
“Scolletta et al. evaluated sublingual microcirculation in infants with cyanotic and acyanotic CHD using SDF, noting that while acyanotic infants maintained stable microvascular parameters, those with cyanotic CHD showed significant reductions in total vessel density (TVD) and proportion of perfused vessels (PPV) [86].”
§ Trends and Challenges in Noninvasive Hemodynamic Monitoring of Neonates Following Cardiac Surgery: A Narrative Review > 4. Direct Hemodynamic Monitoring > 4.6. Microcirculatory Monitoring Devices
What the sources say
- Disputedcites [86] Scolletta S. Marianello D. Isgrò G. Dapoto A. Terranova V. Franchi F. Baryshnikova E. Carlucci C. Ranucci M. Microcirculatory changes in children undergoing cardiac surgery: A prospective observational study Br. J. Anaesth. 2016 117 206 213 10.1093/bja/aew187 27440632 sourceExperimental — accuracy not yet measured
The Table 2 results report a significant increase in PPV over time in cyanotic children and an increase, not a reduction, in TVD. The paper does describe comparatively stable PPV in acyanotic children, but its findings run contrary to the claimed reductions in the cyanotic group. (LLM, conf=0.90)
Checked by a reasoning model — may be wrong.
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Disputed
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Low cardiac output state is accurately clinically recognized in only 25% of cases in pediatric intensive care units [10].”
§ Trends and Challenges in Noninvasive Hemodynamic Monitoring of Neonates Following Cardiac Surgery: A Narrative Review > 3. Indirect Hemodynamic Monitoring
What the sources say
- Partly supportedcites [10] Tibby S.M. Hatherill M. Marsh M.J. Murdoch I.A. Clinicians’ abilities to estimate cardiac index in ventilated children and infants Arch. Dis. Child. 1997 77 516 518 10.1136/adc.77.6.516 9496187 PMC1717412 sourceExperimental — accuracy not yet measured
Table 1 (Discussion; page unavailable) shows that clinicians correctly classified 12 of 47 measured low-cardiac-index estimates as low—about 25%. The Patients and methods section specifies that these estimates concerned ventilated, sedated children, so the result does not establish a 25% recognition rate for all pediatric intensive care patients. (LLM, conf=0.85)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“However, aggressive efforts to normalize lactate concentrations may lead to fluid overload, especially in the absence of other signs of tissue hypoperfusion [36,37].”
§ Trends and Challenges in Noninvasive Hemodynamic Monitoring of Neonates Following Cardiac Surgery: A Narrative Review > 3. Indirect Hemodynamic Monitoring > 3.6. Lactate
What the sources say
- Supportedcites [36] Kattan E. Hernández G. Ospina-Tascón G. Valenzuela E.D. Bakker J. Castro R. A lactate-targeted resuscitation strategy may be associated with higher mortality in patients with septic shock and normal capillary refill time: A post hoc analysis of the ANDROMEDA-SHOCK study Ann. Intensive Care 2020 10 114 10.1186/s13613-020-00732-1 32845407 PMC7450018 source
The Discussion reports that patients with normal capillary refill time who received lactate-targeted resuscitation received more fluid boluses. It also describes an increased burden of over-resuscitation and cautions against pursuing lactate normalization without considering peripheral perfusion. These passages support the claim’s hedged warning in the septic-shock setting studied; page numbers were not available in our copy. (LLM, conf=0.88)
- Supportedcites [37] Bakker J. de Backer D. Hernandez G. Lactate-guided resuscitation saves lives: We are not sure Intensive Care Med. 2016 42 472 474 10.1007/s00134-016-4220-z 26831675 source
In the paper’s discussion (page not provided), the authors warn that pursuing lactate normalization with further fluids or inotropes after other signs of tissue hypoperfusion have disappeared may expose patients to the toxicity of overresuscitation without clear benefit. This supports the citing sentence’s qualified warning about aggressive fluid administration. (LLM, conf=0.88)
hedgedstates a possible cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“TNE can lead to changes in clinical management in approximately 30–60% of patients with hemodynamic instability [50].”
§ Trends and Challenges in Noninvasive Hemodynamic Monitoring of Neonates Following Cardiac Surgery: A Narrative Review > 4. Direct Hemodynamic Monitoring > 4.1. Echocardiography
What the sources say
- Supportedcites [50] Singh Y. Echocardiographic evaluation of hemodynamics in neonates and children Front. Pediatr. 2017 5 201 10.3389/fped.2017.00201 28966921 PMC5605552 source
The Introduction reports that clinical management changed in 30–60% of cases after point-of-care echocardiography. The Abstract places its use in clinical decision-making for patients with hemodynamic instability, supporting the paraphrase. (LLM, conf=0.90)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
83 more claims · 26 with their cited sources checked
Create a free account to see all 195 claims and to verify the 3 references still open.
- It may state two separate claims as one. 77
- Probably not a substantial claim. 45
- The quote only partly states this claim. 2
Citations
Cites
Diagnosis, evaluation, and management of acute kidney injury: a KDIGO summary (Part 1)
Kellum, J. A., Lameire, N., & for the KDIGO AKI Guideline Work Group (2013) · Critical Care · doi:10.1186/cc11454 · Cited by 2,827 · cited 1×
“According to the KDIGO guidelines, acute kidney injury (AKI) is diagnosed based on elevated serum creatinine levels and/or diminished urine output [19].”
Gaies, M. G., Gurney, J. G., Yen, A. H., Napoli, M. L., Gajarski, R. J., Ohye, R. G., Charpie, J. R., & Hirsch, J. C. (2010) · Pediatric Critical Care Medicine · doi:10.1097/pcc.0b013e3181b806fc · Cited by 1,325 · cited 1×
“Gaies et al. applied VIS to neonates and small infants after cardiac surgery, defining high VIS as ≥ 20 during the first 24 h or ≥ 15 during the 24–48 h postoperative period [43].”
Wernovsky, G., Wypij, D., Jonas, R. A., Mayer, J. E., Hanley, F. L., Hickey, P. R., Walsh, A. Z., Chang, A. C., Castañeda, A. R., Newburger, J. W., et al. (1995) · Circulation · doi:10.1161/01.cir.92.8.2226 · Cited by 1,010
Jansen, T. C., van Bommel, J., Schoonderbeek, F. J., Sleeswijk Visser, S. J., van der Klooster, J. M., Lima, A. P., Willemsen, S. P., & Bakker, J. (2010) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.200912-1918oc · Cited by 894 · cited 1×
“Strategies aimed at reducing lactate levels have been linked to improved clinical outcomes, reduced organ dysfunction and lower mortality rates in various shock states [35].”
Marik, P. E., & Cavallazzi, R. (2013) · Critical Care Medicine · doi:10.1097/ccm.0b013e31828a25fd · Cited by 866 · cited 1×
“Central venous pressure (CVP), measured at the junction of the superior vena cava and right atrium, remains a vital part of the hemodynamic monitoring post-cardiopulmonary bypass, despite its limited utility in assessing fluid responsiveness [15,16].”
Cecconi, M., Rhodes, A., Poloniecki, J., Della Rocca, G., & Grounds, R. M. (2009) · Critical Care · doi:10.1186/cc7129 · Cited by 351 · cited 1×
“For instance, echocardiography shows a precision error of around 30% when compared to gold standard dilution techniques, which falls within the clinically accepted range [54,55].”
Vasoactive-Inotropic Score: Evolution, Clinical Utility, and Pitfalls
Belletti, A., Lerose, C. C., Zangrillo, A., & Landoni, G. (2021) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1053/j.jvca.2020.09.117 · Cited by 295
Clinicians’ abilities to estimate cardiac index in ventilated children and infants
Tibby, S. M., Hatherill, M., Marsh, M. J., & Murdoch, I. A. (1997) · Archives of Disease in Childhood · doi:10.1136/adc.77.6.516 · Cited by 224 · cited 1×
“Low cardiac output state is accurately clinically recognized in only 25% of cases in pediatric intensive care units [10].”
Alderliesten, T., Dix, L., Baerts, W., Caicedo, A., van Huffel, S., Naulaers, G., Groenendaal, F., van Bel, F., & Lemmers, P. (2016) · Pediatric Research · doi:10.1038/pr.2015.186 · Cited by 199 · cited 1×
“Average ScO 2 values range between approximately 65 and 70% in both term and preterm neonates [68,69], in comparison to 46–57% in patients with cyanotic congenital heart disease [12].”
Charpie, J. R., Dekeon, M. K., Goldberg, C. S., Mosca, R. S., Bove, E. L., & Kulik, T. J. (2000) · The Journal of Thoracic and Cardiovascular Surgery · doi:10.1067/mtc.2000.106838 · Cited by 198 · cited 1×
“Elevated lactate concentrations have been associated with increased mortality and a higher risk for requiring ECMO post-cardiac surgery [34].”
Echocardiographic Evaluation of Ventricular Function—For the Neonatologist and Pediatric Intensivist
Tissot, C., Singh, Y., & Sekarski, N. (2018) · Frontiers in Pediatrics · doi:10.3389/fped.2018.00079 · Cited by 181 · cited 1×
“In neonates without shunting, echocardiographic measurement of cardiac output is generally reliable, with values ranging between 150 and 300 mL/kg/min [53,54].”
Prophylactic intravenous use of milrinone after cardiac operation in pediatrics (PRIMACORP) study
Hoffman, T. M., Wernovsky, G., Atz, A. M., Bailey, J. M., Akbary, A., Kocsis, J. F., Nelson, D. P., Chang, A. C., Kulik, T. J., Spray, T. L., et al. (2002) · American Heart Journal · doi:10.1067/mhj.2002.120305 · Cited by 167 · cited 1×
“…as a clinical syndrome with tachycardia, oliguria, cold extremities, and/or cardiac arrest, with or without an arterial–mixed venous oxygen saturation difference ≥ 30%, or metabolic acidosis (base deficit > 4 mmol/L or lactate concentration > 2 mg/dL) on two consecutive blood gas measurements [5].”
Effective hemodynamic monitoring
Pinsky, M. R., Cecconi, M., Chew, M. S., De Backer, D., Douglas, I., Edwards, M., Hamzaoui, O., Hernandez, G., Martin, G., Monnet, X., et al. (2022) · Critical Care · doi:10.1186/s13054-022-04173-z · Cited by 167 · cited 2×
“CRT has been validated as a rapid response marker during septic shock resuscitation and is considered useful for evaluating the effectiveness of fluid therapy or vasoactive medication [23].”
Early markers of major adverse events in children after cardiac operations
Duke, T., Butt, W., South, M., & Karl, T. R. (1997) · The Journal of Thoracic and Cardiovascular Surgery · doi:10.1016/s0022-5223(97)70018-7 · Cited by 152 · cited 1×
“Nonetheless, other studies did not find any statistically significant associations [28,29].”
Tibby, S. M., Hatherill, M., & Murdoch, I. A. (1999) · Archives of Disease in Childhood · doi:10.1136/adc.80.2.163 · Cited by 152
Chew, M. S., & Poelaert, J. (2003) · Intensive Care Medicine · doi:10.1007/s00134-003-1967-9 · Cited by 151 · cited 1×
“Despite its known limitations, including relatively high intra-and interobserver variability (coefficient of variation ranging from 2.1% to 22% and 3.1% to 22%, respectively) [58], echocardiography remains the most widely adopted modality in neonatal and pediatric intensive care.”
InfluentialNoori, S., Drabu, B., Soleymani, S., & Seri, I. (2012) · Archives of Disease in Childhood - Fetal and Neonatal Edition · doi:10.1136/fetalneonatal-2011-301090 · Cited by 146 · cited 1×
“…al. compared stroke volume measurements obtained through EC and TTE in preterm infants with low and very low birth weights [72], while Noori et al. examined left ventricular output in preterm neonates with and without patent ductus arteriosus, finding good agreement between the two modalities [73].”
Jacobs, J. P., O'Brien, S. M., Pasquali, S. K., Jacobs, M. L., Lacour-Gayet, F. G., Tchervenkov, C. I., Austin, E. H., Pizarro, C., Pourmoghadam, K. K., Scholl, F. G., et al. (2012) · The Annals of Thoracic Surgery · doi:10.1016/j.athoracsur.2012.01.105 · Cited by 136 · cited 1×
“In a multicenter study focusing exclusively on neonates, a VVR score ≥ 35 at 48 h post-surgery was more predictive of prolonged mechanical ventilation than VIS, the ventilation index (VI), or lactate levels [46].”
Park, S.-K., Hur, M., Kim, E., Kim, W. H., Park, J. B., Kim, Y., Yang, J.-H., Jun, T.-G., & Kim, C. S. (2016) · PLOS ONE · doi:10.1371/journal.pone.0166328 · Cited by 133 · cited 1×
“Several perioperative factors contribute to cardiac surgery-associated AKI development, including cardiopulmonary bypass, aortic cross-clamping, low cardiac output syndrome, administration of vasoactive agents and albumin, transfusions, inflammation, and clotting abnormalities [22].”
Alten, J. A., Cooper, D. S., Blinder, J. J., Selewski, D. T., Tabbutt, S., Sasaki, J., Gaies, M. G., Bertrandt, R. A., Smith, A. H., Reichle, G., et al. (2021) · Critical Care Medicine · doi:10.1097/ccm.0000000000005165 · Cited by 126 · cited 1×
“Postoperative urine output monitoring facilitates the early detection of AKI, which has been reported in approximately 53.8% of neonates undergoing congenital heart surgery [21].”
InfluentialHoffman, G. M., Ghanayem, N. S., Scott, J. P., Tweddell, J. S., Mitchell, M. E., & Mussatto, K. A. (2017) · The Annals of Thoracic Surgery · doi:10.1016/j.athoracsur.2016.09.100 · Cited by 123 · cited 1×
“Similarly, in a large-scale investigation, Hoffman et al. demonstrated that NIRS measurements at 6 h after palliative surgery for hypoplastic left heart syndrome were strong predictors of early mortality and the need for extracorporeal membrane oxygenation [67].”
Echocardiographic Evaluation of Hemodynamics in Neonates and Children
Singh, Y. (2017) · Frontiers in Pediatrics · doi:10.3389/fped.2017.00201 · Cited by 116 · cited 1×
“TNE can lead to changes in clinical management in approximately 30–60% of patients with hemodynamic instability [50].”
Ficial, B., Finnemore, A. E., Cox, D. J., Broadhouse, K. M., Price, A. N., Durighel, G., Ekitzidou, G., Hajnal, J. V., Edwards, A. D., & Groves, A. M. (2013) · Journal of the American Society of Echocardiography · doi:10.1016/j.echo.2013.08.019 · Cited by 102 · cited 1×
“Similar levels of variability have been noted in comparison to cardiac MRI [56].”
Ranucci, M., Isgrò, G., Carlucci, C., De La Torre, T., Enginoli, S., Frigiola, A., & Surgical and Clinical Outcome REsearch (SCORE) Group (2010) · Critical Care · doi:10.1186/cc9217 · Cited by 100 · cited 1×
“In pediatric patients undergoing cardiac surgery, ScvO 2 levels < 68% associated with lactate concentrations > 3 mmol/L during cardiopulmonary bypass have been correlated with increased morbidity and mortality [38].”
Kattan, E., Hernández, G., Ospina-Tascón, G., Valenzuela, E. D., Bakker, J., & Castro, R. (2020) · Annals of Intensive Care · doi:10.1186/s13613-020-00732-1 · Cited by 100 · cited 1×
“However, aggressive efforts to normalize lactate concentrations may lead to fluid overload, especially in the absence of other signs of tissue hypoperfusion [36,37].”
Clinical monitoring of systemic hemodynamics in critically ill newborns
de Boode, W.-P. (2010) · Early Human Development · doi:10.1016/j.earlhumdev.2010.01.031 · Cited by 87
Singh, Y., Villaescusa, J. U., da Cruz, E. M., Tibby, S. M., Bottari, G., Saxena, R., Guillén, M., Herce, J. L., Di Nardo, M., Cecchetti, C., et al. (2020) · Critical Care · doi:10.1186/s13054-020-03326-2 · Cited by 86 · cited 6×
“…Pediatric and Neonatal Intensive Care (ESPNIC) advises against relying solely on blood pressure as a therapeutic target in hemodynamically unstable children, instead advocating for a comprehensive assessment that integrates blood pressure with a range of clinical and paraclinical parameters [12].”
InfluentialAly, S. A., Zurakowski, D., Glass, P., Skurow-Todd, K., Jonas, R. A., & Donofrio, M. T. (2017) · Congenital Heart Disease · doi:10.1111/chd.12426 · Cited by 77 · cited 1×
“Aly et al. reported that a cerebral oxygen saturation threshold of <58%, when combined with a lactate concentration > 7.4 mmol/L at 24 h post-neonatal cardiac surgery, served as a reliable predictor of survival and neurodevelopmental outcomes [66].”
InfluentialSchoonen, A., van Klei, W. A., van Wolfswinkel, L., & van Loon, K. (2022) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2022.926957 · Cited by 74
Understanding the venous–arterial CO2 to arterial–venous O2 content difference ratio
Ospina-Tascón, G. A., Hernández, G., & Cecconi, M. (2016) · Intensive Care Medicine · doi:10.1007/s00134-016-4233-7 · Cited by 60
Weber, F., & Scoones, G. P. (2019) · Pediatric Anesthesia · doi:10.1111/pan.13726 · Cited by 53 · cited 1×
“Cerebral oxygen saturation (ScO 2 ) acts as a marker of cardiac output if ventilation, oxygenation, and hemoglobin levels remain stable [65].”
Non-Invasive Cardiac Output Measurement in Low and Very Low Birth Weight: A Method Comparison
Grollmuss, O., & Gonzalez, P. (2014) · Frontiers in Pediatrics · doi:10.3389/fped.2014.00016 · Cited by 52
Roeleveld, P. P., & de Klerk, J. C. A. (2018) · World Journal for Pediatric and Congenital Heart Surgery · doi:10.1177/2150135117731725 · Cited by 51 · cited 1×
“According to a survey study, 99% of pediatric intensivists include it in their assessments [32].”
Lactate-guided resuscitation saves lives: We are not sure
Bakker, J., de Backer, D., & Hernandez, G. (2016) · Intensive Care Medicine · doi:10.1007/s00134-016-4220-z · Cited by 48 · cited 1×
“However, aggressive efforts to normalize lactate concentrations may lead to fluid overload, especially in the absence of other signs of tissue hypoperfusion [36,37].”
Growth restriction and gender influence cerebral oxygenation in preterm neonates
Cohen, E., Baerts, W., Alderliesten, T., Derks, J., Lemmers, P., & van Bel, F. (2016) · Archives of Disease in Childhood - Fetal and Neonatal Edition · doi:10.1136/archdischild-2015-308843 · Cited by 46 · cited 1×
“Average ScO 2 values range between approximately 65 and 70% in both term and preterm neonates [68,69], in comparison to 46–57% in patients with cyanotic congenital heart disease [12].”
Ulate, K. P., Yanay, O., Jeffries, H., Baden, H., Di Gennaro, J. L., & Zimmerman, J. (2017) · Pediatric Critical Care Medicine · doi:10.1097/pcc.0000000000000979 · Cited by 45 · cited 3×
“Both Ulate et al. [30] and Aslan et al. [33] integrated lactate measurement into their LCOS scoring systems as a marker of tissue perfusion.”
Mottling score and skin temperature in septic shock: Relation and impact on prognosis in ICU
Ferraris, A., Bouisse, C., Mottard, N., Thiollière, F., Anselin, S., Piriou, V., & Allaouchiche, B. (2018) · PLOS ONE · doi:10.1371/journal.pone.0202329 · Cited by 45 · cited 1×
“IRT, a noninvasive imaging modality, offers objective and reproducible measurements, which makes it suitable for evaluating peripheral perfusion and microcirculation [26].”
Perturbation of the microvascular glycocalyx and perfusion in infants after cardiopulmonary bypass
Nussbaum, C., Haberer, A., Tiefenthaller, A., Januszewska, K., Chappell, D., Brettner, F., Mayer, P., Dalla Pozza, R., & Genzel-Boroviczény, O. (2015) · The Journal of Thoracic and Cardiovascular Surgery · doi:10.1016/j.jtcvs.2015.08.050 · Cited by 41 · cited 1×
“Nussbaum et al. employed SDF imaging to study the effects of cardiopulmonary bypass on skin microcirculation, reporting transient disturbances in glycocalyx thickness, microvascular flow index (MFI), and perfused vessel density (PVD), which normalized within a week postoperatively [85].”
Urine Output Assessment in Acute Kidney Injury: The Cheapest and Most Impactful Biomarker
Goldstein, S. L. (2020) · Frontiers in Pediatrics · doi:10.3389/fped.2019.00565 · Cited by 41 · cited 1×
“Urine output is regarded as a sensitive and early marker of renal dysfunction [20].”
Schey, B. M., Williams, D. Y., & Bucknall, T. (2010) · Heart & Lung · doi:10.1016/j.hrtlng.2009.04.002 · Cited by 40 · cited 3×
“Additionally, a central– peripheral temperature difference > 5 ◦ C has been used as a marker in several studies, showing associations with decreased cardiac output and elevated peripheral vascular resistance [25,31].”
Miletic, K. G., Spiering, T. J., Delius, R. E., Walters, H. L., & Mastropietro, C. W. (2015) · Interactive CardioVascular and Thoracic Surgery · doi:10.1093/icvts/ivu409 · Cited by 40
Narula, J., Chauhan, S., Ramakrishnan, S., & Gupta, S. K. (2017) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1053/j.jvca.2016.12.009 · Cited by 39 · cited 1×
“Multiple studies have evaluated EC in neonates and infants with congenital heart disease, reporting statistically significant variations in cardiac index and left ventricular ejection fraction among heart failure subgroups [70].”
Thoracic fluid content: A novel parameter for predicting failed weaning from mechanical ventilation
Fathy, S., Hasanin, A. M., Raafat, M., Mostafa, M. M. A., Fetouh, A. M., Elsayed, M., Badr, E. M., Kamal, H. M., & Fouad, A. Z. (2020) · Journal of Intensive Care · doi:10.1186/s40560-020-00439-2 · Cited by 36 · cited 1×
“…as predicting outcomes in critically ill children [76], evaluating fluid responsiveness in children with shock [77], detecting pulmonary congestion [78], monitoring fluid overload during hemodialysis [79], and distinguishing between hemodynamically significant and closing/restrictive patent ductus…”
New technologies in pediatric anesthesia
Holtby, H., Skowno, J. J., Kor, D. J., Flick, R. P., & Uezono, S. (2012) · Pediatric Anesthesia · doi:10.1111/pan.12007 · Cited by 35 · cited 1×
“Additionally, EC has shown low bias when compared with TTE in infants postoperatively [74].”
Pees, C., Glagau, E., Hauser, J., & Michel-Behnke, I. (2013) · Pediatric Cardiology · doi:10.1007/s00246-012-0628-6 · Cited by 35 · cited 1×
“There are normative data for neonatal VTI and cardiac index, but these measurements are highly dependent on correct ultrasound probe positioning and accurate estimation of the outflow tract diameter (see Figure 2) [52].”
Advanced Hemodynamic Monitoring in the Neonatal Intensive Care Unit
de Boode, W.-P. (2020) · Clinics in Perinatology · doi:10.1016/j.clp.2020.05.001 · Cited by 34
Neonatal Blood Pressure Standards
Batton, B. (2020) · Clinics in Perinatology · doi:10.1016/j.clp.2020.05.008 · Cited by 31 · cited 1×
“A universally accepted definition of neonatal hypotension is still lacking [11].”
Seminal Postoperative Complications and Mode of Death After Pediatric Cardiac Surgical Procedures
Gaies, M., Pasquali, S. K., Donohue, J. E., Dimick, J. B., Limbach, S., Burnham, N. B., Ravishankar, C., Ohye, R. G., Gaynor, J. W., & Mascio, C. E. (2016) · The Annals of Thoracic Surgery · doi:10.1016/j.athoracsur.2016.02.043 · Cited by 29
Hernández, G., Castro, R., & Bakker, J. (2020) · Journal of Thoracic Disease · doi:10.21037/jtd.2019.12.102 · Cited by 27
Sumbel, L., Wats, A., Salameh, M., Appachi, E., & Bhalala, U. (2021) · Frontiers in Pediatrics · doi:10.3389/fped.2020.564902 · Cited by 27 · cited 1×
“It has proven useful in a variety of clinical scenarios, such as predicting outcomes in critically ill children [76], evaluating fluid responsiveness in children with shock [77], detecting pulmonary congestion [78], monitoring fluid overload during hemodialysis [79], and distinguishing between…”
Erdem, Ö., Ince, C., Tibboel, D., & Kuiper, J. W. (2019) · Frontiers in Pediatrics · doi:10.3389/fped.2019.00273 · Cited by 25 · cited 2×
“In neonates, commonly used imaging sites include the sublingual and buccal mucosa, as well as transcutaneous visualization of the upper inner arm and axilla [82].”
Neonatal hemodynamics: Monitoring, data acquisition and analysis
Soleymani, S., Borzage, M., Noori, S., & Seri, I. (2012) · Expert Review of Medical Devices · doi:10.1586/erd.12.32 · Cited by 23
Targeted Neonatal Echocardiography in Patients With Hemodynamic Instability
Rahde Bischoff, A., Bhombal, S., Altman, C. A., Fraga, M. V., Punn, R., Rohatgi, R. K., Lopez, L., & McNamara, P. J. (2022) · PEDIATRICS · doi:10.1542/peds.2022-056415i · Cited by 21 · cited 1×
“However, accurate image acquisition and interpretation require training of the neonatologist, as well as a close collaboration with pediatric cardiologists [51].”
Vasoactive-ventilation-renal score in predicting outcome postcardiac surgery in children
Akunuri, S., Alam, S., Jain, A., Mazahir, R., & Hegde, R. (2018) · International Journal of Critical Illness and Injury Science · doi:10.4103/ijciis.ijciis_1_18 · Cited by 20 · cited 1×
“The VVR score has shown utility in predicting postoperative outcomes and mortality in pediatric cardiac surgery patients [45].”
Department of Pediatric Intensive Care, M. T. A. R. H., Aslan, N., Yildizdas, D., & Department of Pediatric Intensive Care, C. U. F. O. M. (2022) · Turk Kardiyoloji Dernegi Arsivi-Archives of the Turkish Society of Cardiology · doi:10.5543/tkda.2022.21212 · Cited by 20 · cited 1×
“NIRS was proved efficient in identifying low cardiac output syndrome (LCOS), leading to its routine use after cardiopulmonary bypass [64].”
Wang, P., Fu, C., Bai, G., Cuan, L., Tang, X., Jin, C., Jin, H., Zhu, J., & Xie, C. (2023) · Frontiers in Pediatrics · doi:10.3389/fped.2022.954427 · Cited by 20 · cited 1×
“It is the most common postoperative complication in children with congenital heart disease (CHD), with an incidence ranging between 25 and 60% and typically manifesting within the first 6 to 18 h following surgery [3].”
InfluentialSchiefenhövel, F., Trauzeddel, R. F., Sander, M., Heringlake, M., Groesdonk, H. V., Grubitzsch, H., Kruppa, J., Berger, C., Treskatsch, S., & Balzer, F. (2021) · Journal of Clinical Medicine · doi:10.3390/jcm10173945 · Cited by 19
Evaluation of Hemodynamics in Infants and Small Babies after Open Heart Surgery
Alexi-Meskhishvili, V., Popov, S., & Nikoljuk, A. (1984) · The Thoracic and Cardiovascular Surgeon · doi:10.1055/s-2007-1023335 · Cited by 17
Murdoch, I., Qureshi, S., Mitchell, A., & Huggon, I. (1993) · Acta Paediatrica · doi:10.1111/j.1651-2227.1993.tb12556.x · Cited by 17 · cited 1×
“Multiple studies have demonstrated a positive correlation between CPTd and systemic vascular resistance, as well as inverse relationships with cardiac output, stroke index, and urine output [25,27].”
Microcirculatory changes in children undergoing cardiac surgery: A prospective observational study
Scolletta, S., Marianello, D., Isgrò, G., Dapoto, A., Terranova, V., Franchi, F., Baryshnikova, E., Carlucci, C., & Ranucci, M. (2016) · British Journal of Anaesthesia · doi:10.1093/bja/aew187 · Cited by 17 · cited 1×
“…sublingual microcirculation in infants with cyanotic and acyanotic CHD using SDF, noting that while acyanotic infants maintained stable microvascular parameters, those with cyanotic CHD showed significant reductions in total vessel density (TVD) and proportion of perfused vessels (PPV) [86].”
A recommendation for the use of electrical biosensing technology in neonatology
van Wyk, L., Austin, T., Barzilay, B., Bravo, M. C., Breindahl, M., Czernik, C., Dempsey, E. M., de Boode, W. P., de Vries, W. B., Eriksen, B. H., et al. (2024) · Pediatric Research · doi:10.1038/s41390-024-03369-z · Cited by 15
Martini, S., Gatelli, I. F., Vitelli, O., Vitali, F., De Rienzo, F., Parladori, R., Corvaglia, L., & Martinelli, S. (2023) · European Journal of Pediatrics · doi:10.1007/s00431-023-05106-w · Cited by 14 · cited 1×
“…in critically ill children [76], evaluating fluid responsiveness in children with shock [77], detecting pulmonary congestion [78], monitoring fluid overload during hemodialysis [79], and distinguishing between hemodynamically significant and closing/restrictive patent ductus arteriosus (PDA) [80].”
Non-invasive Cardiac Output Monitoring in Congenital Heart Disease
Tandon, A., Bhattacharya, S., Morca, A., Inan, O. T., Munther, D. S., Ryan, S. D., Latifi, S. Q., Lu, N., Lasa, J. J., Marino, B. S., et al. (2023) · Current Treatment Options in Pediatrics · doi:10.1007/s40746-023-00274-1 · Cited by 12 · cited 1×
“VIS serves as an indirect measure of clinical severity and of circulatory support needed [42].”
Cerebral blood flow dynamics during cardiac surgery in infants
Leth-Olsen, M., Døhlen, G., Torp, H., & Nyrnes, S. A. (2025) · Pediatric Research · doi:10.1038/s41390-024-03161-z · Cited by 12
Menger, J., Edlinger-Stanger, M., Dworschak, M., & Steinlechner, B. (2018) · Wiener klinische Wochenschrift · doi:10.1007/s00508-018-1403-3 · Cited by 11
Backer, D. D., Hajjar, L. A., & Monnet, X. (2024) · Intensive Care Medicine · doi:10.1007/s00134-024-07566-6 · Cited by 11
Validation of the Vasoactive-Ventilation-Renal Score for Neonatal Heart Surgery
Zubarioglu, A. U., Yıldırım, Ö., Zeybek, C., Balaban, İ., Yazıcıoglu, V., & Aliyev, B. (2021) · Cureus · doi:10.7759/cureus.15110 · Cited by 10 · cited 1×
“Another study identified a peak VVR ≥ 46.5 within the first 72 h after surgery as a marker of increased risk for adverse outcomes in neonates with congenital heart disease [47].”
InfluentialAwadhare, P., Patel, R., McCallin, T., Mainali, K., Jackson, K., Starke, H., & Bhalala, U. (2022) · Frontiers in Pediatrics · doi:10.3389/fped.2022.857106 · Cited by 9 · cited 1×
“…variety of clinical scenarios, such as predicting outcomes in critically ill children [76], evaluating fluid responsiveness in children with shock [77], detecting pulmonary congestion [78], monitoring fluid overload during hemodialysis [79], and distinguishing between hemodynamically significant…”
Microcirculatory Differences in Children With Congenital Heart Disease According to Cyanosis and Age
González, R., Urbano, J., Solana, M. J., Hervías, M., Pita, A., Pérez, R., Álvarez, R., Teigell, E., Gil-Jaurena, J.-M., Zamorano, J., et al. (2019) · Frontiers in Pediatrics · doi:10.3389/fped.2019.00264 · Cited by 8 · cited 1×
“One small pediatric study found that patients with cyanotic CHD show higher vessel densities than those with acyanotic CHD [83].”
Effect of hemodialysis on impedance cardiography (electrical velocimetry) parameters in children
Wilken, M., Oh, J., Pinnschmidt, H. O., Singer, D., & Blohm, M. E. (2020) · Pediatric Nephrology · doi:10.1007/s00467-019-04409-1 · Cited by 8 · cited 1×
“…in critically ill children [76], evaluating fluid responsiveness in children with shock [77], detecting pulmonary congestion [78], monitoring fluid overload during hemodialysis [79], and distinguishing between hemodynamically significant and closing/restrictive patent ductus arteriosus (PDA) [80].”
Taiana, M., Tomasella, I., Russo, A., Lerose, A., Graziadei, M. C., Corubolo, L., Rama, J., Schweiger, V., Vignola, A., Polati, E., et al. (2023) · Journal of Clinical Medicine · doi:10.3390/jcm12175700 · Cited by 8
Cardiac output monitoring in paediatric cardiac surgery: A review
Woodman, H. M., Lee, C., Ahmed, A. N., Malik, B. A., Mellor, S., Brown, L. J., Gentle, L., & Harky, A. (2021) · Cardiology in the Young · doi:10.1017/s1047951120004680 · Cited by 7 · cited 1×
“Simultaneous assessment of blood pressure and cardiac output allows a pathophysiology-oriented management of the circulatory failure [13].”
Karunanithi, Z., Jegatheeswaran, A., Goodfellow, S. D., Dixon, W., Hjortdal, V. E., Eytan, D., & Mazwi, M. L. (2022) · International Journal of Cardiology Congenital Heart Disease · doi:10.1016/j.ijcchd.2022.100333 · Cited by 7 · cited 1×
“Reduced HRV is associated with impaired autonomic regulation and elevated risk of hemodynamic instability and has been observed in patients with congenital heart defects before and after surgeries involving cardiopulmonary bypass [17,18].”
Gürdoğan, M., Ustabaşıoğlu, F. E., Kula, O., & Korkmaz, S. (2019) · Medicina · doi:10.3390/medicina55060260 · Cited by 6
Bridier, A., Shcherbakova, M., Kawaguchi, A., Poirier, N., Said, C., Noumeir, R., & Jouvet, P. (2023) · Frontiers in Pediatrics · doi:10.3389/fped.2023.1083962 · Cited by 6 · cited 1×
“Low cardiac output syndrome (LCOS) is characterized by insufficient delivery of oxygenated blood to meet the metabolic demands of tissues, leading to tissue hypoxia and metabolic acidosis [2].”
Taiana, M., Tomasella, I., Russo, A., Lerose, A., Graziadei, M. C., Corubolo, L., Rama, J., Schweiger, V., Vignola, A., Polati, E., et al. (2023) · doi:10.20944/preprints202308.1985.v1 · Cited by 6
Erdem, Ö., de Graaff, J. C., Hilty, M. P., Kraemer, U. S., de Liefde, I. I., van Rosmalen, J., Ince, C., Tibboel, D., & Kuiper, J. W. (2023) · Journal of Cardiovascular Translational Research · doi:10.1007/s12265-023-10407-4 · Cited by 5 · cited 1×
“A recent single-center study investigated the microcirculation of patients with heart disease, before and after cardiac surgery [84].”
Ferreira, F. M., Araujo, D. D., Dantas, G. M., Cunha, L. C. C., Zeferino, S. P., & Galas, F. B. (2025) · Brazilian Journal of Anesthesiology (English Edition) · doi:10.1016/j.bjane.2025.844614 · Cited by 5 · cited 1×
“Therapeutic strategies aimed at optimizing ScvO 2 have been associated with improved clinical outcomes, reduced organ dysfunction, decreased need for vasoactive drugs, decreased mechanical ventilation duration, and enhanced survival rates [39].”
Sumbel, L., Nagaraju, L., Ogbeifun, H., Agarwal, A., & Bhalala, U. (2022) · Progress in Pediatric Cardiology · doi:10.1016/j.ppedcard.2022.101551 · Cited by 4 · cited 2×
“Cardiac output estimated via EC has also shown good correlation with cardiac magnetic resonance imaging (CMRI) in pediatric patients with congenital heart defects, although EC tends to slightly overestimate CO [60].”
The Gold Standard of Pediatric Hemodynamic Monitoring: Not All That Glitters Is Gold
Ricci, Z., Iacobelli, R., & Romagnoli, S. (2023) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000006215 · Cited by 4
Menif, K., Ayari, A., Louati, A., Ibn Haj Hassine, S., Bouziri, A., & Borgi, A. (2024) · La Tunisie Médicale · doi:10.62438/tunismed.v102i9.5095 · Cited by 3
Research Progress of Noninvasive Cardiac Output Monitor (NICOTM) and Hemodynamic Monitor in Neonates
Li, F., & Dong, W. (2023) · Clinical and Experimental Obstetrics & Gynecology · doi:10.31083/j.ceog5010211 · Cited by 2 · cited 1×
“Moreover, EC has shown strong agreement with transthoracic echocardiography regarding stroke volume, stroke volume variation (SVV), and the inferior vena cava distensibility index (dIVC) [71].”
Hemodynamic monitoring in pediatric cardiac critical care
Kumar, A., & Joshi, R. K. (2025) · JOURNAL OF PEDIATRIC CRITICAL CARE · doi:10.4103/jpcc.jpcc_21_25 · Cited by 2 · cited 3×
“…(ECG) monitoring following cardiac surgery allows the assessment of heart rate, pacemaker function, and detection of arrhythmias, myocardial ischemia, and acute ST-segment changes, particularly after coronary artery reimplantation during procedures such as arterial switch operation [14].”
InfluentialAslan, N. (2020) · Turk Kardiyoloji Dernegi Arsivi-Archives of the Turkish Society of Cardiology · doi:10.5543/tkda.2020.13844 · Cited by 2
Heart Rate Variability in Children with Congenital Heart Disease Before and After Open Heart Surgery
Özeren, M., Lu, O. H., Makharoblidze, K., & AnkaralI, H. (2009) · Journal of the Hong Kong College of Cardiology · doi:10.55503/2790-6744.1091 · Cited by 1
Leth-Olsen, M., Døhlen, G., Torp, H., & Nyrnes, S. A. (2024) · Journal of Clinical Medicine · doi:10.3390/jcm13113115 · Cited by 1 · cited 1×
“In a small, single-center study by Olsen et al., this technology was found to be a potentially valuable tool for real-time cerebral monitoring during transcatheter interventions in infants with CHD [63].”
Ibrahim, H. M., Eldin Mansor, M. G., Kamel, W. I., Salah El Din Soliman, N. M., El Din Abo El Wafa, M. S., & Bakry, N. A. (2025) · International Journal of Child Health and Nutrition · doi:10.6000/1929-4247.2025.14.01.06 · Cited by 1 · cited 1×
“In a more recent investigation, Ibrahim et al. examined the relationship between cardiac output measurements by EC and TTE, and clinical status in infants following cardiac surgery [75].”
Cited by
Li, D., & Yao, D. (2026) · Frontiers in Immunology · doi:10.3389/fimmu.2026.1776897 · Cited by 1
“Based on available pediatric data and expert consensus, we propose following provisional thresholds (7, 93, 132, 137, 146, 147): (cid:129) NT-proBNP/BNP: A change ≥ 30% from baseline (or ≥ 2-fold increase if no baseline available) is considered signi fi cant.”
Nedelcu, C., Cirstoveanu, C. G., Filip, C. I., et al. (2026) · Life · doi:10.3390/life16061001
Self-citation
Metadata
- Authors
- Nedelcu, C., Ionescu, N. S., Bizubac, A. M., Filip, C., & Cirstoveanu, C.
- Affiliations
- Carol Davila University of Medicine and Pharmacy (RO)
- Department of Neonatal Intensive Care, “Carol Davila” University of Medicine and Pharmacy, 020021 Bucharest, Romania
- Corresponding author
- Nedelcu, Carmina
- Ionescu, Nicolae Sebastian
- Venue
- Life
- Publisher
- MDPI AG
- Published
- 2025-10-17
- Volume
- 15
- Issue
- 10
- Pages
- 1621
- ISSN
- 2075-1729 (ISSN-L)
- PMC
- PMC12565398
- OpenAlex
- W4415299593
- Cited by
- 2
- References
- 85
- Citations by year
- 2026: 2
- FWCI
- 0.48
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Life (publishedVersion, cc-by)
- Preprints.org (acceptedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Topics
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Congenital Heart Disease Studies
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- noninvasive hemodynamic monitoring, neonates, neonatal cardiac surgery, postoperative care, circulatory failure, near-infrared spectroscopy, electrical biosensing, multimodal monitoring, regional perfusion, oxygen delivery, lactate, mixed venous oxygen saturation, capillary refill time, urine output, echocardiography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.