Neurological Tremor: Sensors, Signal Processing and Emerging Applications
Open accessThe DOI matches its Crossref record; 78 of the 82 references it cites were found in a registry or an index; 2 of the 70 sources its claims cite were checked for support.
Abstract
Neurological tremor is the most common movement disorder, affecting more than 4% of elderly people. Tremor is a non linear and non stationary phenomenon, which is increasingly recognized. The issue of selection of sensors is central in the characterization of tremor. This paper reviews the state-of-the-art instrumentation and methods of signal processing for tremor occurring in humans. We describe the advantages and disadvantages of the most commonly used sensors, as well as the emerging wearable sensors being developed to assess tremor instantaneously. We discuss the current limitations and the future applications such as the integration of tremor sensors in BCIs (brain-computer interfaces) and the need for sensor fusion approaches for wearable solutions.
Overview
CoChat Summary
## TL;DR
This review surveys sensors and signal-processing methods used to characterize neurological tremor, emphasizing accelerometers, EMG, and emerging wearable/MEMS solutions. It argues that clinical usefulness depends on sensors that are stable, lightweight, and correlated with clinical and functional scores, with sensor fusion and BCI applications highlighted as future directions.
## Key findings
- Neurological tremor affects “more than 4%” of people older than 65 years, and more than two-thirds of people with upper-limb tremor face serious daily-life difficulties.
- The paper identifies accelerometers and EMG as the most commonly used tools for tremor assessment, while noting that wearable sensors, MEMS devices, and hybrid/sensor-fusion approaches are expanding quickly.
- Accelerometers are described as simple and relatively reliable for measuring oscillation frequency and amplitude, but they cannot directly separate linear acceleration from gravity and require calibration.
- The paper notes reported associations between instrumentation and clinical use cases, including:
- accelerometers for estimating Parkinson’s disease symptom severity and for intra-operative DBS electrode placement,
- long-term EMG as a valid and reliable method for quantifying pathological tremor and classifying parkinsonian versus essential tremor,
- haptic/force-sensor devices achieving “high accuracy” in reproducibility of frequency and amplitudes of imposed oscillations.
- For signal processing, the review lists common tremor descriptors such as peak frequency, median frequency, PSD in the 1–20 Hz and 1–33 Hz bands, crest factor, harmonic index, and frequency dispersion; it also highlights wavelets, FFT, autoregression, and EMD as useful methods for nonstationary tremor signals.
## Methods
- This is a narrative review of the state of the art in tremor instrumentation and signal processing; no original experimental sample or dataset is reported.
- The paper synthesizes published evidence on sensors (EMG, accelerometers, gyroscopes, goniometers, video, force sensors, wearable orthoses, optoelectronic devices) and on signal-processing approaches (editing, filtering, spectral estimation, wavelets, cross-spectrum/coherence).
- The analysis is descriptive/comparative, focusing on advantages, disadvantages, and proposed applications rather than a formal meta-analysis.
## Limitations
- The paper states that several future studies are still required, including better sensor fusion approaches for wearable solutions, reliable/stable hybrid sensors, and integration of sensors and actuators into comfortable textiles.
- It also explicitly notes practical limitations of current methods: accelerometers need calibration and cannot distinguish gravity from acceleration without a validated model; video analysis can be time-consuming; haptic devices can be nonportable and expensive; wearable orthoses face aesthetic/cosmetic and affordability issues.
- Evident from the text: no systematic review protocol, search strategy, inclusion criteria, or quantitative synthesis is described, so the review’s scope and completeness cannot be assessed from the paper itself.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures









Verification
Claims the paper makes
2 Valid · 1 Not valid · 56 Not checked · 14 Own results and interpretations
Every source the paper cites for this claim was checked and supports it.
In the paper
“The feasibility of using data recorded with accelerometers to estimate the severity of symptoms and motor complications in patients with Parkinson’s disease has been shown recently [27].”
§ Neurological Tremor: Sensors, Signal Processing and Emerging Applications > Sensors and Instrumentation for the Characterization of Tremor > Accelerometers
What the sources say
- Supportedcites [27] Patel S. Lorincz K. Hughes R. Huggins N. Growdon J. Standaert D. Akay M. Dy J. Welsh M. Bonato P. Monitoring motor fluctuations in patients with Parkinson’s disease using wearable sensors IEEE Trans. Inf. Technol. Biomed 2009 13 864 873 19846382 PMC5432434 source
The Discussion and Conclusions states that the study estimated tremor, bradykinesia, and dyskinesia severity from accelerometer data. The Results section reports estimation-error analyses for those clinical scores, supporting the claimed feasibility in patients with Parkinson’s disease. Page numbers are not provided in our copy. (LLM, conf=0.94)
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“Data from accelerometers are composed of a combination of linear acceleration, gravity, and additive noise [16].”
§ Neurological Tremor: Sensors, Signal Processing and Emerging Applications > Sensors and Instrumentation for the Characterization of Tremor > Accelerometers
What the sources say
- Supportedcites [16] Luinge H.J. Veltink P.H. Measuring orientation of human body segments using miniature gyroscopes and accelerometers Med. Biol. Eng. Comput 2005 43 273 282 15865139 source
Section 2.2 models accelerometer signals as linear acceleration, a gravity contribution, and additive white Gaussian measurement noise. The gravity term is subtracted under the paper’s sign convention, but this supports the claim that all three contribute to the signal. (LLM, conf=0.95)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
In the paper
“Measuring devices based on video imaging are effective in quantifying motor impairments in clinical settings [34].”
§ Neurological Tremor: Sensors, Signal Processing and Emerging Applications > Sensors and Instrumentation for the Characterization of Tremor > Videos
What the sources say
- Not checked — no full text of the sourcecites [34] Swider M. The application of video image processing to quantitative analysis of extremity tremor in humans J. Neurosci. Methods 1998 84 167 172 9821648 source
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
In the paper
“Recent works indicate that these recordings allow a rater-independent classification of parkinsonian versus essential tremor [23].”
§ Neurological Tremor: Sensors, Signal Processing and Emerging Applications > Sensors and Instrumentation for the Characterization of Tremor > Electromyography (EMG)
What the sources say
- Not checked — no full text of the sourcecites [23] Breit S. Spieker S. Schulz J.B. Gasser T. Long-term EMG recordings differentiate between parkinsonian and essential tremor J. Neurol 2008 255 103 11 18204805 source
hedgedstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
In the paper
“However, the videotaped motor examination may present some limitations when patients have mild symptoms [35].”
§ Neurological Tremor: Sensors, Signal Processing and Emerging Applications > Sensors and Instrumentation for the Characterization of Tremor > Videos
What the sources say
- Not checked — no full text of the sourcecites [35] Louis E.D. Levy G. Côte L.J. Mejia H. Fahn S. Marder K. Diagnosing Parkinson’s disease using videotaped neurological examinations: validity and factors that contribute to incorrect diagnoses Mov. Disord 2002 17 513 517 12112199 source
hedgedstates a possible cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
22 more claims
Create a free account to see all 73 claims and every source verdict.
- Probably not a substantial claim. 32
- It may state two separate claims as one. 25
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, kinematic data have been shown to lack precision since they are affected by increased segment acceleration or the presence of drifts (algorithms have been developed to decrease the effect of drifting in accelerometers) [4,5,6].”
From Validity and Reliability of Wearable Sensors for Joint Angle Estimation: A Systematic Review (2019) · cites this paper as [4]
Experimental — accuracy not yet measured
Citations
Cites
Adult Norms for the Box and Block Test of Manual Dexterity
Mathiowetz, V., Volland, G., Kashman, N., & Weber, K. (1985) · The American Journal of Occupational Therapy · doi:10.5014/ajot.39.6.386 · Cited by 2,105 · cited 1×
“C-TES (Clinical-TES) Anamnesis Assessment of disability (Activities of Daily Living scales / ADL-T24) Physical and Neurological examination Tremor evaluation - Body segments involved (head, trunk, upper limbs, lower limbs) - Enhancing/reducing effect (effect of mental calculation/contralateral contractions) - Distribu…”
MethodConsensus Statement of the Movement Disorder Society on Tremor
Deuschl, G., Bain, P., & Brin, M. (2008) · Movement Disorders · doi:10.1002/mds.870131303 · Cited by 1,981
Human movement analysis using stereophotogrammetry
Leardini, A., Chiari, L., Croce, U. D., & Cappozzo, A. (2005) · Gait & Posture · doi:10.1016/j.gaitpost.2004.05.002 · Cited by 1,239 · cited 3×
“…axis of hips, knees and ankles can be determined reliably, but motion about other axes at those joints should be interpreted cautiously, as the soft tissue artefact produces spurious effects with amplitudes comparable to the amount of motion actually occurring in those joints [46].”
BackgroundInfluentialWhat is the Bereitschaftspotential?
Shibasaki, H., & Hallett, M. (2006) · Clinical Neurophysiology · doi:10.1016/j.clinph.2006.04.025 · Cited by 1,169 · cited 2×
“The early BP might reflect slowly increasing cortical excitability and subconscious readiness for the forthcoming movement [64].”
BackgroundAdult Norms for the Nine Hole Peg Test of Finger Dexterity
Mathiowetz, V., Weber, K., Kashman, N., & Volland, G. (1985) · The Occupational Therapy Journal of Research · doi:10.1177/153944928500500102 · Cited by 972 · cited 1×
“C-TES (Clinical-TES) Anamnesis Assessment of disability (Activities of Daily Living scales / ADL-T24) Physical and Neurological examination Tremor evaluation - Body segments involved (head, trunk, upper limbs, lower limbs) - Enhancing/reducing effect (effect of mental calculation/contralateral contractions) - Distribu…”
MethodMeasuring orientation of human body segments using miniature gyroscopes and accelerometers
Luinge, H. J., & Veltink, P. H. (2005) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345966 · Cited by 647 · cited 1×
“Data from accelerometers are composed of a combination of linear acceleration, gravity, and additive noise [16].”
BackgroundMonitoring Motor Fluctuations in Patients With Parkinson's Disease Using Wearable Sensors
Patel, S., Lorincz, K., Hughes, R., Huggins, N., Growdon, J., Standaert, D., Akay, M., Dy, J., Welsh, M., & Bonato, P. (2009) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2009.2033471 · Cited by 623 · cited 1×
“The feasibility of using data recorded with accelerometers to estimate the severity of symptoms and motor complications in patients with Parkinson's disease has been shown recently [27].”
MethodRoetenberg, D., Luinge, H. J., Baten, C. T. M., & Veltink, P. H. (2005) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2005.847353 · Cited by 588 · cited 1×
“Low-pass filtering is commonly used [25].”
Human movement analysis using stereophotogrammetry
Chiari, L., Croce, U. D., Leardini, A., & Cappozzo, A. (2005) · Gait & Posture · doi:10.1016/j.gaitpost.2004.04.004 · Cited by 441 · cited 4×
“Recognition of passive markers can be performed either via recognition software or with a dedicated hardware circuit [42-44].”
BackgroundMethodInfluentialSalarian, A., Russmann, H., Wider, C., Burkhard, P. R., Vingerhoets, F. J. G., & Aminian, K. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2006.886670 · Cited by 416 · cited 1×
“Flexible angular sensors are commonly used in rehabilitation and sports, but have not been disseminated at a large scale for studies of tremor [8,31,32].”
Background
Research works printed but not linked: 1
- Avizzano C. A. Bergamasco M. Technological aids for the treatment of the tremor International Conference on Rehabilitation Robotics (ICORR), Stanford University Stanford, CA, USA 1999
Cited by
Items 101–140 of 140
A Smartphone-Based Clinical Decision Support System for Tremor Assessment
Zamora, G., Fuchs, C., Degeneffe, A., Kubben, P., & Kaymak, U. (2020) · Lecture notes in computer science · doi:10.1007/978-3-030-63061-4_1 · Cited by 3
MODEL OF AN ADAPTIVE ENERGY HARVESTER WITH ELECTRO-RHEOLOGICAL FLUID
Teodorescu, H.-N. (2015) · Cited by 1
Distributed Monitoring and Supervising System for E-Health Applications
Folea, S. C., Hulea, M., Avram, C. C., & Aştilean, A. (2012) · Advances in healthcare information systems and administration book series · doi:10.4018/978-1-4666-0888-7.ch011 · Cited by 1
Towards a multi-sensor system for the diagnosis of neurological disorders
Tenner, F., Schramm, A., Sohle, M., Regensburger, M., Wirthmann, E., Zalevsky, Z., & Schmidt, M. (2016) · IEEE International Conference on Advanced Intelligent Mechatronics (AIM) · doi:10.1109/aim.2016.7576816 · Cited by 2
“fingers, hands, limbs, head etc) [1], fasciculations are irregular, short and circumscribed muscle twitches which are generated by spontaneous involuntary discharges of single motor units [2].”
BackgroundPRASAD, G., Siddaiah, N., Rao, K. S., & Babu, P. S. S. (2017) · International Journal of Pharma and Bio Sciences · doi:10.22376/ijpbs.2017.8.1.b390-395 · Cited by 2
Evaluation of smartphone usage in neurological pathologies diagnosis
Cardellicchio, A., Dario, R., Lecce, V. D., Guaragnella, C., Lombardi, A., Mongelli, L., Quarto, A., & Soldo, D. (2017) · IEEE Workshop on Environmental, Energy, and Structural Monitoring Systems (EESMS) · doi:10.1109/eesms.2017.8052697 · Cited by 1
“For a comprehensive review of these methods, refer to [11].”
BackgroundQUANTITATIVE ANALYSIS OF ESSENTIAL TREMOR DURING CLINICAL SPIRAL DRAWING TASK USING GYRO SENSORS
Kwon, Y., Eom, G., Ko, J., & Kim, J.-W. (2021) · Journal of Mechanics in Medicine and Biology · doi:10.1142/s0219519421400509 · Cited by 2
Does Cup-Grip Type Affect Tremor among People with Essential Tremor?
Roth, N., & Rosenblum, S. (2021) · Sensors · doi:10.3390/s21237797 · Cited by 1
Study of a measuring and attenuation device prototype for essential and Parkinsonian tremors
Molina, T., Castilho, M., Tsurumaki, A., et al. (2020) · The Academic Society Journal · doi:10.32640/tasj.2020.1.20 · Cited by 1
Validation study of low-cost sensors
Nandy, A., Chakraborty, S., Chakraborty, J., & Venture, G. (2021) · Elsevier eBooks · doi:10.1016/b978-0-323-85245-6.00005-9 · Cited by 1
Towards an automated weight lifting coach: introducing LIFT
Lady, M. A. (2014) · DigitalCommons - CalPoly (California State Polytechnic University) · doi:10.15368/theses.2014.87 · Cited by 1
Wireless Health Monitoring using Passive WiFi SensingPreprint
Khan, U. M., Kabir, Z., & Hassan, S. A. (2017) · arXiv (Cornell University) · doi:10.48550/arxiv.1704.00620 · Cited by 1
“Essential tremor monitoring is mostly unexplored in the wireless domain and is typically done using electromyography sensors, accelerometers and gyroscopes [18].”
Optimization of passive cantilever-type tuned mass damper to reduce the hand postural tremor
Gebai, S. (2020) · theses.fr (ABES) · doi:10.70675/994e0848ze136z4e48z8db8z9981215dcb5b · Cited by 5
Research of hand tremor vibrations and interference with external me-chanical oscillations source
Venslauskas, M., Litvinas, E., Juknevičius, A. R., et al. (2016) · Mechanika · doi:10.5755/j01.mech.22.5.16351
Grimaldi, G. (2016) · Springer eBooks · doi:10.1007/978-3-319-24551-5_66
Self-citationAssessment of Essential Tremors on Clinical and Functional Performance Tests
Kumar, M., Kumar, D., & Tewary, K. (2016) · Journal of Neurology and Neuroscience · doi:10.21767/2171-6625.1000157
“Effect of tremor on daily living activities (ADL) can be evaluated using specific questionnaires such as the ADL-T24, which has a good inter-session reproducibility [4].”
Mesa, A. R., Barco-Diaz, C., Torres, R., & Jaramillo, V. (2017) · IEEE Colombian Conference on Automatic Control (CCAC) · doi:10.1109/ccac.2017.8276437
“There are diverse methods for the characterization of tremors through signal acquisition: electromyography (EMG), accelerometers, gyroscopes and force sensors [10].”
MethodStroke Rehabilitation and Parkinson's Disease Tremor Reduction Using BCIs Combined With FES
Adama, S. V., & Bogdan, M. (2018) · International Journal of Privacy and Health Information Management · doi:10.4018/ijphim.2018010102 · Cited by 1
The Effect of Tactile Biofeedback on the Wrist to Suppress Hand Tremors
Gorospe, Y. (2020) · Advances in intelligent systems and computing · doi:10.1007/978-3-030-51828-8_118
Therapeutic maps for a sensor-based evaluation of deep brain stimulation programming
Bremm, R. P., Berthold, C., Krüger, R., Koch, K. P., Gonçalves, J. M., & Hertel, F. (2021) · Biomedizinische Technik/Biomedical Engineering · doi:10.1515/bmt-2020-0210
Frequency Detection Of Hand Tremors Using Customized Accelerometer
Al-Kazwini, A. T., Halilah, A., Hendi, A., Al-Saeed, I., Al-Azamat, A., & Al-Zyoud, W. A. (2021) · IEEE National Biomedical Engineering Conference (NBEC) · doi:10.1109/nbec53282.2021.9618761
Stroke Rehabilitation and Parkinson's Disease Tremor Reduction Using BCIs Combined With FES
Adama, S. V., & Bogdan, M. (2020) · IGI Global eBooks · doi:10.4018/978-1-7998-3432-8.ch033
Wearing sensors inside and outside of the human body for the early detection of diseases
Zheng, Y., Jiang, Y., Luo, N., Ding, X., Zhang, Y., & Poon, C. C. Y. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-819246-7.00003-6 · Cited by 1
Stroke Rehabilitation and Parkinson's Disease Tremor Reduction Using BCIs Combined With FES
Adama, S. V., & Bogdan, M. (2020) · IGI Global eBooks · doi:10.4018/978-1-7998-3441-0.ch024
Involuntary breathing movement pattern recognition and classification via force based sensorsPreprint
Singh, R. E., Fleury, J. M., Gupta, S., et al. (2022) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2022.07.12.499777
“A sensor review study validates it [17].”
BackgroundGrimaldi, G. (2023) · Springer eBooks · doi:10.1007/978-3-031-15070-8_72
Self-citationKovács, A., & Kamondi, A. (2023) · Contemporary clinical neuroscience · doi:10.1007/978-3-031-26128-2_11
McNames, J. (2023) · Contemporary clinical neuroscience · doi:10.1007/978-3-031-26128-2_17
Deep Learning Models for Parkinson’s Disease Severity Evaluation
Channa, A., & Popescu, N. (2023) · SpringerBriefs in computer science · doi:10.1007/978-3-031-45003-7_4 · Cited by 1
Foot Clearance Prediction using Wrist Acceleration and Gait Speed
Kitagawa, K., Wada, C., & Toya, N. (2023) · Journal of Electronics Electromedical Engineering and Medical Informatics · doi:10.35882/jeeemi.v6i1.345
“In the case using actual accelerometer, accuracy of the proposed method might be decreased from this simulation study since accelerometer has specific errors for dynamic measurement [47, 48].”
MethodMorikawa, T., Mura, N., Sato, T., & Katoh, H. (2024) · Bio-Medical Materials and Engineering · doi:10.3233/bme-240031 · Cited by 5
“However, it tends to be less reliable due to the drift effect, although, various algorithms have been developed to minimize this [19–21].”
Fall Detection Wearables in Parkinson's Disease Patients
Ikenaga, S. (2024) · doi:10.32920/26866591.v1
Essential Tremor Severity Classification Using a Multi-Layer Perceptron and the TETRAS Scale
Ting, J., Basyal, S., Mishra, K., Rose, C. G., & Allen, B. C. (2024) · IEEE/RAS-EMBS International Conference on Biomedical Robotics and Biomechatronics (BioRob) · doi:10.1109/biorob60516.2024.10719853 · Cited by 1
“The diagnosis of ET severity often relies on subjective methods, such as simple clinical observation, standardized rating scales, or the subjective assessment of drawn figures [4].”
BackgroundVibrational Activation of Blood Flow
Ostaševičius, V., Jūrėnas, V., Venslauskas, M., & Kižauskienė, L. (2025) · Synthesis lectures on biomedical engineering · doi:10.1007/978-3-031-79025-6_3
Ouellet, R., Turcot, K., Séguin, N., Campeau‐Lecours, A., & Bouffard, J. (2025) · Sensors · doi:10.3390/s25061680
Vaida, C. L., Vanta, O. M., Rus, G., et al. (2025) · Bioengineering · doi:10.3390/bioengineering12121340
“Classical approaches employ digital filtering, adaptive Kalman smoothing, and wavelet-based denoising to reduce involuntary oscillations without degrading voluntary motion [18,19].”
Lavalley, A., Lorton, O., Constanthin, P. E., Fleury, V., & Momjian, S. (2026) · Stereotactic and Functional Neurosurgery · doi:10.1159/000552899
Preeti, M., Sastry, A., Guha, K., Iannacci, J., & Baishnab, K. L. (2026) · Lecture notes in electrical engineering · doi:10.1007/978-981-92-0217-1_5
Interrater reliability of Tremor CharacterizationPreprint
Joshi, P. N., Borges, Y. F. F., Pandey, S., Vidailhet, M., Fung, V. S. C., Hallett, M., Jinnah, H. A., Shaikh, A. G., & Fasano, A. A. (2026) · medRxiv · doi:10.64898/2026.09.21.26363551
Metadata
- Authors
- Grimaldi, G., & Manto, M.
- Affiliations
- FNRS, Neurologie ULB-Erasme, 808 Route de Lennik, 1070 Bruxelles, Belgium
- Corresponding author
- Manto, Mario
- ORCID
- Manto, Mario
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2010-02-24
- Volume
- 10
- Issue
- 2
- Pages
- 1399-1422
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC3244020
- OpenAlex
- W2123354643
- Cited by
- 138
- References
- 82
- Citations by year
- 2026: 4 · 2025: 5 · 2024: 5 · 2023: 7 · 2022: 12 · 2021: 17 · 2020: 19 · 2019: 8 · 2018: 7 · 2017: 12 · 2016: 10 · 2015: 5 · 2014: 7 · 2013: 8 · 2012: 9
- FWCI
- 0.46
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Dépôt institutionnel de l'Université libre de Bruxelles (Université Libre de Bruxelles) (submittedVersion, other-oa)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- Fonds De La Recherche Scientifique - FNRS
- Topics
- Life Sciences › Neuroscience › Cognitive Neuroscience › Motor Control and Adaptation
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Context-Aware Activity Recognition Systems
- Life Sciences › Neuroscience › Cognitive Neuroscience › EEG and Brain-Computer Interfaces
- Keywords
- movement disorders, wearable sensors, sensor fusion, brain-computer interface, signal processing, non-stationary signals, older adults
- MeSH terms
- Biofeedback, Psychology, Electrodes, Humans, Man-Machine Systems, Movement, Signal Processing, Computer-Assisted, Therapy, Computer-Assisted, Tremor, User-Computer Interface
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.