Optimal Magnetic Sensor Vests for Cardiac Source Imaging
Open accessThe DOI matches its Crossref record; 74 of the 75 references it cites were found in a registry or an index.
Abstract
Magnetocardiography (MCG) non-invasively provides functional information about the heart. New room-temperature magnetic field sensors, specifically magnetoresistive and optically pumped magnetometers, have reached sensitivities in the ultra-low range of cardiac fields while allowing for free placement around the human torso. Our aim is to optimize positions and orientations of such magnetic sensors in a vest-like arrangement for robust reconstruction of the electric current distributions in the heart. We optimized a set of 32 sensors on the surface of a torso model with respect to a 13-dipole cardiac source model under noise-free conditions. The reconstruction robustness was estimated by the condition of the lead field matrix. Optimization improved the condition of the lead field matrix by approximately two orders of magnitude compared to a regular array at the front of the torso. Optimized setups exhibited distributions of sensors over the whole torso with denser sampling above the heart at the front and back of the torso. Sensors close to the heart were arranged predominantly tangential to the body surface. The optimized sensor setup could facilitate the definition of a standard for sensor placement in MCG and the development of a wearable MCG vest for clinical diagnostics.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures









Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Large Magnetoresistance at Room Temperature in Ferromagnetic Thin Film Tunnel Junctions
Moodera, J. S., Kinder, L. R., Wong, T. M., & Meservey, R. (1995) · Physical Review Letters · doi:10.1103/physrevlett.74.3273 · Cited by 3,707
Budker, D., & Romalis, M. (2007) · Nature Physics · doi:10.1038/nphys566 · Cited by 1,916
Clustering by means of medoids
Kaufman, L., & Rousseeuw, P. J. (1987) · Lirias · Cited by 1,311
Statistical Analysis of Spherical Data
Fisher, N. I., Lewis, T., & Embleton, B. J. J. (1987) · Cambridge University Press eBooks · doi:10.1017/cbo9780511623059 · Cited by 1,013
Ultrahigh sensitivity magnetic field and magnetization measurements with an atomic magnetometer
Dang, H. B., Maloof, A. C., & Romalis, M. V. (2010) · Applied Physics Letters · doi:10.1063/1.3491215 · Cited by 891 · cited 1×
“While many experimental OPMs are 1–5 cm large to increase sensitivity [9,12,30,31,34,35], it is possible to miniaturize them to 5 mm diameter [28,29] and to reduce power consumption to approximately 5 mW [28].”
BackgroundGallagher, J. J., Pritchett, E. L. C., Sealy, W. C., Kasell, J., & Wallace, A. G. (1978) · Progress in Cardiovascular Diseases · doi:10.1016/0033-0620(78)90015-4 · Cited by 757
Statistical Analysis of Spherical Data.
Jupp, P. E., Fisher, N. I., Lewis, T., & Embleton, B. J. J. (1988) · Biometrics · doi:10.2307/2531880 · Cited by 704
Freitas, P. P., Ferreira, R., Cardoso, S., & Cardoso, F. (2007) · Journal of Physics: Condensed Matter · doi:10.1088/0953-8984/19/16/165221 · Cited by 422
A compact, high performance atomic magnetometer for biomedical applications
Shah, V. K., & Wakai, R. T. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/22/8153 · Cited by 322 · cited 1×
“While many experimental OPMs are 1–5 cm large to increase sensitivity [9,12,30,31,34,35], it is possible to miniaturize them to 5 mm diameter [28,29] and to reduce power consumption to approximately 5 mW [28].”
BackgroundMagnetoencephalography with a chip-scale atomic magnetometer
Sander, T. H., Preusser, J., Mhaskar, R., Kitching, J., Trahms, L., & Knappe, S. (2012) · Biomedical Optics Express · doi:10.1364/boe.3.000981 · Cited by 293
Limited Lead Selection for Estimation of Body Surface Potential Maps in Electrocardiography
Lux, R. L., Smith, C. R., Wyatt, R. F., & Abildskov, J. A. (1978) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1978.326332 · Cited by 216 · cited 2×
“This quantity is a comparable to previous MCG studies [50–52] and ECG studies [43,45,46].”
BackgroundResultMagnetism in Medicine: A Handbook
(2006) · Wiley eBooks · doi:10.1002/9783527610174 · Cited by 200
Scaling for Numerical Stability in Gaussian Elimination
Skeel, R. D. (1979) · Journal of the ACM · doi:10.1145/322139.322148 · Cited by 192 · cited 2×
“The second goal function is the Skeel condition number, proposed by Skeel [55] for squ matrices and then generalized to rectangular matrices [56].”
BackgroundA room temperature 19-channel magnetic field mapping device for cardiac signals
Bison, G., Castagna, N., Hofer, A., Knowles, P., Schenker, J.-L., Kasprzak, M., Saudan, H., & Weis, A. (2009) · Applied Physics Letters · doi:10.1063/1.3255041 · Cited by 186
Femtotesla atomic magnetometry in a microfabricated vapor cell
Griffith, W. C., Knappe, S., & Kitching, J. (2010) · Optics Express · doi:10.1364/oe.18.027167 · Cited by 183 · cited 4×
“Consequently, a sensor footprint of 1 cm or less [8,18,23,28] should allow for sufficient spatial sampling density.”
BackgroundMethodInfluentialSelection of the Number and Positions of Measuring Locations for Electrocardiography
Barr, R. C., Spach, M. S., & Herman-Giddens, G. S. (1971) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1971.4502813 · Cited by 182
Alem, O., Sander, T. H., Mhaskar, R., LeBlanc, J., Eswaran, H., Steinhoff, U., Okada, Y., Kitching, J., Trahms, L., & Knappe, S. (2015) · Physics in Medicine and Biology · doi:10.1088/0031-9155/60/12/4797 · Cited by 182
Lux, R. L., Evans, A. K., Burgess, M. J., Wyatt, R. F., & Abildskov, J. A. (1981) · Circulation Research · doi:10.1161/01.res.49.1.186 · Cited by 163
Optical magnetometer array for fetal magnetocardiography
Wyllie, R., Kauer, M., Wakai, R. T., & Walker, T. G. (2012) · Optics Letters · doi:10.1364/ol.37.002247 · Cited by 148 · cited 1×
“While many experimental OPMs are 1–5 cm large to increase sensitivity [9,12,30,31,34,35], it is possible to miniaturize them to 5 mm diameter [28,29] and to reduce power consumption to approximately 5 mW [28].”
BackgroundKnappe, S., Sander, T. H., Kosch, O., Wiekhorst, F., Kitching, J., & Trahms, L. (2010) · Applied Physics Letters · doi:10.1063/1.3491548 · Cited by 122 · cited 1×
“While many experimental OPMs are 1–5 cm large to increase sensitivity [9,12,30,31,34,35], it is possible to miniaturize them to 5 mm diameter [28,29] and to reduce power consumption to approximately 5 mW [28].”
BackgroundEdelstein, A. S., Fischer, G., Pedersen, M., Nowak, E. R., Cheng, S. F., & Nordman, C. A. (2006) · Journal of Applied Physics · doi:10.1063/1.2170067 · Cited by 111
Magnetic tunnel junction sensors with pTesla sensitivity
Cardoso, S., Leitão, D. C., Gameiro, L., Cardoso, F. A., Ferreira, R., Paz, E., & Freitas, P. P. (2014) · Microsystem Technologies · doi:10.1007/s00542-013-2035-1 · Cited by 94 · cited 3×
“Consequently, a sensor footprint of 1 cm or less [8,18,23,28] should allow for sufficient spatial sampling density.”
BackgroundMagnetoencephalography: From Signals to Dynamic Cortical Networks
(2014) · Springer eBooks · doi:10.1007/978-3-642-33045-2 · Cited by 84
Kornreich, F., Rautaharju, P. M., Warren, J. D., Montague, T. J., & Horáček, B. M. (1985) · The American Journal of Cardiology · doi:10.1016/0002-9149(85)90768-4 · Cited by 79 · cited 1×
“[47] added that the right sub-clavicular and left posterior axillary regions are diagnostically valuable.”
BackgroundMultichannel SQUID system detecting tangential components of the cardiac magnetic field
Tsukada, K., Haruta, Y., Adachi, A., Ogata, H., Komuro, T., Ito, T., Takada, Y., Kandori, A., Noda, Y., Terada, Y., et al. (1995) · Review of Scientific Instruments · doi:10.1063/1.1146524 · Cited by 69
Schultze, V., IJsselsteijn, R. P. J., Scholtes, T., Woetzel, S., & Meyer, H. (2012) · Optics Express · doi:10.1364/oe.20.014201 · Cited by 55
Optical multichannel room temperature magnetic field imaging system for clinical application
Lembke, G., Erné, S. N., Nowak, H., Menhorn, B., Pasquarelli, A., & Bison, G. (2014) · Biomedical Optics Express · doi:10.1364/boe.5.000876 · Cited by 53 · cited 1×
“During signal processing, residual interferences may be further eliminated by spectral filtering [79], averaging of heartbeats [32], signal decomposition approaches, and common mode rejection.”
BackgroundPaz, E., Serrano-Guisan, S., Ferreira, R., & Freitas, P. P. (2014) · Journal of Applied Physics · doi:10.1063/1.4859036 · Cited by 50
Improved Error Bounds for Underdetermined System Solvers
Demmel, J. W., & Higham, N. J. (1993) · SIAM Journal on Matrix Analysis and Applications · doi:10.1137/0614001 · Cited by 50
Non-invasive imaging of arrhythmogenic left-ventricular myocardium after infarction
Leder, U., Haueisen, J., Huck, M., & Nowak, H. (1998) · The Lancet · doi:10.1016/s0140-6736(98)00082-8 · Cited by 48
Self-citationOptically-Pumped Magnetometers for MEG
Knappe, S. A., Sander, T. H., & Trahms, L. (2014) · Springer eBooks · doi:10.1007/978-3-642-33045-2_49 · Cited by 47
Valadeiro, J. P., Amaral, J., Leitão, D. C., Ferreira, R., Cardoso, S., & Freitas, P. P. (2015) · IEEE Transactions on Magnetics · doi:10.1109/tmag.2014.2352115 · Cited by 45 · cited 4×
“Consequently, a sensor footprint of 1 cm or less [8,18,23,28] should allow for sufficient spatial sampling density.”
BackgroundInfluentialKamada, K., Ito, Y., Ichihara, S., Mizutani, N., & Kobayashi, T. (2015) · Optics Express · doi:10.1364/oe.23.006976 · Cited by 43 · cited 3×
“Current developments aim at increasing the dynamic range, developing vectorial OPMs [36], and integrating the OPM technology into multi-channel systems [27].”
BackgroundA sensor configuration for a 304 SQUID vector magnetometer.
Schnabel, A., Burghoff, M., Hartwig, S., Petsche, F., Steinhoff, U., Drung, D., & Koch, H. (2004) · PubMed · Cited by 43
Magnetoencephalography with diversely oriented and multicomponent sensors
Hochwald, B. M., & Nehorai, A. (1997) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.553711 · Cited by 40 · cited 1×
“A similar observation was reported by Hochwald and Nehorai [71] for the magnetoencephalography-based reconstruction of cortical sources in the center of a spherical magnetoencephalography sensor setup.”
MethodTabu Search Optimization of Magnetic Sensor Systems for Magnetocardiography
Lau, S., Eichardt, R., Rienzo, L. D., & Haueisen, J. (2008) · IEEE Transactions on Magnetics · doi:10.1109/tmag.2007.915911 · Cited by 38 · cited 4×
“The algorithm is described in [53] and implemented as an extension of the software toolbox SimBio [59].”
BackgroundMethodInfluentialSelf-citationEffect of torso boundaries on electric potential and magnetic field of a dipole
Purcell, C. J., Stroink, G., & Horacek, B. M. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.7268 · Cited by 36
Finlay, D., Nugent, C., Donnelly, M. P., Lux, R. L., McCullagh, P. J., & Black, N. D. (2006) · BMC Medical Informatics and Decision Making · doi:10.1186/1472-6947-6-9 · Cited by 32
Hren, R., Zhang, X., & Stroink, G. (1996) · Medical & Biological Engineering & Computing · doi:10.1007/bf02520014 · Cited by 32
Schneider, U., Haueisen, J., Loeff, M., Bondarenko, N., & Schleußner, E. (2005) · Prenatal Diagnosis · doi:10.1002/pd.1205 · Cited by 32 · cited 1×
“A common approach is to measure the gradient of the magnetic field with gradiometers, thereby eliminating most external interference [76,77].”
BackgroundSelf-citationIntegrating magnetoresistive sensors with microelectromechanical systems for noise reduction
Hu, J., Pan, M., Tian, W., Chen, D., & Luo, F. (2012) · Applied Physics Letters · doi:10.1063/1.4769903 · Cited by 26
Information Content in Single-Component Versus Three-Component Cardiomagnetic Fields
Arturi, C. M., Rienzo, L. D., & Haueisen, J. (2004) · IEEE Transactions on Magnetics · doi:10.1109/tmag.2004.824891 · Cited by 26
Self-citationKamada, K., Taue, S., & Kobayashi, T. (2011) · Japanese Journal of Applied Physics · doi:10.1143/jjap.50.056602 · Cited by 25 · cited 1×
“While many experimental OPMs are 1–5 cm large to increase sensitivity [9,12,30,31,34,35], it is possible to miniaturize them to 5 mm diameter [28,29] and to reduce power consumption to approximately 5 mW [28].”
BackgroundLead selection: old and new methods for locating the most electrocardiogram information
Donnelly, M. P., Finlay, D., Nugent, C., & Black, N. D. (2008) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2008.02.004 · Cited by 25 · cited 2×
“The ECG investigations concur in the finding that sets of diagnostically optimal electrodes can be found, but that there is no unique optimal arrangement [49].”
BackgroundReproducibility of HTS-SQUID magnetocardiography in an unshielded clinical environment
LEDER, U., Schrey, F., Haueisen, J., Dörrer, L., Schreiber, J., Liehr, M., Schwarz, G., Solbrig, O., Figulla, H. R., & Seidel, P. (2001) · International Journal of Cardiology · doi:10.1016/s0167-5273(01)00440-5 · Cited by 24 · cited 1×
“A common approach is to measure the gradient of the magnetic field with gradiometers, thereby eliminating most external interference [76,77].”
BackgroundSelf-citationOptimal sensor distribution for measuring the tangential field components in MCG.
Kim, K., Lee, Y. H., Kwon, H. C., Kim, J. M., Kim, I. S., & Park, Y. K. (2004) · PubMed · Cited by 23
Magnetic-Field Dependence of the Noise in a Magnetoresistive Sensor Having MEMS Flux Concentrators
Ozbay, A., Nowak, E. R., Edelstein, A. S., Fischer, G., Nordman, C. A., & Cheng, S. F. (2006) · IEEE Transactions on Magnetics · doi:10.1109/tmag.2006.879752 · Cited by 22 · cited 2×
“The flux guides are made to mechanically oscillate at frequencies above 10 kHz [15].”
BackgroundReconstruction of two-dimensional current distribution from tangential MCG measurement
Kandori, A., Tsukada, K., Haruta, Y., Noda, Y., Terada, Y., Mitsui, T., & Sekihara, K. (1996) · Physics in Medicine and Biology · doi:10.1088/0031-9155/41/9/010 · Cited by 19 · cited 1×
“[68] based on measurements with their custom tangential sensor system further advocate the value of tangential sensor orientations.”
Noninvasive Imaging of Cardiac Excitation: Current Status and Future Perspective
van der Graaf, A. W. M., Bhagirath, P., Ramanna, H., van Driel, V. J. H. M., de Hooge, J., de Groot, N. M. S., & Götte, M. (2014) · Annals of Noninvasive Electrocardiology · doi:10.1111/anec.12140 · Cited by 19
Wang, Y., Li, J., & Viehland, D. D. (2015) · Journal of Applied Physics · doi:10.1063/1.4936263 · Cited by 19 · cited 2×
“Currently, active research aims to reduce this noise with: (1) new materials, sensor designs and fabrication methods [16,17]; (2) combining series of sensor elements [6,18]; and (3) using microelectromechanical flux guides [18–24].”
BackgroundThree component magnetic field data
Rienzo, L. D., Haueisen, J., & Arturi, C. M. (2005) · COMPEL The International Journal for Computation and Mathematics in Electrical and Electronic Engineering · doi:10.1108/03321640510598193 · Cited by 18
Self-citationOptically pumped alkali magnetometers for biomedical applications
Weis, A. (2012) · Europhysics news · doi:10.1051/epn/2012301 · Cited by 17 · cited 1×
“An OPM utilizes the Zeeman Effect using a cell filled with a heated gaseous alkali metal, such as cesium or potassium [9,10,13,27].”
MethodMagnetic noise in a low-power picotesla magnetoresistive sensor
Liou, S. H., Sellmyer, D. J., Russek, S. E., Heindl, R., Silva, F. C. S. D., Moreland, J., Pappas, D. P., Yuan, L., & Shen, J. (2009) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/icsens.2009.5398414 · Cited by 16 · cited 1×
“Chip-scale MR sensors are well suited for a magnetic sensor vest, because they are less than or around 1 cm in diameter [8,18,23,25], require only a few μW of power [8,14,17,25,26], are electrically safe with operating voltages around 1 V [14,20,25,26], emit little heat, and can be mass-produced at low cost [17].”
BackgroundTian, W., Hu, J., Pan, M., Chen, D., & Zhao, J. (2013) · IEEE Transactions on Magnetics · doi:10.1109/tmag.2013.2264061 · Cited by 15
Comparison of sensor arrangements of MCG and ECG with respect to information content
Nalbach, M., & Dössel, O. (2002) · Physica C Superconductivity · doi:10.1016/s0921-4534(02)00683-4 · Cited by 15
Reduction of magnetic 1/f noise in miniature anisotropic magnetoresistive sensors
Guo, Y., Wang, J., White, R. M., & Wang, S. X. (2015) · Applied Physics Letters · doi:10.1063/1.4921764 · Cited by 14 · cited 1×
“Currently, active research aims to reduce this noise with: (1) new materials, sensor designs and fabrication methods [16,17]; (2) combining series of sensor elements [6,18]; and (3) using microelectromechanical flux guides [18–24].”
BackgroundTuning magnetic nanostructures and flux concentrators for magnetoresistive sensors
Yin, X., Liu, Y.-F., Ewing, D., Ruder, C. K., Rego, P. J. D., Edelstein, A. S., & Liou, S. H. (2015) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2188712 · Cited by 13 · cited 3×
“Tunnel junction MR sensors, for example, can have 1/f noise of 47 pT/Hz1/2 at 1 Hz [14].”
BackgroundMagnetic tunnel junction sensors with pTesla sensitivity for biomedical imaging
Cardoso, S., Gameiro, L., Leitão, D. C., Cardoso, F. A., Ferreira, R., Paz, E., & Freitas, P. P. (2013) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2018070 · Cited by 13 · cited 1×
“Available MR sensors are already capable of detecting a heartbeat [4,6,7,18,23].”
BackgroundEichardt, R., Baumgarten, D., Petković, B. R., Wiekhorst, F., Trahms, L., & Haueisen, J. (2012) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-012-0950-4 · Cited by 12
Self-citationDössel, O., Schneider, F., & Muller, M. (2002) · Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.1998.745825 · Cited by 12 · cited 1×
“[48] using a finite element simulation showed that a numerical optimization of electrode positions maximizing the condition of the lead field matrix produces comparable electrode arrangements.”
MethodComani, S., Mantini, D., Alleva, G., Luzio, S. D., & Romani, G. L. (2005) · Physics in Medicine and Biology · doi:10.1088/0031-9155/50/23/006 · Cited by 11 · cited 1×
“During signal processing, residual interferences may be further eliminated by spectral filtering [79], averaging of heartbeats [32], signal decomposition approaches, and common mode rejection.”
BackgroundDiekmann, V., Becker, W., Grözinger, B., Jürgens, R., & Kornhuber, C. (1991) · Clinical Physics and Physiological Measurement · doi:10.1088/0143-0815/12/a/011 · Cited by 11 · cited 1×
“[67] who showed in simulations and measurements that the SNR can be improved by a factor of up to 20 if the total field, including the tangential components above the heart, was considered rather than only radial components.”
BackgroundVisualization of atrial excitation by magnetocardiogram
Sato, M., Terada, Y., Mitsui, T., Miyashita, T., Kandori, A., & Tsukada, K. (2002) · International journal of cardiac imaging · doi:10.1023/a:1015597910933 · Cited by 10 · cited 1×
“[63], add that even the smaller atrial excitation can be observed with MCG sensors at the back of the torso.”
BackgroundMultichannel fetal magnetocardiography using SQUID bootstrap circuit
Zhang, S., Zhang, G., Wang, Y., Liu, M., Li, H., Qiu, Y., Zeng, J., Kong, X., & Xie, X. (2013) · Chinese Physics B · doi:10.1088/1674-1056/22/12/128501 · Cited by 9
Jazbinšek, V., Hren, R., & Trontelj, Z. (2007) · International Congress Series · doi:10.1016/j.ics.2007.01.060 · Cited by 8 · cited 3×
“[50,51] simulated and measured the magnetic field of the heart at the front and back of the torso.”
MethodSchneider, F., Dössel, O., & Muller, M. (2002) · Computers in cardiology · doi:10.1109/cic.1998.731967 · Cited by 8 · cited 2×
“[64] reported necessary sensor distances from 10 mm at the apex to 100 mm at less informative locations.”
BackgroundAchieving 1/f noise reduction with the MEMS flux concentrator
Edelstein, A. S., Fischer, G., Burnette, J. E., Egelhoff, W. E., & Cheng, S. F. (2009) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/icsens.2009.5398411 · Cited by 7 · cited 2×
“Given the prototypes and simulations, the detectivity is expected to reach a few pT/Hz1/2 at 1 Hz at room temperature [8,19,21].”
BackgroundLee, Y.-H., Kim, J. M., Kim, K., Kwon, H.-C., Kim, I. S., Park, Y. K., Ko, Y., & Chung, N. (2005) · IEEE Transactions on Applied Superconductivity · doi:10.1109/tasc.2005.849983 · Cited by 7 · cited 1×
“[65,69,70] point out that the tangential components of the MCG have a higher confidence of the source reconstruction of deep sources than radial components.”
BackgroundNovel magnetic nanostructured multilayer for high sensitive magnetoresistive sensor
Yin, X., & Liou, S. H. (2012) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/icsens.2012.6411027 · Cited by 6 · cited 1×
“Chip-scale MR sensors are well suited for a magnetic sensor vest, because they are less than or around 1 cm in diameter [8,18,23,25], require only a few μW of power [8,14,17,25,26], are electrically safe with operating voltages around 1 V [14,20,25,26], emit little heat, and can be mass-produced at low cost [17].”
BackgroundRienzo, L. D., & Haueisen, J. (2007) · COMPEL The International Journal for Computation and Mathematics in Electrical and Electronic Engineering · doi:10.1108/03321640710727719 · Cited by 6 · cited 1×
“A complete sensor arrangement for measuring cardiac magnetic fields would consist of three-axial sensors lying on the faces of a box enclosing the whole torso [37].”
BackgroundSelf-citationPerformance Investigation of a Three-Dimensional SQUID MCG System by Using a Computer Simulation
Kang, C. S., Lee, S.-G., Kim, K., Kwon, H., Lee, Y., Yu, K. K., Lim, H. K., & Park, Y.-K. (2008) · Journal of the Korean Physical Society · doi:10.3938/jkps.53.3444 · Cited by 4 · cited 1×
“[42] showed that regular arrangements of 3D sensors with larger inter-sensor distance can have a spatial resolving power and localization accuracy comparable to a dense regular arrangement of 3D sensors.”
BackgroundMagnetic modeling of a MEMS flux concentrator
Fischer, G., & Edelstein, A. S. (2010) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/icsens.2010.5690690 · Cited by 4 · cited 1×
“Given the prototypes and simulations, the detectivity is expected to reach a few pT/Hz1/2 at 1 Hz at room temperature [8,19,21].”
BackgroundMulyadi, I. H., Fiedler, P., Eichardt, R., Haueisen, J., & Supriyanto, E. (2021) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-021-02319-9 · Cited by 4
Self-citationSimultaneous use of multiple MCG sensor arrays-a study on the localization accuracy
TILG, B., Wach, P., Lafer, G., Nenonen, J., & Katila, T. E. (2002) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.1996.647497 · Cited by 3
Research works printed but not linked: 8
- Dössel, O. Inverse Problem of Electro- and Magnetocardiography: Review and Recent Progress. Available online: http://www.ijbem.org/volume2/number2/doessel/paper_ijbem.htm.
- Burghoff, 2004, Selection of optimal recording sites in electrocardiography and magnetocardiography, Biomed. Eng. Biomed. Te., 48, 174
- Golub, G.H., and Loan, C.F.V. (1996). Matrix Computations, John Hopkins University Press.
- Engelbrecht, A.P. (2005). Fundamentals of Swarm Intelligence, Wiley.
- Dodge, Y. (1987). Statistical Data Analysis Based on the L1 Norm and Related Methods, North Holland/Elsevier.
- Jazbinšek, V., Kosch, O., Meindl, P., Steinhoff, U., Trontelj, Z., and Trahms, L. (2000, January 13–17). Multichannel vector MFM and BSPM of chest and back. Proceedings of the 12th International Conference on Biomagnetism Biomag, Espoo, Finland.
- Kim, 2004, Design of a SQUID sensor array measuring the tangential field components in magnetocardiogram, Prog. Supercond., 45, 56
- Mulyadi, I.H., Fiedler, P., Eichardt, R., Zelle, D., Haueisen, J., and Supriyanto, E. (2016). Optimal electrode placement for smart clothing: Compromises between electrophysiological and practical aspects. IEEE J. Biomed. Health Inform., in press.
Cited by
Ultrasensitive Magnetic Field Sensors for Biomedical Applications
Murzin, D., Mapps, D. J., Levada, K., Belyaev, V., Omelyanchik, A., Panina, L., & Rodionova, V. (2020) · Sensors · doi:10.3390/s20061569 · Cited by 277
Clinical magnetocardiography: the unshielded bet—past, present, and future
Brisinda, D., Fenici, P., & Fenici, R. (2023) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2023.1232882 · Cited by 80
Recording the heart beat of cattle using a gradiometer system of optically pumped magnetometers
Sutter, J. U., Lewis, O., Robinson, C., McMahon, A., Boyce, R., Bragg, R., Macrae, A., Orton, J., Shah, V., Ingleby, S. J., et al. (2020) · Computers and Electronics in Agriculture · doi:10.1016/j.compag.2020.105651 · Cited by 35
“Recent work on the identification of optimum sensor positioning has shown that a wide range of heart beat characteristics, such as the spatial and temporal magnetic signature of the heart action, are becoming measurable with OPMs (Lau et al., 2016).”
InfluentialBîcă, I., & Anitas, E. M. (2018) · Composites Part B Engineering · doi:10.1016/j.compositesb.2018.09.058 · Cited by 22
Sorbo, A. R., Lombardi, G., La Brocca, L., Guida, G., Fenici, R., & Brisinda, D. (2018) · Annals of Noninvasive Electrocardiology · doi:10.1111/anec.12526 · Cited by 18
“, 2017; Shah, 2013) or even for dynamic monitoring (Lau et al., 2016).”
Background
Metadata
- Authors
- Lau, S., Petković, B., & Haueisen, J.
- Affiliations
- Technische Universität Ilmenau (DE)
- Corresponding author
- Lau, Stephan
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2016-05-24
- Volume
- 16
- Issue
- 6
- Pages
- 754
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC4934180
- OpenAlex
- W2402812596
- Cited by
- 17
- References
- 75
- Citations by year
- 2026: 2 · 2025: 3 · 2024: 2 · 2023: 2 · 2022: 1 · 2021: 1 · 2020: 2 · 2018: 3 · 2017: 1
- FWCI
- 0.63
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Minerva Access (University of Melbourne) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Common Library Network (Der Gemeinsame Bibliotheksverbund) (submittedVersion, cc-by)
- Funders
- German Research Council (GK 1567-C4)
- Topics
- Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Atomic and Subatomic Physics Research
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Magnetic Field Sensors Techniques
- Keywords
- magnetocardiography, sensor placement optimization, magnetoresistive sensors, optically pumped magnetometers, torso model, cardiac magnetic field
- MeSH terms
- Cardiovascular Diseases, Humans, Magnetics, Biosensing Techniques, Magnetocardiography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›