Reliability of Kinovea® Software and Agreement with a Three-Dimensional Motion System for Gait Analysis in Healthy Subjects
Open accessAbstract
Gait analysis is necessary to diagnose movement disorders. In order to reduce the costs of three-dimensional motion capture systems, new low-cost methods of motion analysis have been developed. The purpose of this study was to evaluate the inter- and intra-rater reliability of Kinovea® and the agreement with a three-dimensional motion system for detecting the joint angles of the hip, knee and ankle during the initial contact phase of walking. Fifty healthy subjects participated in this study. All participants were examined twice with a one-week interval between the two appointments. The motion data were recorded using the VICON Motion System® and digital video cameras. The intra-rater reliability showed a good correlation for the hip, the knee and the ankle joints (Intraclass Correlation Coefficient, ICC > 0.85) for both observers. The ICC for the inter-rater reliability was >0.90 for the hip, the knee and the ankle joints. The Bland–Altman plots showed that the magnitude of disagreement was approximately ±5° for intra-rater reliability, ±2.5° for inter-rater reliability and around ±2.5° to ±5° for Kinovea® versus Vicon®. The ICC was good for the hip, knee and ankle angles registered with Kinovea® during the initial contact of walking for both observers (intra-rater reliability) and higher for the agreement between observers (inter-rater reliability). However, the Bland–Altman plots showed disagreement between observers, measurements and systems (Kinovea® vs. three-dimensional motion system) that should be considered in the interpretation of clinical evaluations.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Based on previous Kinovea reliability studies in lower-limb kinematics, with α = 0.05, β = 0.20, a minimum acceptable ICC of 0.70 and an expected ICC of 0.90, 18.4 subjects were required.
“Assuming values for type I and type II errors of α = 0.05 and β = 0.20, respectively, a minimum acceptable ICC of 0.70 and an expected ICC of 0.90, it was estimated that 18.4 subjects were required.”
From Intra-Rater and Test–Retest Reliability of Kinovea for the Kinematic Analysis of Squatting in Healthy Active Women (2026) · cites this paper as [32]
The Sample Size Calculation section (2.6) gives a minimum acceptable ICC of 0.6, an expected ICC of 0.8, and a required sample of 43—not 0.70, 0.90, and 18.4. It does not give the claimed α and β values for that calculation. (LLM, conf=0.92)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measuredThe validity and reliability of the Kinovea software was previously studied.
“The validity and reliability of the Kinovea software was previously studied (34–36).”
From Kinematic analysis of the back squat at different load intensities in powerlifters and weightlifters (2024) · cites this paper as [34]
Previous literature evaluating sagittal plane kinematics with Kinovea uniformly reports robust reliability coefficients (ICC > 0.80) for hip, knee, and ankle angles, pelvic alignment, lumbar lordosis, and trunk tilt.
“previous literature evaluating sagittal plane kinematics with Kinovea, which uniformly reports robust reliability coefficients (ICC > 0.80) for hip, knee, and ankle angles [32], pelvic alignment [31], lumbar lordosis [46,47], and trunk tilt [48].”
From Intra-Rater and Test–Retest Reliability of Kinovea for the Kinematic Analysis of Squatting in Healthy Active Women (2026) · cites this paper as [32]
Previous Kinovea test-retest studies report ICCs > 0.85.
“previous Kinovea test–retest studies reporting ICCs > 0.85 [31,32].”
From Intra-Rater and Test–Retest Reliability of Kinovea for the Kinematic Analysis of Squatting in Healthy Active Women (2026) · cites this paper as [32]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
A guideline of selecting and reporting intraclass correlation coefficients for reliability research
Koo, T. K., & Li, M. Y. (2016) · Journal of Chiropractic Medicine · doi:10.1016/j.jcm.2016.02.012 · Cited by 29,639 · see citing works
Measurement of lower extremity kinematics during level walking
Kadaba, M. P., Ramakrishnan, H. K., & Wootten, M. E. (1990) · Journal of Orthopaedic Research · doi:10.1002/jor.1100080310 · Cited by 3,216 · see citing works
A gait analysis data collection and reduction technique
Davis, R. B., Õunpuu, S., Tyburski, D., & Gage, J. R. (1991) · Human Movement Science · doi:10.1016/0167-9457(91)90046-z · Cited by 3,086 · see citing works
The reliability of three-dimensional kinematic gait measurements: A systematic review
McGinley, J. L., Baker, R. J., Wolfe, R. S., & Morris, M. E. (2008) · Gait & Posture · doi:10.1016/j.gaitpost.2008.09.003 · Cited by 1,155 · see citing works
Perry, J., Slac, K., & Davids, J. R. (1992) · Journal of Pediatric Orthopaedics · doi:10.1097/01241398-199211000-00023 · Cited by 1,125 · see citing works
Bartlett, J. W., & Frost, C. D. (2008) · Ultrasound in Obstetrics and Gynecology · doi:10.1002/uog.5256 · Cited by 943 · see citing works
Reliability and minimal detectible change values for gait kinematics and kinetics in healthy adults
Wilken, J. M., Rodriguez, K. M., Brawner, M., & Darter, B. J. (2011) · Gait & Posture · doi:10.1016/j.gaitpost.2011.09.105 · Cited by 340 · see citing works
Balsalobre‐Fernández, C., Tejero‐González, C. M., del Campo-Vecino, J., & Bavaresco, N. (2013) · The Journal of Strength and Conditioning Research · doi:10.1519/jsc.0b013e318299a52e · Cited by 242 · see citing works
Gait analysis: clinical facts.
Baker, R. J., Esquenazi, A., Benedetti, M. G., et al. (2016) · PubMed · Cited by 181 · see citing works
Observational Gait Assessments in People With Neurological Disorders: A Systematic Review
Gor-García-Fogeda, M. D., Cano‐de‐la‐Cuerda, R., Carratalá‐Tejada, M., Diego, I. M. A., & Rueda, F. M. (2015) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/j.apmr.2015.07.018 · Cited by 116 · see citing works
Self-citationMeldrum, D., Shouldice, C., Conroy, R. M., Jones, K., & Forward, M. (2013) · Gait & Posture · doi:10.1016/j.gaitpost.2013.07.130 · Cited by 115 · see citing works
de Almeida, M. O., Saragiotto, B. T., Yamato, T. P., et al. (2014) · Physical Therapy in Sport · doi:10.1016/j.ptsp.2014.02.005 · Cited by 96 · see citing works
Daly, J. J., Nethery, J., McCabe, J. P., Brenner, I., Rogers, J., Gansen, J., Butler, K., Burdsall, R., Roenigk, K., & Holcomb, J. (2009) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2008.12.016 · Cited by 90 · see citing works
Reliability of video-based quantification of the knee- and hip angle at foot strike during running.
Damsted, C., Nielsen, R. O., & Larsen, L. H. (2015) · PubMed Central · Cited by 89 · see citing works
Reliability of Kinovea Computer Program in Measuring Cervical Range of Motion in Sagittal Plane
Elwardany, S. H., Elsayed, W. H., & Ali, M. F. (2015) · OALib · doi:10.4236/oalib.1101916 · Cited by 74 · see citing works
Footstrike patterns among novice runners wearing a conventional, neutral running shoe
Bertelsen, M. L., Jensen, J. F., Nielsen, M., Nielsen, R. O., & Rasmussen, S. (2012) · Gait & Posture · doi:10.1016/j.gaitpost.2012.11.022 · Cited by 63 · see citing works
Yeo, S. S., & Park, G. Y. (2020) · Sensors · doi:10.3390/s20051343 · Cited by 56 · see citing works
Video gait analysis for ambulatory children with cerebral palsy: Why, when, where and how!
Harvey, A. R., & Gorter, J. W. (2010) · Gait & Posture · doi:10.1016/j.gaitpost.2010.11.025 · Cited by 53 · see citing works
Brown, C. R., Hillman, S. J., Richardson, A., Herman, J., & Robb, J. (2007) · Gait & Posture · doi:10.1016/j.gaitpost.2007.08.008 · Cited by 49 · see citing works
Damsted, C., Larsen, L. H., & Nielsen, R. O. (2015) · Gait & Posture · doi:10.1016/j.gaitpost.2015.01.029 · Cited by 39 · see citing works
Real-Time Gait Event Detection Based on Kinematic Data Coupled to a Biomechanical Model †
Lambrecht, S., Harutyunyan, A., Tanghe, K., Afschrift, M., Schutter, J. D., & Jonkers, I. (2017) · Sensors · doi:10.3390/s17040671 · Cited by 39 · see citing works
Reliability of Using Kinovea Program in Measuring Dominant Wrist Joint Range of Motion
Ali, M., El-Raheem, R. M. A., & Kamel, R. M. (2015) · Trends in Applied Sciences Research · doi:10.3923/tasr.2015.224.230 · Cited by 35 · see citing works
Development and Assessment of a Low-Cost Clinical Gait Analysis System
Littrell, M. E., Chang, Y., & Selgrade, B. P. (2018) · Journal of Applied Biomechanics · doi:10.1123/jab.2017-0370 · Cited by 31 · see citing works
Motion analysis in the clinic: There's an app for that
Mills, K. A. G. (2014) · Journal of physiotherapy · doi:10.1016/j.jphys.2014.11.014 · Cited by 12 · see citing works
Walter (1998) · doi:10.1002/(sici)1097-0258(19980115)17:1
Flagged first, then by citations
Cited by
Sorted by citations
Shishov, N., Elabd, K., Komisar, V., Chong, H., & Robinovitch, S. N. (2021) · PLOS ONE · doi:10.1371/journal.pone.0258923 · Cited by 42 · see citing works
Vicente-Pina, L., Sánchez-Rodríguez, R., Ferrández-Laliena, L., Heredia-Jimenez, J., Müller-Thyssen-Uriarte, J., Monti-Ballano, S., Hidalgo-García, C., Tricás-Moreno, J. M., & Lucha-López, M. O. (2025) · Sensors · doi:10.3390/s25010250 · Cited by 16 · see citing works
Reliability and applicability of a low-cost, camera-based gait evaluation method for clinical use
Spanos, S., Kanellopoulos, A., Petropoulakos, K., Dimitriadis, Z., Siasios, I., & Poulis, I. (2023) · Expert Review of Medical Devices · doi:10.1080/17434440.2023.2171289 · Cited by 15 · see citing works
The effects of fatigue on linear and angular kinematics during bilateral squat
Erman, B., Vural, F., Dopsaj, M., Ozkol, M. Z., Kose, D. E., & Aksit, T. (2023) · PLoS ONE · doi:10.1371/journal.pone.0289089 · Cited by 10 · see citing works
Soylu, C., Acar, G., Uzumcu, B., Demir, P., Seyhan, S., & Biyikli, T. (2025) · Life · doi:10.3390/life15010080 · Cited by 7 · see citing works
Kinematic analysis of the back squat at different load intensities in powerlifters and weightlifters
Giustino, V., Vicari, D. S. S., Figlioli, F., Gervasi, M., Fernández Peña, E., Schifaudo, N., Tedesco, M., Drid, P., Paoli, A., Battaglia, G., et al. (2024) · Frontiers in Sports and Active Living · doi:10.3389/fspor.2024.1454309 · Cited by 6 · see citing works
Yodchaisarn, W., Rungruangbaiyok, S., de Lourdes Pereira, M., & Rungruangbaiyok, C. (2023) · PeerJ · doi:10.7717/peerj.16148 · Cited by 3 · see citing works
Vicente-Loren, C., Lucha-López, M. O., Monti-Ballano, S., Hijazo-Larrosa, S., Vicente-Pina, L., Ferrández-Laliena, L., Tricás-Moreno, J. M., & Hidalgo-García, C. (2026) · Sensors · doi:10.3390/s26123749
- Authors
- Fernández-González, P., Koutsou, A., Cuesta-Gómez, A., Carratalá-Tejada, M., Miangolarra-Page, J. C., & Molina-Rueda, F.
- Affiliations
- Universidad Rey Juan Carlos (ES)
- Motion Analysis, Ergonomics, Biomechanics and Motor Control Laboratory (LAMBECOM), Department of Physical Therapy, Occupational Therapy, Rehabilitation and Physical Medicine, Faculty of Health Sciences, Universidad Rey Juan Carlos, 28922 Madrid, Spain
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2020-06-02
- Volume
- 20
- Issue
- 11
- Pages
- 3154
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC7308968
- OpenAlex
- W3033183326
- Cited by
- 185
- References
- 26
- Citations by year
- 2026: 41 · 2025: 41 · 2024: 36 · 2023: 30 · 2022: 19 · 2021: 15 · 2020: 3
- FWCI
- 5.26
- Type
- journal article
- Language
- en
- Open access
- gold
- dx.doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- BURJC Digital (King Juan Carlos University) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Hispana (submittedVersion, cc-by)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Lower Extremity Biomechanics and Pathologies
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Physical Sciences › Engineering › Biomedical Engineering › Gait Recognition and Analysis
- Keywords
- gait analysis, Kinovea, video-based motion analysis, 3D motion capture, inter-rater reliability, intra-rater reliability, joint angles, measurement agreement, low-cost motion analysis
- MeSH terms
- Gait Analysis, Biomechanical Phenomena, Humans, Software, Reproducibility of Results, Healthy Volunteers
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review