Effects of Different Hematocrit Levels on Glucose Measurements With Handheld Meters for Point-of-Care Testing
The DOI matches its Crossref record; 23 of the 23 references it cites were found in a registry or an index.
Abstract
Objectives. —To determine the effects of low, normal, and high hematocrit levels on glucose meter measurements and to assess the clinical risks of hematocrit errors. Design. —Changes in glucose measurements between low and high hematocrit levels were calculated to determine hematocrit effects. The differences between glucose measured with meters and with a plasma glucose method (YSI 2300) also were compared. Setting. —Six handheld glucose meters were assessed in vitro at low (19.1%), normal (38.5%), and high (58.3%) hematocrit levels, and at 6 glucose concentrations ranging from 2.06 mmol/L (37.1 mg/dL) to 30.24 mmol/L (544.7 mg/dL). Results. —Most systems, regardless of the reference to which they were calibrated, demonstrated positive bias at lower hematocrit levels and negative bias at higher hematocrit levels. Low, normal, and high hematocrit levels progressively lowered Precision G and Precision QID glucose measurements. Hematocrit effects on the other systems were more dependent on the glucose concentration. Overall, Accu-Chek Comfort Curve showed the least sensitivity to hematocrit changes, except at the lowest glucose concentration. Conclusions. —We strongly recommend that clinical professionals choose glucose systems carefully and interpret glucose measurements with extreme caution when the patient's hematocrit value changes, particularly if there is a simultaneous change in glucose level.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“In strip-based glucose assays, high hematocrit values, due to increased blood viscosity, can reduce the diffusion of plasma and determine underestimated results, whereas low hematocrit levels, like in anemia, are associated with overestimated readings 101,102.”
From Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices (2022) · cites this paper as [102]
In the citing paper
“Glucose extraction is approximately in linear relation with the density and duration of iontophoretic current (<0.5 mA/cm2) 102.”
From Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices (2022) · cites this paper as [102]
In the citing paper
“Important variables that may influence the results of bedside glucose monitoring include changes in hematocrit (114), altitude, environmental temperature or humidity, hypotension, hypoxia and high triglyceride concentrations (115), and various drugs.”
From Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus (2011) · cites this paper as [114]
In the citing paper
“Quite often the relevance of factors such as HCT is not considered by the treating physician and, subsequently, by the PwD [28].”
From Self-Monitoring of Blood Glucose as an Integral Part in the Management of People with Type 2 Diabetes Mellitus (2022) · cites this paper as [28]
Citations
Cites
A Text-Book of Physical Chemistry
(1936) · Nature · doi:10.1038/138864b0 · Cited by 1,146
A textbook of physical chemistry
Adamson, A. W. (1973) · Cited by 343
Tang, Z., Du, X., Louie, R. F., & Kost, G. J. (2000) · American Journal of Clinical Pathology · doi:10.1309/qaw1-x5xw-bvrq-5lkq · Cited by 207
Self-citationFingerstick Glucose Determination in Shock
Atkin, S. H., Dasmahapatra, A., Jaker, M. A., Chorost, M. I., & Reddy, S. (1991) · Annals of Internal Medicine · doi:10.7326/0003-4819-114-12-1020 · Cited by 189
Louie, R. F., Tang, Z., Sutton, D. V., Lee, J. H., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-0257-pocgt · Cited by 180
Self-citationBurrin, J. M., & Price, C. P. (1985) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/000456328502200401 · Cited by 150
Kost, G. J., Vu, H.-T., Lee, J. H., Bourgeois, P., Kiechle, F. L., Martin, C., Miller, S. S., Okorodudu, A. O., Podczasy, J. J., Webster, R., et al. (1998) · Critical Care Medicine · doi:10.1097/00003246-199803000-00036 · Cited by 120
Self-citationCommercial biosensors : applications to clinical, bioprocess, and environmental samples
Ramsay, G. (1998) · Wiley eBooks · Cited by 109
Accuracy of fingerstick glucose values in shock patients
Sylvain, H., Pokorny, M., English, S., Benson, N., Whitley, T., Ferenczy, C., & Harrison, J. (1995) · American Journal of Critical Care · doi:10.4037/ajcc1995.4.1.44 · Cited by 89
Neonatal Hyperviscosity: I. Incidence
Wirth, F. H., Goldberg, K. E., & Lubchenco, L. O. (1979) · PEDIATRICS · doi:10.1542/peds.63.6.833 · Cited by 79
Research works printed but not linked: 4
- Hills, L., G.Azurin, and X.Wang. et al. Glutathione as an interferent in near patient whole blood glucose devices.In: Proceedings of the 17th International Symposium of the International Federation of Clinical Chemistry. Madison, Wis: Omnipress. 1998. 207–219.
- Ramsay, G. Commercial Biosensors: Applications to Clinical, Bioprocess, and Environmental Samples. New York, NY: John Wiley & Sons Inc. 1998. 23.
- Adamson, A. W. A Textbook of Physical Chemistry. New York, NY: Academic Press. 1973. 339–437.
- Hills, L., G.Azurin, and X.Wang. et al. Glutathione as an interferent in near patient whole blood glucose devices.In: Proceedings of the 17th International Symposium of the International Federation of Clinical Chemistry. Madison, Wis: Omnipress. 1998. 207–219.
Cited by
Items 101–200 of 294
Factors interfering with the accuracy of five blood glucose meters used in Chinese hospitals
Lv, H., Zhang, G., Kang, X., Yuan, H., Lv, Y., Wang, W., & Randall, R. (2013) · Journal of Clinical Laboratory Analysis · doi:10.1002/jcla.21611 · Cited by 20
Orsonneau, J., Fraissinet, F., Sebille, V., Dudouet, D., & Bigot‐Corbel, E. (2012) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2012-0201 · Cited by 21
Clinical performance of the new glucometer in the nursery and neonatal intensive care unit
Nuntnarumit, P., Chittamma, A., Pongmee, P., Tangnoo, A., & Goonthon, S. (2010) · Pediatrics International · doi:10.1111/j.1442-200x.2010.03214.x · Cited by 20
Hermayer, K. L., Loftley, A., Reddy, S. B., Narla, S. N., Epps, N. A., & Zhu, Y. (2015) · Current Diabetes Reports · doi:10.1007/s11892-015-0582-9 · Cited by 23
Biswas, S. K., Chatterjee, S., Laha, S., Pakira, V., Som, N. K., Saha, S., & Chakraborty, S. K. (2022) · Lab on a Chip · doi:10.1039/d2lc00824f · Cited by 24
Acierno, M. J., Schnellbacher, R., & Tully, T. N. (2012) · Journal of Avian Medicine and Surgery · doi:10.1647/2011-038r1.1 · Cited by 23
Solnica, B., SKUPIEN, J. K., Kuśnierz‐Cabala, B., Słowińska-Solnica, K., Witek, P., Cempa, A., & Malecki, M. T. (2011) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2011.770 · Cited by 21
St-Louis, P., & Éthier, J. (2002) · Clinica Chimica Acta · doi:10.1016/s0009-8981(02)00160-2 · Cited by 21
Tendl, K. A., Christoph, J., Bohn, A., Herkner, K. R., Pollak, A., & Prusa, A.-R. (2013) · Clinical Chemistry and Laboratory Medicine · doi:10.1515/cclm-2012-0864 · Cited by 18
Burdick, S., Mitchell, M. A., Neil, J., Heggem, B., Whittington, J. K., & Acierno, M. J. (2012) · Journal of the American Veterinary Medical Association · doi:10.2460/javma.240.5.596 · Cited by 22
Tran, N. K., Godwin, Z. R., Steele, A. N., Wolf, S. E., & Palmieri, T. L. (2016) · Pediatric Critical Care Medicine · doi:10.1097/pcc.0000000000000877 · Cited by 18
Sirkin, A., Jalloh, T., & Lee, L. (2002) · Journal for Specialists in Pediatric Nursing · doi:10.1111/j.1744-6155.2002.tb00159.x · Cited by 18
Jendrike, N., Baumstark, A., Pleus, S., Liebing, C., Beer, A., Flacke, F., Haug, C., & Freckmann, G. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0391 · Cited by 17
Floré, K., Fiers, T., & Delanghe, J. (2008) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2008.332 · Cited by 17
Summa, N., Eshar, D., Lee-Chow, B., Larrat, S., & Brown, D. C. (2014) · PubMed Central · Cited by 25
Hypoglycemia - Causes and Occurrences
Rigobelo, E. C. (2011) · InTech eBooks · doi:10.5772/864 · Cited by 17
A Perspective on the Accuracy of Blood Glucose Meters During Pregnancy
Immanuel, J., & Simmons, D. S. (2018) · Diabetes Care · doi:10.2337/dc18-0833 · Cited by 18
Kunze, P. E., Perrault, J. R., Chang, Y., Manire, C. A., Clark, S. A., & Stacy, N. I. (2020) · PLoS ONE · doi:10.1371/journal.pone.0229800 · Cited by 17
Scully, M. S., Ort, T., James, I. E., Bugelski, P. J., Makropoulos, D., Deutsch, H., Pieterman, E. J., van den Hoek, A. M., Havekes, L. M., duBell, W. H., et al. (2011) · Experimental Diabetes Research · doi:10.1155/2011/910159 · Cited by 17
Accuracy of Glucometers Should Not Be Assumed
Mann, E. A., Pidcoke, H. F., Salinas, J., Wade, C. E., Holcomb, J. B., & Wolf, S. E. (2007) · American Journal of Critical Care · doi:10.4037/ajcc2007.16.6.531-a · Cited by 17
Mann, E. A., Pidcoke, H. F., Salinas, J., Holcomb, J. B., Wolf, S. E., & Wade, C. E. (2008) · Journal of Burn Care & Research · doi:10.1097/bcr.0b013e3181848c74 · Cited by 17
Lascaris, B., Freling, H. G., Edens, M. A., Fokkert, M. J., Olthof, C. G., & Slingerland, R. J. (2021) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-021-00656-6 · Cited by 16
Perioperative glucose control: living in uncertain times—Continuing Professional Development
Girard, M., & Schricker, T. (2011) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/s12630-010-9449-3 · Cited by 14
Pfützner, A., Schipper, C., Ramljak, S., Flacke, F., Sieber, J., Forst, T., & Musholt, P. B. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700122 · Cited by 16
Datte, K., Guillaumin, J., Barrett, S., Monnig, A. A., & Cooper, E. S. (2016) · Journal of Veterinary Emergency and Critical Care · doi:10.1111/vec.12513 · Cited by 16
Self-monitoring of blood glucose with a focus on analytical quality: an overview
Kristensen, G. B. B., & Sandberg, S. (2010) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2010.186 · Cited by 17
Glycemic Control in the Burn Intensive Care Unit: Focus on the Role of Anemia in Glucose Measurement
Mann, E. A., Mora, A. G., Pidcoke, H. F., Wolf, S. E., & Wade, C. E. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300612 · Cited by 19
Higbie, C. T., Eshar, D., & Bello, N. M. (2015) · American Journal of Veterinary Research · doi:10.2460/ajvr.76.6.532 · Cited by 16
Huang, T.-Y., Chang, H.-W., Tsao, M.-F., Chuang, S.-M., Ni, C.-C., Sue, J.-W., Lin, H.-C., & Hsu, C.-T. (2014) · Clinica Chimica Acta · doi:10.1016/j.cca.2014.02.023 · Cited by 15
Evaluation of a New Blood Glucose Monitoring System with Auto-Calibration
Kilo, C., Pinson, M., Joynes, J. O., Joseph, H., Monhaut, N., Parkes, J. L., & Baum, J. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.283 · Cited by 15
Kilo, C., Dickey, W. T., Joynes, J. O., Pinson, M. B., Baum, J. M., Parkes, J. L., & Parker, D. R. (2006) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2006.02.012 · Cited by 15
Diabetes-Technology and the Environment: What Do We Have to Consider?
Heinemann, L. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221146194 · Cited by 15
Point of care blood glucose devices in the hospital setting
Tran, N. K., LaValley, C., Bagley, B., & Rodrigo, J. (2023) · Critical Reviews in Clinical Laboratory Sciences · doi:10.1080/10408363.2023.2170316 · Cited by 17
The Equine Neonatal Intensive Care Laboratory: Point-of-Care Testing
Wilkins, P. A. (2011) · Clinics in Laboratory Medicine · doi:10.1016/j.cll.2010.11.001 · Cited by 14
Critical Care Glucose Point-of-Care Testing
Narla, S. N., Jones, M., Hermayer, K. L., & Zhu, Y. (2016) · Advances in clinical chemistry · doi:10.1016/bs.acc.2016.05.002 · Cited by 17
Plasma-Generating Glucose Monitor Accuracy Demonstrated in an Animal Model
Magarian, P., & Sterling, B. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300622 · Cited by 14
Evaluation of two glucose meters and interference corrections for screening neonatal hypoglycemia
Wada, Y., Nakamura, T., Kaneshige, M., Takahashi, S., Fujinaga, H., Tsukamoto, K., Ito, Y., & Sago, H. (2015) · Pediatrics International · doi:10.1111/ped.12543 · Cited by 12
Comparison of three point-of-care blood glucose meters for use in adult and juvenile alpacas
Tennent‐Brown, B. S., Koenig, A., Williamson, L. H., & Boston, R. C. (2011) · Journal of the American Veterinary Medical Association · doi:10.2460/javma.239.3.380 · Cited by 13
Fiabilidad de la determinación de la glucemia a la cabecera del paciente en pacientes críticos
Arias-Rivera, S., Copete-Vega, A., Vadillo-Obesso, P., Corrochano-Varas, S., Sánchez-Izquierdo, R., Sánchez-Sánchez, M. M., Sáiz-Sanz, A. I., Frutos–Vivar, F., & Durán, T. P. (2007) · Enfermería Intensiva · doi:10.1016/s1130-2399(07)74385-7 · Cited by 12
Hsu, C., Hsiao, H.-C., Lee, M., Chang, S.-F., Lee, T.-C., Tsai, Y.-S., & Zen, J. (2009) · Clinica Chimica Acta · doi:10.1016/j.cca.2008.12.029 · Cited by 12
Kim, K. O., & Kim, B. (2017) · Journal of Fiber Science and Technology · doi:10.2115/fiberst.fiberst.2017-0004 · Cited by 12
Comparison in the Performance of Glucose Meters in Blood Glucose Monitoring during Pregnancy
Kong, G. W. S., Tam, W. H., Chan, M. H. M., So, W. Y., Lam, C. W. K., Yiu, I. P. C., Loo, K. M., & Li, C. Y. (2010) · Gynecologic and Obstetric Investigation · doi:10.1159/000276572 · Cited by 12
Wolfenden, G., James, F., Hung, L. H. T., Bruce, M., & Thompson, M. (2022) · Journal of Small Animal Practice · doi:10.1111/jsap.13491 · Cited by 13
Blood Cell Classification Using the Hough Transform and Convolutional Neural Networks
Molina‐Cabello, M. A., López‐Rubio, E., Luque‐Baena, R. M., Rodríguez-Espinosa, M. J., & Thurnhofer‐Hemsi, K. (2018) · Advances in intelligent systems and computing · doi:10.1007/978-3-319-77712-2_62 · Cited by 14
Pfützner, A., Musholt, P. B., Schipper, C., Demircik, F., Hengesbach, C., Flacke, F., Sieber, J., & Forst, T. M. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700613 · Cited by 14
Tight Glycemic Control and Point-of-Care Testing
Alter, D. N., & Deines, G. (2009) · Clinics in Laboratory Medicine · doi:10.1016/j.cll.2009.06.009 · Cited by 12
Wu, M., Fang, M.-Y., Jen, L.-N., Hsiao, H.-C., Müller, A., & Hsu, C. (2008) · Clinical Chemistry · doi:10.1373/clinchem.2008.106328 · Cited by 12
Proulx, M.-P., Vergneau‐Grosset, C., Hébert, J., Bédard, C., & Maccolini, É. (2022) · Journal of Exotic Pet Medicine · doi:10.1053/j.jepm.2022.07.004 · Cited by 11
Sharma, K., Tiwari, P. K., & Sinha, S. K. (2025) · Journal of Chemical Information and Modeling · doi:10.1021/acs.jcim.4c01423 · Cited by 13
Bewley, B., O’Rahilly, S., Tassell, R., DuBois, J. A., & Donald, E. C. (2009) · Point of Care The Journal of Near-Patient Testing & Technology · doi:10.1097/poc.0b13e3181a4c4f4 · Cited by 10
Patientennahe Labordiagnostik. Point‐of‐Care‐Testing
Gauglitz, G., & Luppa, P. B. (2009) · Chemie in unserer Zeit · doi:10.1002/ciuz.200900483 · Cited by 14
Daves, M., Cemin, R., Fattor, B., Cosio, G., Salvagno, G. L., Rizza, F., & Lippi, G. (2011) · Biochemia Medica · doi:10.11613/bm.2011.041 · Cited by 9
Weng, C.-W., Hsieh, B., Hou, Y., & Cheng, T. (2015) · The Analyst · doi:10.1039/c5an01251a · Cited by 9
Glucagon therapy in canines with an insulinoma: A retrospective descriptive study of 11 dogs.
Harris, M. E., Weatherton, L. K., & Bloch, C. P. (2020) · Europe PMC (PubMed Central) · Cited by 9
Chemonges, S. (2020) · F1000Research · doi:10.12688/f1000research.24927.1 · Cited by 9
Voulgari, C., & Tentolouris, N. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0035 · Cited by 9
Weng, C.-W., Cheng, T., Chen, R. L. C., & Hsieh, B. (2017) · International Journal of Electrochemical Science · doi:10.20964/2017.06.87 · Cited by 9
Les conséquences de l’hypoglycémie
Lachérade, J.-C., & Jacqueminet, S. (2009) · Annales Françaises d Anesthésie et de Réanimation · doi:10.1016/j.annfar.2009.02.037 · Cited by 14
Hattemer, A., & Wardat, S. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818757550 · Cited by 8
Rizvi, A. A., & Sanders, M. B. (2006) · Journal of the American Academy of Nurse Practitioners · doi:10.1111/j.1745-7599.2006.00092.x · Cited by 18
Shin, J., Park, H., Cho, S., Nam, H., & Lee, K.-J. (2014) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2014.06.005 · Cited by 10
Stadelmann, C., Piol, N., Urfer, J., Werner, D., & Roth‐Kleiner, M. (2013) · Clinica Chimica Acta · doi:10.1016/j.cca.2013.07.021 · Cited by 9
Precision and accuracy of a point-of-care glucometer in horses and the effects of sample type
Rendle, D. I., Armstrong, S. K., Heller, J., & Hughes, K. (2019) · The Veterinary Journal · doi:10.1016/j.tvjl.2019.105359 · Cited by 8
Hematocrit estimation using online sequential extreme learning machine
Huynh, H. T., Won, Y., & Kim, J. (2015) · Bio-Medical Materials and Engineering · doi:10.3233/bme-151507 · Cited by 8
Point of Care Glucose Testing in Neonatal Hypoglycemia
Movalia, M. K., & Ogino, M. (2006) · Point of Care The Journal of Near-Patient Testing & Technology · doi:10.1097/01.poc.0000232027.69785.f5 · Cited by 8
Jamaluddin, F. A. (2012) · Asian Biomedicine · doi:10.5372/9704 · Cited by 11
Yakushiji, F., Fujita, H., Suzuki, H., Joukyu, R., Yasuda, M., Terayama, Y., Nagasawa, K., Ohwada, A., Taniguchi, K., Fujiki, K., et al. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0095 · Cited by 8
Kost, G. J., Tran, N. K., Sifontes, J. R., Mecozzi, D. M., Louie, R. F., & Abad, V. J. (2009) · Point of Care The Journal of Near-Patient Testing & Technology · doi:10.1097/poc.0b013e3181a4951d · Cited by 8
Self-citationRed blood cells serve as a primary glucose sink to improve glucose tolerance at altitude
Martí-Mateos, Y., Safari, Z., Bevers, S. G., Midha, A. D., Flanigan, W., Joshi, T., Huynh, H., Desousa, B. R., Blume, S. Y., Baik, A. H., et al. (2026) · Cell Metabolism · doi:10.1016/j.cmet.2026.01.019 · Cited by 12
Huynh, H. T., & Won, Y. (2008) · IEEE Congress on Evolutionary Computation (CEC) · doi:10.1109/cec.2008.4631197 · Cited by 7
Clemmons, E. A., Stovall, M. I., Owens, D. C., Scott, J. A., Wilkes, A. C., Kempf, D., & Ethun, K. F. (2016) · PubMed · Cited by 6
Wavesense technology glucometer Linus for routine self-monitoring and clinical practice
Chlup, R., Doubravová, B., Peterson, K., Zapletalová, J., & Bartek, J. (2010) · Acta Diabetologica · doi:10.1007/s00592-010-0213-5 · Cited by 8
Huynh, H. T., & Won, Y. (2007) · International Conference on Computer Sciences and Convergence Information Technology · doi:10.1109/iccit.2007.224 · Cited by 7
Technical and clinical evaluation of a glucose meter employing amperometric biosensor technology
Demir, S., Yilmazturk, G. C., & Aslan, D. (2007) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2007.10.006 · Cited by 8
Evaluation of GLUCOCARD X-METER Glucose Monitoring System
Kwon, M., & Lee, S. (2008) · Annals of Laboratory Medicine · doi:10.3343/kjlm.2008.28.1.8 · Cited by 7
Perrault, J. R., Arendt, M. D., Schwenter, J. A., Byrd, J., Tuxbury, K. A., & Stacy, N. I. (2021) · Journal of Veterinary Diagnostic Investigation · doi:10.1177/10406387211001830 · Cited by 6
Validación de un glucómetro en una unidad de cuidados intensivos
López, M. Á. C., Serrano, J. L. F. D. L., Rodríguez, J. L. R., & Márquez, T. M. (2011) · Endocrinología y Nutrición · doi:10.1016/j.endonu.2011.08.003 · Cited by 11
Evaluation of a handheld point-of-care analyser for measurement of creatinine in cats
Reeve, J. A., Warman, S. M., Lewis, D. H., Watson, N., & Papasouliotis, K. (2015) · Journal of Feline Medicine and Surgery · doi:10.1177/1098612x15622676 · Cited by 6
Kim, H. N., Moon, K., & Yoon, S. (2020) · Journal of Clinical Laboratory Analysis · doi:10.1002/jcla.23356 · Cited by 6
Correlation between blood glucose and hematocrit: A new estimation methodology
Tiwari, P. K., Nam, H., & Sohn, Y. (2012) · BioChip Journal · doi:10.1007/s13206-012-6302-5 · Cited by 6
Cutler, D. C., Koenig, A., Girolamo, N. D., & Mayer, J. (2020) · American Journal of Veterinary Research · doi:10.2460/ajvr.81.8.642 · Cited by 7
Vesper, H. W., & Myers, G. L. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100202 · Cited by 7
Hematocrit causes the most significant error in point of care glucometers
Mann, E. A., Pidcoke, H. F., Salinas, J., Wolf, S. E., Wade, C. E., & Holcomb, J. B. (2009) · Critical Care Medicine · doi:10.1097/ccm.0b013e31819d2e8a · Cited by 6
A Critical Review of Blood Glucose Testing
Nichols, J. H. (2003) · Point of Care The Journal of Near-Patient Testing & Technology · doi:10.1097/00134384-200303000-00008 · Cited by 6
Common but Missed Errors in Blood Glucose Monitoring
Sreelakshmi, R., Ashok, A. D., Unes, Y., & Basanth, A. (2023) · International Journal of Diabetes and Technology · doi:10.4103/ijdt.ijdt_15_24 · Cited by 6
Inman, M., Lyon, A. W., Lyon, O. A. S., & Lyon, M. E. (2020) · Archives of Pathology & Laboratory Medicine · doi:10.5858/arpa.2020-0101-ra · Cited by 6
Albloui, F., John, J., Alghamdi, O., Alseraye, F., Alqahtani, A., Tamimi, W. G., Albloshi, A. M. K., Aldakheel, F. M., Mateen, A., & Syed, R. (2022) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365513.2022.2138779 · Cited by 7
Durner, J. (2009) · Angewandte Chemie · doi:10.1002/ange.200903363 · Cited by 5
Precision-to-Tolerance Capability: An Important Consideration in Tight Glucose Control
Prinzing, C., Rosenlund, S., Sukeena, V., Malinowski, C., & Wise, L. C. (2009) · American Journal of Critical Care · doi:10.4037/ajcc2009262 · Cited by 4
Louie, R. F., Curtis, C. M., Toffaletti, J. G., Handel, E. A., Slingerland, R. J., Fokkert, M. J., Muller, W., Weinert, S. E., Lee, D. M., & Kotagiri, S. (2015) · Point of Care The Journal of Near-Patient Testing & Technology · doi:10.1097/poc.0000000000000045 · Cited by 6
Self-citationDiabetes diagnostics including analytical methods for glucose monitoring
Wahl, H. G., & Koschinsky, T. (2018) · Springer eBooks · doi:10.1007/978-3-662-54497-6_12 · Cited by 5
Evaluation of a point‐of‐care blood glucose monitor in healthy goats
QUANDT, J. E., Barletta, M., Cornell, K. K., Giguère, S., & Hofmeister, E. H. (2018) · Journal of Veterinary Emergency and Critical Care · doi:10.1111/vec.12686 · Cited by 5
Lane, S. L., & Koenig, A. (2019) · Journal of the American Veterinary Medical Association · doi:10.2460/javma.254.10.1180 · Cited by 6
Validation of a glucose meter at an intensive care unit
López, M. Á. C., Serrano, J. L. F. D. L., Rodríguez, J. L. R., & Márquez, T. M. (2012) · Endocrinología y Nutrición (English Edition) · doi:10.1016/j.endoen.2011.08.003 · Cited by 6
Analytical evaluation of the Optium Xido blood glucose meter
Solnica, B., Naskalski, J. W., & Gernand, W. (2007) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2008.022 · Cited by 7
Diverse Influences on Blood Glucose Measurements in the ICU Setting
Binette, T. M., & Cembrowski, G. S. (2005) · CHEST Journal · doi:10.1378/chest.128.4.3084 · Cited by 5
Grzych, G., Defauwes, I., de Tullio, P., Pekar, J. D., Brousseau, T., Lippi, G., & Cavalier, E. (2025) · Critical Reviews in Clinical Laboratory Sciences · doi:10.1080/10408363.2025.2533855 · Cited by 4
Huynh, H. T., Kim, J., & Won, Y. (2008) · Cited by 4
Impact of Errors and Variability on Clinical Laboratory Test Interpretation
Nerenz, R. D., Pittman, M. E., & Scott, M. G. (2014) · Elsevier eBooks · doi:10.1016/b978-0-12-386456-7.06303-6 · Cited by 4
Dunseath, G. J., Bright, D., & Luzio, S. D. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818821389 · Cited by 5
Metadata
- Authors
- Tang, Z., Lee, J. H., Louie, R. F., & Kost, G. J.
- Affiliations
- University of California, Davis (US)
- ORCID
- Kost, Gerald J.
- Venue
- Archives of Pathology & Laboratory Medicine
- Published
- 2000-08
- Volume
- 124
- Issue
- 8
- Pages
- 1135-1140
- ISSN
- 0003-9985 (ISSN-L)
- 1543-2165
- OpenAlex
- W1883317613
- Cited by
- 293
- References
- 23
- Citations by year
- 2026: 10 · 2025: 6 · 2024: 1 · 2023: 5 · 2022: 10 · 2021: 8 · 2020: 9 · 2019: 10 · 2018: 13 · 2017: 10 · 2016: 10 · 2015: 14 · 2014: 14 · 2013: 21 · 2012: 11
- FWCI
- 3.06
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Archives of Pathology & Laboratory Medicine (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Keywords
- hematocrit, glucose measurement, point-of-care testing, measurement bias, plasma glucose, meter calibration
- MeSH terms
- Blood Glucose, California, Diagnostic Techniques, Endocrine, Evaluation Studies as Topic, Hematocrit, Humans, Predictive Value of Tests, Reproducibility of Results, Biosensing Techniques, Point-of-Care Systems, Diagnostic Equipment
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.