Recalls and Safety Alerts Involving Pacemakers and Implantable Cardioverter-Defibrillator Generators
The DOI matches its Crossref record; 30 of the 30 references it cites were found in a registry or an index.
Abstract
CONTEXT: Unanticipated pacemaker and implantable cardioverter-defibrillator (ICD) generator malfunctions sometimes warrant recall by the US Food and Drug Administration (FDA). Despite increasingly frequent device implantation, pacemaker and ICD recalls and safety alerts (advisories) remain poorly characterized. OBJECTIVES: To determine pacemaker and ICD generator advisory rates in the United States, to identify trends in these rates, and to examine their clinical and financial implications. DESIGN AND SETTING: Analysis of weekly FDA Enforcement Reports issued between January 1990 and December 2000 to identify all advisories involving pacemaker or ICD generators in the United States. Recalls and safety alerts involving lead malfunctions were not included. MAIN OUTCOME MEASURES: Number of pacemakers and ICD generators in the United States subject to FDA recall or safety alert in 1990-2000; annual pacemaker and ICD advisory rates in the United States in 1990-2000; and estimated cost of device advisories. RESULTS: During the study period, 52 advisories (median [25th and 75th percentiles], 4 [4 and 7] per year) involving 408 500 pacemakers and 114 645 ICDs (523 145 total devices) were issued. Hardware malfunctions (35 advisories affecting 280 641 devices) and computer errors (10 advisories affecting 216 533 devices) accounted for 95% of device recalls. Implantable cardioverter-defibrillators were recalled more frequently than pacemakers (mean [SD], 16.4 [1.6] vs 6.7 [0.8] advisories per 100 person-years; P<.001). Between 1995 and 2000, the annual advisory rate increased for both pacemakers (P for trend <.001) and ICDs (P for trend =.02). An estimated 1.3 million device checks and analyses and 36 187 device replacements resulted from the advisories and cost approximately $870 million. CONCLUSIONS: Pacemaker and ICD recalls and safety alerts occur frequently, affect many patients, and appear to be increasing in number and rate. With the growing number of device implants and expanding indications for device therapy, the number of patients affected by device advisories will likely continue to increase.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Moss, A. J., Hall, W. J., Cannom, D. S., Daubert, J. P., Higgins, S. L., Klein, H., Levine, J. H., Saksena, S., Waldo, A. L., Wilber, D., et al. (1996) · New England Journal of Medicine · doi:10.1056/nejm199612263352601 · Cited by 4,165
(1997) · New England Journal of Medicine · doi:10.1056/nejm199711273372202 · Cited by 3,232
A Randomized Study of the Prevention of Sudden Death in Patients with Coronary Artery Disease
Buxton, A. E., Lee, K. L., Fisher, J. D., Josephson, M. E., Prystowsky, E. N., & Hafley, G. (1999) · New England Journal of Medicine · doi:10.1056/nejm199912163412503 · Cited by 2,489
Connolly, S. J., Gent, M., Roberts, R. S., Dorian, P., Roy, D., Sheldon, R. S., Mitchell, L. B., Green, M. S., Klein, G. J., & O’Brien, B. (2000) · Circulation · doi:10.1161/01.cir.101.11.1297 · Cited by 1,607
National Hospital Discharge Survey, 1988: Archival Version
United States Department of Health and Human Services. National Center for Health Statistics (1991) · ICPSR Data Holdings · doi:10.3886/icpsr09521 · Cited by 1,310
Andersen, H. R., Nielsen, J. C., Thomsen, P. E. B., Thuesen, L., Mortensen, P. T., Vesterlund, T., & Pedersen, A. K. (1997) · The Lancet · doi:10.1016/s0140-6736(97)03425-9 · Cited by 841
Connolly, S. J., Kerr, C. R., Gent, M., Roberts, R. S., Yusuf, S., Gillis, A. M., Sami, M. H., Talajic, M., Tang, A. S. L., Klein, G. J., et al. (2000) · New England Journal of Medicine · doi:10.1056/nejm200005113421902 · Cited by 775
Health, United States, 1998; with socioeconomic status and health chartbook
Pamuk, E. R., Makuc, D. M., Reuben, C. A., & Lochner, K. A. (1998) · DigitalGeorgetown (Georgetown University Library) · Cited by 577
Feldman, J. J., Kleinman, J. C., Makuc, D. M., Mościcki, E. K., Pierre, M. B., & Wagener, D. K. (1989) · Cited by 401
GRAS, D., MABO, P., TANG, T., LUTTIKUIS, O., CHATOOR, R., PEDERSEN, A., TSCHELIESSNIGG, H., DEHARO, J., PUGLISI, A., SILVESTRE, J., et al. (1998) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1998.tb01162.x · Cited by 398
Cited by
Gersh, B. J., Maron, B. J., Bonow, R. O., Dearani, J. A., Fifer, M. A., Link, M. S., Naidu, S. S., Nishimura, R. A., Ommen, S. R., Rakowski, H., et al. (2011) · Circulation · doi:10.1161/cir.0b013e318223e2bd · Cited by 2,104
2011 ACCF/AHA Guideline for the Diagnosis and Treatment of Hypertrophic Cardiomyopathy
Gersh, B. J., Maron, B. J., Bonow, R. O., Dearani, J. A., Fifer, M. A., Link, M. S., Naidu, S. S., Nishimura, R. A., Ommen, S. R., Rakowski, H., et al. (2011) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2011.06.011 · Cited by 1,276
Gersh, B. J., Maron, B. J., Bonow, R. O., Dearani, J. A., Fifer, M. A., Link, M. S., Naidu, S. S., Nishimura, R. A., Ommen, S. R., Rakowski, H., et al. (2011) · Circulation · doi:10.1161/cir.0b013e318223e230 · Cited by 818
Security and Privacy for Implantable Medical Devices
Halperin, D., Heydt-Benjamin, T. S., Fu, K., Kohno, T., & Maisel, W. H. (2008) · IEEE Pervasive Computing · doi:10.1109/mprv.2008.16 · Cited by 470
Self-citationBerul, C. I., Van Hare, G. F., Kertesz, N. J., Dubin, A. M., Cecchin, F., Collins, K. K., Cannon, B. C., Alexander, M. E., Triedman, J. K., Walsh, E. P., et al. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2008.01.033 · Cited by 399
Embedded Software: Facts, Figures, and Future
Ebert, C., & Jones, C. (2009) · Computer · doi:10.1109/mc.2009.118 · Cited by 367
Tung, R., Zimetbaum, P., & Josephson, M. E. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2008.05.058 · Cited by 362
Pacemaker and ICD Generator Malfunctions: Analysis of Food and Drug Administration Annual Reports
Maisel, W. H., Moynahan, M., Zuckerman, B. D., Gross, T. P., Tovar, O. H., Tillman, D.-B., & Schultz, D. B. (2006) · JAMA · doi:10.1001/jama.295.16.1901 · Cited by 353
Self-citation2011 ACCF/AHA guideline for the diagnosis and treatment of hypertrophic cardiomyopathy
Gersh, B. J., Maron, B. J., Bonow, R. O., Dearani, J. A., Fifer, M. A., Link, M. S., Naidu, S. S., Nishimura, R. A., Ommen, S. R., Rakowski, H., et al. (2011) · The Journal of Thoracic and Cardiovascular Surgery · doi:10.1016/j.jtcvs.2011.10.020 · Cited by 322
Gould, P. A. (2006) · JAMA · doi:10.1001/jama.295.16.1907 · Cited by 288
Gersh, B. J., Maron, B. J., Bonow, R. O., Dearani, J. A., Fifer, M. A., Link, M. S., Naidu, S. S., Nishimura, R. A., Ommen, S. R., Rakowski, H., et al. (2011) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2011.10.825 · Cited by 279
Implantable Cardioverter-Defibrillators: Expanding Indications and Technologies
Goldberger, Z., & Lampert, R. (2006) · JAMA · doi:10.1001/jama.295.7.809 · Cited by 271
“Death due to device malfunction is rare(110,111) but emotionally devastating when it occurs, as recently illustrated by the death of a college student with hypertrophic cardiomyopathy whose ICD failed to deliver therapy due to a short-circuiting device.(54,109,111) Since February 2005, there have been advisories invol…”
BackgroundCardiac Device Implantation in the United States from 1997 through 2004: A Population-based Analysis
Zhan, C., Baine, W. B., Sedrakyan, A., & Steiner, C. (2008) · Journal of General Internal Medicine · doi:10.1007/s11606-007-0392-0 · Cited by 266
Cyber–Physical Modeling of Implantable Cardiac Medical Devices
Jiang, Z., Pajic, M., & Mangharam, R. (2012) · Proceedings of the IEEE · doi:10.1109/jproc.2011.2161241 · Cited by 172
“Of these, 200,000 or 41%, were due to firmware issues and their effect continues to increase in frequency [1].”
KORHONEN, P., HUSA, T., TIERALA, I., VÄÄNÄNEN, H., MÄKIJÄRVI, M., KATILA, T., & TOIVONEN, L. (2006) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2005.00332.x · Cited by 51
Guidelines for Managing Pacemaker and Implantable Defibrillator Advisories
Kramer, D. B., & Maisel, W. H. (2011) · Clinical Cardiac Pacing, Defibrillation and Resynchronization Therapy · doi:10.1016/b978-1-4377-1616-0.00034-5 · Cited by 8
Self-citationRobustness evaluation of computer-aided clinical trials for medical devices
Jang, K. J., Pant, Y. V., Zhang, B., Weimer, J., & Mangharam, R. (2019) · Proceedings of the 10th ACM/IEEE International Conference on Cyber-Physical Systems · doi:10.1145/3302509.3311058 · Cited by 6
“Even with a successful trial, medical devices are still subject to recalls and safety alerts [21].”
Background
Metadata
- Authors
- Maisel, W. H.
- Venue
- JAMA
- Publisher
- American Medical Association (AMA)
- Published
- 2001-08-15
- Volume
- 286
- Issue
- 7
- Pages
- 793
- ISSN
- 0002-9955 (ISSN-L)
- 0098-7484
- OpenAlex
- W1994999406
- Cited by
- 210
- References
- 30
- FWCI
- 5.22
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.