Low‐field MRI: An MR physics perspective
Open accessAbstract
Historically, clinical MRI started with main magnetic field strengths in the ∼0.05–0.35T range. In the past 40 years there have been considerable developments in MRI hardware, with one of the primary ones being the trend to higher magnetic fields. While resulting in large improvements in data quality and diagnostic value, such developments have meant that conventional systems at 1.5 and 3T remain relatively expensive pieces of medical imaging equipment, and are out of the financial reach for much of the world. In this review we describe the current state‐of‐the‐art of low‐field systems (defined as 0.25–1T), both with respect to its low cost, low foot‐print, and subject accessibility. Furthermore, we discuss how low field could potentially benefit from many of the developments that have occurred in higher‐field MRI. In the first section, the signal‐to‐noise ratio (SNR) dependence on the static magnetic field and its impact on the achievable contrast, resolution, and acquisition times are discussed from a theoretical perspective. In the second section, developments in hardware (eg, magnet, gradient, and RF coils) used both in experimental low‐field scanners and also those that are currently in the market are reviewed. In the final section the potential roles of new acquisition readouts, motion tracking, and image reconstruction strategies, currently being developed primarily at higher fields, are presented. Level of Evidence: 5 Technical Efficacy Stage: 1 J. Magn. Reson. Imaging 2019.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
1 retracted
SQUID-detected ultra-low field MRI
Espy, M. A., Matlashov, A. N., & Volegov, P. L. (2013) · Journal of Magnetic Resonance · doi:10.1016/j.jmr.2013.02.009 · Cited by 60
RetractedSparse MRI: The application of compressed sensing for rapid MR imaging
Lustig, M., Donoho, D., & Pauly, J. M. (2007) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21391 · Cited by 6,952
SENSE: Sensitivity encoding for fast MRI
Pruessmann, K. P., Weiger, M., Scheidegger, M. B., & Boesiger, P. (1999) · Magnetic Resonance in Medicine · doi:10.1002/(sici)1522-2594(199911)42:5<952::aid-mrm16>3.0.co;2-s · Cited by 6,251
Generalized autocalibrating partially parallel acquisitions (GRAPPA)
Griswold, M. A., Jakob, P. M., Heidemann, R. M., Nittka, M., Jellus, V., Wang, J., Kiefer, B., & Haase, A. (2002) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10171 · Cited by 5,303
The Alzheimer’s disease neuroimaging initiative (ADNI): MRI methods
Jack, C. R., Bernstein, M. A., Fox, N. C., Thompson, P., Alexander, G., Harvey, D., Borowski, B., Britson, P. J., L. Whitwell, J., Ward, C., et al. (2008) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.21049 · Cited by 4,347
Roemer, P. B., Edelstein, W. A., Hayes, C. E., Souza, S. P., & Mueller, O. M. (1990) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910160203 · Cited by 2,322
RARE imaging: A fast imaging method for clinical MR
Hennig, J., Nauerth, A., & Friedburg, H. (1986) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910030602 · Cited by 2,081
Image reconstruction by domain-transform manifold learning
Zhu, B., Liu, J. Z., Cauley, S. F., Rosen, B. R., & Rosen, M. S. (2018) · Nature · doi:10.1038/nature25988 · Cited by 1,694
Rohrer, M., Bauer, H., Mintorovitch, J., Requardt, M., & Weinmann, H.-J. (2005) · Investigative Radiology · doi:10.1097/01.rli.0000184756.66360.d3 · Cited by 1,682
Magnetic resonance fingerprinting
Ma, D., Gulani, V., Seiberlich, N., Liu, K., Sunshine, J. L., Duerk, J. L., & Griswold, M. A. (2013) · Nature · doi:10.1038/nature11971 · Cited by 1,577
Bottomley, P. A., Foster, T. H., Argersinger, R. E., & Pfeifer, L. M. (1984) · Medical Physics · doi:10.1118/1.595535 · Cited by 1,192
Uğurbil, K., Xu, J., Auerbach, E. J., Moeller, S., Vu, A. T., Duarte-Carvajalino, J. M., Lenglet, C., Wu, X., Schmitter, S., Van de Moortele, P. F., et al. (2013) · NeuroImage · doi:10.1016/j.neuroimage.2013.05.012 · Cited by 812
Otazo, R., Candès, E., & Sodickson, D. K. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25240 · Cited by 697
Magnetic field and tissue dependencies of human brain longitudinal 1H2O relaxation in vivo
Rooney, W. D., Johnson, G., Li, X., Cohen, E. R., Kim, S., Ugurbil, K., & Springer, C. S. (2007) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21122 · Cited by 652
Breuer, F. A., Blaimer, M., Heidemann, R. M., Mueller, M. F., Griswold, M. A., & Jakob, P. M. (2005) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20401 · Cited by 633
Simultaneous multislice (SMS) imaging techniques
Barth, M., Breuer, F., Koopmans, P. J., Norris, D. G., & Poser, B. A. (2016) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25897 · Cited by 534
Pohmann, R., Speck, O., & Scheffler, K. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25677 · Cited by 468
Tisdall, M. D., Hess, A. T., Reuter, M., Meintjes, E. M., Fischl, B., & van der Kouwe, A. J. W. (2012) · Magnetic Resonance in Medicine · doi:10.1002/mrm.23228 · Cited by 443
Deep learning for undersampled MRI reconstruction
Hyun, C. M., Kim, H. P., Lee, S. M., Lee, S., & Seo, J. K. (2018) · Physics in Medicine & Biology · doi:10.1088/1361-6560/aac71a · Cited by 428
Macovski, A. (1996) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910360327 · Cited by 420
Controlled aliasing in volumetric parallel imaging (2D CAIPIRINHA)
Breuer, F. A., Blaimer, M., Mueller, M. F., Seiberlich, N., Heidemann, R. M., Griswold, M. A., & Jakob, P. M. (2006) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20787 · Cited by 413
Multiecho imaging sequences with low refocusing flip angles
Hennig, J. (1988) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(88)90128-x · Cited by 403
Prospective motion correction in brain imaging: A review
Maclaren, J., Herbst, M., Speck, O., & Zaitsev, M. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24314 · Cited by 401
GRASE (Gradient‐and Spin‐Echo) imaging: A novel fast MRI technique
Oshio, K., & Feinberg, D. A. (1991) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910200219 · Cited by 340
T2* measurements in human brain at 1.5, 3 and 7 T
Peters, A. M., Brookes, M. J., Hoogenraad, F. G., Gowland, P. A., Francis, S. T., Morris, P. G., & Bowtell, R. (2007) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2007.02.014 · Cited by 245
Wave‐CAIPI for highly accelerated 3D imaging
Bilgic, B., Gagoski, B. A., Cauley, S. F., Fan, A. P., Polimeni, J. R., Grant, P. E., Wald, L. L., & Setsompop, K. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25347 · Cited by 231
Measurement and Correction of Microscopic Head Motion during Magnetic Resonance Imaging of the Brain
Maclaren, J., Armstrong, B. S. R., Barrows, R. T., Danishad, K. A., Ernst, T., Foster, C. L., Gumus, K., Herbst, M., Kadashevich, I. Y., Kusik, T. P., et al. (2012) · PLoS ONE · doi:10.1371/journal.pone.0048088 · Cited by 215
Fast quantitative susceptibility mapping using 3D EPI and total generalized variation
Langkammer, C., Bredies, K., Poser, B. A., Barth, M., Reishofer, G., Fan, A. P., Bilgic, B., Fazekas, F., Mainero, C., & Ropele, S. (2015) · NeuroImage · doi:10.1016/j.neuroimage.2015.02.041 · Cited by 189
Two‐dimensional imaging in a lightweight portable MRI scanner without gradient coils
Cooley, C. Z., Stockmann, J., Armstrong, B. D., Sarracanie, M., Lev, M. H., Rosen, M. S., & Wald, L. L. (2014) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25147 · Cited by 175
Strategies for brain shift evaluation
Hastreiter, P., Rezk‐Salama, C., Soza, G., Bauer, M., Greiner, G., Fahlbusch, R., Ganslandt, O., & Nimsky, C. (2004) · Medical Image Analysis · doi:10.1016/j.media.2004.02.001 · Cited by 154
Rapid, high-resolution, whole-brain, susceptibility-based MRI of multiple sclerosis
Sati, P., Thomasson, D. M., Li, N., Pham, D. L., Biassou, N., Reich, D., & Butman, J. A. (2014) · Multiple Sclerosis Journal · doi:10.1177/1352458514525868 · Cited by 136
Magnetic resonance fingerprinting – An overview
Panda, A., Mehta, B. B., Coppo, S., Jiang, Y., Ma, D., Seiberlich, N., Griswold, M. A., & Gulani, V. (2017) · Current Opinion in Biomedical Engineering · doi:10.1016/j.cobme.2017.11.001 · Cited by 117
Yang, Y., Cao, M., Sheng, K., Gao, Y., Chen, A. R., Kamrava, M., Lee, P. P., Agazaryan, N., Lamb, J., Thomas, D. H., et al. (2016) · Medical Physics · doi:10.1118/1.4942381 · Cited by 112
Vesanen, P. T., Nieminen, J. O., Zevenhoven, K. C. J., Dabek, J., Parkkonen, L. T., Zhdanov, A. V., Luomahaara, J., Hassel, J., Penttilä, J. S., Simola, J., et al. (2012) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24413 · Cited by 110
Four-dimensional respiratory motion-resolved whole heart coronary MR angiography
Piccini, D., Feng, L., Bonanno, G., Coppo, S., Yerly, J., Lim, R., Schwitter, J., Sodickson, D. K., Otazo, R., & Stuber, M. (2016) · Magnetic Resonance in Medicine · doi:10.1002/mrm.26221 · Cited by 105
Tarantino, U., Fanucci, E., Iundusi, R., Celi, M., Altobelli, S., Gasbarra, E., Simonetti, G., & Manenti, G. (2012) · Journal of Orthopaedics and Traumatology · doi:10.1007/s10195-012-0213-z · Cited by 89
Insertable biplanar gradient coils for magnetic resonance imaging
Martens, M., Petropoulos, L. S., Brown, R. W., Andrews, J. H., Morich, M. A., & Patrick, J. L. (1991) · Review of Scientific Instruments · doi:10.1063/1.1142245 · Cited by 86
RARE/turbo spin echo imaging with simultaneous multislice Wave-CAIPI
Gagoski, B. A., Bilgiç, B., Eichner, C., Bhat, H., Grant, P. E., Wald, L. L., & Setsompop, K. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25615 · Cited by 75
Motion-Correction Enabled Ultra-High Resolution In-Vivo 7T-MRI of the Brain
Federau, C., & Gallichan, D. (2016) · PLoS ONE · doi:10.1371/journal.pone.0154974 · Cited by 71
Zahneisen, B., Poser, B. A., Ernst, T., & Stenger, V. A. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24875 · Cited by 67
Large T1 contrast enhancement using superparamagnetic nanoparticles in ultra-low field MRI
Yin, X., Russek, S. E., Zabow, G., Sun, F., Mohapatra, J., Keenan, K. E., Boss, M. A., Zeng, H., Liu, J. P., Viert, A., et al. (2018) · Scientific Reports · doi:10.1038/s41598-018-30264-5 · Cited by 63
Narsude, M., Gallichan, D., van der Zwaag, W., Rolf Gruetter, A., & Marques, J. P. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25835 · Cited by 61
Self-citationLopez, H. S., Liu, F., Weber, E., et al. (2008) · IEEE Transactions on Magnetics · doi:10.1109/tmag.2007.914770 · Cited by 60
Schulz, J., Boyacıoğlu, R., & Norris, D. G. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25774 · Cited by 25
Terada, Y., Kono, S., Ishizawa, K., Inamura, S., Uchiumi, T., Tamada, D., & Kose, K. (2013) · Journal of Magnetic Resonance · doi:10.1016/j.jmr.2013.02.005 · Cited by 23
Schulz, J., Marques, J. P., ter Telgte, A., van Dorst, A., de Leeuw, F., Meijer, F. J. A., & Norris, D. G. (2017) · European Journal of Radiology · doi:10.1016/j.ejrad.2017.11.022 · Cited by 14
Self-citationMinimum Field Strength Simulator for Proton Density Weighted MRI
Wu, Z., Chen, W., & Nayak, K. S. (2016) · PLoS ONE · doi:10.1371/journal.pone.0154711 · Cited by 12
Monoplanar gradient system for imaging with nonlinear gradients
Littin, S., Gallichan, D., Welz, A. M., Jia, F., Dewdney, A., Weber, H., Schultz, G., Hennig, J., & Zaitsev, M. (2015) · Magnetic Resonance Materials in Physics Biology and Medicine · doi:10.1007/s10334-015-0481-8 · Cited by 10
Methodology for 3D image reconstruction of the female pelvis from upright open MRI (MRO) 2D imaging
Abdulaziz, M., Stothers, L. S., & Macnab, A. J. (2018) · Biomedical Spectroscopy and Imaging · doi:10.3233/bsi-180178 · Cited by 4
Wansapura (1999) · doi:10.1002/(sici)1522-2586(199904)9:4
Larkman (2001) · doi:10.1002/1522-2586(200102)13:2<313::aid-jmri1045>3.0.co;2-w
Cited by
Low‐field MRI: Clinical promise and challenges
Arnold, T. C., Freeman, C. W., Litt, B., & Stein, J. M. (2023) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.28408 · Cited by 374
Wald, L. L., McDaniel, P. C., Witzel, T., Stockmann, J. P., & Cooley, C. Z. (2020) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.26942 · Cited by 272
A low-cost and shielding-free ultra-low-field brain MRI scanner
Liu, Y., Leong, A. T. L., Zhao, Y., Xiao, L., Mak, H. K. F., Tsang, A. C. O., Lau, G. K. K., Leung, G. K. K., & Wu, E. X. (2021) · Nature Communications · doi:10.1038/s41467-021-27317-1 · Cited by 261
Top-Level Design and Simulated Performance of the First Portable CT-MR Scanner
Peng, Y., Li, M., Grandinetti, J., Wang, G., & Jia, X. (2022) · IEEE Access · doi:10.1109/access.2022.3208278 · Cited by 4
“The medical relevance of ultra-low-field MRI has recently been demonstrated in several examples [9]–[12].”
Background
Metadata
- Authors
- Marques, J. P., Simonis, F. F. J., & Webb, A. G.
- Venue
- Journal of Magnetic Resonance Imaging
- Publisher
- Wiley
- Published
- 2019-01-13
- Volume
- 49
- Issue
- 6
- Pages
- 1528-1542
- ISSN
- 1053-1807 (ISSN-L)
- 1522-2586
- OpenAlex
- W2911060328
- Cited by
- 386
- References
- 51
- FWCI
- 19.50
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.