Validity of the Polar V800 heart rate monitor to measure RR intervals at rest
Open accessAbstract
PURPOSE: To assess the validity of RR intervals and short-term heart rate variability (HRV) data obtained from the Polar V800 heart rate monitor, in comparison to an electrocardiograph (ECG). METHOD: Twenty participants completed an active orthostatic test using the V800 and ECG. An improved method for the identification and correction of RR intervals was employed prior to HRV analysis. Agreement of the data was assessed using intra-class correlation coefficients (ICC), Bland-Altman limits of agreement (LoA), and effect size (ES). RESULTS: A small number of errors were detected between ECG and Polar RR signal, with a combined error rate of 0.086 %. The RR intervals from ECG to V800 were significantly different, but with small ES for both supine corrected and standing corrected data (ES <0.001). The bias (LoA) were 0.06 (-4.33 to 4.45 ms) and 0.59 (-1.70 to 2.87 ms) for supine and standing intervals, respectively. The ICC was >0.999 for both supine and standing corrected intervals. When analysed with the same HRV software no significant differences were observed in any HRV parameters, for either supine or standing; the data displayed small bias and tight LoA, strong ICC (>0.99) and small ES (≤0.029). CONCLUSIONS: The V800 improves over previous Polar models, with narrower LoA, stronger ICC and smaller ES for both the RR intervals and HRV parameters. The findings support the validity of the Polar V800 and its ability to produce RR interval recordings consistent with an ECG. In addition, HRV parameters derived from these recordings are also highly comparable.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Statistical power analysis for the behavioral sciences
(1990) · Computers, Environment and Urban Systems · doi:10.1016/0198-9715(90)90050-4 · Cited by 65,269
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Statistical Power Analysis for the Behavioral Sciences
Cohen, J. (2013) · doi:10.4324/9780203771587 · Cited by 25,373
Akselrod, S., Gordon, D., Ubel, F. A., Shannon, D. C., Berger, A. C., & Cohen, R. J. (1981) · Science · doi:10.1126/science.6166045 · Cited by 5,079
QUANTIFYING TEST-RETEST RELIABILITY USING THE INTRACLASS CORRELATION COEFFICIENT AND THE SEM
WEIR, J. P. (2005) · Journal of Strength and Conditioning Research · doi:10.1519/00124278-200502000-00038 · Cited by 4,842
Atkinson, G., & Nevill, A. M. (1998) · Sports Medicine · doi:10.2165/00007256-199826040-00002 · Cited by 3,779
Thayer, J. F., Åhs, F., Fredrikson, M., Sollers, J. J., & Wager, T. D. (2012) · Neuroscience & Biobehavioral Reviews · doi:10.1016/j.neubiorev.2011.11.009 · Cited by 3,224 · cited 3×
“The use and analysis of HRV has become increasingly common, as it is simple, non-invasive and sensitive to physiological and psychological changes (Thayer et al. 2012).”
BackgroundInfluentialHeart rate variability: A review
Rajendra Acharya, U., Paul Joseph, K., Kannathal, N., Lim, C. M., & Suri, J. S. (2006) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-006-0119-0 · Cited by 2,820
Kubios HRV–heart rate variability analysis software.
Tarvainen, M. P., Niskanen, J.-P., Lipponen, J. A., Ranta-aho, P. O., & Karjalainen, P. A. (2014) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2013.07.024 · Cited by 2,720 · cited 5×
“…be cautious about making comparisons between HRV parameters derived from different software packages, particularly when software packages such as Kubios HRV are freely available that support data exported directly from a large number of ECGs, as well as RR interval data (Tarvainen et al. 2014).”
InfluentialThayer, J. F., Yamamoto, S. S., & Brosschot, J. F. (2010) · International Journal of Cardiology · doi:10.1016/j.ijcard.2009.09.543 · Cited by 2,276 · cited 1×
“In a clinical setting, reduced HRV has been shown to unfavourably reflect prognoses for cardiovascular disease, diabetic neuropathy, arterial hypertension, acute myocardial infarction and other heart conditions (Spallone et al. 2011; Thayer et al. 2010; Yi et al. 2014).”
BackgroundResearch Methods in Physical Activity
Thomas, J. R., & Nelson, J. K. (1990) · Cited by 1,562
Heart Rate Monitoring: Applications and Limitations
Achten, J., & Jeukendrup, A. E. (2003) · Sports Medicine · doi:10.2165/00007256-200333070-00004 · Cited by 1,166
Do existing measures of Poincaré plot geometry reflect nonlinear features of heart rate variability
Brennan, M., Palaniswami, M., & Kamen, P. (2001) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.959330 · Cited by 1,044
Spallone, V., Ziegler, D., Freeman, R., Bernardi, L., Frontoni, S., Pop‐Busui, R., Stevens, M., Kempler, P., Hilsted, J., Tesfaye, S., et al. (2011) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.1239 · Cited by 957 · cited 1×
“In a clinical setting, reduced HRV has been shown to unfavourably reflect prognoses for cardiovascular disease, diabetic neuropathy, arterial hypertension, acute myocardial infarction and other heart conditions (Spallone et al. 2011; Thayer et al. 2010; Yi et al. 2014).”
BackgroundValidity of the Polar S810 Heart Rate Monitor to Measure R-R Intervals at Rest
GAMELIN, F. X., BERTHOIN, S., & BOSQUET, L. (2006) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000218135.79476.9c · Cited by 509
Assessment of parasympathetic control of heart rate by a noninvasive method
Fouad, F. M., Tarazi, R. C., Ferrario, C. M., Fighaly, S., & Alicandri, C. (1984) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.1984.246.6.h838 · Cited by 309
Validity and Reliability of Short-Term Heart-Rate Variability from the Polar S810
NUNAN, D., DONOVAN, G., JAKOVLJEVIC, D. G., HODGES, L. D., SANDERCOCK, G. R. H., & BRODIE, D. A. (2009) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0b013e318184a4b1 · Cited by 302
Weippert, M., Kumar, M., Kreuzfeld, S., Arndt, D., Rieger, A., & Stoll, R. (2010) · European Journal of Applied Physiology · doi:10.1007/s00421-010-1415-9 · Cited by 300
Vanderlei, L. C. M., Silva, R. A., Pastre, C. M., Azevedo, F. M., & Godoy, M. F. (2008) · Brazilian Journal of Medical and Biological Research · doi:10.1590/s0100-879x2008005000039 · Cited by 296
The effects of specific respiratory rates on heart rate and heart rate variability
Song, H.-S., & Lehrer, P. M. (2003) · Applied Psychophysiology and Biofeedback · doi:10.1023/a:1022312815649 · Cited by 292
Influence of physical training on heart rate variability and baroreflex circulatory control
Seals, D. R., & Chase, P. B. (1989) · Journal of Applied Physiology · doi:10.1152/jappl.1989.66.4.1886 · Cited by 241
Kingsley, M., Lewis, M. J., & Marson, R. E. (2005) · International Journal of Sports Medicine · doi:10.1055/s-2004-817878 · Cited by 230
The effect of carbon dioxide, respiratory rate and tidal volume on human heart rate variability
Pöyhönen, M. J., Syväoja, S., Hartikainen, J. E. K., Ruokonen, E., & Takala, J. A. (2003) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.2004.00272.x · Cited by 163
Possibilities and limitations of the polar RS800 in measuring heart rate variability at rest
Wallén, M. B., Hasson, D., Theorell, T., Canlon, B., & Osika, W. (2012) · European Journal of Applied Physiology · doi:10.1007/s00421-011-2079-9 · Cited by 135 · cited 7×
“It has been shown that software differences in signal processing and calculation of HRV indices results in unacceptable variation (Nunan et al. 2008; Radespiel-Tröger et al. 2003; Sandercock et al. 2004; Wallén et al. 2012; Weippert et al. 2010).”
InfluentialPorto, L. G. G., & Junqueira, L. F. (2008) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2009.02175.x · Cited by 135
Agreement of two different methods for measurement of heart rate variability
Radespiel‐Tröger, M., Rauh, R. D., Mahlke, C., Gottschalk, T., & M�ck-Weymann, M. (2003) · Clinical Autonomic Research · doi:10.1007/s10286-003-0085-7 · Cited by 132
Validity of the Polar S810 to Measure R-R Intervals in Children
Gamelin, F.-X., Baquet, G., Berthoin, S., & Bosquet, L. (2008) · International Journal of Sports Medicine · doi:10.1055/s-2007-964995 · Cited by 132
Nunan, D., Jakovljević, D. G., Donovan, G., Hodges, L. D., SANDERCOCK, G. R. H., & Brodie, D. A. (2008) · European Journal of Applied Physiology · doi:10.1007/s00421-008-0742-6 · Cited by 130
On the validity of using the Polar RS800 heart rate monitor for heart rate variability research
Quintana, D. S., Heathers, J. A. J., & Kemp, A. H. (2012) · European Journal of Applied Physiology · doi:10.1007/s00421-012-2453-2 · Cited by 118 · cited 4×
“2010) and more recently the RS800 (Quintana et al. 2012; Wallén et al. 2012).”
BackgroundInfluentialHeart rate variability and pre-competitive anxiety in BMX discipline.
Mateo, M., Blasco-Lafarga, C., Martínez-Navarro, I., Guzmán, J. F., & Zabala, M. (2012) · European Journal of Applied Physiology · doi:10.1007/s00421-011-1962-8 · Cited by 116
Theoretical issues for critical currents of bulk polycrystalline YBa2Cu3O7−δ superconductors
Chen, L., & Zhang, Y. (1989) · Journal of Applied Physics · doi:10.1063/1.344346 · Cited by 83
Weippert, M., Behrens, M., Rieger, A., & Behrens, K. (2014) · Entropy · doi:10.3390/e16115698 · Cited by 62
SANDERCOCK, G. R. H., Bromley, P. D., & Brodie, D. A. (2004) · Clinical Physiology and Functional Imaging · doi:10.1111/j.1475-097x.2004.00584.x · Cited by 38
Yi, H.-T., Hsieh, Y., Wu, T., Huang, J., Lin, W., Liang, K., Su, C., Tsai, W., & Wang, K. (2014) · Heart & Lung · doi:10.1016/j.hrtlng.2014.05.010 · Cited by 36
Cited by
Goldsack, J. C., Coravos, A., Bakker, J. P., Bent, B., Dowling, A. V., Fitzer-Attas, C., Godfrey, A., Godino, J. G., Gujar, N., Izmailova, E., et al. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0260-4 · Cited by 557
Henriksen, A., Haugen Mikalsen, M., Woldaregay, A. Z., Muzny, M., Hartvigsen, G., Hopstock, L. A., & Grimsgaard, S. (2018) · Journal of Medical Internet Research · doi:10.2196/jmir.9157 · Cited by 542
“Misfit devices were used in 12 studies [9,36,42,43,46,61-63,85,95-97]; Polar devices were used in 6 studies [36,43,46,62,98,99]; Withings [63,85,89,100,101], Mio [29,30,54,102,103], and Samsung [29,30,58,62,96] devices were used in 5 studies; PulseOn devices were used in 4 studies [29,104-106]; TomTom devices were use…”
MethodRR interval signal quality of a heart rate monitor and an ECG Holter at rest and during exercise
Gilgen-Ammann, R., Schweizer, T., & Wyss, T. (2019) · European Journal of Applied Physiology · doi:10.1007/s00421-019-04142-5 · Cited by 438
Can Wearable Devices Accurately Measure Heart Rate Variability? A Systematic Review
Georgiou, K., Larentzakis, A. V., Khamis, N. N., Alsuhaibani, G. I., Alaska, Y. A., & Giallafos, E. J. (2018) · Folia Medica · doi:10.2478/folmed-2018-0012 · Cited by 244
Schuurmans, A. A. T., de Looff, P., Nijhof, K. S., Rosada, C., Scholte, R. H. J., Popma, A., & Otten, R. (2020) · Journal of Medical Systems · doi:10.1007/s10916-020-01648-w · Cited by 210
Metadata
- Authors
- Giles, D., Draper, N., & Neil, W.
- Venue
- European Journal of Applied Physiology
- Published
- 2015-12-26
- Volume
- 116
- Issue
- 3
- Pages
- 563-571
- ISSN
- 0020-9376 (ISSN-L)
- 1439-6319
- 1439-6327
- OpenAlex
- W2210949426
- Cited by
- 373
- References
- 34
- FWCI
- 15.17
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.