Measurement of Tear Glucose Levels with Amperometric Glucose Biosensor/Capillary Tube Configuration
The DOI matches its Crossref record; 26 of the 26 references it cites were found in a registry or an index.
Abstract
An amperometric needle-type electrochemical glucose sensor intended for tear glucose measurements is described and employed in conjunction with a 0.84 mm i.d. capillary tube to collect microliter volumes of tear fluid. The sensor is based on immobilizing glucose oxidase on a 0.25 mm o.d. platinum/iridium (Pt/Ir) wire and anodically detecting the liberated hydrogen peroxide from the enzymatic reaction. Inner layers of Nafion and an electropolymerized film of 1,3-diaminobenzene/resorcinol greatly enhance the selectivity for glucose over potential interferences in tear fluid, including ascorbic acid and uric acid. Further, the new sensor is optimized to achieve very low detection limits of 1.5 ± 0.4 μM of glucose (S/N = 3) that is required to monitor glucose levels in tear fluid with a glucose sensitivity of 0.032 ± 0.02 nA/μM (n = 6). Only 4-5 μL of tear fluid in the capillary tube is required when the needle sensor is inserted into the capillary. The glucose sensor was employed to measure tear glucose levels in anesthetized rabbits over an 8 h period while also measuring the blood glucose values. A strong correlation between tear and blood glucose levels was found, suggesting that measurement of tear glucose is a potential noninvasive substitute for blood glucose measurements, and the new sensor configuration could aid in conducting further research in this direction.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“The same source of variability, if present in human tears, may be at least partly responsible for the inability to establish a clinically sound blood-to-tear correlation in human subjects (247).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [247]
In the citing paper
“Variability is reported in the ratio between glucose concentrations in blood and tear fluid for individual rabbits (247).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [247]
In the citing paper
“Despite the demonstrated correlations, sampling tears for in vitro diagnoses is associated with several errors that are due to the following: first, small sample volumes102; second, ease of evaporation during sample collection; third, variations in tear production among individuals and throughout the day103; and lastly, challenging collection methods that can affect the sampled analyte concentra”
From Wearable biosensors for healthcare monitoring (2019) · cites this paper as [104]
Citations
Cites
Design and in vitro studies of a needle-type glucose sensor for subcutaneous monitoring
Bindra, D. S., Zhang, Y., Wilson, G. S., Sternberg, R., Thevenot, D. R., Moatti, D., & Reach, G. (1991) · Analytical Chemistry · doi:10.1021/ac00017a008 · Cited by 309
Saha, J. K., Xia, J., Grondin, J. M., Engle, S. K., & Jakubowski, J. A. (2005) · Experimental Biology and Medicine · doi:10.1177/153537020523001012 · Cited by 274
Current development in non-invasive glucose monitoring
Ferrante do Amaral, C. E., & Wolf, B. (2008) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2007.06.003 · Cited by 269
Glucose monitoring by reverse iontophoresis
Potts, R. O., A. Tamada, J., & J. Tierney, M. (2002) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.210 · Cited by 243
Tear glucose levels in normal people and in diabetic patients.
Sen, D. K., & Sarin, G. S. (1980) · British Journal of Ophthalmology · doi:10.1136/bjo.64.9.693 · Cited by 203
Tear glucose analysis for the noninvasive detection and monitoring of diabetes mellitus
Baca, J. T., Finegold, D. N., & Asher, S. A. (2007) · The Ocular Surface · doi:10.1016/s1542-0124(12)70094-0 · Cited by 187
Electropolymerized films to prevent interferences and electrode fouling in biosensors
Geise, R. J., Adams, J. M., Barone, N. J., & Yacynych, A. M. (1991) · Biosensors and Bioelectronics · doi:10.1016/0956-5663(91)87039-e · Cited by 169
Major, T. C., Brant, D. O., Reynolds, M. M., Bartlett, R. H., Meyerhoff, M. E., Handa, H., & Annich, G. M. (2010) · Biomaterials · doi:10.1016/j.biomaterials.2009.12.028 · Cited by 133
Self-citationMass spectral determination of fasting tear glucose concentrations in nondiabetic volunteers
Baca, J. T., Taormina, C. R., Feingold, E., Finegold, D. N., Grabowski, J. J., & Asher, S. A. (2007) · Clinical Chemistry · doi:10.1373/clinchem.2006.078543 · Cited by 110
A glucose-sensing contact lens: from bench top to patient
Badugu, R., Lakowicz, J. R., & Geddes, C. D. (2005) · Current Opinion in Biotechnology · doi:10.1016/j.copbio.2004.12.007 · Cited by 105
Research works printed but not linked: 1
- Choy C. K. M., 2000, Invest. Ophthalmol. Vis. Sci., 41, 3293
Cited by
Items 101–187 of 187
Wang, C., Yang, X., Zhu, G., Wang, T., Yu, D., Lu, Y., & Yu, H. (2022) · Colloids and Surfaces A Physicochemical and Engineering Aspects · doi:10.1016/j.colsurfa.2022.130672 · Cited by 18
Park, S., Nam, D. Y., Jeon, H., Han, J. H., Jang, D., Hwang, J., Park, Y.-S., Han, Y., Choy, Y. B., & Lee, D. Y. (2023) · Biomaterials Research · doi:10.1186/s40824-023-00469-5 · Cited by 18
“To induce diabetic condition in rabbits, we intentionally elevated blood glucose levels via subcutaneous injection of a cocktail of 15 mg/kg ketamine and 5 mg/kg xylazine, as reported in previous studies [37, 46, 47].”
MethodHan, J. H., Cho, Y. C., Koh, W., & Choy, Y. B. (2020) · PLoS ONE · doi:10.1371/journal.pone.0239317 · Cited by 17
“To estimate blood glucose concentrations more accurately, the inherent differences in correlations between the glucose concentrations in tear and blood between individuals need to be considered (Fig 6A) [19, 22, 23], which could also be ascribed to the variation in their temporal change [23, 45].”
BackgroundMethodResultInfluentialFuture of Healthcare—Sensor Data-Driven Prognosis
Pal, A., Mukherjee, A., & Dey, S. (2016) · Springer series in wireless technology · doi:10.1007/978-3-319-42141-4_9 · Cited by 20
Jang, S. (2018) · International Journal of Biosensors & Bioelectronics · doi:10.15406/ijbsbe.2018.04.00087 · Cited by 20
“Research groups working towards non-invasive monitoring methods of glucose in the ocular fluids include Chronoamperometric technique, Electrode/electrochemical embedded contact lens, CMOS/ Amperometric needle-type electrochemical method, Optical coherence tomography (OCT), Fluorescence spectrophotometry, Ocular spectr…”
MethodA Disposable Tear Glucose Biosensor—Part 4
Belle, J. T. L., Engelschall, E., Lan, K., Shah, P., Saez, N., Maxwell, S., Adamson, T., Abou-Eid, M., McAferty, K., Patel, D. R., et al. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296813511741 · Cited by 16
Ultrasensitive detection of proteins and sugars at single-cell level
Watabe, S., Morikawa, M., Kaneda, M., Nakaishi, K., Nakatsuma, A., Ninomiya, M., Yoshimura, T., Miura, T., & Ito, E. (2016) · Communicative & Integrative Biology · doi:10.1080/19420889.2015.1124201 · Cited by 15
“recent study demonstrated that an amperometric needletype electrochemical glucose sensor can detect the mMorder glucose in the tear of mL fluid.(60) Our FCS method also has a sufficient sensitivity for a tear.”
MethodAdvances in Wearable Chemosensors
Korallı, P., & Mouzakis, D. E. (2021) · Chemosensors · doi:10.3390/chemosensors9050099 · Cited by 15
“Nevertheless, in vitrotechniques encountered several problems like the sample’s evaporation during the sensing process [93].”
BackgroundKim, H., Lee, S., Lee, K. W., Kim, E. S., Kim, H., Im, H., Yoon, H. C., Ko, J., & Kim, E. (2023) · Dyes and Pigments · doi:10.1016/j.dyepig.2023.111287 · Cited by 14
Bioinspired assemblies and plasmonic interfaces for electrochemical biosensing
Hinman, S. S., & Cheng, Q. J. (2016) · Journal of Electroanalytical Chemistry · doi:10.1016/j.jelechem.2016.05.014 · Cited by 14
“While blood glucose has been the medium of choice for glucose testing, tear glucose levels have seen a recent increase in popularity due to their high correlation with blood glucose values and lower invasiveness for sample collection [4-6].”
BackgroundKim, H., Rim, Y. S., & Kwon, J. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.2988269 · Cited by 16
“Among the substrates in tear fluids, ascorbic acid (AA) and uric acid (UA) can cause a malfunction in the detection of glucose, which are present in tears at 10-30 μM and 20-100 μM levels, respectively [8], [64], [65].”
BackgroundHybrid hydrogel films with graphene oxide for continuous saliva-level monitoring
Zhang, Z., Dou, Q., Wang, S., Hu, D., Guo, X., Liao, B., Zhao, Z., Liu, H., & Dai, Q. (2020) · Journal of Materials Chemistry C · doi:10.1039/d0tc00725k · Cited by 14
Ratiometric glucose sensing based on fluorescent oxygen films and glucose oxidase
Su, F., Zhang, L., Kong, X., Lee, F., Tian, Y., & Meldrum, D. R. (2017) · Sensing and Bio-Sensing Research · doi:10.1016/j.sbsr.2017.03.004 · Cited by 16
Xu, X., Niu, X., Wu, S., Zou, X., & Pan, J. (2018) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2018.04.132 · Cited by 14
Manbohi, A., & Ahmadi, S. H. (2020) · Chemical Papers · doi:10.1007/s11696-020-01105-5 · Cited by 14
Wang, X., & Qin, W. (2012) · Chemical Communications · doi:10.1039/c2cc31020a · Cited by 14
Vasimalla, Y., Singh, L., Pradhan, H. S., Kumar, R., & Belwal, P. (2022) · Hybrid Advances · doi:10.1016/j.hybadv.2022.100005 · Cited by 13
The effects of blood glucose regulation on tear function tests in diabetic patients
Kan, S., Acar, U., Kızılgül, M., Beyazyıldız, E., Çankaya, A. B., Özçelik, O., & Özbek, M. (2017) · Journal Français d Ophtalmologie · doi:10.1016/j.jfo.2016.10.019 · Cited by 13
Mohammadinejad, A., Heydari, M., Oskuee, R. K., & Rezayi, M. (2021) · Critical Reviews in Analytical Chemistry · doi:10.1080/10408347.2021.1919986 · Cited by 13
“) was used as a cell to collect low volumes of tear fluid.([135])According to the problems of enzymatic methods and glucose measurement such as dependency of enzyme activity to environmental condition and fluctuation of glucose amount by lifestyle, as a proposal the combination of aptamer with this technique lead to p…”
MethodDevi, L. S., Paily, R. P., & Dasmahapatra, A. K. (2024) · Polymer · doi:10.1016/j.polymer.2024.127947 · Cited by 12
Antifouling hydrogel film based on a sandwich array for salivary glucose monitoring
Zhang, Z., Wang, S., Liu, G., Hu, D., Yang, B., Dai, Q., & Dou, Q. (2021) · RSC Advances · doi:10.1039/d1ra03517g · Cited by 12
Electrochemical Sensors and Biosensors: Their Basic Working Principles and Some Applications
Amaral, E. M. F., de Andrade, J. C. B., de Freitas, L. B., Teixeira, G. T. L., do Nascimento, J. P. L., Gelamo, R. V., Leite, N. B., & Moreto, J. A. (2022) · Revista Virtual de Química · doi:10.21577/1984-6835.20220112 · Cited by 12
“Devido à sua alta seletividade, sensibilidade, estabilidade, tempo de resposta rápido e baixo custo, os biossensores apresentam aplicações em diversos setores, tais como: alimentos (patógeno, aditivos), monitoramento ambiental (poluição tóxica) e diagnósticos clínicos (glicose no sangue, colesterol, dopamina).(30-40)…”
BackgroundOh, J.-M., & Chow, K. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b00552 · Cited by 12
Dhakal, R., Kim, E., Jo, Y.-H., Kim, S. S., & Kim, N. (2017) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2017.01.008 · Cited by 12
Paper integrated microfluidic contact lens for colorimetric glucose detection
İşgör, P. K., Abbasiasl, T., Das, R., Istif, E., Yener, U. C., & Beker, L. (2024) · Sensors & Diagnostics · doi:10.1039/d4sd00135d · Cited by 13
Ho, K. K. Y., Peng, Y.-W., Ye, M., Tchouta, L., Schneider, B., Hayes, M., Toomasian, J. M., Cornell, M. S., Rojas-Peña, Á., Charpie, J. R., et al. (2020) · Chemosensors · doi:10.3390/chemosensors8030056 · Cited by 11
Trends in Epidermal Stretchable Electronics for Noninvasive Long-term Healthcare Applications
Han, H., Reichmuth, A. M., Renz, A. F., Stauffer, F., Thielen, M., & Vörös, J. (2017) · International Journal of Automation and Smart Technology · doi:10.5875/ausmt.v7i2.1336 · Cited by 10
Bamatraf, S. M., Aldhaeebi, M. A., & Ramahi, O. M. (2022) · Progress In Electromagnetics Research Letters · doi:10.2528/pierl21110905 · Cited by 10
“Those fluids include blood [3], saliva [4], sweat [5, 6], urine [7], tears [8, 9], breath [10], and interstitial fluid [11].”
BackgroundZhou, Y., Ding, Z., Sun, Q., Chen, L., Wang, D., Bao, X., & Feng, X. (2024) · Journal of the American Chemical Society · doi:10.1021/jacs.4c03341 · Cited by 10
Graphene oxide–mediated electrochemistry of glucose oxidase on glassy carbon electrodes
Castrignanò, S., Valetti, F., Gilardi, G., & Sadeghi, S. J. (2015) · Biotechnology and Applied Biochemistry · doi:10.1002/bab.1392 · Cited by 10
Li, X., Ju, Y., Xu, Y., Wang, W., Dong, Y., Ma, Y., & Chen, X. (2013) · Analytica Chimica Acta · doi:10.1016/j.aca.2013.06.031 · Cited by 10
Lab‐on‐a‐Chip Platforms for Disease Detection and Diagnosis
Işıksaçan, Z., Guler, M. T., Kalantarifard, A., Asghari, M. H., & Elbüken, Ç. (2017) · Wiley eBooks · doi:10.1002/9781119065036.ch8 · Cited by 10
A Review: Non Invasive Sensing System for Detection Glucose Level
Bachache, L. N., Hasan, J. A., & Al-Neami, A. Q. (2021) · Journal of Physics Conference Series · doi:10.1088/1742-6596/1963/1/012125 · Cited by 12
“The volume of the tearing sample must be fixed with 1microletter for every sample [18, 19].”
BackgroundGuo, X., Mandelis, A., & Zinman, B. (2012) · Journal of Biophotonics · doi:10.1002/jbio.201200103 · Cited by 10
Echem methods and electrode types of the current in vivo electrochemical sensing
Song, Q., Li, Q., Yan, J., & Song, Y. (2022) · RSC Advances · doi:10.1039/d2ra01273a · Cited by 9
Deng, H., Shen, W., & Gao, Z. (2013) · ChemPhysChem · doi:10.1002/cphc.201200961 · Cited by 9
Zhang, F., Xu, W., Deng, Z., & Huang, J. (2025) · Frontiers in Medicine · doi:10.3389/fmed.2024.1487981 · Cited by 7
Sani, M. H., Sami, P., Shen, C. Q., & Behzadnia, H. (2021) · Journal of Research in Science Engineering and Technology · doi:10.24200/jrset.vol9iss03pp64-75 · Cited by 7
“65 Shokouhmand, 2020), filters (Hosseinzadeh Sani, Ghanbari, & Saghaei, 2020; Robinson & Nakkeeran, 2013), logical gate (Jiang, Liu, Zhang, & Kong, 2015; Karkhanehchi, Parandin, & Zahedi, 2017; Sani, Tabrizi, Saghaei, & Karimzadeh, 2020), decoder (Mehdizadeh, Soroosh, & Alipour-Banaei, 2016; Parandin, Karkhanehchi, Na…”
Glucose Sensing for Diabetes Monitoring: From Invasive to Wearable Device
Chhiba, L., Zaher, B., Sidqui, M., & Marzak, A. (2020) · Lecture notes in intelligent transportation and infrastructure · doi:10.1007/978-3-030-37629-1_27 · Cited by 10
Designing Minimally Invasive Preocular Contact Tips for Potential Application in Tear Collection
Lee, S. H., Cho, Y. C., Nam, D. Y., Huh, B. K., Kim, S.-N., Lee, D. H., Park, C. G., & Choy, Y. B. (2018) · Cornea · doi:10.1097/ico.0000000000001603 · Cited by 7
Quantitative Determination of Tear Glucose Using Paper Based Microfluidic Devices
Kim, H.-J., Jeong, S., & Noh, H. (2015) · Journal of the Korean Chemical Society · doi:10.5012/jkcs.2015.59.1.88 · Cited by 6
Xie, B., Wang, K., Li, B., Huang, R., Xu, Z., & Li, X. (2023) · Journal of Materials Chemistry B · doi:10.1039/d3tb00034f · Cited by 7
Cavity enhanced lab-on-fiber optrode for ultra-sensitive pH monitoring
Gambino, F., Cicatiello, P., Giaquinto, M., Micco, A., Aliberti, A., Cusano, A. M., Ricciardi, A., & Cusano, A. (2022) · Sensors & Diagnostics · doi:10.1039/d1sd00071c · Cited by 6
Hu, J., Brendle, S. A., Balogh, K. K., Bywaters, S. M., & Christensen, N. D. (2014) · Journal of General Virology · doi:10.1099/vir.0.064154-0 · Cited by 5
“A recent study demonstrated a correlation between glucose levels from matching tear and serum samples in diabetic patients (Yan et al., 2011).”
BackgroundSmart Health – Potential and Pathways: A Survey
Arulananthan, C., & Hanifa, S. M. (2017) · IOP Conference Series Materials Science and Engineering · doi:10.1088/1757-899x/225/1/012065 · Cited by 7
“Reports on measuring blood glucose from tear drops [25].”
BackgroundA highly sensitive non-enzymatic glucose electrochemical sensor based on NiO nanohives
Vu, T. O., Chu, T., & Nguyen, D. H. (2021) · Advances in Natural Sciences Nanoscience and Nanotechnology · doi:10.1088/2043-6262/ac4995 · Cited by 5
Design of a non-invasive blood glucose meter connected to an Android diabetes monitoring application
Chhiba, L., Sidqui, M., & Marzak, A. (2019) · International Conference on Smart City Applications (SCA) · doi:10.1145/3368756.3369022 · Cited by 5
Mondal, S., & Subramaniam, C. (2021) · Nanoscale Advances · doi:10.1039/d0na01042a · Cited by 5
Tear glucose is associated with the presence and severity of diabetic retinopathy
Cao, N., Feng, L., Xu, W., He, F., Xiao, C., Liu, Y., Xu, W., Cui, J., Guo, Y., Yao, L., et al. (2025) · International Journal of Retina and Vitreous · doi:10.1186/s40942-025-00636-x · Cited by 4
“The correlation between TG and BG concentration had been investigated using the amperometric sensors [20].”
Li, Y., Han, J. G., Jiang, L., Li, F., Li, K., & Dong, Y. (2014) · Analytical Letters · doi:10.1080/00032719.2014.974056 · Cited by 4
“Consequently, many glucose monitoring technologies have been developed to help in the management of diabetes (Yan et al. 2011).”
BackgroundMuñoz-Hernández, A. M., Galbis-Estrada, C., Santos-Bueso, E., Cuiña‐Sardiña, R., Díaz‐Valle, D., Gegúndez-Fernández, J. A., Pinazo-Duran, M. D., & Benitez-del-Castillo, J. M. (2016) · Archivos de la Sociedad Española de Oftalmología (English Edition) · doi:10.1016/j.oftale.2015.09.007 · Cited by 4
Jang, S., Wang, Y., & Jang, A. (2022) · Springer series on bio- and neurosystems · doi:10.1007/978-3-030-99728-1_2 · Cited by 4
Sensor Ecosystems for Health Self‐Monitoring: Focus on Diabetes Management
Ganguly, A., Churcher, N. K. M., Upasham, S., Dhamu, V. N., & Prasad, S. (2023) · Wiley eBooks · doi:10.1002/9783527834266.ch44 · Cited by 4
Iyer, M. (2018) · RIT Scholar Works (Rochester Institute of Technology) · Cited by 5
A Potentially Versatile Enzyme Sensor Platform: Enzyme-Loaded, Tagged, Porous Polymeric Nanocapsules
Hambly, B. P., Sears, C. K., Pendley, B. D., Thompson, L. L., & Lindner, E. (2024) · ACS Sensors · doi:10.1021/acssensors.3c01980 · Cited by 3
Status of Soft Contact Lens Biosensor Development for Tear Sugar Monitoring: A Review
Kudo, H., Arakawa, T., & Mitsubayashi, K. (2014) · Electronics and Communications in Japan · doi:10.1002/ecj.11591 · Cited by 3
Muñoz-Hernández, A. M., Galbis-Estrada, C., Santos-Bueso, E., Cuiña‐Sardiña, R., Díaz‐Valle, D., Gegúndez-Fernández, J. A., Pinazo-Duran, M. D., & Benitez-del-Castillo, J. M. (2015) · Archivos de la Sociedad Española de Oftalmología · doi:10.1016/j.oftal.2015.09.010 · Cited by 3
Noninvasive and Ultrasensitive Glucose Detection via Electrolyte-Gated IGZO Thin-Film Transistors
Hwang, C., Park, S., & Kim, D. Y. (2026) · IEEE Sensors Journal · doi:10.1109/jsen.2026.3661008 · Cited by 2
“Although a quantifiable correlation between blood and tear glucose concentrations has been established, glucose levels in tear fluid are approximately 30–50 times lower than those in blood [18], [21], [22], [23].”
Status of Soft-Contact Lens Biosensor Development towards Tear Sugar Monitoring: A Review
Kudo, H., Arakawa, T., & Mitsubayashi, K. (2012) · IEEJ Transactions on Sensors and Micromachines · doi:10.1541/ieejsmas.132.451 · Cited by 2
Enhanced non-enzymatic glucose biosensor of Ga-doped ZnO nanorods
Peng, W.-C., Wang, Z.-H., Yang, C.-C., Huang, C.-S., Su, Y., & Ruan, J. (2017) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2274916 · Cited by 2
Fundamentals and applications of enzymatic bioelectrocatalysis
Simoska, O., Lee, Y. S., & Minteer, S. D. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-823144-9.00057-1 · Cited by 2
Miniaturized bioelectrochemical devices
Ghoorchian, A., Moradi, M., Aghajani, S., Afkhami, A., Madrakian, T., Thomas, S., Nguyen, T. A., & Ahmadi, M. (2022) · Elsevier eBooks · doi:10.1016/b978-0-12-823727-4.00009-2 · Cited by 2
Scoping review of subcutaneous glucose monitoring techniques
Hrubá, E., Kubíček, J., & Augustýnek, M. (2025) · Measurement · doi:10.1016/j.measurement.2025.119940 · Cited by 1
Ghosh, T. N., Kumar, S., Jalali, S., Belenje, A., Mahajan, A., Kaur, I., & Singh, S. G. (2025) · IEEE Sensors Journal · doi:10.1109/jsen.2025.3645254 · Cited by 1
“Recent advancements in the development of electrochemical biosensors [4], [5], [6], [7], [8], fluorescence-based sensors [9], [10], [11], surface plasmon resonance (SPR)-based biosensors [12], [13], [14], and colorimetric sensors [15], [16] represent a cutting-edge approach for noninvasive…”
Recent advances in smart contact lens biosensors for non-invasive tear glucose detection
An, S., Duan, J., Zhou, C., & Zeng, D. (2026) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2026.118122 · Cited by 1
Non‐invasive Glucose Biosensors Based on Nanomaterials
Faridbod, F., Ganjali, M. R., Larijani, B. A., & Norouzil, P. (2015) · Wiley eBooks · doi:10.1002/9781118998922.ch10 · Cited by 1
Development of paper-gate transistor toward direct detection from microbiological fluids
Kajisa, T., & Sakata, T. (2017) · Japanese Journal of Applied Physics · doi:10.7567/jjap.56.04cm07 · Cited by 1
Perspectives on Non-Invasive Glucose Sensing using Flexible Hybrid-Printed Electronics Systems
Bassant, H., Refaei, A., Fayomi, C. J. B., & Boukadoum, M. (2019) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2019.8702818 · Cited by 1
“Sensing lenses with direct contact with tears can be used for continuous monitoring, exploiting the correlation between the lacrimal and sanguine levels [10, 11].”
Technological Innovations and the Advancement of Preventive Healthcare for Society 5.0
Mkhize‐Kwitshana, Z. L., Naidoo, P., Fourie, L., Ecker, R., & Dlamini, Z. (2023) · Springer eBooks · doi:10.1007/978-3-031-36461-7_4 · Cited by 1
Miniaturized electrochemical devices
Afkhami, A., Madrakian, T., & Ahmadi, M. (2023) · Elsevier eBooks · doi:10.1016/b978-0-323-91741-4.00010-5 · Cited by 1
Marie, H. (2013) · theses.fr (ABES) · doi:10.70675/677e4647zbea5z4f16za939zb3923b04284f · Cited by 1
Glucose Sensor and Its Potential Directions
Cheng, C., Kuan, C.-M., & Chen, C. (2015) · Springer eBooks · doi:10.1007/978-3-319-19737-1_4
Simple Methods of Detecting Disease Factors in Tears and Bloods
Jeong, S., Kim, K., & Noh, H. (2016) · Journal of Life Sciences and Technologies · doi:10.18178/jolst.4.1.26-29
Soft mechanical and biochemical sensors
Koh, A. S., & Vikas, V. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-819246-7.00004-8
三林, 浩. (2014) · JAPANES JOURNAL OF MEDICAL INSTRUMENTATION · doi:10.4286/jjmi.84.631
Correlating the Effect of Dynamic Variability in the Sensor Environment on Sensor Design
Bhushan, P. (2021) · FIU - Digital Commons (Florida International University) · doi:10.25148/etd.fidc009698 · Cited by 1
The Clinical Accuracy of Non-Invasive Glucose Monitoring for ex vivo Artificial Pancreas
Levy, M. M., & Levy, Y. M. (2016) · Journal of Emerging Investigators · doi:10.59720/16-008
“the glucose concentration in the tears (10).”
BackgroundDesagani, D., Cetó, X., del Valle, M., & Ben‐Yoav, H. (2024) · Elsevier eBooks · doi:10.1016/b978-0-443-15684-7.00025-7
Park, S., Nam, D. Y., 전희재, Han, J. H., Jang, D., Hwang, J., Park, Y.-S., Han, Y.-G., Choy, Y. B., & Lee, D. Y. (2023) · Figshare · doi:10.6084/m9.figshare.24864169
Perovskite-Based Wearable Sensors
Darab, M., Bakhshayesh, A., & Tajabadi, M. (2025) · Engineering materials · doi:10.1007/978-3-032-03931-6_5
Advances in Smart Contact Lenses for Tear Glucose Monitoring: A Review
An, S., & Zeng, D. (2026) · Frontiers in Science and Engineering · doi:10.54691/h3rp8p51
“As Yan et al. fixed GOx on a chitosan-carbon paste electrode, the detection limit reached 0.1 mM, but enzyme activity decreased by 40% after 24 hours of wearing [27].”
InfluentialDevelopment of a selective and energy-autonomous lactic acid monitoring system
Lin, S.-H. (2024) · theses.fr (ABES) · doi:10.70675/4cf84d2cz15b0z4e58z8e1fz23980ac846ec
Peng, Y., Huang, X., Zhang, X., Jiang, Y., Zhang, X., Zeng, X., Sun, F., Liu, P., Zhou, J., Luo, Z., et al. (2026) · Interdisciplinary medicine · doi:10.1002/inmd.70132
Electrochemical Tablet-based Sensor for Stable and Reproducible Biosensing
Saebi, R. (2025) · Spectrum Research Repository (Concordia University)
Tear biomarkers in veterinary species: An updated review with special focus on dogs and cats
Vaquer-Alcaraz, L., Muñoz-Prieto, A., Pardo-Marín, L., & del Río, A. Á. B. (2026) · The Veterinary Journal · doi:10.1016/j.tvjl.2026.106786 · Cited by 1
Metadata
- Authors
- Yan, Q., Peng, B., Su, G., Cohan, B. E., Major, T. C., & Meyerhoff, M. E.
- Affiliations
- University of Michigan (US)
- Corresponding author
- Meyerhoff, Mark E.
- Venue
- Analytical Chemistry
- Publisher
- American Chemical Society (ACS)
- Published
- 2011-10-12
- Volume
- 83
- Issue
- 21
- Pages
- 8341-8346
- ISSN
- 0003-2700 (ISSN-L)
- 1520-6882
- OpenAlex
- W2018860363
- Semantic Scholar
- 2bac2226ca694c079ac35a58255ced63f9971be4
- Cited by
- 185
- References
- 26
- Citations by year
- 2026: 5 · 2025: 8 · 2024: 10 · 2023: 13 · 2022: 13 · 2021: 21 · 2020: 15 · 2019: 14 · 2018: 13 · 2017: 20 · 2016: 14 · 2015: 11 · 2014: 17 · 2013: 6 · 2012: 5
- FWCI
- 6.27
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Analytical Chemistry (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Physical Sciences › Chemistry › Electrochemistry › Electrochemical Analysis and Applications
- Physical Sciences › Chemistry › Analytical Chemistry › Analytical Chemistry and Sensors
- Keywords
- tear glucose, amperometric glucose sensor, glucose oxidase immobilization, hydrogen peroxide detection, Nafion, ascorbic acid interference, uric acid interference, non-invasive glucose monitoring, rabbits
- MeSH terms
- Animals, Ascorbic Acid, Blood Glucose, Enzymes, Immobilized, Glucose, Glucose Oxidase, Hydrogen Peroxide, Iridium, Platinum, Rabbits, Resorcinols, Tears, Uric Acid, Biosensing Techniques, Electrochemical Techniques
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.