Multi-laboratory assessment of reproducibility, qualitative and quantitative performance of SWATH-mass spectrometry
Open accessPublished version of a preprint
The DOI matches its Crossref record; 82 of the 82 references it cites were found in a registry or an index; 19 of the 48 sources its claims cite were checked for support.
Abstract
Quantitative proteomics employing mass spectrometry is an indispensable tool in life science research. Targeted proteomics has emerged as a powerful approach for reproducible quantification but is limited in the number of proteins quantified. SWATH-mass spectrometry consists of data-independent acquisition and a targeted data analysis strategy that aims to maintain the favorable quantitative characteristics (accuracy, sensitivity, and selectivity) of targeted proteomics at large scale. While previous SWATH-mass spectrometry studies have shown high intra-lab reproducibility, this has not been evaluated between labs. In this multi-laboratory evaluation study including 11 sites worldwide, we demonstrate that using SWATH-mass spectrometry data acquisition we can consistently detect and reproducibly quantify >4000 proteins from HEK293 cells. Using synthetic peptide dilution series, we show that the sensitivity, dynamic range and reproducibility established with SWATH-mass spectrometry are uniformly achieved. This study demonstrates that the acquisition of reproducible quantitative proteomics data by multiple labs is achievable, and broadly serves to increase confidence in SWATH-mass spectrometry data acquisition as a reproducible method for large-scale protein quantification.
Overview
CoChat Summary
TL;DR
This multi-laboratory study shows that SWATH-MS data acquisition can reproducibly detect and quantify thousands of proteins across 11 sites, with quantitative performance approaching targeted proteomics. The authors conclude that SWATH-MS is suitable for large-scale, reproducible protein quantification across laboratories.
Key findings
- Across 229 SWATH-MS files from 11 laboratories, 4,984 proteins were detected at 1% FDR; the median per file was 4,548 proteins, and 4,077 proteins were detected in >80% of all samples.
- Median repeatability was 91.6% at the protein level and 79.5% at the peptide level across all sites; within-site median repeatability ranged from 90.0–98.2% (protein) and 79.5–95.5% (peptide).
- For SIS peptides in sample S4, median intra-day CV was 5.5 ± 2.9%, inter-day CV was 8.9 ± 11.1%, and inter-site CV was 47.3 ± 13.9% without normalization, reduced to 21.3 ± 10.3% after normalization.
- Protein-level CVs for the 4,077 proteins detected in >80% of samples were 8.3 ± 16.2% intra-day, 11.9 ± 17.2% inter-day, and 22.0 ± 17.4% inter-site; median Pearson correlation across all runs was 0.940, with within-site median 0.971 and minimum pairwise 0.868.
Methods
- Benchmarking sample: 30 stable isotope-labeled standard peptides partitioned into five groups and serially diluted into 1 µg HEK293 cell lysate, creating samples S1–S5 spanning 0.012–10,000 fmol on column; distributed to 11 laboratories worldwide.
- Acquisition: SWATH-MS on SCIEX TripleTOF 5600/5600+ systems using 64 variable-width Q1 windows; samples S1–S5 acquired with S4 in technical triplicate on days 1, 3, and 5, yielding 229 SWATH-MS files.
- Analysis: OpenSWATH with a 10,000+ human protein spectral library, PyProphet for 1% FDR control at peptide query and protein levels, aLFQ for protein abundance inference, MultiQuant for SIS peptide XIC analysis, and median normalization based on HEK293 background peptides.
Limitations
- Stated: the study did not evaluate variance in sample preparation between sites, so no conclusions can be made on that topic; only one instrument type/model was used, limiting tested variables mainly to site of data acquisition.
- Stated: high peptide loads at the upper end of the dilution series caused some MS signal saturation; a newer instrument platform (TripleTOF 6600) would be expected to reduce this.
- Stated/evident: no internal standard was available for protein-level dynamic range, so U2OS copy-number data from a different cell line was used as a surrogate; the study is also limited to HEK293 background and SIS peptides, so generalizability to other sample types and platforms is not established.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures






Verification
Claims the paper makes
10 Valid · 1 Not valid · 16 Not checked · 81 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“In general, these studies confirmed that a high degree of reproducibility is difficult to achieve for complex samples10.”
§ Multi-laboratory assessment of reproducibility, qualitative and quantitative performance of SWATH-mass spectrometry > Introduction
What the sources say
- Partly supportedcites [10] Rudnick PA Performance metrics for liquid chromatography-tandem mass spectrometry systems in proteomics analyses Mol. Cell. Proteomics 2010 9 225 241 10.1074/mcp.M900223-MCP200 19837981 PMC2830836 sourceExperimental — accuracy not yet measured
The Abstract identifies stochastic sampling of lower-abundance peptides as a source of variability in LC-MS/MS identifications, which supports the DDA-related concern. But the Results report stable metrics across analytical replicates and highly reproducible chromatography metrics across instruments; they do not establish that high reproducibility is generally difficult to achieve for complex samples. Page numbers are unavailable in our copy. (LLM, conf=0.78)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“SWATH-MS and related DIA approaches have achieved a high degree of reproducibility in intra-laboratory studies in a variety of research questions such as interaction proteomics40, 41, plasma proteomics42, tissue proteomics43, microbial proteomics44, 45, pre-clinical toxicology9, analysis of genetic reference strains46, and many others.”
§ Multi-laboratory assessment of reproducibility, qualitative and quantitative performance of SWATH-mass spectrometry > Introduction
What the sources say
- Supportedcites [40] Collins BC Quantifying protein interaction dynamics by SWATH mass spectrometry: application to the 14-3-3 system Nat. Methods 2013 10 1246 1253 10.1038/nmeth.2703 24162925 source
Our copy of Collins et al. (2013) is precisely the intra-laboratory interaction-proteomics application the citing sentence offers this reference for: SWATH-MS (a DIA implementation) applied to affinity-purified 14-3-3β complexes, yielding an essentially complete quantitative matrix for 1,967 proteins across 27 samples in biological triplicate (Generation and analysis of AP-SWATH maps; Abstract: dynamics 'robustly followed'). The Introduction states SWATH-MS 'provides SRM-like performance in terms of quantitative accuracy, data completeness and dynamic range', against the explicit contrast that DDA identification is not reproducible while SRM reproducibly targets a predetermined peptide set. The paper's detailed quantitative-reproducibility figures are referred to its Supplementary Results from the same section, so the magnitude is asserted here qualitatively rather than numerically, but the source plainly supports the exemplified point. (LLM, conf=0.80)
- Supportedcites [41] Lambert J-P Mapping differential interactomes by affinity purification coupled with data-independent mass spectrometry acquisition Nat. Methods 2013 10 1239 1245 10.1038/nmeth.2702 24162924 PMC3882083 source
The citation is attached to the 'interaction proteomics' instance of the citing sentence, and our copy is exactly an interaction-proteomics (AP-SWATH) study reporting reproducibility. Results ('A pipeline for quantitative interactome monitoring', p. ~3) give 79% of peptides and 87% of proteins quantified within 20% CV across CDK4 AP biological replicates (plus three technical replicates each), a higher proportion of peptides at CV ≤ 20% than DDA (82.1% vs 74.5%), and conclude the extraction/normalization pipeline 'is reliable'. The Discussion states the pipeline studies interactome changes 'in an unbiased and reproducible manner', and the Abstract calls AP-SWATH 'a robust label-free approach'. The 'related DIA approaches … variety of research questions' breadth rests on the other citations, not this one, so nothing here is unsupported. (LLM, conf=0.85)
- Supportedcites [42] Liu Y Quantitative variability of 342 plasma proteins in a human twin population Mol. Syst. Biol. 2015 11 786 10.15252/msb.20145728 25652787 PMC4358658 source
The citation is offered as an exemplar of SWATH-MS/DIA applied to plasma proteomics with high reproducibility, and our copy is exactly that: the Results subsection 'The reproducible quantification of plasma proteins across a longitudinal twin cohort by SWATH-MS' reports median CVs of 7.2% (technical) and 14.2% (whole-process) replicates with 84.5%/76.0% of proteins below 25% CV, and states the study 'significantly exceeds the previous studies in terms of reproducibility and quantitative accuracy'. All measurements were run on the same in-house LC-MS/MS system and analysed on an internal computing cluster (Materials and Methods), i.e. a single-laboratory study, matching the 'intra-laboratory' qualifier. The list of other application areas (interaction, tissue, microbial proteomics, toxicology, genetic reference strains) belongs to the other citations and the citing authors' own framing, not to this source. (LLM, conf=0.88)
- Supportedcites [43] Guo T Rapid mass spectrometric conversion of tissue biopsy samples into permanent quantitative digital proteome maps Nat. Med. 2015 21 407 413 10.1038/nm.3807 25730263 PMC4390165 source
The cited sentence attaches this source to the 'tissue proteomics' instance of high reproducibility in intra-laboratory SWATH-MS/DIA studies, and our copy is exactly that: the Abstract (p. 1) reports converting 18 renal biopsy samples into SWATH-MS maps and quantifying >2,000 proteins 'with a high degree of reproducibility across all samples', and the Discussion states the pipeline performs 'high throughput, quantitative and highly reproducible proteomic analyses of clinical tumor biopsies' with 'a median CV of technical replicates below 20%'. The Results (Generation of SWATH-MS maps) quantify that reproducibility as a single-laboratory technical-replicate experiment (median CV 10-12% for repeat MS injections, 24% overall including PCT digestion), matching the 'intra-laboratory' qualifier. The wider list of application areas (interaction, plasma, microbial proteomics, toxicology, genetic reference strains) is carried by the sentence's other citations, not by this one, so nothing here is overstated. (LLM, conf=0.85)
boosted- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“Notably, the spectral library and peptide query parameters we used to perform the analysis of the SWATH-MS data were previously published49 and built by a single lab independent of the current study, illustrating the generic applicability of such spectral libraries.”
§ Multi-laboratory assessment of reproducibility, qualitative and quantitative performance of SWATH-mass spectrometry > Discussion
What the sources say
- Supportedcites [49] Rosenberger G A repository of assays to quantify 10,000 human proteins by SWATH-MS Sci. Data 2014 1 140031 10.1038/sdata.2014.31 25977788 PMC4322573 source
The Background & Summary of Rosenberger et al. (2014) presents 'a generic large-scale human assay library' built by the authors' group from 331 fraction measurements, stating its results are 'applicable across laboratories', and Data Record 2 confirms that both spectral libraries (SpectraST format) and assay libraries (CSV, TraML, converted to formats for OpenSWATH/PeakView/Skyline/Spectronaut) were published. The Technical Validation section 'Portability of the combined assay library to different sample types and laboratories' explicitly demonstrates the library working on data acquired in a different laboratory, at a different time and under different chromatographic conditions — i.e., the generic applicability the citing sentence attributes to it. Only the exact term 'peptide query parameters' is absent from our copy (the paper describes the equivalent assay-library content: precursor/fragment ion m/z, normalized RT and relative ion intensities), which is a wording difference rather than an unsupported element. (LLM, conf=0.85)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Here, we highlight an advantage of SWATH-MS data; i.e., as with MS1/DDA-based quantification and, unlike more classical targeted methods such as SRM, there are large numbers of peptides available for global normalization that can be used in sample types where the assumptions underlying this type of normalization are valid65, 66.”
§ Multi-laboratory assessment of reproducibility, qualitative and quantitative performance of SWATH-mass spectrometry > Discussion
What the sources say
- Partly supportedcites [65] Välikangas, T., Suomi, T. & Elo, L. L. A systematic evaluation of normalization methods in quantitative label-free proteomics. Brief Bioinform. 10.1093/bib/bbw095 (2016). PMC5862339 27694351 sourceExperimental — accuracy not yet measured
Our copy of Välikangas et al. (2016) supports only the qualifier of the claim: the Discussion states that all the normalization methods studied, excluding EigenMS, 'assume that only a small portion of the proteins are differentially expressed between samples and force the total intensity levels of the samples to be on the same level', which 'might be problematic if in fact a large number of proteins are differentially expressed' — i.e. global normalization is usable only where its assumptions hold. The Introduction and Materials and methods (Description of the data sets) frame this entirely in shotgun/DDA label-free data (UPS1, CPTAC, SGSD, mouse liver). The distinctive part of the claim — that SWATH-MS, unlike classical targeted SRM, yields large numbers of peptides available for global normalization — is absent: no passage in our copy mentions SWATH, DIA-based targeted quantification or SRM/MRM at all (searched: SWATH, data-independent acquisition targeted, SRM/MRM, number of peptides for global normalization). (LLM, conf=0.65)
- Partly supportedcites [66] Karpievitch YV Dabney AR Smith RD Normalization and missing value imputation for label-free LC-MS analysis BMC Bioinformatics 2012 13 S5 10.1186/1471-2105-13-S16-S5 23176322 PMC3489534 sourceExperimental — accuracy not yet measured
Our copy supports the normalization half of the attributed point: Methods > Normalization states that global adjustment centers log-intensity distributions on a constant and rests on 'an assumption... that most peptide abundances do not change', and the Background notes high-throughput MS quantifies 'thousands of peptides', which is the label-free shotgun setting the citing authors equate with MS1/DDA and SWATH. However, the paper never mentions SWATH-MS, DIA, or a comparison with targeted SRM/MRM (searched: 'SWATH data independent acquisition targeted MS/MS SRM MRM'; only Background/Methods normalization passages returned), so the specific claim of an advantage over classical targeted methods is not in our copy. It also frames validity of the normalization assumption as a property of the data rather than of particular 'sample types'. (LLM, conf=0.72)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“In this multi-laboratory evaluation study including 11 sites worldwide, we demonstrate that using SWATH-mass spectrometry data acquisition we can consistently detect and reproducibly quantify >4000 proteins from HEK293 cells.”
§ Multi-laboratory assessment of reproducibility, qualitative and quantitative performance of SWATH-mass spectrometry > Abstract
boostedstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
46 more claims
Create a free account to see all 108 claims and to verify the 29 references still open.
- It may state two separate claims as one. 42
- Probably not a substantial claim. 21
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Proteomic technologies, including mass spectrometry–based proteomics, aptamer-based arrays (SomaLogic), and proximity extension assay technology (Olink), are now available for the large-scale quantification and validation of thousands of proteins in human biological fluids or tissues (122–127).”
From Connecting aging biology and inflammation in the omics era (2022) · cites this paper as [124]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Multi-laboratory assessment of reproducibility, qualitative and quantitative performance of SWATH-mass spectrometry > Results > Reproducibility of quantification
“The median for site intra-day and inter-day CVs (expressed as median ± standard deviation) were 5.5 ± 2.9% and 8.9 ± 11.1%, respectively”
§ Multi-laboratory assessment of reproducibility, qualitative and quantitative performance of SWATH-mass spectrometry > Results > Reproducibility of quantification
“Normalization also reduced the median within site inter-day CV from 8.9 ± 11.0 to 5.8 ± 5.4%”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Multi-laboratory assessment of reproducibility, qualitative and quantitative performance of SWATH-mass spectrometry > Results > Linearity and dynamic range
“A representative example for a single site is shown in Fig 4a (remaining sites in Supplementary Fig. 15; equivalent plots separated by peptide are shown in Supplementary Fig. 16; source data in Supplementary Data 8).”
§ Multi-laboratory assessment of reproducibility, qualitative and quantitative performance of SWATH-mass spectrometry > Results > Linearity and dynamic range
“a The response curves for each of the 30 SIS peptides for Site 1 were determined and plotted together (corresponding plots for all other sites are shown in Supplementary Fig. 13).”
Citations
Cites
Statistical significance for genomewide studies
Storey, J. D., & Tibshirani, R. (2003) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1530509100 · Cited by 9,692
Cox, J., Hein, M. Y., Luber, C. A., Paron, I., Nagaraj, N., & Mann, M. (2014) · Molecular & Cellular Proteomics · doi:10.1074/mcp.m113.031591 · Cited by 5,180
Raise standards for preclinical cancer research
Begley, C. G., & Ellis, L. M. (2012) · Nature · doi:10.1038/483531a · Cited by 2,979
Gillet, L. C., Navarro, P., Tate, S., Röst, H., Selevsek, N., Reiter, L., Bonner, R., & Aebersold, R. (2012) · Molecular & Cellular Proteomics · doi:10.1074/mcp.o111.016717 · Cited by 2,877
Self-citationSemi-supervised learning for peptide identification from shotgun proteomics datasets
Käll, L., Canterbury, J. D., Weston, J., Noble, W. S., & MacCoss, M. J. (2007) · Nature Methods · doi:10.1038/nmeth1113 · Cited by 2,424
A draft map of the human proteome
Kim, M.-S., Pinto, S. M., Getnet, D., Nirujogi, R. S., Manda, S. S., Chaerkady, R., Madugundu, A. K., Kelkar, D. S., Isserlin, R., Jain, S., et al. (2014) · Nature · doi:10.1038/nature13302 · Cited by 2,198
The Proteomics Identifications (PRIDE) database and associated tools: status in 2013
Vizcaíno, J. A., Côté, R. G., Csordas, A., Dianes, J. A., Fabregat, A., Foster, J. M., Griss, J., Alpi, E., Birim, M., Contell, J., et al. (2012) · Nucleic Acids Research · doi:10.1093/nar/gks1262 · Cited by 1,918
Believe it or not: how much can we rely on published data on potential drug targets?
Prinz, F., Schlange, T., & Asadullah, K. (2011) · Nature Reviews Drug Discovery · doi:10.1038/nrd3439-c1 · Cited by 1,888
Mass-spectrometry-based draft of the human proteome
Wilhelm, M., Schlegl, J., Hahne, H., Gholami, A. M., Lieberenz, M., Savitski, M. M., Ziegler, E., Butzmann, L., Gessulat, S., Marx, H., et al. (2014) · Nature · doi:10.1038/nature13319 · Cited by 1,820
Peterson, A. C., Russell, J. D., Bailey, D. J., Westphall, M. S., & Coon, J. J. (2012) · Molecular & Cellular Proteomics · doi:10.1074/mcp.o112.020131 · Cited by 1,288
Research works printed but not linked: 1
- Ting YS Peptide-centric proteome analysis: an alternative strategy for the analysis of tandem mass spectrometry data Mol. Cell. Proteomics 2015 14 2301 2307 10.1074/mcp.O114.047035 26217018 PMC4563716 · DOI · Paper page
Cited by
Items 101–200 of 601
Cross-compartment signal propagation in the mitotic exit network
Zhou, X., Li, W., Liu, Y., & Amon, A. (2021) · eLife · doi:10.7554/elife.63645 · Cited by 53
Self-citationTurnover atlas of proteome and phosphoproteome across mouse tissues and brain regions
Li, W., Dasgupta, A., Yang, K., Wang, S., Kumar, N. H., Chepyala, S. R., Yarbro, J. M., Hu, Z., Šalovská, B., Fornasiero, E. F., et al. (2025) · Cell · doi:10.1016/j.cell.2025.02.021 · Cited by 58
Self-citationPérez–Sánchez, R., Carnero-Morán, Á., Valero, M. L., & Oleaga, A. (2022) · Parasites & Vectors · doi:10.1186/s13071-021-05118-1 · Cited by 53
“Several studies have shown that SWATH-MS may outperform DDA LC–MS/MS by increasing the sensitivity and reproducibility of protein and peptide identification across multiple replicates [25, 48, 49].”
BackgroundIntegrating phosphoproteomics into kinase-targeted cancer therapies in precision medicine
Wu, X., Xing, X., Dowlut, D., Zeng, Y., Liu, J., & Liu, X. (2018) · Journal of Proteomics · doi:10.1016/j.jprot.2018.03.033 · Cited by 60
Schilling, B., Basisty, N., Christensen, D. G., Sorensen, D. J., Orr, J. S., Wolfe, A. J., & Rao, C. V. (2019) · Journal of Bacteriology · doi:10.1128/jb.00768-18 · Cited by 49
“could be integrated in other biological replicates (19, 20).”
BackgroundSelf-citationReview of Batch Effects Prevention, Diagnostics, and Correction Approaches
Čuklina, J., Pedrioli, P. G. A., & Aebersold, R. H. (2019) · Methods in molecular biology · doi:10.1007/978-1-4939-9744-2_16 · Cited by 54
Self-citationNeri, F., Basisty, N., Desprez, P., Campisi, J., & Schilling, B. (2021) · Current Protocols · doi:10.1002/cpz1.32 · Cited by 48
“…and validation of cultured senescent cells, collection and processing of conditioned medium containing the SASP, and discovery and quantification of SASP proteins using a comprehensive DIA-MS-based acquisition workflow (Collins et al., 2017; Gillet et al., 2012; Schilling, Gibson, & Hunter, 2017).”
MethodSelf-citationKalia, N. P., Singh, S., Hards, K., Cheung, C., Sviriaeva, E. N., Banaei‐Esfahani, A., Aebersold, R. H., Berney, M., Cook, G. M., & Pethe, K. (2023) · Cell Reports · doi:10.1016/j.celrep.2023.112444 · Cited by 51
Self-citationEGFR targeting PhosTACs as a dual inhibitory approach reveals differential downstream signaling
Hu, Z., Chen, P., Li, W., Krone, M. W., Zheng, S., Saarbach, J., Velasco, I. U., Hines, J. F., Liu, Y., & Crews, C. M. (2024) · Science Advances · doi:10.1126/sciadv.adj7251 · Cited by 52
“As described in the previous study ( 52 ), the cell samples were lysed by sonication at 4 ° for two cycles (1 min per cycle) using a VialTweeter device (Hielscher-Ultrasound Technology) ( 75 , 76 ) and then centri-fuged at 20,000 g for 1 hour to remove the insoluble material.”
Self-citationTurner, N. P., Abeysinghe, P., Cheung, K. A. K., Vaswani, K., Logan, J., Sadowski, P. G., & Mitchell, M. D. (2022) · Proteomes · doi:10.3390/proteomes10020019 · Cited by 50
Mass Spectrometry–Based Proteomics of Epithelial Ovarian Cancers: A Clinical Perspective
Qian, L., Sun, R., Xue, Z., & Guo, T. (2023) · Molecular & Cellular Proteomics · doi:10.1016/j.mcpro.2023.100578 · Cited by 54
Huang, P., Kong, Q., Gao, W., Chu, B., Li, H., Mao, Y., Cai, Z., Xu, R., & Tian, R. (2020) · Analytica Chimica Acta · doi:10.1016/j.aca.2020.06.049 · Cited by 47
Gharibi, H., Ashkarran, A. A., Jafari, M. T., Voke, E., Landry, M. P., Saei, A. A., & Mahmoudi, M. (2024) · Nature Communications · doi:10.1038/s41467-023-44678-x · Cited by 47
GproDIA enables data-independent acquisition glycoproteomics with comprehensive statistical control
Yang, Y., Yan, G., Kong, S., Wu, M., Yang, P., Cao, W., & Qiao, L. (2021) · Nature Communications · doi:10.1038/s41467-021-26246-3 · Cited by 46
A comprehensive spectral assay library to quantify the Escherichia coli proteome by DIA/SWATH-MS
Midha, M. K., Kusebauch, U., Shteynberg, D. D., Kapil, C., Bader, S. L., Reddy, P. J., Campbell, D. S., Baliga, N. S., & Moritz, R. L. (2020) · Scientific Data · doi:10.1038/s41597-020-00724-7 · Cited by 47
“The reproducibility and quantitative performance of SWATH-MS has been assessed in an exemplary multi-laboratory study 19 .”
Self-citationProteomics and the microbiome: pitfalls and potential
Lin, H., He, Q., Shi, L., Sleeman, M. W., Baker, M. S., & Nice, E. C. (2018) · Expert Review of Proteomics · doi:10.1080/14789450.2018.1523724 · Cited by 47
Dipali, S. S., King, C. D., Rose, J., Burdette, J. E., Campisi, J., Schilling, B., & Duncan, F. E. (2023) · Aging · doi:10.18632/aging.205190 · Cited by 53
“Each sample was acquired in data-independent acquisition (DIA) mode [27–29].”
MethodSelf-citationBlank, H. M., Papoulas, O., Maitra, N., Garge, R. K., Kennedy, B. K., Schilling, B., Marcotte, E. M., & Polymenis, M. (2020) · Molecular Biology of the Cell · doi:10.1091/mbc.e19-12-0708 · Cited by 44
“, 2012), using 64 variable-width isolation windows (Collins et al., 2017; Schilling et al., 2017).”
MethodSelf-citationKučera, J., Lochman, J., Bouchal, P., Pakostová, E., Mikulášek, K., Hedrich, S., Janiczek, O., Mandl, M., & Johnson, D. B. (2020) · Frontiers in Microbiology · doi:10.3389/fmicb.2020.610836 · Cited by 45
“MS and MS/MS data were acquired by performing survey MS scan followed by 64 MS/MS scans variable SWATH windows (Collins et al., 2017) between 400 and 1,200 m/z (1 m/z overlap).”
MethodHallmarks of Basidiomycete Soft- and White-Rot in Wood-Decay -Omics Data of Two Armillaria Species
Sahu, N., Merényi, Z., Bálint, B., Kiss, B., Sipos, G., Owens, R. A., & Nagy, L. G. (2021) · Microorganisms · doi:10.3390/microorganisms9010149 · Cited by 51
Hodge, B. A., Meyerhof, G. T., Katewa, S. D., Lian, T., Lau, C., Bar, S., Leung, N. Y., Li, M., Li‐Kroeger, D., Melov, S., et al. (2022) · Nature Communications · doi:10.1038/s41467-022-30975-4 · Cited by 46
Self-citationGeneration of a zebrafish SWATH-MS spectral library to quantify 10,000 proteins
Blattmann, P., Stutz, V., Lizzo, G., Richard, J., Gut, P., & Aebersold, R. H. (2019) · Scientific Data · doi:10.1038/sdata.2019.11 · Cited by 45
Self-citationTian, S., Zhan, D., Yu, Y., Wang, Y., Liu, M., Tan, S., Li, Y., Song, L., Qin, Z., Li, X., et al. (2023) · Genome Biology · doi:10.1186/s13059-023-03048-y · Cited by 43
“…longitudinal quality assessment [25], the generation of reference xenograft proteomes for evaluating variability in differentially expression analysis [26], the assessment of reproducibility using SWATH-MS data acquisition mode [27], and the establishment of a quality control framework [28], etc.”
MethodSelf-citationInter-Laboratory Comparison of Metabolite Measurements for Metabolomics Data Integration
Izumi, Y., Matsuda, F., Hirayama, A., Ikeda, K., Kita, Y., Horie, K., Saigusa, D., Saito, K., Sawada, Y., Nakanishi, H., et al. (2019) · Metabolites · doi:10.3390/metabo9110257 · Cited by 43
Tshilenge, K., Aguirre, C. G., Bons, J., Gerencser, A. A., Basisty, N., Song, S., Rose, J., Lopez‐Ramirez, A., Naphade, S. B., Loureiro, A., et al. (2023) · Molecular & Cellular Proteomics · doi:10.1016/j.mcpro.2023.100534 · Cited by 44
“All samples were analyzed by DIA, specifically using variable window DIA acquisitions (41).”
MethodSelf-citationData-Independent Acquisition for the Orbitrap Q Exactive HF: A Tutorial
Reubsaet, L., Sweredoski, M. J., & Moradian, A. (2018) · Journal of Proteome Research · doi:10.1021/acs.jproteome.8b00845 · Cited by 42
Histone malonylation is regulated by SIRT5 and KAT2A
Zhang, R., Bons, J., Scheidemantle, G. M., Liu, X., Bielska, O., Carrico, C., Rose, J., Heckenbach, I., Scheibye‐Knudsen, M., Schilling, B., et al. (2023) · iScience · doi:10.1016/j.isci.2023.106193 · Cited by 47
“We used MS and dataindependent acquisitions (Collins et al., 2017; Gillet et al., 2012) to profile the malonylation proteome in mouse brain after immunoaffinity enrichment of malonylated peptides.”
MethodSelf-citationHülsmann, J., Kravić, B., Weith, M., Gstaiger, M., Aebersold, R. H., Collins, B. C., & Meyer, H. H. (2018) · Molecular & Cellular Proteomics · doi:10.1074/mcp.ra117.000471 · Cited by 46
“MS data in DDA and SWATH-MS modes were acquired using a Sciex 5600 TripleTOF mass spectrometer interfaced to an Eksigent NanoLC Ultra using acquisition parameters as previously described (30) with the exception that for SWATH analysis 64 variable width precursor isolation windows were used (31).”
MethodSelf-citationAging impairs the osteocytic regulation of collagen integrity and bone quality
Schurman, C. A., Kaya, S., Dole, N. S., Luna, N. M. M., Castillo, N., Potter, R., Rose, J., Bons, J., King, C. D., Burton, J. B., et al. (2024) · Bone Research · doi:10.1038/s41413-023-00303-7 · Cited by 44
Self-citationAlphaDIA enables DIA transfer learning for feature-free proteomics
Wallmann, G., Skowronek, P., Brennsteiner, V., Lebedev, M., Thielert, M., Steigerwald, S., Kotb, M., Despard, O., Heymann, T., Zhou, X., et al. (2025) · Nature Biotechnology · doi:10.1038/s41587-025-02791-w · Cited by 41
“In contrast to DDA, data-independent acquisition (DIA) allows the selection of multiple peptides in parallel, originally in the form of cycles of fixed-width, relatively wide selection windows 9,10 .”
Mass spectrometry-based proteomics techniques and their application in ovarian cancer research
Światły‐Błaszkiewicz, A., Plewa, S., Matysiak, J., & Kokot, Z. J. (2018) · Journal of Ovarian Research · doi:10.1186/s13048-018-0460-6 · Cited by 44
Li, W., Chi, H., Šalovská, B., Wu, C., Sun, L., Rosenberger, G., & Liu, Y. (2019) · Journal of the American Society for Mass Spectrometry · doi:10.1007/s13361-019-02243-1 · Cited by 43
Self-citationMehnert, M., Li, W., Wu, C., Šalovská, B., & Liu, Y. (2019) · PROTEOMICS · doi:10.1002/pmic.201800438 · Cited by 41
Self-citationDeLaney, K., & Li, L. (2019) · Analytical Chemistry · doi:10.1021/acs.analchem.8b05734 · Cited by 40
The Core Proteome of Biofilm-Grown Clinical Pseudomonas aeruginosa Isolates
Erdmann, J., Thöming, J. G., Pohl, S., Pich, A., Lenz, C., & Häußler, S. (2019) · Cells · doi:10.3390/cells8101129 · Cited by 43
“It outperforms standard shotgun approaches using data dependent acquisition (DDA) in terms of detectable peptides and associated proteins as well as measurement reproducibility [28,29].”
BackgroundPark, S., Helsley, R. N., Fadhul, T., Willoughby, J. L. S., Noetzli, L. J., Tu, H.-C., Solheim, M. H., Fujisaka, S., Pan, H., Dreyfuss, J. M., et al. (2023) · Metabolism · doi:10.1016/j.metabol.2023.155591 · Cited by 40
Self-citationStrasser, M. K., Gibbs, D. L., Gascard, P. D., Bons, J., Hickey, J. W., Pan, D., Caruso, J. A., Wang, X., Chen‐Tanyolac, C., Bai, R., et al. (2023) · Developmental Cell · doi:10.1016/j.devcel.2025.06.034 · Cited by 40
Self-citationProteomics for species authentication of cod and corresponding fishery products
Chien, H., Huang, Y., Zheng, Y., Wang, W., Kuo, C.-Y., Wei, G.-J., & Lai, C. (2021) · Food Chemistry · doi:10.1016/j.foodchem.2021.131631 · Cited by 39
Reference Materials for Improving Reliability of Multiomics Profiling
Ren, L., Shi, L. M., & Zheng, Y. (2024) · Phenomics · doi:10.1007/s43657-023-00153-7 · Cited by 40
Nguyen, E. V., Centenera, M. M., Moldovan, M. V., Das, R., Irani, S., Vincent, A. D., Chan, H., Horvath, L. G., Lynn, D. J., Daly, R. J., et al. (2018) · Molecular & Cellular Proteomics · doi:10.1074/mcp.ra118.000633 · Cited by 41
“Quantification was based on the top 3 proteotypic peptide for each protein (24) and exported as an excel file with Spectronaut 8 software (23).”
MethodData-Independent Acquisition for the Quantification and Identification of Metabolites in Plasma
van der Laan, T., Boom, I., Maliepaard, J. C. L., Dubbelman, A.-C., Harms, A. C., & Hankemeier, T. (2020) · Metabolites · doi:10.3390/metabo10120514 · Cited by 38
Deep Proteomics Using Two Dimensional Data Independent Acquisition Mass Spectrometry
Cho, K.-C., Clark, D. J., Schnaubelt, M., Teo, G. C., Leprevost, F. D. V., Bocik, W., Boja, E. S., Hiltke, T., Nesvizhskii, A. I., & Zhang, H. (2020) · Analytical Chemistry · doi:10.1021/acs.analchem.9b04418 · Cited by 46
Self-citationEmerging mass spectrometry-based proteomics methodologies for novel biomedical applications
Pino, L. K., Rose, J., O’Broin, A., Shah, S., & Schilling, B. (2020) · Biochemical Society Transactions · doi:10.1042/bst20191091 · Cited by 40
Self-citationGao, E., Li, W., Wu, C., Shao, W., Di, Y., & Liu, Y. (2021) · Molecular Omics · doi:10.1039/d0mo00188k · Cited by 38
Self-citationGarcía‐Quintanilla, L., Martinez, L. J. R., Bandín‐Vilar, E., Gil‐Martínez, M., González‐Barcia, M., Mondelo‐García, C., Fernández‐Ferreiro, A., & Mateos, J. (2022) · International Journal of Molecular Sciences · doi:10.3390/ijms232314759 · Cited by 36
“However, they have inherent drawbacks related to the stochastic nature of peptide ionization and fragmentation, such as lack of reproducibility/accuracy in the quantification [89].”
BackgroundA proximity proteomics pipeline with improved reproducibility and throughput
Zhong, X., Li, Q., Polacco, B. J., Patil, T., Marley, A., Foussard, H., Khare, P., Vartak, R. S., Xu, J., DiBerto, J. F., et al. (2024) · Molecular Systems Biology · doi:10.1038/s44320-024-00049-2 · Cited by 35
“In contrast, owing to the systematic and unbiased fragmentation of all precursor ions within a prede fi ned mass range, data-independent acquisition (DIA) achieves higher quantitative consistency at comparable proteome coverage (Gillet et al, 2012; Collins et al, 2017; Ludwig et al, 2018).”
Schurman, C. A., Burton, J. B., Rose, J., Ellerby, L., Alliston, T. N., & Schilling, B. (2023) · Journal of Bone Metabolism · doi:10.11005/jbm.2023.30.1.1 · Cited by 35
“Data-independent acquisition (DIA) were established as a label-free mass spectrometry approach to increase the number of identified and quantified proteins in a highly quantitative and accurate fashion, avoiding so-called ‘missing values’.[215-218] DIA approaches to measure and quantify all detectable peptides to prov…”
Self-citationMeyer, J. G., Garcia, T. Y., Schilling, B., Gibson, B. W., & Lamba, D. A. (2019) · Scientific Reports · doi:10.1038/s41598-019-51777-7 · Cited by 36
Self-citationA mass spectrometry workflow for measuring protein turnover rates in vivo
Alevra, M., Mandad, S., Ischebeck, T., Urlaub, H., Rizzoli, S. O., & Fornasiero, E. F. (2019) · Nature Protocols · doi:10.1038/s41596-019-0222-y · Cited by 36
Revealing Dynamic Protein Acetylation across Subcellular Compartments
Baeza, J., Lawton, A. J., Fan, J., Smallegan, M. J., Lienert, I., Gandhi, T., Bernhardt, O. M., Reiter, L., & Denu, J. M. (2020) · Journal of Proteome Research · doi:10.1021/acs.jproteome.0c00088 · Cited by 35
Thomas, S. N., Friedrich, B., Schnaubelt, M., Chan, D. W., Zhang, H., & Aebersold, R. H. (2020) · iScience · doi:10.1016/j.isci.2020.101079 · Cited by 44
Self-citationFuentealba, M., KIPROV, D. D., Schneider, K., Mu, W., Kumaar, P. V. A., Kasler, H. G., Burton, J. B., Watson, M. A., Halaweh, H., King, C. D., et al. (2025) · Aging Cell · doi:10.1111/acel.70103 · Cited by 35
“All human samples were acquired in data-independent acquisition mode (DIA) analysis with two technical replicates for each biological sample replicate (Gillet et al. 2012; Collins et al. 2017; Schilling et al. 2017).”
Self-citationBanaei‐Esfahani, A., Nicod, C., Aebersold, R. H., & Collins, B. C. (2017) · Current Opinion in Microbiology · doi:10.1016/j.mib.2017.09.013 · Cited by 36
“For example, SWATH-MS, an implementation of the DIA strategy, combines the strengths of both aforementioned strategies, and can systematically and reproducibly analyze up of thousands proteins across various conditions and samples (Figure 1d) [31,34].”
Self-citationSystems-Wide Approaches in Induced Pluripotent Stem Cell Models
Lau, E., Paik, D. T., & Wu, J. C. (2018) · Annual Review of Pathology Mechanisms of Disease · doi:10.1146/annurev-pathmechdis-012418-013046 · Cited by 34
“Recent DIA applications have demonstrated consistently quantification of over 4,000 proteins across experiments in a manner that is reproducible across laboratories and operators with low CV (87).”
BackgroundBian, J., Sze, Y. H., Tse, D. Y., To, C. H., McFadden, S. A., Lam, C. S. Y., Li, K. K., & Lam, T. C. (2021) · International Journal of Molecular Sciences · doi:10.3390/ijms22094721 · Cited by 33
“In summary, different from the conventional tools of closed target methods [42,101], the overall results in the current study suggest that the SWATH-based proteomic method used was a large-scale open target method.”
MethodKim, Y. J., Sweet, S. M. M., Egertson, J. D., Sedgewick, A. J., Woo, S., Liao, W., Merrihew, G. E., Searle, B. C., Vaske, C. J., Heaton, R. K., et al. (2018) · Journal of Proteome Research · doi:10.1021/acs.jproteome.8b00699 · Cited by 35
Hausmann, A., Russo, G., Grossmann, J., Zünd, M., Schwank, G., Aebersold, R. H., Liu, Y., Sellin, M. E., & Hardt, W. (2020) · Cellular Microbiology · doi:10.1111/cmi.13191 · Cited by 35
“For SWATH mode, 64-variable window schema was used (Collins et al., 2017; Liu et al., 2019; Ludwig et al., 2018).”
MethodSelf-citationWalzer, M., García‐Seisdedos, D., Prakash, A., Brack, P., Crowther, P., Graham, R. L., George, N., Mohammed, S., Moreno, P. A., Papatheodorou, I., et al. (2022) · Scientific Data · doi:10.1038/s41597-022-01380-9 · Cited by 35
Strategic outline of interventions targeting extracellular matrix for promoting healthy longevity
Park, J. Y. C., King, A., Björk, V., English, B. W., Fedintsev, A., & Ewald, C. Y. (2023) · American Journal of Physiology-Cell Physiology · doi:10.1152/ajpcell.00060.2023 · Cited by 33
Zheng, Y., Wu, M.-C., Chien, H., Wang, W., Kuo, C.-Y., & Lai, C. (2021) · Food Chemistry · doi:10.1016/j.foodchem.2021.129590 · Cited by 32
Oleaga, A., Carnero-Morán, Á., Valero, M. L., & Pérez–Sánchez, R. (2021) · Parasites & Vectors · doi:10.1186/s13071-021-04892-2 · Cited by 32
“Therefore, SWATH-MS outperformed DDA in the reproducibility of proteins identified across all three technical replicate analyses, and these results are in good agreement with those of previous reports [32].”
BackgroundResultSerag, A., Shakkour, Z., Halboup, A. M., Kobeissy, F. H., & Farag, M. A. (2021) · Journal of Proteomics · doi:10.1016/j.jprot.2021.104310 · Cited by 34
Comparing SRM and SWATH Methods for Quantitation of Bovine Muscle Proteomes
Wu, W., Dai, R., & Bendixen, E. (2019) · Journal of Agricultural and Food Chemistry · doi:10.1021/acs.jafc.8b05459 · Cited by 35
Siddiqui, G., Paoli, A. D., MacRaild, C. A., Sexton, A. E., Boulet, C., Shah, A. D., Batty, M. B., Schittenhelm, R. B., Carvalho, T. G., & Creek, D. J. (2022) · GigaScience · doi:10.1093/gigascience/giac008 · Cited by 35
“Although the best comparability is achieved when fragment ion spectra are generated on the same instrument platform [54],fragment ion spectra acquired from instruments using “beam type” collision-induced dissociation such as QTOFs and Q-Orbitrap [55] (used in this study) are sufficiently comparable…”
Optimized data-independent acquisition approach for proteomic analysis at single-cell level
Wang, Y., Lih, T. M., Chen, L., Xu, Y., Kuczler, M. D., Cao, L., Pienta, K. J., Amend, S. R., & Zhang, H. (2022) · Clinical Proteomics · doi:10.1186/s12014-022-09359-9 · Cited by 36
Self-citationŠalovská, B., Gao, E., Müller‐Dott, S., Li, W., Cordon, C. C., Wang, S., Dugourd, A. J. F., Rosenberger, G., Sáez-Rodríguez, J., & Liu, Y. (2023) · Clinical and Translational Medicine · doi:10.1002/ctm2.1179 · Cited by 32
Self-citationHigh-Resolution Mass Spectrometry to Identify and Quantify Acetylation Protein Targets
Schilling, B., Meyer, J. G., Wei, L., Ott, M., & Verdin, E. M. (2019) · Methods in molecular biology · doi:10.1007/978-1-4939-9434-2_1 · Cited by 34
Self-citationRegulation of urea cycle by reversible high-stoichiometry lysine succinylation
Zhang, R., Fang, J., Xie, X., Carrico, C., Meyer, J. G., Wei, L., Bons, J., Rose, J., Riley, R. R., Kwok, R., et al. (2024) · Nature Metabolism · doi:10.1038/s42255-024-01005-y · Cited by 31
“…MS instrument method for DIA, perform a MS1 precursor ion scan from m / z 400 to 1,250 (accumulation time of 250 ms), and MS/MS product ion scans for 64 variable SWATH segments with a MS2 scan range from m / z 100 to 1,250 (accumulation time of 45 ms per each 64-product ion scan per cycle) 38,39 .”
Self-citationCoorssen, J. R., & Padula, M. P. (2024) · Proteomes · doi:10.3390/proteomes12020014 · Cited by 32
β-hydroxybutyrate is a metabolic regulator of proteostasis in the aged and Alzheimer disease brain
Madhavan, S. S., Diaz, S. R., Peralta, S., Nomura, M., King, C. D., Ceyhan, K. E., Lin, A., Bhaumik, D., Foulger, A., Shah, S., et al. (2024) · Cell chemical biology · doi:10.1016/j.chembiol.2024.11.001 · Cited by 33
Self-citationSun, Y., Li, L., Liao, M., Su, M., Wan, C., Zhang, L., & Zhang, H. (2018) · Journal of Pharmaceutical and Biomedical Analysis · doi:10.1016/j.jpba.2018.03.022 · Cited by 32
Self-citationRoot, L., Campo, A., MacNiven, L. M., Con, P., Cnaani, A., & Kültz, D. (2021) · Genomics · doi:10.1016/j.ygeno.2021.07.016 · Cited by 33
Strzepek, R. F., Nunn, B. L., Bach, L. T., Berges, J. A., Young, E. B., & Boyd, P. W. (2022) · Journal of Plankton Research · doi:10.1093/plankt/fbac026 · Cited by 38
“146 Proteomics has advanced methodologically, with more accurate standardized quantitative analyses 147 (Collins et al., 2017; Pino et al., 2020) and protein identifications that allow metabolic profiling 148 (Nunn et al.”
MethodRapid Profiling of Protein Complex Reorganization in Perturbed Systems
Bludau, I., Nicod, C., Martelli, C., Xue, P., Heusel, M., Fossati, A., Uliana, F., Frommelt, F., Aebersold, R. H., & Collins, B. C. (2023) · Journal of Proteome Research · doi:10.1021/acs.jproteome.3c00125 · Cited by 30
Self-citationAn optogenetic-phosphoproteomic study reveals dynamic Akt1 signaling profiles in endothelial cells
Zhou, W., Li, W., Wang, S., Šalovská, B., Hu, Z., Tao, B., Di, Y., Punyamurtula, U., Turk, B. E., Sessa, W. C., et al. (2023) · Nature Communications · doi:10.1038/s41467-023-39514-1 · Cited by 31
Self-citationKIBRA repairs synaptic plasticity and promotes resilience to tauopathy-related memory loss
Kauwe, G., Pareja-Navarro, K. A., Yao, L., Chen, J. H., Wong, I., Saloner, R., Cifuentes, H., Nana, A. L., Shah, S., Li, Y., et al. (2024) · Journal of Clinical Investigation · doi:10.1172/jci169064 · Cited by 33
Self-citationMasuda, T., Hoshiyama, T., Uemura, T., Hirayama‐Kurogi, M., Ogata, S., Furukawa, A., Couraud, P., Furihata, T., Ito, S., & Ohtsuki, S. (2019) · Molecular Pharmaceutics · doi:10.1021/acs.molpharmaceut.9b00114 · Cited by 30
Self-citationThe CUL5 E3 ligase complex negatively regulates central signaling pathways in CD8+ T cells
Liao, X., Li, W., Zhou, H., Rajendran, B. K., Li, A., Ren, J., Luan, Y., Calderwood, D., Turk, B. E., Tang, W., et al. (2024) · Nature Communications · doi:10.1038/s41467-024-44885-0 · Cited by 29
“The cell tubes were ultrasonicated at 4 °C for two cycle (1 min per cycle) usingaVialTweeterdevice(Hielscher-UltrasoundTechnology) 94,95 ,and thencentrifugedat20,000xgfor1 htoremovetheinsolublematerial.”
MethodSelf-citationAcuña-Catalán, D., Shah, S., Wehrfritz, C., Nomura, M., Acevedo, A., Olmos, C., Quiroz, G., Huerta, H., Bons, J., Ampuero, E., et al. (2024) · Cell Reports Medicine · doi:10.1016/j.xcrm.2024.101593 · Cited by 34
“Subsequently, samples were analyzed by label-free data-in-dependent acquisition proteomics, 31–33 which enabled the sensitive and accurate quantification of synaptic proteins by using tandem mass spectrometry (MS2) ion chromatograms.”
Self-citationCrawford, A. J., Forjaz, A., Bons, J., Bhorkar, I., Roy, T., Schell, D., Queiroga, V., Ren, K., Kramer, D., Huang, W., et al. (2024) · Science Advances · doi:10.1126/sciadv.adp6285 · Cited by 30
Self-citationDistler, U., Yoo, H. B., Kardell, O., Hein, D., Sielaff, M., Scherer, M., Józefowicz, A. M., Leps, C., Gomez‐Zepeda, D., von Toerne, C., et al. (2025) · Nature Communications · doi:10.1038/s41467-025-64501-z · Cited by 29
Lucas, N., Robinson, A. B., Espersen, M. L. M., Mahboob, S., Xavier, D., Xue, J. W., Balleine, R. L., DeFazio, A., Hains, P. G., & Robinson, P. J. (2018) · Journal of Proteome Research · doi:10.1021/acs.jproteome.8b00684 · Cited by 29
Dong, M., Lih, T. M., Ao, M., Hu, Y., Chen, S., Eguez, R. V., & Zhang, H. (2021) · Analytical Chemistry · doi:10.1021/acs.analchem.1c01659 · Cited by 29
Self-citationA method to identify and quantify the complete peptide composition in protein hydrolysates
Vreeke, G. J. C., Lubbers, W. D., Vincken, J., & Wierenga, P. A. (2022) · Analytica Chimica Acta · doi:10.1016/j.aca.2022.339616 · Cited by 28
Opportunities for pharmacoproteomics in biomarker discovery
Poulos, R. C., Cai, Z., Robinson, P. J., Reddel, R. R., & Zhong, Q. (2022) · PROTEOMICS · doi:10.1002/pmic.202200031 · Cited by 30
Liu, H., Binoy, A., Ren, S., Martino, T., Miller, A. E., Willis, C. R. G., Veerabhadraiah, S. R., Sukul, A., Bons, J., Rose, J., et al. (2025) · Arthritis & Rheumatology · doi:10.1002/art.43164 · Cited by 29
Self-citationExercise, healthy ageing, and the potential role of small extracellular vesicles
McIlvenna, L. C., & Whitham, M. (2022) · The Journal of Physiology · doi:10.1113/jp282468 · Cited by 28
“The reproducibility of proteomics between laboratories has been demonstrated by the ability to detect minor differences (Collins et al., 2017; Poulos et al., 2020).”
BackgroundShen, S., Wang, X., Zhu, X., Rasam, S. S., Ma, M., Huo, S., Qian, S., Zhang, M., Qu, M., Hu, C., et al. (2022) · Nature Protocols · doi:10.1038/s41596-022-00780-w · Cited by 28
Multi-omics in Crohn's disease: New insights from inside
Mu, C., Zhao, Q., Zhao, Q. C., Yang, L., Pang, X., Liu, T., Li, X., Wang, B., Fung, S., & Cao, H. (2023) · Computational and Structural Biotechnology Journal · doi:10.1016/j.csbj.2023.05.010 · Cited by 29
Identification and characterization of CHO host‐cell proteins in monoclonal antibody bioprocessing
Oh, Y. H., Mendola, K. M., Choe, L. H., Min, L., Lavoie, R. A., Sripada, S. A., Williams, T. I., Lee, K. H., Yigzaw, Y., Seay, A., et al. (2023) · Biotechnology and Bioengineering · doi:10.1002/bit.28568 · Cited by 27
“The reproducibility of DDA measurements was noticeably lower (Figure 3b), with a tendency for reproducibility to increase with HCP abundance, which can be attributed to reduced accuracy of peptide detection at lower concentration (Collins et al., 2017; Kramer et al., 2015).”
Holtz, A. N., Basisty, N., & Schilling, B. (2021) · Methods in molecular biology · doi:10.1007/978-1-0716-1024-4_16 · Cited by 27
Self-citationDabke, K., Kreimer, S. D., Jones, M. R., & Parker, S. J. (2021) · Journal of Proteome Research · doi:10.1021/acs.jproteome.1c00070 · Cited by 29
Immunoproteomics Methods and Techniques
Fulton, K. M., Baltat, I., & Twine, S. M. (2019) · Methods in molecular biology · doi:10.1007/978-1-4939-9597-4_2 · Cited by 27
Zhang, H., Cai, Y., Ding, Y., Zhang, G., Liu, Y., Sun, J., Yang, Y. S. H., Zhan, Z., Iliuk, A. B., Gu, Z., et al. (2022) · Cells · doi:10.3390/cells11132070 · Cited by 26
Self-citationThe status of proteomics as we enter the 2020s: Towards personalised/precision medicine
Nice, E. C. (2020) · Analytical Biochemistry · doi:10.1016/j.ab.2020.113840 · Cited by 29
Frontiers in mass spectrometry–based clinical proteomics for cancer diagnosis and treatment
Haga, Y., Minegishi, Y., & Ueda, K. (2023) · Cancer Science · doi:10.1111/cas.15731 · Cited by 28
Oncoproteomics: Current status and future opportunities
He, Y., Mohamedali, A., Huang, C., Baker, M. S., & Nice, E. C. (2019) · Clinica Chimica Acta · doi:10.1016/j.cca.2019.06.006 · Cited by 27
DIA-MS proteome analysis of formalin-fixed paraffin-embedded glioblastoma tissues
Weke, K., Kote, S., Faktor, J., Shboul, S. A., Uwugiaren, N., Brennan, P. M., Goodlett, D. R., Hupp, T. R., & Đapić, I. (2022) · Analytica Chimica Acta · doi:10.1016/j.aca.2022.339695 · Cited by 25
Salladay-Perez, I. A., Avila, I., Estrada, L., Alexandru, A. C., Ponce, C., Dhingra, A., Torres, G., Deng, C. Y., Hegde, R., Gensheimer, J., et al. (2026) · Nature Aging · doi:10.1038/s43587-026-01101-6 · Cited by 32
Self-citationDiscovery and Qualification of Candidate Urinary Biomarkers of Disease Activity in Lupus Nephritis
Anania, V. G., Yu, K., Pingitore, F., Li, Q., Rose, C. M., Liu, P., Sandoval, W. N., Herman, A. E., Lill, J. R., & Mathews, W. R. (2018) · Journal of Proteome Research · doi:10.1021/acs.jproteome.8b00874 · Cited by 25
Metadata
- Authors
- Collins, B. C., Hunter, C. L., Liu, Y., Schilling, B., Rosenberger, G., Bader, S. L., Chan, D. W., Gibson, B. W., Gingras, A.-C., Held, J. M., Hirayama-Kurogi, M., Hou, G., Krisp, C., Larsen, B., Lin, L., Liu, S., Molloy, M. P., Moritz, R. L., Ohtsuki, S., Schlapbach, R., Selevsek, N., Thomas, S. N., Tzeng, S.-C., Zhang, H., & Aebersold, R.
- Affiliations
- ETH Zurich (CH)
- Buck Institute for Research on Aging (US)
- University of Zurich (CH)
- Institute for Systems Biology (US)
- Johns Hopkins University (US)
- University of Toronto (CA)
- Washington University in St. Louis (US)
- BGI Group (China) (CN)
- Macquarie University (AU)
- Lunenfeld-Tanenbaum Research Institute (CA)
- Kumamoto University (JP)
- Corresponding author
- Aebersold, Ruedi
- ORCID
- Collins, Ben C.
- Hunter, Christie L.
- Liu, Yansheng
- Schilling, Birgit
- Rosenberger, George
- Bader, Samuel L.
- Chan, Daniel W.
- Gibson, Bradford W.
- Gingras, Anne-Claude
- Held, Jason M.
- Hou, Guixue
- Krisp, Christoph
- Liu, Siqi
- Molloy, Mark P.
- Moritz, Robert L.
- Ohtsuki, Sumio
- Schlapbach, Ralph
- Thomas, Stefani N.
- Tzeng, Shin-Cheng
- Zhang, Hui
- Aebersold, Ruedi
- Venue
- Nature Communications
- Published
- 2017-08-21
- Volume
- 8
- Issue
- 1
- Pages
- 291-291
- ISSN
- 2041-1723 (ISSN-L)
- PMC
- PMC5566333
- OpenAlex
- W2950557897
- Semantic Scholar
- 9c55acff92bfc3dfc5b12eddbb845c075896d866
- Cited by
- 576
- References
- 80
- Citations by year
- 2026: 37 · 2025: 60 · 2024: 77 · 2023: 81 · 2022: 66 · 2021: 71 · 2020: 60 · 2019: 66 · 2018: 45 · 2017: 11
- FWCI
- 25.37
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- creativecommons.org/licenses/by/4.0 (tdm)
- creativecommons.org/licenses/by/4.0 (vor)
- Locations
- Nature Communications (publishedVersion, cc-by)
- bioRxiv (Cold Spring Harbor Laboratory) (acceptedVersion, cc-by-nc-nd)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Research Portal (Queen's University Belfast) (publishedVersion, cc-by)
- Digital Commons@Becker (Washington University School of Medicine) (submittedVersion)
- RePEc: Research Papers in Economics (submittedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Repository for Publications and Research Data (ETH Zurich) (submittedVersion, cc-by)
- ETH Zürich Research Collection (cc-by)
- Funders
- National Science Foundation (1330912, 1S10 OD016281)
- Genome Canada (OGI-097)
- Japan Agency for Medical Research and Development (CREST)
- Ontario Genomics (OGI-097)
- Buck Institute for Research on Aging
- Australian Government
- Government of Canada (FDN 143301, 143301)
- European Commission (233226, 602156)
- Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung (143301, 166435, 161435)
- SystemsX.ch
- National Institutes of Health (1s10od016281-01)
- Canadian Institutes of Health Research (14330_1, FDN143301)
- Functional Genomics Center Zurich
- National Human Genome Research Institute (RC2HG005805, HG005805)
- National Cancer Institute (U24CA160036)
- National Institute of General Medical Sciences (R01GM087221, R01GM087221, S10RR027584 and 2P50GM076547, P50GM076547)
- National Center for Research Resources (S10RR027584)
- Core Research for Evolutional Science and Technology
- H2020 European Research Council (670821)
- Topics
- Physical Sciences › Chemistry › Spectroscopy › Advanced Proteomics Techniques and Applications
- Physical Sciences › Chemistry › Spectroscopy › Mass Spectrometry Techniques and Applications
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Molecular Biology › Advanced Biosensing Techniques and Applications
- Keywords
- SWATH mass spectrometry, reproducibility, quantitative proteomics, targeted proteomics, data-independent acquisition, protein quantification, HEK293 cells, sensitivity, dynamic range
- MeSH terms
- Humans, Laboratories, Mass Spectrometry, Reproducibility of Results, Proteome, Proteomics, HEK293 Cells, Laboratory Proficiency Testing
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-05, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-05 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol and qwen3.8-omni-flash).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.