New insights into the origin of remote PPG signals in visible light and infrared
Open accessThe DOI matches its Crossref record; 50 of the 50 references it cites were found in a registry or an index.
Abstract
Remote photoplethysmography (PPG) is an optical measurement technique with established applications in vital signs monitoring. Recently, the consensual understanding of blood volume variations (BVVs) as the origin of PPG signals was challenged, raising validity concerns about the remote SpO 2 methodology. Recognizing the imperative for new opto-physiological evidence, this investigation supports the volumetric hypothesis with living skin experiments and Monte Carlo simulations of remote PPG-amplitude in visible light (VIS) and infrared (IR). Multilayered models of the skin were developed to simulate the separate contributions from skin layers containing pulsatile arterioles to the PPG signal in the 450–1000 nm range. The simulated spectra were qualitatively compared with observations of the resting and compressed finger pad, and complemented with videocapillaroscopy. Our results indicate that remote PPG systems indeed probe arterial blood. Green wavelengths probe dermal arterioles while red-IR wavelengths also reach subcutaneous BVVs. Owing to stable penetration depths, the red-IR diagnostic window promotes the invariance of SpO 2 measurements to skin non-homogeneities.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Optical properties of biological tissues: A review
Jacques, S. L. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/11/r37 · Cited by 4,008 · cited 2×
“Glaring examples are the absorption and scattering parameters, which are mostly determined in ex vivo tissue samples and may differ by an order of magnitude 36,37 .”
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967
Anderson, R. R., & Parrish, J. A. (1981) · Journal of Investigative Dermatology · doi:10.1111/1523-1747.ep12479191 · Cited by 2,456 · cited 1×
“The skin was modeled as multilayered media with optical properties that translate its anatomophysiol-ogy 19 .”
Optical properties of the clearest natural waters (200–800 nm)
Smith, R. C., & Baker, K. S. (1981) · Applied Optics · doi:10.1364/ao.20.000177 · Cited by 1,756 · cited 1×
“The epidermal absorption coefficient, μ a , EPI , was estimated as a combination of background tissue, μ a , base , EPI , and water: ) a EPI w a water w a base EPI , , , , μ a , water was determined from Palmer 52 and Smith 53 .”
Remote plethysmographic imaging using ambient light
Verkruysse, W., Svaasand, L. O., & Nelson, J. S. (2008) · Optics Express · doi:10.1364/oe.16.021434 · Cited by 1,719 · cited 2×
“Processing pipeline: (1, 2) Remote-PPG signals acquired at the finger pad are low-pass-filtered and AC/DC-normalized; (3, 4) The AC/DC streams are ensemble-averaged (EA) (template size, 31 samples per cardiac cycle), including (e) multiple of the pulse-rate frequency and (f,g) the fundamental only.”
Robust pulse rate from chrominance-based rPPG
de Haan, G., & Jeanne, V. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2266196 · Cited by 1,378 · cited 1×
“Remote PPG allows that the pulse-rate is extracted, preferentially, at green wavelengths, though a quest for motion robustness deems necessary that multiple wavelength bands (color-channels) are combined 7 .”
OPTICAL PROPERTIES OF SKIN, SUBCUTANEOUS, AND MUSCLE TISSUES: A REVIEW
BASHKATOV, A. N., GENINA, E. A., & TUCHIN, V. V. (2011) · Journal of Innovative Optical Health Sciences · doi:10.1142/s1793545811001319 · Cited by 761 · cited 1×
“Glaring examples are the absorption and scattering parameters, which are mostly determined in ex vivo tissue samples and may differ by an order of magnitude 36,37 .”
Optical properties of water in the near infrared*
Palmer, K. F., & Williams, D. (1974) · Journal of the Optical Society of America · doi:10.1364/josa.64.001107 · Cited by 709 · cited 1×
“The epidermal absorption coefficient, μ a , EPI , was estimated as a combination of background tissue, μ a , base , EPI , and water: ) a EPI w a water w a base EPI , , , , μ a , water was determined from Palmer 52 and Smith 53 .”
Simpson, C. R., Kohl, M., Essenpreis, M., & Cope, M. (1998) · Physics in Medicine and Biology · doi:10.1088/0031-9155/43/9/003 · Cited by 651
Salomatina, E. V., Jiang, B., Novak, J., & Yaroslavsky, A. N. (2006) · Journal of Biomedical Optics · doi:10.1117/1.2398928 · Cited by 599 · cited 1×
“3 was taken from Salomatina et al .”
The dynamic anatomy and patterning of skin
Wong, R., Geyer, S., Weninger, W., Guimberteau, J., & Wong, J. K. (2016) · Experimental Dermatology · doi:10.1111/exd.12832 · Cited by 484 · cited 1×
“We further remark that modeling the skin as a structure with a discrete number of stacked horizontal layers is a valid simplification 33 , though not without the possible risk that the microanatomic description of the skin into 2 plexuses may not be representative of all skin sites 34,35 .”
A literature review and novel theoretical approach on the optical properties of whole blood
Bosschaart, N., Edelman, G. J., Aalders, M. C. G., van Leeuwen, T. G., & Faber, D. J. (2014) · Lasers in Medical Science · doi:10.1007/s10103-013-1446-7 · Cited by 459 · cited 1×
“57 , the diastolic absorption coefficient of the total tissue was given by Systole is modeled as fractional pulsatile increases, p , of arterial blood in pulsating layers.”
Pulse oximetry: Analysis of theory, technology, and practice
Wukitsch, M. W., Petterson, M. T., Tobler, D. R., & Pologe, J. A. (1988) · Journal of Clinical Monitoring · doi:10.1007/bf01617328 · Cited by 306 · cited 2×
“( ) 0 244 85 3 exp 154 66 2 (3) a base EPI , ,”
New principle for measuring arterial blood oxygenation, enabling motion-robust remote monitoring
van Gastel, M., Stuijk, S., & de Haan, G. (2016) · Scientific Reports · doi:10.1038/srep38609 · Cited by 253 · cited 1×
“μ μ = − + ( ) f r Cb f v SvO SvO (1 ) , (8) v d v d a Hb a HbO ( ) 2 , 2 , 2”
Self-citationMonte Carlo Modeling of Light Transport in Tissues
Jacques, S. L., & Wang, L. (1995) · Optical-Thermal Response of Laser-Irradiated Tissue · doi:10.1007/978-1-4757-6092-7_4 · Cited by 252
A new look at the essence of the imaging photoplethysmography
Kamshilin, A. A., Nippolainen, E., Sidorov, I. S., Vasilev, P. V., Erofeev, N. P., Podolian, N. P., & Romashko, R. V. (2015) · Scientific Reports · doi:10.1038/srep10494 · Cited by 248 · cited 1×
“1038/s41598-018-26068-2 μ′ = λ λ c l ( , ) , (15) s b 0”
Melanocytes and Their Diseases
Yamaguchi, Y., & Hearing, V. J. (2014) · Cold Spring Harbor Perspectives in Medicine · doi:10.1101/cshperspect.a017046 · Cited by 238 · cited 1×
“The palm also features a fivefold lower density of melanocytes than at other skin areas 47 .”
Microvascular imaging: techniques and opportunities for clinical physiological measurements
Allen, J., & Howell, K. (2014) · Physiological Measurement · doi:10.1088/0967-3334/35/7/r91 · Cited by 208 · cited 2×
“Multispectral PPG data, in VIS–IR, may find clinical value in skin health assessments 9,10 .”
Multilayer model of photon diffusion in skin
Schmitt, J. M., Wall, R. T., Zhou, G. X., & Walker, E. C. (1990) · Journal of the Optical Society of America A · doi:10.1364/josaa.7.002141 · Cited by 204
Remote heart rate measurement using low-cost RGB face video: a technical literature review
Rouast, P. V., Adam, M. T. P., Chiong, R., Cornforth, D., & Lux, E. (2018) · Frontiers of Computer Science · doi:10.1007/s11704-016-6243-6 · Cited by 187 · cited 2×
“Processing pipeline: (1, 2) Remote-PPG signals acquired at the finger pad are low-pass-filtered and AC/DC-normalized; (3, 4) The AC/DC streams are ensemble-averaged (EA) (template size, 31 samples per cardiac cycle), including (e) multiple of the pulse-rate frequency and (f,g) the fundamental only.”
Computer simulation of the skin reflectance spectra
Meglinski, I. V., & Matcher, S. J. (2003) · Computer Methods and Programs in Biomedicine · doi:10.1016/s0169-2607(02)00099-8 · Cited by 179 · cited 3×
“The dermis was subdivided into four layers with different blood concentrations, which are as follows 21 : capillary loops (CL; 150–200 μ m thick); upper plexus (UP; 80 μ m thick), reticular dermis (RD; 1400–3000 μ m thick); and deep plexus (DP; 80–700 μ m thick).”
Next-generation acceleration and code optimization for light transport in turbid media using GPUs
Alerstam, E., Yip Lo, W. C., Han, T. D., Rose, J., Andersson-Engels, S., & Lilge, L. (2010) · Biomedical Optics Express · doi:10.1364/boe.1.000658 · Cited by 172 · cited 1×
“We used the publicly available GPU-MCML code package, which enables simulating photon propagation in a multi-layered turbid media with adjustable spatial dimensions and resolution 45 .”
Photoplethysmographic imaging of high spatial resolution
Kamshilin, A. A., Miridonov, S. V., Teplov, V., Saarenheimo, R., & Nippolainen, E. (2011) · Biomedical Optics Express · doi:10.1364/boe.2.000996 · Cited by 153 · cited 1×
“15 reported conflicting observations with the volumetric model 16 .”
Light scattering spectroscopy of human skin in vivo
Zonios, G. I., & Dimou, A. D. (2009) · Optics Express · doi:10.1364/oe.17.001256 · Cited by 124 · cited 1×
“Moreover, the scattering coefficients of living skin can be much lower than those of ex vivo samples 38,39 .”
Reif, R., Amorosino, M. S., Calabro, K. W., A’Amar, O. M., Singh, S. K., & Bigio, I. J. (2008) · Journal of Biomedical Optics · doi:10.1117/1.2870115 · Cited by 118 · cited 1×
“DR measurements were contact-based and may contain probe pressure artifacts 40,41 .”
Verkruysse, W., Lucassen, G. W., de Boer, J. F., Smithies, D. J., Nelson, J. S., & van Gemert, M. J. C. (1997) · Physics in Medicine and Biology · doi:10.1088/0031-9155/42/1/003 · Cited by 112 · cited 1×
“This effect is called self-shielding and is in conflict with the assumption of homogeneous mixture between bloodless skin tissue and blood 56 .”
Calibration of Contactless Pulse Oximetry
Verkruysse, W., Bartula, M., Bresch, E., Rocque, M., Meftah, M., & Kirenko, I. (2017) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000001381 · Cited by 92 · cited 1×
“C 0 244 16 82 exp 400 80 5 (5) a base w”
Microbubbles as a scattering contrast agent for grating-based x-ray dark-field imaging
Velroyen, A., Bech, M., Malecki, A., Tapfer, A., Yaroshenko, A., Ingrisch, M., Cyran, C. C., Auweter, S., Nikolaou, K., Reiser, M. F., et al. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/4/n37 · Cited by 91
Stroke volume response during exercise measured by acetylene uptake and MRI
Groepenhoff, H., Holverda, S., Marcus, J. T., Postmus, P. E., Boonstra, A., & Vonk-Noordegraaf, A. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/1/001 · Cited by 88
Estimation of skin optical parameters for real-time hyperspectral imaging applications
Bjørgan, A., Milanič, M., & Randeberg, L. L. (2014) · Journal of Biomedical Optics · doi:10.1117/1.jbo.19.6.066003 · Cited by 85 · cited 3×
“42 .”
InfluentialBlood pressure fluctuations in human nailfold capillaries
Mahler, F., Muheim, M. H., Intaglietta, M., Bollinger, A., & Anliker, M. (1979) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.1979.236.6.h888 · Cited by 80 · cited 1×
“31 , who performed direct cannulation of human finger nailfold capillaries.”
Ballistocardiographic Artifacts in PPG Imaging
Moço, A., Stuijk, S., & de Haan, G. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2502398 · Cited by 76 · cited 2×
“One of such observations were counter-phase PPG signals in the vicinity of the radial or brachial artery, which are simply motion artifacts 17,18 .”
Self-citationMultilayer Modeling of Reflectance Pulse Oximetry
Reuss, J. L. (2005) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2004.840188 · Cited by 72 · cited 1×
“λ = λ λ ε C RdT uDR n ( ) ( )/ ( , ), (22) d d d”
Potential use of OCT ‐based microangiography in clinical dermatology
Baran, U., Choi, W. J., & Wang, R. K. (2015) · Skin Research and Technology · doi:10.1111/srt.12255 · Cited by 67 · cited 1×
“…acquisition is insightful to investigate a possible contribution of capillaries to PPG. Owing to the low epidermal thickness at the nail fold, the capillary loops and arterioles are found as close to the surface as 0.28–0.43 mm and > 0.43 mm, respectively 26 , and can be reached using green light.”
InfluentialVideo capillaroscopy clarifies mechanism of the photoplethysmographic waveform appearance
Volkov, M. V., Борисович, М. Н., Potemkin, A. V., Volynsky, M. A., Gurov, I. P., Mamontov, O. V., & Kamshilin, A. A. (2017) · Scientific Reports · doi:10.1038/s41598-017-13552-4 · Cited by 67 · cited 1×
“30 showed that the capillary flow speed of RBCs of the fingernail fold have a pronounced pulsatile flow component overlapped with an asynchronous component.”
Optical transmission of blood: effect of erythrocyte aggregation
Shvartsman, L. D., & Fine, I. (2003) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.814532 · Cited by 65
Capillary blood flow imaging within human finger cuticle using optical microangiography
Baran, U., Shi, L., & Wang, R. K. (2013) · Journal of Biophotonics · doi:10.1002/jbio.201300154 · Cited by 59 · cited 2×
“28 applied Doppler optical microangiography (DOMAG) method to map RBC absolute velocity in the arterial and side of the finger cuticle capillaries loops and obtained about 0.67 mm/s for arteriole-end capillaries.”
InfluentialShimada, M., Yamada, Y., Masahide, I. T. O. H., & Yatagai, T. (2001) · Physics in Medicine and Biology · doi:10.1088/0031-9155/46/9/309 · Cited by 58 · cited 2×
“Moreover, the scattering coefficients of living skin can be much lower than those of ex vivo samples 38,39 .”
Retrieving skin properties fromin vivospectral reflectance measurements
Yudovsky, D., & Pilon, L. (2011) · Journal of Biophotonics · doi:10.1002/jbio.201000069 · Cited by 58
Blood flow measurements at different depths using photoplethysmography and laser Doppler techniques
Bergstrand, S., Lindberg, L. G., Ek, A., Lindén, M., & Lindgren, M. (2009) · Skin Research and Technology · doi:10.1111/j.1600-0846.2008.00337.x · Cited by 57 · cited 1×
“λ = λ λ DR C uDR n ( ) ( ) ( , ), (24) OFP s s s r ( ) ( ) ( ) 0”
Non-invasive continuous estimation of blood flow changes in human patellar bone
Näslund, J., Pettersson, J., Lundeberg, T., Linnarsson, D., & Lindberg, L. (2006) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-006-0070-0 · Cited by 57 · cited 1×
“( ) Cb Cb p f v SpO SpO (1 ) (12) s d a Hb a HbO ( ) 2 , 2 , 2”
The pulse in reflectance pulse oximetry: Modeling and experimental studies
Reuss, J. L., & Siker, D. (2004) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-005-2909-6 · Cited by 56 · cited 1×
“Either pulsatile changes are compensated by water displacements 50,58 (WD) or there is layer expansion (LE) to accommodate the additional fluid; i.e., micro-modulations of layers thickness 59 .”
Motion robust PPG-imaging through color channel mapping
Moço, A., Stuijk, S., & de Haan, G. (2016) · Biomedical Optics Express · doi:10.1364/boe.7.001737 · Cited by 55 · cited 1×
“∑ Δ = − F z F z F z w ( ) ( ( ) ( )) , (18) l d l s l T ( ) ( )”
Self-citationAmelard, R., Scharfenberger, C., Kazemzadeh, F., Pfisterer, K. J., Lin, B. S., Clausi, D. A., & Wong, A. K. C. (2015) · Scientific Reports · doi:10.1038/srep14637 · Cited by 53 · cited 1×
“Camera-based systems can operate in transmission-mode 6 , but are better suited for remote reflectance measurements.”
The effect of vascular changes on the photoplethysmographic signal at different hand elevations
Hickey, M., Phillips, J. P., & Kyriacou, P. A. (2015) · Physiological Measurement · doi:10.1088/0967-3334/36/3/425 · Cited by 47 · cited 1×
“This exercise should only hold for illustrative purposes because PPG is highly influenced by individual and contextual factors like posture 32 and skin site (see Section 2 in the Supplementary file for details).”
Finger tissue model and blood perfused skin tissue phantom
Tuchin, V. V., Bashkatov, A. N., Genina, E. A., Kochubey, V. I., Валерьевич, Л. В., Portnov, S. A., Trunina, N. A., Miller, D. R., Cho, S.-M., Oh, H., et al. (2011) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.881604 · Cited by 39 · cited 1×
“For the remaining layers, the optical and anatomical properties of our skin geometry are similar to previous work 21,23,48 , though the average blood concentration in the dermis, C b , is lower.”
Thatcher, J. E., Li, W., Rodriguez‐Vaqueiro, Y., Squiers, J. J., Mo, W., Lu, Y., Plant, K. D., Sellke, E. W., King, D. R., Fan, W., et al. (2015) · Journal of Burn Care & Research · doi:10.1097/bcr.0000000000000317 · Cited by 38 · cited 1×
“Multispectral PPG data, in VIS–IR, may find clinical value in skin health assessments 9,10 .”
The Wavelength Dependence Of The Photoplethysmogram And Its Implication To Pulse Oximetry
Crowe, J. A., & Damianou, D. (1992) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.1992.5761522 · Cited by 37 · cited 1×
“The enabling principle is the dependency between normalized PPG-amplitude ratios at red–IR wavelengths, and the relative proportion between oxygenated and non-oxygenated haemoglobin absorption 3,11 .”
How Tissue Optics Influences Reflectance Pulse Oximetry
Graaff, R., Dassel, A. C. M., Zijlstra, W. G., de Mul, F. F. M., & Aarnoudse, J. G. (1996) · Advances in experimental medicine and biology · doi:10.1007/978-1-4613-0333-6_14 · Cited by 17
Blackford, E. B., Estepp, J. R., & McDuff, D. (2018) · Optical Diagnostics and Sensing: Toward Point-of-Care Diagnostics · doi:10.1117/12.2291073 · Cited by 17 · cited 2×
“44 .”
Influential
Cited by
The 2023 wearable photoplethysmography roadmap
Charlton, P. H., Allen, J., Bailón, R., Baker, S., Behar, J. A., Chen, F., Clifford, G. D., Clifton, D. A., Davies, H. J., Ding, C., et al. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acead2 · Cited by 187
Kaisti, M., Panula, T., Leppänen, J., Punkkinen, R., Jafari Tadi, M., Vasankari, T., Jaakkola, S., Kiviniemi, T., Airaksinen, J., Kostiainen, P., et al. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0117-x · Cited by 164
Photoplethysmography signal processing and synthesis
Mejía-Mejía, E., Allen, J., Budidha, K., El-Hajj, C., Kyriacou, P. A., & Charlton, P. H. (2022) · Photoplethysmography · doi:10.1016/b978-0-12-823374-0.00015-3 · Cited by 138
Chatterjee, S., & Kyriacou, P. A. (2019) · Sensors · doi:10.3390/s19040789 · Cited by 86
“This consideration manifests the physiological hypothesis that the systolic increase in blood volume in the arterioles increases the transmural pressure, resulting in a compression in the elastic fibres in the dermal connective tissue and the overall density of the capillaries present in the dermal sublayers [21].”
BackgroundAjmal, Boonya-Ananta, T., Rodriguez, A. J., Du Le, V. N., & Ramella-Roman, J. C. (2021) · Biomedical Optics Express · doi:10.1364/boe.439893 · Cited by 79
Metadata
- Authors
- Moço, A. V., Stuijk, S., & de Haan, G.
- Venue
- Scientific Reports
- Published
- 2018-05-31
- Volume
- 8
- Issue
- 1
- Pages
- 8501-8501
- ISSN
- 2045-2322 (ISSN-L)
- OpenAlex
- W2805165045
- Cited by
- 168
- References
- 50
- FWCI
- 5.10
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.