Effect of Particle Size on the Performance of an N95 Filtering Facepiece Respirator and a Surgical Mask at Various Breathing Conditions
Open accessAbstract
The effect of aerosol particle size on the performance of an N95 filtering facepiece respirator (FFR) and a surgical mask (SM) was evaluated under different breathing conditions, including breathing frequency and mean inspiratory flow (MIF) rate. The FFR and SM were sealed on a manikin headform and challenged with charge-equilibrated NaCl aerosol. Filter penetration (P filter) was determined as the ratio of aerosol concentrations inside and outside the FFR/SM size-selectively (28 channels) within a range of 20 to 500 nm. In addition, the same models of the FFR and SM were donned, but not sealed, on an advanced manikin headform covered with skin-like material. Total inward leakage (TIL), which represents the total particle penetration, was measured under conditions identical to the filter penetration experiment. Testing was conducted at four mean MIFs (15, 30, 55, and 85 L/min) combined with five breathing frequencies (10, 15, 20, 25, and 30 breaths/min). The results show that SM produced much higher P filter and TIL values, and thus provide little protection against aerosols in the size range tested. P filter was significantly affected by particle size and breathing flow rate (p < 0.05) for the tested FFR and SM. Surprisingly, for both devices, P filter as a function of the particle size exhibited more than one peak under all tested breathing conditions. The effect of breathing frequency on P filter was generally less pronounced, especially for lower MIFs. For the FFR and SM, TIL increased with increasing particle size up to about 50 nm; for particles above 50 nm, the total penetration was not significantly affected by particle size and breathing frequency; however, the effect of MIF remained significant. Copyright 2013 American Association for Aerosol Research
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The lack of significant change in filtration efficiency between H2O2 and heat treatment on the N95 respirator is because of the involvement of mechanical filtration and particle size.
“because of the involvement of mechanical filtration 34 and the particle size. 36”
From Assessment of N95 and K95 respirator decontamination: fiber integrity, filtration efficiency, and dipole charge density (2020) · cites this paper as [36]
The Results and Discussion section reports that particle size significantly affected filter penetration in the tested N95 respirator and discusses possible mechanical and electrostatic contributions. The paper does not compare H2O2 with heat treatment or explain a lack of significant difference between those treatments. (LLM, conf=0.87)
Experimental — accuracy not yet measuredFiltration efficiency depends on the particle size and air flow in the air.
“illustrating that filtration efficiency depends on the particle size and air flow in the air. 36”
From Assessment of N95 and K95 respirator decontamination: fiber integrity, filtration efficiency, and dipole charge density (2020) · cites this paper as [36]
Mechanical filtration and particle size are involved in determining filtration efficiency.
“because of the involvement of mechanical filtration 34 and the particle size. 36”
From Assessment of N95 and K95 respirator decontamination: fiber integrity, filtration efficiency, and dipole charge density (2020) · cites this paper as [36]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Kulkarni, P., Baron, P. A., & Willeke, K. (2011) · Wiley eBooks · doi:10.1002/9781118001684 · Cited by 1,363 · see citing works
Bałazy, A., Toivola, M., Adhikari, A., Sivasubramani, S. K., Reponen, T., & Grinshpun, S. A. (2006) · American Journal of Infection Control · doi:10.1016/j.ajic.2005.08.018 · Cited by 376 · see citing works
Self-citationGrinshpun, S. A., Haruta, H., Eninger, R. M., Reponen, T., McKay, R. T., & Lee, S.-A. (2009) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459620903120086 · Cited by 353 · see citing works
Self-citationSurgical mask filter and fit performance
Oberg, T., & Brosseau, L. M. (2008) · American Journal of Infection Control · doi:10.1016/j.ajic.2007.07.008 · Cited by 310 · see citing works
Bałazy, A., Toivola, M., Reponen, T. A., Podgórski, A., Zimmer, A. T., & Grinshpun, S. A. (2005) · The Annals of Occupational Hygiene · doi:10.1093/annhyg/mei058 · Cited by 237 · see citing works
Self-citationLee, S.-A., Grinshpun, S. A., & Reponen, T. (2008) · The Annals of Occupational Hygiene · doi:10.1093/annhyg/men005 · Cited by 235 · see citing works
Self-citationFilter Performance of N99 and N95 Facepiece Respirators Against Viruses and Ultrafine Particles
Eninger, R. M., Honda, T., Adhikari, A., Heinonen-Tanski, H., Reponen, T., & Grinshpun, S. A. (2008) · The Annals of Occupational Hygiene · doi:10.1093/annhyg/men019 · Cited by 165 · see citing works
Self-citationRengasamy, S., Eimer, B. C., & Shaffer, R. E. (2009) · The Annals of Occupational Hygiene · doi:10.1093/annhyg/men086 · Cited by 157 · see citing works
Electrostatic Respirator Filter Media: Filter Efficiency and Most Penetrating Particle Size Effects
Martin, S. B., & Moyer, E. S. (2000) · Applied Occupational and Environmental Hygiene · doi:10.1080/10473220050075617 · Cited by 156 · see citing works
Rengasamy, S., King, W. P., Eimer, B. C., & Shaffer, R. E. (2008) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459620802275387 · Cited by 125 · see citing works
Penetration of 4.5nm to aerosol particles through fibrous filters
Huang, S.-H., Chen, C.-W., Chang, C.-P., Lai, C.-Y., & Chen, C.-C. (2007) · Journal of Aerosol Science · doi:10.1016/j.jaerosci.2007.05.007 · Cited by 108 · see citing works
Effect of exposure to diesel exhaust particles on the susceptibility of the lung to infection.
Castranova, V., Ma, J. Y., Yang, H. M., et al. (2001) · Environmental Health Perspectives · doi:10.1289/ehp.01109s4609 · Cited by 106 · see citing works
Total Inward Leakage of Nanoparticles Through Filtering Facepiece Respirators
Rengasamy, S., & Eimer, B. C. (2011) · The Annals of Occupational Hygiene · doi:10.1093/annhyg/meq096 · Cited by 89 · see citing works
Penetration of Airborne Microorganisms Through a Surgical Mask and a Dust/Mist Respirator
Willeke, K., Qian, Y., Donnelly, J., Grinshpun, S. A., & Ulevičius, V. (1996) · American Industrial Hygiene Association Journal · doi:10.1080/15428119691014882 · Cited by 84 · see citing works
Self-citationCHARACTERISTICS OF FACE SEAL LEAKAGE IN FILTERING FACEPIECES
Chen, C. C., & Willeke, K. (1992) · American Industrial Hygiene Association Journal · doi:10.1080/15298669291360120 · Cited by 79 · see citing works
Filtration Performance of FDA-Cleared Surgical Masks
Rengasamy*, S., Miller, A., Eimer, B. C., & Shaffer, R. E. (2009) · Journal of the International Society for Respiratory Protection · doi:10.66828/001c.167200 · Cited by 69 · see citing works
FILTER AND LEAK PENETRATION CHARACTERISTICS OF A DUST AND MIST FILTERING FACEPIECE
Chen, C. C., Ruuskanen, J., Pilacinski, W., & Willeke, K. (1990) · American Industrial Hygiene Association Journal · doi:10.1080/15298669091370275 · Cited by 53 · see citing works
Development of an Advanced Respirator Fit-Test Headform
Bergman, M. S., Zhuang, Z., Hanson, D. F., Heimbuch, B. K., McDonald, M. J., Palmiero, A. J., Shaffer, R. E., Harnish, D. A., Husband, M., & Wander, J. D. (2013) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2013.816434 · Cited by 52 · see citing works
Large Particle Penetration through N95 Respirator Filters and Facepiece Leaks with Cyclic Flow
(2009) · The Annals of Occupational Hygiene · doi:10.1093/annhyg/mep062 · Cited by 42 · cited 6× · see citing works
“Other FFR performance studies have reported decreased inward leakage with increasing the breathing flow (Cho et al. 2010b; Rengasamy and Eimer 2011; He et al. 2013c).”
BackgroundMethodResultInfluentialSelf-citationCho, K. J., Jones, S., Jones, G., McKay, R. T., Grinshpun, S. A., Dwivedi, A. K., Shukla, R., Singh, U. P., & Reponen, T. A. (2010) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2010.513910 · Cited by 40 · see citing works
Self-citationHe, X., Grinshpun, S. A., Reponen, T. A., McKay, R. T., Bergman, M. S., & Zhuang, Z. (2013) · The Annals of Occupational Hygiene · doi:10.1093/annhyg/met053 · Cited by 34 · see citing works
Self-citationHe, X., Reponen, T. A., McKay, R. T., & Grinshpun, S. A. (2013) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2013.848037 · Cited by 29 · see citing works
Self-citationManikin-Based Performance Evaluation of Elastomeric Respirators Against Combustion Particles
He, X., Yermakov, M. V., Reponen, T. A., McKay, R. T., James, K., & Grinshpun, S. A. (2013) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2012.760063 · Cited by 29 · see citing works
Self-citationPerformance of Dust Respirators with Facial Seal Leaks: II. Predictive Model
Hinds, W. C., & Bellin, P. (1987) · American Industrial Hygiene Association Journal · doi:10.1080/15298668791385688 · Cited by 28 · see citing works
He, X., Grinshpun, S. A., Reponen, T. A., Yermakov, M. V., McKay, R. T., Haruta, H., & Kimura, K. (2013) · The Annals of Occupational Hygiene · doi:10.1093/annhyg/met014 · Cited by 28 · see citing works
Self-citationEffect of Exposure to Diesel Exhaust Particles on the Susceptibility of the Lung to Infection
Castranova, V., Ma, J. Y. C., Yang, H.-M., et al. (2001) · Environmental Health Perspectives · doi:10.2307/3454678 · Cited by 26 · see citing works
Enhancement of EAP actuated facial expressions by designed chamber geometry in elastomers
Hanson, D. F., Bergs, R. A., Tadesse, Y. T., et al. (2006) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.659000 · Cited by 24 · see citing works
Instruments Based on Electrical Detection of Aerosols
Dhaniyala, S., Fierz, M., Keskinen, J., et al. (2011) · doi:10.1002/9781118001684.ch18 · Cited by 21 · see citing works
Humanlike Articulate Robotic Headform to Replace Human Volunteers in Respirator Fit Testing
Wander, J. D., Hanson, D. F., & Margolin, R. (2012) · doi:10.21236/ada570781 · Cited by 4 · see citing works
Flagged first, then by citations
Research works printed but not linked: 3
- Grafe T, Gogins M, Barris M, Schaefer J and Canepa R (2001). “ Nanofibers in Filtration Applications in Transportation”. Filtration 2001 Conference Proceedings, Chicago, IL; 1– 15.
- Grinshpun SA, Haruta H, Eninger RM, Reponen T, McKay RT and Lee SA (2009b). Performance of an N95 Filtering Facepiece Particulate Respirator and a Surgical Mask During Human Breathing: Two Pathways for Particle Penetration. J. Occup. Environ. Hyg 6: 593– 603. 19598054 PMC7196699
- NIOSH (1995). National Institute for Occupational Safety and Health. Us Dhhs, Public Health Service. “ Respiratory Protective Devices; Final Rules and Notices” Federal Register 60: 110, pp. 30335– 30393.
Cited by
Sorted by citations
Smith, J. D., MacDougall, C. C., Johnstone, J., Copes, R. A., Schwartz, B., & Garber, G. E. (2016) · Canadian Medical Association Journal · doi:10.1503/cmaj.150835 · Cited by 328 · see citing works
Evaluating the Efficacy of Cloth Facemasks in Reducing Particulate Matter Exposure
Shakya, K. M., Noyes, A., Kallin, R., & Peltier, R. E. (2017) · Journal of Exposure Science & Environmental Epidemiology · doi:10.1038/jes.2016.42 · Cited by 202 · see citing works
Yim, W., Cheng, D., Patel, S., Kui, R., Meng, Y. S., & Jokerst, J. V. (2020) · medRxiv · doi:10.1101/2020.07.07.20148551
- Authors
- He, X., Reponen, T., McKay, R. T., & Grinshpun, S. A.
- Affiliations
- University of Cincinnati (US)
- Venue
- Aerosol Science and Technology
- Publisher
- Informa UK Limited
- Published
- 2013-11
- Volume
- 47
- Issue
- 11
- Pages
- 1180-1187
- ISSN
- 0278-6826 (ISSN-L)
- 1521-7388
- PMC
- PMC6756464
- OpenAlex
- W2073215115
- Cited by
- 64
- References
- 29
- Citations by year
- 2025: 1 · 2024: 1 · 2023: 5 · 2022: 9 · 2021: 18 · 2020: 19 · 2019: 2 · 2018: 4 · 2016: 3 · 2015: 2
- FWCI
- 1.36
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Aerosol Science and Technology (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- University of Cincinnati
- National Institute for Occupational Safety and Health (T42-OH008432, OH008432)
- Topics
- Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Infection Control and Ventilation
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Aerosol Filtration and Electrostatic Precipitation
- Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Inhalation and Respiratory Drug Delivery
- Keywords
- N95 respirators, surgical masks, particle size, filter penetration, breathing conditions, respiratory rate, NaCl aerosol, respiratory protection, aerosol filtration
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review