Comparison of calibration methods for atomic-force microscopy cantilevers
The DOI matches its Crossref record; 20 of the 20 references it cites were found in a registry or an index.
Abstract
The scientific community needs a rapid and reliable way of accurately determining the stiffness of atomic-force microscopy cantilevers. We have compared the experimentally determined values of stiffness for ten cantilever probes using four different methods. For rectangular silicon cantilever beams of well defined geometry, the approaches all yield values within 17% of the manufacturer’s nominal stiffness. One of the methods is new, based on the acquisition and analysis of thermal distribution functions of the oscillator’s amplitude fluctuations. We evaluate this method in comparison to the three others and recommend it for its ease of use and broad applicability.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Similarly, the force F was determined by recording the AFM cantilever deflection during sample deformation and knowing the spring constant of the AFM cantilever, calibrated using the thermal noise method [25]. The error in force measurements using this calibration method has been shown to be lower than 5% [26].”
From Extreme strength observed in limpet teeth (2015) · cites this paper as [26]
Citations
Cites
Calibration of atomic-force microscope tips
Hutter, J. L., & Bechhoefer, J. (1993) · Review of Scientific Instruments · doi:10.1063/1.1143970 · Cited by 4,194
Calibration of rectangular atomic force microscope cantilevers
Sader, J. E., Chon, J. W. M., & Mulvaney, P. (1999) · Review of Scientific Instruments · doi:10.1063/1.1150021 · Cited by 2,145
Biological Applications of Optical Forces
Svoboda, K., & Block, S. M. (1994) · Annual Review of Biophysics and Biomolecular Structure · doi:10.1146/annurev.bb.23.060194.001335 · Cited by 1,792
Cleveland, J. P., Manne, S., Bocek, D., & Hansma, P. K. (1993) · Review of Scientific Instruments · doi:10.1063/1.1144209 · Cited by 1,642
Calculation of thermal noise in atomic force microscopy
Butt, H. J., & Jaschke, M. (1995) · Nanotechnology · doi:10.1088/0957-4484/6/1/001 · Cited by 1,589
Sader, J. E. (1998) · Journal of Applied Physics · doi:10.1063/1.368002 · Cited by 1,410
Method for the calibration of atomic force microscope cantilevers
Sader, J. E., Larson, I., Mulvaney, P., & White, L. R. (1995) · Review of Scientific Instruments · doi:10.1063/1.1145439 · Cited by 1,000
Short cantilevers for atomic force microscopy
Walters, D. A., Cleveland, J. P., Thomson, N. H., Hansma, P. K., Wendman, M. A., Gurley, G., & Elings, V. (1996) · Review of Scientific Instruments · doi:10.1063/1.1147177 · Cited by 494
Chon, J. W. M., Mulvaney, P., & Sader, J. E. (2000) · Journal of Applied Physics · doi:10.1063/1.372455 · Cited by 338
Lateral, normal, and longitudinal spring constants of atomic force microscopy cantilevers
Neumeister, J. M., & Ducker, W. A. (1994) · Review of Scientific Instruments · doi:10.1063/1.1144646 · Cited by 334
Research works printed but not linked: 1
- Hazel J L, 1996, Polym. Preparation, 37, 567
Cited by
Items 501–522 of 522
Molecular Adsorption Studies at Heterogeneous Oxide/Electrolyte Interface
Özkaya, B. (2011) · Dokumentenrepositorium der RUB (Ruhr University Bochum)
A computationally designed protein-ligand interaction is mechanically robust
Patten, W. J. V. (2016) · CU Scholar (University of Colorado Boulder)
Forces and Dynamics in Protein Translocation through the Bacterial Transolcon
Fis, A. (2018) · University Library Linz repository (Johannes Kepler Universitat Linz)
Martínez Cuadrado, N. F. (2009) · DIGITAL.CSIC (Spanish National Research Council (CSIC))
Experimental studies of damping in microresonators using laser Doppler vibrometry
Kuter-Arnebeck, O. (2016) · doi:10.82308/1314
Staphylococcus and Biomaterials
Strauss, J. A. (2006) · Digital WPI
Mapping Interfacial Solution Forces that Drive Fluorescent Nanoparticle Incorporation into CrystalsPreprint
Zhang, M., Liu, K.-T., Jang, D. J., Bernat, V., Xia, Y., Li, D., Kao, H.-M., Zhu, Z., Chun, J., Wiesner, U. B., et al. (2026) · ChemRxiv · doi:10.26434/chemrxiv-2026-dnj1k
Coalescence de gouttes dans l'air : du millimètre au nanomètre
Incerti, V. (2017) · theses.fr (ABES) · doi:10.70675/e491c17bz6cedz4683zbfcbz9f0a0a581e0d
Towards a non-contact determination of the mechanical properties of soft condensed matter using AFM
Abegg, E. (2020) · theses.fr (ABES) · doi:10.70675/b7a312fbz0ec6z4474z9f92zb4063fb502dc
Etude de la dynamique des liquides par microscopie à sonde locale
Mortagne, C. (2017) · theses.fr (ABES) · doi:10.70675/8d3e340ez350cz4497z8244z1da93af84124
Dynamique de la ligne de contact à l'échelle nanoscopique
Franiatte, S. (2023) · theses.fr (ABES) · doi:10.70675/a45fdb04z43f3z4216z8188zcdeb916c39d5
Experimental study of information engines at the mesoscopic scale
Archambault, A. (2023) · theses.fr (ABES) · doi:10.70675/27aedf76zf87ez4f7cz98f5z2b8bca372c9b
Kinoshita, Y., & Ito, T. (2026) · Mechanical Engineering Journal · doi:10.1299/mej.25-00359
Matryce dźwigni magnetoelektrycznych w metrologii oddziaływań molekularnych
Majstrzyk, W. (2023) · Prace Naukowe Uniwersytetu Ekonomicznego we Wrocławiu
AFM to Study Pore-Forming Proteins
Aufdermauer, J., Unsay, J. D., & García‐Sáez, A. J. (2026) · Methods in molecular biology · doi:10.1007/978-1-0716-5555-9_6
Accuracy of AFM cantilever thermal calibration in fluid of varying viscosities
Ayangbemi, A. O., & Jacobson, D. R. (2026) · Measurement Science and Technology · doi:10.1088/1361-6501/aebb53
QUANTITATIVE APPLICATION OF LATERAL FORCE MICROSCOPY FOR CARBON NANOTUBES INVESTI GATION
Baturin, A. S., Chouprik, A. A., & Sheshin, E. P. (2007) · Springer eBooks · doi:10.1007/978-1-4020-5514-0_51
Fonctionnalisation de surface pour la micromanipulation : modélisation et expérimentation.
Cot, A. (2014) · theses.fr (ABES) · doi:10.70675/3992a9d6z97bez4a0az8e30z4a2bc6acacf7
Thick-film piezoelectric bimorph actuators for MEMS devices
Owens, S. (2011) · OpenGrey (Institut de l'Information Scientifique et Technique)
Metadata
- Authors
- Burnham, N. A., Chen, X., Hodges, C. S., Matei, G. A., Thoreson, E. J., Roberts, C. J., Davies, M. C., & Tendler, S. J. B.
- Affiliations
- Worcester Polytechnic Institute (US)
- University of Nottingham (GB)
- University of Leeds (GB)
- Venue
- Nanotechnology
- Publisher
- IOP Publishing
- Published
- 2002-12-04
- Volume
- 14
- Issue
- 1
- Pages
- 1-6
- ISSN
- 0957-4484 (ISSN-L)
- OpenAlex
- W2105665577
- Semantic Scholar
- 12f0d9ba4e490824f14c52b5888c8d74f0a80bb4
- Cited by
- 458
- References
- 20
- Citations by year
- 2026: 3 · 2025: 4 · 2024: 8 · 2023: 7 · 2022: 9 · 2021: 9 · 2020: 14 · 2019: 17 · 2018: 20 · 2017: 22 · 2016: 16 · 2015: 20 · 2014: 13 · 2013: 26 · 2012: 34
- FWCI
- 8.47
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Nanotechnology (publishedVersion)
- Funders
- Biotechnology and Biological Sciences Research Council
- Topics
- Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Force Microscopy Techniques and Applications
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Advanced MEMS and NEMS Technologies
- Physical Sciences › Physics and Astronomy › Instrumentation › Scientific Measurement and Uncertainty Evaluation
- Keywords
- atomic force microscopy cantilevers, cantilever stiffness, thermal noise, amplitude fluctuations, calibration method comparison
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.