Rodent model for assessing the long term safety and performance of peripheral nerve recording electrodes

Open access
Vasudevan, Srikanth, Patel, Kunal, Welle, Cristin · 2017 · Journal of Neural Engineering
Cited by 37

The DOI matches its Crossref record; 33 of the 33 references it cites were found in a registry or an index.

Abstract

OBJECTIVE: In the US alone, there are approximately 185 000 cases of limb amputation annually, which can reduce the quality of life for those individuals. Current prosthesis technology could be improved by access to signals from the nervous system for intuitive prosthesis control. After amputation, residual peripheral nerves continue to convey motor signals and electrical stimulation of these nerves can elicit sensory percepts. However, current technology for extracting information directly from peripheral nerves has limited chronic reliability, and novel approaches must be vetted to ensure safe long-term use. The present study aims to optimize methods to establish a test platform using rodent model to assess the long term safety and performance of electrode interfaces implanted in the peripheral nerves. APPROACH: Floating Microelectrode Arrays (FMA, Microprobes for Life Sciences) were implanted into the rodent sciatic nerve. Weekly in vivo recordings and impedance measurements were performed in animals to assess performance and physical integrity of electrodes. Motor (walking track analysis) and sensory (Von Frey) function tests were used to assess change in nerve function due to the implant. Following the terminal recording session, the nerve was explanted and the health of axons, myelin and surrounding tissues were assessed using immunohistochemistry (IHC). The explanted electrodes were visualized under high magnification using scanning electrode microscopy (SEM) to observe any physical damage. MAIN RESULTS: Recordings of axonal action potentials demonstrated notable session-to-session variability. Impedance of the electrodes increased upon implantation and displayed relative stability until electrode failure. Initial deficits in motor function recovered by 2 weeks, while sensory deficits persisted through 6 weeks of assessment. The primary cause of failure was identified as lead wire breakage in all of animals. IHC indicated myelinated and unmyelinated axons near the implanted electrode shanks, along with dense cellular accumulations near the implant site. Scanning electron microscopy (SEM) showed alterations of the electrode insulation and deformation of electrode shanks. SIGNIFICANCE: We describe a comprehensive testing platform with applicability to electrodes that record from the peripheral nerves. This study assesses the long term safety and performance of electrodes in the peripheral nerves using a rodent model. Under this animal test platform, FMA electrodes record single unit action potentials but have limited chronic reliability due to structural weaknesses. Future work will apply these methods to other commercially-available and novel peripheral electrode technologies.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1088/1741-2552/14/1/016008

Cites

33 of 33 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 33 references

Research works printed but not linked: 30

  1. Agnew, 1999, Evolution and resolution of stimulation-induced axonal injury in peripheral nerve, Muscle Nerve, 22, 1393, 10.1002/(SICI)1097-4598(199910)22:10<1393::AID-MUS9>3.0.CO;2-E · DOI · Paper page
  2. Austin, 2012, Chronic constriction of the sciatic nerve and pain hypersensitivity testing in rats, J. Vis. Exp., 10.3791/3393 · DOI
  3. Badia, 2011, Comparative analysis of transverse intrafascicular multichannel, longitudinal intrafascicular and multipolar cuff electrodes for the selective stimulation of nerve fascicles, J. Neural Eng., 8, 10.1088/1741-2560/8/3/036023 · DOI · Paper page
  4. Bain, 1989, Functional evaluation of complete sciatic, peroneal, and posterior tibial nerve lesions in the rat, Plast. Reconstr. Surg., 83, 129, 10.1097/00006534-198901000-00024 · DOI · Paper page
  5. Biddiss, 2007a, Upper-limb prosthetics: critical factors in device abandonment, Am. J. Phys. Med. Rehabil., 86, 977, 10.1097/PHM.0b013e3181587f6c · DOI · Paper page
  6. Biddiss, 2007b, Upper limb prosthesis use and abandonment: a survey of the last 25 years, Prosthet. Orthot. Int., 31, 236, 10.1080/03093640600994581 · DOI · Paper page
  7. Davis, 2016, Restoring motor control and sensory feedback in people with upper extremity amputations using arrays of 96 microelectrodes implanted in the median and ulnar nerves, J. Neural Eng., 13, 10.1088/1741-2560/13/3/036001 · DOI · Paper page
  8. Debnath, 2014, Microelectrode array recordings from the ventral roots in chronically implanted cats, Front. Neurol., 5, 104, 10.3389/fneur.2014.00104 · DOI
  9. Durand, 2014, Time to address the problems at the neural interface, J. Neural Eng., 11, 10.1088/1741-2560/11/2/020201 · DOI
  10. Faroni, 2015, Peripheral nerve regeneration: experimental strategies and future perspectives, Adv. Drug Deliv. Rev., 82–3, 160, 10.1016/j.addr.2014.11.010 · DOI · Paper page
  11. Grill, 1994, Electrical properties of implant encapsulation tissue, Ann. Biomed. Eng., 22, 23, 10.1007/BF02368219 · DOI · Paper page
  12. Grill, 2000, Neural and connective tissue response to long-term implantation of multiple contact nerve cuff electrodes, J. Biomed. Mater. Res., 50, 215, 10.1002/(SICI)1097-4636(200005)50:2<215::AID-JBM17>3.0.CO;2-A · doi:10.1002/(sici)1097-4636(200005)50:2<215::aid-jbm17>3.0.co;2-a
  13. Kejlaa, 1993, Consumer concerns and the functional value of prostheses to upper limb amputees, Prosthet. Orthot. Int., 17, 157, 10.3109/03093649309164376 · DOI · Paper page
  14. Krueger, 2012, Ten years at war: comprehensive analysis of amputation trends, J. Trauma Acute Care Surg., 73, S438, 10.1097/TA.0b013e318275469c · doi:10.1097/ta.0b013e318275469c
  15. Malaga, 2016, Data-driven model comparing the effects of glial scarring and interface interactions on chronic neural recordings in non-human primates, J. Neural Eng., 13, 10.1088/1741-2560/13/1/016010 · DOI · Paper page
  16. McConnell, 2009, Bioimpedance modeling to monitor astrocytic response to chronically implanted electrodes, J. Neural Eng., 6, 10.1088/1741-2560/6/5/055005 · DOI · Paper page
  17. Mercanzini, 2009, In vivo electrical impedance spectroscopy of tissue reaction to microelectrode arrays, IEEE Trans. Biomed. Eng., 56, 1909, 10.1109/TBME.2009.2018457 · DOI · Paper page
  18. Millstein, 1986, Prosthetic use in adult upper limb amputees: a comparison of the body powered and electrically powered prostheses, Prosthet. Orthot. Int., 10, 27, 10.3109/03093648609103076 · DOI · Paper page
  19. Pezzin, 2000, Rehabilitation and the long-term outcomes of persons with trauma-related amputations, Arch. Phys. Med. Rehabil., 81, 292, 10.1016/S0003-9993(00)90074-1 · DOI · Paper page
  20. Pitcher, 1999, Paw withdrawal threshold in the von Frey hair test is influenced by the surface on which the rat stands, J Neurosci. Methods, 87, 185, 10.1016/S0165-0270(99)00004-7 · DOI · Paper page
  21. Schmalbruch, 1986, Fiber composition of the rat sciatic nerve, Anat. Rec., 215, 71, 10.1002/ar.1092150111 · DOI · Paper page
  22. Stinner, 2010, Return to duty rate of amputee soldiers in the current conflicts in Afghanistan and Iraq, J. Trauma, 68, 1476, 10.1097/TA.0b013e3181bb9a6c · doi:10.1097/ta.0b013e3181bb9a6c
  23. Takmakov, 2015, Rapid evaluation of the durability of cortical neural implants using accelerated aging with reactive oxygen species, J. Neural Eng., 12, 10.1088/1741-2560/12/2/026003 · DOI · Paper page
  24. Tan, 2015, Stability and selectivity of a chronic, multi-contact cuff electrode for sensory stimulation in human amputees, J. Neural Eng., 12, 10.1088/1741-2560/12/2/026002 · DOI · Paper page
  25. Vasudevan, 2014, Detergent-free decellularized nerve grafts for long-gap peripheral nerve reconstruction, Plast. Reconstr. Surg. Glob. Open, 2, e201, 10.1097/GOX.0000000000000118 · DOI
  26. Vasudevan, 2013, A rat model for long-gap peripheral nerve reconstruction, Plast. Reconstr. Surg., 132, 871, 10.1097/PRS.0b013e31829fe515 · DOI
  27. Vince, 2005, Anti-TNF-alpha reduces the inflammatory reaction associated with cuff electrode implantation around the sciatic nerve, J. Neuroimmunol., 165, 121, 10.1016/j.jneuroim.2005.04.019 · DOI
  28. Vince, 2004, Biocompatibility of platinum-metallized silicone rubber: in vivo and in vitro evaluation, J. Biomater. Sci. Polym. Ed., 15, 173, 10.1163/156856204322793566 · DOI
  29. Wang, 2013, Characteristics of electrode impedance and stimulation efficacy of a chronic cortical implant using novel annulus electrodes in rat motor cortex, J. Neural Eng., 10, 10.1088/1741-2560/10/4/046010 · DOI · Paper page
  30. Williams, 2007, Complex impedance spectroscopy for monitoring tissue responses to inserted neural implants, J. Neural Eng., 4, 410, 10.1088/1741-2560/4/4/007 · DOI · Paper page

Cited by

See all 37

Metadata

Authors
Vasudevan, S., Patel, K., & Welle, C.
Venue
Journal of Neural Engineering
Published
2016-12-09
Volume
14
Issue
1
Pages
016008
ISSN
  • 1741-2552 (ISSN-L)
  • 1741-2560
OpenAlex
W2560794058
Cited by
37
References
33
FWCI
2.77
Type
journal article
Language
en
Open access
bronze
Licence
Funders
  • US Food and Drug Administration Center for Devices and Radiological Health
  • Defense Advanced Research Projects Agency (DARPA-FDA IAA 224-14-6009)

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Rodent model for assessing the long term safety and performance of peripheral nerve recording electrodes. https://app.cochat.ai/papers/10.1088/1741-2552/14/1/016008
The paper
Vasudevan, S., Patel, K., & Welle, C. (2017). Rodent model for assessing the long term safety and performance of peripheral nerve recording electrodes. Journal of Neural Engineering. https://doi.org/10.1088/1741-2552/14/1/016008

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.