Rodent model for assessing the long term safety and performance of peripheral nerve recording electrodes
Open accessThe DOI matches its Crossref record; 33 of the 33 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: In the US alone, there are approximately 185 000 cases of limb amputation annually, which can reduce the quality of life for those individuals. Current prosthesis technology could be improved by access to signals from the nervous system for intuitive prosthesis control. After amputation, residual peripheral nerves continue to convey motor signals and electrical stimulation of these nerves can elicit sensory percepts. However, current technology for extracting information directly from peripheral nerves has limited chronic reliability, and novel approaches must be vetted to ensure safe long-term use. The present study aims to optimize methods to establish a test platform using rodent model to assess the long term safety and performance of electrode interfaces implanted in the peripheral nerves. APPROACH: Floating Microelectrode Arrays (FMA, Microprobes for Life Sciences) were implanted into the rodent sciatic nerve. Weekly in vivo recordings and impedance measurements were performed in animals to assess performance and physical integrity of electrodes. Motor (walking track analysis) and sensory (Von Frey) function tests were used to assess change in nerve function due to the implant. Following the terminal recording session, the nerve was explanted and the health of axons, myelin and surrounding tissues were assessed using immunohistochemistry (IHC). The explanted electrodes were visualized under high magnification using scanning electrode microscopy (SEM) to observe any physical damage. MAIN RESULTS: Recordings of axonal action potentials demonstrated notable session-to-session variability. Impedance of the electrodes increased upon implantation and displayed relative stability until electrode failure. Initial deficits in motor function recovered by 2 weeks, while sensory deficits persisted through 6 weeks of assessment. The primary cause of failure was identified as lead wire breakage in all of animals. IHC indicated myelinated and unmyelinated axons near the implanted electrode shanks, along with dense cellular accumulations near the implant site. Scanning electron microscopy (SEM) showed alterations of the electrode insulation and deformation of electrode shanks. SIGNIFICANCE: We describe a comprehensive testing platform with applicability to electrodes that record from the peripheral nerves. This study assesses the long term safety and performance of electrodes in the peripheral nerves using a rodent model. Under this animal test platform, FMA electrodes record single unit action potentials but have limited chronic reliability due to structural weaknesses. Future work will apply these methods to other commercially-available and novel peripheral electrode technologies.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Restoring natural sensory feedback in real-time bidirectional hand prostheses
Raspopovic, S., Capogrosso, M., Petrini, F. M., Bonizzato, M., Rigosa, J., Di Pino, G., Carpaneto, J., Controzzi, M., Boretius, T., Fernandez, E., et al. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3006820 · Cited by 1,001
Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P. (2005) · Journal of the Peripheral Nervous System · doi:10.1111/j.1085-9489.2005.10303.x · Cited by 845
A neural interface provides long-term stable natural touch perception
Tan, D. W., Schiefer, M. A., Keith, M. W., Anderson, J. R., Tyler, J., & Tyler, D. J. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3008669 · Cited by 751 · cited 3×
“Similarly, electrical stimulation of the residual nerves can elicit realistic sensory percepts (Davis et al., 2016, Raspopovic et al., 2014, Tan et al., 2014).”
BackgroundMethodInfluentialTransformation of nonfunctional spinal circuits into functional states after the loss of brain input
Courtine, G., Gerasimenko, Y., van den Brand, R., Yew, A., Musienko, P., Zhong, H., Song, B., Ao, Y., Ichiyama, R. M., Lavrov, I., et al. (2009) · Nature Neuroscience · doi:10.1038/nn.2401 · Cited by 706
Chronic recording capability of the Utah Intracortical Electrode Array in cat sensory cortex
Rousche, P. J., & Normann, R. A. (1998) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(98)00031-4 · Cited by 646
Failure mode analysis of silicon-based intracortical microelectrode arrays in non-human primates
Barrese, J. C., Rao, N., Paroo, K., Triebwasser, C., Vargas-Irwin, C., Franquemont, L., & Donoghue, J. P. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/6/066014 · Cited by 600
Simeral, J. D., Kim, S.-P., Black, M. J., Donoghue, J. P., & Hochberg, L. R. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/2/025027 · Cited by 496
Implanted Neural Interfaces: Biochallenges and Engineered Solutions
Grill, W. M., Norman, S. E., & Bellamkonda, R. V. (2009) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-061008-124927 · Cited by 488
A transverse intrafascicular multichannel electrode (TIME) to interface with the peripheral nerve
Boretius, T., Badia, J., Pascual-Font, A., Schuettler, M., Navarro, X., Yoshida, K., & Stieglitz, T. (2010) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2010.05.010 · Cited by 468
A spiral nerve cuff electrode for peripheral nerve stimulation
Naples, G. G., Mortimer, J. T., Scheiner, A., & Sweeney, J. D. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.8670 · Cited by 415 · cited 1×
“Extraneural electrodes, such as the spiral cuff electrodes, allow the epineurium to remain intact, and have a long history of use (Fisher et al., 2009, Naples et al., 1988, Polasek et al., 2007).”
Background
Research works printed but not linked: 30
- Agnew, 1999, Evolution and resolution of stimulation-induced axonal injury in peripheral nerve, Muscle Nerve, 22, 1393, 10.1002/(SICI)1097-4598(199910)22:10<1393::AID-MUS9>3.0.CO;2-E · DOI · Paper page
- Austin, 2012, Chronic constriction of the sciatic nerve and pain hypersensitivity testing in rats, J. Vis. Exp., 10.3791/3393 · DOI
- Badia, 2011, Comparative analysis of transverse intrafascicular multichannel, longitudinal intrafascicular and multipolar cuff electrodes for the selective stimulation of nerve fascicles, J. Neural Eng., 8, 10.1088/1741-2560/8/3/036023 · DOI · Paper page
- Bain, 1989, Functional evaluation of complete sciatic, peroneal, and posterior tibial nerve lesions in the rat, Plast. Reconstr. Surg., 83, 129, 10.1097/00006534-198901000-00024 · DOI · Paper page
- Biddiss, 2007a, Upper-limb prosthetics: critical factors in device abandonment, Am. J. Phys. Med. Rehabil., 86, 977, 10.1097/PHM.0b013e3181587f6c · DOI · Paper page
- Biddiss, 2007b, Upper limb prosthesis use and abandonment: a survey of the last 25 years, Prosthet. Orthot. Int., 31, 236, 10.1080/03093640600994581 · DOI · Paper page
- Davis, 2016, Restoring motor control and sensory feedback in people with upper extremity amputations using arrays of 96 microelectrodes implanted in the median and ulnar nerves, J. Neural Eng., 13, 10.1088/1741-2560/13/3/036001 · DOI · Paper page
- Debnath, 2014, Microelectrode array recordings from the ventral roots in chronically implanted cats, Front. Neurol., 5, 104, 10.3389/fneur.2014.00104 · DOI
- Durand, 2014, Time to address the problems at the neural interface, J. Neural Eng., 11, 10.1088/1741-2560/11/2/020201 · DOI
- Faroni, 2015, Peripheral nerve regeneration: experimental strategies and future perspectives, Adv. Drug Deliv. Rev., 82–3, 160, 10.1016/j.addr.2014.11.010 · DOI · Paper page
- Grill, 1994, Electrical properties of implant encapsulation tissue, Ann. Biomed. Eng., 22, 23, 10.1007/BF02368219 · DOI · Paper page
- Grill, 2000, Neural and connective tissue response to long-term implantation of multiple contact nerve cuff electrodes, J. Biomed. Mater. Res., 50, 215, 10.1002/(SICI)1097-4636(200005)50:2<215::AID-JBM17>3.0.CO;2-A · doi:10.1002/(sici)1097-4636(200005)50:2<215::aid-jbm17>3.0.co;2-a
- Kejlaa, 1993, Consumer concerns and the functional value of prostheses to upper limb amputees, Prosthet. Orthot. Int., 17, 157, 10.3109/03093649309164376 · DOI · Paper page
- Krueger, 2012, Ten years at war: comprehensive analysis of amputation trends, J. Trauma Acute Care Surg., 73, S438, 10.1097/TA.0b013e318275469c · doi:10.1097/ta.0b013e318275469c
- Malaga, 2016, Data-driven model comparing the effects of glial scarring and interface interactions on chronic neural recordings in non-human primates, J. Neural Eng., 13, 10.1088/1741-2560/13/1/016010 · DOI · Paper page
- McConnell, 2009, Bioimpedance modeling to monitor astrocytic response to chronically implanted electrodes, J. Neural Eng., 6, 10.1088/1741-2560/6/5/055005 · DOI · Paper page
- Mercanzini, 2009, In vivo electrical impedance spectroscopy of tissue reaction to microelectrode arrays, IEEE Trans. Biomed. Eng., 56, 1909, 10.1109/TBME.2009.2018457 · DOI · Paper page
- Millstein, 1986, Prosthetic use in adult upper limb amputees: a comparison of the body powered and electrically powered prostheses, Prosthet. Orthot. Int., 10, 27, 10.3109/03093648609103076 · DOI · Paper page
- Pezzin, 2000, Rehabilitation and the long-term outcomes of persons with trauma-related amputations, Arch. Phys. Med. Rehabil., 81, 292, 10.1016/S0003-9993(00)90074-1 · DOI · Paper page
- Pitcher, 1999, Paw withdrawal threshold in the von Frey hair test is influenced by the surface on which the rat stands, J Neurosci. Methods, 87, 185, 10.1016/S0165-0270(99)00004-7 · DOI · Paper page
- Schmalbruch, 1986, Fiber composition of the rat sciatic nerve, Anat. Rec., 215, 71, 10.1002/ar.1092150111 · DOI · Paper page
- Stinner, 2010, Return to duty rate of amputee soldiers in the current conflicts in Afghanistan and Iraq, J. Trauma, 68, 1476, 10.1097/TA.0b013e3181bb9a6c · doi:10.1097/ta.0b013e3181bb9a6c
- Takmakov, 2015, Rapid evaluation of the durability of cortical neural implants using accelerated aging with reactive oxygen species, J. Neural Eng., 12, 10.1088/1741-2560/12/2/026003 · DOI · Paper page
- Tan, 2015, Stability and selectivity of a chronic, multi-contact cuff electrode for sensory stimulation in human amputees, J. Neural Eng., 12, 10.1088/1741-2560/12/2/026002 · DOI · Paper page
- Vasudevan, 2014, Detergent-free decellularized nerve grafts for long-gap peripheral nerve reconstruction, Plast. Reconstr. Surg. Glob. Open, 2, e201, 10.1097/GOX.0000000000000118 · DOI
- Vasudevan, 2013, A rat model for long-gap peripheral nerve reconstruction, Plast. Reconstr. Surg., 132, 871, 10.1097/PRS.0b013e31829fe515 · DOI
- Vince, 2005, Anti-TNF-alpha reduces the inflammatory reaction associated with cuff electrode implantation around the sciatic nerve, J. Neuroimmunol., 165, 121, 10.1016/j.jneuroim.2005.04.019 · DOI
- Vince, 2004, Biocompatibility of platinum-metallized silicone rubber: in vivo and in vitro evaluation, J. Biomater. Sci. Polym. Ed., 15, 173, 10.1163/156856204322793566 · DOI
- Wang, 2013, Characteristics of electrode impedance and stimulation efficacy of a chronic cortical implant using novel annulus electrodes in rat motor cortex, J. Neural Eng., 10, 10.1088/1741-2560/10/4/046010 · DOI · Paper page
- Williams, 2007, Complex impedance spectroscopy for monitoring tissue responses to inserted neural implants, J. Neural Eng., 4, 410, 10.1088/1741-2560/4/4/007 · DOI · Paper page
Cited by
Toward higher-performance bionic limbs for wider clinical use
Farina, D., Vujaklija, I., Brånemark, R., Bull, A. M. J., Dietl, H., Graimann, B., Hargrove, L. J., Hoffmann, K.-P., Huang, H., Ingvarsson, T., et al. (2021) · Nature Biomedical Engineering · doi:10.1038/s41551-021-00732-x · Cited by 263
A review for the peripheral nerve interface designer
Larson, C. E., & Meng, E. (2020) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108523 · Cited by 133
Frost, C. M., Ursu, D. C., Flattery, S. M., Nedic, A., Hassett, C. A., Moon, J. D., Buchanan, P. J., Brent Gillespie, R., Kung, T. A., Kemp, S. W. P., et al. (2018) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-018-0452-1 · Cited by 68
Frederick, R. A., Troyk, P. R., & Cogan, S. F. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac2bb8 · Cited by 7
“Rows represent each of the 16 electrodes in the implanted array and columns represent each time point of recorded data (week 1, 2, 3, 4, 5, 6, 7, 8, 10, 12, 14, 16, 18, 20, 22, 24, 26, 28, 30, 32, 34, 36, 38).”
Influential
Metadata
- Authors
- Vasudevan, S., Patel, K., & Welle, C.
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2016-12-09
- Volume
- 14
- Issue
- 1
- Pages
- 016008
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- OpenAlex
- W2560794058
- Cited by
- 37
- References
- 33
- FWCI
- 2.77
- Type
- journal article
- Language
- en
- Open access
- bronze
- Licence
- publishingsupport.iopscience.iop.org/iop-standard/v1 (vor)
- iopscience.iop.org/info/page/text-and-data-mining (tdm)
- Funders
- US Food and Drug Administration Center for Devices and Radiological Health
- Defense Advanced Research Projects Agency (DARPA-FDA IAA 224-14-6009)
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Rodent model for assessing the long term safety and performance of peripheral nerve recording electrodes. https://app.cochat.ai/papers/10.1088/1741-2552/14/1/016008
- The paper
- Vasudevan, S., Patel, K., & Welle, C. (2017). Rodent model for assessing the long term safety and performance of peripheral nerve recording electrodes. Journal of Neural Engineering. https://doi.org/10.1088/1741-2552/14/1/016008
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.