Comparison between red, green and blue light reflection photoplethysmography for heart rate monitoring during motion
Open accessAbstract
Reflection photoplethysmography (PPG) using 530 nm (green) wavelength light has the potential to be a superior method for monitoring heart rate (HR) during normal daily life due to its relative freedom from artifacts. However, little is known about the accuracy of pulse rate (PR) measured by 530 nm light PPG during motion. Therefore, we compared the HR measured by electrocadiography (ECG) as a reference with PR measured by 530, 645 (red), and 470 nm (blue) wavelength light PPG during baseline and while performing hand waving in 12 participants. In addition, we examined the change of signal-to-noise ratio (SNR) by motion for each of the three wavelengths used for the PPG. The results showed that the limit of agreement in Bland-Altman plots between the HR measured by ECG and PR measured by 530 nm light PPG (±0.61 bpm) was smaller than that achieved when using 645 and 470 nm light PPG (±3.20 bpm and ±2.23 bpm, respectively). The ΔSNR (the difference between baseline and task values) of 530 and 470nm light PPG was significantly smaller than ΔSNR for red light PPG. In conclusion, 530 nm light PPG could be a more suitable method than 645 and 470nm light PPG for monitoring HR in normal daily life.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Ambient artifacts are for example light artifacts interfering with the light of the PPG sensor43, weather conditions, such as high humidity which increases the moisture level of the skin surface and thus its conductivity44, as well as coverage of the device for example by a blanket which increases the temperature and thereby potentially EDA levels45.”
From Data quality evaluation in wearable monitoring (2022) · cites this paper as [43]
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967
Anderson, R. R., & Parrish, J. A. (1981) · Journal of Investigative Dermatology · doi:10.1111/1523-1747.ep12479191 · Cited by 2,456
Maeda, Y., Sekine, M., & Tamura, T. (2011) · Journal of Medical Systems · doi:10.1007/s10916-010-9505-0 · Cited by 203
Reddy, K. A., George, B., & Kumar, V. J. (2008) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2008.2009136 · Cited by 165
In vivo reflectance of blood and tissue as a function of light wavelength
Cui, W., Ostrander, L. E., & Lee, B. Y. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.55667 · Cited by 149
Spīgulis, J., Gailite, L., Lihachev, A., & Erts, R. (2007) · Applied Optics · doi:10.1364/ao.46.001754 · Cited by 98
Han, H., & Kim, J. (2011) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2011.12.005 · Cited by 92
Matsumura, K., & Yamakoshi, T. (2013) · Behavior Research Methods · doi:10.3758/s13428-012-0312-z · Cited by 87
Self-citationPulsed multifrequency photoplethysmograph
Giltvedt, J., Sira, A., & Helme, P. (1984) · Medical & Biological Engineering & Computing · doi:10.1007/bf02442745 · Cited by 47
Research works printed but not linked: 1
- challoner, 1979, Photoelectric plethysmography for estimating cutaneous blood flow, Noninvasive Physiological Measurements, 1, 125
Cited by
Sorted by citations
Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2441951 · Cited by 935
“For reflectance-mode PPG, the green wavelength produces greater signal amplitude and lower motion artifact than IR and red [69], [75]–[77].”
BackgroundPhotoplethysmography Revisited: From Contact to Noncontact, From Point to Imaging
Sun, Y., & Thakor, N. V. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2476337 · Cited by 589
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348
An Internet-of-Things (IoT) Network System for Connected Safety and Health Monitoring Applications
Wu, F., Wu, T., & Yuce, M. R. (2018) · Sensors · doi:10.3390/s19010021 · Cited by 302
Jinno, H., Yokota, T., Koizumi, M., Yukita, W., Saito, M., Osaka, I., Fukuda, K., & Someya, T. (2021) · Nature Communications · doi:10.1038/s41467-021-22558-6 · Cited by 253
Can Wearable Devices Accurately Measure Heart Rate Variability? A Systematic Review
Georgiou, K., Larentzakis, A. V., Khamis, N. N., Alsuhaibani, G. I., Alaska, Y. A., & Giallafos, E. J. (2018) · Folia Medica · doi:10.2478/folmed-2018-0012 · Cited by 244
A Rigid-Flex Wearable Health Monitoring Sensor Patch for IoT-Connected Healthcare Applications
Wu, T., Wu, F., Qiu, C., Redouté, J., & Yuce, M. R. (2020) · IEEE Internet of Things Journal · doi:10.1109/jiot.2020.2977164 · Cited by 230
Kim, K. B., & Baek, H. J. (2023) · Electronics · doi:10.3390/electronics12132923 · Cited by 260
Wearable-band type visible-near infrared optical biosensor for non-invasive blood glucose monitoring
Rachim, V. P., & Chung, W.-Y. (2019) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2019.01.121 · Cited by 200
Noncontact Monitoring of Blood Oxygen Saturation Using Camera and Dual-Wavelength Imaging System
Shao, D., Liu, C., Tsow, F., Yang, Y., Du, Z., Iriya, R., Yu, H., & Tao, N. J. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2481896 · Cited by 151
Liu, J., Yan, B. P., Zhang, Y., Ding, X., Su, P., & Zhao, N. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2874957 · Cited by 139
Wang, G., Atef, M., & Lian, Y. (2018) · IEEE Circuits and Systems Magazine · doi:10.1109/mcas.2018.2849261 · Cited by 121
Franklin, D., Tzavelis, A., Lee, J. Y., Chung, H. U., Trueb, J. T., Arafa, H., Kwak, S. S., Huang, I. A., Liu, Y., Rathod, M., et al. (2023) · Nature Biomedical Engineering · doi:10.1038/s41551-023-01098-y · Cited by 119
Self-citationScardulla, F., Cosoli, G., Spinsante, S., Poli, A., Iadarola, G., Pernice, R., Busacca, A., Pasta, S., Scalise, L., & D’Acquisto, L. (2023) · Measurement · doi:10.1016/j.measurement.2023.113150 · Cited by 134
A Multi-Sensor and Parallel Processing SoC for Miniaturized Medical Instrumentation
Schönle, P., Glaser, F., Burger, T., Rovere, G., Luca Benini, C., & Huang, Q. (2018) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2018.2815653 · Cited by 98
A Real-time PPG Quality Assessment Approach for Healthcare Internet-of-Things
Naeini, E. K., Azimi, I., Rahmani, A. M., Liljeberg, P., & Dutt, N. D. (2019) · Procedia Computer Science · doi:10.1016/j.procs.2019.04.074 · Cited by 92
Data quality evaluation in wearable monitoring
Böttcher, S., Vieluf, S., Bruno, E., Joseph, B., Epitashvili, N., Biondi, A., Zabler, N., Glasstetter, M., Dümpelmann, M., Van Laerhoven, K., et al. (2022) · Scientific Reports · doi:10.1038/s41598-022-25949-x · Cited by 91
iPhys: An Open Non-Contact Imaging-Based Physiological Measurement Toolbox
McDuff, D., & Blackford, E. B. (2019) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2019.8857012 · Cited by 84
Stone, J. D., Rentz, L. E., Forsey, J. D., Ramadan, J. H., Markwald, R. R., Finomore, V. S., Galster, S. M., Rezai, A. R., & Hagen, J. A. (2020) · Nature and Science of Sleep · doi:10.2147/nss.s270705 · Cited by 85
A Review of Skin-Wearable Sensors for Non-Invasive Health Monitoring Applications
Mao, P., Li, H., & Yu, Z. (2023) · Sensors · doi:10.3390/s23073673 · Cited by 87
Driver Drowsiness Detection Using Multi-Channel Second Order Blind Identifications
Zhang, C., Wu, X., Zheng, X., & Yu, S. (2019) · IEEE Access · doi:10.1109/access.2019.2891971 · Cited by 81
An Overview of the Sensors for Heart Rate Monitoring Used in Extramural Applications
Galli, A., Montree, R. J. H., Que, S., Peri, E., & Vullings, R. (2022) · Sensors · doi:10.3390/s22114035 · Cited by 87
Zhou, Z.-B., Cui, T., Li, D., Jian, J., Li, Z., Ji, S., Li, X., Xu, J., Liu, H., Yang, Y., et al. (2023) · Materials · doi:10.3390/ma16062133 · Cited by 85
Matsumura, K., Rolfe, P. J., Lee, J., & Yamakoshi, T. (2014) · PLoS ONE · doi:10.1371/journal.pone.0091205 · Cited by 72
Self-citationSmart-watch-programmed green-light-operated percutaneous control of therapeutic transgenes
Mansouri, M., Hussherr, M., Strittmatter, T., Buchmann, P., Xue, S., Camenisch, G., & Fussenegger, M. (2021) · Nature Communications · doi:10.1038/s41467-021-23572-4 · Cited by 71
Blood Glucose Level Regression for Smartphone PPG Signals Using Machine Learning
Islam, T. T., Ahmed, M. S., Hassanuzzaman, M., Amir, S. A. B., & Rahman, T. (2021) · Applied Sciences · doi:10.3390/app11020618 · Cited by 72
Lee, H., Chung, H., Ko, H., & Lee, J. (2018) · IEEE Sensors Journal · doi:10.1109/jsen.2018.2801385 · Cited by 61
Self-citationValidation of a Photoplethysmographic Heart Rate Monitor: Polar OH1
Hermand, E., Cassirame, J., Ennequin, G., & Hue, O. (2019) · International Journal of Sports Medicine · doi:10.1055/a-0875-4033 · Cited by 59
Alzahrani, A., Hu, S., Azorin-Peris, V., Barrett, L. A., Esliger, D., Hayes, M., Akbare, S., Achart, J., & Kuoch, S. (2015) · Sensors · doi:10.3390/s151025681 · Cited by 56
Flexible and Printed PPG Sensors for Estimation of Drowsiness
Ryu, G., You, J., Kostianovskii, V., Lee, E.-B., Kim, Y., Park, C., & Noh, Y. (2018) · IEEE Transactions on Electron Devices · doi:10.1109/ted.2018.2833477 · Cited by 56
Image Enhancement for Remote Photoplethysmography in a Low-Light Environment
Lin, X., Chen, W., Zhao, C., Wu, X., & Wang, J. (2020) · International Conference on Automatic Face and Gesture Recognition · doi:10.1109/fg47880.2020.00076 · Cited by 55
Harju, J., Tarniceriu, A., Parák, J., Vehkaoja, A. T., Yli‐Hankala, A., & Korhonen, I. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aac9a9 · Cited by 55
Fall prediction using behavioural modelling from sensor data in smart homes
Forbes, G., Massie, S., & Craw, S. (2019) · Artificial Intelligence Review · doi:10.1007/s10462-019-09687-7 · Cited by 55
Blood pressure measurement using only a smartphone
Frey, L., Menon, C., & Elgendi, M. (2022) · npj Digital Medicine · doi:10.1038/s41746-022-00629-2 · Cited by 55
Prabha, A., Yadav, J., Rani, A., & Singh, V. (2022) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2022.103876 · Cited by 58
Li, J., Chu, H., Chen, Z., Yiu, C. K., Qu, Q., Li, Z.-Y., & Yu, X. (2024) · ACS Nano · doi:10.1021/acsnano.4c04291 · Cited by 50
Photoplethysmogram at green light: Where does chaos arise from?
Sviridova, N., Zhao, T., Aihara, K., Nakamura, K., & Nakano, A. (2018) · Chaos Solitons & Fractals · doi:10.1016/j.chaos.2018.09.016 · Cited by 52
Morelli, D., Bartoloni, L., Colombo, M., Plans, D., & Clifton, D. A. (2017) · Healthcare Technology Letters · doi:10.1049/htl.2017.0039 · Cited by 51
Wei, B., He, X., Zhang, C., & Wu, X. (2017) · BioMedical Engineering OnLine · doi:10.1186/s12938-016-0300-0 · Cited by 50
Convertino, V. A., Schauer, S. G., Weitzel, E. K., Cardin, S., Stackle, M. E., Talley, M. J., Sawka, M. N., & Inan, O. T. (2020) · Sensors · doi:10.3390/s20226413 · Cited by 49
A Review of Methods for Non-Invasive Heart Rate Measurement on Wrist
De Pinho Ferreira, N., Gehin, C., & Massot, B. (2021) · IRBM · doi:10.1016/j.irbm.2020.04.001 · Cited by 49
Tautan, A.-M., Young, A., Wentink, E., & Wieringa, F. (2015) · 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7319795 · Cited by 47
Rodrigues, L. M., Rocha, C., Ferreira, H. A., & Silva, H. (2019) · Scientific Reports · doi:10.1038/s41598-019-53213-2 · Cited by 45
Rentz, L. E., Ulman, H. K., & Galster, S. M. (2021) · Sensors · doi:10.3390/s21155071 · Cited by 52
Automated loss of pulse detection on a consumer smartwatch
Shah, K., Wang, A., Chen, Y., Munjal, J., Chhabra, S., Stange, A., Wei, E., Phan, T., Giest, T., Hawkins, B., et al. (2025) · Nature · doi:10.1038/s41586-025-08810-9 · Cited by 44
Gnacek, M., Broulidakis, M. J., Mavridou, I. I., Fatoorechi, M., Seiss, E., Κωστούλας, Θ., Balaguer‐Ballester, E., Kiprijanovska, I., Rosten, C., & Nduka, C. C. (2022) · Frontiers in Virtual Reality · doi:10.3389/frvir.2022.781218 · Cited by 44
Non-contact remote estimation of cardiovascular parameters
Fan, Q., & Li, K. (2017) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2017.09.022 · Cited by 44
Using photoplethysmography in heart rate monitoring of patients with epilepsy
van Andel, J., Ungureanu, C., Aarts, R., Leijten, F., & Arends, J. (2015) · Epilepsy & Behavior · doi:10.1016/j.yebeh.2015.02.018 · Cited by 39
Estimation of HRV and SpO2 from wrist-worn commercial sensors for clinical settings
Jarchi, D., Salvi, D., Velardo, C., Mahdi, A., Tarassenko, L., & Clifton, D. A. (2018) · 2018 IEEE 15th International Conference on Wearable and Implantable Body Sensor Networks (BSN) · doi:10.1109/bsn.2018.8329679 · Cited by 38
The Importance of Data Quality Control in Using Fitbit Device Data From the Research Program
Lederer, L. E., Breton, A., Jeong, H., Master, H., Roghanizad, A. R., & Dunn, J. (2023) · JMIR mhealth and uhealth · doi:10.2196/45103 · Cited by 42
Wang, Y., Liu, Z., & Ma, S. (2017) · Physiological Measurement · doi:10.1088/1361-6579/aa996d · Cited by 36
Zenko, J., Kos, M., & Kramberger, I. (2016) · International Conference on Systems, Signals and Image Processing · doi:10.1109/iwssip.2016.7502766 · Cited by 36
Video pulse rate variability analysis in stationary and motion conditions
Rodríguez, A., & Ramos-Castro, J. J. (2018) · BioMedical Engineering OnLine · doi:10.1186/s12938-018-0437-0 · Cited by 35
A Survey of Photoplethysmography and Imaging Photoplethysmography Quality Assessment Methods
Desquins, T., Bousefsaf, F., Pruski, A., & Maaoui, C. (2022) · Applied Sciences · doi:10.3390/app12199582 · Cited by 35
Accuracy and Reliability of Pulse O2 Saturation Measured by a Wrist-worn Oximeter
Hermand, É., Coll, C., Richalet, J., & Lhuissier, F. (2021) · International Journal of Sports Medicine · doi:10.1055/a-1337-2790 · Cited by 33
MW-PPG Sensor: An on-Chip Spectrometer Approach
Chang, C.-C., Wu, C.-T., Choi, B. I., & Fang, T.-J. (2019) · Sensors · doi:10.3390/s19173698 · Cited by 30
Simultaneous detection of blink and heart rate using multi-channel ICA from smart phone videos
Zhang, C., Wu, X., Zhang, L., He, X., & Lv, Z. (2016) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2016.11.022 · Cited by 30
Analysis and improvement of non-contact SpO2 extraction using an RGB webcam
Wei, B., Wu, X., Zhang, C., & Lv, Z. (2021) · Biomedical Optics Express · doi:10.1364/boe.423508 · Cited by 30
Guede-Fernández, F., Ferrer-Mileo, V., Ramos-Castro, J. J., Fernández-Chimeno, M., & Garcia-Gonzalez, M. A. (2015) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7320085 · Cited by 30
Using imaging photoplethysmography for heart rate estimation in non-human primates
Михайлович, У. А., Moeller, S., Kagan, I., Gail, A., Treue, S., & Wolf, F. (2018) · PLoS ONE · doi:10.1371/journal.pone.0202581 · Cited by 29
Wang, Y., Liu, Z., Yang, J., & Ma, S. (2017) · BioMedical Engineering OnLine · doi:10.1186/s12938-017-0389-9 · Cited by 29
Wearables in Cardiovascular Disease
Kumar, S., Victoria‐Castro, A. M., Melchinger, H. C., O’Connor, K. D., Psotka, M. A., Desai, N. R., Ahmad, T., & Wilson, F. P. (2022) · Journal of Cardiovascular Translational Research · doi:10.1007/s12265-022-10314-0 · Cited by 27
Yamakoshi, Y., Matsumura, K., Yamakoshi, T., Lee, J., Rolfe, P. J., Kato, Y., Shimizu, K., & Yamakoshi, K.-I. (2017) · Journal of Biomedical Optics · doi:10.1117/1.jbo.22.6.067001 · Cited by 27
Self-citationMatsumura, K., Toda, S., & Kato, Y. (2020) · IEEE Access · doi:10.1109/access.2020.2990438 · Cited by 27
Self-citationImaging PPG for In Vivo Human Tissue Perfusion Assessment during Surgery
Lai, M., van der Stel, S. D., Groen, H. C., van Gastel, M., Kuhlmann, K. F. D., Ruers, T. J. M., & Hendriks, B. H. W. (2022) · Journal of Imaging · doi:10.3390/jimaging8040094 · Cited by 25
Uncertainties in the Analysis of Heart Rate Variability: A Systematic Review
Lü, L., Zhu, T., Morelli, D., Creagh, A. P., Liu, Z., Yang, J., Liu, F., Zhang, Y., & Clifton, D. A. (2023) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2023.3271595 · Cited by 28
Wearable Motion Tolerant PPG Sensor for Instant Heart Rate in Daily Activity
Ishikawa, T., Hyodo, Y., Miyashita, K., Yoshifuji, K., Komoriya, Y., & Imai, Y. (2017) · International Joint Conference on Biomedical Engineering Systems and Technologies (BIOSTEC) · doi:10.5220/0006109901260133 · Cited by 24
Alharbi, S., Hu, S., Mulvaney, D. J., Barrett, L. A., Yan, L., Blanos, P., Elsahar, Y., & Adema, S. (2018) · Sensors · doi:10.3390/s19010118 · Cited by 24
Park, P., Lee, W., & Cho, S. (2023) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2023.3315297 · Cited by 24
Shooshtari, L., Ghods, S., Mohammadpour, R., Esfandiar, A., & zad, A. I. (2022) · Scientific Reports · doi:10.1038/s41598-022-11327-0 · Cited by 22
Non-Contact Reflectance Photoplethysmography: Progress, Limitations, and Myths
Špetlík, R., Čech, J., & Matas, J. (2018) · International Conference on Automatic Face and Gesture Recognition · doi:10.1109/fg.2018.00111 · Cited by 22
OpenEarable 2.0: Open-Source Earphone Platform for Physiological Ear Sensing
Röddiger, T., Küttner, M., Lepold, P., King, T., Moschina, D., Bagge, O., Paradiso, J. A., Clarke, C., & Beigl, M. (2025) · Proceedings of the ACM on Interactive Mobile Wearable and Ubiquitous Technologies · doi:10.1145/3712069 · Cited by 34
Reynolds, J., Ahmmed, P., & Bozkurt, A. Y. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2923153 · Cited by 19
A novel temporal adaptive fuzzy neural network for facial feature based fatigue assessment
Zhang, Z., Wang, H., You, Q., Chen, L., & Ning, H. (2024) · Expert Systems with Applications · doi:10.1016/j.eswa.2024.124124 · Cited by 19
Health analysis of bicycle rider and security of bicycle using IoT
Nath, S., Sinha, S., Gladence, L. M., BevishJinila, Y., & Rajalakshmi, V. (2017) · doi:10.1109/iccsp.2017.8286475 · Cited by 18
Carek, A. M., Jung, H., & Inan, O. T. (2019) · IEEE Sensors Journal · doi:10.1109/jsen.2019.2960063 · Cited by 18
Sinichi, M., Gevonden, M. J., & Krabbendam, L. (2025) · Psychophysiology · doi:10.1111/psyp.70004 · Cited by 26
Yamakoshi, T., Lee, J., Matsumura, K., Yamakoshi, Y., Rolfe, P. J., Kiyohara, D., & Yamakoshi, K.-I. (2015) · PLoS ONE · doi:10.1371/journal.pone.0143506 · Cited by 17
Self-citationPont, M. P. D., & Marques, J. L. B. (2020) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2019.2963508 · Cited by 17
Evaluating RGB channels in remote photoplethysmography: a comparative study with contact-based PPG
Ontiveros, R. C., Elgendi, M., Missale, G., & Menon, C. (2023) · Frontiers in Physiology · doi:10.3389/fphys.2023.1296277 · Cited by 16
Bauer, M., Bräuer, C., Schuldt, J., Niemann, M., & Krömker, H. (2018) · Advances in intelligent systems and computing · doi:10.1007/978-3-319-94619-1_4 · Cited by 16
iPhysioMeter: A Smartphone Photoplethysmograph for Measuring Various Physiological Indices
Matsumura, K., Rolfe, P. J., & Yamakoshi, T. (2014) · Methods in molecular biology · doi:10.1007/978-1-4939-2172-0_21 · Cited by 17
Self-citationWearable Blood Pressure Monitoring Devices: Understanding Heterogeneity in Design and Evaluation
Cheung, M. Y., Sabharwal, A., Coté, G. L., & Veeraraghavan, A. (2024) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2024.3434344 · Cited by 23
Biometrie authentication using hand movement information from wrist-worn PPG sensors
Ohtsuki, T. O., & Hiroto, K. (2016) · IEEE International Symposium on Personal, Indoor and Mobile Radio Communications (PIMRC) · doi:10.1109/pimrc.2016.7794969 · Cited by 15
Giovanni, E. D., Srinivasan Murali, A., Rincón, F. J., & Atienza, D. (2016) · Euromicro Conference on Digital System Design (DSD) · doi:10.1109/dsd.2016.101 · Cited by 15
Noncontact measurement of emotional and physiological changes in heart rate from a webcam
Madan, C. R., Harrison, T., & Mathewson, K. E. (2017) · Psychophysiology · doi:10.1111/psyp.13005 · Cited by 16
Pretty, E. J., Guarese, R., Dziego, C. A., Fayek, H. M., & Zambetta, F. (2024) · IEEE Transactions on Games · doi:10.1109/tg.2024.3406723 · Cited by 15
Low-cost portable heart rate monitoring based on photoplethysmography and decision tree
Tunggal, T. P., Latif, A., & Iswanto, I. (2016) · AIP conference proceedings · doi:10.1063/1.4958522 · Cited by 14
Blažek, V., Blanik, N., Blazek, C., Paul, M., Pereira, C. B., Köny, M., Venema, B., & Leonhardt, S. K. (2016) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000001388 · Cited by 14
Remote measurement of heart rate from facial video in different scenarios
Zheng, X., Zhang, C., Chen, H., Zhang, Y., & Yang, X. (2021) · Measurement · doi:10.1016/j.measurement.2021.110243 · Cited by 14
Hayirlioglu, Y. Z., & Semiz, B. (2024) · IEEE Sensors Journal · doi:10.1109/jsen.2024.3403846 · Cited by 13
Preclinical Evaluation of a Wearable Wristband With Compressed-Sensing Based Photoplethysmography
Ahmmed, P., Garceau, E. L., Latif, T., Brewer, A., Dieffenderfer, J. P., Valero-Sarmiento, J. M., Pamula, V. R., Helleputte, N. V., Hoof, C. A. V., Verhelst, M., et al. (2024) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2024.3513462 · Cited by 15
Embedded system in handheld water turbidity meter for smallholders
Putra, B. T. W., Rocelline, L. A., & Syahputra, W. N. H. (2022) · Microprocessors and Microsystems · doi:10.1016/j.micpro.2022.104603 · Cited by 13
Blanos, P., Hu, S., Mulvaney, D. J., & Alharbi, S. (2018) · Design and Quality for Biomedical Technologies · doi:10.1117/12.2287854 · Cited by 13
Ahmmed, P., Dieffenderfer, J. P., Valero-Sarmiento, J. M., Pamula, V. R., Helleputte, N. V., Hoof, C. A. V., Verhelst, M., & Bozkurt, A. Y. (2019) · Wearable and Implantable Body Sensor Networks · doi:10.1109/bsn.2019.8771074 · Cited by 13
Ceugniez, M., Devanne, H., & Hermand, É. (2024) · JMIR mhealth and uhealth · doi:10.2196/54871 · Cited by 13
Low-cost wearable bluetooth sensor for epileptic episodes detection
Ramirez-Alaminos, J. M., Sendra, S., Lloret, J., & Navarro‐Ortiz, J. (2017) · International Conference on Communications · doi:10.1109/icc.2017.7997413 · Cited by 12
Watson, A., Kendell, C., Lingamoorthy, A., Lee, I., & Weimer, J. E. (2022) · Proceedings of the ACM on Interactive Mobile Wearable and Ubiquitous Technologies · doi:10.1145/3569502 · Cited by 12
Bioelectronic devices for light-based diagnostics and therapies
Kim, H., Sritandi, W., Xiong, Z., & Ho, J. S. (2023) · Biophysics Reviews · doi:10.1063/5.0102811 · Cited by 12
Wearable edge machine learning with synthetic photoplethysmograms
Sirkiä, J., Panula, T., & Kaisti, M. (2023) · Expert Systems with Applications · doi:10.1016/j.eswa.2023.121523 · Cited by 12
Metadata
- Authors
- Lee, J., Matsumura, K., Yamakoshi, K.-I., Rolfe, P., Tanaka, S., & Yamakoshi, T.
- Affiliations
- Kanazawa University (JP)
- Harbin Institute of Technology (CN)
- Venue
- 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
- Publisher
- IEEE
- Published
- 2013-07
- Volume
- 2013
- Issue
- 6609852
- Pages
- 1724-1727
- ISSN
- 2375-7477 (ISSN-L)
- OpenAlex
- W1988770599
- Cited by
- 202
- References
- 12
- Citations by year
- 2026: 4 · 2025: 8 · 2024: 23 · 2023: 26 · 2022: 19 · 2021: 16 · 2020: 14 · 2019: 19 · 2018: 23 · 2017: 19 · 2016: 11 · 2015: 15 · 2014: 5
- FWCI
- 20.48
- Type
- conference paper
- Language
- en
- Open access
- Locations
- Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) (publishedVersion)
- PubMed (publishedVersion)
- Institutional Repositories DataBase (IRDB) (submittedVersion)
- Institutional Repositories DataBase (IRDB) (submittedVersion, cc-by-nc-nd)
- Kanazawa University Repository for Academic Resources (DSpace) (Kanazawa University) (submittedVersion, cc-by-nc-nd)
- Funders
- Kissei
- Japan Society for the Promotion of Science (24650443)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Keywords
- heart rate monitoring, motion artifacts, signal-to-noise ratio, hand waving, ECG comparison, wearable sensors, non-contact vital sign monitoring
- MeSH terms
- Adult, Heart Rate, Humans, Light, Male, Movement, Photoplethysmography, Monitoring, Ambulatory, Young Adult, Signal-To-Noise Ratio
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review