Active Books: The Design of an Implantable Stimulator That Minimizes Cable Count Using Integrated Circuits Very Close to Electrodes
Abstract
This paper presents an integrated stimulator that can be embedded in implantable electrode books for interfacing with nerve roots at the cauda equina. The Active Book overcomes the limitation of conventional nerve root stimulators which can only support a small number of stimulating electrodes due to cable count restriction through the dura. Instead, a distributed stimulation system with many tripole electrodes can be configured using several Active Books which are addressed sequentially. The stimulator was fabricated in a 0.6-μm high-voltage CMOS process and occupies a silicon area of 4.2 × 6.5 mm(2). The circuit was designed to deliver up to 8 mA stimulus current to tripole electrodes from an 18 V power supply. Input pad count is limited to five (two power and three control lines) hence requiring a specific procedure for downloading stimulation commands to the chip and extracting information from it. Supported commands include adjusting the amplitude of stimulus current, varying the current ratio at the two anodes in each channel, and measuring relative humidity inside the chip package. In addition to stimulation mode, the chip supports quiescent mode, dissipating less than 100 nA current from the power supply. The performance of the stimulator chip was verified with bench tests including measurements using tripoles in saline.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
For humidity sensing, capacitive measurement techniques have been reported with advanced compensatory methods to reduce parasitics and increase the readout sensitivity to achieve sub-fF resolution using nA-scale currents for charging and discharging the capacitors.
“For humidity sensing, capacitive measurement techniques have been reported [80] with advanced compensatory methods to reduce parasitics and increase the readout sensitivity to achieve sub-fF resolution using nA-scale currents for charging and discharging the capacitors [127].”
From Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review (2024) · cites this paper as [80]
Platinum black and nanostructured platinum grass coatings have both shown significant improvements in electrode impedance, due to an increased surface area.
“More specifically, platinum black (Tand et al. 2014) and nanostructured platinum grass (Liu et al. 2012) coatings have both shown significant improvements in electrode impedance, due to an increased surface area.”
From Realizing flexible bioelectronic medicines for accessing the peripheral nerves – technology considerations (2018) · cites this paper as Liu et al. 2012
Cites
Harkema, S., Gerasimenko, Y., Hodes, J., Burdick, J., Angeli, C., Chen, Y., Ferreira, C., Willhite, A., Rejc, E., Grossman, R. G., et al. (2011) · The Lancet · doi:10.1016/s0140-6736(11)60547-3 · Cited by 1,229 · see citing works
Transformation of nonfunctional spinal circuits into functional states after the loss of brain input
Courtine, G., Gerasimenko, Y., van den Brand, R., Yew, A., Musienko, P., Zhong, H., Song, B., Ao, Y., Ichiyama, R. M., Lavrov, I., et al. (2009) · Nature Neuroscience · doi:10.1038/nn.2401 · Cited by 706 · see citing works
Microelectrodes, Microelectronics, and Implantable Neural Microsystems
Wise, K. D., Sodagar, A. M., Yao, Y., Gulari, M. N., Perlin, G. E., & Najafi, K. (2008) · Proceedings of the IEEE · doi:10.1109/jproc.2008.922564 · Cited by 336 · see citing works
The Effect of Stimulus Parameters on the Recruitment Characteristics of Direct Nerve Stimulation
Gorman, P. H., & Mortimer, J. T. (1983) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1983.325041 · Cited by 313 · see citing works
Sacral anterior root stimulators for bladder control in paraplegia
Brindley, G. S., Polkey, C. E., & Rushton, D. N. (1982) · Spinal Cord · doi:10.1038/sc.1982.65 · Cited by 311 · see citing works
Minassian, K., Persy, I., Rattay, F., Pinter, M. M., Kern, H., & Dimitrijevic, M. R. (2007) · Human Movement Science · doi:10.1016/j.humov.2007.01.005 · Cited by 203 · see citing works
Implant Size and Fixation Mode Strongly Influence Tissue Reactions in the CNS
Thelin, J., Jörntell, H., Psouni, E., Garwicz, M., Schouenborg, J., Danielsen, N., & Linsmeier, C. E. (2011) · PLoS ONE · doi:10.1371/journal.pone.0016267 · Cited by 188 · see citing works
An implantable CMOS circuit interface for multiplexed microelectrode recording arrays
Ji, J., & Wise, K. D. (1992) · IEEE Journal of Solid-State Circuits · doi:10.1109/4.121568 · Cited by 173 · see citing works
A CMOS Chip With Active Pixel Array and Specific Test Features for Subretinal Implantation
Rothermel, A., Liu, L., Aryan, N. P., Fischer, M. D., Wuenschmann, J., Kibbel, S., & Harscher, A. (2009) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2008.2007436 · Cited by 109 · see citing works
The first 500 sacral anterior root stimulators: implant failures and their repair
Brindley, G. S. (1995) · Spinal Cord · doi:10.1038/sc.1995.3 · Cited by 89 · see citing works
Kirkham, A. P. S., Knight, S. L., Craggs, M. D., Casey, A. T. H., & Shah, P. J. R. (2002) · Spinal Cord · doi:10.1038/sj.sc.3101278 · Cited by 85 · see citing works
Donaldson, N., & Donaldson, P. E. K. (1986) · Medical & Biological Engineering & Computing · doi:10.1007/bf02441604 · Cited by 82 · see citing works
Self-citationArchitecture tradeoffs in high-density microstimulators for retinal prosthesis
Sivaprakasam, M., Liu, W., Wang, G., Weiland, J. D., & Humayun, M. S. (2005) · IEEE Transactions on Circuits and Systems I Fundamental Theory and Applications · doi:10.1109/tcsi.2005.857554 · Cited by 74 · see citing works
BION microstimulators: A case study in the engineering of an electronic implantable medical device
Kane, M. J., Breen, P. P., Quondamatteo, F., & OLaighin, G. M. (2010) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2010.08.010 · Cited by 65 · see citing works
Perkins, T. A., Donaldson, N., Hatcher, N. A. C., Swain, I., & Wood, D. E. (2002) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2002.802860 · Cited by 64 · see citing works
Self-citationThe frequency response of frog muscle spindles under various conditions
Kirkwood, P. A. (1972) · The Journal of Physiology · doi:10.1113/jphysiol.1972.sp009791 · Cited by 63 · see citing works
Donaldson, N., & Donaldson, P. E. K. (1986) · Medical & Biological Engineering & Computing · doi:10.1007/bf02441605 · Cited by 47 · see citing works
Self-citation(1972) · PubMed · Cited by 43 · see citing works
BION™ Implants for Therapeutic and Functional Electrical Stimulation
Loeb, G. E., & RICHMOND, F. J. R. (2000) · Frontiers in neuroscience · doi:10.1201/9781420039054.ch3 · Cited by 35 · see citing works
Neuroprostheses for leg function after spinal-cord injury
Donaldson, N., Rushton, D., & Tromans, T. (1997) · The Lancet · doi:10.1016/s0140-6736(97)24036-5 · Cited by 29 · see citing works
Self-citationDesign of a CMOS oscillator with magnetic-field frequency modulation
Nathan, A., McKay, I., Filanovsky, I. M., et al. (1987) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.1987.1052707 · Cited by 27 · see citing works
Lumbar Root Stimulation for Restoring Leg Function: Results in Paraplegia
Rushton, D. N., Donaldson, N., Barr, F. M. D., Harper, V., Perkins, T. A., Taylor, P. N., & Tromans, A. M. (1997) · Artificial Organs · doi:10.1111/j.1525-1594.1997.tb04647.x · Cited by 26 · see citing works
Self-citationAn Integrated Stimulator With DC-Isolation and Fine Current Control for Implanted Nerve Tripoles
Liu, X., Demosthenous, A., & Donaldson, N. (2011) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2011.2144310 · Cited by 21 · see citing works
Self-citationLanglois, P. J., Demosthenous, A., Pachnis, I., & Donaldson, N. (2009) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2009.2034138 · Cited by 20 · see citing works
Self-citationPassive Neutralization of Myoelectric Interference From Neural Recording Tripoles
Pachnis, I., Demosthenous, A., & Donaldson, N. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.891170 · Cited by 20 · see citing works
Self-citationTemperature and perfusion responses of muscle and lung tissue during chronic heatingin vivo
Saidel, G. M., Davies, C. R., Liu, E. H., et al. (2001) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345276 · Cited by 17 · see citing works
Design of an Implant for Preventing Incontinence After Spinal Cord Injury
Donaldson, N., Perkins, T., Pachnis, I., et al. (2008) · Artificial Organs · doi:10.1111/j.1525-1594.2008.00606.x · Cited by 13 · see citing works
Self-citationA programmable ENG amplifier with passive EMG neutralization for FES applications
Demosthenous, A., Jiang, D., Pachnis, I., Liu, X., Rahal, M., & Donaldson, N. (2008) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2008.4541727 · Cited by 8 · see citing works
Self-citationThe Effects of Fabrication Process on the Performance of a CMOS Based Capacitive Humidity Sensor
Saeidi, N., Blake, A., Colinge, C., et al. (2011) · ECS Transactions · doi:10.1149/1.3653925 · Cited by 8 · see citing works
Implanted devices: Improved methods for nerve root stimulation
Vanhoestenberghe, A. (2008) · OpenGrey (Institut de l'Information Scientifique et Technique) · Cited by 8 · see citing works
Self-citationSaeidi, N., Demosthenous, A., Donaldson, N., et al. (2009) · UCL Discovery (University College London) · Cited by 7 · see citing works
Self-citationAn implantable humidity-to-frequency sensor in CMOS technology
Cirmirakis, D., Demosthenous, A., Saeidi, N., Vanhoestenberghe, A., & Donaldson, N. (2011) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/icsens.2011.6127363 · Cited by 6 · see citing works
Self-citationSchüettler, M., Vanhoestenberghe, A., Saeidi, N., et al. (2011) · doi:10.1109/iembs.2011.6090814 · Cited by 5 · see citing works
Self-citationDesign of a stimulator ASIC for active electrode books
Liu, X., Demosthenous, A., Vanhoestenberghe, A., & Donaldson, N. (2010) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2010.5709597 · Cited by 1 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Giagka, V., & Serdijn, W. A. (2018) · Bioelectronic Medicine · doi:10.1186/s42234-018-0010-y · Cited by 58 · see citing works
A Multichannel High-Frequency Power-Isolated Neural Stimulator With Crosstalk Reduction
Jiang, D., & Demosthenous, A. (2018) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2018.2832541 · Cited by 36 · see citing works
Self-citationAn ASIC for Recording and Stimulation in Stacked Microchannel Neural Interfaces
Lancashire, H. T., Jiang, D., Demosthenous, A., & Donaldson, N. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2891284 · Cited by 15 · see citing works
Self-citationZhou, Y., Wang, K., Yin, S., Li, W.-Y., Meng, F., Wang, Z.-G., & Ma, K. (2024) · IEEE Transactions on Circuits and Systems I: Regular Papers · doi:10.1109/tcsi.2024.3376370 · Cited by 12 · see citing works
Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review
Habibollahi, M., Jiang, D., Lancashire, H. T., & Demosthenous, A. (2024) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2024.3430038 · Cited by 11 · see citing works
“They check for available headroom with respect to the interface impedance [44], [129], and detect excess power or leakage of fluids via temperature and humidity sensing [67], [80].”
InfluentialSelf-citation
- Authors
- Liu, X., Demosthenous, A., Vanhoestenberghe, A., Jiang, D., & Donaldson, N.
- Affiliations
- University College London (GB)
- Venue
- IEEE Transactions on Biomedical Circuits and Systems
- Published
- 2012-06
- Volume
- 6
- Issue
- 3
- Pages
- 216-227
- ISSN
- 1932-4545 (ISSN-L)
- 1940-9990
- OpenAlex
- W2147707331
- Cited by
- 39
- References
- 34
- Citations by year
- 2026: 1 · 2025: 1 · 2024: 3 · 2022: 3 · 2021: 1 · 2020: 2 · 2019: 1 · 2018: 5 · 2017: 2 · 2016: 6 · 2015: 3 · 2014: 7 · 2013: 4
- FWCI
- 1.84
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- IEEE Transactions on Biomedical Circuits and Systems (publishedVersion)
- PubMed (publishedVersion)
- UCL Discovery (University College London) (submittedVersion)
- Funders
- Engineering and Physical Sciences Research Council (EP/F009593/1)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Biomedical Engineering › Analog and Mixed-Signal Circuit Design
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Keywords
- implantable neural stimulator, nerve root stimulation, high-voltage CMOS, integrated circuit design, relative humidity measurement
- MeSH terms
- Biomedical Engineering, Cauda Equina, Computers, Electric Stimulation, Electrodes, Implanted, Electronics, Medical, Equipment Design, Humans, Humidity, Male, Motor Neurons, Electric Power Supplies, Signal Processing, Computer-Assisted, Implantable Neurostimulators, Wireless Technology
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review