A Multichannel High-Frequency Power-Isolated Neural Stimulator With Crosstalk Reduction
Open accessAbstract
In neuroprostheses applications requiring simultaneous stimulations on a multielectrode array, electric crosstalk, the spatial interaction between electric fields from various electrodes is a major limitation to the performance of multichannel stimulation. This paper presents a multichannel stimulator design that combines high-frequency current stimulation (using biphasic charge-balanced chopped pulse profile) with a switched-capacitor power isolation method. The approach minimizes crosstalk and is particularly suitable for fully integrated realization. A stimulator fabricated in a 0.6 μm CMOS high-voltage technology is presented. It is used to implement a multichannel, high-frequency, power-isolated stimulator. Crosstalk reduction is demonstrated with electrodes in physiological media while the efficacy of the high-frequency stimulator chip is proven in vivo. The stimulator provides fully independent operation on multiple channels and full flexibility in the design of neural modulation protocols.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
High compliance voltage for current stimulation consumes large amounts of power and may result in redundant voltage overhead on the electrodes, as their impedance may vary over time and position by several orders of magnitude.
“This consumes large amounts of power and may result in redundant voltage overhead on the electrodes, as their impedance may vary over time and position by several orders of magnitude [78], [128].”
From Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review (2024) · cites this paper as [78]
The Discussion (G. Discussion; page unavailable) says dynamic supply-voltage control could improve power efficiency by reducing redundant voltage headroom on the electrodes. It does not say that electrode impedance varies over time and position by several orders of magnitude, or specifically attribute large power consumption to high compliance voltage. (LLM, conf=0.84)
Experimental — accuracy not yet measuredHV technologies provide thick gate oxide devices that withstand higher voltages and superior isolation between stimulating channels, particularly via shallow or deep trench isolation, which separates the module substrates, thus limiting the occurrence of crosstalk.
“HV technologies provide thick gate oxide devices that withstand higher voltages and superior isolation between stimulating channels, particularly via shallow or deep trench isolation, which separates the module substrates, thus limiting the occurrence of crosstalk [78].”
From Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review (2024) · cites this paper as [78]
Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measuredProximity of interface sites on tissue can result in crosstalk between channels due to current spread, which can compromise efficacy of the neural interface solution, requiring electrical or physical isolation between interface channels.
“However, the proximity of interface sites on the tissue can result in crosstalk between channels due to current spread. This can compromise the efficacy of the neural interface solution [77], requiring electrical [78] or physical isolation between the interface channel [71].”
From Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review (2024) · cites this paper as [78]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Restoring natural sensory feedback in real-time bidirectional hand prostheses
Raspopovic, S., Capogrosso, M., Petrini, F. M., Bonizzato, M., Rigosa, J., Di Pino, G., Carpaneto, J., Controzzi, M., Boretius, T., Fernandez, E., et al. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3006820 · Cited by 1,001 · see citing works
Cochlear Implants: System Design, Integration, and Evaluation
Zeng, F.-G., Rebscher, S., Harrison, W., Sun, X., & Feng, H. (2008) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2008.2008250 · Cited by 878 · see citing works
Weiland, J. D., Liu, W., & Humayun, M. S. (2005) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.7.060804.100435 · Cited by 371 · see citing works
A 232-Channel Epiretinal Stimulator ASIC
Ortmanns, M., Rocke, A., Gehrke, M., & Tiedtke, H. J. (2007) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2007.908693 · Cited by 225 · see citing works
Shepherd, R. K., & Javel, E. (1999) · Hearing Research · doi:10.1016/s0378-5955(99)00011-8 · Cited by 189 · see citing works
Santina, C. C. D., Migliaccio, A. A., & Patel, A. H. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.894629 · Cited by 173 · see citing works
Sit, J.-J., & Sarpeshkar, R. (2007) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2007.911631 · Cited by 155 · see citing works
Prototype Neural Semicircular Canal Prosthesis using Patterned Electrical Stimulation
Gong, W., & Merfeld, D. M. (2000) · Annals of Biomedical Engineering · doi:10.1114/1.293 · Cited by 141 · see citing works
Arfin, S. K., & Sarpeshkar, R. (2012) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2166072 · Cited by 133 · see citing works
A Microchannel Neuroprosthesis for Bladder Control After Spinal Cord Injury in Rat
Chew, D. J., Zhu, L., Delivopoulos, E., Minev, I. R., Musick, K. M., Mosse, C. A., Craggs, M., Donaldson, N., Lacour, S. P., McMahon, S. B., et al. (2013) · Science Translational Medicine · doi:10.1126/scitranslmed.3007186 · Cited by 123 · see citing works
Wilson, B. S., Finley, C. C., Lawson, D. T., Wolford, R. D., & Zerbi, M. (1993) · PubMed · Cited by 109 · see citing works
A Power-Efficient Neural Tissue Stimulator With Energy Recovery
Kelly, S. K., & Wyatt, J. L. (2011) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2010.2076384 · Cited by 107 · see citing works
An Energy-Efficient, Dynamic Voltage Scaling Neural Stimulator for a Proprioceptive Prosthesis
Williams, I. A., & Constandinou, T. G. (2013) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2013.2256906 · Cited by 101 · see citing works
Jiang, D., Cirmirakis, D., Schormans, M., Perkins, T. A., Donaldson, N., & Demosthenous, A. (2016) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2016.2580513 · Cited by 94 · see citing works
Self-citationAcclimation to Chronic Constant-Rate Peripheral Stimulation Provided by a Vestibular Prosthesis
Merfeld, D. M., Gong, W., Morrissey, J. H., Saginaw, M. A., Haburcakova, C., & Lewis, R. F. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2006.883645 · Cited by 93 · see citing works
A three-dimensional self-opening intraneural peripheral interface (SELINE)
Cutrone, A., Valle, J. D., Santos, D., Badia, J., Filippeschi, C., Micera, S., Navarro, X., & Bossi, S. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/1/016016 · Cited by 82 · see citing works
An advanced multiple channel cochlear implant
McDermott, H. J. (1989) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.32112 · Cited by 81 · see citing works
Retinal Neurostimulator for a Multifocal Vision Prosthesis
Wong, Y. T., Dommel, N., Preston, P., Hallum, L. E., Lehmann, T., Lovell, N. H., & Suaning, G. J. (2007) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2007.903958 · Cited by 79 · see citing works
A Partial-Current-Steering Biphasic Stimulation Driver for Vestibular Prostheses
Constandinou, T. G., Georgiou, J., & Toumazou, C. (2008) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2008.927238 · Cited by 76 · see citing works
Chiang, B., Fridman, G. Y., Dai, C., Rahman, M. A., & Santina, C. C. D. (2011) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2011.2164937 · Cited by 65 · see citing works
A Stimulator ASIC Featuring Versatile Management for Vestibular Prostheses
Jiang, D., Demosthenous, A., Perkins, T. A., Liu, X., & Donaldson, N. (2011) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2138139 · Cited by 54 · see citing works
Self-citationHsu, W.-Y., & Schmid, A. (2017) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2017.2694144 · Cited by 49 · see citing works
Davidovics, N. S., Fridman, G. Y., & Santina, C. C. D. (2012) · Experimental Brain Research · doi:10.1007/s00221-012-3025-8 · Cited by 48 · see citing works
van Dongen, M. N., & Serdijn, W. A. (2014) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2014.2363736 · Cited by 40 · see citing works
Wilson, B. S., & Dorman, M. F. (2008) · IEEE Sensors Journal · doi:10.1109/jsen.2007.912917 · Cited by 40 · see citing works
Liu, X., Demosthenous, A., Vanhoestenberghe, A., Jiang, D., & Donaldson, N. (2012) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2174360 · Cited by 39 · see citing works
Self-citationFlexible Charge Balanced Stimulator With 5.6 fC Accuracy for 140 nC Injections
Nag, S., Jia, X., Thakor, N. V., & Sharma, D. (2012) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2012.2205574 · Cited by 38 · see citing works
A CMOS Neural Interface for a Multichannel Vestibular Prosthesis
Hageman, K. N., Kalayjian, Z. K., Tejada, F. J., Chiang, B., Rahman, M. A., Fridman, G. Y., Dai, C., Pouliquen, P. O., Georgiou, J., Santina, C. C. D., et al. (2015) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2015.2409797 · Cited by 36 · see citing works
Lu, Y., Yan, Y., Chai, X., Ren, Q., Chen, Y., & Li, L. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/3/036022 · Cited by 32 · see citing works
An Implantable Versatile Electrode-Driving ASIC for Chronic Epidural Stimulation in Rats
Giagka, V., Eder, C. F., Donaldson, N., & Demosthenous, A. (2014) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2014.2330859 · Cited by 25 · see citing works
Self-citationDevelopment, manufacturing and application of double-sided flexible implantable microelectrodes
Poppendieck, W., Sossalla, A., Krob, M.-O., Welsch, C., Nguyen, T. A. K., Gong, W., DiGiovanna, J., Micera, S., Merfeld, D. M., & Hoffmann, K. (2014) · Biomedical Microdevices · doi:10.1007/s10544-014-9887-8 · Cited by 25 · see citing works
A Vestibular Prosthesis With Highly-Isolated Parallel Multichannel Stimulation
Jiang, D., Cirmirakis, D., & Demosthenous, A. (2015) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2014.2323310 · Cited by 20 · see citing works
Self-citationA custom-designed receiver-stimulator chip for an advanced multiple-channel hearing prosthesis
McDermott, H. J. (1991) · IEEE Journal of Solid-State Circuits · doi:10.1109/4.90069 · Cited by 15 · see citing works
van Dongen, M. N., Hoebeek, F. E., Koekkoek, S. K. E., De Zeeuw, C. I., & Serdijn, W. A. (2015) · Frontiers in Neuroengineering · doi:10.3389/fneng.2015.00002 · Cited by 15 · see citing works
Microchannel neural interface manufacture by stacking silicone and metal foil laminae
Lancashire, H. T., Vanhoestenberghe, A., Pendegrass, C. J., Ajam, Y. A., Magee, E., Donaldson, N., & Blunn, G. W. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/3/034001 · Cited by 13 · see citing works
Advances in Scalable Implantable Systems for Neurostimulation Using Networked ASICs
Liu, X., Zong, Z., Jiang, D., Bougaila, B., Donaldson, N., & Demosthenous, A. (2016) · IEEE Design and Test · doi:10.1109/mdat.2016.2533358 · Cited by 7 · see citing works
Self-citation
Flagged first, then by citations
Research works printed but not linked: 1
- B. Chiang, G. Fridman, C. Dai, M. Rahman, and C. Della Santina, "Design and performance of a multichannel vestibular prosthesis," IEEE Trans. Neural Syst. Rehabil. Eng., vol. 19, no. 5, pp. 588-598, Oct. 2011.
Cited by
Sorted by citations
Zeng, Q., & Huang, Z. (2023) · Advanced Functional Materials · doi:10.1002/adfm.202301223 · Cited by 103 · see citing works
Neural interfaces: Bridging the brain to the world beyond healthcare
Xu, S., Liu, Y., Lee, H., & Li, W. (2024) · Exploration · doi:10.1002/exp.20230146 · Cited by 96 · see citing works
Son, J.-Y., & Cha, H. (2020) · IEEE Access · doi:10.1109/access.2020.3012028 · Cited by 35 · see citing works
Wu, Y., Jiang, D., & Demosthenous, A. (2021) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2021.3139996 · Cited by 26 · see citing works
Self-citationDesign of Dual-Mode Stimulus Chip With Built-In High Voltage Generator for Biomedical Applications
Yen, T.-Y., & Ker, M. (2020) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2020.2999398 · Cited by 26 · see citing works
An Ultra High-Frequency 8-Channel Neurostimulator Circuit With $\text{68}\%$ Peak Power Efficiency
Urso, A., Giagka, V., van Dongen, M. N., & Serdijn, W. A. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2920294 · Cited by 25 · see citing works
Wu, Y., Jiang, D., & Demosthenous, A. (2022) · UCL Discovery (University College London) · Cited by 20 · see citing works
Self-citationKatebi, M., Erfanian, A., Karami, M. A., & Sawan, M. (2024) · IEEE Access · doi:10.1109/access.2024.3432611 · Cited by 15 · see citing works
IC-Based Neuro-Stimulation Environment for Arbitrary Waveform Generation
Kölbl, F., Bornat, Y., Castelli, J., Regnacq, L., N’Kaoua, G., Renaud, S., & Lewis, N. (2021) · Electronics · doi:10.3390/electronics10151867 · Cited by 13 · see citing works
Eom, K., Lee, H.-S., Park, M., Yang, S. M., Choe, J. C., Hwang, S., Suh, Y.-W., & Lee, H. (2023) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2023.3334753 · Cited by 12 · see citing works
Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review
Habibollahi, M., Jiang, D., Lancashire, H. T., & Demosthenous, A. (2024) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2024.3430038 · Cited by 11 · see citing works
“This can compromise the efficacy of the neural interface solution [77], requiring electrical [78] or physical isolation between the interface channel [71].”
Self-citationEom, K., Lee, H.-S., Ku, S.-B., Kang, J., Jung, H., Kim, T., Joo, J., Kim, T.-K., Shon, Y., & Lee, H. (2025) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2025.3542022 · Cited by 8 · see citing works
An Active Microchannel Neural Interface for Implantable Electrical Stimulation and Recording
Habibollahi, M., Jiang, D., Lancashire, H. T., & Demosthenous, A. (2025) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2025.3533612 · Cited by 8 · see citing works
Self-citationJiang, D., Liu, F., Lancashire, H. T., Perkins, T. A., Schormans, M., Vanhoestenberghe, A., Donaldson, N., & Demosthenous, A. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.681021 · Cited by 8 · see citing works
Self-citationLee, S.-Y., Liao, Z.-X., Feng, I.-T., Lee, H.-Y., & Lin, C. K. (2024) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2024.3380055 · Cited by 6 · see citing works
An Active Microchannel Neural Interface with Artifact Reduction
Habibollahi, M., Hanzaee, F. F., Jiang, D., Lancashire, H., & Demosthenous, A. (2021) · 2021 IEEE International Symposium on Circuits and Systems (ISCAS) · doi:10.1109/iscas51556.2021.9401318 · Cited by 5 · see citing works
Self-citationLiu, F., Habibollahi, M., Wu, Y., Neshatvar, N., Zhang, J., Zinno, C., Akouissi, O., Bernini, F., Alibrandi, L., Gabisonia, K., et al. (2024) · Bioelectronic Medicine · doi:10.1186/s42234-024-00148-3 · Cited by 4 · see citing works
Self-citationMassaioli, S., Carlino, M. F., & Gielen, G. E. (2025) · IEEE Transactions on Circuits and Systems I Regular Papers · doi:10.1109/tcsi.2025.3555895 · Cited by 3 · see citing works
Palomeque-Mangut, D., Rodriguez-Vazquez, A., & Delgado‐Restituto, M. (2022) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas54905.2022.9948547 · Cited by 3 · see citing works
Eom, K., Lee, H.-S., Park, M., Yang, S. M., Choe, J. C., Hwang, S., Suh, Y.-W., & Lee, H. (2024) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2024.3460814 · Cited by 3 · see citing works
Moganti, G. L. K., Praneeth, V. N. S., & Vanjari, S. R. K. (2021) · IEEE International Conference on Intelligent Data Acquisition and Advanced Computing Systems: Technology and Applications (IDAACS) · doi:10.1109/idaacs53288.2021.9660895 · Cited by 3 · see citing works
Moganti, G. L. K., Praneeth, V. N. S., & Vanjari, S. R. K. (2022) · Sensors · doi:10.3390/s22051756 · Cited by 3 · see citing works
An Efficient and Secure Transmission Model Based on Compressive Sensing
Li, L., Peng, H., Liu, L., & Yang, Y. (2018) · International Symposium on Security and Privacy in Social Networks and Big Data (SocialSec) · doi:10.1109/socialsec.2018.8760382 · Cited by 2 · see citing works
A Bidirectional ASIC for Active Microchannel Neural Interfaces
Habibollahi, M., Jiang, D., Lancashire, H. T., & Demosthenous, A. (2022) · IEEE International Conference on Electronics, Circuits and Systems (ICECS) · doi:10.1109/icecs202256217.2022.9970803 · Cited by 2 · see citing works
Self-citationA Two Channels Fully-Programmable Integrated Neurostimulator with a 0.07% Charge Mismatch
Contardi, S., Ria, A., Nannipieri, I., Scognamiglio, M., & Piotto, M. (2024) · Conference on Ph.D. Research in Microelectronics and Electronics · doi:10.1109/prime61930.2024.10559689 · Cited by 1 · see citing works
A High-Density Silicon-Based Microelectrode Array for Chronic In Vivo Neural Recording
Geng, L., Wang, Y., Gu, W., Yang, Y., Xie, D., Ma, X., Chen, Z., Xu, L., Li, C., & Wu, F. H. (2025) · Journal of Microelectromechanical Systems · doi:10.1109/jmems.2025.3621151 · Cited by 1 · see citing works
Current-Based Neurostimulation Circuit and System Techniques
Shirafkan, R., & Shoaei, O. (2020) · Springer eBooks · doi:10.1007/978-1-4614-6623-9_42-1 · Cited by 1 · see citing works
Wireless Programmable Healthy Neural Stimulator Prototype
Moganti, G. L. K., & Rishita, P. Y. (2021) · International Conference on Smart Computing and Communications (ICSCC) · doi:10.1109/icscc51209.2021.9528004 · Cited by 1 · see citing works
A High-Voltage, Implantable ASIC for Active Interfaces of the Vagus Nerve
Habibollahi, M., Li, J., Almarri, N., Jiang, D., Lancashire, H. T., & Demosthenous, A. (2023) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas58349.2023.10389014 · Cited by 1 · see citing works
Self-citationCurrent-Based Neurostimulation Circuit and System Techniques
Shirafkan, R., & Shoaei, O. (2022) · Springer eBooks · doi:10.1007/978-1-4614-3447-4_42 · Cited by 1 · see citing works
Strategies on Deep Brain Stimulation Devices for Effective Behavioral Studies in Rodents
Plocksties, F. V., Lüttig, A., Niemann, C., Uster, F., Franz, D., Kober, M., Koschay, M., Perl, S., Richter, A., Köhling, R., et al. (2022) · IEEE-EMBS Conference on Biomedical Engineering and Sciences · doi:10.1109/iecbes54088.2022.10079260
Wu, J., Yin, S., Zhou, Y., Li, J., Li, K., Shen, X., Sun, T., Zhou, Z., Ma, K., Meng, F., et al. (2025) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2025.3623450
Koolani, M., Mansoury, H. S., Salehian, M., & Atarodi, M. (2026) · AEU - International Journal of Electronics and Communications · doi:10.1016/j.aeue.2026.156257
Wu, Z., & Liu, X. (2025) · IEEE Asia Pacific Conference on Circuits and Systems (APCCAS) · doi:10.1109/apccas67402.2025.11376911
An Energy-Efficient Neurostimulator with Closed-Loop Boost Converter for Dynamic Supply Regulation
Chang, Q., & Liu, X. (2026) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas66217.2026.11562774
A Fully Programmable Dual-Channel CMOS Neurostimulator With Ultra-Low Charge Mismatch
Contardi, S., Ria, A., Gagliardi, F., Bruschi, P., & Piotto, M. (2026) · IEEE Access · doi:10.1109/access.2026.3726766
- Authors
- Jiang, D., & Demosthenous, A.
- Affiliations
- University College London (GB)
- Venue
- IEEE Transactions on Biomedical Circuits and Systems
- Published
- 2018-08
- Volume
- 12
- Issue
- 4
- Pages
- 940-953
- ISSN
- 1932-4545 (ISSN-L)
- 1940-9990
- OpenAlex
- W2886606287
- Cited by
- 36
- References
- 36
- Citations by year
- 2026: 3 · 2025: 6 · 2024: 7 · 2023: 3 · 2022: 6 · 2021: 6 · 2020: 3 · 2019: 1 · 2018: 1
- FWCI
- 1.24
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Locations
- IEEE Transactions on Biomedical Circuits and Systems (publishedVersion, cc-by)
- PubMed (publishedVersion)
- UCL Discovery (University College London) (submittedVersion)
- Funders
- Engineering and Physical Sciences Research Council (EP/F009593/1)
- European Commission
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Biomedical Engineering › Analog and Mixed-Signal Circuit Design
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Radio Frequency Integrated Circuit Design
- Keywords
- neural stimulator, multichannel stimulation, microelectrode arrays, neural prostheses, in vivo stimulation
- MeSH terms
- Animals, Electric Stimulation, Electrodes, Implanted, Equipment Design, Humans, Neurons, Implantable Neurostimulators
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review