Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review
Open accessThe DOI matches its Crossref record; 228 of the 228 references it cites were found in a registry or an index; 66 of the 166 sources its claims cite were checked for support.
Abstract
Interfaces with peripheral nerves have been widely developed to enable bioelectronic control of neural activity. Peripheral nerve neuromodulation shows great potential in addressing motor dysfunctions, neurological disorders, and psychiatric conditions. The integration of high-density neural electrodes with stimulation and recording circuits poses a challenge in the design of neural interfaces. Recent advances in active electrode strategies have achieved improved reliability and performance by implementing in-situ control, stimulation, and recording of neural fibers. This paper presents an overview of state-of-the-art neural interface systems that comprise a range of neural electrodes, neurostimulators, and bio-amplifier circuits, with a special focus on interfaces for the peripheral nerves. A discussion on the efficacy of active electrode systems and recommendations for future directions conclude this paper.
Overview
CoChat Summary
TL;DR
This paper is a review of active neural interface circuits and systems for selective peripheral nerve control, covering stimulation, recording, artifact suppression, and integration challenges. It argues that placing electronics close to the electrodes improves selectivity, reliability, signal quality, and supports closed-loop neuromodulation.
Key findings
- Passive neural interfaces rely on remote electronic control via cables, but they face limits in selectivity, scaling, interconnect reliability, signal quality, and device size; active interfaces address these by in-situ signal conditioning and control.
- The paper identifies three active interface architectures: stand-alone, tethered, and distributed; it notes that distributed systems can enable flexible multisite, high-density coverage, but can increase surgical complexity and face micromotion challenges.
- For stimulation, the paper reports that peripheral-nerve interfaces commonly use current amplitudes from tens of μA to a few mA and pulse widths up to a few hundred μs, with high compliance voltage often needed for small, high-impedance electrodes.
- For recording, neural spike signals are described as μV-level with bandwidths of 200 Hz to 10 kHz, typically requiring about 40 dB to 60 dB preamplification and high CMRR; the paper also notes that implantable microelectrode arrays have achieved noise levels below 1 μVrms.
Methods
- Narrative review of published active neural interface systems and circuit techniques, focused on peripheral nerve applications.
- Synthesizes literature on electrode types, stimulation driver architectures, recording front-ends, artifact-reduction methods, fabrication/packaging, and wireless power/communication.
- Compares representative state-of-the-art systems in tables, including process nodes, channel counts, area, power/data protocol, and artifact/charge-balancing performance where available.
Limitations
- The paper does not present a new experimental dataset or head-to-head empirical evaluation; this is a review article.
- It states that selecting a suitable active interface is highly application-dependent, and that miniaturization, high-voltage compliance, crosstalk, biocompatibility, chronic stability, and power constraints remain unresolved challenges.
- Evident from the text: many performance claims are drawn from heterogeneous prior studies with different nerves, devices, and metrics, so direct comparisons across systems are limited.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
50 Valid · 7 Not valid · 62 Not checked · 23 Own results and interpretations
Every source the paper cites for this claim was checked and supports it.
In the paper
“The electrode's degree of invasiveness is an important factor in the neural response [5].”
What the sources say
- Supportedcites [5] [5] X. Navarro, T. B. Krueger, N. Lago, S. Micera, T. Stieglitz, and P. Dario, 'A critical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems,' J. Peripheral Nervous Syst. , vol. 10, no. 3, pp. 229-258, 2005. - source
The Electrical section (p. 2) contrasts electrodes placed adjacent to, around, or within a nerve and explains that placement affects activation thresholds and the risk of nerve damage. The Penetrating microelectrodes section (p. 14) further reports differences in neural conduction and motor recruitment between invasive and extraneural electrodes. Together, these support the claim that invasiveness matters to neural response. (LLM, conf=0.87)
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Interfaces of the peripheral nerves such as the ultrasonic neural dust, StimDust, and the NeuroTalk active Utah electrode array demonstrated good performance in monitoring and activating the sciatic nerve [42], [52], [53].”
What the sources say
- Partly supportedcites [42] [42] D. Seo et al., 'Wireless recording in the peripheral nervous system with ultrasonic neural dust,' Neuron , vol. 91, no. 3, pp. 529-539, 2016. - sourceExperimental — accuracy not yet measured
The Introduction and Results report that ultrasonic neural dust recorded signals from the sciatic nerve in anesthetized rats, and the Discussion says its recording performance was equivalent to conventional wired recordings. The paper does not show neural dust activating the sciatic nerve; stimulation used separate electrodes. The other named devices are associated with other citations in the sentence, not this source. (LLM, conf=0.87)
- Not checked — no full text of the sourcecites [52] [52] D. K. Piech et al., 'A wireless millimetre-scale implantable neural stimulator with ultrasonically powered bidirectional communication,' Nature Biomed. Eng. , vol. 4, no. 2, pp. 207-222, 2020. - source
- Partly supportedcites [53] [53] R. A. Frederick, P. R. Troyk, and S. F. Cogan, 'Wireless microelectrode arrays for selective and chronically stable peripheral nerve stimulation for hindlimb movement,' J. Neural Eng. , vol. 18, no. 5, 2021, Art. no. 056058. - sourceExperimental — accuracy not yet measured
The Methods (Device design and assembly) describe a wireless microelectrode array using the NeuroTalk interface, and the Main results report reliable activation of the rat sciatic nerve to evoke hindlimb movement. The paper does not demonstrate monitoring sciatic-nerve activity with that array; it reports stimulation and observation of movement. The other named interfaces are not evaluated in this paper. (LLM, conf=0.86)
boostedstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Passive neural interfaces have proved as effective bioelectronic therapies in managing neurological and psychiatric disorders, as well as treatments for sensory and motor dysfunctions [1].”
What the sources say
- Partly supportedcites [1] [1] J. A. Frank, M. J. Antonini, and P. Anikeeva, 'Next-generation interfaces for studying neural function,' Nature Biotechnol. , vol. 37, no. 9, pp. 1013-1023, 2019. - sourceExperimental — accuracy not yet measured
The Abstract (page unavailable) says implanted electrical stimulation devices are used to treat neurological disorders, supporting that part of the claim. But the Conclusions and perspective section describes therapies for psychiatric conditions as something future insights will inform, while the Multi-modal interaction section reports restored locomotion only in rodents after spinal cord injury. Our copy does not establish that passive interfaces have proved effective across psychiatric, sensory, and motor dysfunctions. (LLM, conf=0.86)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“For example, non-selective vagus nerve stimulation (VNS) results in coughs, headaches, throat pain, nausea and apnea.”
What the sources say
- Supportedcites [3] [3] A. Fitchett, S. Mastitskaya, and K. Aristovich, 'Selective neuromodulation of the vagus nerve,' Frontiers Neurosci. , vol. 15, no. 5, May 2021, Art. no. 685872. - source
The Introduction identifies cough, headache, throat and tonsil pain, nausea, and apnea among VNS side effects, and says these result primarily from stimulating the whole vagus nerve rather than stimulating it selectively. The section ‘Side Effects of Non-selective VNS’ further discusses throat pain, cough, and acute apnea. (LLM, conf=0.94)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Mechanical modes of neurostimulation enable non-invasive technologies such as ultrasound and magnetic forces by operating on the basis of mechanotransduction to convert mechanical stimuli into electrophysiological signals through compression, expansion, bending and tension [108].”
What the sources say
- Partly supportedcites [108] [108] H. Song et al., 'Neuromodulation of the peripheral nervous system: Bioelectronic technology and prospective developments,' BMEMat , vol. 2, no. 1, Sep. 2023, Art. no. e12048. - sourceExperimental — accuracy not yet measured
Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
63 more claims · 22 with their cited sources checked
Create a free account to see all 142 claims and to verify the 3 references still open.
- It may state two separate claims as one. 46
- Probably not a substantial claim. 37
Citations
Cites
Design of Analog CMOS Integrated Circuits
Razavi, B. (1999) · Cited by 6,609
Neural Stimulation and Recording Electrodes
Cogan, S. F. (2008) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.10.061807.160518 · Cited by 2,243 · cited 1×
“This reduces the occurrence of irreversible electrochemical processes at the tissue such as water electrolysis, pH changes, and electrode dissolution or corrosion [102].”
Electrical stimulation of excitable tissue: design of efficacious and safe protocols
Merrill, D. R., Bikson, M., & Jefferys, J. G. R. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.10.020 · Cited by 2,078 · cited 5×
“Faradaic reactions where products are not available for the reverse reaction are not reversed, for example hydrolysis of water or “water window” which places safety limits on electrode-electrolyte interface potentials [101].”
InfluentialTranscranial Magnetic Stimulation: A Primer
Hallett, M. (2007) · Neuron · doi:10.1016/j.neuron.2007.06.026 · Cited by 1,891 · cited 1×
“Primarily developed as a painless therapy for the brain as transcranial magnetic stimulation, this technique has also been used to diagnose peripheral damage through direct stimulation to the nerves [110].”
Enz, C. C., & Temes, G. C. (1996) · Proceedings of the IEEE · doi:10.1109/5.542410 · Cited by 1,877 · cited 1×
“It results in minimum low-frequency noise, while limiting the signal or noise aliasing and maintaining a zero dc gain [175], [176].”
The development and application of optogenetics
Fenno, L., Yizhar, O., & Deisseroth, K. (2011) · Annual Review of Neuroscience · doi:10.1146/annurev-neuro-061010-113817 · Cited by 1,822 · cited 1×
“The emergence of optogenetic technologies has enabled the genetic encoding of neurons with light-gated ion channels such as Channelrhodopsin-2 (ChR2) and Halorhodopsin (HR), that serve as sensory photoreceptors [106].”
A low-power low-noise cmos for amplifier neural recording applications
Harrison, R. R., & Charles, C. (2003) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2003.811979 · Cited by 1,695 · cited 3×
“Provided that C I , C L (cid:2) C F , the high-pass corner, ω H is approximately g m / ( A M C L ) , where g m , A M , and C L represent the OTA transconductance, the midband gain, and the load capacitance, respectively [168].”
InfluentialDeep brain stimulation: current challenges and future directions
Lozano, A. M., Lipsman, N., Bergman, H., Brown, P., Chabardes, S., Chang, J. W., Matthews, K., McIntyre, C. C., Schlaepfer, T. E., Schulder, M., et al. (2019) · Nature Reviews Neurology · doi:10.1038/s41582-018-0128-2 · Cited by 1,450 · cited 1×
“…to manage neurological and psychiatric disorders such as Parkinson’s disease and essential tremor, Alzheimer’s disease, epilepsy, depression via deep brain stimulation [6], and those associated with the spinal cord to treat chronic neuropathic pain, motor dysfunction, and sensory paralysis [7].”
Neuropixels 2.0: A miniaturized high-density probe for stable, long-term brain recordings.
Steinmetz, N. A., Aydin, C., Lebedeva, A., Okun, M., Pachitariu, M., Bauza, M., Beau, M., Bhagat, J., Böhm, C., Broux, M., et al. (2021) · Science · doi:10.1126/science.abf4588 · Cited by 1,173 · cited 1×
“Neural probes enable high-density interfacing with minimum invasiveness using a narrow, rigid shank, which can accommodate hundreds to thousands of pixels or electrode sites for recording [17], [73] or stimulating [74], [75] neural activity.”
Flexible Electronics: Stretchable Electrodes and Their Future
Huang, S., Liu, Y., Zhao, Y., Ren, Z., & Guo, C. F. (2019) · Advanced Functional Materials · doi:10.1002/adfm.201805924 · Cited by 855 · cited 2×
“Biomechanical energy harvesting systems such as triboelectric and piezoelectric generators harness the electrical field resulting from charge separation and mechanical strain, respectively, enabling in-vivo powering from the natural movement of the organs [208], [216].”
Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P. (2005) · Journal of the Peripheral Nervous System · doi:10.1111/j.1085-9489.2005.10303.x · Cited by 845 · cited 1×
“The electrode’s degree of invasiveness is an important factor in the neural response [5].”
Angeli, C. A., Edgerton, V. R., Gerasimenko, Y. P., & Harkema, S. J. (2014) · Brain · doi:10.1093/brain/awu038 · Cited by 750 · cited 1×
“in detection and modulation of neural activity; this is critical for restoring sensory and motor functions, e.g., for amputees and patients with spinal cord injuries [2], and for mitigating the side effects of non-selective stimulation.”
Materials and technologies for soft implantable neuroprostheses
Lacour, S. P., Courtine, G., & Guck, J. (2016) · Nature Reviews Materials · doi:10.1038/natrevmats.2016.63 · Cited by 738 · cited 1×
“[204], [205].”
A micropower low-noise monolithic instrumentation amplifier for medical purposes
Steyaert, M. S. J., & Sansen, W. M. C. (1987) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.1987.1052869 · Cited by 577 · cited 1×
“The noise efficiency factor (NEF) is commonly used as a figure of merit [165].”
Analog Integrated Circuit Design
(2011) · Selected Semiconductor Research · doi:10.1142/9781848164079_0004 · Cited by 568
A review of vagus nerve stimulation as a therapeutic intervention
Johnson, R. L., & Wilson, C. G. (2018) · Journal of Inflammation Research · doi:10.2147/jir.s163248 · Cited by 562 · cited 1×
“…of the autonomic nerves, which regulates non-volitional bodily processes such as those of the cardiovascular, gastrointestinal, and immune systems, has shown therapeutic benefits in diseases including heart failure, lung injury, inflammation, rheumatoid arthritis, obesity, and diabetes [11], [12].”
Wireless Recording in the Peripheral Nervous System with Ultrasonic Neural Dust
Seo, D., Neely, R. M., Shen, K., Singhal, U., Alon, E., Rabaey, J. M., Carmena, J. M., & Maharbiz, M. M. (2016) · Neuron · doi:10.1016/j.neuron.2016.06.034 · Cited by 557 · cited 2×
“Interfaces of the peripheral nerves such as the ultrasonic neural dust, StimDust, and the NeuroTalk active Utah electrode array demonstrated good performance in monitoring and activating the sciatic nerve [42], [52], [53].”
A model of safe levels for electrical stimulation
Shannon, R. V. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.126616 · Cited by 526 · cited 1×
“Safety limits must also consider stimulation charge density [117], [118], in particular where neural interfaces are miniaturised, leading to small electrode surface areas.”
A wireless closed-loop system for optogenetic peripheral neuromodulation
Mickle, A. D., Won, S. M., Noh, K. N., Yoon, J., Meacham, K. W., Xue, Y., McIlvried, L. A., Copits, B. A., Samineni, V. K., Crawford, K. E., et al. (2019) · Nature · doi:10.1038/s41586-018-0823-6 · Cited by 514
(2006) · The International Series in Engineering and Computer Science · doi:10.1007/b135984 · Cited by 511
Materials for flexible bioelectronic systems as chronic neural interfaces
Song, E., Li, J., Won, S. M., Bai, W., & Rogers, J. A. (2020) · Nature Materials · doi:10.1038/s41563-020-0679-7 · Cited by 498 · cited 1×
“The measurements can suffer from a low signal-to-noise ratio (SNR), limiting the efficacy of feedback [4].”
Denison, T., Consoer, K., Santa, W., Avestruz, A.-T., Cooley, J., & Kelly, A. (2007) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2007.908664 · Cited by 451
Piech, D. K., Johnson, B. C., Shen, K., Ghanbari, M. M., Li, K. Y., Neely, R. M., Kay, J. E., Carmena, J. M., Maharbiz, M. M., & Muller, R. (2020) · Nature Biomedical Engineering · doi:10.1038/s41551-020-0518-9 · Cited by 442 · cited 2×
“Interfaces of the peripheral nerves such as the ultrasonic neural dust, StimDust, and the NeuroTalk active Utah electrode array demonstrated good performance in monitoring and activating the sciatic nerve [42], [52], [53].”
Shin, G., Gomez, A. M., Al-Hasani, R., Jeong, Y. R., Kim, J., Xie, Z., Banks, A., Lee, S. M., Han, S. Y., Yoo, C. J., et al. (2017) · Neuron · doi:10.1016/j.neuron.2016.12.031 · Cited by 396 · cited 1×
“Standalone: ( g ) neural dust [42] and ( h ) free-floating wireless optical stimulator [43].”
A Materials Roadmap to Functional Neural Interface Design
Wellman, S. M., Eles, J. R., Ludwig, K. A., Seymour, J. P., Michelson, N. J., McFadden, W. E., Vazquez, A. L., & Kozai, T. D. Y. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201701269 · Cited by 389 · cited 1×
“Real-time nerve response to stimuli is collected to detect electrode micromotion and change the stimulation site, if necessary [202].”
Tissue damage thresholds during therapeutic electrical stimulation
Cogan, S. F., Ludwig, K. A., Welle, C. G., & Takmakov, P. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/2/021001 · Cited by 386 · cited 1×
“Safety limits must also consider stimulation charge density [117], [118], in particular where neural interfaces are miniaturised, leading to small electrode surface areas.”
Spinal Cord Stimulation: Clinical Efficacy and Potential Mechanisms
Sdrulla, A. D., Guan, Y., & Raja, S. N. (2018) · Pain Practice · doi:10.1111/papr.12692 · Cited by 375
Anti-inflammatory effect of vagus nerve stimulation in a rat model of inflammatory bowel disease
Meregnani, J., Clarençon, D., Vivier, M., Peinnequin, A., Mouret, C., Sinniger, V., Picq, C., Job, A., Canini, F., Jacquier-Sarlin, M., et al. (2011) · Autonomic Neuroscience · doi:10.1016/j.autneu.2010.10.007 · Cited by 310
Ultra Low Power Bioelectronics: Fundamentals, Biomedical Applications, and Bio-Inspired Systems
Sarpeshkar, R. (2010) · Cambridge University Press eBooks · doi:10.1017/cbo9780511841446 · Cited by 288
An Integrated 256-Channel Epiretinal Prosthesis
Chen, K., Yang, Z., Hoang, L., Weiland, J., Humayun, M., & Liu, W. (2010) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2010.2055371 · Cited by 266 · cited 2×
“The integration of variable-gain current mirrors in the output path provides flexibility in the range of output currents, with low-power digital control to minimize the area used [140], [141].”
BION ™ system for distributed neural prosthetic interfaces
Loeb, G. E., Peck, R. A., Moore, W. H., & Hood, K. (2001) · Medical Engineering & Physics · doi:10.1016/s1350-4533(01)00011-x · Cited by 243 · cited 1×
“Early developments in wireless stimulators include the cylindrical, injectable Bionic Neuron (BION) electrode [48], demonstrating successful application in the peripheral nerves and muscles using hemispherical, ring or disc electrodes.”
Next-generation interfaces for studying neural function
Frank, J. A., Antonini, M.-J., & Anikeeva, P. (2019) · Nature Biotechnology · doi:10.1038/s41587-019-0198-8 · Cited by 241
Chandrakumar, H., & Markovic, D. (2017) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2016.2645611 · Cited by 237 · cited 1×
“Chopping significantly reduces the amplifier’s input impedance due to the combination of the switched capacitor C I and the passive mixer at the input [177].”
Zhang, M., Tang, Z., Liu, X., & Van der Spiegel, J. (2020) · Nature Electronics · doi:10.1038/s41928-020-0390-3 · Cited by 231 · cited 2×
“To address lower stimulation efficiency, switched-capacitor stimulating topologies have been developed to maintain high efficiency with adequate control over the charge; however, the circuit implementation requires large capacitors, which may exceed the available area of an implantable device [8].”
A wireless millimetric magnetoelectric implant for the endovascular stimulation of peripheral nerves
Chen, J. C., Kan, P., Yu, Z., Alrashdan, F., Garcia, R., Singer, A., Lai, C. S. E., Avants, B., Crosby, S., Li, Z., et al. (2022) · Nature Biomedical Engineering · doi:10.1038/s41551-022-00873-7 · Cited by 231
A mm-Sized Wireless Implantable Device for Electrical Stimulation of Peripheral Nerves
Charthad, J., Chang, T. C., Liu, Z., Sawaby, A., Weber, M. J., Baker, S., Gore, F., Felt, S. A., & Arbabian, A. (2018) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2018.2799623 · Cited by 230 · cited 7×
“Each electrode operates as either a cathode or an anode from a single supply current, maximising flexibility and limiting the reliance on storage capacitors [57], [134].”
InfluentialBadia, J., Boretius, T., Andreu, D., Azevedo-Coste, C., Stieglitz, T., & Navarro, X. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/3/036023 · Cited by 227 · cited 1×
“…is made above the outer layer of the nerve (epineurium), (2) intra-fascicular or intra-neural, which tap into the nerve fascicles with penetrative structures [20], [21], and (3) regenerative, where the interface architecture requires axonal growth through fine-scale device geometries [22], [23].”
Lago, N., Yoshida, K., Koch, K. P., & Navarro, X. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2006.886617 · Cited by 227 · cited 2×
“Intraneural: ( c ) longitudinal (LIFE) [25], transverse (TIME) [26], ( d ) self-opening (SELINE) [28], and Utah slanted (USEA) [27] electrode arrays.”
Noorsal, E., Sooksood, K., Xu, H., Hornig, R., Becker, J., & Ortmanns, M. (2012) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2011.2164667 · Cited by 212 · cited 4×
“They check for available headroom with respect to the interface impedance [44], [129], and detect excess power or leakage of fluids via temperature and humidity sensing [67], [80].”
InfluentialRecent Advances in Materials, Devices, and Systems for Neural Interfaces
Won, S. M., Song, E., Zhao, J., Li, J., Rivnay, J., & Rogers, J. A. (2018) · Advanced Materials · doi:10.1002/adma.201800534 · Cited by 206 · cited 2×
“To maintain a balance between structural conformity and strength, material enhancements have been proposed by combining several layers of parylene C, or by integrating the surface with poly-imide to form the electrode substrate or flexible interconnect to external systems [207].”
A review on mechanical considerations for chronically-implanted neural probes
Lecomte, A., Descamps, E., & Bergaud, C. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa8b4f · Cited by 205 · cited 1×
“This improves longevity by addressing interface compromising factors such as the foreign body reaction, which can deteriorate the neural interface and isolate the nerve fibers through scarring [12], [206].”
A Neural Probe With Up to 966 Electrodes and Up to 384 Configurable Channels in 0.13 $\mu$m SOI CMOS
Mora Lopez, C., Putzeys, J., Raducanu, B. C., Ballini, M., Wang, S., Andrei, A., Rochus, V., Vandebriel, R., Severi, S., Van Hoof, C., et al. (2017) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2016.2646901 · Cited by 201 · cited 2×
“Neural probes enable high-density interfacing with minimum invasiveness using a narrow, rigid shank, which can accommodate hundreds to thousands of pixels or electrode sites for recording [17], [73] or stimulating [74], [75] neural activity.”
Safety of long-term electrical peripheral nerve stimulation: review of the state of the art
Günter, C., Delbeke, J., & Ortiz-Catalan, M. (2019) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-018-0474-8 · Cited by 198 · cited 2×
“Activation of motor regions is another use of peripheral nerve interfaces to manage respiratory ventilation and urinary incontinence, correct foot drop, and restore or achieve control of prosthetic limbs with tactile feedback [8], [9], [10].”
Neural recording and stimulation using wireless networks of microimplants
Lee, J., Leung, V., Lee, A.-H., Huang, J., Asbeck, P., Mercier, P. P., Shellhammer, S., Larson, L., Laiwalla, F., & Nurmikko, A. (2021) · Nature Electronics · doi:10.1038/s41928-021-00631-8 · Cited by 195 · cited 1×
“…system can include a collection of independent ‘motes’ of a wireless floating microelectrode array (WFMA), such as the neurograin interface [81], [82], the mi-crobead stimulating system [83], and the intraspinal microstimulation (ISMS) network [47], [84], [85], as well as the injectable neural…”
Burton, C., Sajja, A., & Latthe, P. M. (2012) · Neurourology and Urodynamics · doi:10.1002/nau.22251 · Cited by 186 · cited 1×
“Technological advancements in percutaneous tibial stimulation for managing lower urinary tract symptoms and chronic pain as an alternative therapy to sacral neuromodulation have led to the development and commercialisation of stand-alone tibial nerve stimulators in minimally invasive packages [58].”
Weiland, J. D., & Humayun, M. S. (2008) · Proceedings of the IEEE · doi:10.1109/jproc.2008.922589 · Cited by 179 · cited 1×
“For instance, retinal prostheses require a high-density stimulator implanted on the retina, visual cortex, and optic nerve for precise mapping of the image pixels collected from an external camera [65].”
Electrical stimulation enhances sensory recovery: A randomized controlled trial
Wong, J. N., Olson, J. L., Morhart, M. J., & Chan, K. M. (2015) · Annals of Neurology · doi:10.1002/ana.24397 · Cited by 177 · cited 1×
“3(a)) in restoring sensory perception and motor functions has been shown in numerous studies [103], [104].”
A Low-Power Integrated Bioamplifier With Active Low-Frequency Suppression
Gosselin, B., Sawan, M., & Chapman, C. A. (2007) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2007.914490 · Cited by 174 · cited 1×
“This may increase power consumption, particularly when improving the noise level [173].”
Biointegrated and Wirelessly Powered Implantable Brain Devices: A Review
Das, R., Moradi, F., & Heidari, H. (2020) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2020.2966920 · Cited by 172 · cited 1×
“They can also directly contact the white matter with improved resolution using penetrative neural probes [17], Utah electrode arrays [18], and flexible mesh electronics [19].”
Neuroprosthetics theory and practice
Horch, K., & Dhillon, G. (2004) · doi:10.1142/10368 · Cited by 172
Borton, D., Micera, S., Millán, J. D. R., & Courtine, G. (2013) · Science Translational Medicine · doi:10.1126/scitranslmed.3005968 · Cited by 171 · cited 1×
“Patient-specific methods of addressing the peripheral nerves require close inspection of neural activity to optimise the position of electrodes [12], the stimulation algorithm, and analysis methods of neural activity according to the individual needs of patients [218].”
An active approach for charge balancing in functional electrical stimulation.
Sooksood, K., Stieglitz, T., & Ortmanns, M. (2010) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2010.2040277 · Cited by 170 · cited 2×
“…by continuous adjustment of the source/sink currents with dynamic current copying in push-pull architectures, [44], [66], and by insertion of a short pulse train that instantaneously corrects the residual charge, or via [137] compensation of the excess charge as an offset current [135], [144].”
Basic mechanisms of peripheral nerve injury and treatment via electrical stimulation
Chu, X.-L., Song, X.-Z., Li, Q., Li, Y.-R., He, F., Gu, X.-S., & Ming, D. (2022) · Neural Regeneration Research · doi:10.4103/1673-5374.335823 · Cited by 164 · cited 1×
“However, peripheral nerves have the [97] potential to self-repair and regenerate, aided by conditions such as anatomical continuity of the injured site, myelination, and the extent of the injury [112].”
Lee, H.-M., Kwon, K. Y., Li, W., & Ghovanloo, M. (2015) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2014.2355814 · Cited by 161
Sivaprakasam, M., Liu, W., Humayun, M. S., & Weiland, J. D. (2005) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2005.843630 · Cited by 155 · cited 2×
“The integration of variable-gain current mirrors in the output path provides flexibility in the range of output currents, with low-power digital control to minimize the area used [140], [141].”
Sit, J.-J., & Sarpeshkar, R. (2007) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2007.911631 · Cited by 155 · cited 1×
“For cochlear implants this has been defined at 15 nC/phase or 100 nA dc error for a 0.01% phase mismatch [126]; many designs strive for precise charge matching using passive and active circuit topologies.”
InfluentialA low-power, low-noise CMOS amplifier for neural recording applications
Harrison, R. R. (2003) · 2002 IEEE International Symposium on Circuits and Systems. Proceedings (Cat. No.02CH37353) · doi:10.1109/iscas.2002.1010674 · Cited by 155
Shin, U., Ding, C., Zhu, B., Vyza, Y., Trouillet, A., Revol, E. C. M., Lacour, S. P., & Shoaran, M. (2022) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2022.3204508 · Cited by 153 · cited 1×
“This can be achieved via the integration of the recording ADC with the feedback path [197] addressing multiple electrodes through a time-multiplexed con-trol structure [198], [199].”
A Modular 32-site wireless neural stimulation microsystem
Ghovanloo, M., & Najafi, K. (2004) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2004.837026 · Cited by 152 · cited 1×
“This may comprise a number of modules connected to a single, implanted wireless transmitter through wired connections, such as the scalable Interstim-2B microstimulator [79], the RDM2 network of epidural stimulators, and the Active Book electrodes [46], [80].”
Principles of electrical stimulation of neural tissue
(2013) · Handbook of Clinical Neurology · doi:10.1016/b978-0-444-53497-2.00001-2 · Cited by 151 · cited 2×
“Charge injection at an electrode occurs through capacitive charge transfer, with no electron transfer across the electrode-electrolyte interface, and Faradaic charge transfer, where electron transfer and reduction and oxidation (redox) reactions occur [101], [114].”
Luan, L., Robinson, J. T., Aazhang, B., Chi, T., Yang, K., Li, X., Rathore, H., Singer, A., Yellapantula, S., Fan, Y., et al. (2020) · Neuron · doi:10.1016/j.neuron.2020.10.011 · Cited by 150 · cited 2×
“[204], [205].”
Neuropixels 2.0: A miniaturized high-density probe for stable, long-term brain recordingsPreprint
Steinmetz, N. A., Aydın, Ç., Lebedeva, A., et al. (2020) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2020.10.27.358291 · Cited by 150
Corrosion of silicon integrated circuits and lifetime predictions in implantable electronic devices
Vanhoestenberghe, A., & Donaldson, N. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/3/031002 · Cited by 140 · cited 1×
“Implant packaging methods to provide a biocompatiblity and corrosion protection through hermetic packages and coatings and non-hermetic encapsulants aim to offer device lifetimes of years to decades [213], [214], [215].”
Recent Advances in Neural Recording Microsystems
Gosselin, B. (2011) · Sensors · doi:10.3390/s110504572 · Cited by 138 · cited 3×
“…of neural recording analog front-ends implemented using ( a ) an ac-coupled capacitive feedback network [155], ( b ) a dc-coupled amplifier with active low-frequency suppression [159], ( c ) a dc-coupled fully differential difference amplifier [170], and ( d ) a chopper-stabilised architecture [171].”
Ghovanloo, M., & Najafi, K. (2005) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2004.839797 · Cited by 137 · cited 1×
“This consumes large amounts of power and may result in redundant voltage overhead on the electrodes, as their impedance may vary over time and position by several orders of magnitude [78], [128].”
A Sub-mm3 Ultrasonic Free-Floating Implant for Multi-Mote Neural Recording
Ghanbari, M. M., Piech, D. K., Shen, K., Faraji Alamouti, S., Yalcin, C., Johnson, B. C., Carmena, J. M., Maharbiz, M. M., & Muller, R. (2019) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2019.2936303 · Cited by 137 · cited 1×
“Current technologies aim for sub-mm-scale devices to enable a minimally invasive implantation procedure and reduce long-term tissue scarring, and wireless power and communication capabilities of free-floating, multi-mote systems to enhance deep tissue accessibility [94], [98].”
Arfin, S. K., & Sarpeshkar, R. (2012) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2166072 · Cited by 133 · cited 1×
“Dynamic power supply regulators have been implemented using high-efficiency dc-dc converters, such as the forward-buck / reverse-boost [131] and switched-capacitor [132] topologies to modulate the stimulator supply based on the electrode voltage measurements.”
A review for the peripheral nerve interface designer
Larson, C. E., & Meng, E. (2020) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108523 · Cited by 133 · cited 1×
“These are largely dictated by the device geometry and material choice, and determine stimulation efficiency and noise levels, but are restricted by the implant application [23], [36].”
Geddes, L. A., & Bourland, J. D. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325456 · Cited by 128 · cited 1×
“The relationship between charge and neuron activation is defined by the strength-duration curve, the threshold current amplitude required for neuron stimulation for a given pulse width [115], [116], thus sufficient charge injection is required for neuron activation.”
Miniature battery-free bioelectronics
Nair, V., Dalrymple, A. N., Yu, Z., Balakrishnan, G., Bettinger, C. J., Weber, D. J., Yang, K., & Robinson, J. T. (2023) · Science · doi:10.1126/science.abn4732 · Cited by 128 · cited 1×
“Biomechanical energy harvesting systems such as triboelectric and piezoelectric generators harness the electrical field resulting from charge separation and mechanical strain, respectively, enabling in-vivo powering from the natural movement of the organs [208], [216].”
Robust Wireless Power Transmission to mm-Sized Free-Floating Distributed Implants
Mirbozorgi, S. A., Yeon, P., & Ghovanloo, M. (2017) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2017.2663358 · Cited by 127
A Microchannel Neuroprosthesis for Bladder Control After Spinal Cord Injury in Rat
Chew, D. J., Zhu, L., Delivopoulos, E., Minev, I. R., Musick, K. M., Mosse, C. A., Craggs, M., Donaldson, N., Lacour, S. P., McMahon, S. B., et al. (2013) · Science Translational Medicine · doi:10.1126/scitranslmed.3007186 · Cited by 123 · cited 1×
“Activation of motor regions is another use of peripheral nerve interfaces to manage respiratory ventilation and urinary incontinence, correct foot drop, and restore or achieve control of prosthetic limbs with tactile feedback [8], [9], [10].”
Non-invasive peripheral nerve stimulation via focused ultrasound in vivo
Downs, M. E., Lee, S. A., Yang, G., Kim, S., Wang, Q., & Konofagou, E. E. (2018) · Physics in Medicine & Biology · doi:10.1088/1361-6560/aa9fc2 · Cited by 121 · cited 1×
“Targeted therapies using focused ultrasound energy have been developed to elicit or inhibit specific neural activity in deep tissues [111].”
Johnson, B. C., Gambini, S., Izyumin, I., Moin, A., Zhou, A., Alexandrov, G., Santacruz, S. R., Rabaey, J. M., Carmena, J. M., & Muller, R. (2017) · 2017 Symposium on VLSI Circuits · doi:10.23919/vlsic.2017.8008543 · Cited by 119 · cited 2×
“Techniques such as adaptive filtering methods rely on the least-mean-squares (LMS) algorithm to implement infinite impulse response (IIR) and finite impulse response (FIR) digital filters to address and cancel out individual samples of artifacts [132], [194].”
Thin films and microelectrode arrays for neuroprosthetics
Ordonez, J., Schuettler, M., Boehler, C., Boretius, T., & Stieglitz, T. (2012) · MRS Bulletin · doi:10.1557/mrs.2012.117 · Cited by 118 · cited 1×
“Extra-neural interfaces such as cylindrical nerve cuffs and epi-neural electrodes contact large groups of axons with limited selectivity but preserve tissue integrity and cause minimum nerve injury [24].”
An Integrated Implantable Stimulator That is Fail-Safe Without Off-Chip Blocking-Capacitors
Liu, X., Demosthenous, A., & Donaldson, N. (2008) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2008.2003199 · Cited by 118 · cited 1×
“In addition to the implementation of thick-oxide devices to achieve a fully isolated active output [140], [150], dynamic bulk biasing techniques have been studied to enable a higher stimulator compliance voltage [151] that is not limited by the diode breakdown level with the aid of charge pumps…”
Self-citationCochlear-implant spatial selectivity with monopolar, bipolar and tripolar stimulation
Zhu, Z., Tang, Q., Zeng, F.-G., Guan, T., & Ye, D. (2012) · Hearing Research · doi:10.1016/j.heares.2011.11.005 · Cited by 115 · cited 2×
“Electrode configuration may be used to focus extracellular fields and improve stimulation selectivity and efficacy [121], [122].”
A Low-Noise Chopper Amplifier Designed for Multi-Channel Neural Signal Acquisition
Luo, D., Zhang, M., & Wang, Z. (2019) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2019.2913101 · Cited by 114 · cited 2×
“…of neural recording analog front-ends implemented using ( a ) an ac-coupled capacitive feedback network [155], ( b ) a dc-coupled amplifier with active low-frequency suppression [159], ( c ) a dc-coupled fully differential difference amplifier [170], and ( d ) a chopper-stabilised architecture [171].”
High-Value Tunable Pseudo-Resistors Design
Guglielmi, E., Toso, F., Zanetto, F., Sciortino, G., Mesri, A., Sampietro, M., & Ferrari, G. (2020) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2020.2973639 · Cited by 114 · cited 1×
“Multi-cell structures have shown improvements in the linear range of the pseudo-resistor [187].”
Lee, C., Jeon, T., Jang, M., Park, S., Kim, J., Lim, J., Ahn, J., Huh, Y., & Chae, Y. (2020) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2020.3018478 · Cited by 114 · cited 2×
“(2) Front-end design techniques that improve artifact immunity [190], [191] via the amplifier’s dynamic range, frequency response, and dishcharge mechanisms.”
Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review
Yildiz, K. A., Shin, A. Y., & Kaufman, K. R. (2020) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-020-00667-5 · Cited by 113 · cited 1×
“Chronic studies of intrafascicular and regenerative electrodes have been carried out in vivo to assess the long-term reliability of peripheral nerve interfaces implemented in prosthetic applications [212].”
320-Channel Active Probe for High-Resolution Neuromonitoring and Responsive Neurostimulation
Shulyzki, R., Abdelhalim, K., Bagheri, A., Salam, M. T., Florez, C. M., Velazquez, J. L. P., Carlen, P. L., & Genov, R. (2015) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2014.2312552 · Cited by 106 · cited 1×
“…[129], ( c ) dynamic current copying in a source-sink structure [66], ( d ) high-frequency switched capacitor configuration [142]. have been made to combine the regulated cascode topology with a current-copying stimulator in [133], which minimises the required power and area and reduces mismatch.”
Optical Stimulation of Neurons
C. Thompson, A., R. Stoddart, P., & Jansen, E. D. (2015) · Current Molecular Imaging · doi:10.2174/2211555203666141117220611 · Cited by 105 · cited 1×
“ChR2 responds to blue (470 nm) lights to depolarise a neuron, and HR exhibits sensitivity toward green light (578 nm) that hyperpolarises the cell, thus inhibiting the propagation of action potentials [107].”
Kim, C., Joshi, S., Courellis, H., Wang, J., Miller, C., & Cauwenberghs, G. (2018) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2018.2870555 · Cited by 104
22.7 A 966-electrode neural probe with 384 configurable channels in 0.13µm SOI CMOS
López, C. M., Mitra, S., Putzeys, J., Raducanu, B., Ballini, M., Andrei, A., Severi, S., Welkenhuysen, M., Hoof, C. A. V., Musa, S., et al. (2016) · Digest of technical papers/Digest of technical papers - IEEE International Solid-State Circuits Conference · doi:10.1109/isscc.2016.7418072 · Cited by 104
Analysis and Design of Tunable Amplifiers for Implantable Neural Recording Applications
Rezaee-Dehsorkh, H., Ravanshad, N., Lotfi, R., Mafinezhad, K., & Sodagar, A. M. (2011) · IEEE Journal on Emerging and Selected Topics in Circuits and Systems · doi:10.1109/jetcas.2011.2174492 · Cited by 103 · cited 1×
“These can be divided into non-tunable (constant) and tunable topologies, which can themselves be classified as voltage-or current-controlled structures [186].”
A 1-V 0.25- $\mu \text{W}$ Inverter Stacking Amplifier With 1.07 Noise Efficiency Factor
Shen, L., Lu, N., & Sun, N. (2018) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2017.2786724 · Cited by 101
Anderson, D. N., Duffley, G., Vorwerk, J., Dorval, A. D., & Butson, C. R. (2019) · Journal of Neural Engineering · doi:10.1088/1741-2552/aae590 · Cited by 100 · cited 1×
“However, orientation selective stimulation has been simulated with anodic first pulses [120].”
Toward Bioelectronic Medicine—Neuromodulation of Small Peripheral Nerves Using Flexible Neural Clip
Lee, S., Peh, W. Y. X., Wang, J., Yang, F., Ho, J. S., Thakor, N. V., Yen, S., & Lee, C. (2017) · Advanced Science · doi:10.1002/advs.201700149 · Cited by 96 · cited 2×
“…is made above the outer layer of the nerve (epineurium), (2) intra-fascicular or intra-neural, which tap into the nerve fascicles with penetrative structures [20], [21], and (3) regenerative, where the interface architecture requires axonal growth through fine-scale device geometries [22], [23].”
Selective Neuromodulation of the Vagus Nerve
Fitchett, A., Mastitskaya, S., & Aristovich, K. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.685872 · Cited by 94 · cited 2×
“However, state-of-the-art neuromodulation systems are approaching more selective solutions through novel neural interface designs [3].”
A Complete 256-Electrode Retinal Prosthesis Chip
Tran, N., Bai, S., Yang, J., Chun, H., Kavehei, O., Yang, Y., Muktamath, V., Ng, D., Meffin, H., Halpern, M., et al. (2014) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2014.2298037 · Cited by 90 · cited 7×
“Tethered: ( i ) retinal prosthesis [44] and ( j ) neural probe [45].”
InfluentialThe BION devices: injectable interfaces with peripheral nerves and muscles
Loeb, G. E., Richmond, F. J. R., & Baker, L. L. (2006) · Neurosurgical Focus · doi:10.3171/foc.2006.20.5.3 · Cited by 88 · cited 1×
“…interface [81], [82], the mi-crobead stimulating system [83], and the intraspinal microstimulation (ISMS) network [47], [84], [85], as well as the injectable neural stimulator (InNS) technologies, which enable interfacing with regions that are inaccessible for implantation [86], [87], [88].”
Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review.
Ghafoor, U., Kim, S., & Hong, K.-S. (2017) · Frontiers in Neurorobotics · doi:10.3389/fnbot.2017.00059 · Cited by 88 · cited 1×
“…between neural interface selectivity and invasiveness, the impact of crosstalk on the stimulation efficacy, and the presence of stimulus artifacts on the recording capability of bidirectional systems, requiring real-time feedback to optimise the stimulation method, site, and intensity [105].”
An Inductively-Powered Wireless Neural Recording and Stimulation System for Freely-Behaving Animals
Lee, B., Jia, Y., Mirbozorgi, S. A., Connolly, M., Tong, X., Zeng, Z., Mahmoudi, B., & Ghovanloo, M. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2891303 · Cited by 88
A Fully Intraocular High-Density Self-Calibrating Epiretinal Prosthesis
Monge, M., Raj, M., Nazari, M. H., Chang, H.-C., Zhao, Y., Weiland, J. D., Humayun, M. S., Tai, Y.-C., & Emami, A. (2013) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2014.2298334 · Cited by 85 · cited 3×
“Another shortcoming of all passive solutions is the lack of control over the success of the discharge method [135], [145].”
Implantable neurotechnologies: a review of integrated circuit neural amplifiers
Ng, K. A., Greenwald, E., Xu, Y. P., & Thakor, N. V. (2016) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-015-1431-3 · Cited by 83 · cited 4×
“…and narrow bandwidths (200 Hz to 10 kHz) demand high-gain pre-amplification (typically of about 40 dB to 60 dB) and filtering prior to digitisation and processing [156], and are recorded in the presence of large common mode signals requiring high common-mode rejection ratio (CMRR) [157], [158].”
InfluentialA three-dimensional self-opening intraneural peripheral interface (SELINE)
Cutrone, A., Valle, J. D., Santos, D., Badia, J., Filippeschi, C., Micera, S., Navarro, X., & Bossi, S. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/1/016016 · Cited by 82 · cited 1×
“Implantation in a rat sciatic nerve showed good selectivity and enhanced stability, despite an encapsulating layer caused by fibrotic scarring of tissue [29].”
Uehlin, J. P., Smith, W. A., Pamula, V. R., Pepin, E. P., Perlmutter, S., Sathe, V., & Rudell, J. C. (2020) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2020.2991524 · Cited by 81 · cited 2×
“…addition to the implementation of thick-oxide devices to achieve a fully isolated active output [140], [150], dynamic bulk biasing techniques have been studied to enable a higher stimulator compliance voltage [151] that is not limited by the diode breakdown level with the aid of charge pumps [152].”
Electrical stimulation for the treatment of lower urinary tract dysfunction after spinal cord injury
McGee, M. J., Amundsen, C. L., & Grill, W. M. (2015) · The Journal of Spinal Cord Medicine · doi:10.1179/2045772314y.0000000299 · Cited by 80 · cited 1×
“Similarly, a treatment for urinary tract dysfunction or overactive bladder involves neuromodulation of the sacral nerve controlled by a pulse generator implanted on the abdominal or buttock region [15], [16].”
Retinal Neurostimulator for a Multifocal Vision Prosthesis
Wong, Y. T., Dommel, N., Preston, P., Hallum, L. E., Lehmann, T., Lovell, N. H., & Suaning, G. J. (2007) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2007.903958 · Cited by 79 · cited 1×
“Epi-retinal interfaces have incorporated a hexagonal mosaic of guard electrodes to reduce crosstalk on a stimulator IC using a push-pull topology [149].”
Electrical Stimulation and Motor Recovery
Young, W. (2015) · Cell Transplantation · doi:10.3727/096368915x686904 · Cited by 76 · cited 1×
“3(a)) in restoring sensory perception and motor functions has been shown in numerous studies [103], [104].”
The Microbead: A Highly Miniaturized Wirelessly Powered Implantable Neural Stimulating System
Khalifa, A., Karimi, Y., Wang, Q., Montlouis, W., Garikapati, S., Stanacevic, M., Thakor, N., & Etienne-Cummings, R. (2018) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2018.2802443 · Cited by 76 · cited 1×
“…‘motes’ of a wireless floating microelectrode array (WFMA), such as the neurograin interface [81], [82], the mi-crobead stimulating system [83], and the intraspinal microstimulation (ISMS) network [47], [84], [85], as well as the injectable neural stimulator (InNS) technologies, which…”
Long Micro-Channel Electrode Arrays: A Novel Type of Regenerative Peripheral Nerve Interface
Lacour, S. P., Fitzgerald, J. J., Lago, N., Tarte, E., McMahon, S., & Fawcett, J. (2009) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2009.2031241 · Cited by 75 · cited 2×
“Regenerative: ( e ) sieve [30], microchannel [34], and ( f ) scaffold [32] electrodes.”
Track-and-Zoom Neural Analog-to-Digital Converter With Blind Stimulation Artifact Rejection
Reza Pazhouhandeh, M., Chang, M., Valiante, T. A., & Genov, R. (2020) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2020.2991526 · Cited by 74
FitzGerald, J. J., Lago, N., Benmerah, S., Serra, J., Watling, C. P., Cameron, R. E., Tarte, E., Lacour, S. P., McMahon, S. B., & Fawcett, J. W. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/1/016010 · Cited by 73
Butz, N., Taschwer, A., Nessler, S., Manoli, Y., & Kuhl, M. (2018) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2018.2828823 · Cited by 70 · cited 2×
“To overcome this, pulse-width-adaptive charge balancing has been developed as a stand-alone solution or to operate in parallel with offset compensation [147], [148].”
Neural interface systems with on-device computing: machine learning and neuromorphic architectures
Yoo, J., & Shoaran, M. (2021) · Current Opinion in Biotechnology · doi:10.1016/j.copbio.2021.10.012 · Cited by 70 · cited 1×
“Power-efficient implementation of online classification engines in situ aims to reduce the burden of data telemetry and respond to symptoms in a timely manner [219].”
Neuroprosthetics: Theory and Practice
Horch, K. W., & Dhillon, G. S. (2004) · Series on Bioengineering and Biomedical Engineering · doi:10.1142/4987 · Cited by 70
A distributed architecture for activating the peripheral nervous system
Andreu, D., Guiraud, D., & Souquet, G. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/2/026001 · Cited by 69
A mm-Sized Free-Floating Wirelessly Powered Implantable Optical Stimulation Device
Jia, Y., Mirbozorgi, S. A., Lee, B., Khan, W., Madi, F., Inan, O. T., Weber, A., Li, W., & Ghovanloo, M. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2918761 · Cited by 68 · cited 1×
“Examples include the free-floating, wirelessly-powered, implantable optical stimulation device and the free-floating, wireless implantable neural recording system-on-chip [51].”
Regenerative Electrode Interfaces for Neural Prostheses.
Thompson, C. H., Zoratti, M. J., Langhals, N. B., & Purcell, E. K. (2016) · Tissue Engineering Part B: Reviews · doi:10.1089/ten.teb.2015.0279 · Cited by 67 · cited 3×
“…is made above the outer layer of the nerve (epineurium), (2) intra-fascicular or intra-neural, which tap into the nerve fascicles with penetrative structures [20], [21], and (3) regenerative, where the interface architecture requires axonal growth through fine-scale device geometries [22], [23].”
Zhu, B., Shin, U., & Shoaran, M. (2021) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2021.3112756 · Cited by 67 · cited 1×
“An important challenge in real time neural decoding involves balancing the accuracy-latency trade-off, which requires careful consideration according to the system application [201].”
Multi-Channel Neural Recording Implants: A Review
Hashemi Noshahr, F., Nabavi, M., & Sawan, M. (2020) · Sensors · doi:10.3390/s20030904 · Cited by 66 · cited 3×
“Schematics of neural recording analog front-ends implemented using ( a ) an ac-coupled capacitive feedback network [155], ( b ) a dc-coupled amplifier with active low-frequency suppression [159], ( c ) a dc-coupled fully differential difference amplifier [170], and ( d ) a chopper-stabilised…”
Burton, A., Wang, Z., Song, D., Tran, S., Hanna, J., Ahmad, D., Bakall, J. A., Clausen, D. M., Anderson, J., Peralta, R., et al. (2023) · Nature Communications · doi:10.1038/s41467-023-43669-2 · Cited by 66 · cited 1×
“However, peripheral nerves have the [97] potential to self-repair and regenerate, aided by conditions such as anatomical continuity of the injured site, myelination, and the extent of the injury [112].”
Hartmann, C., Dosen, S., Amsuess, S., & Farina, D. (2015) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2014.2357175 · Cited by 64 · cited 3×
“3) Back-End Processing: Digital processing techniques such as reconstruction of the biopotentials use interpolation methods that can be applied in parallel to signal blanking to fill in the blocking phases of the synchronised sample-and-hold process [192], [195].”
MLean, N. A., & Verge, V. M. K. (2016) · Glia · doi:10.1002/glia.23021 · Cited by 62 · cited 1×
“Local neuromodulation of damaged nerve fibers can help support peripheral nerve axonal regeneration through the activation or re-programming of myelinating Schwann cells [113].”
Song, H., Kim, M., Kim, E., Lee, J., Jeong, I., Lim, K., Ryu, S. Y., Oh, M., Kim, Y., & Park, J. (2024) · BMEMat · doi:10.1002/bmm2.12048 · Cited by 61 · cited 3×
“Mechanical modes of neurostimulation enable non-invasive technologies such as ultrasound and magnetic forces by operating on the basis of mechanotransduction to convert mechanical stimuli into electrophysiological signals through compression, expansion, bending and tension [108].”
InfluentialRecent advances in neural interfaces—Materials chemistry to clinical translation
Bettinger, C. J., Ecker, M., Kozai, T. D. Y., Malliaras, G. G., Meng, E. F., & Voit, W. (2020) · MRS Bulletin · doi:10.1557/mrs.2020.195 · Cited by 60 · cited 1×
“Advanced methods of developing parallelised fiber arrays, minimally invasive devices, and biodegradable elastomers that support implantation aim to address the challenges of biocompatibility with improved tissue integration [209].”
Compliant peripheral nerve interfaces.
Paggi, V., Akouissi, O., Micera, S., & Lacour, S. P. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/abcdbe · Cited by 60 · cited 3×
“This improves longevity by addressing interface compromising factors such as the foreign body reaction, which can deteriorate the neural interface and isolate the nerve fibers through scarring [12], [206].”
A Chopper Stabilized, Current Feedback, Neural Recording Amplifier
Samiei, A., & Hashemi, H. (2019) · IEEE Solid-State Circuits Letters · doi:10.1109/lssc.2019.2916754 · Cited by 54 · cited 1×
“Attempts to improve the input impedance such as off-chip coupling capacitors or a positive feedback loop have either proven impractical for implantable solutions or suffer from sensitivity to parasitics and poor stability [178].”
An Implantable Wireless Network of Distributed Microscale Sensors for Neural Applications
Lee, J., Lee, F., Huang, J., Cui, L., Lee, A.-H., Leung, V., Mercier, P., Shellhammer, S., Larson, L., Asbeck, P., et al. (2019) · 2019 9th International IEEE/EMBS Conference on Neural Engineering (NER) · doi:10.1109/ner.2019.8717023 · Cited by 54 · cited 2×
“[81].”
On-Probe Neural Interface ASIC for Combined Electrical Recording and Optogenetic Stimulation
Ramezani, R., Jackson, A., Constandinou, T. G., Degenaar, P., Liu, Y., Dehkhoda, F., Soltan, A., Haci, D., Zhao, H., Firfilionis, D., et al. (2018) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2018.2818818 · Cited by 53 · cited 1×
“Neural probes enable high-density interfacing with minimum invasiveness using a narrow, rigid shank, which can accommodate hundreds to thousands of pixels or electrode sites for recording [17], [73] or stimulating [74], [75] neural activity.”
A bioelectric neural interface towards intuitive prosthetic control for amputees
Nguyen, A. T., Xu, J., Jiang, M., Luu, D. K., Wu, T., Tam, W.-K., Zhao, W., Drealan, M. W., Overstreet, C. K., Zhao, Q., et al. (2020) · Journal of Neural Engineering · doi:10.1088/1741-2552/abc3d3 · Cited by 52 · cited 1×
“The integration of deep learning-based artificial intelligence (AI) with existing interface structures has led to the development of intuitive prosthetic control, which aims to address complex movement patterns supported by an AI decoder [220].”
ReStore: A wireless peripheral nerve stimulation system
Sivaji, V., Grasse, D. W., Hays, S. A., Bucksot, J. E., Saini, R., Kilgard, M. P., & Rennaker, R. L. (2019) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.02.010 · Cited by 51 · cited 1×
“Similarly, the ReStore stimulation system and the endovascular neural stimulators have enabled wireless control of motor units with largely commercial off-the-shelf components, reducing the costs and risks of development [49], [50].”
Yuan, X., Hierlemann, A., & Frey, U. (2021) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2021.3066043 · Cited by 50 · cited 1×
“Devices implemented using active structures have enabled channel counts orders of magnitude larger than that of passive interface structures [37], ranging up to tens of thousands of channels on a single probe [38].”
Hsu, W.-Y., & Schmid, A. (2017) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2017.2694144 · Cited by 49 · cited 2×
“…compliance monitoring [129], ( c ) dynamic current copying in a source-sink structure [66], ( d ) high-frequency switched capacitor configuration [142]. have been made to combine the regulated cascode topology with a current-copying stimulator in [133], which minimises the required power and…”
Zhao, Y., Shang, Z., & Lian, Y. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2924416 · Cited by 49 · cited 2×
“6(c)) employ a three-amplifier topology to maximise the amplifier’s CMRR and noise/area efficiency [170], [174].”
Ortmanns, M., Unger, N., Rocke, A., Gehrke, M., & Tietdke, H. J. (2006) · 2006 IEEE International Solid State Circuits Conference - Digest of Technical Papers · doi:10.1109/isscc.2006.1696037 · Cited by 48
ElAnsary, M., Xu, J., Filho, J. S., Dutta, G., Long, L., Tejeiro, C., Shoukry, A., Tang, C., Kilinc, E., Joshi, J., et al. (2021) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2021.3113354 · Cited by 48 · cited 2×
“Recording circuits developed for the PNS have strict requirements of noise cancellation due to the proximity of sources of biosignal interference including electromyogram and electrocardiogram which are orders of magnitude larger than the signals of interest [54].”
A Wireless Multi-Channel Peripheral Nerve Signal Acquisition System-on-Chip
Ng, K. A., Yuan, C., Rusly, A., Do, A.-T., Zhao, B., Liu, S.-C., Peh, W. Y. X., Thow, X. Y., Voges, K., Lee, S., et al. (2019) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2019.2909158 · Cited by 47 · cited 1×
“Alternative structures such as the fascicle-selective, die-on-cuff interface, flat-interface cuff electrodes (FINE) and thin-film longitudinal intra-fascicular electrodes (tfLIFE) have also been developed with ICs to achieve selective, active neural interfaces [54], [55], [56], [57].”
Ramachandran, A., Schuettler, M., Lago, N., Doerge, T., Koch, K. P., Navarro, X., Hoffmann, K.-P., & Stieglitz, T. (2006) · Journal of Neural Engineering · doi:10.1088/1741-2560/3/2/005 · Cited by 46 · cited 1×
“An external tube may be used to provide mechanical support, join multiple sieves, or enhance performance, such as the cuff and sieve electrode (CASE), as a hybrid method [31].”
Regenerative Scaffold Electrodes for Peripheral Nerve Interfacing
Clements, I. P., Mukhatyar, V. J., Srinivasan, A., Bentley, J. T., Andreasen, D. S., & Bellamkonda, R. V. (2013) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2012.2217352 · Cited by 46 · cited 2×
“…as the regenerative multielectrode interface, regenerative scaffold electrodes (RSE), double-aisle electrodes, and tissue-engineered electronic nerve interfaces (TEENI) have been developed, where supporting biomolecules and hydrogel scaffolds can be integrated to help repair injured nerves [32].”
Joshi‐Imre, A., Black, B., Abbott, J., Kanneganti, A., Rihani, R. T., Chakraborty, B., Danda, V. R., Maeng, J., Sharma, R., Rieth, L. W., et al. (2019) · Journal of Neural Engineering · doi:10.1088/1741-2552/ab1bc8 · Cited by 46 · cited 1×
“They can also directly contact the white matter with improved resolution using penetrative neural probes [17], Utah electrode arrays [18], and flexible mesh electronics [19].”
Uehlin, J. P., Smith, W. A., Pamula, V. R., Perlmutter, S. I., Rudell, J. C., & Sathe, V. S. (2020) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2963174 · Cited by 45 · cited 1×
“This can be achieved via the integration of the recording ADC with the feedback path [197] addressing multiple electrodes through a time-multiplexed con-trol structure [198], [199].”
Thakor, N. V. (2023) · Springer eBooks · doi:10.1007/978-981-16-5540-1 · Cited by 42
Mapping of Small Nerve Trunks and Branches Using Adaptive Flexible Electrodes
Xiang, Z., Sheshadri, S., Lee, S., Wang, J., Xue, N., Thakor, N. V., Yen, S., & Lee, C. (2016) · Advanced Science · doi:10.1002/advs.201500386 · Cited by 41
Optical stimulation of neural tissue
Richardson, R. T., Ibbotson, M. R., Thompson, A. C., Wise, A. K., & Fallon, J. B. (2020) · Healthcare Technology Letters · doi:10.1049/htl.2019.0114 · Cited by 40
Boretius, T., Yoshida, K., Badia, J., Harreby, K., Kundu, A., Navarro, X., Jensen, W., & Stieglitz, T. (2012) · 2012 4th IEEE RAS & EMBS International Conference on Biomedical Robotics and Biomechatronics (BioRob) · doi:10.1109/biorob.2012.6290787 · Cited by 39 · cited 2×
“Intraneural: ( c ) longitudinal (LIFE) [25], transverse (TIME) [26], ( d ) self-opening (SELINE) [28], and Utah slanted (USEA) [27] electrode arrays.”
Liu, X., Demosthenous, A., Vanhoestenberghe, A., Jiang, D., & Donaldson, N. (2012) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2174360 · Cited by 39 · cited 4×
“They check for available headroom with respect to the interface impedance [44], [129], and detect excess power or leakage of fluids via temperature and humidity sensing [67], [80].”
InfluentialSelf-citation1225-Channel Neuromorphic Retinal-Prosthesis SoC With Localized Temperature-Regulation
Park, J. H., Tan, J. S. Y., Wu, H., Dong, Y., & Yoo, J. (2020) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2020.3036091 · Cited by 39 · cited 4×
“They check for available headroom with respect to the interface impedance [44], [129], and detect excess power or leakage of fluids via temperature and humidity sensing [67], [80].”
InfluentialFlexible Charge Balanced Stimulator With 5.6 fC Accuracy for 140 nC Injections
Nag, S., Jia, X., Thakor, N., & Sharma, D. (2013) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2012.2205574 · Cited by 38 · cited 1×
“A discrete implementation of floating current sources with steering diodes has shown fC-range accuracy via precise control over the stimulating phases [146].”
The injectable neurostimulator: an emerging therapeutic device
Li, X., Serdijn, W. A., Zheng, W., Tian, Y., & Zhang, B. (2015) · Trends in biotechnology · doi:10.1016/j.tibtech.2015.04.001 · Cited by 38 · cited 1×
“…interface [81], [82], the mi-crobead stimulating system [83], and the intraspinal microstimulation (ISMS) network [47], [84], [85], as well as the injectable neural stimulator (InNS) technologies, which enable interfacing with regions that are inaccessible for implantation [86], [87], [88].”
Recent Trends and Future Prospects of Neural Recording Circuits and Systems: A Tutorial Brief
Chen, J., Tarkhan, M., Wu, H., Noshahr, F. H., Yang, J., & Sawan, M. (2022) · IEEE Transactions on Circuits and Systems II: Express Briefs · doi:10.1109/tcsii.2022.3171689 · Cited by 38 · cited 2×
“Pre-processing techniques such as compressive sensing and adaptive sampling enable efficient digitisation by reducing the amount of recorded data to the biosignals using thresholding methods, thus taking advantage of the sparsity of neural signals in the time or frequency domains [172].”
Niederhoffer, T., Vanhoestenberghe, A., & Lancashire, H. T. (2023) · Journal of Neural Engineering · doi:10.1088/1741-2552/acb084 · Cited by 38 · cited 1×
“In addition, a large input impedance with respect to electrode impedance reduces signal attenuation [156], [158], [160].”
Self-citationAl-Danakh, A., Safi, M., Alradhi, M., Almoiliqy, M., Chen, Q., Al-Nusaif, M., Yang, X., Al-Dherasi, A., Zhu, X., & Yang, D. (2022) · Parkinson's Disease · doi:10.1155/2022/2700227 · Cited by 37 · cited 1×
“Such technologies address some of the challenges involved in transcutaneous tibial nerve stimulation, for example, reduced efficiency arising from the skin impedance and improper placement of electrodes by patients, decreasing the efficacy of the therapy [62].”
A Multichannel High-Frequency Power-Isolated Neural Stimulator With Crosstalk Reduction
Jiang, D., & Demosthenous, A. (2018) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2018.2832541 · Cited by 36 · cited 3×
“This can compromise the efficacy of the neural interface solution [77], requiring electrical [78] or physical isolation between the interface channel [71].”
Self-citationA Fully Integrated RF-Powered Energy-Replenishing Current-Controlled Stimulator
Ha, S., Kim, C., Park, J., Cauwenberghs, G., & Mercier, P. P. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2018.2881800 · Cited by 35 · cited 1×
“Adiabatic constant-current stimulation has been developed to achieve superior power efficiency by recycling the un-used energy [130].”
Vallone, F., Ottaviani, M. M., Dedola, F., Cutrone, A., Romeni, S., Panarese, A. M., Bernini, F., Cracchiolo, M., Strauss, I., Gabisonia, K., et al. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac0d42 · Cited by 35 · cited 1×
“Advanced methods that incorporate biopotential decoding algorithms have enabled a superior understanding of the anatomical organisation of the vagus nerve with respect to cardiac and respiratory functions [200].”
Guo, Y., Sun, L., Zhong, W., Zhang, N., Zhao, Z., & Tian, W. (2024) · PubMed Central · doi:10.4103/1673-5374.380909 · Cited by 35
Fathy, N. S. K., Huang, J., & Mercier, P. P. (2021) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2021.3116021 · Cited by 34 · cited 1×
“This can be achieved via the integration of the recording ADC with the feedback path [197] addressing multiple electrodes through a time-multiplexed con-trol structure [198], [199].”
A 430nW 64nV/vHz current-reuse telescopic amplifier for neural recording applications
Song, S., Rooijakkers, M. J., Harpe, P., Rabotti, C., Mischi, M., van Roermund, A. H. M., & Cantatore, E. (2013) · 2013 IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2013.6679704 · Cited by 33 · cited 1×
“A superior trade-off between noise and power has been reported (NEF 2) using the stacked and orthogonal current-reuse methods with OTA topologies such as the telescopic and folded cascode amplifiers [183].”
Bionic Intrafascicular Interfaces for Recording and Stimulating Peripheral Nerve Fibers
Jung, R., Abbas, J. J., Kuntaegowdanahalli, S. S., & Thota, A. K. (2017) · Bioelectronics in Medicine · doi:10.2217/bem-2017-0009 · Cited by 32 · cited 2×
“While the small diameter of LIFE and its longitudinal placement on a nerve reduces the mechanical stress it imposes on the tissue, multichannel operation across the nerve cross-section requires several electrode arrays, which increases fascicular damage [21].”
Retinal Stimulator ASIC Architecture Based on a Joint Power and Data Optical Link
Lemaire, W., Benhouria, M., Koua, K., Besrour, M., Gauthier, L.-P., Martin-Hardy, G., Rossignol, T., Roy, S., & Fontaine, R. (2021) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2020.3045141 · Cited by 32 · cited 7×
“Voltage controlled stimulation is the simplest and can ensure electrode potentials remain within the water window; however, variations in stimulus current with respect to the electrode load lead to uncertainty in efficacy and challenges with charge-balancing [66].”
InfluentialCuff electrodes for very small diameter nerves — Prototyping and first recordings in vivo
Ordonez, J. S., Pikov, V., Wiggins, H., Patten, C., Stieglitz, T., Rickert, J., & Schuettler, M. (2014) · 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc.2014.6945201 · Cited by 30 · cited 1×
“Extraneural: ( a ) cuff [40], ( b ) ribbon and split-ring electrodes [41].”
Xu, J., Nguyen, A. T., Wu, T., Zhao, W., Luu, D. K., & Yang, Z. (2020) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2020.2972013 · Cited by 29 · cited 1×
“Techniques such as a wide input dynamic range reduce the risk of front-end saturation and prevent loss of data [161].”
In-vivo tests of a 16-channel implantable wireless neural stimulator
Troyk, P. R., Bredeson, S. D., Cogan, S. F., Romero‐Ortega, M. I., Suh, S., Hu, Z., Kanneganti, A., Granja-Vazquez, R., Seifert, J. L., & Bak, M. (2015) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2015.7146662 · Cited by 28 · cited 1×
“…such as the neurograin interface [81], [82], the mi-crobead stimulating system [83], and the intraspinal microstimulation (ISMS) network [47], [84], [85], as well as the injectable neural stimulator (InNS) technologies, which enable interfacing with regions that are inaccessible for…”
Perspectives on electrical neural recording: a revisit to the fundamental concepts
Guo, L. (2020) · Journal of Neural Engineering · doi:10.1088/1741-2552/ab702f · Cited by 28 · cited 3×
“…and narrow bandwidths (200 Hz to 10 kHz) demand high-gain pre-amplification (typically of about 40 dB to 60 dB) and filtering prior to digitisation and processing [156], and are recorded in the presence of large common mode signals requiring high common-mode rejection ratio (CMRR) [157], [158].”
Lamont, C., Grego, T., Nanbakhsh, K., Shah Idil, A., Giagka, V, Vanhoestenberghe, A., Cogan, S., & Donaldson, N. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/abf0d6 · Cited by 28 · cited 1×
“Implant packaging methods to provide a biocompatiblity and corrosion protection through hermetic packages and coatings and non-hermetic encapsulants aim to offer device lifetimes of years to decades [213], [214], [215].”
An ultra-low-power neural signal acquisition analog front-end IC
Tran, L., & Cha, H.-K. (2021) · Microelectronics Journal · doi:10.1016/j.mejo.2020.104950 · Cited by 26 · cited 1×
“Sub-threshold transistor operation enables the currents to scale down exponentially, improving the transconductance-to-drain-current ( g m /I D ) ratio for greater power efficiency with a lower supply voltage [180], [181].”
An Implantable Versatile Electrode-Driving ASIC for Chronic Epidural Stimulation in Rats
Giagka, V., Eder, C., Donaldson, N., & Demosthenous, A. (2015) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2014.2330859 · Cited by 25 · cited 3×
“Distributed: ( k ) spinal cord stimulator [46] and ( l ) intraspinal microstimulator [47].”
Self-citationA 32-Channel Time-Multiplexed Artifact-Aware Neural Recording System
Pérez-Prieto, N., Rodriguez-Vazquez, A., Álvarez‐Dolado, M., & Delgado‐Restituto, M. (2021) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2021.3108725 · Cited by 25 · cited 1×
“Several transistor structures have been proposed for implementing pseudo-resistors that aim to obtain a stable and uniform equivalent resistance across the expected range of voltages [196] using transistors operating in the sub-threshold region with an acceptable tolerance to PVT variations.”
Shadmani, A., Viswam, V., Chen, Y., Bounik, R., Dragas, J., Radivojevic, M., Geissler, S., Sitnikov, S., Muller, J., & Hierlemann, A. (2019) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2890530 · Cited by 24 · cited 3×
“Alteration of the poles is often achieved using a variable pseudo-resistor in the amplifier’s feedback path, which determines its high-pass characteristics [190].”
A modular 32-site wireless neural stimulation microsystem
Ghovanloo, M., & Najafi, K. (2004) · Digest of technical papers/Digest of technical papers - IEEE International Solid-State Circuits Conference · doi:10.1109/isscc.2004.1332676 · Cited by 24
A Low Noise Amplifier for Neural Spike Recording Interfaces
Ruiz-Amaya, J., Rodriguez-Perez, A., & Delgado-Restituto, M. (2015) · Sensors · doi:10.3390/s151025313 · Cited by 23 · cited 1×
“A common element of integrated recording systems is an operational transconductance amplifier (OTA), used as the active amplifying element and can be implemented using a variety of techniques [157], [167].”
Harnessing the Inflammatory Reflex for the Treatment of Inflammation-Mediated Diseases
Levine, Y. A., Faltys, M., & Chernoff, D. (2020) · Cold Spring Harbor Perspectives in Medicine · doi:10.1101/cshperspect.a034330 · Cited by 22
A Fully Implantable Opto-Electro Closed-Loop Neural Interface for Motor Neuron Disease Studies
Liu, F., Wu, Y., Almarri, N., Habibollahi, M., Lancashire, H. T., Bryson, J. B., Greensmith, L., Jiang, D., & Demosthenous, A. (2022) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2022.3202026 · Cited by 22 · cited 3×
“…to provide sensory perception and motor functionality has been proposed with high-density transverse intra-fascicular micro-electrodes (hd-TIME) [68], [69], [70] and active microchannel neural interface (MNI) systems [71], [72], where the electrodes are integrated into the IC and additional…”
InfluentialSelf-citationFascicle-Selective Ultrasound-Powered Bidirectional Wireless Peripheral Nerve Interface IC
Xu, J., Filho, J. S., Nag, S., Long, L., Hwang, E., Tejeiro, C., O’Leary, G., Huang, Y. T., Kanchwala, M., Abdolrazzaghi, M., et al. (2023) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2023.3332258 · Cited by 22 · cited 1×
“Alternative structures such as the fascicle-selective, die-on-cuff interface, flat-interface cuff electrodes (FINE) and thin-film longitudinal intra-fascicular electrodes (tfLIFE) have also been developed with ICs to achieve selective, active neural interfaces [54], [55], [56], [57].”
Langlois, P. J., Demosthenous, A., Pachnis, I., & Donaldson, N. (2009) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2009.2034138 · Cited by 21 · cited 1×
“Other techniques include the clamping of transistor gate-source voltages to safely drive the gate terminals using depletion-mode transistors for minimum power consumption [153], and the use of stacked standard transistors at the expense of a greater area [154].”
Self-citationImplantable stimulator for bipolar stimulation without charge balancing circuits
Chun, H., Lehmann, T., & Yang, Y. (2010) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2010.5709606 · Cited by 21
Development of an active high-density transverse intrafascicular micro-electrode probe
Verplancke, R., Cauwe, M., Schaubroeck, D., Cuypers, D., Vandecasteele, B., Mader, L., Vanhaverbeke, C., Ballini, M., O’Callaghan, J., Goikoetxea, E., et al. (2020) · Journal of Micromechanics and Microengineering · doi:10.1088/1361-6439/ab5df2 · Cited by 21 · cited 1×
“…sensory perception and motor functionality has been proposed with high-density transverse intra-fascicular micro-electrodes (hd-TIME) [68], [69], [70] and active microchannel neural interface (MNI) systems [71], [72], where the electrodes are integrated into the IC and additional circuits, such…”
Advances in Neural Recording and Stimulation Integrated Circuits
Li, J., Liu, X., Mao, W., Chen, T., & Yu, H. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.663204 · Cited by 21 · cited 3×
“…and narrow bandwidths (200 Hz to 10 kHz) demand high-gain pre-amplification (typically of about 40 dB to 60 dB) and filtering prior to digitisation and processing [156], and are recorded in the presence of large common mode signals requiring high common-mode rejection ratio (CMRR) [157], [158].”
A Vestibular Prosthesis With Highly-Isolated Parallel Multichannel Stimulation
Jiang, D., Cirmirakis, D., & Demosthenous, A. (2015) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2014.2323310 · Cited by 20 · cited 2×
“This can compromise the efficacy of the neural interface solution [77], requiring electrical [78] or physical isolation between the interface channel [71].”
Self-citationRecent advances in charge balancing for functional electrical stimulation
Sooksood, K., Stieglitz, T., & Ortmanns, M. (2009) · 2009 Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2009.5333181 · Cited by 20 · cited 1×
“The large capacitance required is unsuitable for miniaturisation and may introduce additional offsets due to secondary effects [143].”
A Multifunctional Neural Electrode Stimulation ASIC using NeuroTalkTM Interface
Troyk, P. R., Detlefsen, D. E. A., & DeMichele, G. A. (2006) · Conference proceedings · doi:10.1109/iembs.2006.260630 · Cited by 20
Charkhkar, H., Christie, B. P., Pinault, G. J., Tyler, D. J., & Triolo, R. J. (2019) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108414 · Cited by 19 · cited 1×
“In addition to the verification of interface safety, quality, and impact, the customisation to neural targets helps mitigate the potential risks in the process [217].”
Abdi, A., & Cha, H. (2016) · Microelectronics Journal · doi:10.1016/j.mejo.2016.10.013 · Cited by 17 · cited 1×
“Other techniques include the clamping of transistor gate-source voltages to safely drive the gate terminals using depletion-mode transistors for minimum power consumption [153], and the use of stacked standard transistors at the expense of a greater area [154].”
Miniature electroparticle-cuff for wireless peripheral neuromodulation
Hernandez-Reynoso, A. G., Nandam, S., O’Brien, J. M., Kanneganti, A., Cogan, S. F., Freeman, D. K., & Romero‐Ortega, M. I. (2019) · Journal of Neural Engineering · doi:10.1088/1741-2552/ab1c36 · Cited by 17
High-density neural recording system design
Lee, H.-S., Eom, K., Park, M., Ku, S.-B., Lee, K., & Lee, H.-M. (2022) · Biomedical Engineering Letters · doi:10.1007/s13534-022-00233-z · Cited by 17 · cited 3×
“Devices implemented using active structures have enabled channel counts orders of magnitude larger than that of passive interface structures [37], ranging up to tens of thousands of channels on a single probe [38].”
Bihr, U., & Ortmanns, M. (2012) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2012.6271735 · Cited by 16 · cited 1×
“Multiple amplifier stages are typically cascaded to achieve the necessary gain and bandwidth characteristics, though much effort is directed toward the design and optimisation of the preamplifier, which has the greatest impact on noise [37], [169].”
Yao, L., Li, P., & Je, M. (2015) · 2015 IEEE Asian Solid-State Circuits Conference (A-SSCC) · doi:10.1109/asscc.2015.7387478 · Cited by 16 · cited 1×
“To overcome this, pulse-width-adaptive charge balancing has been developed as a stand-alone solution or to operate in parallel with offset compensation [147], [148].”
Intraneural active probe for bidirectional peripheral nerve interface
Ballini, M., Bae, J., Marrocco, N., Verplancke, R., Schaubroeck, D., Cuypers, D., Cauwe, M., O'Callaghan, J., Fahmy, A., Maghari, N., et al. (2017) · 2017 Symposium on VLSI Circuits · doi:10.23919/vlsic.2017.8008544 · Cited by 16 · cited 3×
“…provide sensory perception and motor functionality has been proposed with high-density transverse intra-fascicular micro-electrodes (hd-TIME) [68], [69], [70] and active microchannel neural interface (MNI) systems [71], [72], where the electrodes are integrated into the IC and additional circuits,…”
Huang, C.-W., Lai, C.-K., Hung, C.-C., Wu, C.-Y., & Ker, M.-D. (2023) · IEEE Transactions on Circuits and Systems I: Regular Papers · doi:10.1109/tcsi.2023.3254891 · Cited by 16 · cited 1×
“3) Back-End Processing: Digital processing techniques such as reconstruction of the biopotentials use interpolation methods that can be applied in parallel to signal blanking to fill in the blocking phases of the synchronised sample-and-hold process [192], [195].”
Nielsen, T. N., Sevcencu, C., & Struijk, J. J. (2014) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2013.2278479 · Cited by 15 · cited 1×
“Electrode configuration may be used to focus extracellular fields and improve stimulation selectivity and efficacy [121], [122].”
Activities on PNS neural interfaces for the control of hand prostheses
Carpaneto, J., Cutrone, A., Bossi, S., Sergi, P. N., Citi, L., Rigosa, J., Rossini, P. M., & Micera, S. (2011) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2011.6091148 · Cited by 15 · cited 2×
“Intraneural: ( c ) longitudinal (LIFE) [25], transverse (TIME) [26], ( d ) self-opening (SELINE) [28], and Utah slanted (USEA) [27] electrode arrays.”
An ASIC for Recording and Stimulation in Stacked Microchannel Neural Interfaces
Lancashire, H. T., Jiang, D., Demosthenous, A., & Donaldson, N. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2891284 · Cited by 15 · cited 4×
“A switch driver is also necessary to provide control signals for the stimulus pulses and a level shifter if operating at a higher voltage [57], [71].”
InfluentialSelf-citationEshaghi, F., Najafiaghdam, E., & Kassiri, H. (2021) · IEEE Access · doi:10.1109/access.2021.3094766 · Cited by 15 · cited 1×
“…by continuous adjustment of the source/sink currents with dynamic current copying in push-pull architectures, [44], [66], and by insertion of a short pulse train that instantaneously corrects the residual charge, or via [137] compensation of the excess charge as an offset current [135], [144].”
García-Moreno, A., Comerma-Montells, A., Tudela-Pi, M., Minguillón, J., Becerra-Fajardo, L., & Ivorra, A. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac8dc4 · Cited by 15 · cited 1×
“…interface [81], [82], the mi-crobead stimulating system [83], and the intraspinal microstimulation (ISMS) network [47], [84], [85], as well as the injectable neural stimulator (InNS) technologies, which enable interfacing with regions that are inaccessible for implantation [86], [87], [88].”
Hwang, Y.-C. E., Long, L., Filho, J. S., Genov, R., & Zariffa, J. (2024) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2024.3355063 · Cited by 15 · cited 1×
“Using the sensory and proprioceptive input as feedback, FES interfaces have shown promising results in restoring motor functions of paralysed limbs [203].”
Jeong, J., Jung, W., Kim, O., Chu, J.-U., Youn, I., Kim, K., Oh, S. R., Park, J. W., & Kim, J. (2016) · 2016 6th IEEE International Conference on Biomedical Robotics and Biomechatronics (BioRob) · doi:10.1109/biorob.2016.7523786 · Cited by 14 · cited 2×
“Initially proposed as a planar disc in 1974, sieve electrodes comprise several metal holes that form a physical barrier between axon groups with enclosed electrode contacts [30].”
Tawakol, O., Herman, M. D., Foxley, S., et al. (2023) · Artificial Organs · doi:10.1111/aor.14562 · Cited by 13 · cited 2×
“Distributed: ( k ) spinal cord stimulator [46] and ( l ) intraspinal microstimulator [47].”
NeuroBus - Architecture for an Ultra-Flexible Neural Interface
Sporer, M., Vasilaş, I.-G., Adžemović, A., Graber, N., Reich, S., Gueli, C., Eickenscheidt, M., Diester, I., Stieglitz, T., & Ortmanns, M. (2024) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2024.3354785 · Cited by 13 · cited 1×
“…sites helps overcome the safety limitations of wireless powering and communication due to tissue heating and enables easy access to deeper tissue regions, while the active, in-situ control of high-density electrodes enables the use of fewer wires that improves interconnect reliability [76].”
Charge Balance Circuit for Constant Current Neural Stimulation with Less than 8 nC Residual Charge
Yigit, H. A., Ulusan, H., Chamanian, S., & Kulah, H. (2019) · 2019 IEEE International Symposium on Circuits and Systems (ISCAS) · doi:10.1109/iscas.2019.8702790 · Cited by 12 · cited 1×
“The exact limit for safe injected residual charge varies in the literature [124], [125].”
A Closed-Loop Optogenetic Platform
Firfilionis, D., Hutchings, F., Tamadoni, R., Walsh, D. A., Turnbull, M., Escobedo‐Cousin, E., Bailey, R. G., Gausden, J., Patel, A., Haci, D., et al. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.718311 · Cited by 12 · cited 1×
“Neural probes enable high-density interfacing with minimum invasiveness using a narrow, rigid shank, which can accommodate hundreds to thousands of pixels or electrode sites for recording [17], [73] or stimulating [74], [75] neural activity.”
Zhou, Y., Wang, K., Yin, S., Li, W.-Y., Meng, F., Wang, Z.-G., & Ma, K. (2024) · IEEE Transactions on Circuits and Systems I: Regular Papers · doi:10.1109/tcsi.2024.3376370 · Cited by 12 · cited 1×
“…addition to the implementation of thick-oxide devices to achieve a fully isolated active output [140], [150], dynamic bulk biasing techniques have been studied to enable a higher stimulator compliance voltage [151] that is not limited by the diode breakdown level with the aid of charge pumps [152].”
Recent advances in power efficient output stage for high density implantable stimulators
Sooksood, K., Noorsal, E., Bihr, U., & Ortmanns, M. (2012) · 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc.2012.6346066 · Cited by 11 · cited 1×
“However, this method introduces additional mismatch from the VCRs [139].”
Analysis of passive charge balancing for safe current-mode neural stimulation
Rueda, G. L. E., Ballini, M., Van Hellepute, N., & Mitra, S. (2017) · 2017 IEEE International Symposium on Circuits and Systems (ISCAS) · doi:10.1109/iscas.2017.8050621 · Cited by 11 · cited 1×
“The exact limit for safe injected residual charge varies in the literature [124], [125].”
Jiang, D., Wu, Y., Almarri, N., Habibollahi, M., Liu, F., Bryson, J. B., Greensmith, L., & Demosthenous, A. (2021) · 2021 IEEE International Symposium on Circuits and Systems (ISCAS) · doi:10.1109/iscas51556.2021.9401220 · Cited by 11 · cited 1×
“Each electrode operates as either a cathode or an anode from a single supply current, maximising flexibility and limiting the reliance on storage capacitors [57], [134].”
Self-citationElterman, D., Ehlert, M. J., Ridder, D. J. M. K. D., et al. (2021) · Neurourology and Urodynamics · doi:10.1002/nau.24648 · Cited by 10 · cited 1×
“Similarly, a treatment for urinary tract dysfunction or overactive bladder involves neuromodulation of the sacral nerve controlled by a pulse generator implanted on the abdominal or buttock region [15], [16].”
Overview of On-Chip Stimulator Designs for Biomedical Applications
Lin, C.-Y., & Ker, M.-D. (2012) · Journal of Neuroscience and Neuroengineering · doi:10.1166/jnsne.2012.1023 · Cited by 9 · cited 1×
“…the amplitude control bits can be either translated directly to a stimulus current via a current-mode converter using binary-weighted and current-steering architectures or by means of a voltage-mode converter operating a transistor in triode mode as a voltage-controlled resistor (VCR) [138].”
Lyu, L., Ye, D., & Shi, C. ‐. R. (2019) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2019.8702781 · Cited by 8 · cited 1×
“The implementation of replica-biasing circuits, stacked differential inputs, and inverter-based input pairs has enabled biopotential recording with supply voltages 0.5 V for nW-level power consumption [182].”
Influential
Research works printed but not linked: 3
- [13] J. P. O'Reardon, P. Cristancho, and A. D. Peshek, 'Vagus nerve stimulation (VNS) and treatment of depression: To the brainstem and beyond,' Psychiatry (edgmont) , vol. 3, no. 5, pp. 54-63, 2006. -
- [33] E. Atkinson, 'Histological, electrophysiological, and electrochemical analysis of the tissue-engineered electronic nerve interface (TEENI),' Ph.D. thesis, University of Florida, 2021. -
- [89] M. Elansary et al., 'Transmission and 64 2nd-order opamp-less ΣΔ ADCs,' Proc. Dig. Tech. Papers IEEE Int. Solid-State Circuits Conf. , vol. 31, 2021, pp. 400-402. -
Cited by
An Active Microchannel Neural Interface for Implantable Electrical Stimulation and Recording
Habibollahi, M., Jiang, D., Lancashire, H. T., & Demosthenous, A. (2025) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2025.3533612 · Cited by 8
“Fewer integrated devices have been reported for the PNS [17].”
Self-citationCMOS Circuits in the Design and Development of Implantable Medical Devices
Laskar, M. J., Banerjee, A., Prasad, V., & Das, D. (2025) · Biomedical Materials & Devices · doi:10.1007/s44174-024-00267-0 · Cited by 12
“CMOS circuits produce minimal heat due to their efficient power usage, making them safer for long-term implantation [31].”
Biotic‐Abiotic Interface Engineering for Peripheral Nerve Modulation and Repair
Jekal, J., Park, J. T., Kim, E., Lee, Y., & Jang, K. (2025) · Small Methods · doi:10.1002/smtd.202501620 · Cited by 2
Olorocisimo, J. P., Nag, S., Zhang, H., Yang, S. M., Prytula, M., Liu, S., Kanchwala, M., Sun, Y., Zariffa, J., & Genov, R. (2025) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-025-01825-3 · Cited by 3
“Current peripheral nerve interfaces are generally grouped into 3 categories, which from least to most invasive are: extraneural, intraneural, and regenerative [1, 18, 19].”
Adhesion Promotion for Non-Hermetic Encapsulations of Implantable Ceramic Printed Circuit Boards
Martens, J., Henneken, L., & Stieglitz, T. (2025) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner61569.2025.11588992 · Cited by 1
“Thanks to modern medical technology, sensory abilities can also be restored by stimulating nerves of the sensory branches in the peripheral nervous system (PNS) [1, 3].”
Metadata
- Authors
- Habibollahi, M., Jiang, D., Lancashire, H. T., & Demosthenous, A.
- Affiliations
- University College London (GB)
- Venue
- IEEE Transactions on Biomedical Circuits and Systems
- Published
- 2024-10
- Volume
- 18
- Issue
- 5
- Pages
- 954-975
- ISSN
- 1932-4545 (ISSN-L)
- 1940-9990
- OpenAlex
- W4400727368
- Cited by
- 11
- References
- 215
- Citations by year
- 2026: 5 · 2025: 6
- FWCI
- 0.69
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Locations
- IEEE Transactions on Biomedical Circuits and Systems (publishedVersion, cc-by)
- PubMed (publishedVersion)
- UCL Discovery (University College London) (submittedVersion, cc-by)
- Funders
- Engineering and Physical Sciences Research Council (EPSRC), U.K. Research and Innovation (EP/R513143/1)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Analog and Mixed-Signal Circuit Design
- Life Sciences › Neuroscience › Neurology › Vagus Nerve Stimulation Research
- Keywords
- peripheral nerve interfaces, nerve stimulator, motor impairment, neurological disorders, psychiatric conditions
- MeSH terms
- Animals, Electric Stimulation, Equipment Design, Humans, Peripheral Nerves
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-11, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-11 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.