An analysis of observer-rated functional vision in patients implanted with the Argus II retinal prosthesis system at three years
Open accessAbstract
OBJECTIVE: The purpose of this analysis was to compare observer-rated tasks in patients implanted with the Argus II Retinal Prosthesis System, when the device is ON versus OFF. METHODS: The Functional Low-Vision Observer Rated Assessment (FLORA) instrument was administered to 26 blind patients implanted with the Argus II Retinal Prosthesis System at a mean follow-up of 36 months. FLORA is a multi-component instrument that consists in part of observer-rated assessment of 35 tasks completed with the device ON versus OFF. The ease with which a patient completes a task is scored using a four-point scale, ranging from easy (score of 1) to impossible (score of 4). The tasks are evaluated individually and organised into four discrete domains, including 'Visual orientation', 'Visual mobility', 'Daily life and 'Interaction with others'. RESULTS: Twenty-six patients completed each of the 35 tasks. Overall, 24 out of 35 tasks (69 per cent) were statistically significantly easier to achieve with the device ON versus OFF. In each of the four domains, patients' performances were significantly better (p < 0.05) with the device ON versus OFF, ranging from 19 to 38 per cent improvement. CONCLUSION: Patients with an Argus II Retinal Prosthesis implanted for 18 to 44 months (mean 36 months), demonstrated significantly improved completion of vision-related tasks with the device ON versus OFF.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In practice, FLORA is often employed as an external validation tool that constitutes the very last step of the design process.
“While it is encouraging to see increasing adoption of FLORA by the community (Geruschat et al., 2016; Karapanos et al., 2021), in practice it is often employed as an external validation tool that constitutes the very last step of the design process—a proof of concept, so to speak.”
From Towards a <i>Smart Bionic Eye</i> : AI-powered artificial vision for the treatment of incurable blindness (2022) · cites this paper as Geruschat et al., 2016
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Dorn, J. D., Ahuja, A. K., Caspi, A., da Cruz, L., Dagnelie, G., Sahel, J.-A., Greenberg, R. J., McMahon, M. J., & Argus II Study Group, F. T. (2013) · JAMA Ophthalmology · doi:10.1001/2013.jamaophthalmol.221 · Cited by 177 · see citing works
Self-citation
Flagged first, then by citations
Research works printed but not linked: 13
- [1]Dagnelie G. Retinal implants: Emergence of a multidisciplinary field. Curr Opin Neurol 2012; 25: 67– 75. 22185902 PMC3291121
- [2]Humayun MS, Dorn JD, da Cruz L, Dagnelie G, Sahel JA, Stanga PE, Cideciyan AV et al. Interim results from the international trial of Second Sight's visual prosthesis. Ophthalmology 2012; 119: 779– 788. 22244176 PMC3319859
- [3]Ho AC, Humayun MS, Dorn JD, da Cruz L, Dagnelie G, Handa J, et al. Long‐term results from an epiretinal prosthesis to restore sight to the blind. Ophthalmology 2015; 122: 1547– 1554. 26162233 PMC4516690
- [5]Stelmack JA, Stelmack, TR, Massof, RW. Measuring low‐vision rehabilitation outcomes with the NEI VFQ‐25. Invest Ophthalmol Vis Sci 2002; 43: 2859– 2868. 12202503
- [6]Stelmack, JA, Szlyk JP, Stelmack TR, Demers‐Turco P, Williams RT, Moran D. Psychometric properties of the Veterans Affairs Low‐Vision Visual Functioning Questionnaire. Invest Ophthalmol Vis Sci 2004; 45: 3919– 3928. 15505037
- [7]Sugawara T, Hagiwara A, Hiramatsu A, Ogata K, Mitamura Y, Yamamoto S. Relationship between peripheral visual field loss and vision‐related quality of life in patients with retinitis pigmentosa. Eye 2010; 24: 535– 539. 19590526
- [8]Seo J, Yu H, Lee B. Assessment of functional vision score and vision‐specific quality of life in individuals with retinitis pigmentosa. Korean J Ophthalmol 2009; 23: 164– 168. 19794942 PMC2739964
- [9]Geruschat DR, Flax M, Tanna N, Bianchi M, Fisher A, Goldschmidt M, et al. FLORA™: Phase I development of a functional vision assessment for prosthetic vision users. Clin Exp Optom 2015; 98: 342– 347. 25675964 PMC4481161
- [10]Evans RL, Werkhoven W, Fox HR. Treatment of social isolation and loneliness in a sample of visually impaired elderly persons. Psychol Rep 1982; 51: 103– 108. 7134334
- [11]Caspi A, Dorn JD, McClure KH, Humayun MS, Greenberg RJ, McMahon MJ. Feasibility study of a retinal prosthesis: spatial vision with a 16‐electrode implant. Arch Ophthalmol 2009; 127: 398– 401. 19365014
- [13]Karanicolas P, Farrokhyar F, Bhandari M. Blinding: Who, what, when, why, how? Can J Surg 2010; 53: 345– 348. 20858381 PMC2947122
- [14]Kaptchuk TJ, Stason WB, Davis RB, Legedza AR, Schnyer RN, Kerr CE, et al. Sham device v inert pill: randomized controlled trial of two placebo treatments. BMJ 2006; 332: 391– 397. 16452103 PMC1370970
- [15]Newton M, Sanderson A. The effect of visual impairment on patients’ falls risk. Nurs Older People 2013; 25: 16– 21. 24067043
Cited by
Sorted by citations
Five-year safety and performance results from the argus II retinal prosthesis system clinical trial.
da Cruz, L., Dorn, J. D., Humayun, M. S., Dagnelie, G., Handa, J., Barale, P.-O., Sahel, J.-A., Stanga, P. E., Hafezi, F., Safran, A. B., et al. (2016) · Ophthalmology · doi:10.1016/j.ophtha.2016.06.049 · Cited by 359 · see citing works
Self-citationThe Argus® II Retinal Prosthesis System
Luo, Y. H.-L., & da Cruz, L. (2016) · Progress in Retinal and Eye Research · doi:10.1016/j.preteyeres.2015.09.003 · Cited by 351 · see citing works
An update on retinal prostheses
Ayton, L. N., Barnes, N., Dagnelie, G., Fujikado, T., Goetz, G., Hornig, R., Jones, B. W., Muqit, M. M. K., Rathbun, D. L., Stingl, K., et al. (2020) · Clinical Neurophysiology · doi:10.1016/j.clinph.2019.11.029 · Cited by 223 · see citing works
Self-citationAdvances in retinal prosthesis systems
Bloch, E., Luo, Y., & da Cruz, L. (2019) · Therapeutic Advances in Ophthalmology · doi:10.1177/2515841418817501 · Cited by 201 · see citing works
Bioengineering strategies for restoring vision
Cehajic-Kapetanovic, J., Singh, M. S., Zrenner, E., & MacLaren, R. E. (2022) · Nature Biomedical Engineering · doi:10.1038/s41551-021-00836-4 · Cited by 79 · see citing works
A Second-Generation (44-Channel) Suprachoroidal Retinal Prosthesis: Interim Clinical Trial Results
Petoe, M. A., Titchener, S. A., Kolic, M., Kentler, W. G., Abbott, C. J., Nayagam, D. A. X., Baglin, E. K., Kvansakul, J., Barnes, N., Walker, J. G., et al. (2021) · Translational Vision Science & Technology · doi:10.1167/tvst.10.10.12 · Cited by 63 · see citing works
Beyeler, M., & Sanchez-Garcia, M. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/aca69d · Cited by 34 · see citing works
“Below we expand on these ideas.”
Karapanos, L., Abbott, C. J., Ayton, L. N., Kolic, M., McGuinness, M. B., Baglin, E. K., Titchener, S. A., Kvansakul, J., Johnson, D., Kentler, W. G., et al. (2021) · Translational Vision Science & Technology · doi:10.1167/tvst.10.10.7 · Cited by 30 · see citing works
The argus ii retinal prosthesis system
Bloch, E., & da Cruz, L. (2020) · Prosthesis · doi:10.5772/intechopen.84947 · Cited by 31 · see citing works
Petoe, M. A., McCarthy, C. D., Shivdasani, M. N., Sinclair, N. C., Scott, A. F., Ayton, L. N., Barnes, N. M., Guymer, R. H., Allen, P. J., Blamey, P. J., et al. (2017) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.16-21041 · Cited by 23 · see citing works
Improvement in reading performance through training with simulated thalamic visual prostheses
Rassia, K. E. K., & Pezaris, J. S. (2018) · Scientific Reports · doi:10.1038/s41598-018-31435-0 · Cited by 13 · see citing works
- Authors
- Geruschat, D. R., Richards, T. P., Arditi, A., Cruz, L., Dagnelie, G., Dorn, J. D., Duncan, J. L., Ho, A. C., Olmos de koo, L. C., Sahel, J., Stanga, P. E., Thumann, G., Wang, V., & Greenberg, R. J.
- Affiliations
- Johns Hopkins University (US)
- IM3 Inc, Vancouver, Washington, USA,
- Lighthouse Guild (US)
- Moorfields Eye Hospital (GB)
- Second Sight Medical Products (US)
- University of California, San Francisco (US)
- Wills Eye Hospital (US)
- University of Southern California, Los Angeles, California, USA,
- Centre hospitalier national d'ophtalmologie des Quinze-Vingts (FR)
- Manchester Royal Eye Hospital (GB)
- University Hospital of Geneva (CH)
- Retina Foundation of the Southwest (US)
- Corresponding author
- Geruschat, Duane R
- ORCID
- Venue
- Clinical and Experimental Optometry
- Publisher
- Informa UK Limited
- Published
- 2016-05-01
- Volume
- 99
- Issue
- 3
- Pages
- 227-232
- ISSN
- 0816-4622 (ISSN-L)
- 1444-0938
- PMC
- PMC5066820
- OpenAlex
- W2255805712
- Cited by
- 77
- References
- 1
- Citations by year
- 2026: 19 · 2025: 7 · 2024: 2 · 2023: 7 · 2022: 5 · 2021: 7 · 2020: 10 · 2019: 5 · 2018: 5 · 2017: 5 · 2016: 4 · 2015: 1
- FWCI
- 3.56
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Locations
- Clinical and Experimental Optometry (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Archive ouverte UNIGE (University of Geneva) (publishedVersion)
- eScholarship (California Digital Library) (submittedVersion)
- eScholarship (California Digital Library) (submittedVersion, cc-by)
- Archive ouverte UNIGE (University of Geneva) (submittedVersion, cc-by)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Health Sciences › Health Professions › Occupational Therapy › Assistive Technology in Communication and Mobility
- Health Sciences › Medicine › Epidemiology › Ophthalmology and Visual Impairment Studies
- Keywords
- Argus II retinal prosthesis, functional vision, visual impairment, blind patients, visual orientation, activities of daily living, long-term outcomes
- MeSH terms
- Humans, Vision, Ocular, Prosthesis Implantation, Visual Prosthesis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review