Combined photoplethysmographic monitoring of respiration rate and pulse: A comparison between different measurement sites in spontaneously breathing subjects

Nilsson, L., Goscinski, T., Kalman, S., Lindberg, L.‐G., Johansson, A.·2007·Acta Anaesthesiologica Scandinavica

Abstract

BACKGROUND: The non-invasive photoplethysmographic (PPG) signal reflects blood flow and volume in a tissue. The PPG signal shows variation synchronous with heartbeat (PPGc), as used in pulse oximetry, and variations synchronous with breathing (PPGr). PPGr has been used for non-invasive monitoring of respiration with promising results. Our aim was to investigate PPG signals recorded from different skin sites in order to find suitable locations for parallel monitoring of variations synchronous with heartbeat and breathing. METHODS: PPG sensors were applied to the forearm, finger, forehead, wrist and shoulder on 48 awake healthy volunteers. From these sites, seven PPG signals were simultaneously recorded during normal spontaneous breathing over 10 min. Capnometry served as respiration and electrocardiogram (ECG) as pulse reference signals. PPG signals were compared with respect to power spectral content and squared coherence. RESULTS: Forearm PPG measurement showed significantly higher power within the respiratory region of the power spectrum [median (quartile range) 42 (26)%], but significantly lower power within the cardiac region [9 (10)%] compared with the other skin sites. PPG finger measurement showed the opposite; in transmission mode, the power within the respiratory region was significantly lower [4 (10)%] and within the cardiac region significantly higher [45 (25)%] than the other sites. PPGc coherence values were generally high [>0.96 (0.08)], and PPGr coherence values lower [0.83 (0.35)-0.94 (0.17)]. CONCLUSION: Combined PPG respiration and pulse monitoring is possible, but there are significant differences between the respiratory and cardiac components of the PPG signal at different sites.

Integrity signals

Registry

Crossref · 10.1111/j.1399-6576.2007.01375.x

Cites

27 of 27 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Venous pressure can add low-frequency noise to the PPG waveform, often between 0 and 0.5 Hz, contributing to baseline oscillations.

    “Venous pressure can also add low-frequency noise to the PPG waveform, the same effects seen by respiration, which is caused by venous anatomy [129,130,131]. Often between 0 and 0.5 Hz, this noise contributes to be baseline oscillations.”
  • The forearm had a high respiratory rate spectral power content compared to anatomies such as the finger, which had signal primarily derived from heart rate.

    “It was found that the forearm had a high respiratory rate spectral power content compared to anatomies such as the finger, which had signal primarily derived from heart rate [130].”

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Nilsson, L., Goscinski, T., Kalman, S., Lindberg, L. ‐., & Johansson, A.
Affiliations
  • Linköping University Hospital (SE)
  • Linköping University (SE)
Corresponding author
Nilsson, L.
Venue
Acta Anaesthesiologica Scandinavica
Publisher
Wiley
Published
2007-08-15
Volume
51
Issue
9
Pages
1250-1257
ISSN
  • 0001-5172 (ISSN-L)
  • 1399-6576
OpenAlex
W1976734044
Cited by
88
References
27
Citations by year
2026: 2 · 2025: 4 · 2024: 7 · 2023: 9 · 2022: 8 · 2021: 10 · 2020: 8 · 2019: 6 · 2018: 4 · 2017: 7 · 2016: 2 · 2015: 1 · 2014: 4 · 2013: 3 · 2012: 2
FWCI
1.43
Type
journal article
Language
en
Open access
closed
Locations
  • Acta Anaesthesiologica Scandinavica (publishedVersion)
  • PubMed (publishedVersion)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
Keywords
photoplethysmography, respiratory monitoring, pulse monitoring, measurement site comparison, power spectrum analysis, coherence analysis, spontaneous breathing, finger PPG, non-invasive monitoring
MeSH terms
Adult, Aged, Analysis of Variance, Electrocardiography, Female, Heart Rate, Humans, Male, Middle Aged, Signal Processing, Computer-Assisted, Skin, Respiratory Mechanics, Photoplethysmography, Capnography

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review