Pulse Transit Time as an Indicator of Arterial Blood Pressure
Abstract
The relationship between pulse‐arrival times and diastolic blood pressure was measured in 10 anesthetized dogs. The pulse‐arrival time was measured using the R‐wave of the electrocardiogram (ECG) as a time reference. Pulse‐transit time was also measured between the carotid and femoral pulses. Blood pressure was raised with epinephrine injected intravenously and lowered with vagal stimulation. In all cases, pulse arrival and transit times decreased with an increase in diastolic pressure for diastolic pressures ranging from 15 to 250 mmHg. The correlation between pulse‐arrival time and pressure was poorest when the ECG was used as a timing reference. The best correlation was found with true pulse‐transit time and diastolic pressure. When pulse‐transit time was used to compute pulse‐wave velocity, it was found to increase nearly linearly with blood pressure. From 90–100 mmHg, the pulse‐wave velocity increased typically by slightly less than six percent.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“A key example is the measurement of pulse arrival time (PAT), defined by the time lapse between the maximum fiducial point in the ECG signal (R peak) and the corresponding minimal fiducial point in the PPG signal at valley as in Fig. 6H, as a direct correlate to systolic blood pressure (29 , 30).”
From Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care (2019) · cites this paper as [30]
Citations
Cites
Pulse Wave Velocity as a Measure of Blood Pressure Change
Gribbin, B., Steptoe, A., & Sleight, P. (1976) · Psychophysiology · doi:10.1111/j.1469-8986.1976.tb03344.x · Cited by 227
Pulse Wave Velocity and Blood Pressure Change: Calibration and Applications
Steptoe, A., Smulyan, H., & Gribbin, B. (1976) · Psychophysiology · doi:10.1111/j.1469-8986.1976.tb00866.x · Cited by 172
THE EFFECT OF BLOOD PRESSURE ALTERATION ON THE PULSE WAVE VELOCITY
Nye, E. R. (1964) · Heart · doi:10.1136/hrt.26.2.261 · Cited by 103
Steptoe, A. (1976) · Psychophysiology · doi:10.1111/j.1469-8986.1976.tb00875.x · Cited by 45
Bramwell, J. C. (1924) · QJM · doi:10.1093/qjmed/os-17.67.225 · Cited by 8
Research works printed but not linked: 4
- Bazett H. C, Measurement of pulse wave velocity, American Journal of Physiology, 63, 1922
- Bramwell J. G., 1923, The effect of blood pressure on the extensibility of the human artery, Heart, 10, 289
- Geddes L. A., 1969, Pulse wave velocity as a continuous indicator of indirect blood pressure, Proceedings of the 14th Annual AA Ml Meeting, 14, 89
- McSwiney B. A., 1924, The effect of variations of blood pressure on pulse wave‐velocity (in man), Journal of Physiology, 49, 217
Cited by
Sorted by citations
Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2441951 · Cited by 935
“Since the level of the waveform feet is diastolic BP, such PTT measurements may best correspond to diastolic BP [44], [45].”
BackgroundInfluentialBinodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care
Chung, H. U., Kim, B. H., Lee, J. Y., Lee, J., Xie, Z., Ibler, E. M., Lee, K., Banks, A., Jeong, J. Y., Kim, J., et al. (2019) · Science · doi:10.1126/science.aau0780 · Cited by 807
Chen, W., Kobayashi, T., Ichikawa, S., Takeuchi, Y., & Togawa, T. (2000) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345755 · Cited by 425
Smart Vest: Wearable multi-parameter remote physiological monitoring system
Pandian, P. S., Mohanavelu, K., Safeer, K. P., Kotresh, T. M., Shakunthala, D. T., Gopal, P., & Padaki, V. C. (2007) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2007.05.014 · Cited by 420
Pulse transit time measured from the ECG: an unreliable marker of beat-to-beat blood pressure
Payne, R. A., Symeonides, C. N., Webb, D. J., & Maxwell, S. R. J. (2006) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00657.2005 · Cited by 418
Blood Pressure Estimation from Photoplethysmogram Using a Spectro-Temporal Deep Neural Network
Slapničar, G., Mlakar, N., & Luštrek, M. (2019) · Sensors · doi:10.3390/s19153420 · Cited by 387
El-Hajj, C., & Kyriacou, P. A. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2020.101870 · Cited by 300
Optical blood pressure estimation with photoplethysmography and FFT-based neural networks
Xing, X., & Sun, M. (2016) · Biomedical Optics Express · doi:10.1364/boe.7.003007 · Cited by 263
Teng, X. F., & Zhang, Y. (2004) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2003.1280811 · Cited by 241
Management of Hypertension in the Digital Era
Kario, K. (2020) · Hypertension · doi:10.1161/hypertensionaha.120.14742 · Cited by 247
Blood pressure estimation from appropriate and inappropriate PPG signals using A whole-based method
Mousavi, S. S., Firouzmand, M., Charmi, M., Hemmati, M. R., Moghadam, M., & Ghorbani, Y. (2018) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2018.08.022 · Cited by 234
Katz, E. S., Lutz, J., Black, C., & Marcus, C. L. (2003) · Pediatric Research · doi:10.1203/01.pdr.0000057206.14698.47 · Cited by 215
Non-constrained Blood Pressure Monitoring Using ECG and PPG for Personal Healthcare
Yoon, Y.-Z., Cho, J. H., & Yoon, G. (2008) · Journal of Medical Systems · doi:10.1007/s10916-008-9186-0 · Cited by 199
Wong, M. Y. M., Poon, C. C. Y., & Zhang, Y. (2009) · Cardiovascular Engineering An International Journal · doi:10.1007/s10558-009-9070-7 · Cited by 191
Tanveer, M. S., & Hasan, M. K. (2019) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2019.02.028 · Cited by 200
Huynh, T. H., Jafari, R., & Chung, W. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2865751 · Cited by 181
Ding, X., Zhao, N., Yang, G., Pettigrew, R. I., Lo, B. P. L., Miao, F., Li, Y., Liu, J., & Zhang, Y. (2016) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2016.2620995 · Cited by 184
Photoplethysmography and Deep Learning: Enhancing Hypertension Risk Stratification
Liang, Y., Chen, Z., Ward, R. K., & Elgendi, M. (2018) · Biosensors · doi:10.3390/bios8040101 · Cited by 176
Pitson, D., Sandell, A., van den Hout, R., & Stradling (1995) · European Respiratory Journal · doi:10.1183/09031936.95.08101669 · Cited by 167
Introducing Contactless Blood Pressure Assessment Using a High Speed Video Camera
Jeong, I. C., & Finkelstein, J. (2016) · Journal of Medical Systems · doi:10.1007/s10916-016-0439-z · Cited by 161
El-Hajj, C., & Kyriacou, P. A. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2020.102301 · Cited by 156
Secondary Peak Detection of PPG Signal for Continuous Cuffless Arterial Blood Pressure Measurement
He, X., Goubran, R. A., & Liu, X. P. (2014) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2014.2299524 · Cited by 149
Printing Flexible and Hybrid Electronics for Human Skin and Eye‐Interfaced Health Monitoring Systems
Kim, K., Kim, B., & Lee, C. H. (2019) · Advanced Materials · doi:10.1002/adma.201902051 · Cited by 148
Shin, H., & Min, S. D. (2017) · BioMedical Engineering OnLine · doi:10.1186/s12938-016-0302-y · Cited by 145
Esmaelpoor, J., Moradi, M. H., & Kadkhodamohammadi, A. (2020) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2020.103719 · Cited by 150
A survey on signals and systems in ambulatory blood pressure monitoring using pulse transit time
Buxi, D., Redouté, J., & Yuce, M. R. (2015) · Physiological Measurement · doi:10.1088/0967-3334/36/3/r1 · Cited by 140
Muehlsteff, J., Aubert, X. L., & Schuett, M. (2006) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2006.260275 · Cited by 137
Cuffless blood pressure estimation from PPG signals and its derivatives using deep learning models
El-Hajj, C., & Kyriacou, P. A. (2021) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2021.102984 · Cited by 140
Monitoring vital signs over multiplexed radio by near-field coherent sensing
Hui, X., & Kan, E. C.-C. (2017) · Nature Electronics · doi:10.1038/s41928-017-0001-0 · Cited by 140
Ding, X., & Zhang, Y. (2019) · Biomedical Engineering Letters · doi:10.1007/s13534-019-00096-x · Cited by 142
PPG2ABP: Translating photoplethysmogram (PPG) signals to arterial blood pressure (ABP) waveforms
Ibtehaz, N., Mahmud, S., Chowdhury, M. E. H., Khandakar, A., Salman Khan, M., Ayari, M. A., Tahir, A. M., & Rahman, M. S. (2022) · Bioengineering · doi:10.3390/bioengineering9110692 · Cited by 133
Pulse arrival time is not an adequate surrogate for pulse transit time as a marker of blood pressure
Zhang, G., Gao, M., Xu, D. Z., Olivier, N. B., & Mukkamala, R. (2011) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00980.2011 · Cited by 133
Wang, G., Atef, M., & Lian, Y. (2018) · IEEE Circuits and Systems Magazine · doi:10.1109/mcas.2018.2849261 · Cited by 121
Naschitz, J. E., Bezobchuk, S., Priselac, R. M., Sundick, S., Dreyfuss, D., Khorshidi, I., Karidis, A., Manor, H., Nagar, M., Peck, E. R., et al. (2004) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-005-4300-z · Cited by 114
Aguirre, N. A., Grall‐Maës, E., Cymberknop, L. J., & Armentano, R. L. (2021) · Sensors · doi:10.3390/s21062167 · Cited by 115
Pollak, M. H., & Obrist, P. A. (1983) · Psychophysiology · doi:10.1111/j.1469-8986.1983.tb00895.x · Cited by 94
Proença, J., Muehlsteff, J., Aubert, X. L., & Carvalho, P. (2010) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2010.5626627 · Cited by 92
Non-constrained monitoring of systolic blood pressure on a weighing scale
Shin, J. H., Lee, K. M., & Park, K. S. (2009) · Physiological Measurement · doi:10.1088/0967-3334/30/7/011 · Cited by 86
Relationship between pulse-wave velocity and arterial elasticity
CALLAGHAN, F., Geddes, L. A., Babbs, C. F., & BOURLAND, J. D. (1986) · Medical & Biological Engineering & Computing · doi:10.1007/bf02441620 · Cited by 83
Self-citationMaqsood, S., Xu, S., Springer, M., & Mohawesh, R. (2021) · IEEE Access · doi:10.1109/access.2021.3117969 · Cited by 84
Monitoring physiological signals using nonintrusive sensors installed in daily life equipment
Lim, Y. G., Hong, K. H., Kim, K. K., Shin, J. H., Lee, S. M., Chung, G. S., Baek, H. J., Jeong, D., & Park, K. S. (2011) · Biomedical Engineering Letters · doi:10.1007/s13534-011-0012-0 · Cited by 81
Pulse Transit Time and Blood Pressure During Cardiopulmonary Exercise Tests
Wibmer, T., DOERING, K., Kropf-Sanchen, C., RÜDIGER, S., Blanta, I., Stoiber, K., Rottbauer, W., & SCHUMANN, C. (2014) · Physiological Research · doi:10.33549/physiolres.932581 · Cited by 81
An Accurate Bioimpedance Measurement System for Blood Pressure Monitoring
Huynh, T. H., Jafari, R., & Chung, W. (2018) · Sensors · doi:10.3390/s18072095 · Cited by 80
Systolic blood pressure estimation using PPG and ECG during physical exercise
Sun, S., Bezemer, R., Long, X., Muehlsteff, J., & Aarts, R. M. (2016) · Physiological Measurement · doi:10.1088/0967-3334/37/12/2154 · Cited by 78
Wearable ballistocardiogram and seismocardiogram systems for health and performance
Etemadi, M., & Inan, O. T. (2017) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00298.2017 · Cited by 77
Cheng, J., Xu, Y., Song, R., Liu, Y., Li, C., & Chen, X. (2021) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2021.104877 · Cited by 80
Alignment‐Free Liquid‐Capsule Pressure Sensor for Cardiovascular Monitoring
Fan, X., Huang, Y., Ding, X., Luo, N., Li, C., Zhao, N., & Chen, S. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201805045 · Cited by 73
The relationship between arterial pulse-wave velocity and pulse frequency at different pressures
CALLAGHAN, F., Babbs, C. F., BOURLAND, J. D., & Geddes, L. A. (1984) · Journal of Medical Engineering & Technology · doi:10.3109/03091908409032067 · Cited by 72
Self-citationAdvances in Non-Invasive Blood Pressure Measurement Techniques
Panula, T., Sirkiä, J., Wong, D., & Kaisti, M. (2022) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2022.3141877 · Cited by 70
Wong, M. Y. M., Pickwell‐MacPherson, E., Zhang, Y., & Cheng, J. C. Y. (2010) · European Journal of Applied Physiology · doi:10.1007/s00421-010-1626-0 · Cited by 69
Lee, J., Yang, S.-M., Lee, S., & Kim, H. C. (2019) · Journal of Clinical Medicine · doi:10.3390/jcm8111773 · Cited by 72
A new approach for non-intrusive monitoring of blood pressure on a toilet seat
Kim, J. S., Chee, Y. J., Park, J. W., Choi, J. W., & Park, K. S. (2006) · Physiological Measurement · doi:10.1088/0967-3334/27/2/010 · Cited by 64
Marie, G. V., Lo, C. R., Jones, J. V., & Johnston, D. W. (1984) · Psychophysiology · doi:10.1111/j.1469-8986.1984.tb00235.x · Cited by 63
Yang, S.-M., Sohn, J., Lee, S., Lee, J., & Kim, H. C. (2020) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2020.3009658 · Cited by 61
Enhancing the estimation of blood pressure using pulse arrival time and two confounding factors
Baek, H. J., Kim, K. K., Kim, J. S., Lee, B., & Park, K. S. (2009) · Physiological Measurement · doi:10.1088/0967-3334/31/2/002 · Cited by 59
Continuous blood pressure measurement using pulse transit time
Hennig, A., & Patzak, A. R. (2013) · Somnologie - Schlafforschung und Schlafmedizin · doi:10.1007/s11818-013-0617-x · Cited by 58
Ochiai, R., Takeda, J., Hosaka, H., Sugo, Y., Tanaka, R., & Soma, T. (1999) · Journal of Clinical Monitoring and Computing · doi:10.1023/a:1009950731297 · Cited by 58
Young, C. C., Mark, J. B., White, W. D., DeBree, A., Vender, J. S., & Fleming, A. J. (1995) · The Journal of Clinical Monitoring · doi:10.1007/bf01617519 · Cited by 56
Pulse Transit Time as a Screening Test for Pediatric Sleep-Related Breathing Disorders
Brietzke, S. E., Katz, E. S., & Roberson, D. W. (2007) · Archives of Otolaryngology - Head and Neck Surgery · doi:10.1001/archotol.133.10.980 · Cited by 55
Kim, S., Song, J., Park, J.-H., Kim, J., Park, Y., & Hwang, G. (2012) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e318270a6d9 · Cited by 54
Measurement of pulse-wave velocity using a beat-sampling technique
Pruett, J. D., BOURLAND, J. D., & Geddes, L. A. (1988) · Annals of Biomedical Engineering · doi:10.1007/bf02364622 · Cited by 53
Self-citationHuynh, S., Balan, R. K., Ko, J., & Lee, Y. (2019) · ACM Conference on Embedded Networked Sensor Systems (SenSys) · doi:10.1145/3356250.3360036 · Cited by 52
Wang, D., Yang, X., Liu, X., Ma, L., Li, L., & Wang, W. (2021) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2021.3109986 · Cited by 50
Acute Cardiovascular Responses to Arousal from Non-REM Sleep During Normoxia and Hypoxia
CATCHESIDE, P. G., Chiong, S. C., Orr, R. S., Mercer, J., Saunders, N. A., & McEvoy, R. D. (2001) · SLEEP · doi:10.1093/sleep/24.8.895 · Cited by 48
Bersi, M. R., Ferruzzi, J., Eberth, J. F., Gleason, R. L., & Humphrey, J. D. (2014) · Annals of Biomedical Engineering · doi:10.1007/s10439-014-0988-6 · Cited by 45
Kim, J. Y., Cho, B. H., Im, S. M., Jeon, M.-J., Kim, I. Y., & Kim, S. I. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2005.1616102 · Cited by 45
Past, present and future of blood pressure measuring instruments and their calibration
Kumar, R., Dubey, P. K., Zafer, A., Kumar, A., & Yadav, S. (2020) · Measurement · doi:10.1016/j.measurement.2020.108845 · Cited by 47
Effect of confounding factors on blood pressure estimation using pulse arrival time
Kim, J. S., Kim, K. K., Baek, H. J., & Park, K. S. (2008) · Physiological Measurement · doi:10.1088/0967-3334/29/5/007 · Cited by 44
The effect of local mild cold exposure on pulse transit time
Zhang, X.-Y., & Zhang, Y.-T. (2006) · Physiological Measurement · doi:10.1088/0967-3334/27/7/008 · Cited by 40
Blood Pressure Measurement: From Cuff-Based to Contactless Monitoring
Man, P.-K., Cheung, K.-L., Sangsiri, N., Shek, W., Wong, K. L., Chin, J. W., Chan, T. T., & So, R. H. Y. (2022) · Healthcare · doi:10.3390/healthcare10102113 · Cited by 40
Lee, J., Sohn, J., Park, J., Yang, S.-M., Lee, S., & Kim, H. C. (2018) · BioMedical Engineering OnLine · doi:10.1186/s12938-018-0510-8 · Cited by 40
Chen, W., Yi, Z., Lim, L. J. R., Lim, R. Q. R., Zhang, A., Qian, Z., Huang, J., He, J., & Лю, Б. (2024) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2024.1420100 · Cited by 40
O’Brien, L. M., & Gozal, D. (2007) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.26744 · Cited by 39
Home geriatric physiological measurements
TAMURA, T. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/10/r47 · Cited by 38
Reconstructing QRS Complex From PPG by Transformed Attentional Neural Networks
Chiu, H.-Y., Shuai, H.-H., & Chao, P. C.-P. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.3000344 · Cited by 44
Luštrek, M., Bohanec, M., Barca, C. C., Ciancarelli, M. C., Clays, E., Dawodu, A. A., Derboven, J., Smedt, D. D., Dovgan, E., Lampe, J., et al. (2021) · JMIR Medical Informatics · doi:10.2196/24501 · Cited by 39
Design and development of continuous cuff-less blood pressure monitoring devices
Griggs, D. J., Sharma, M., Naghibi, A., Wallin, C., Ho, V., Barbosa, K., Ghirmai, T., Cao, H., & Krishnan, S. K. (2016) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/icsens.2016.7808908 · Cited by 37
Nyvad, J. T., Christensen, K. L., Buus, N. H., & Reinhard, M. (2020) · Journal of Clinical Hypertension · doi:10.1111/jch.14135 · Cited by 37
Ferroelectret film‐based patch‐type sensor for continuous blood pressure monitoring
Noh, S., Yoon, C., Hyun, E., Yoon, H. N., Chung, T. J., Park, K. S., & Kim, H. C. (2014) · Electronics Letters · doi:10.1049/el.2013.3715 · Cited by 35
Non-Invasive Blood Pressure Sensing via Machine Learning
Attivissimo, F., D’Alessandro, V. I., Palma, L. D., Lanzolla, A. M. L., & Nisio, A. D. (2023) · Sensors · doi:10.3390/s23198342 · Cited by 35
Wearable PPG Based BP Estimation Methods: A Systematic Review and Meta-Analysis
Sastimoglu, Z., Subramaniam, S., Faisal, A. I., Jiang, W., Ye, A., & Deen, M. J. (2024) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2024.3499834 · Cited by 35
Escobar, B., Villa, R. A. T., & Kyriacou, P. A. (2018) · Frontiers in Physiology · doi:10.3389/fphys.2018.01848 · Cited by 35
Hellhammer, D. H., Vogt, D., Franz, N., Freitas, U., & Rutenberg, D. (2014) · Lipids in Health and Disease · doi:10.1186/1476-511x-13-121 · Cited by 34
van Velzen, M. H. N., Loeve, A. J., Niehof, S. P., & Mik, E. G. (2017) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-017-1642-x · Cited by 34
Noninvasive Estimation of Blood Pressure Using Photoplethysmographic Signals in the Period Domain
Yan, Y. S., & Zhang, Y. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2005.1617255 · Cited by 33
Latest hypertension research to inform clinical practice in Asia
Kario, K., Mogi, M., & Hoshide, S. (2022) · Hypertension Research · doi:10.1038/s41440-022-00874-8 · Cited by 32
Kao, Y.-H., Chao, P. C.-P., & Wey, C. (2018) · Microsystem Technologies · doi:10.1007/s00542-018-3877-3 · Cited by 32
Mohammadi, H., Tarvirdizadeh, B., Alipour, K., & Ghamari, M. (2025) · Scientific Reports · doi:10.1038/s41598-025-07087-2 · Cited by 33
Attivissimo, F., Palma, L. D., Nisio, A. D., Scarpetta, M., & Lanzolla, A. M. L. (2023) · Sensors · doi:10.3390/s23042321 · Cited by 31
Development of A Nonintrusive Blood Pressure Estimation System for Computer Users
Kim, J., Park, J., Kim, K., Chee, Y., Lim, Y., & Park, K. (2007) · Telemedicine Journal and e-Health · doi:10.1089/tmj.2006.0034 · Cited by 31
Estimation of pulse transit time using two diametric blood pressure waveform measurements
Hahn, J., Reisner, A. T., & Asada, H. H. (2010) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2010.04.017 · Cited by 30
Piezoelectric approaches for wearable continuous blood pressure monitoring: a review
Yi, Z., Zhang, W., & Yang, B. T. (2022) · Journal of Micromechanics and Microengineering · doi:10.1088/1361-6439/ac87ba · Cited by 29
Xu, D. Z., Ryan, K. L., Rickards, C. A., Zhang, G., Convertino, V. A., & Mukkamala, R. (2011) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.00443.2011 · Cited by 29
Development of central and peripheral smoking effects over time
Hasenfratz, M., Nil, R., & Bättig, K. (1990) · Psychopharmacology · doi:10.1007/bf02244054 · Cited by 29
Balmer, J., Pretty, C. G., Davidson, S. M., Desaive, T., Kamoi, S., Pironet, A., Morimont, P., Janssen, N., Lambermont, B. C., Shaw, G. M., et al. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aada72 · Cited by 28
The Role of Pulse Transit Time as an Index of Arterial Stiffness During Exercise
Kounalakis, S. N., & Geladas, N. D. (2009) · Cardiovascular Engineering An International Journal · doi:10.1007/s10558-009-9081-4 · Cited by 28
Remote Blood Pressure Estimation via the Spatiotemporal Mapping of Facial Videos
Chen, Y., Zhuang, J., Li, B., Zhang, Y., & Zheng, X. (2023) · Sensors · doi:10.3390/s23062963 · Cited by 27
An Arterial to Peripheral Pulse Wave Velocity Measure
Jennings, J. R., & Choi, S. (1983) · Psychophysiology · doi:10.1111/j.1469-8986.1983.tb00922.x · Cited by 27
Blood pressure prediction from speech recordings
Ankışhan, H. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2019.101842 · Cited by 25
Long, N. M. H., & Chung, W. (2022) · IEEE photonics journal · doi:10.1109/jphot.2022.3153506 · Cited by 24
Metadata
- Authors
- Geddes, L. A., Voelz, M. H., Babbs, C. F., Bourland, J. D., & Tacker, W. A.
- Affiliations
- Purdue University West Lafayette (US)
- Corresponding author
- Geddes, L. A.
- Venue
- Psychophysiology
- Publisher
- Wiley
- Published
- 1981-01
- Volume
- 18
- Issue
- 1
- Pages
- 71-74
- ISSN
- 0048-5772 (ISSN-L)
- 1469-8986
- OpenAlex
- W2163704183
- Cited by
- 293
- References
- 5
- Citations by year
- 2026: 10 · 2025: 13 · 2024: 13 · 2023: 14 · 2022: 15 · 2021: 22 · 2020: 22 · 2019: 18 · 2018: 23 · 2017: 12 · 2016: 14 · 2015: 8 · 2014: 11 · 2013: 3 · 2012: 8
- FWCI
- 3.13
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Psychophysiology (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Keywords
- pulse transit time, arterial blood pressure, diastolic blood pressure, pulse arrival time, pulse wave velocity, epinephrine, vagus nerve stimulation, dogs
- MeSH terms
- Animals, Blood Pressure, Dogs, Electrocardiography, Pulse
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review