A neural interface provides long-term stable natural touch perception
Open accessThe DOI matches its Crossref record; 29 of the 29 references it cites were found in a registry or an index; 5 of the 23 sources its claims cite were checked for support.
Abstract
Touch perception on the fingers and hand is essential for fine motor control, contributes to our sense of self, allows for effective communication, and aids in our fundamental perception of the world. Despite increasingly sophisticated mechatronics, prosthetic devices still do not directly convey sensation back to their wearers. We show that implanted peripheral nerve interfaces in two human subjects with upper limb amputation provided stable, natural touch sensation in their hands for more than 1 year. Electrical stimulation using implanted peripheral nerve cuff electrodes that did not penetrate the nerve produced touch perceptions at many locations on the phantom hand with repeatable, stable responses in the two subjects for 16 and 24 months. Patterned stimulation intensity produced a sensation that the subjects described as natural and without "tingling," or paresthesia. Different patterns produced different types of sensory perception at the same location on the phantom hand. The two subjects reported tactile perceptions they described as natural tapping, constant pressure, light moving touch, and vibration. Changing average stimulation intensity controlled the size of the percept area; changing stimulation frequency controlled sensation strength. Artificial touch sensation improved the subjects' ability to control grasping strength of the prosthesis and enabled them to better manipulate delicate objects. Thus, electrical stimulation through peripheral nerve electrodes produced long-term sensory restoration after limb loss.
Overview
CoChat Summary
TL;DR
Implanted nonpenetrating peripheral nerve cuff electrodes provided stable, natural touch sensations in the phantom hands of two upper-limb amputees for more than 1 year. Patterned stimulation intensity produced natural pressure, tapping, vibration, and moving touch, and sensory feedback improved grasping control and confidence.
Key findings
- Two subjects with upper limb amputation received implanted nerve cuff electrodes and reported stable touch percepts for 16 and 24 months; stimulation produced sensation from 19 of 20 channels at 15 unique locations in subject 1 and 14 of 16 channels at 9 unique locations in subject 2.
- Constant stimulation intensity typically produced paresthesia, described as “electrical,” in 96% of 151 trials over 10 months; time-variant pulse-width modulation instead produced perceptions described as natural, including pulsing pressure, constant pressure, light moving touch, tapping, and vibration.
- Perceived pressure intensity correlated with maximum pulse width (P < 0.05, R² = 0.85, n = 22), and matched pressure sensation correlated with log stimulation frequency in subject 1 (ANOVA, P < 0.05, R² = 0.70, n = 25).
- In a cherry-stem pulling task, success improved from 43% without sensory or audiovisual feedback to 93% with sensory feedback alone (P < 0.001, n = 15 per condition), and from 77% with audiovisual feedback alone to 100% with both (P < 0.005); self-reported confidence was higher with sensory feedback (one-tailed t test, P = 0.0305).
Methods
- Two male subjects with upper limb amputation received nonpenetrating peripheral nerve cuff electrodes: subject 1 had cuffs on median, radial, and ulnar nerves in the mid-forearm (20 channels total), and subject 2 had cuffs on median and radial nerves in the mid-upper arm (16 channels total).
- During monthly 4- to 6-hour visits, experimenters delivered trains of electrical pulses, identified sensory thresholds, and had subjects describe and trace perceived sensations on a hand schematic; stimulation parameters included pulse amplitude, pulse width, frequency, and time-variant pulse-width modulation.
- Functional testing included a cherry-stem pulling task under four sensory/audiovisual feedback conditions, force matching with a sensor, and pain surveys using the Trinity Amputation and Prosthesis Experience Scales (TAPES).
Limitations
- The study reports only two human subjects, so findings are based on a very small sample and no controlled comparison group.
- The paper states that the phantom pain findings “warrant more rigorous investigation” and that direct comparison with prior natural tactile sensation work is difficult because natural sensation had not been explicitly quantified or detailed.
- The paper notes that higher-density contact arrays may further improve hand coverage; the text also makes evident that many outcomes rely on subjective verbal reports, although some tests blinded subjects to stimulation level.
Model-written summary of a shortened version of the paper’s full text (it is longer than we read). Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
4 Valid · 1 Not valid · 12 Not checked · 82 Own results and interpretations
Every source the paper cites for this claim was checked and supports it.
In the paper
“Nerve electrodes activate populations of axons on the basis of size and location relative to the stimulating contact (23).”
What the sources say
- Supportedcites [23] matched to “Probabilistic modeling of selective stimulation of the human sciatic nerve with a flat interface nerve electrode” (2012) source
In Methods §2.2 (page unavailable), the paper models axons with varying diameters and positions relative to a stimulating contact, then determines which propagate an action potential. Results §3.1 also describes the assumption that larger-diameter axons are recruited before smaller ones. Together, these passages support the paraphrase for the nerve electrode studied. (LLM, conf=0.88)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“When the peak pulse width was set to a first noticeable level, B = Bth, the sensation was described as the back of a pen repeatedly pushing “very lightly” on a small, localized area of the skin (Fig. 3A).”
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“There was a psychometric correlation between perceived intensity of the pressure pulse and the maximum pulse width.”
plainstates an associationn = 22 ·- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Of the 16 available channels implanted in subject 2, stimulation produced sensation from 14 channels at 9 unique locations (fig. S1).”
plainstates a causeSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The tapping matched the frequency of the intensity modulation, fmod, as fmod varied between 0.5 and 4 Hz.”
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
61 more claims
Create a free account to see all 99 claims and every source verdict.
- Probably not a substantial claim. 10
- It may state two separate claims as one. 24
- The quote only partly states this claim. 1
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“It is proven that FINE and the spiral cuff electrode, when used for stimulation of the afferent axons, produce stable and somatotopically selective results in human subjects in long-term applications, providing appropriate sensory information to the prosthesis wearer [73].”
From Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review (2020) · cites this paper as [73]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Complex qualities of touch such as vibration on the skin, tingling, tapping, stinging, brushing, and itch have been observed during experimentation (Tan et al., 2014).”
From Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review. (2017) · cites this paper as Tan et al., 2014
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“and that long-term stability and selectivity has demonstrated using FINE (Ortiz-Catalan et al., 2014; Tan et al., 2014).”
From Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review. (2017) · cites this paper as Tan et al., 2014
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Stimulation of the nerve has been shown to induce a variety of sensations with complex qualities, such as tapping on the skin, vibration, tingling, stinging, brushing, and itch (Clark et al., 2014;Tan et al., 2015;2014).”
From Biomimetic approaches to bionic touch through a peripheral nerve interface (2015) · cites this paper as Tan et al., 2014
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Differences in evoked sensations across electrodes likely reflect differences in the submodality distribution of afferents near the electrode, as evidenced by the fact that stimulation through electrodes with glabrous and non-glabrous receptive fields elicits distinct sensations, reflecting differences in the submodalities that innervate these skin regions (Tan et al., 2014).”
From Biomimetic approaches to bionic touch through a peripheral nerve interface (2015) · cites this paper as Tan et al., 2014
Experimental — accuracy not yet measured
14 more claims about this paper
Create a free account to see every claim other papers make about this one.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated major: it would touch a result — a value a result rests on, reported and recomputed p-values on opposite sides of the significance level, two passages that conflict with no sign which is right, or a cited work retracted before this paper was published.
In the paper
§ 151 trials over a 10-month period. We systematically mapped the sensory perceptions
“When sensory feedback was not provided, the subject successfully plucked 43 and 77% of the cherries without (S−/AV−) or with (S−/AV+) audiovisual feedback, respectively.”
§ 151 trials over a 10-month period. We systematically mapped the sensory perceptions
“n = 15 per condition”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ RESULTS
“stimulation produced sensation from 19 channels at 15 unique locations on the perceived hand”
§ DISCUSSION
“There was independent control of 19 different locations on the hand with only three implanted cuffs”
Citations
Cites
Synaesthesia in phantom limbs induced with mirrors
(1996) · Proceedings of the Royal Society of London. Series B: Biological Sciences · doi:10.1098/rspb.1996.0058 · Cited by 1,246
Restoring natural sensory feedback in real-time bidirectional hand prostheses
Raspopovic, S., Capogrosso, M., Petrini, F. M., Bonizzato, M., Rigosa, J., Di Pino, G., Carpaneto, J., Controzzi, M., Boretius, T., Fernandez, E., et al. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3006820 · Cited by 1,001
A spiral nerve cuff electrode for peripheral nerve stimulation
Naples, G. G., Mortimer, J. T., Scheiner, A., & Sweeney, J. D. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.8670 · Cited by 415 · cited 2×
“Here, nonpenetrating peripheral nerve cuff electrodes (14, 15) were shown to provide multiple tactile perceptions at multiple locations on”
BackgroundMethodDouble nerve intraneural interface implant on a human amputee for robotic hand control
Rossini, P. M., Micera, S., Benvenuto, A., Carpaneto, J., Cavallo, G., Citi, L., Cipriani, C., Denaro, L., Denaro, V., Di Pino, G., et al. (2010) · Clinical Neurophysiology · doi:10.1016/j.clinph.2010.01.001 · Cited by 396 · cited 4×
“There is mounting evidence that peripheral nerves are somatotopically organized (24–26), and hence, multi-channel cuff electrodes are able to produce somatotopically selective results similar to those reported in acute trials with intrafascicular electrodes (5, 12, 16).”
BackgroundResultInfluentialDhillon, G. S., Lawrence, S. M., Hutchinson, D. T., & Horch, K. W. (2004) · The Journal of Hand Surgery · doi:10.1016/j.jhsa.2004.02.006 · Cited by 354 · cited 2×
“There is mounting evidence that peripheral nerves are somatotopically organized (24–26), and hence, multi-channel cuff electrodes are able to produce somatotopically selective results similar to those reported in acute trials with intrafascicular electrodes (5, 12, 16).”
ResultRobotic touch shifts perception of embodiment to a prosthesis in targeted reinnervation amputees
Marasco, P. D., Kim, K., Colgate, J. E., Peshkin, M. A., & Kuiken, T. A. (2011) · Brain · doi:10.1093/brain/awq361 · Cited by 335 · cited 1×
“Sensory perception is important for control of the prosthetic limb, a sense of embodiment (3), and in the reduction of phantom limb pain (4, 5).”
BackgroundAugurelle, A.-S., Smith, A. M., Lejeune, T., & Thonnard, J.-L. (2003) · Journal of Neurophysiology · doi:10.1152/jn.00249.2002 · Cited by 330
The Importance of the Sense of Touch in Virtual and Real Environments
Robles-De-La-Torre, G. (2006) · IEEE Multimedia · doi:10.1109/mmul.2006.69 · Cited by 311
PARÆSTHESIÆ FROM ECTOPIC IMPULSE GENERATION IN HUMAN SENSORY NERVES
OCHOA, J. L., & TOREJÖK, H. E. (1980) · Brain · doi:10.1093/brain/103.4.835 · Cited by 213
Horch, K., Meek, S., Taylor, T. G., & Hutchinson, D. T. (2011) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2011.2162635 · Cited by 194 · cited 1×
“single-subject case studies (5, 18) and warrant more rigorous investigation into the benefits of sensory feedback and neural interfaces in the management of phantom pain.”
Background
Research works printed but not linked: 1
- C. D. Murray in Psychoprosthetics P. Gallagher D. Desmond M. MacLachlan Eds. (Springer London 2008) pp. 119–129.
Cited by
Items 601–700 of 746
High-fidelity interfacing for bionic rehabilitation
Vujaklija, I. (2024) · Elsevier eBooks · doi:10.1016/b978-0-443-23987-8.00003-1 · Cited by 3
Zhu, C., Esteller, R., Block, J., Lechleiter, K. M., Frey, R., & Moffitt, M. A. (2024) · Frontiers in Pain Research · doi:10.3389/fpain.2023.1339892 · Cited by 2
“Peripheral stimulation for a prosthetic neural interface using time varying dynamic pulses has been shown to induce sensation that was perceived as more natural than the “ tingling ” paresthesia (33, 34).”
BackgroundOur research path toward the restoration of natural sensations in hand prostheses
Micera, S., & Shokur, S. (2024) · Artificial Organs · doi:10.1111/aor.14823 · Cited by 7
Kiele, P., Laengle, G., Schmoll, M., Pasluosta, C. F., Pfeifer, R., Schüettler, M., Aszmann, O. C., & Stieglitz, T. (2024) · IEEE Open Journal of Engineering in Medicine and Biology · doi:10.1109/ojemb.2024.3477267 · Cited by 3
Visual Neuroprostheses for Impaired Human Nervous System: State‐of‐the‐Art and Future Outlook
Mahadiuzzaman, A. S. M., Hoque, M. E., Alam, S., Chawdhury, Z. T., Hasan, M., & Rashid, A. B. (2024) · International Journal of Cell Biology · doi:10.1155/ijcb/2651763 · Cited by 5
Scarpelli, A., Cordella, F., & Zollo, L. (2025) · Journal of Neural Engineering · doi:10.1088/1741-2552/ade503 · Cited by 2
“In the first evidence [46], the algorithm was based on microneurography studies of sensation, which showed that axonal firing frequency encodes intensity of pressure sensation [74].”
InfluentialSakurada, K., Kondo, R., Nakamura, F., Fukuoka, M., Kitazaki, M., & Sugimoto, M. (2022) · Augmented Humans International Conference (AHs) · doi:10.1145/3519391.3522754 · Cited by 3
“These results showed that the robotic limbs with high SoA and SoBO improved dexterous performance [16, 27, 29].”
ResultHuman–machine interfaces expand the functionality of prosthetic limbs
Wood, H. B. (2014) · Nature Reviews Neurology · doi:10.1038/nrneurol.2014.212 · Cited by 2
Automatic hand phantom map detection methods
Huang, H., Li, T., Antfolk, C., Bruschini, C. E., Enz, C., Justiz, J., & Koch, V. M. (2015) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2015.7348315 · Cited by 2
“Invasive feedback stimulates the central nervous system using cortical electrodes [8] or the peripheral nervous system using for example cuff electrodes [9], [10] or transversal intrafascicular multichannel electrodes [11].”
MethodVLSI circuits for bidirectional interface to peripheral and visceral nerves
Greenwald, E., Wang, Q., & Thakor, N. V. (2015) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7318818 · Cited by 2
“both sensory input as well as the motor output [1].”
Texture Recognition Using Force Sensitive Resistors
Sayed, M. M. M., García, J. C. D., & Alboul, L. (2016) · Lecture notes in computer science · doi:10.1007/978-3-319-40379-3_30 · Cited by 2
“The user reported the ability to perceive “texture” of surfaces [8].”
BackgroundNeural electrodes for long-term tissue interfaces
del Valle, J., Rodríguez‐Meana, B., & Navarro, X. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-822828-9.00009-5 · Cited by 2
Bjånes, D. A., & Moritz, C. T. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-822828-9.00018-6 · Cited by 2
Engels, L. F., Cappello, L., Fischer-Janzen, A., & Cipriani, C. (2021) · PLoS ONE · doi:10.1371/journal.pone.0256753 · Cited by 3
Sensory Percepts Elicited by Chronic Macro-Sieve Electrode Stimulation of the Rat Sciatic Nerve
Chandra, N. S., McCarron, W. M., Yan, Y., Ruiz, L. C., Sallinger, E. G., Birenbaum, N. K., Burton, H., Green, L., Moran, D. W., Ray, W. Z., et al. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.758427 · Cited by 2
Katic, N., Valle, G. D., & Raspopović, S. (2022) · Handbook of Neuroengineering · doi:10.1007/978-981-15-2848-4_100-1 · Cited by 3
Gozzi, N., & Valle, G. D. (2023) · Elsevier eBooks · doi:10.1016/b978-0-443-18498-7.00014-4 · Cited by 4
Design and assessment of stimulation parameters for a novel peripheral nerve interface
Patrick, E. E., Currlin, S., Kundu, A., Delgado, F., Fahmy, A., Madler, R., Maghari, N., Bashirullah, R., Gunduz, A., & Otto, K. J. (2018) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2018.8513582 · Cited by 3
“have published and have a patent on a “population encoding” scheme that was able to encode pressure sensation by small-scale sinusoidal modulation of the pulse width of a constantamplitude pulse train [7].”
BackgroundEvaluation of Time-Domain Features of Sensory ENG Signals
Silveira, C. R. A., Brunton, E., Khushaba, R. N., & Nazarpour, K. (2018) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2018.8512798 · Cited by 2
“in such devices could substantially increase their functionality, usability, and acceptability [1]–[7].”
BackgroundNguyen, T. D., & Timko, B. P. (2017) · Wiley eBooks · doi:10.1002/9783527689934.ch21 · Cited by 4
Discrimination of Neural Sources using Spatiotemporal Signatures in Nerve Cuff Electrode Recordings
Koh, R. G. L. (2019) · TSpace (University of Toronto) · Cited by 2
Somatosensory signals from the controllers of an extra robotic finger support motor learningPreprint
Amoruso, E., Dowdall, L., Kollamkulam, M., Ukaegbu, O. C., Kieliba, P., Ng, T., Dempsey‐Jones, H., Clode, D., & Makin, T. R. (2021) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2021.05.18.444661 · Cited by 2
“…the neural pathways originally supporting the sensory function, e.g. the peripheral nerves (Dhillon and Horch, 2005; Raspopovic et al., 2014; Tan et al., 2014), the spinal cord (Chandrasekaran et al., 2020), or the somatosensory cortex (Romo et al., 1998; Tabot et al., 2013; see Bensmaia &…”
BackgroundDisruption of Proprioceptive Information During Electrical Stimulation of the Cutaneous AfferentsPreprint
Katic, N., Balaguer, J.-M., Gorskii, O., Pavlova, N., Карал-оглы, Д. Д., Bulgin, D. V., Orlov, S., Евгеньевич, М. П., Raspopović, S., & Capogrosso, M. (2021) · Research Square · doi:10.21203/rs.3.rs-1104392/v1 · Cited by 2
Chandra, N. S., Birenbaum, N. K., Green, L., Ray, W. Z., & MacEwan, M. R. (2022) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2022.109528 · Cited by 1
Valle, G. D., Carpaneto, J., & Micera, S. (2023) · Handbook of Neuroengineering · doi:10.1007/978-981-16-5540-1_27 · Cited by 5
Jabbari, K., Orfahli, L. M., & Iorio, M. L. (2025) · Artificial Intelligence Surgery · doi:10.20517/ais.2024.77 · Cited by 1
An Ultrathin and Lightweight Soft Inflatable Actuator for Natural Tactile Sensory Feedback
Scherer, H., Iberite, F., Caserta, G., Boccardo, N., Carpaneto, J., Gruppioni, E., Micera, S., & Proietti, T. (2025) · Advanced Intelligent Systems · doi:10.1002/aisy.202500262 · Cited by 4
The Cutting Edge of Peripheral Nerve Surgery: A Look at Emerging Techniques
Garozzo, D., Llorian, E., Henriques, V. M., Sanches, G. E., Torrão, F., & Guedes, F. (2025) · Springer eBooks · doi:10.1007/978-3-031-92445-3_6 · Cited by 1
Pan, L., Lun, J., Ren, Z., Zhao, H., Mu, R., & Li, J. (2025) · Journal of Bionic Engineering · doi:10.1007/s42235-025-00810-9 · Cited by 2
From Restoration to Augmentation: New Approaches to Haptic Feedback for Artificial Limbs
Dowdall, L., Foster, C., & Makin, T. R. (2025) · IEEE Transactions on Haptics · doi:10.1109/toh.2025.3641707 · Cited by 1
“Peripheral nerve implants have been developed using two main approaches: nerve cuffs that are placed around the nerve, and transversal multichannel intrafascicular electrodes (TIMEs) that penetrate the nerves [11], [12].”
Perilesional neuromodulation replaces lost sensorimotor function in persons with spinal cord injury
Calvert, J. S., Parker, S. R., Govindarajan, L. N., Darie, R., Shaaya, E. A., Solinsky, R. J., Valle, L. M. D., Miranda, P., Jang, J., Tiwari, E., et al. (2026) · Nature Biomedical Engineering · doi:10.1038/s41551-026-01627-5 · Cited by 9
So, A., Pejkovic, I., Devecıoğlu, İ., Aplin, F. P., Birznieks, I., & Vickery, R. M. (2026) · Journal of Applied Physiology · doi:10.1152/japplphysiol.01172.2025 · Cited by 2
Neural Interface Activates a Sense of Touch in Prosthetic Limbs
Talan, J. (2014) · Neurology Today · doi:10.1097/01.nt.0000457142.45399.80 · Cited by 1
Artificial Limbs for Upper Extremity Amputation
Marasco, P. D., Hebert, J. S., & Orzell, B. M. (2015) · Springer eBooks · doi:10.1007/978-1-4939-2671-8_49 · Cited by 1
Directionally-sensitive peripheral nerve recording: Bipolar nerve cuff design
Sabetian, P., Popović, M. R., & Yoo, P. B. (2016) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7592139 · Cited by 1
Optimisation of the spatial discretisation of myelinated axon models
Juan, M. C., Kölbl, F., & Sepulveda, F. (2016) · Computer Science and Electronic Engineering Conference · doi:10.1109/ceec.2016.7835916 · Cited by 1
“Recently, attempts at enhancing such prosthetics by providing sensory feedback have been successfully conducted [1] [2] [3].”
BackgroundAnalysis of temporal firing patterns of primary afferent C-fibers for different sensations in mice
Cho, K., Jang, J. H., Kim, S.-P., Choi, J., Song, M. K., Kim, I. Y., Jung, S. J., & Jang, D. P. (2017) · International Journal of Precision Engineering and Manufacturing · doi:10.1007/s12541-017-0088-2 · Cited by 1
Liquid Metal as Electronic Medium to Recover Damaged Nerves’ Function
Liu, J., & Yi, L. (2018) · Springer series in biomaterials science and engineering · doi:10.1007/978-981-10-5607-9_9 · Cited by 1
Silveira, C. R. A., Brunton, E., Spendiff, S., & Nazarpour, K. (2018) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2018.8512751 · Cited by 1
Biomimetic bidirectional hand neuroprostheses for restoring somatosensory and motor functions
Iberite, F., Mendez, V., Mazzoni, A., Shokur, S., & Micera, S. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-822828-9.00011-3 · Cited by 2
Smart sensing and actuators for people with hand motion impairment
Izdrui, D., Geman, O., Postolache, O. A., Elahi, H., Vicoveanu, D. I., & Curea, S. (2022) · Elsevier eBooks · doi:10.1016/b978-0-323-85174-9.00002-9 · Cited by 1
Neuromorphic Tactile Sensing and Encoding
Kumar, D., Silva, A. N., & Thakor, N. V. (2021) · Handbook of Neuroengineering · doi:10.1007/978-981-15-2848-4_117-1 · Cited by 2
Artificial Sensory Feedback to the Brain: Somatosensory Feedback for Neural Devices and BCI
Bjånes, D. A., & Moritz, C. T. (2023) · Handbook of Neuroengineering · doi:10.1007/978-981-16-5540-1_111 · Cited by 1
Somatosensory Neuromodulation with a Focus Towards Clinical Systems
Graczyk, E. L., & Tyler, D. J. (2023) · Handbook of Neuroengineering · doi:10.1007/978-981-16-5540-1_92 · Cited by 1
Self-citationA Self-powered Neural Stimulator Based on Programmable Triboelectric Nanogenerators
Yue, W., Yu, S., Guo, T., & Wang, H. (2023) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc40787.2023.10340669 · Cited by 1
“Modulation of neural signals using implantable bioelectronics, which includes neuroprosthesis for restoring sensory/motor functions [1-2] and bioelectronic medicine for peripheral visceral nerves to control bodily functions or treat diseases [3-4], are intensively investigated as a promising…”
BackgroundHarnoncourt, L., Gstoettner, C., Laengle, G., Boesendorfer, A., & Aszmann, O. C. (2024) · Handchirurgie · Mikrochirurgie · Plastische Chirurgie · doi:10.1055/a-2260-9842 · Cited by 3
“Erste Ansätze diesbezüglich sind zurzeit Gegenstand der Forschung [36, 37], eine Lösung zur Integration einer ansatzweise na-türlichen, multimodalen Sensibilität mit akzeptabler Latenzzeit in die prothetische Versorgung ist zum jetzigen Zeitpunkt allerdings noch nicht absehbar.”
BackgroundEmergence of Nerve Interfaces with Robotic Applications
Burke, K. L., Dumanian, G. A., & Cederna, P. S. (2024) · Springer eBooks · doi:10.1007/978-3-031-57132-9_6 · Cited by 1
Pitkin, M. R., Park, H., Frossard, L. A., Klishko, A. N., & Prilutsky, B. I. (2024) · Military Medicine · doi:10.1093/milmed/usae149 · Cited by 1
Jakes, R. S., Mesias, L., Santos, V. J., et al. (2024) · doi:10.1109/telepresence63209.2024.10841536 · Cited by 1
Self-citationA biosynthetic thumb prosthesis
Bansal, S., Lai, G. V., Belingheri, M., Kashay, A. Q., Kim, J., Tomkinson, A., Herman, S., Bollavaram, K., Zukotynski, B. K., Thomas, S., et al. (2025) · npj Biomedical Innovations. · doi:10.1038/s44385-025-00031-z · Cited by 1
Antonello, M., Collu, R., Roberto, P., Andrea, D., Cordella, F., Zollo, L., Massimo, B., & Cosseddu, P. (2025) · Scientific Reports · doi:10.1038/s41598-025-21599-x · Cited by 2
Utilising reinforcement learning to develop strategies for driving auditory neural implants
Lee, G. W., Zambetta, F., Li, X., & Paolini, A. G. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/4/046027 · Cited by 2
Skeletal Motor Neuroprostheses
Kilgore, K. L., Kirsch, R. F., & Peckham, P. H. (2017) · Series on bioengineering and biomedical engineering · doi:10.1142/9789813207158_0016 · Cited by 1
Optimizing User Integration for Individualized Rehabilitation
Nataraj, R. (2018) · InTech eBooks · doi:10.5772/intechopen.70267 · Cited by 1
“Such an example is a prosthetic hand following amputation that connects mechanical sensors to transduce physical sensations to neural activations and allow users to perceive touch or movement [7, 8].”
BackgroundInfluentialA Computational Model for the Design of Lower-Limb Sensorimotor Neuroprostheses
Raspopović, S., & Petrini, F. M. (2018) · Biosystems & biorobotics · doi:10.1007/978-3-030-01845-0_10 · Cited by 2
Zhang, P., Li, X., Hu, D., Lai, Q., Wang, Y. Y., Ma, X., Xu, Q., Li, W., Huang, J., & He, J. (2019) · Advances in experimental medicine and biology · doi:10.1007/978-981-13-2050-7_4 · Cited by 1
One-step Implantation of a 3D Neural Microelectrode Array
Yim, S., Jeong, J., Ihn, Y., Hwang, D., Yang, S., Oh, S., & Kim, K. (2020) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc44109.2020.9175643 · Cited by 1
FeetBack–Redirecting touch sensation from a prosthetic hand to the human foot
Morand, R., Brusa, T., Schnüriger, N., Catanzaro, S., Berli, M., & Koch, V. M. (2022) · Frontiers in Neuroscience · doi:10.3389/fnins.2022.1019880 · Cited by 1
EEG Feature in Amputee with Five Projected Fingers: A Comparison Study
Liang, W., Qin, C., Dong, Y., Tang, X., Li, G., Fang, P., & Bi, S. (2022) · IEEE International Conference on Robotics and Biomimetics (ROBIO) · doi:10.1109/robio55434.2022.10011722 · Cited by 1
“Second, in view of the ways of the interaction, the restoration of sensory rebuild [8], [9], [12], [16], [17] focuses on connecting the stimulus by means of the reserved natural nerve system, and the restoration of sensory substitution [13], [14] focus on building a new sensory stream to the brain somatosensory area.”
BackgroundKatic, N., Valle, G. D., & Raspopović, S. (2023) · Handbook of Neuroengineering · doi:10.1007/978-981-16-5540-1_100 · Cited by 1
US regulators move on thought-controlled prosthetics
Reardon, S. (2014) · Nature · doi:10.1038/515476a · Cited by 1
Electrical system for interfacing with muscle for use in prosthetic devices
Keyes, S. R. (2017) · DSpace@MIT (Massachusetts Institute of Technology) · Cited by 1
Providing Non-Invasive Sensory Feedback for Transradial Prosthetic Hand Users
Stephens-Fripp, B. (2020) · Research Online (University of Wollongong) · Cited by 1
Cortical reorganization in the adult primary sensorimotor cortex
Kikkert, S., Root, V., Buehler, S., & Makin, T. R. (2022) · Elsevier eBooks · doi:10.1016/b978-0-12-820480-1.00004-8 · Cited by 2
Bionics: Creating the Twenty-Four Million Dollar Man or Woman
Stratmann, H. G. (2015) · Science and fiction · doi:10.1007/978-3-319-16015-3_14
Method of Fabrication for Nerve Cuff Electrodes for Use in Animal Models
Sanner, B. M. (2015) · SMARTech Repository (Georgia Institute of Technology)
Acute changes associated with electrode insertion measured with optical coherence microscopy
Hammer, D. X., Lozzi, A., Boretsky, A. R., Agrawal, A., & Welle, C. G. (2016) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2213073
Artificial arms get closer to the real thing
Shen, H. (2014) · Nature · doi:10.1038/nature.2014.16111 · Cited by 1
Analysis of the efficacy of EPIONE therapies to treat phantom limb pain
Comoglio, C. C. (2017) · Purdue e-Pubs (Purdue University System) · doi:10.7912/c2594m
“Prior research in the field suggests a potential successful treatment may require the incorporation of sensory feedback with cognitive reinforcement [174].”
BackgroundAlgorithms for Neural Prosthetic Applications
Padmanaban, S. (2017)
Utilization of fNIRS signals as feedback for sensory decoding
Bhutta, M. R., Woo, S.-W., Yoo, S.-H., Park, J., & Hong, K. (2017) · International Conference on Control, Automation and Systems (ICCAS) · doi:10.23919/iccas.2017.8204392
Neuroprostheses for Restoring Sensation
Tyler, D. J. (2018) · Elsevier eBooks · doi:10.1016/b978-0-12-805353-9.00103-0 · Cited by 1
Self-citationNeural Prostheses for Reaching and Grasping
Scherberger, H. (2018) · Elsevier eBooks · doi:10.1016/b978-0-12-809324-5.23377-4
Norrlid, J., Enander, J. M. D., Mogensen, H., & Jörntell, H. (2019) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/810770
Popovic, I., Igor, S., Popovic, S., & Vekić, B. (2019) · Serbian Journal of Experimental and Clinical Research · doi:10.2478/sjecr-2019-0065
Page, D. M., Wendelken, S., Davis, T. S., Kluger, D. T., Hutchinson, D. T., George, J. A., & Clark, G. A. (2020) · arXiv (Cornell University) · doi:10.48550/arxiv.2003.03626
“Past studies indicate that the rate of nominal habituation varies as a function of stimulus frequency and inversely as a function of stimulus strength for neural interface stimulation [14], [52], suggesting that algorithms that use either frequency or charge per pulse to encode stimulus intensity may have different co…”
BackgroundGranata, G., Jensen, W., Divoux, J., Guiraud, D., Micera, S., Navarro, X., Stieglitz, T., Yoshida, K., & Rossini, P. M. (2022) · River Publishers eBooks · doi:10.1201/9781003337973-10
Shoffstall, A. J., & Capadona, J. R. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816137-1.00073-8
Directionally Sensitive Low-noise Neural Recording Using Tetrapolar Nerve Cuff Electrode
Sabetian, P. (2020) · TSpace (University of Toronto) · Cited by 2
A multi-channel peripheral nerve stimulator with integrate-and-fire encoding
Kundu, A., Fahmy, A., Madler, R., Otto, K. J., Patrick, E. E., Príncipe, J. C., Maghari, N., & Bashirullah, R. (2021) · Journal of Medical Engineering & Technology · doi:10.1080/03091902.2021.1891311
“amplitude [3] or pulse width [15] has been effective in”
BackgroundStimulus interaction in transcutaneous electrical stimulation
Dupan, S. S. G., Jabban, L., Metcalfe, B., & Nazarpour, K. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-822828-9.00014-9
A Software Tool for the Real-Time in Vivo Evaluation of Neural Electrodes' Selectivity
Strauss, I., Luca, D. D., Panarese, A. M., Bernini, F., Gabisonia, K., Petrini, F. M., Recchia, F. A., Raspopović, S., & Micera, S. (2021) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner49283.2021.9441334 · Cited by 1
“Other groups restored sensory feedback using epineural electrodes such as FINEs or cuff electrodes [5], [6].”
MethodWells, E. D. (2021) · ERA: Education and Research Archive (University of Alberta) · doi:10.7939/r3-h59b-p194
Modern approaches of signal processing for bidirectional neural interfaces
Cimolato, A., Katic, N., & Raspopović, S. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-822828-9.00016-2 · Cited by 1
Kyberd, P. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-820544-0.00020-2
Kyberd, P. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-820544-0.00015-9
Katic, N., Balaguer, J.-M., Gorskii, O., Pavlova, N., Liang, L., Ho, J. C., Grigsby, E. M., Gerszten, P. C., Карал-оглы, Д. Д., Bulgin, D. V., et al. (2021) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2021.11.19.469209
“Peripheral nerve stimulation of upper-limb amputees using CUFF 15,16 , FINE 7,17 , TIME 18,19 , Wire LIFE 20 – 22 , tf - LIFE 16,23 and USEA 8,22,24,25 electrodes.”
BackgroundMethodMablekos-Alexiou, A., Kontogiannopoulos, S., Bertos, G. A., & Papadopoulos, E. G. (2021) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2021.103454 · Cited by 1
(2021) · Elsevier eBooks · doi:10.1016/b978-0-12-820544-0.00025-1
Osborn, L. E., Betthauser, J., & Thakor, N. V. (2019) · Wiley eBooks · doi:10.1002/047134608x.w1424.pub2 · Cited by 4
Peripheral nervous system control for neuroprostheses
Buil, J. (2017) · eDiss (Georg-August-Universität Göttingen) · doi:10.53846/goediss-6491
Kellmeyer, P. (2019) · J.B. Metzler eBooks · doi:10.1007/978-3-476-05604-7_51
Opinions on noninvasive sensory feedback of upper limb prosthetic users
Smither, F. C., Andrews, K. L., Scrabeck, T. L., Lennon, R. J., & Zhao, K. D. (2022) · Prosthetics and Orthotics International · doi:10.1097/pxr.0000000000000160 · Cited by 2
A Skin-Inspired, Self-Powered Tactile SensorPreprint
Kim, S., Lee, S., & Park, J. (2022) · SSRN Electronic Journal · doi:10.2139/ssrn.4124165
Bionic Human: A Review of Interface Modalities for Externally Powered Prosthetic Limbs
Mundkur, N. (2019) · McGill Science Undergraduate Research Journal · doi:10.26443/msurj.v14i1.53 · Cited by 1
“However, this retrograde transfer of information has proven to be considerably more challenging than in the usual human-to-prosthesis direction (23).”
BackgroundBionic limb replacement: an evolving concept in lower extremity reconstruction
Amin, K. (2022) · Plastic and Aesthetic Research · doi:10.20517/2347-9264.2021.119 · Cited by 1
Soghoyan, G. A., Рамилевич, Б. А., Matvienko, Y., Chekh, I., Sintsov, M., & Lebedev, M. (2022) · SSRN Electronic Journal · doi:10.31219/osf.io/jwqtx · Cited by 4
Metadata
- Authors
- Tan, D. W., Schiefer, M. A., Keith, M. W., Anderson, J. R., Tyler, J., & Tyler, D. J.
- Affiliations
- Louis Stokes Cleveland VA Medical Center (US)
- MetroHealth Medical Center (US)
- Corresponding author
- Tan, Daniel W.
- Schiefer, Matthew A.
- Tyler, Dustin J.
- Venue
- Science Translational Medicine
- Published
- 2014-10-08
- Volume
- 6
- Issue
- 257
- Pages
- 257ra138-257ra138
- ISSN
- 1946-6234 (ISSN-L)
- 1946-6242
- PMC
- PMC5517305
- OpenAlex
- W2157094999
- Semantic Scholar
- 308d067b912e90f2bd3958d5248e2d960435ea5c
- Cited by
- 751
- References
- 29
- Citations by year
- 2026: 24 · 2025: 41 · 2024: 57 · 2023: 59 · 2022: 81 · 2021: 75 · 2020: 74 · 2019: 87 · 2018: 82 · 2017: 64 · 2016: 64 · 2015: 37 · 2014: 5
- FWCI
- 25.89
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Science Translational Medicine (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- U.S. Department of Veterans Affairs (I01 RX001334, IK1 RX000724)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Cognitive Neuroscience › Tactile and Sensory Interactions
- Health Sciences › Medicine › Anesthesiology and Pain Medicine › Pain Management and Treatment
- Keywords
- peripheral nerve interfaces, electrical stimulation, tactile perception, upper limb amputation, prosthetic hand, sensory feedback, phantom hand, long-term stability, paresthesia, grasp control, fine motor control, sensory recovery
- MeSH terms
- Amputees, Artificial Limbs, Electric Stimulation, Electrodes, Implanted, Hand, Humans, Male, Middle Aged, Motor Skills, Paresthesia, Peripheral Nerves, Phantom Limb, Pressure, Sensation, Touch Perception, Feedback, Sensory, Translational Research, Biomedical
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-07, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-07 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.