A neural interface provides long-term stable natural touch perception
Open accessThe DOI matches its Crossref record; 29 of the 29 references it cites were found in a registry or an index; 5 of the 23 sources its claims cite were checked for support.
Abstract
Touch perception on the fingers and hand is essential for fine motor control, contributes to our sense of self, allows for effective communication, and aids in our fundamental perception of the world. Despite increasingly sophisticated mechatronics, prosthetic devices still do not directly convey sensation back to their wearers. We show that implanted peripheral nerve interfaces in two human subjects with upper limb amputation provided stable, natural touch sensation in their hands for more than 1 year. Electrical stimulation using implanted peripheral nerve cuff electrodes that did not penetrate the nerve produced touch perceptions at many locations on the phantom hand with repeatable, stable responses in the two subjects for 16 and 24 months. Patterned stimulation intensity produced a sensation that the subjects described as natural and without "tingling," or paresthesia. Different patterns produced different types of sensory perception at the same location on the phantom hand. The two subjects reported tactile perceptions they described as natural tapping, constant pressure, light moving touch, and vibration. Changing average stimulation intensity controlled the size of the percept area; changing stimulation frequency controlled sensation strength. Artificial touch sensation improved the subjects' ability to control grasping strength of the prosthesis and enabled them to better manipulate delicate objects. Thus, electrical stimulation through peripheral nerve electrodes produced long-term sensory restoration after limb loss.
Overview
CoChat Summary
TL;DR
Implanted nonpenetrating peripheral nerve cuff electrodes provided stable, natural touch sensations in the phantom hands of two upper-limb amputees for more than 1 year. Patterned stimulation intensity produced natural pressure, tapping, vibration, and moving touch, and sensory feedback improved grasping control and confidence.
Key findings
- Two subjects with upper limb amputation received implanted nerve cuff electrodes and reported stable touch percepts for 16 and 24 months; stimulation produced sensation from 19 of 20 channels at 15 unique locations in subject 1 and 14 of 16 channels at 9 unique locations in subject 2.
- Constant stimulation intensity typically produced paresthesia, described as “electrical,” in 96% of 151 trials over 10 months; time-variant pulse-width modulation instead produced perceptions described as natural, including pulsing pressure, constant pressure, light moving touch, tapping, and vibration.
- Perceived pressure intensity correlated with maximum pulse width (P < 0.05, R² = 0.85, n = 22), and matched pressure sensation correlated with log stimulation frequency in subject 1 (ANOVA, P < 0.05, R² = 0.70, n = 25).
- In a cherry-stem pulling task, success improved from 43% without sensory or audiovisual feedback to 93% with sensory feedback alone (P < 0.001, n = 15 per condition), and from 77% with audiovisual feedback alone to 100% with both (P < 0.005); self-reported confidence was higher with sensory feedback (one-tailed t test, P = 0.0305).
Methods
- Two male subjects with upper limb amputation received nonpenetrating peripheral nerve cuff electrodes: subject 1 had cuffs on median, radial, and ulnar nerves in the mid-forearm (20 channels total), and subject 2 had cuffs on median and radial nerves in the mid-upper arm (16 channels total).
- During monthly 4- to 6-hour visits, experimenters delivered trains of electrical pulses, identified sensory thresholds, and had subjects describe and trace perceived sensations on a hand schematic; stimulation parameters included pulse amplitude, pulse width, frequency, and time-variant pulse-width modulation.
- Functional testing included a cherry-stem pulling task under four sensory/audiovisual feedback conditions, force matching with a sensor, and pain surveys using the Trinity Amputation and Prosthesis Experience Scales (TAPES).
Limitations
- The study reports only two human subjects, so findings are based on a very small sample and no controlled comparison group.
- The paper states that the phantom pain findings “warrant more rigorous investigation” and that direct comparison with prior natural tactile sensation work is difficult because natural sensation had not been explicitly quantified or detailed.
- The paper notes that higher-density contact arrays may further improve hand coverage; the text also makes evident that many outcomes rely on subjective verbal reports, although some tests blinded subjects to stimulation level.
Model-written summary of a shortened version of the paper’s full text (it is longer than we read). Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
4 Valid · 1 Not valid · 12 Not checked · 82 Own results and interpretations
Every source the paper cites for this claim was checked and supports it.
In the paper
“Nerve electrodes activate populations of axons on the basis of size and location relative to the stimulating contact (23).”
What the sources say
- Supportedcites [23] matched to “Probabilistic modeling of selective stimulation of the human sciatic nerve with a flat interface nerve electrode” (2012) source
In Methods §2.2 (page unavailable), the paper models axons with varying diameters and positions relative to a stimulating contact, then determines which propagate an action potential. Results §3.1 also describes the assumption that larger-diameter axons are recruited before smaller ones. Together, these passages support the paraphrase for the nerve electrode studied. (LLM, conf=0.88)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“When the peak pulse width was set to a first noticeable level, B = Bth, the sensation was described as the back of a pen repeatedly pushing “very lightly” on a small, localized area of the skin (Fig. 3A).”
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“There was a psychometric correlation between perceived intensity of the pressure pulse and the maximum pulse width.”
plainstates an associationn = 22 ·- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Of the 16 available channels implanted in subject 2, stimulation produced sensation from 14 channels at 9 unique locations (fig. S1).”
plainstates a causeSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The tapping matched the frequency of the intensity modulation, fmod, as fmod varied between 0.5 and 4 Hz.”
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
61 more claims
Create a free account to see all 99 claims and every source verdict.
- Probably not a substantial claim. 10
- It may state two separate claims as one. 24
- The quote only partly states this claim. 1
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“It is proven that FINE and the spiral cuff electrode, when used for stimulation of the afferent axons, produce stable and somatotopically selective results in human subjects in long-term applications, providing appropriate sensory information to the prosthesis wearer [73].”
From Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review (2020) · cites this paper as [73]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Complex qualities of touch such as vibration on the skin, tingling, tapping, stinging, brushing, and itch have been observed during experimentation (Tan et al., 2014).”
From Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review. (2017) · cites this paper as Tan et al., 2014
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“and that long-term stability and selectivity has demonstrated using FINE (Ortiz-Catalan et al., 2014; Tan et al., 2014).”
From Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review. (2017) · cites this paper as Tan et al., 2014
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Stimulation of the nerve has been shown to induce a variety of sensations with complex qualities, such as tapping on the skin, vibration, tingling, stinging, brushing, and itch (Clark et al., 2014;Tan et al., 2015;2014).”
From Biomimetic approaches to bionic touch through a peripheral nerve interface (2015) · cites this paper as Tan et al., 2014
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Differences in evoked sensations across electrodes likely reflect differences in the submodality distribution of afferents near the electrode, as evidenced by the fact that stimulation through electrodes with glabrous and non-glabrous receptive fields elicits distinct sensations, reflecting differences in the submodalities that innervate these skin regions (Tan et al., 2014).”
From Biomimetic approaches to bionic touch through a peripheral nerve interface (2015) · cites this paper as Tan et al., 2014
Experimental — accuracy not yet measured
14 more claims about this paper
Create a free account to see every claim other papers make about this one.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated major: it would touch a result — a value a result rests on, reported and recomputed p-values on opposite sides of the significance level, two passages that conflict with no sign which is right, or a cited work retracted before this paper was published.
In the paper
§ 151 trials over a 10-month period. We systematically mapped the sensory perceptions
“When sensory feedback was not provided, the subject successfully plucked 43 and 77% of the cherries without (S−/AV−) or with (S−/AV+) audiovisual feedback, respectively.”
§ 151 trials over a 10-month period. We systematically mapped the sensory perceptions
“n = 15 per condition”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ RESULTS
“stimulation produced sensation from 19 channels at 15 unique locations on the perceived hand”
§ DISCUSSION
“There was independent control of 19 different locations on the hand with only three implanted cuffs”
Citations
Cites
Synaesthesia in phantom limbs induced with mirrors
(1996) · Proceedings of the Royal Society of London. Series B: Biological Sciences · doi:10.1098/rspb.1996.0058 · Cited by 1,246
Restoring natural sensory feedback in real-time bidirectional hand prostheses
Raspopovic, S., Capogrosso, M., Petrini, F. M., Bonizzato, M., Rigosa, J., Di Pino, G., Carpaneto, J., Controzzi, M., Boretius, T., Fernandez, E., et al. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3006820 · Cited by 1,001
A spiral nerve cuff electrode for peripheral nerve stimulation
Naples, G. G., Mortimer, J. T., Scheiner, A., & Sweeney, J. D. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.8670 · Cited by 415 · cited 2×
“Here, nonpenetrating peripheral nerve cuff electrodes (14, 15) were shown to provide multiple tactile perceptions at multiple locations on”
BackgroundMethodDouble nerve intraneural interface implant on a human amputee for robotic hand control
Rossini, P. M., Micera, S., Benvenuto, A., Carpaneto, J., Cavallo, G., Citi, L., Cipriani, C., Denaro, L., Denaro, V., Di Pino, G., et al. (2010) · Clinical Neurophysiology · doi:10.1016/j.clinph.2010.01.001 · Cited by 396 · cited 4×
“There is mounting evidence that peripheral nerves are somatotopically organized (24–26), and hence, multi-channel cuff electrodes are able to produce somatotopically selective results similar to those reported in acute trials with intrafascicular electrodes (5, 12, 16).”
BackgroundResultInfluentialDhillon, G. S., Lawrence, S. M., Hutchinson, D. T., & Horch, K. W. (2004) · The Journal of Hand Surgery · doi:10.1016/j.jhsa.2004.02.006 · Cited by 354 · cited 2×
“There is mounting evidence that peripheral nerves are somatotopically organized (24–26), and hence, multi-channel cuff electrodes are able to produce somatotopically selective results similar to those reported in acute trials with intrafascicular electrodes (5, 12, 16).”
ResultRobotic touch shifts perception of embodiment to a prosthesis in targeted reinnervation amputees
Marasco, P. D., Kim, K., Colgate, J. E., Peshkin, M. A., & Kuiken, T. A. (2011) · Brain · doi:10.1093/brain/awq361 · Cited by 335 · cited 1×
“Sensory perception is important for control of the prosthetic limb, a sense of embodiment (3), and in the reduction of phantom limb pain (4, 5).”
BackgroundAugurelle, A.-S., Smith, A. M., Lejeune, T., & Thonnard, J.-L. (2003) · Journal of Neurophysiology · doi:10.1152/jn.00249.2002 · Cited by 330
The Importance of the Sense of Touch in Virtual and Real Environments
Robles-De-La-Torre, G. (2006) · IEEE Multimedia · doi:10.1109/mmul.2006.69 · Cited by 311
PARÆSTHESIÆ FROM ECTOPIC IMPULSE GENERATION IN HUMAN SENSORY NERVES
OCHOA, J. L., & TOREJÖK, H. E. (1980) · Brain · doi:10.1093/brain/103.4.835 · Cited by 213
Horch, K., Meek, S., Taylor, T. G., & Hutchinson, D. T. (2011) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2011.2162635 · Cited by 194 · cited 1×
“single-subject case studies (5, 18) and warrant more rigorous investigation into the benefits of sensory feedback and neural interfaces in the management of phantom pain.”
Background
Research works printed but not linked: 1
- C. D. Murray in Psychoprosthetics P. Gallagher D. Desmond M. MacLachlan Eds. (Springer London 2008) pp. 119–129.
Cited by
Items 401–500 of 746
Neuroengineering: Deciphering neural drive
Ortiz-Catalan, M. (2017) · Nature Biomedical Engineering · doi:10.1038/s41551-017-0034 · Cited by 13
Vascular and Neural Response to Focal Vibration, Sensory Feedback, and Piezo Ion Channel Signaling
Penasso, H., Petersen, F., & Peternell, G. (2023) · Journal of Vascular Diseases · doi:10.3390/jvd2010006 · Cited by 21
Cesari, V., Melfi, F. M. A., Gemignani, A., & Menicucci, D. (2023) · Heliyon · doi:10.1016/j.heliyon.2023.e21665 · Cited by 14
“…favors the enhancement of performance in the following clinical/non clinical context: motor learning using a virtual limb in perceptual decision-making task [48], control of prosthesis [49], prosthetic object discrimination and manipulation [50], and prosthetic manual accuracy and sensitivity [51].”
BackgroundTang, W., Lu, X., Chen, S., Ge, S., Jing, X., Wang, X., Liu, R., & Zhu, H. J. (2019) · Journal of the Textile Institute · doi:10.1080/00405000.2019.1661067 · Cited by 16
Hammer, D. X., Lozzi, A., Boretsky, A. R., & Welle, C. G. (2016) · Neurophotonics · doi:10.1117/1.nph.3.2.025002 · Cited by 13
Natural Sensations Evoked in Distal Extremities Using Surface Electrical Stimulation
Slopsema, J. P., Boss, J. M., Heyboer, L. A., Tobias, C. M., Draggoo, B. P., Finn, K. E., Hoff, P. J., & Polasek, K. H. (2018) · The Open Biomedical Engineering Journal · doi:10.2174/1874120701812010001 · Cited by 16
“The first two waveforms were based on promising results using electrodes implanted around the nerve [8].”
BackgroundInfluentialKang, Y. N., Chou, N., Jang, J., Byun, D., Kang, H., Moon, D.-J., Kim, J., & Kim, S. (2019) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2019.2917916 · Cited by 15
Joint Speed Discrimination and Augmentation For Prosthesis Feedback
Earley, E. J., Johnson, R. E., Hargrove, L. J., & Sensinger, J. W. (2018) · Scientific Reports · doi:10.1038/s41598-018-36126-4 · Cited by 12
Temporal Modulation of the Response of Sensory Fibers to Paired-Pulse Stimulation
Brunton, E., Silveira, C. R. A., Rosenberg, J. V., Schiefer, M. A., Riddell, J. S., & Nazarpour, K. (2019) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2019.2935813 · Cited by 13
“A sense of touch is vital when it comes to interacting and experiencing the world around us [1], [2].”
BackgroundResultInfluentialSelf-citationKrom, B. N., Catoire, M., Toet, A., van Dijk, R. J. E., & van Erp, J. B. F. (2019) · IEEE World Haptics Conference (WHC) · doi:10.1109/whc.2019.8816112 · Cited by 18
“Ownership over a robotic or prosthetic hand can increase dexterous performance [1], [2], [3].”
BackgroundKarczewski, A. M., Dingle, A. M., & Poore, S. O. (2021) · Frontiers in Neurorobotics · doi:10.3389/fnbot.2021.711028 · Cited by 13
MXene-Based Microneedle Electrode for Brain–Computer Interface in Diverse Scenarios
Chen, Y., Fan, Z., Shi, N., Cheng, B., Huang, C., Liu, X., Gao, X., & Liu, R. (2025) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.5c03798 · Cited by 16
Electrospun nanofibers: promising nanomaterials for biomedical applications
Mohamed, M. A., El‐Rahman, M. K. A., & Mousavi, M. P. S. (2023) · Royal Society of Chemistry eBooks · doi:10.1039/bk9781839169366-00225 · Cited by 14
Katic, N., Balaguer, J.-M., Gorskii, O., Pavlova, N., Liang, L., Ho, J. C., Grigsby, E. M., Gerszten, P. C., Карал-оглы, Д. Д., Bulgin, D. V., et al. (2024) · Cell Reports · doi:10.1016/j.celrep.2024.113695 · Cited by 16
Sensory feedback in upper limb prosthetics: advances and challenges
Malešević, N., & Antfolk, C. (2024) · Nature Reviews Neurology · doi:10.1038/s41582-024-00987-3 · Cited by 18
“In a study of this approach published around 10 years ago, an interface consisting of flat interface electrodes was implanted in the median and ulnar nerves of two people with upper limb amputations for a period of more than 1 year 5 .”
Flexible sling electrode for bidirectional neural signal recording and selective stimulation
Lee, S., Yen, S., Liao, L., Gammad, G. G. L., Thakor, N. V., & Lee, C. (2016) · IEEE International Conference on Micro Electro Mechanical Systems (MEMS) · doi:10.1109/memsys.2016.7421639 · Cited by 12
A Microclip Peripheral Nerve Interface (μcPNI) for Bioelectronic Interfacing with Small Nerves
Rowan, C. C., Graudejus, O., & Otchy, T. M. (2021) · Advanced Science · doi:10.1002/advs.202102945 · Cited by 12
Luo, Q., Niu, C. M., Chou, C.-H., Liang, W., Deng, X., Hao, M., & Lan, N. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.783505 · Cited by 13
“…recruitment of spinal motoneurons (Henneman et al., 1965), to generation of muscle forces to move joints (Bernstein, 1967; McNeill Alexander, 2002; Valero-Cuevas et al., 2007), and finally the formation of proprioceptive feedback to inform the brain of motor consequences (Weiler et al., 2019;…”
BackgroundMotion intention recognition of the affected hand based on the sEMG and improved DenseNet network
Niu, Q., Shi, L., Niu, Y., Jia, K., Fan, G., Gui, R., & Wang, L. (2024) · Heliyon · doi:10.1016/j.heliyon.2024.e26763 · Cited by 12
“Several recent studies, including Ortiz-Catalan, M [5]and Tan, D. W. et al. [6], have demonstrated the presence of sEMG in residual muscle tissue even in cases of physical disability or amputation, unless the amputation is accompanied by nerve damage or other motor deficits.”
ResultGaetani, F., Zappatore, G. A., Visconti, P., et al. (2018) · International Conference on IC Design & Technology (ICICDT) · doi:10.1109/icicdt.2018.8399787 · Cited by 12
“Refined electrical impulses are sent to the central nervous system using other electrodes which are surgically implanted into the upper arm nerves: when the patient grips an object, he is able to feel the strength of his grasp as well as the object shape, roughness, consistency, etc [4].”
BackgroundInnovations in Electrical Stimulation Harness Neural Plasticity to Restore Motor Function
Peng, X., Hickman, J. L., Bowles, S. G., Donegan, D., & Welle, C. G. (2018) · Bioelectronics in Medicine · doi:10.2217/bem-2019-0002 · Cited by 13
Vibration Induced Proprioceptive Modulation in Surface-EMG Based Control of a Robotic Arm
Rangwani, R., & Park, H. (2019) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2019.8717117 · Cited by 12
Dingle, A. M., Ness, J. P., Novello, J., Israel, J. S., Sanchez, R., Millevolte, A. X. T., Brodnick, S. K., Krugner-Higby, L. A., Nemke, B. W., Lü, Y., et al. (2019) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108504 · Cited by 12
Huggins, J. E., Guger, C., Aarnoutse, E. J., Allison, B. Z., Anderson, C. W., Bedrick, S. D., Besio, W. G., Chavarriaga, R., Collinger, J. L., Do, A. H., et al. (2019) · Brain-Computer Interfaces · doi:10.1080/2326263x.2019.1697163 · Cited by 18
“The sensations are appropriate in perceived magnitude and are described as having tactile and proprioceptive qualities [163,166].”
BackgroundČvančara, P., Valle, G., Müller, M., Guiho, T., Hiairrassary, A., Petrini, F., Raspopovic, S., Strauss, I., Granata, G., Fernandez, E., et al. (2019) · doi:10.1101/653964 · Cited by 12
The Next Frontier in Neuroprosthetics: Integration of Biomimetic Somatosensory Feedback
Tian, Y., Valle, G. D., Cederna, P. S., & Kemp, S. W. P. (2025) · Biomimetics · doi:10.3390/biomimetics10030130 · Cited by 18
“Numerous implanted electrodes have been developed and used to stimulate neurons for somatosensory feedback [24,35,37,46,102–106]. somatosensory feedback [24,35,37,46,102–106].”
Scarpelli, A., Demofonti, A., Cordella, F., Coffa, U., Mereu, F., Gruppioni, E., & Zollo, L. (2024) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2024.3443398 · Cited by 14
“On one hand, invasive feedback systems demonstrated to be an effective solution to elicit tactile and proprioceptive sensations in amputees [14], [15], [16], [17].”
BackgroundHighly Flexible Precisely Braided Multielectrode Probes and Combinatorics for Future Neuroprostheses
Kim, T., Schmidt, K., Deemie, C., Wycech, J., Liang, H., & Giszter, S. F. (2019) · Frontiers in Neuroscience · doi:10.3389/fnins.2019.00613 · Cited by 9
Novel Neurostimulation-Based Haptic Feedback Platform for Grasp Interactions With Virtual Objects
Shell, A. K., Pena, A. E., Abbas, J. J., & Jung, R. (2022) · Frontiers in Virtual Reality · doi:10.3389/frvir.2022.910379 · Cited by 16
“While intensity is encoded by rate and population recruitment, modality seems to be encoded by the spatiotemporal patterning of this activity (Tan et al., 2014).”
BackgroundPeripheral Nerve Stimulation for the Treatment of Postamputation Pain: A Systematic Review
Smith, B. J., Twohey, E. E., Dean, K. P., & D'Souza, R. S. (2023) · American Journal of Physical Medicine & Rehabilitation · doi:10.1097/phm.0000000000002237 · Cited by 14
“Of the studies involving neuroprosthetics, two used TIMEs,(35,36) and one study used flat interface nerve electrodes.(37) Petrini et al.”
BackgroundMethodInfluentialTissue Engineering: Then, Now, and the Future
Caplan, A. I. (2019) · Tissue Engineering Part A · doi:10.1089/ten.tea.2019.0011 · Cited by 11
“Indeed, such prosthesis now have small servomotors that serve as contractors instead of live innervated muscle tissue.(2,3) Yet no tissue-engineered muscle is available to replace excised or inoperative tissue.”
BackgroundManoharan, S., & Park, H. (2019) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2019.8717127 · Cited by 14
“The bi-directional system presented in this paper is designed to provide a person with artificial tactile feedback from the surface of racket through electrical stimulation [12],[13] along with intact visual feedback, which fulfills the criteria of multidimensional sensory feedback.”
BackgroundBrain reactions to the use of sensorized hand prosthesis in amputees
Granata, G., Iorio, R. D., Miraglia, F., Caulo, M., Iodice, F., Vecchio, F., Valle, G., Strauss, I., D’Anna, E., Iberite, F., et al. (2020) · Brain and Behavior · doi:10.1002/brb3.1734 · Cited by 11
“…of devices giving to the amputee the ongoing possibility of performing complex movements and receiving real-time somatosensory feedback in an open loop (Raspopovic et al., 2014; Schiefer, Tan, Sidek, & Tyler, 2016; Tan, Schiefer, Keith, Anderson, & Tyler, 2015; Tan et al., 2014; Tyler, 2015).”
BackgroundSensory Feedback in Upper Limb Prostheses
Dimante, D., Logina, I., Sinisi, M., & Krūmiņa, A. (2020) · Proceedings of the Latvian Academy of Sciences Section B Natural Exact and Applied Sciences · doi:10.2478/prolas-2020-0047 · Cited by 15
“In fact, an increased number of synchronously activated fascicles can lead to unnatural sensation and paraesthesia (Tan et al., 2014).”
BackgroundGuiho, T., López-Álvarez, V. M., Čvančara, P., Hiairrassary, A., Andreu, D., Stieglitz, T., Navarro, X., & Guiraud, D. (2021) · Sensors · doi:10.3390/s21217219 · Cited by 13
“During this trial, stimulating with TIME electrodes improved the richness and accuracy of the evoked sensations compared to those elicited via multicontact cuff electrodes [14,15] in upper limb amputees.”
BackgroundMethodZhang, P., Zhang, M., Li, C., Liu, S., & Zhang, F. (2022) · Neural Regeneration Research · doi:10.4103/1673-5374.335838 · Cited by 12
“In addition, a humanmachine interface has been used for receiving and translating neural electrical signals in amputation patients to assist with movement (Ortiz-Catalan et al., 2014; Raspopovic et al., 2014; Tan et al., 2014; Oddo et al., 2016; Valle et al., 2018; Petrini et al., 2019) and to feed back touch and prop…”
BackgroundMethodTactile display on the remaining hand for unilateral hand amputees
Li, T., Huang, H., Antfolk, C., Justiz, J., & Koch, V. M. (2016) · Current Directions in Biomedical Engineering · doi:10.1515/cdbme-2016-0089 · Cited by 11
“With the recent breakthroughs, invasive tactile feedback approaches to provide tactile feedback [2, 3] would provide”
BackgroundA cosmetic prosthetic digit with bioinspired embedded touch feedback
Barone, D., D’Alonzo, M., Controzzi, M., Clemente, F., & Cipriani, C. (2017) · International Conference on Rehabilitation Robotics · doi:10.1109/icorr.2017.8009402 · Cited by 15
“Besides the possibility of improving the closed-loop control of a prosthesis [5], [6] sensory feedback was shown functional in reducing phantom limb pain [7].”
BackgroundLambrecht, J. M., Cady, S. R., Peterson, E. J., Dunning, J. L., Dinsmoor, D. A., Pape, F., Graczyk, E. L., & Tyler, D. J. (2024) · Journal of Neural Engineering · doi:10.1088/1741-2552/ad56c9 · Cited by 12
“Sinusoidal modulation of PW may reduce the paresthesia frequently associated with electrical stimulation and result in a more natural pressure sensation [15].”
BackgroundMethodSelf-citationNg, K. K. W., Snow, I. N., Birznieks, I., & Vickery, R. M. (2021) · Journal of Neurophysiology · doi:10.1152/jn.00662.2020 · Cited by 12
Liang, W., Qin, C., Sun, A., Zhang, X., Lan, N., & Bi, S. (2022) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2022.3229271 · Cited by 11
“Second, considering the type of restoration, the artificial interface could have two types, sensory rebuild [15], [16], [17], [18], [21] and sensory substitution [22], [23].”
BackgroundBiorealistic hand prosthesis with compliance control and noninvasive somatotopic sensory feedback
Lan, N., Zhang, J., Zhang, Z., Chou, C.-H., Rymer, W. Z., Niu, C. M., & Fang, P. (2023) · Progress in Biomedical Engineering · doi:10.1088/2516-1091/acc625 · Cited by 11
“…can transmit sensory information to amputees by implanted regenerative electrodes [89], intraneural electrodes (such as the longitudinal intrafascicular electrode [90], transverse intrafascicular multichannel electrode [91], Utah slanted electrode array [92], and nerve cuff electrodes [93, 94]).”
Hybrid Bionic Nerve Interface for Application in Bionic Limbs
Cho, Y., Jeong, H. H., Shin, H., Pak, C. J., Cho, J., Kim, Y., Kim, D.-G., Kim, T. K., Kim, H., Kim, S., et al. (2023) · Advanced Science · doi:10.1002/advs.202303728 · Cited by 11
Resnik, L. J., Benz, H. L., Borgia, M. L., & Clark, M. A. (2019) · Expert Review of Medical Devices · doi:10.1080/17434440.2019.1619453 · Cited by 10
And Yet It Moves: What We Currently Know about Phantom Arm Movements
Scaliti, E., Gruppioni, E., & Becchio, C. (2020) · The Neuroscientist · doi:10.1177/1073858420904326 · Cited by 13
Graham, R. D., Bruns, T. M., Duan, B., & Lempka, S. F. (2020) · Neuromodulation Technology at the Neural Interface · doi:10.1111/ner.13211 · Cited by 16
Towards optimizing the non-invasive sensory feedback interfaces in a neural prosthetic control
Su, S., Chai, G., Meng, J., Sheng, X., Mouraux, A., & Zhu, X. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac4e1b · Cited by 17
Palatal Electrotactile Display Outperforms Visual Display in Tongue Motor Learning
Jiang, B., Kim, J., & Park, H. (2022) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2022.3156398 · Cited by 10
Ultrathin Conformable Electronic Tattoo for Tactile Sensations
Mazzotta, A., & Mattoli, V. (2023) · Advanced Electronic Materials · doi:10.1002/aelm.202201327 · Cited by 10
Spike detection: The first step towards an ENG-based neuroprosheses
Noce, E., Ciancio, A. L., & Zollo, L. (2018) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2018.07.008 · Cited by 12
Dingle, A. M., Ness, J. P., Novello, J., Millevolte, A. X. T., Zeng, W., Sanchez, R., Nemke, B. W., Lü, Y., Suminski, A. J., Markel, M. D., et al. (2020) · Military Medicine · doi:10.1093/milmed/usz246 · Cited by 10
Liu, M., Batista, A. P., Bensmaïa, S. J., & Weber, D. J. (2020) · Journal of Neurophysiology · doi:10.1152/jn.00725.2019 · Cited by 11
Berman, A., Shi, B., Zaluska, T. J., Yong, A., Clees, S., Xu, C., Beker, L., & Bao, Z. (2025) · npj Flexible Electronics · doi:10.1038/s41528-025-00503-7 · Cited by 13
High-precision all-in-one dual robotic arm strategy in oral implant surgery
Tang, G., Liu, S., Sun, M., et al. (2024) · BDJ Open · doi:10.1038/s41405-024-00231-6 · Cited by 9
“Extensive research efforts have enabled medical robots to be used in a variety of industries, including laparoscopic intestinal anastomosis [7, 8], tumor ablation [9], assistive wearable robots [10 – 12], and capsule robots [13, 14].”
BackgroundCady, S. R., Lambrecht, J. M., Dsouza, K. T., Dunning, J. L., Anderson, J. R., Malone, K. J., Chepla, K. J., Graczyk, E. L., & Tyler, D. J. (2025) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-025-01613-z · Cited by 13
“…antimicrobial baths [63]. iSens ® ‑evoked somatosensory percepts demonstrated stable qualities and a combination of stable and shifting locations Similar to previous implanted nerve stimulation studies, iSens ® -evoked somatosensory percept locations spanned various hand locations [17, 28, 32, 33].”
InfluentialSelf-citationDRG microstimulation evokes postural responses in awake, standing felines
King, K., Cusack, W. F., Nanivadekar, A. C., Ayers, C. A., Urbin, M. A., Gaunt, R. A., Fisher, L. E., & Weber, D. J. (2019) · Journal of Neural Engineering · doi:10.1088/1741-2552/ab50f4 · Cited by 12
“As most somatosensory neuroprostheses rely on electrical stimulation of mixed nerves to evoke sensations [19, 21, 22], it is unclear whether any evoked proprioceptive sensations result from direct activation of sensory afferents or are an indirect result of efferent activation giving rise to…”
Mechanism of peripheral nerve modulation and recent applications
Shin, H., Kang, M., & Lee, S. (2021) · International Journal of Optomechatronics · doi:10.1080/15599612.2021.1978601 · Cited by 9
“Since the electrode at the invasive method is adjacent to the axon, it is possible to exceed the threshold value of the axon even using less current and has higher spatial selectiveness than the non-invasive method.([42,43]) This point is also closely related to long-term implantation.”
Manoharan, S., & Park, H. (2023) · Biomedical Engineering Letters · doi:10.1007/s13534-023-00308-5 · Cited by 15
Human-in-the-Loop Control of Robotic Leg Prostheses With Sensory Feedback
Pi, M., Li, Z., Li, Q., Kang, Y., Kan, Z., & Song, R. (2023) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2023.3321403 · Cited by 10
Quality of life and reconstructive surgery efforts in severe hand injuries
Alawi, S. A., Werner, D., Könneker, S., Vogt, P. M., & Jokuszies, A. (2018) · Innovative Surgical Sciences · doi:10.1515/iss-2018-0002 · Cited by 14
“[32] Tan DW, Schiefer MA, Keith MW, Anderson JR, Tyler J, Tyler DJ.”
BackgroundMuramatsu, S., Toda, M., Nishikawa, J., & Tateno, T. (2019) · Brain Research · doi:10.1016/j.brainres.2019.146312 · Cited by 9
Spatio-temporal feature extraction in sensory electroneurographic signals
Silveira, C. R. A., Khushaba, R. N., Brunton, E., & Nazarpour, K. (2022) · Philosophical Transactions of the Royal Society A Mathematical Physical and Engineering Sciences · doi:10.1098/rsta.2021.0268 · Cited by 8
“In stimulation studies, the flat interface nerve electrode (FINE) has proved to be stable and selective over time [2,55,58,59].”
BackgroundA legal perspective on body implants for therapy and enhancement
Palmerini, E. (2015) · International Review of Law Computers & Technology · doi:10.1080/13600869.2015.1055664 · Cited by 14
“Sensate prosthetics, which convey sensation back to their wearers and produce long-term sensory restoration after limb loss, could make everyday life much easier for people with this kind of disability (Tan et al. 2014).”
Background(2022) · PubMed
Taghlabi, K. M., Cruz-Garza, J. G., Hassan, T., Potnis, O., Bhenderu, L. S. S., Guerrero, J. R., Whitehead, R. E., Wu, Y. C., Luan, L., Xie, C., et al. (2024) · Journal of Neural Engineering · doi:10.1088/1741-2552/ad200f · Cited by 9
“Better control over the prosthesis’s grasping force and handling of delicate objects was made possible by these sensations, which were described as vibration, light-moving touch, constant pressure, and natural tapping [188].”
BackgroundMethodInfluentialToward more naturalistic tactile sensors
Micera, S. (2024) · Science · doi:10.1126/science.adp2623 · Cited by 9
Ng, K. K. W., So, A., Fang, J. Y., Birznieks, I., & Vickery, R. M. (2024) · Frontiers in Neuroscience · doi:10.3389/fnins.2024.1125597 · Cited by 9
“Nevertheless, we expect that the results would also be applicable when using electrodes implanted directly in the nerve to deliver sensory feedback (Tan et al., 2014; Graczyk et al., 2016).”
BackgroundRecent advances in facilitating the translation of bioelectronic medicine therapies
Baldwin, A., States, G., Pikov, V., Gunalan, P. K., Elyahoodayan, S., Kilgore, K. L., & Meng, E. F. (2024) · Current Opinion in Biomedical Engineering · doi:10.1016/j.cobme.2024.100575 · Cited by 10
Long-term implantable flexible neural interfaces for electrophysiological monitoring
Chen, B., Lu, J., Chen, Z., Han, X., Sun, Y., Lin, X., Tang, Z., Jia, S., Xie, G., Huo, F., et al. (2025) · Journal of Materials Chemistry C · doi:10.1039/d4tc04289a · Cited by 10
Trout, M. A., Mino, F. R., Olsen, C. D., Hansen, T. C., Teramoto, M., Warren, D. J., Segil, J. L., & George, J. A. (2025) · Nature Communications · doi:10.1038/s41467-025-65965-9 · Cited by 14
Borkowska, V. R., McConnell, A. C., Vijayakumar, S., Stokes, A. A., & Roche, A. D. (2022) · Frontiers in Rehabilitation Sciences · doi:10.3389/fresc.2022.806479 · Cited by 10
“They carry a number of risks, such as nerve damage (29) and paraesthesia (30), have been tested on a limited number of volunteers and face much reluctance from amputees (31).”
BackgroundA feasibility study on AI-controlled closed-loop electrical stimulation implants
Eickhoff, S., García-Agúndez, A., Haidar, D., Zaidat, B., Adjei-Mosi, M., Li, P., & Eickhoff, C. (2023) · Scientific Reports · doi:10.1038/s41598-023-36384-x · Cited by 12
“Although those existing systems have demonstrated successful and reliable stimulation performance over several years of implantation 2–4 , the degree of invasiveness of placing those large devices as well as the risk of electrode lead breakage represent major challenges to ES implantation in…”
BackgroundToni, L., Pierantoni, L., Verardo, C., Romeni, S., & Micera, S. (2024) · Bioelectromagnetics · doi:10.1002/bem.22535 · Cited by 10
Digital Extensions with Bi-axial Fingertip Sensors for Supplementary Tactile Feedback Studies
Engels, L. F., Cappello, L., & Cipriani, C. (2018) · IEEE/RAS-EMBS International Conference on Biomedical Robotics and Biomechatronics (BioRob) · doi:10.1109/biorob.2018.8487812 · Cited by 8
Merged Haptic Sensation in the Hand during Concurrent Non-Invasive Proximal Nerve Stimulation
Vargas, L., Huang, H., Zhu, Y., & Hu, X. (2018) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2018.8512707 · Cited by 10
“For example, Tan et al, using peripherally implanted cuff electrode, has shown repeatable and stable sensory responses in two human amputee subjects [10-11].”
Dosen, S. S., Prahm, C., Amsuess, S., Vujaklija, I., & Farina, D. (2021) · Springer eBooks · doi:10.1007/978-3-030-60746-3_15 · Cited by 10
Liu, Y., Xi, P., Li, B., Zhang, M., Liu, H., Tang, R., Shan, X., Huang, Q., He, J., Liu, Z., et al. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac5bf6 · Cited by 8
“TENS (Johnson 2012) has been generally used to elicit artificial tactile perception in the tactile feed-back system (Antfolk et al 2013, Tan et al 2014, Valle et al 2018).”
A Prototype of a Neural, Powered, Transtibial Prosthesis for the Cat: Benchtop Characterization
Park, H., Islam, M. S., Grover, M. A., Klishko, A. N., Prilutsky, B. I., & DeWeerth, S. P. (2018) · Frontiers in Neuroscience · doi:10.3389/fnins.2018.00471 · Cited by 9
Freeberg, M. J., Ansari, R., Pinault, G. C. J., Lombardo, L. M., Miller, M. E., Tyler, D. J., & Triolo, R. J. (2019) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2019.2951079 · Cited by 8
“long-term neuroprosthesis performance [9], [14].”
BackgroundSelf-citationWang, H., Fang, P., Tian, L., Zheng, Y., Zhou, H. D., Li, G., & Zhang, X. (2015) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7319114 · Cited by 9
“nerves are possible approaches to regenerate intuitive and accurate perception feedback for limb amputees [4-6].”
BackgroundElectro-tactile Feedback System for a Prosthetic Hand
Pamungkas, D. S., & Ward, K. (2015) · Research Online (University of Wollongong) · Cited by 11
An actor-model framework for visual sensory encoding
Leong, F., Rahmani, B., Psaltis, D., Moser, C., & Ghezzi, D. (2024) · Nature Communications · doi:10.1038/s41467-024-45105-5 · Cited by 8
Directed stimulation with interfascicular interfaces for peripheral nerve stimulation
Koppaka, S., Hess-Dunning, A., & Tyler, D. J. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac33e8 · Cited by 7
Self-citationIdentifying Health-Related Quality of Life Domains After Upper Extremity Transplantation
Tulsky, D. S., Kisala, P. A., Tyner, C. E., Slotkin, J., Kaufman, C. L., Dearth, C. L., Horan, A. D., Talbot, S. G., Shores, J. T., Azari, K. K., et al. (2023) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/j.apmr.2023.01.001 · Cited by 9
Merging Humans and Neuroprosthetics through Regenerative Peripheral Nerve Interfaces
Tian, Y., Vaskov, A. K., Adidharma, W., Cederna, P. S., & Kemp, S. W. P. (2024) · Seminars in Plastic Surgery · doi:10.1055/s-0044-1779028 · Cited by 8
“31,34 However, these peripheral nerve interfaces are not chronically stable for sensory stimulation due to the complexity of surgical implantation, lack of stimulation speci fi city, and potential nerve damage.”
BackgroundNeural mechanisms underlying intracortical microstimulation for sensory restoration
Hughes, C. L., Chen, X., Grill, W. M., & Kozai, T. D. Y. (2026) · Nature Biomedical Engineering · doi:10.1038/s41551-025-01583-6 · Cited by 14
“Dynamic frequency modulation may improve perception and activity as does dynamic amplitude modulation 28,92 by breaking up synchronization of neural activity and allowing recovery of neuronal excitability, although this requires further study.”
Wang, X., Zhang, Y., Guo, T., Wu, S., Zhong, J., Cheng, C., & Sui, X. (2024) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2024.108556 · Cited by 7
Ciancio, A. L., Cordella, F., Hoffmann, K., Schneider, A., Guglielmelli, E., & Zollo, L. (2017) · Studies in applied philosophy, epistemology and rational ethics · doi:10.1007/978-3-319-66881-9_5 · Cited by 11
Implantable Device Fabrication and Packaging
Stieglitz, T. (2023) · Handbook of Neuroengineering · doi:10.1007/978-981-16-5540-1_102 · Cited by 7
Stimulation artifact rejection in closed-loop, distributed neural interfaces
Peterson, E. J., Dinsmoor, D. A., Tyler, D. J., & Denison, T. (2016) · European Solid-State Circuits Conference (ESSCIRC) · doi:10.1109/esscirc.2016.7598285 · Cited by 7
“Stimulation to restore a single sensation can require patterned stimulation at 20600 Hz and pulse widths ranging 10 to 260 us to provide natural sensation [4].”
BackgroundSelf-citationBioelectric Medicine and Devices for the Treatment of Spinal Cord Injury
Torregrosa, T., & Koppes, R. A. (2016) · Cells Tissues Organs · doi:10.1159/000446698 · Cited by 8
“Relaying crude indications of grasp pressure back to the user has an unprecedented impact on functionality, enabling delicate tasks [Tan et al., 2014].”
BackgroundHargrove, L. J. (2017) · Springer eBooks · doi:10.1007/978-1-4939-7247-0_8 · Cited by 7
Functional Frequency Discrimination From Cortical Somatosensory Stimulation in Humans
Kramer, D. R., Lamorie‐Foote, K., Barbaro, M. F., Lee, M., Peng, T. D., Gogia, A. S., Liu, C. Y., Kellis, S., & Lee, B. (2019) · Frontiers in Neuroscience · doi:10.3389/fnins.2019.00832 · Cited by 6
Kolarcik, C. L., Castro, C. A., Lesniak, A. J., Demetris, A. J., Fisher, L. E., Gaunt, R. A., Weber, D. J., & Cui, X. T. (2020) · Journal of Neural Engineering · doi:10.1088/1741-2552/ab94d7 · Cited by 8
W:Ti Flexible Transversal Electrode Array for Peripheral Nerve Stimulation: A Feasibility Study
Silveira, C. R. A., Brunton, E., Escobedo‐Cousin, E., Gupta, G., Whittaker, R. G., O'Neill, A. G., & Nazarpour, K. (2020) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2020.3014812 · Cited by 7
“peripheral nervous system [2], [15]–[17]; invasive [2], [7], [8], [10], [18] vs.”
BackgroundAnxiety and the problem of “inattentive” animals in water maze tests
Bondarenko, N. (2017) · The Russian Journal of Cognitive Science · doi:10.47010/17.4.7 · Cited by 8
Artificial sensory feedback for bionic hands
Downey, J. E., Brooks, J., & Bensmaïa, S. J. (2019) · Elsevier eBooks · doi:10.1016/b978-0-12-814942-3.00008-8 · Cited by 7
Somatosensory Feedback in BCIs: Why Aiming for Naturalness Raises Ethical Concerns
van Balen, B. (2025) · AJOB Neuroscience · doi:10.1080/21507740.2025.2478427 · Cited by 6
“…(Beauchamp et al. 2020; Chen et al. 2020), tactile neural prostheses that aim to generate a sense of touch (Flesher et al. 2016; Salas et al. 2018; Tan et al. 2014) and cochlear implants that aim to generate a sense of hearing (Djourno and Eyriès 1957) by electrically stimulating relevant paths…”
Xiong, Q., Yao, H., Wang, J., Huang, Z., Luo, J., Zhong, C., Lin, Y., & Shu, L. (2025) · Frontiers in Neuroscience · doi:10.3389/fnins.2025.1519758 · Cited by 7
Su, S., Chai, G., Xu, W., Meng, J., Sheng, X., Mouraux, A., & Zhu, X. (2023) · Journal of Neural Engineering · doi:10.1088/1741-2552/acab32 · Cited by 9
Metadata
- Authors
- Tan, D. W., Schiefer, M. A., Keith, M. W., Anderson, J. R., Tyler, J., & Tyler, D. J.
- Affiliations
- Louis Stokes Cleveland VA Medical Center (US)
- MetroHealth Medical Center (US)
- Corresponding author
- Tan, Daniel W.
- Schiefer, Matthew A.
- Tyler, Dustin J.
- Venue
- Science Translational Medicine
- Published
- 2014-10-08
- Volume
- 6
- Issue
- 257
- Pages
- 257ra138-257ra138
- ISSN
- 1946-6234 (ISSN-L)
- 1946-6242
- PMC
- PMC5517305
- OpenAlex
- W2157094999
- Semantic Scholar
- 308d067b912e90f2bd3958d5248e2d960435ea5c
- Cited by
- 751
- References
- 29
- Citations by year
- 2026: 24 · 2025: 41 · 2024: 57 · 2023: 59 · 2022: 81 · 2021: 75 · 2020: 74 · 2019: 87 · 2018: 82 · 2017: 64 · 2016: 64 · 2015: 37 · 2014: 5
- FWCI
- 25.89
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Science Translational Medicine (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- U.S. Department of Veterans Affairs (I01 RX001334, IK1 RX000724)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Cognitive Neuroscience › Tactile and Sensory Interactions
- Health Sciences › Medicine › Anesthesiology and Pain Medicine › Pain Management and Treatment
- Keywords
- peripheral nerve interfaces, electrical stimulation, tactile perception, upper limb amputation, prosthetic hand, sensory feedback, phantom hand, long-term stability, paresthesia, grasp control, fine motor control, sensory recovery
- MeSH terms
- Amputees, Artificial Limbs, Electric Stimulation, Electrodes, Implanted, Hand, Humans, Male, Middle Aged, Motor Skills, Paresthesia, Peripheral Nerves, Phantom Limb, Pressure, Sensation, Touch Perception, Feedback, Sensory, Translational Research, Biomedical
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-07, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-07 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.