A neural interface provides long-term stable natural touch perception
Open accessThe DOI matches its Crossref record; 29 of the 29 references it cites were found in a registry or an index; 5 of the 23 sources its claims cite were checked for support.
Abstract
Touch perception on the fingers and hand is essential for fine motor control, contributes to our sense of self, allows for effective communication, and aids in our fundamental perception of the world. Despite increasingly sophisticated mechatronics, prosthetic devices still do not directly convey sensation back to their wearers. We show that implanted peripheral nerve interfaces in two human subjects with upper limb amputation provided stable, natural touch sensation in their hands for more than 1 year. Electrical stimulation using implanted peripheral nerve cuff electrodes that did not penetrate the nerve produced touch perceptions at many locations on the phantom hand with repeatable, stable responses in the two subjects for 16 and 24 months. Patterned stimulation intensity produced a sensation that the subjects described as natural and without "tingling," or paresthesia. Different patterns produced different types of sensory perception at the same location on the phantom hand. The two subjects reported tactile perceptions they described as natural tapping, constant pressure, light moving touch, and vibration. Changing average stimulation intensity controlled the size of the percept area; changing stimulation frequency controlled sensation strength. Artificial touch sensation improved the subjects' ability to control grasping strength of the prosthesis and enabled them to better manipulate delicate objects. Thus, electrical stimulation through peripheral nerve electrodes produced long-term sensory restoration after limb loss.
Overview
CoChat Summary
TL;DR
Implanted nonpenetrating peripheral nerve cuff electrodes provided stable, natural touch sensations in the phantom hands of two upper-limb amputees for more than 1 year. Patterned stimulation intensity produced natural pressure, tapping, vibration, and moving touch, and sensory feedback improved grasping control and confidence.
Key findings
- Two subjects with upper limb amputation received implanted nerve cuff electrodes and reported stable touch percepts for 16 and 24 months; stimulation produced sensation from 19 of 20 channels at 15 unique locations in subject 1 and 14 of 16 channels at 9 unique locations in subject 2.
- Constant stimulation intensity typically produced paresthesia, described as “electrical,” in 96% of 151 trials over 10 months; time-variant pulse-width modulation instead produced perceptions described as natural, including pulsing pressure, constant pressure, light moving touch, tapping, and vibration.
- Perceived pressure intensity correlated with maximum pulse width (P < 0.05, R² = 0.85, n = 22), and matched pressure sensation correlated with log stimulation frequency in subject 1 (ANOVA, P < 0.05, R² = 0.70, n = 25).
- In a cherry-stem pulling task, success improved from 43% without sensory or audiovisual feedback to 93% with sensory feedback alone (P < 0.001, n = 15 per condition), and from 77% with audiovisual feedback alone to 100% with both (P < 0.005); self-reported confidence was higher with sensory feedback (one-tailed t test, P = 0.0305).
Methods
- Two male subjects with upper limb amputation received nonpenetrating peripheral nerve cuff electrodes: subject 1 had cuffs on median, radial, and ulnar nerves in the mid-forearm (20 channels total), and subject 2 had cuffs on median and radial nerves in the mid-upper arm (16 channels total).
- During monthly 4- to 6-hour visits, experimenters delivered trains of electrical pulses, identified sensory thresholds, and had subjects describe and trace perceived sensations on a hand schematic; stimulation parameters included pulse amplitude, pulse width, frequency, and time-variant pulse-width modulation.
- Functional testing included a cherry-stem pulling task under four sensory/audiovisual feedback conditions, force matching with a sensor, and pain surveys using the Trinity Amputation and Prosthesis Experience Scales (TAPES).
Limitations
- The study reports only two human subjects, so findings are based on a very small sample and no controlled comparison group.
- The paper states that the phantom pain findings “warrant more rigorous investigation” and that direct comparison with prior natural tactile sensation work is difficult because natural sensation had not been explicitly quantified or detailed.
- The paper notes that higher-density contact arrays may further improve hand coverage; the text also makes evident that many outcomes rely on subjective verbal reports, although some tests blinded subjects to stimulation level.
Model-written summary of a shortened version of the paper’s full text (it is longer than we read). Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
4 Valid · 1 Not valid · 12 Not checked · 82 Own results and interpretations
Every source the paper cites for this claim was checked and supports it.
In the paper
“Nerve electrodes activate populations of axons on the basis of size and location relative to the stimulating contact (23).”
What the sources say
- Supportedcites [23] matched to “Probabilistic modeling of selective stimulation of the human sciatic nerve with a flat interface nerve electrode” (2012) source
In Methods §2.2 (page unavailable), the paper models axons with varying diameters and positions relative to a stimulating contact, then determines which propagate an action potential. Results §3.1 also describes the assumption that larger-diameter axons are recruited before smaller ones. Together, these passages support the paraphrase for the nerve electrode studied. (LLM, conf=0.88)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“When the peak pulse width was set to a first noticeable level, B = Bth, the sensation was described as the back of a pen repeatedly pushing “very lightly” on a small, localized area of the skin (Fig. 3A).”
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“There was a psychometric correlation between perceived intensity of the pressure pulse and the maximum pulse width.”
plainstates an associationn = 22 ·- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Of the 16 available channels implanted in subject 2, stimulation produced sensation from 14 channels at 9 unique locations (fig. S1).”
plainstates a causeSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The tapping matched the frequency of the intensity modulation, fmod, as fmod varied between 0.5 and 4 Hz.”
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
61 more claims
Create a free account to see all 99 claims and every source verdict.
- Probably not a substantial claim. 10
- It may state two separate claims as one. 24
- The quote only partly states this claim. 1
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“It is proven that FINE and the spiral cuff electrode, when used for stimulation of the afferent axons, produce stable and somatotopically selective results in human subjects in long-term applications, providing appropriate sensory information to the prosthesis wearer [73].”
From Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review (2020) · cites this paper as [73]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Complex qualities of touch such as vibration on the skin, tingling, tapping, stinging, brushing, and itch have been observed during experimentation (Tan et al., 2014).”
From Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review. (2017) · cites this paper as Tan et al., 2014
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“and that long-term stability and selectivity has demonstrated using FINE (Ortiz-Catalan et al., 2014; Tan et al., 2014).”
From Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review. (2017) · cites this paper as Tan et al., 2014
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Stimulation of the nerve has been shown to induce a variety of sensations with complex qualities, such as tapping on the skin, vibration, tingling, stinging, brushing, and itch (Clark et al., 2014;Tan et al., 2015;2014).”
From Biomimetic approaches to bionic touch through a peripheral nerve interface (2015) · cites this paper as Tan et al., 2014
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Differences in evoked sensations across electrodes likely reflect differences in the submodality distribution of afferents near the electrode, as evidenced by the fact that stimulation through electrodes with glabrous and non-glabrous receptive fields elicits distinct sensations, reflecting differences in the submodalities that innervate these skin regions (Tan et al., 2014).”
From Biomimetic approaches to bionic touch through a peripheral nerve interface (2015) · cites this paper as Tan et al., 2014
Experimental — accuracy not yet measured
14 more claims about this paper
Create a free account to see every claim other papers make about this one.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated major: it would touch a result — a value a result rests on, reported and recomputed p-values on opposite sides of the significance level, two passages that conflict with no sign which is right, or a cited work retracted before this paper was published.
In the paper
§ 151 trials over a 10-month period. We systematically mapped the sensory perceptions
“When sensory feedback was not provided, the subject successfully plucked 43 and 77% of the cherries without (S−/AV−) or with (S−/AV+) audiovisual feedback, respectively.”
§ 151 trials over a 10-month period. We systematically mapped the sensory perceptions
“n = 15 per condition”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ RESULTS
“stimulation produced sensation from 19 channels at 15 unique locations on the perceived hand”
§ DISCUSSION
“There was independent control of 19 different locations on the hand with only three implanted cuffs”
Citations
Cites
Synaesthesia in phantom limbs induced with mirrors
(1996) · Proceedings of the Royal Society of London. Series B: Biological Sciences · doi:10.1098/rspb.1996.0058 · Cited by 1,246
Restoring natural sensory feedback in real-time bidirectional hand prostheses
Raspopovic, S., Capogrosso, M., Petrini, F. M., Bonizzato, M., Rigosa, J., Di Pino, G., Carpaneto, J., Controzzi, M., Boretius, T., Fernandez, E., et al. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3006820 · Cited by 1,001
A spiral nerve cuff electrode for peripheral nerve stimulation
Naples, G. G., Mortimer, J. T., Scheiner, A., & Sweeney, J. D. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.8670 · Cited by 415 · cited 2×
“Here, nonpenetrating peripheral nerve cuff electrodes (14, 15) were shown to provide multiple tactile perceptions at multiple locations on”
BackgroundMethodDouble nerve intraneural interface implant on a human amputee for robotic hand control
Rossini, P. M., Micera, S., Benvenuto, A., Carpaneto, J., Cavallo, G., Citi, L., Cipriani, C., Denaro, L., Denaro, V., Di Pino, G., et al. (2010) · Clinical Neurophysiology · doi:10.1016/j.clinph.2010.01.001 · Cited by 396 · cited 4×
“There is mounting evidence that peripheral nerves are somatotopically organized (24–26), and hence, multi-channel cuff electrodes are able to produce somatotopically selective results similar to those reported in acute trials with intrafascicular electrodes (5, 12, 16).”
BackgroundResultInfluentialDhillon, G. S., Lawrence, S. M., Hutchinson, D. T., & Horch, K. W. (2004) · The Journal of Hand Surgery · doi:10.1016/j.jhsa.2004.02.006 · Cited by 354 · cited 2×
“There is mounting evidence that peripheral nerves are somatotopically organized (24–26), and hence, multi-channel cuff electrodes are able to produce somatotopically selective results similar to those reported in acute trials with intrafascicular electrodes (5, 12, 16).”
ResultRobotic touch shifts perception of embodiment to a prosthesis in targeted reinnervation amputees
Marasco, P. D., Kim, K., Colgate, J. E., Peshkin, M. A., & Kuiken, T. A. (2011) · Brain · doi:10.1093/brain/awq361 · Cited by 335 · cited 1×
“Sensory perception is important for control of the prosthetic limb, a sense of embodiment (3), and in the reduction of phantom limb pain (4, 5).”
BackgroundAugurelle, A.-S., Smith, A. M., Lejeune, T., & Thonnard, J.-L. (2003) · Journal of Neurophysiology · doi:10.1152/jn.00249.2002 · Cited by 330
The Importance of the Sense of Touch in Virtual and Real Environments
Robles-De-La-Torre, G. (2006) · IEEE Multimedia · doi:10.1109/mmul.2006.69 · Cited by 311
PARÆSTHESIÆ FROM ECTOPIC IMPULSE GENERATION IN HUMAN SENSORY NERVES
OCHOA, J. L., & TOREJÖK, H. E. (1980) · Brain · doi:10.1093/brain/103.4.835 · Cited by 213
Horch, K., Meek, S., Taylor, T. G., & Hutchinson, D. T. (2011) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2011.2162635 · Cited by 194 · cited 1×
“single-subject case studies (5, 18) and warrant more rigorous investigation into the benefits of sensory feedback and neural interfaces in the management of phantom pain.”
Background
Research works printed but not linked: 1
- C. D. Murray in Psychoprosthetics P. Gallagher D. Desmond M. MacLachlan Eds. (Springer London 2008) pp. 119–129.
Cited by
Items 301–400 of 746
Segil, J. L., Cuberovic, I., Graczyk, E. L., Weir, R. F. F., & Tyler, D. J. (2020) · Scientific Reports · doi:10.1038/s41598-020-62970-4 · Cited by 30
Self-citationA Fully Integrated Sensor-Brain–Machine Interface System for Restoring Somatosensation
Liu, X., Zhu, H., Qiu, T., Sritharan, S. Y., Ge, D., Yang, S. M., Zhang, M., Richardson, A. G., Lucas, T. H., Engheta, N., et al. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.3030899 · Cited by 34
“The sensor output can then be encoded into the brain through electrical stimulation at a point above the injury along the neural pathway that normally processes this information: peripheral nerves [6], [7],”
BackgroundWells, E. D., Shehata, A. W., Dawson, M. R. W., Carey, J. P., & Hebert, J. S. (2022) · Sensors · doi:10.3390/s22103892 · Cited by 24
“They have also demonstrated functional performance increases in longitudinal studies [29–32] and stability in sensory percepts over long periods [21].”
BackgroundIenca, M., Valle, G. D., & Raspopović, S. (2025) · The Lancet Digital Health · doi:10.1016/s2589-7500(24)00222-x · Cited by 34
Nataraj, R., Audu, M. L., & Triolo, R. J. (2017) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2017.01.023 · Cited by 26
Regenerative Peripheral Nerve Interfaces for Advanced Control of Upper Extremity Prosthetic Devices
Kumar, N. G., Kung, T. A., & Cederna, P. S. (2021) · Hand Clinics · doi:10.1016/j.hcl.2021.04.005 · Cited by 26
Multi-modal prosthetic fingertip sensor with proximity, contact, and force localization capabilities
Segil, J. L., Patel, R., Klingner, J., Weir, R. F. F., & Correll, N. (2019) · Advances in Mechanical Engineering · doi:10.1177/1687814019844643 · Cited by 26
“10,11 During grasping tasks, the digits conform around objects using a variety of grasp postures.”
BackgroundMethodNeural engineering: the process, applications, and its role in the future of medicine
Ereifej, E. S., Shell, C. E., Schofield, J. S., Charkhkar, H., Cuberovic, I., Dorval, A. D., Graczyk, E. L., Kozai, T. D. Y., Otto, K. J., Tyler, D. J., et al. (2019) · Journal of Neural Engineering · doi:10.1088/1741-2552/ab4869 · Cited by 30
Self-citationGreenwald, E., Masters, M. R., & Thakor, N. V. (2016) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-016-1452-6 · Cited by 24
Recording nerve signals in canine sciatic nerves with a flexible penetrating microelectrode array
Byun, D., Cho, S.-J., Lee, B. H., Min, J., Lee, J.-H., & Kim, S. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa7493 · Cited by 23
“These non-penetrating electrodes have been reported to provide effective electrical stimulation with an adequate selectivity in chronic trials [25], but accompany mechanically-induced nerve damage due to tight cuff wrapping around the nerve, low signal-to-noise ratio (SNR), and low selectivity in recordings [19, 26].”
BackgroundAssessing the quality of supplementary sensory feedback using the crossmodal congruency task
Blustein, D. H., Wilson, A. W., & Sensinger, J. W. (2018) · Scientific Reports · doi:10.1038/s41598-018-24560-3 · Cited by 32
Sensing and Control for Prosthetic Hands in Clinical and Research Applications
Osborn, L. E., Iskarous, M. M., & Thakor, N. V. (2019) · Elsevier eBooks · doi:10.1016/b978-0-12-814659-0.00022-9 · Cited by 34
“Through tactile perceptions, prosthesis users with implanted stimulating electrodes in their median and ulnar nerves have been shown to differentiate between object stiffness [14], perform fine motor movements such as pulling a stem off a cherry [78], and improve performance in functional tasks [79] and activities of…”
BackgroundChronic nerve health following implantation of femoral nerve cuff electrodes
Freeberg, M. J., Pinault, G. C. J., Tyler, D. J., Triolo, R. J., & Ansari, R. (2020) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-020-00720-3 · Cited by 23
“While penetrating electrodes offer the potential of higher selectivity [27–29], non-penetrating nerve cuff electrodes (NCEs) on nerves in the upper [2, 4, 5, 30] and lower [3, 26, 31–35] extremities have been operational and stable for more than 11 years postimplantation [31] in terms of stimulation threshold and”
BackgroundSelf-citationDweiri, Y. M., Stone, M. A., Tyler, D. J., McCallum, G. A., & Durand, D. M. (2016) · Journal of Visualized Experiments · doi:10.3791/54388 · Cited by 23
Self-citationLoutit, A. J., & Potas, J. R. (2020) · Frontiers in Neuroscience · doi:10.3389/fnins.2020.00156 · Cited by 24
The Neural Representation of Force across Grasp Types in Motor Cortex of Humans with Tetraplegia
Rastogi, A., Willett, F. R., Abreu, J., Crowder, D. C., Murphy, B. A., Memberg, W. D., Vargas-Irwin, C. E., Miller, J. P., Sweet, J. A., Walter, B. L., et al. (2021) · eNeuro · doi:10.1523/eneuro.0231-20.2020 · Cited by 23
“Specifically, since force-related representation relies heavily on proprioceptive and tactile feedback (Tan et al., 2014; Tabot et al., 2015; Schiefer et al., 2018), whose neural pathways are altered during tetraplegia (Solstrand Dahlberg et al., 2018), the current study may have yielded weaker…”
BackgroundAckerley, R., Wasling, H. B., Ortiz-Catalan, M., Brånemark, R., & Wessberg, J. (2018) · Journal of Neurophysiology · doi:10.1152/jn.00909.2017 · Cited by 26
“Other studies in amputees implanted with electrodes between 1 week and 3 years have shown similar feasibility for producing sensations through stimulation, although these studies have focused on more specialized activities, such as motor-functional (Horch et al. 2011; Raspopovic et al. 2014; Tan et al. 2014; Davis et…”
BackgroundInfluentialThe Interaction Between Feedback Type and Learning in Routine Grasping With Myoelectric Prostheses
Wilke, M. A., Hartmann, C., Schimpf, F., Farina, D., & Dosen, S. S. (2019) · IEEE Transactions on Haptics · doi:10.1109/toh.2019.2961652 · Cited by 28
“For example, the studies using different non-invasive and invasive methods to transmit force feedback have reported no difference in performance [8], [16], [23], [27], clear improvement in an abstract electrotactile tracking task [28], or in functional and real-life tasks [11], [29]–[31], or improvement in only some s…”
BackgroundOsseointegrated Metallic Implants for Finger Amputees: A Review of the Literature
Bregoli, C., Biffi, C. A., Morellato, K., Gruppioni, E., Primavera, M., Rampoldi, M., Lando, M., Adani, R., & Tuissi, A. (2022) · Orthopaedic Surgery · doi:10.1111/os.13296 · Cited by 21
Bergmeister, K. D., Hader, M., Lewis, S., Russold, M.-F., Schiestl, M., Manzano-Szalai, K., Roche, A. D., Salminger, S., Dietl, H., & Aszmann, O. C. (2015) · Plastic & Reconstructive Surgery · doi:10.1097/prs.0000000000001926 · Cited by 23
D’Alonzo, M., Engels, L. F., Controzzi, M., & Cipriani, C. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa81e2 · Cited by 24
Giuliano Vanghetti and the innovation of “cineplastic operations”
Tropea, P., Mazzoni, A., Micera, S., & Corbo, M. (2017) · Neurology · doi:10.1212/wnl.0000000000004488 · Cited by 23
“The latest generation of upper limb neuroprostheses makes use of both the remaining motor and sensory functionality of the stump.(24,25) Moreover, recent studies show that it is possible to extend Vanghetti’s principle in patients with tetraplegia, using the residual functionality of the motor cortex to control the re…”
BackgroundIsaković, M. S., Malesevic, J. M., Keller, T., Kostic, M. D., & Strbac, M. D. (2019) · Applied Bionics and Biomechanics · doi:10.1155/2019/9298758 · Cited by 29
Nizamis, K., Stienen, A. H. A., Kamper, D. G., Keller, T., Plettenburg, D. H., Rouse, E. J., Farina, D., Koopman, H. F. J. M., & Sartori, M. (2019) · IEEE Transactions on Medical Robotics and Bionics · doi:10.1109/tmrb.2019.2912453 · Cited by 24
“in terms of coordination and dexterity [121]–[123].”
ThermalGrasp: Enabling Thermal Feedback even while Grasping and Walking
Mazursky, A., Brooks, J., Desta, B., & Lopes, P. (2024) · IEEE Conference on Virtual Reality and 3D User Interfaces (VR) · doi:10.1109/vr58804.2024.00056 · Cited by 31
“Both metrics were evaluated using standardized study designs from psychophysics literature [75], [76].”
Wearable non-invasive neuroprosthesis for targeted sensory restoration in neuropathy
Gozzi, N., Chee, L., Odermatt, I., Kikkert, S., Preatoni, G., Valle, G., Pfender, N., Beuschlein, F., Wenderoth, N., Zipser, C. M., et al. (2024) · Nature Communications · doi:10.1038/s41467-024-55152-7 · Cited by 29
“Near-naturaltactilesensationhasbeenachieved only through invasive neural technologies both at the peripheral level 22,25,66 and at the central level 67 – 69 .”
A Robust Encoding Scheme for Delivering Artificial Sensory Information via Direct Brain Stimulation
Bjånes, D. A., & Moritz, C. T. (2019) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2019.2936739 · Cited by 21
“These open-loop controllers, however, rely on slow visual feedback pathways and may realize greater improvements by incorporating closed-loop sensory feedback [37], [38].”
BackgroundAmoruso, E., Dowdall, L., Kollamkulam, M., Ukaegbu, O. C., Kieliba, P., Ng, T., Dempsey‐Jones, H., Clode, D., & Makin, T. R. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac47d9 · Cited by 27
Wang, Y., Fang, P., Tang, X., Jiang, N., Tian, L., Li, X., Zheng, Y., Huang, J., Samuel, O. W., Wang, H., et al. (2022) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2022.3155756 · Cited by 23
Koppaka, S., Hess‐Dunning, A., & Tyler, D. J. (2022) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2022.111058 · Cited by 23
Self-citationEmerging Medical Technologies and Their Use in Bionic Repair and Human Augmentation
Manero, A. C., Rivera, V., Fu, Q., Schwartzman, J. D., Gibbs, H. P., Shah, N., Gandhi, D. A., White, E., Crawford, K. E., & Coathup, M. J. (2024) · Bioengineering · doi:10.3390/bioengineering11070695 · Cited by 28
Christie, B. P., Charkhkar, H., Shell, C. E., Marasco, P. D., Tyler, D. J., & Triolo, R. J. (2019) · Scientific Reports · doi:10.1038/s41598-019-47867-1 · Cited by 22
Self-citationSensitivity to temporal parameters of intraneural tactile sensory feedback
Valle, G., Strauss, I., D’Anna, E., Granata, G., Di Iorio, R., Stieglitz, T., Rossini, P. M., Raspopovic, S., Petrini, F. M., & Micera, S. (2020) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-020-00737-8 · Cited by 20
Intensity Modulation: A Novel Approach to Percept Control in Spinal Cord Stimulation
Tan, D. W., Tyler, D. J., Sweet, J. A., & Miller, J. P. (2015) · Neuromodulation Technology at the Neural Interface · doi:10.1111/ner.12358 · Cited by 24
“In previous studies using IMS (8), we were able to identify a small window for IMS (e.”
BackgroundMethodInfluentialSelf-citationTessellation of artificial touch via microstimulation of human somatosensory cortexPreprint
Greenspon, C. M., Shelchkova, N. D., Valle, G., Hobbs, T. G., Berger-Wolf, E. I., Hutchison, B. C., Dogruoz, E., Verbaarschot, C., Callier, T., Sobinov, A. R., et al. (2023) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2023.06.23.545425 · Cited by 24
“Indeed, electrical activation of tactile nerve fibers evokes sensations experienced at a specific location on the skin, known as the projected field (PF) (Ochoa and Torebjörk, 1983; Ortiz-Catalan et al., 2020b; Overstreet et al., 2019; Petrini et al., 2019; Tan et al., 2014; Wendelken et al., 2017).”
Gherardini, M., Masiero, F., Ianniciello, V., & Cipriani, C. (2023) · Current Opinion in Biomedical Engineering · doi:10.1016/j.cobme.2023.100460 · Cited by 24
Importance of Breast Sensation After Mastectomy: Evidence from Three Sources
Lindau, S. T., Pinkerton, E. A., Abramsohn, E., Fuller, C. M., Grubb, D., Mendoza, T., & Siston, A. K. (2023) · Women s Health Reports · doi:10.1089/whr.2023.0106 · Cited by 21
“32,33 The Bionic Breast Project has been developed with extra-mural research support, including funding from the National Cancer Institute.”
MethodAsymmetric Sensory-Motor Regeneration of Transected Peripheral Nerves Using Molecular Guidance Cues
Anand, S., Desai, V. H., Alsmadi, N. Z., Kanneganti, A., Nguyen, D. H., Tran, M., Patil, L. S., Vasudevan, S., Xu, C., Hong, Y., et al. (2017) · Scientific Reports · doi:10.1038/s41598-017-14331-x · Cited by 20
Charkhkar, H., Christie, B. P., Pinault, G. J., Tyler, D. J., & Triolo, R. J. (2019) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108414 · Cited by 19
Self-citationSelectivity of afferent microstimulation at the DRG using epineural and penetrating electrode arrays
Nanivadekar, A. C., Ayers, C. A., Gaunt, R. A., Weber, D. J., & Fisher, L. E. (2019) · Journal of Neural Engineering · doi:10.1088/1741-2552/ab4a24 · Cited by 21
The Short-Term Repeatability of Subdermal Electrical Stimulation for Sensory Feedback
Dong, J. C., Kamavuako, E. N., Dosen, S. S., Jensen, W., & Geng, B. (2020) · IEEE Access · doi:10.1109/access.2020.2984534 · Cited by 20
“Thesemethods have led to promising results [9], [29], [30]; however, a surgical procedure is required which may not be accepted by all amputees.”
BackgroundQ-PINE: A quick to implant peripheral intraneural electrode
Strauss, I., Niederhoffer, T., Giannotti, A., Panarese, A. M., Bernini, F., Gabisonia, K., Ottaviani, M. M., Petrini, F. M., Recchia, F. A., Raspopović, S., et al. (2020) · Journal of Neural Engineering · doi:10.1088/1741-2552/abc52a · Cited by 19
A neurocognitive pathway for engineering artificial touch
Nisky, I., & Makin, T. R. (2024) · Science Advances · doi:10.1126/sciadv.adq6290 · Cited by 25
The Agonist-antagonist Myoneural Interface
Herr, H., & Carty, M. J. (2021) · Techniques in Orthopaedics · doi:10.1097/bto.0000000000000552 · Cited by 22
Decoding Vagus-Nerve Activity with Carbon Nanotube Sensors in Freely Moving Rodents
Marmerstein, J. T., McCallum, G. A., & Durand, D. M. (2022) · Biosensors · doi:10.3390/bios12020114 · Cited by 18
“Extraneural cuff electrodes have proven to be very effective peripheral nerve interfaces, allowing for selective stimulation [16] and some selectivity in recording [17–20].”
BackgroundArtificial referred sensation in upper and lower limb prosthesis users: a systematic review
Gonzalez, M. A., Bismuth, A., Lee, C., Chestek, C. A., & Gates, D. H. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac8c38 · Cited by 27
“Additionally, no human trials were cited in determining electrode limits, and all papers reached pre-set parameter limits or subjective pain/discomfort limits before reaching their charge limitations [31, 44, 61].”
InfluentialSchmitt, M. S., Wright, J. D., Triolo, R. J., Charkhkar, H., & Graczyk, E. L. (2023) · Frontiers in Human Neuroscience · doi:10.3389/fnhum.2022.1074033 · Cited by 21
“Peripheral nerve stimulation to restore sensation has been tested in both upper and lower limb amputees (Clark et al., 2014; Ortiz-Catalan et al., 2014; Raspopovic et al., 2014; Tan et al., 2014; Charkhkar et al., 2018; Petrini et al., 2019), and it has been shown to produce sensation perceived as originating from the…”
BackgroundThe mechanisms of electrical neuromodulation
Balbinot, G. C., Milosevic, M., Morshead, C. M., Iwasa, S. N., Zariffa, J., Milosevic, L., Valiante, T. A., Hoffer, J. A., & Popović, M. R. (2024) · The Journal of Physiology · doi:10.1113/jp286205 · Cited by 32
Development of a Prototype Over-Actuated Biomimetic Prosthetic Hand
Williams, M. R., & Walter, W. (2015) · PLoS ONE · doi:10.1371/journal.pone.0118817 · Cited by 21
Micera, S. (2016) · IEEE Pulse · doi:10.1109/mpul.2016.2539760 · Cited by 18
A third arm — Design of a bypass prosthesis enabling incorporation
Wilson, A. W., Blustein, D. H., & Sensinger, J. W. (2017) · International Conference on Rehabilitation Robotics · doi:10.1109/icorr.2017.8009441 · Cited by 19
Kim, H., Dingle, A. M., Ness, J. P., Baek, D.-H., Bong, J., Lee, I. K., Shulzhenko, N. O., Zeng, W., Israel, J. S., Pisaniello, J. A., et al. (2020) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2020.108602 · Cited by 19
Nguyen, A. T., Xu, J., Luu, D. K., Overstreet, C. K., Cheng, J., Keefer, E. W., & Yang, Z. (2021) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2021.3057041 · Cited by 20
Cracchiolo, M., Panarese, A., Valle, G. D., Strauss, I., Granata, G., Iorio, R. D., Stieglitz, T., Rossini, P. M., Mazzoni, A., & Micera, S. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/abef3a · Cited by 17
“Previous works has also shown that different encoding strategies could be used to successfully restore natural-like sensory feedback [5, 7, 9, 11, 15] and improve hand prosthesis movement control [11, 12].”
Encoding of limb state by single neurons in the cuneate nucleus of awake monkeys
Versteeg, C., Rosenow, J. M., Bensmaïa, S. J., & Miller, L. E. (2021) · Journal of Neurophysiology · doi:10.1152/jn.00568.2020 · Cited by 19
Park, H., Oh, K., Prilutsky, B. I., & DeWeerth, S. P. (2016) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2016.7833733 · Cited by 20
“The electrical stimulation of sensory nerves with the appropriate stimulation parameters and movement phases can evoke a compound action potential of sensory nerve fibers and partially restore the missing sensory feedback [5].”
Inter-Limb Transfer of Grasp Force Perception With Closed-Loop Hand Prosthesis
Fu, Q., Shao, F., & Santello, M. (2019) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2019.2911893 · Cited by 20
“Examples of such feedback methods are peripheral nerve stimulation [6], [7] and targeted reinnervation [8].”
BackgroundShon, A., Brakel, K., Hook, M. A., & Park, H. (2021) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2021.3072894 · Cited by 19
“parameters used in prior studies of animals and humans [22], [44].”
MethodA machine learning framework to optimize optic nerve electrical stimulation for vision restoration
Romeni, S., Zoccolan, D., & Micera, S. (2021) · Patterns · doi:10.1016/j.patter.2021.100286 · Cited by 17
Wearable High Voltage Compliant Current Stimulator for Restoring Sensory Feedback
Collu, R., Paolini, R., Bilotta, M., Demofonti, A., Cordella, F., Zollo, L., & Barbaro, M. (2023) · Micromachines · doi:10.3390/mi14040782 · Cited by 19
A Decade of Haptic Feedback for Upper Limb Prostheses
Dey, A., Basumatary, H., & Hazarika, S. M. (2023) · IEEE Transactions on Medical Robotics and Bionics · doi:10.1109/tmrb.2023.3310045 · Cited by 21
“However, invasive procedures are prone to site rejection, paresthesia and interference with myoelectric prostheses because of their direct routing with the nervous system [18], [96].”
BackgroundSensory restoration for improved motor control of prostheses
Fisher, L. E., Gaunt, R. A., & Huang, H. (2023) · Current Opinion in Biomedical Engineering · doi:10.1016/j.cobme.2023.100498 · Cited by 22
Guo, Y., Sun, L., Zhong, W., Zhang, N., Zhao, Z., & Tian, W. (2024) · PubMed Central · doi:10.4103/1673-5374.380909 · Cited by 35
Zhang, L., Zhao, Y., Shen, C., Lei, L., Dong, J., Zou, D., Zou, J., & Wang, M. (2018) · Frontiers in Physiology · doi:10.3389/fphys.2018.01396 · Cited by 21
“Touch perception in the finger is an indispensable part of fine motor control, which helps contribute to the sense of self, to effective communication, and a clear perception of the world (Tan et al., 2014).”
BackgroundQuick, K. M., Weiss, J. M., Clemente, F., Gaunt, R. A., & Collinger, J. L. (2020) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc44109.2020.9175926 · Cited by 21
“Exploiting this approach, researchers restored the ability to modulate the grip force [8] and improved motor coordination in functional grasping tasks [9].”
BackgroundXu, H., Chai, G., Zhang, N., & Gu, G. (2022) · Soft Science · doi:10.20517/ss.2022.17 · Cited by 19
Muller, L., Rolston, J. D., Fox, N. P., Knowlton, R. C., Rao, V. R., & Chang, E. F. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa9bf9 · Cited by 16
“Future work may include metrics of intensity or perceptual aspects associated with stimulation patterns or modulated stimulation profiles [62, 63].”
BackgroundPayne, S. C., Burns, O., Stebbing, M. J., Thomas, R., de Silva, A., Sedó, A., Weissenborn, F., Hyakumura, T., Huynh, M., May, C. N., et al. (2018) · Bioelectronics in Medicine · doi:10.2217/bem-2018-0011 · Cited by 18
Norrlid, J., Enander, J. M. D., Mogensen, H., & Jörntell, H. (2021) · Frontiers in Cellular Neuroscience · doi:10.3389/fncel.2021.677568 · Cited by 21
“Accordingly, in humans, electrical nerve stimulation with a much lower resolution than in the present set of experiments are known to generate sensory impressions that are in part perceived as unnatural but also to generate diversified and meaningful tactile percepts (Tan et al., 2014; Oddo et al., 2016).”
BackgroundWorkshops of the eighth international brain–computer interface meeting: BCIs: the next frontier
Huggins, J. E., Krusienski, D. J., Vansteensel, M. J., Valeriani, D., Thelen, A., Stavisky, S. D., Norton, J. J. S., Nijholt, A., Müller-Putz, G., Kosmyna, N., et al. (2022) · Brain-Computer Interfaces · doi:10.1080/2326263x.2021.2009654 · Cited by 21
Bidirectional bionic limbs: a perspective bridging technology and physiology
Pasluosta, C. F., Kiele, P., Čvančara, P., Micera, S., Aszmann, O. C., & Stieglitz, T. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac4bff · Cited by 19
“Reprinted by permission from Springer Nature Customer Service Centre GmbH: Springer Nature, Current Surgery Reports [23] Control Strategies: A Clinical Perspective, Roche A D, Rehbaum H, Farina D and Aszmann O C, (c) 2014.”
BackgroundMethodRecent Advances in Implantable Neural Interfaces for Multimodal Electrical Neuromodulation
Wang, L., Liu, S., Zhao, W., Li, J., Zeng, H., Kang, S., Sheng, X., Wang, L., Fan, Y., & Yin, L. (2024) · Advanced Healthcare Materials · doi:10.1002/adhm.202303316 · Cited by 16
Sabetian, P., Popović, M. R., & Yoo, P. B. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa6407 · Cited by 18
Desai, V. H., Spearman, B. S., Shafor, C. S., Natt, S., Teem, B., Graham, J. B., Atkinson, E. W., Wachs, R. A., Nunamaker, E. A., Otto, K. J., et al. (2017) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2017.8008326 · Cited by 17
“Recent advances have shown that electrical stimulation of axons with peripheral-nerve interfaces can successfully provide natural multi-perception sensory feedback and alleviate phantomlimb pain in upper-limb amputees [1-2].”
BackgroundRekant, J., Fisher, L. E., Boninger, M. L., Gaunt, R. A., & Collinger, J. L. (2022) · Assistive Technology · doi:10.1080/10400435.2021.2020935 · Cited by 18
Restoration of Hand Function in Stroke and Spinal Cord Injury
Kamper, D. G. (2016) · Springer eBooks · doi:10.1007/978-3-319-28603-7_15 · Cited by 19
Anikeeva, P., & Koppes, R. A. (2015) · Science · doi:10.1126/science.aad0910 · Cited by 16
Yu, X. P., Su, J., Guo, J. Y., Zhang, X., Li, R. H., Chai, X., Chen, Y., Zhang, D. G., Wang, J., Sui, X., et al. (2019) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108450 · Cited by 17
Sensory information feedback for neural prostheses
Weber, D. J., Hao, M., Urbin, M. A., Schoenewald, C., & Lan, N. (2019) · Elsevier eBooks · doi:10.1016/b978-0-12-816034-3.00021-3 · Cited by 17
Tutorial: a guide to techniques for analysing recordings from the peripheral nervous system
Koh, R. G. L., Zariffa, J., Jabban, L., Yen, S., Donaldson, N., & Metcalfe, B. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac7d74 · Cited by 19
“Good examples of this include PNIs that are used for either controlling upper-limb prosthesis or for proving sensory feedback from a prosthesis—such experimental interfaces have been implanted in humans for several weeks or months [6, 7, 114, 115].”
Regenerative peripheral nerve interfaces (RPNIs): current status and future direction
Burke, K. L., Kung, T. A., Hooper, R. C., Kemp, S. W. P., & Cederna, P. S. (2022) · Plastic and Aesthetic Research · doi:10.20517/2347-9264.2022.26 · Cited by 15
Nishi, Y., Ikuno, K., Minamikawa, Y., Igawa, Y., Osumi, M., & Morioka, S. (2022) · Frontiers in Human Neuroscience · doi:10.3389/fnhum.2022.937319 · Cited by 16
Gonzalez, M. A., Nwokeabia, C., Vaskov, A. K., Vu, P. P., Lu, C. W., Patil, P. G., Cederna, P. S., Chestek, C. A., & Gates, D. H. (2023) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2023.3345164 · Cited by 17
“Using peripheral nerve interfaces, researchers have elicited sensations that are referred to the phantom hand and consistent in location and quality over several months [8], [9], [10].”
BackgroundClinical Applications and Future Translation of Somatosensory Neuroprostheses
Graczyk, E. L., Hutchison, B. C., Valle, G. D., Bjånes, D. A., Gates, D. H., Raspopović, S., & Gaunt, R. A. (2024) · Journal of Neuroscience · doi:10.1523/jneurosci.1237-24.2024 · Cited by 20
Petrini, F. M., Mazzoni, A., Rigosa, J., Giambattistelli, F., Granata, G., Barra, B., Pampaloni, A., Guglielmelli, E., Zollo, L., Capogrosso, M., et al. (2019) · BioMedical Engineering OnLine · doi:10.1186/s12938-019-0659-9 · Cited by 15
Edhi, M. M., Heijmans, L., Vanent, K. N., Bloye, K., Baanante, A., Jeong, K., Leung, J., Zhu, C., Esteller, R., & Saab, C. Y. (2020) · Scientific Reports · doi:10.1038/s41598-020-77212-w · Cited by 16
Vėbraitė, I., & Hanein, Y. (2021) · Frontiers in Medical Technology · doi:10.3389/fmedt.2021.675744 · Cited by 23
Cole, E. R., Eggers, T. E., Weiss, D. A., Connolly, M. J., Gombolay, M., Laxpati, N. G., & Gross, R. E. (2024) · Journal of Neural Engineering · doi:10.1088/1741-2552/ad5407 · Cited by 12
“…of neuronal population responses across a micro-electrode array during pulsatile vs. sinusoidal stimulation and examine whether the biomimetic utility of sinusoidal stimulation can be replicated with electrical stimulation by smoothly adjusting the amplitude of an electrical pulse train [56].”
Neuromorphic Artificial Sense of Touch: Bridging Robotics and Neuroscience
Rongala, U. B., Mazzoni, A., Camboni, D., Carrozza, M. C., & Oddo, C. M. (2017) · Springer proceedings in advanced robotics · doi:10.1007/978-3-319-60916-4_35 · Cited by 19
“This suggests that our neuromorphic spike trains might contain enough information to convey an adequate sensory feedback to upper limb neuroprostheses [32, 37] and to be useful in robotic applications.”
BackgroundNeuromorphic vision and tactile fusion for upper limb prosthesis control
Hays, M. A., Osborn, L. E., Ghosh, R., Iskarous, M. M., Hunt, C. L., & Thakor, N. V. (2019) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2019.8716890 · Cited by 15
“efforts have also led to restoring sensations of touch [9], [10], texture [11], and even pain [12] in amputees.”
Azbell, J., Park, J. K., Chang, S., ENGELEN, M. P. K. J., & Park, H. (2019) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2019.8717004 · Cited by 21
A Study on Biocompatible Polymer-Based Packaging of Neural Interface for Chronic Implantation
Park, H., Choi, W., Oh, S.-H., Kim, Y.-J., Seok, S., & Kim, J. (2022) · Micromachines · doi:10.3390/mi13040516 · Cited by 15
Creation of a biological sensorimotor interface for bionic reconstruction
Festin, C., Ortmayr, J., Maierhofer, U., Tereshenko, V., Blumer, R., Schmoll, M., Rojas, G. C., Luft, M., Laengle, G., Farina, D., et al. (2024) · Nature Communications · doi:10.1038/s41467-024-49580-8 · Cited by 18
Delhaye, B. P., Schluter, E. W., & Bensmaïa, S. J. (2016) · IEEE Transactions on Haptics · doi:10.1109/toh.2016.2573298 · Cited by 17
“[13], [14], [15], [16], [17].”
Earley, E. J., Johnson, R. E., Sensinger, J. W., & Hargrove, L. J. (2021) · Scientific Reports · doi:10.1038/s41598-021-84795-5 · Cited by 13
Sensorimotor integration within the primary motor cortex by selective nerve fascicle stimulation
Ranieri, F., Pellegrino, G., Ciancio, A. L., Musumeci, G., Noce, E., Insola, A., Balzani, L. A. D., Lazzaro, V. D., & Pino, G. D. (2022) · The Journal of Physiology · doi:10.1113/jp282259 · Cited by 19
“Stimulation parameters were chosen to inject a total charge below the safety limit for electrode and nerve damage, based on available data with similar electrodes (Dhillon et al. 2004; Dhillon, 2005; Rossini et al. 2010; Ortiz-Catalan et al. 2014; Tan et al. 2014).”
Eickenscheidt, M., Herrmann, T., Weisshap, M., Mittnacht, A., Rudmann, L., Zeck, G. M., & Stieglitz, T. (2022) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2022.114090 · Cited by 16
Peripheral neurostimulation for encoding artificial somatosensations
Valle, G. D. (2022) · European Journal of Neuroscience · doi:10.1111/ejn.15822 · Cited by 18
“to deliver electrical stimuli to the peripheral nerves that were traditionally proportional to the readouts of artificial sensors in the hand prosthesis (D’Anna et al., 2019; Raspopovic et al., 2014; Tan et al., 2014; Valle, Petrini, et al., 2018).”
BackgroundMethodInfluentialTriboelectric neurostimulator for physiological modulation of leg muscle
Kang, M., Shin, H., Cho, Y., Park, J., Nagwade, P., & Lee, S. (2022) · Nano Energy · doi:10.1016/j.nanoen.2022.107861 · Cited by 15
Alteration in Cortical Activity and Perceived Sensation Following Modulated TENS
Jadidi, A. F., Jensen, W., Zarei, A. A., & Lontis, R. (2023) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2023.3236038 · Cited by 17
“which can replicate a more natural sensation [27], [47], [48].”
Background
Metadata
- Authors
- Tan, D. W., Schiefer, M. A., Keith, M. W., Anderson, J. R., Tyler, J., & Tyler, D. J.
- Affiliations
- Louis Stokes Cleveland VA Medical Center (US)
- MetroHealth Medical Center (US)
- Corresponding author
- Tan, Daniel W.
- Schiefer, Matthew A.
- Tyler, Dustin J.
- Venue
- Science Translational Medicine
- Published
- 2014-10-08
- Volume
- 6
- Issue
- 257
- Pages
- 257ra138-257ra138
- ISSN
- 1946-6234 (ISSN-L)
- 1946-6242
- PMC
- PMC5517305
- OpenAlex
- W2157094999
- Semantic Scholar
- 308d067b912e90f2bd3958d5248e2d960435ea5c
- Cited by
- 751
- References
- 29
- Citations by year
- 2026: 24 · 2025: 41 · 2024: 57 · 2023: 59 · 2022: 81 · 2021: 75 · 2020: 74 · 2019: 87 · 2018: 82 · 2017: 64 · 2016: 64 · 2015: 37 · 2014: 5
- FWCI
- 25.89
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Science Translational Medicine (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- U.S. Department of Veterans Affairs (I01 RX001334, IK1 RX000724)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Cognitive Neuroscience › Tactile and Sensory Interactions
- Health Sciences › Medicine › Anesthesiology and Pain Medicine › Pain Management and Treatment
- Keywords
- peripheral nerve interfaces, electrical stimulation, tactile perception, upper limb amputation, prosthetic hand, sensory feedback, phantom hand, long-term stability, paresthesia, grasp control, fine motor control, sensory recovery
- MeSH terms
- Amputees, Artificial Limbs, Electric Stimulation, Electrodes, Implanted, Hand, Humans, Male, Middle Aged, Motor Skills, Paresthesia, Peripheral Nerves, Phantom Limb, Pressure, Sensation, Touch Perception, Feedback, Sensory, Translational Research, Biomedical
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-07, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-07 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.