Validation of Laboratory-Developed Molecular Assays for Infectious Diseases
Open accessThe DOI matches its Crossref record; 48 of the 48 references it cites were found in a registry or an index.
Abstract
SUMMARY: Molecular technology has changed the way that clinical laboratories diagnose and manage many infectious diseases. Excellent sensitivity, specificity, and speed have made molecular assays an attractive alternative to culture or enzyme immunoassay methods. Many molecular assays are commercially available and FDA approved. Others, especially those that test for less common analytes, are often laboratory developed. Laboratories also often modify FDA-approved assays to include different extraction systems or additional specimen types. The Clinical Laboratory Improvement Amendments (CLIA) federal regulatory standards require clinical laboratories to establish and document their own performance specifications for laboratory-developed tests to ensure accurate and precise results prior to implementation of the test. The performance characteristics that must be established include accuracy, precision, reportable range, reference interval, analytical sensitivity, and analytical specificity. Clinical laboratories are challenged to understand the requirements and determine the types of experiments and analyses necessary to meet the requirements. A variety of protocols and guidelines are available in various texts and documents. Many of the guidelines are general and more appropriate for assays in chemistry sections of the laboratory but are applied in principle to molecular assays. This review presents information that laboratories may consider in their efforts to meet regulatory requirements.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Implementation of mNGS in the clinical laboratory is a complex endeavour that requires customization of research protocols using a quality management approach consistent with regulatory standards90.”
From Clinical metagenomics (2019) · cites this paper as [90]
In the citing paper
“Clinical laboratories are highly regulated, and general laboratory and testing requirements apply to all molecular diagnostic assays reported for patient care90.”
From Clinical metagenomics (2019) · cites this paper as [90]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
International Organization for Standardization
(2015) · Encyclopedia of Astrobiology · doi:10.1007/978-3-662-44185-5_100574 · Cited by 6,716
Tietz Textbook of Clinical Chemistry
(1999) · Archives of Pathology and Laboratory Medicine · doi:10.5858/123.7.646b · Cited by 2,495
Thompson, M., Ellison, S. L. R., & Wood, R. (2002) · Pure and Applied Chemistry · doi:10.1351/pac200274050835 · Cited by 2,478
Real-Time PCR in Clinical Microbiology: Applications for Routine Laboratory Testing
Espy, M. J., Uhl, J. R., Sloan, L. M., Buckwalter, S. P., Jones, M. F., Vetter, E. A., Yao, J. D. C., Wengenack, N. L., Rosenblatt, J. E., Cockerill, F. R., et al. (2006) · Clinical Microbiology Reviews · doi:10.1128/cmr.19.1.165-256.2006 · Cited by 1,318
Highly sensitive real-time PCR for specific detection and quantification of Coxiella burnetii
Klee, S. R., Tyczka, J., Ellerbrok, H., Franz, T., Linke, S., Baljer, G., & Appel, B. (2006) · BMC Microbiology · doi:10.1186/1471-2180-6-2 · Cited by 323
Saah, A. J., & Hoover, D. R. (1997) · Annals of Internal Medicine · doi:10.7326/0003-4819-126-1-199701010-00026 · Cited by 269
Recommended Principles and Practices for Validating Clinical Molecular Pathology Tests
Jennings, L., Van Deerlin, V. M., & Gulley, M. L. (2009) · Archives of Pathology & Laboratory Medicine · doi:10.5858/133.5.743 · Cited by 250
Use of Bacteriophage MS2 as an Internal Control in Viral Reverse Transcription-PCR Assays
Dreier, J., Störmer, M., & Kleesiek, K. (2005) · Journal of Clinical Microbiology · doi:10.1128/jcm.43.9.4551-4557.2005 · Cited by 220
Research works printed but not linked: 16
- Code of Federal Regulations. 2010. Title 21. Food and drugs, vol. 8, chapter 1. Food and Drug Administration, Department of Health and Human Services, part 56. Institutional review boards. U.S. Government Printing Office, Washington, DC.
- Code of Federal Regulations. 2010. Title 21. Food and drugs, vol. 8, chapter 1. Food and Drug Administration, Department of Health and Human Services, part 864. Hematology and pathology devices, section 864.4020. Analyte specific reagents. U.S. Government Printing Office, Washington, DC.
- Code of Federal Regulations. 2010. Title 21. Food and drugs, vol. 8, chapter 1. Food and Drug Administration, Department of Health and Human Services, part 807. Establishment registration and device listing for manufacturers and initial importers of devices. U.S. Government Printing Office, Washington, DC.
- Code of Federal Regulations. 2010. Title 21. Food and drugs, vol. 8, chapter 1. Food and Drug Administration, Department of Health and Human Services, part 809. In vitro diagnostic products for human use, section 809.10. Labeling for in vitro diagnostic products. U.S. Government Printing Office, Washington, DC.
- Code of Federal Regulations. 2010. Title 21. Food and drugs, vol. 8, chapter 1. Food and Drug Administration, Department of Health and Human Services, part 809. In vitro diagnostic products for human use, section 809.20. General requirements for manufacturers and producers of in vitro diagnostic products. U.S. Government Printing Office, Washington, DC.
- Code of Federal Regulations. 2010. Title 21. Food and drugs, vol. 8, chapter 1. Food and Drug Administration, Department of Health and Human Services, part 809. In vitro diagnostic products for human use, section 809.30. Restrictions on the sale, distribution and use of analyte specific reagents. U.S. Government Printing Office, Washington, DC.
- Code of Federal Regulations. 2010. Title 21. Food and drugs, vol. 8, chapter 1. Food and Drug Administration, Department of Health and Human Services, part 812. Investigational device exemptions. U.S. Government Printing Office, Washington, DC.
- Code of Federal Regulations. 2010. Title 21. Food and drugs, vol. 8, chapter 1. Food and Drug Administration, Department of Health and Human Services, part 820. Quality system regulation. U.S. Government Printing Office, Washington, DC.
- Code of Federal Regulations. 2009. Title 42. Public health, vol. 4, chapter V. Health Care Financing Administration, Department of Health and Human Services, part 493. Laboratory requirements, section 493.1455. Standard: clinical consultant qualifications. U.S. Government Printing Office, Washington, DC.
- Code of Federal Regulations. 2009. Title 42. Public health, vol. 4, chapter V. Health Care Financing Administration, Department of Health and Human Services, part 493. Laboratory requirements, section 493.2. Definitions. U.S. Government Printing Office, Washington, DC.
- Code of Federal Regulations. 2009. Title 42. Public health, vol. 4, chapter V. Health Care Financing Administration, Department of Health and Human Services, part 493. Laboratory requirements, section 493.1253. Standard: establishment and verification of performance specifications. U.S. Government Printing Office, Washington, DC.
- Code of Federal Regulations. 2009. Title 42. Public health, vol. 4, chapter V. Health Care Financing Administration, Department of Health and Human Services, part 493. Laboratory requirements, section 493.1255. Standard: calibration and calibration verification procedures. U.S. Government Printing Office, Washington, DC.
- Code of Federal Regulations. 2009. Title 42. Public health, vol. 4, chapter V. Health Care Financing Administration, Department of Health and Human Services, part 493. Laboratory requirements, section 493.1256. Standard: control Procedures. U.S. Government Printing Office, Washington, DC.
- Code of Federal Regulations. 2009. Title 42. Public health, vol. 4, chapter V. Health Care Financing Administration, Department of Health and Human Services, part 493. Laboratory requirements, section 493.1282. Standard: corrective actions. U.S. Government Printing Office, Washington, DC.
- Social Security Act. Section 1862(a) (1), exclusions from coverage and Medicare as a second payer, now codified as Title 42, chapter 7, exclusions from coverage and Medicare as a second payer, U.S. Code. U.S. House of Representatives, Washington, DC.
- U.S. Department of Health and Human Services, Centers for Medicare and Medicaid Services. 1992. Clinical laboratory improvement amendments of 1988 (CLIA). Fed. Regist. 57:7002-7186. [PubMed] [Google Scholar]
Cited by
Items 101–200 of 550
Real-time RT-PCR for Mayaro virus detection in plasma and urine
Waggoner, J. J., Rojas, A., Mohamed-Hadley, A. A., Guillén, Y., & Pinsky, B. A. (2017) · Journal of Clinical Virology · doi:10.1016/j.jcv.2017.11.006 · Cited by 40
Latest Advances in Laboratory Detection of Mycoplasma genitalium
Waites, K. B., Crabb, D. M., Ratliff, A. E., Geisler, W. M., Atkinson, T. P., & Xiao, L. (2023) · Journal of Clinical Microbiology · doi:10.1128/jcm.00790-21 · Cited by 47
“documentation and maintaining copies of the data collected are CLIA requirements (127), and the results of this performance evaluation must be compared to those of another”
ResultPotula, R., & Tang, Y. (2016) · doi:10.1002/9781119021872.ch16 · Cited by 40
Lucero, R. H., Brusés, B. L., Cura, C. I., Formichelli, L. B., Juiz, N. A., Fernández, G. J., Bisio, M. M. C., Deluca, G., Besuschio, S. A., Hernández, D. O., et al. (2016) · Infection Genetics and Evolution · doi:10.1016/j.meegid.2016.03.028 · Cited by 38
“An anticipated linear reportable range (Burd, 2010) was obtained between 0.125 and 105 par.eq.”
BackgroundMethodInfluentialQvarnström, Y., Bishop, H. S., & da Silva, A. J. (2013) · Europe PMC (PubMed Central) · Cited by 38
Stranieri, I., Kanunfre, K. A., Rodrigues, J. C., Yamamoto, L., Nadaf, M. I. V., Palmeira, P., & Okay, T. S. (2018) · Revista do Instituto de Medicina Tropical de São Paulo · doi:10.1590/s1678-9946201860061 · Cited by 43
Zhou, G., & Liu, H. (2026) · Biosensors · doi:10.3390/bios16020112 · Cited by 51
“Senseonics got the CE mark in Europe for the Eversense system beforehand (under the pre-IVDR regime) [84], but getting to the U.S. market needs PMA-level evidence and life cycle controls, and FDA PMA records publicly show a regulated path that cannot be separated from clinical study expectations…”
Piralla, A., Girello, A., Premoli, M., & Baldanti, F. (2015) · Journal of Clinical Microbiology · doi:10.1128/jcm.03691-14 · Cited by 35
“The analytical evaluation of the EV-D68-specific real-time RT-PCR was performed according to previously published recommendations (8).”
MethodCura, C. I., Ramı́rez, J. C., Rodríguez, M. A., Lopez-Albízu, C., Irazú, L., Scollo, K., & Sosa‐Estáni, S. (2017) · Journal of Molecular Diagnostics · doi:10.1016/j.jmoldx.2017.05.010 · Cited by 35
Abràs, A., Ballart, C., Llovet, T., Roig, C., Gutiérrez, C., Tebar, S., Berenguer, P., Pinazo, M., Posada, E., Gascón, J., et al. (2018) · PLoS ONE · doi:10.1371/journal.pone.0195738 · Cited by 35
Carpenter, M. L., Tan, S. K., Watson, T., Bacher, R., Nagesh, V., Watts, A., Bentley, G., Weber, J. M., Huang, C., Sahoo, M. K., et al. (2019) · Journal of Clinical Microbiology · doi:10.1128/jcm.01113-19 · Cited by 51
“typical qPCR cutoffs (29) and previous mNGS approaches (28), may be overly conser-”
BackgroundLowe, C. F., Merrick, L., Harrigan, P. R., Mazzulli, T., Sherlock, C. H., & Ritchie, G. (2015) · Journal of Clinical Microbiology · doi:10.1128/jcm.02229-15 · Cited by 34
“Validation of this HBV NGS assay was based on recommendations for molecular testing in microbiology labs and NGS for genetics (12, 13, 16).”
MethodMentasti, M., Keše, D., Echahidi, F., Uldum, S. A., Afshar, B., David, S., Mrazek, J., Mendonça, R. D., Harrison, T. G., & Chalker, V. J. (2015) · European Journal of Clinical Microbiology & Infectious Diseases · doi:10.1007/s10096-015-2363-4 · Cited by 33
“in 95, 90 or 50 % of specimens tested) Bunder routine laboratory conditions in a defined type of specimens^ [17].”
BackgroundValidation of a laboratory-developed test of human sperm capacitation
Moody, M. A., Cardona, C., Simpson, A. J., Smith, T. T., Travis, A. J., & Ostermeier, G. C. (2017) · Molecular Reproduction and Development · doi:10.1002/mrd.22801 · Cited by 37
“This attribute is known as “precision,” which refers to how well a given measurement can be reproduced when applied repeatedly to multiple aliquots of a single homogeneous sample (Burd, 2010).”
BackgroundSensitivity enhancement of lateral flow assay by embedding cotton threads in paper
Zhang, S., Liu, L.-N., Tang, R., Liu, Z., He, X., Qu, Z., & Li, F. (2019) · Cellulose · doi:10.1007/s10570-019-02677-6 · Cited by 33
“To further validate the repeatability of the thread-embedded LFAs, we performed three tests for the high-concentration samples (50–2.5 nM) and eight tests for the lowconcentration samples (0.05–1 nM) using the prepared LFAs according to the literature (Burd 2010).”
MethodInfluentialEvaluation of a Highly Efficient DNA Extraction Method for Bacillus anthracis Endospores
Knüpfer, M., Braun, P. R., Baumann, K., Rehn, A., Antwerpen, M. H., Grass, G., & Wölfel, A. R. (2020) · Microorganisms · doi:10.3390/microorganisms8050763 · Cited by 37
“The limit of detection (LOD) at 0.95 probability for a positive response was calculated by probit analysis as 33 spores per mL, which corresponds to three copies per PCR-reaction [41].”
InfluentialTrends in Molecular Diagnostics and Genotyping Tools Applied for Emerging Sporothrix Species
de Carvalho, J. A., Monteiro, R. C., Hagen, F., Camargo, Z. P. D., & Rodrigues, A. M. (2022) · Journal of Fungi · doi:10.3390/jof8080809 · Cited by 33
Vandiver, A. R., Hoang, A. N., Herbst, A. J., Lee, C. C., Aiken, J. M., McKenzie, D. I., Teitell, M. A., Timp, W., & Wanagat, J. (2023) · Aging Cell · doi:10.1111/acel.13842 · Cited by 33
Multiplex nucleic acid amplification test for diagnosis of dengue fever, malaria, and leptospirosis.
Waggoner, J. J., Abeynayake, J., Balassiano, I., Lefterova, M., Sahoo, M. K., Liu, Y., Vital-Brazil, J. M., Gresh, L., Balmaseda, A., Harris, E., et al. (2014) · Journal of Clinical Microbiology · doi:10.1128/jcm.00341-14 · Cited by 25
“characterized according to published recommendations (33).”
Saunders, N., Zambon, M. C. F. A., Sharp, I., Siddiqui, R., Bermingham, A., Ellis, J., Vipond, B. B., Sails, A. D., Moran‐Gilad, J., Marsh, P., et al. (2012) · Journal of Clinical Virology · doi:10.1016/j.jcv.2012.11.013 · Cited by 32
Larsen, S. E., Berube, B. J., Pecor, T., Cross, E., Brown, B. P., Williams, B. D., Johnson, E., Qu, P., Carter, L. P., Wrenn, S. P., et al. (2021) · Journal of Immunological Methods · doi:10.1016/j.jim.2021.113160 · Cited by 33
Clinical Applications of Quantitative Real-Time PCR in Virology
Engstrom-Melnyk, J., Rodriguez, P. L., Peraud, O., & Hein, R. C. (2015) · Methods in microbiology · doi:10.1016/bs.mim.2015.04.005 · Cited by 38
“Laboratory developed tests can be grouped into three categories, FDA-cleared or approved test that have been modified, tests that are not subject to FDA clearance or approval, and tests for which no performance specifications have been provided by the manufacturer (e.g. analytespecific reagents or ASRs) (Burd, 2010; C…”
BackgroundInfluentialNovel technologies for the diagnosis of urinary tract infections
Bermudez, T. A., Schmitz, J. E., Boswell, M., & Humphries, R. M. (2025) · Journal of Clinical Microbiology · doi:10.1128/jcm.00306-24 · Cited by 51
Ramı́rez, J. C., Parrado, R., Sulleiro, E., de la Barra, A., Rodríguez, M. A., Villarroel, S., Irazú, L., Alonso‐Vega, C., Alves, F. P., Curto, M. Á., et al. (2017) · PLoS ONE · doi:10.1371/journal.pone.0188550 · Cited by 31
Variably improved microbial source tracking with digital droplet PCR
Nshimyimana, J. P., Cruz, M. C., Wuertz, S., & Thompson, J. R. (2019) · Water Research · doi:10.1016/j.watres.2019.04.056 · Cited by 37
KRAS mutations: Analytical considerations
Herreros‐Villanueva, M., Chen, C.-C., Yuan, S. F., Liu, T., & Er, T. (2014) · Clinica Chimica Acta · doi:10.1016/j.cca.2014.01.049 · Cited by 35
Khumalo, J., Nicol, M., Hardie, D., Muloiwa, R., Mteshana, P., & Bamford, C. (2017) · PLOS ONE · doi:10.1371/journal.pone.0173948 · Cited by 31
Development of a loop-mediated isothermal amplification assay for detection of Trichomonas vaginalis
Reyes, J. C. B., Solon, J. A., & Rivera, W. L. (2014) · Diagnostic Microbiology and Infectious Disease · doi:10.1016/j.diagmicrobio.2014.03.016 · Cited by 34
Colling, R. T., Wang, L. M., & Soilleux, E. J. (2017) · Journal of Clinical Pathology · doi:10.1136/jclinpath-2017-204356 · Cited by 30
Promises and pitfalls of Illumina sequencing for HIV resistance genotyping
Brumme, C. J., & Poon, A. F. Y. (2016) · Virus Research · doi:10.1016/j.virusres.2016.12.008 · Cited by 28
Pérez‐López, E., Rodríguez, D., Olivier, C., Luna‐Rodríguez, M., & Dumonceaux, T. (2017) · Scientific Reports · doi:10.1038/s41598-017-00895-1 · Cited by 31
Short-Term Stability of Pathogen-Specific Nucleic Acid Targets in Clinical Samples
Hasan, M. R., Tan, R., Al‐Rawahi, G. N., Thomas, É., & Tilley, P. A. G. (2012) · Journal of Clinical Microbiology · doi:10.1128/jcm.02659-12 · Cited by 33
“Although these assays provide high sensitivity, high specificity, and rapid turnaround time, special laboratory practice and procedures are required to avoid false-positive and false-negative results (1, 2).”
BackgroundZhang, C., Xiu, L., Xiao, Y., Xie, Z., Ren, L., & Peng, J. (2018) · Frontiers in Cellular and Infection Microbiology · doi:10.3389/fcimb.2018.00107 · Cited by 28
“However, some substances present in blood, body fluids, and sputum may affect polymerase activity and inhibit amplification (Burd, 2010).”
BackgroundKolia-Diafouka, P., Godreuil, S., BOURDIN, A. A., Carrère‐Kremer, S., Kremer, L., de Perre, P. V., & Tuaillon, É. (2018) · Frontiers in Microbiology · doi:10.3389/fmicb.2018.02224 · Cited by 35
Cury, A. P., de Almeida, J. N., Costa, S. F., Salomão, M. C., Boszczowski, Í., Duarte, A. J. S., & Rossi, F. (2020) · Journal of Microbiological Methods · doi:10.1016/j.mimet.2020.105884 · Cited by 28
Dupuis, M., Brunt, S. J., Appler, K. A., Davis, A., & Rudd, R. J. (2015) · Journal of Clinical Microbiology · doi:10.1128/jcm.01227-15 · Cited by 30
“Conventional nested PCR is time and labor-intensive and prone to cross contamination, and in the United States, it is no longer considered to have sufficient specificity for diagnostic use without sequence confirmation (14).”
BackgroundMethodHIV and Hepatitis C Virus Infection in the United States: Whom and How to Test
Panneer, N., Lontok, E., Branson, B. M., Teo, C. G., Dan, C., Parker, M. M., Stekler, J. D., DeMaria, A., & Miller, V. (2014) · Clinical Infectious Diseases · doi:10.1093/cid/ciu396 · Cited by 30
Sherma, N. D., Borges, C. R., Trenchevska, O., Jarvis, J. W., Rehder, D. S., Oran, P. E., Nelson, R. W., & Nedelkov, D. (2014) · Proteome Science · doi:10.1186/s12953-014-0052-3 · Cited by 29
“CLIA has put forth requirements for clinical labs [70], and as proteomics research has brought about novel techniques for quantitative protein analysis, interest has recently emerged in the research community with regard to harmonizing criteria for assay performance parameters.”
New Proteins for a New Perspective on Syphilis Diagnosis
Smith, B. C., Simpson, Y., Morshed, M. G., Cowen, L. L. E., Hof, R., Charmaine, W., & Cameron, C. E. (2012) · Journal of Clinical Microbiology · doi:10.1128/jcm.01390-12 · Cited by 27
“163 for Tp0326, Tp0453, and the Tp0453-Tp0326 chimera, respectively (24).”
Ambagala, A. P. N., Pahari, S., Fisher, M., Lee, P.-Y. A., Pasick, J. M. M., Ostlund, E. N., Johnson, D. J., & Lung, O. (2015) · Transboundary and Emerging Diseases · doi:10.1111/tbed.12388 · Cited by 27
“A percentage coefficient of variation (% CV – the ratio of the standard deviation to the mean) less than or equal to 15% was considered acceptable (Burd, 2010).”
BackgroundWaugh, C., & Clark, G. (2021) · Scientific and Technical Review (World Organisation for Animal Health) · doi:10.20506/rst.40.1.3213 · Cited by 35
“) should be taken into account when undertaking and preparing reproducibility assessments (15).”
BackgroundA case for investment in clinical metagenomics in low-income and middle-income countries
Marais, G. J. K., Hardie, D. R., & Brink, A. J. (2022) · The Lancet Microbe · doi:10.1016/s2666-5247(22)00328-7 · Cited by 29
Validating Real-Time Polymerase Chain Reaction (PCR) Assays
Smith, M. H. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-814515-9.00053-9 · Cited by 26
“There are two approaches to evaluating trueness (Burd, 2010): (1) A comparison of methods study, where split samples are tested in parallel with an appropriate gold standard method.”
MethodThe Case for Laboratory Developed Procedures
Kaul, K. L., Sabatini, L. M., Tsongalis, G. J., Caliendo, A. M., Olsen, R. J., Ashwood, E. R., Bale, S. J., Benirschke, R. C., Carlow, D. C., Funke, B., et al. (2017) · Academic Pathology · doi:10.1177/2374289517708309 · Cited by 29
Design, optimization, and application of multiplex rRT-PCR in the detection of respiratory viruses
Yang, J., Li, D., Wang, J., Zhang, R., & Li, J. (2022) · Critical Reviews in Clinical Laboratory Sciences · doi:10.1080/10408363.2022.2072467 · Cited by 28
“It includes analytical sensitivity, analytical specificity, accuracy and precision [83].”
BackgroundLi, Z., Bruce, J. L., Cohen, B., Cunningham, C. V., Jack, W. E., Kunin, K., Langhorst, B. W., Miller, J., Moncion, R. A., Poole, C. B., et al. (2022) · PLoS ONE · doi:10.1371/journal.pone.0268692 · Cited by 26
“The LoD, defined as the lowest concentration that can be detected with a 95% detection rate [37], was calculated to be 39 copies/ μL using an R script that determines LoD using curvefitting methods [38].”
MethodSchroeder, M. E., Johnson, D. J., Ostlund, E. N., Meier, J., Bounpheng, M. A., & Clavijo, A. (2013) · Journal of Veterinary Diagnostic Investigation · doi:10.1177/1040638713503654 · Cited by 25
“Probit analysis was performed using statistical software j to assess the limit of detection of each assay.(8,9,35) Repeatability of the mRT-qPCR was determined by calculation of the intra- and interassay coefficient of variation (CV) for each pathogen assay.”
MethodThanh, T. T., Anh, N. T., Thắm, N. T., Van, H. M. T., Sabanathan, S., Qui, P. T., Ngan, T. T., Van, T. T. M., Nguyet, L. A., Ny, N. T. H., et al. (2015) · Virology Journal · doi:10.1186/s12985-015-0316-2 · Cited by 24
“Note: +ve = positive; -ve = negative # Rectal swab analysis only when patients had throat-swab PCR negative *The analysis ended when rectal swab PCR is completed Lower limit of 95 % detection, linearity and precision Analytical evaluation of the multiplex assay was done as previously described [17, 18] using both plas…”
MethodKan, C.-M., Tsang, H. F., Pei, X. M., Ng, S. S. M., Yim, A. K., Yu, A. C.-S., & Wong, S. C. C. (2024) · International Journal of Molecular Sciences · doi:10.3390/ijms25063333 · Cited by 32
“An integrated quality management approach, in line with regulatory standards, is required for any NGS-based implementation in the clinical laboratory [45].”
BackgroundInhibition monitoring in veterinary molecular testing
Yan, L., Toohey-Kurth, K. L., Crossley, B. M., Bai, J., Glaser, A. L., Tallmadge, R. L., & Goodman, L. B. (2019) · Journal of Veterinary Diagnostic Investigation · doi:10.1177/1040638719889315 · Cited by 26
“tory testing, recommends 20 separate determinations for establishing the initial mean and SDs, followed by monthly updates over a 3–6 mo period to establish a stable performance range.(6) The number of replicates for veterinary testing is ultimately determined by each laboratory and should be documented as part of the…”
BackgroundBruzek, S., Vestal, G. D., Lasher, A., Lima, A., & Silbert, S. (2020) · Journal of Molecular Diagnostics · doi:10.1016/j.jmoldx.2020.09.003 · Cited by 24
Best practices for performance of real-time PCR assays in veterinary diagnostic laboratories
Toohey-Kurth, K. L., Mulrooney, D. M., Hinkley, S., Killian, M. L., Pedersen, J. C., Bounpheng, M. A., Pogranichniy, R. M., Bolin, S. R., Maes, R. K., Tallmadge, R. L., et al. (2020) · Journal of Veterinary Diagnostic Investigation · doi:10.1177/1040638720962076 · Cited by 33
“Instructions for determining the appropriate concentration for routine use of positive amplification controls (PACs) are described elsewhere.(6,88) Briefly, a provisional control concentration is determined based on a recommended 15 independent determinations.”
BackgroundMölsä, M., Hemmilä, H., Katz, A., Niemimaa, J., Forbes, K. M., Huitu, O., Stuart, P., Henttonen, H. A., & Nikkari, S. (2015) · Journal of Microbiological Methods · doi:10.1016/j.mimet.2015.05.026 · Cited by 23
“One of the problemswithmany diagnostic PCR-based methodologies is the potential accumulation of amplified DNA fragments, which eventually lead to false positive results through contamination of sample materials and reagents (Burd, 2010).”
BackgroundLü, S., Duplat, D., Benitez-Bolivar, P., León, C. M., Villota, S. D., Veloz-Villavicencio, E., Arévalo, V., Jaenes, K., Guo, Y., Cicek, S., et al. (2022) · PLoS ONE · doi:10.1371/journal.pone.0268340 · Cited by 25
“The LoD is commonly defined as the concentration of analyte that can be detected in 95% of replicates [35], but as we used 10 replicates in this phase, we defined LoD as the lowest copy number at which sensitivity was 100% after 30 min.”
MethodSARS-CoV-2 infection and immune responses
Harne, R., Williams, B., Abdelaal, H. F. M., Baldwin, S. L., & Coler, R. N. (2023) · AIMS Microbiology · doi:10.3934/microbiol.2023015 · Cited by 25
Ranjbar, R., & Afshar, D. (2015) · Europe PMC (PubMed Central) · Cited by 22
Grimm, J., Krickl, J., Beck, A., Nell, J., Bergmann, M., Tappe, D., Grüner, B., Barth, T. F. E., & Brehm, K. (2021) · PLoS neglected tropical diseases · doi:10.1371/journal.pntd.0009155 · Cited by 21
Kemp, M. W., Miura, Y., Payne, M. S., Jobe, A. H., Kallapur, S. G., Saito, M., Stock, S. J., Spiller, O. B., Ireland, D. J., Yaegashi, N., et al. (2014) · Antimicrobial Agents and Chemotherapy · doi:10.1128/aac.03721-14 · Cited by 21
“The tissue samples were defrosted on wet ice, and 50 ng roxithromycin and 100 mg of tissue were added to a Precellys MK28-R lysing tube (Bertin Corp., Rockville, MD) containing LC-MS grade acetonitrile (Honeywell-Burdick and Jackson, Morristown, NJ) to a final volume of 1,120 l.”
MethodInfluentialSanosyan, A., de Maudave, A. F., Bolloré, K., Zimmermann, V. S., Foulongne, V., de Perre, P. V., & Tuaillon, É. (2017) · PLoS ONE · doi:10.1371/journal.pone.0183856 · Cited by 21
“65 probability of units was used to determine the LOD for both in-house PCRs as described elsewhere [22].”
MethodAnahtar, M. N., Shaw, B. M., Slater, D., Byrne, E. H., Botti-Lodovico, Y., Adams, G. C., Schaffner, S. F., Eversley, J. S., McGrath, G., Gogakos, T., et al. (2021) · Journal of Clinical Pathology · doi:10.1136/jclinpath-2020-207128 · Cited by 20
Sullivan, B. P., Chou, Y.-S., Bender, A. T., Martin, C. D., Kaputa, Z. G., March, H., Song, M., & Posner, J. D. (2022) · Lab on a Chip · doi:10.1039/d2lc00007e · Cited by 20
Kim, J.-H., Chong, C., Sinniah, M., Sinnadurai, J., Song, H., & Park, H. (2015) · Journal of Clinical Virology · doi:10.1016/j.jcv.2015.01.018 · Cited by 20
“The assay was evaluated according to a previous report with modification [40].”
MethodQuantification of cytomegalovirus viral load
Pillet, S., Roblin, X., Cornillon, J., Mariat, C., & Pozzetto, B. (2013) · Expert Review of Anti-infective Therapy · doi:10.1586/14787210.2014.870887 · Cited by 22
“The requirements for quality certification (International Standardization Organization (ISO) 15189 norm) of diagnosis laboratories (for helpful guidelines and checklist see [64,66,67]) led to the increased commercialization of kits that are in vitro diagnostic (IVD)/Conformité européenne (CE)-labeled or US FDAapproved.”
BackgroundClinical utility – Information about the usefulness of tests
Badrick, T. C., & Bowling, F. G. (2023) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2023.110656 · Cited by 30
Raposo, J. V., Alves, A. D. R., da Silva, A. D. S., Santos, D. C. D., Melgaço, J. G., Moreira, O. C., Pinto, M. A., & de Paula, V. S. (2019) · BMC Infectious Diseases · doi:10.1186/s12879-019-4309-4 · Cited by 19
Greninger, A. L., Bard, J. D., Colgrove, R. C., Graf, E. H., Hanson, K. E., Hayden, M. K., Humphries, R. M., Lowe, C. F., Miller, M. B., Pillai, D. R., et al. (2021) · Clinical Infectious Diseases · doi:10.1093/cid/ciab761 · Cited by 20
“Amendments) high-complexity laboratory (4).”
Manta, F. S. D. N., Jacomasso, T., Rampazzo, . D. C. P., Moreira, S. J. M., Zahra, N., Cole, S. T., Avanzi, C., Leal-Calvo, T., Vasconcellos, S. E. G., Suffys, P. N., et al. (2022) · PLoS neglected tropical diseases · doi:10.1371/journal.pntd.0009850 · Cited by 29
“The “analytical sensitivity” or “limit of detection” of an assay is defined as the ability of the assay to detect very low concentrations of a given substance in a biological specimen [47].”
BackgroundMethodMontone, K. T., & Guarner, J. (2013) · Advances in Anatomic Pathology · doi:10.1097/pap.0b013e31828d187d · Cited by 19
Yeung, M., & Thorsen, T. (2016) · Journal of Visualized Experiments · doi:10.3791/54493 · Cited by 18
Rojas, A., Diagne, C. T., Stittleburg, V. D., Mohamed-Hadley, A. A., Guillén, Y., Balmaseda, Á., Faye, O., Faye, O., Sall, A. A., Harris, E., et al. (2018) · American Journal of Tropical Medicine and Hygiene · doi:10.4269/ajtmh.18-0024 · Cited by 18
Munson, E. L., Morgan, E., Sienkiewicz, L., Thomas, Y., Buehler, K., Ryan, D. T., Clifford, A., & Mustanski, B. S. (2020) · Sexually Transmitted Infections · doi:10.1136/sextrans-2020-054463 · Cited by 19
Graham, K. M., Burgess, E. A., & Rolland, R. M. (2021) · Conservation Physiology · doi:10.1093/conphys/coaa133 · Cited by 22
“Hormone extraction should be kept as simple as possible because additional steps increase the extent of variation, which could potentially impact accuracy of the final measurement (Burd, 2010, Palme, 2019).”
BackgroundFerreira, B. I. D. S., Gomes, N. L. D. S., Coelho, W. L. D. C. N. P., da Costa, V. D., Carneiro, V. C. D. S., Kader, R. L., Amaro, M. P., Villar, L. M., Miyajima, F., Alves‐Leon, S. V., et al. (2021) · PLoS ONE · doi:10.1371/journal.pone.0260087 · Cited by 18
“In this study, the analytical validation of HRM-qPCR assays was performed using international guidelines [16] and synthetic sequence templates of the respective SARS-CoV-2 genes.”
MethodDesign and Optimization of a Monkeypox virus Specific Serological Assay
Taha, T. Y., Townsend, M. B., Pohl, J., Karem, K. L., Damon, I. K., Kingebeni, P. M., Tamfum, J. M., Martin, J. W., Pittman, P. R., Huggins, J. W., et al. (2023) · Pathogens · doi:10.3390/pathogens12030396 · Cited by 20
“The background value was the sum of the average OD450 of duplicate background wells for each sample, plus three times the standard deviation of these wells to account for error [41].”
BackgroundThieulent, C., Carossino, M., Peak, L., Strother, K. O., Wolfson, W., & Balasuriya, U. B. R. (2023) · Viruses · doi:10.3390/v15091881 · Cited by 18
“Cycle threshold (Ct) cut-off values were determined using the following formula: Ct cut-off = Average Ct values of 12 replicates of the endpoint dilution + (3 × standard deviation [SD]) [53,54,65].”
MethodTesting for SARS-CoV-2: lessons learned and current use cases
Theel, E. S., Kirby, J. E., & Pollock, N. R. (2024) · Clinical Microbiology Reviews · doi:10.1128/cmr.00072-23 · Cited by 21
“For molecular assays, the LoD by tradition has been defined as the lowest level of a serially diluted standard at which the analyte (SARS-CoV-2 gene target) is detected greater than 95% of the time with repeat measurement (132).”
MicroRNA levels quantified in whole blood varies from PBMCs
Atarod, S., Smith, H., Dickinson, A. M., & Wang, X. (2015) · F1000Research · doi:10.12688/f1000research.4884.4 · Cited by 18
“Performing a comprehensive statistical analysis is required when two methods are compared (Burd, 2010).”
Ultrasensitive Detection of Shigella Species in Blood and Stool
Luo, J., Wang, J., Mathew, A. S., & Yau, S. ‐. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.5b04242 · Cited by 18
Essentials of filoviral load quantification
Cnops, L., van Griensven, J., Honko, A. N., Bausch, D. G., Sprecher, A., Hill, C. E., Colebunders, R., Johnson, J. C., Griffiths, A., Palacios, G. F., et al. (2016) · The Lancet Infectious Diseases · doi:10.1016/s1473-3099(16)30063-9 · Cited by 18
Wong, S. S. Y., Yip, C. C., Sridhar, S., Leung, K., Cheng, A. K. W., Fung, A. M. Y., Lam, H., Chan, K., Chan, J. F., Cheng, V. C., et al. (2018) · Virology Journal · doi:10.1186/s12985-018-1059-7 · Cited by 18
Laboratory verification of an RT‐PCR assay for SARS‐CoV‐2
Wu, Y., Xu, W., Zhu, Z., & Xia, X. (2020) · Journal of Clinical Laboratory Analysis · doi:10.1002/jcla.23507 · Cited by 22
Shrestha, L., Lin, M. J., Xie, H. J., Mills, M. G., Bakhash, S. A. M., Gaur, V. P., Livingston, R. J., Castor, J., Bruce, E. A., Botten, J., et al. (2022) · Journal of Molecular Diagnostics · doi:10.1016/j.jmoldx.2022.05.007 · Cited by 17
Caza, M., Hogan, C. A., Jassem, A. N., Prystajecky, N., Hadzic, A., & Wilmer, A. (2021) · Journal of Clinical Virology · doi:10.1016/j.jcv.2021.104996 · Cited by 19
“Simulated NP swab and SG specimens with known concentration of DNA (50, 100, 200, 300, 400, 500 and 1000 copies/mL) were extracted and tested at least 20 times for each concentration, with LOD calculated as previously described [28].”
Garland, S. M., Iftner, T., Cuschieri, K. S., Kaufmann, A. M., Arbyn, M., de Sanjosé, S., Poljak, M., Dillner, J., Unger, E. R., Garland, S. M., et al. (2022) · Journal of Clinical Virology · doi:10.1016/j.jcv.2022.105349 · Cited by 22
Thieulent, C., Carossino, M., Peak, L., Wolfson, W., & Balasuriya, U. B. R. (2023) · Viruses · doi:10.3390/v15091890 · Cited by 17
“…using SPSS 14.0 software (SPSS Inc., Chicago, IL, USA) from twelve replicates per dilution ranging from 10 3 to 10 0 copies/ µ L. Ct cut-off values were determined using the following formula: Ct cut-off = Average replicate values of the endpoint dilution + (3 × standard deviation [SD]) [53].”
MethodInfluentialSpecimen Collection, Transport, and Processing: Virology
Dunn, J. J. (2015) · ASM Press eBooks · doi:10.1128/9781555817381.ch79 · Cited by 20
Cheng, A., & Kirby, J. E. (2014) · American Journal of Clinical Pathology · doi:10.1309/ajcpfq25sqvzawhz · Cited by 17
Sam, S. S., Kurpewski, J., Cu‐Uvin, S., & Caliendo, A. M. (2016) · Journal of Clinical Microbiology · doi:10.1128/jcm.03289-15 · Cited by 17
“Since poor precision can affect the linearity of the assay (23), precision, especially at low viremia levels, is an important assay characteristic; the Aptima assay showed a tendency of less variability at low viral loads (1.”
BackgroundGreninger, A. L., Bard, J. D., Colgrove, R. C., Graf, E. H., Hanson, K. E., Hayden, M. K., Humphries, R. M., Lowe, C. F., Miller, M. B., Pillai, D. R., et al. (2021) · Journal of Clinical Microbiology · doi:10.1128/jcm.01659-21 · Cited by 16
Molecular diagnostics in clinical microbiology: Advances, applications, and future directions
Asokan, S., Choudekar, A., Jagadeesan, A., Rudresh, S., Ali, A., Napte, S. U., Selvam, S. A., Verma, G., Das, P., Beniwal, N., et al. (2025) · Diagnostic Microbiology and Infectious Disease · doi:10.1016/j.diagmicrobio.2025.117223 · Cited by 26
Tremblay, M.-A., Rodrigue, M.-A., Deschênes, L., Boivin, G., & Longtin, J. (2015) · Journal of Virological Methods · doi:10.1016/j.jviromet.2015.08.010 · Cited by 16
Kim, K.-H., Choi, K.-M., Kang, G., Woo, W.-S., Sohn, M.-Y., Son, H.-J., Yun, D., Kim, D., & Park, C. (2022) · Fishes · doi:10.3390/fishes7050236 · Cited by 16
Validation of the Cepheid GeneXpert for Detecting Ebola Virus in Semen
Loftis, A. J., Quellie, S., Chason, K. D., Sumo, E., Toukolon, M., Otieno, Y., Ellerbrok, H., Hobbs, M. M., Hoover, D., Dubé, K., et al. (2016) · The Journal of Infectious Diseases · doi:10.1093/infdis/jiw562 · Cited by 16
dos Santos, E. H., Yamamoto, L., Domingues, W., Santi, S. M. D., Kanunfre, K. A., & Okay, T. S. (2020) · Parasitology International · doi:10.1016/j.parint.2020.102069 · Cited by 17
Building up a clinical microbiota profiling: a quality framework proposal
Scherz, V., Greub, G., & Bertelli, C. (2021) · Critical Reviews in Microbiology · doi:10.1080/1040841x.2021.1975642 · Cited by 19
Good Clinical Laboratory Practice (GCLP) for Molecular Based Tests Used in Diagnostic Laboratories
Raquel, V., & Carole, L. (2011) · InTech eBooks · doi:10.5772/23963 · Cited by 16
RT-qPCR Testing and Performance Metrics in the COVID-19 Era
Bustin, S. A. (2024) · International Journal of Molecular Sciences · doi:10.3390/ijms25179326 · Cited by 18
“…unprecedented lack of understanding of the nature of this pandemic [82], the complexities involved in diagnostic test development and evaluation [83,84] and the short time frame in which to achieve these, the scientific community reacted with unprecedented speed to produce a number of RT-qPCR…”
BackgroundKardjadj, M. (2025) · The Journal of Applied Laboratory Medicine · doi:10.1093/jalm/jfaf106 · Cited by 20
Preparedness of European diagnostic microbiology labs for detection of SARS-CoV-2, March 2020
Matheeussen, V., Loens, K., Lammens, C., Vilken, T., Koopmans, M., Goossens, H., & Ieven, M. M. (2020) · Journal of Clinical Virology · doi:10.1016/j.jcv.2020.104432 · Cited by 15
Lau, R., Du, L., Chen, E., Fu, C., Gould, R., Marczyk, M., Sinn, B. V., Layman, R. M., Bedrosian, I., Valero, V., et al. (2020) · Clinical Chemistry · doi:10.1093/clinchem/hvaa105 · Cited by 16
Metadata
- Authors
- Burd, E. M.
- Affiliations
- Emory University Hospital (US)
- Corresponding author
- Burd, Eileen M.
- ORCID
- Burd, Eileen M.
- Venue
- Clinical Microbiology Reviews
- Publisher
- American Society for Microbiology
- Published
- 2010-07
- Volume
- 23
- Issue
- 3
- Pages
- 550-576
- ISSN
- 0893-8512 (ISSN-L)
- 1098-6618
- PMC
- PMC2901657
- OpenAlex
- W2096278562
- Cited by
- 550
- References
- 48
- Citations by year
- 2026: 30 · 2025: 43 · 2024: 36 · 2023: 42 · 2022: 42 · 2021: 41 · 2020: 38 · 2019: 36 · 2018: 39 · 2017: 37 · 2016: 44 · 2015: 35 · 2014: 32 · 2013: 21 · 2012: 18
- FWCI
- 1.74
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- journals.asm.org/non-commercial-tdm-license (tdm)
- Locations
- Clinical Microbiology Reviews (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Microbiology (medical) › Bacterial Identification and Susceptibility Testing
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Healthcare Regulation
- Keywords
- laboratory-developed tests, molecular diagnosis, infectious diseases, assay validation, analytical sensitivity, analytical specificity, precision and accuracy, reference values, assay modification
- MeSH terms
- Communicable Diseases, Humans, Clinical Laboratory Techniques, Molecular Diagnostic Techniques, Validation Studies as Topic
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.