The DOI matches its Crossref record; 16 of the 16 references it cites were found in a registry or an index.
Continuous non-invasive cardiac output measurements in the neonate by electrical velocimetry: a comparison with echocardiography
Abstract
OBJECTIVE: Electrical velocimetry (EV) is a non-invasive method of continuous left cardiac output monitoring based on measurement of thoracic electrical bioimpedance. The objective was to validate EV by investigating the agreement in cardiac output measurements performed by EV and echocardiography. DESIGN: In this prospective observational study, left ventricular output (LVO) was simultaneously measured by EV (LVO(ev)) using Aesculon and by echocardiography (LVO(echo)) in healthy term neonates during the first 2 postnatal days. To determine the agreement between the two methods, we calculated the bias (mean difference) and precision (1.96×SD of the difference). As LVO(echo) has its own limitations, the authors also calculated the 'true precision' of EV adjusted for echocardiography as the reference method. RESULTS: The authors performed 115 paired measurements in 20 neonates. LVO(ev) and LVO(echo) were similar (534±105 vs 538±105 ml/min, p=0.7). The bias and precision of EV were -4 and 234 ml/min, respectively. The authors found the true precision of EV to be similar to the precision of echocardiography (31.6% vs 30%, respectively). There was no difference in bias and precision between the measurements obtained in patients with or without a haemodynamically significant patent ductus arteriosus. CONCLUSIONS: EV is as accurate in measuring LVO as echocardiography and the variation in the agreement between EV and echocardiography among the individual subjects reflects the limitations of both techniques.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“while Noori et al. examined left ventricular output in preterm neonates with and without patent ductus arteriosus, finding good agreement between the two modalities [73].”
From Trends and Challenges in Noninvasive Hemodynamic Monitoring of Neonates Following Cardiac Surgery: A Narrative Review (2025) · cites this paper as [73]
Citations
Cites
1 retracted
Suttner, S. W., Schöllhorn, T., Boldt, J., Mayer, J., Röhm, K. D., Lang, K., & Piper, S. N. (2006) · Intensive Care Medicine · doi:10.1007/s00134-006-0409-x · Cited by 88
RetractedAgreement Between Methods of Measurement with Multiple Observations Per Individual
Bland, J. M., & Altman, D. G. (2007) · Journal of Biopharmaceutical Statistics · doi:10.1080/10543400701329422 · Cited by 1,864
Critchley, L. A. H., & Critchley, J. (1999) · Journal of Clinical Monitoring and Computing · doi:10.1023/a:1009982611386 · Cited by 1,328
Development and evaluation of an impedance cardiac output system.
Kubicek, W. G., Karnegis, J. N., PATTERSON, R. P., Witsoe, D. A., & Mattson, R. H. (1966) · PubMed · Cited by 1,143
Cecconi, M., Rhodes, A., Poloniecki, J., Della Rocca, G., & Grounds, R. M. (2009) · Critical Care · doi:10.1186/cc7129 · Cited by 351
Kluckow, M. R., & Evans, N. (1995) · The Journal of Pediatrics · doi:10.1016/s0022-3476(95)70172-9 · Cited by 265
Noninvasive pulsed Doppler determination of cardiac output in neonates and children
Alverson, D. C., Eldridge, M. W., Dillon, T., Yabek, S. M., & Berman, W. (1982) · The Journal of Pediatrics · doi:10.1016/s0022-3476(82)80178-9 · Cited by 232
Chew, M. S., & Poelaert, J. (2003) · Intensive Care Medicine · doi:10.1007/s00134-003-1967-9 · Cited by 151
Electrical velocimetry for measuring cardiac output in children with congenital heart disease
Norozi, K., Beck, C. E., Osthaus, W. A., Wille, I., Wessel, A., & Bertram, H. (2007) · British Journal of Anaesthesia · doi:10.1093/bja/aem320 · Cited by 128
Zoremba, N., Bickenbach, J., Krauss, B., Rossaint, R., Kuhlen, R., & Schälte, G. (2007) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.2007.01445.x · Cited by 106
Cited by
Sorted by citations
Neonatal Hemodynamics: From Developmental Physiology to Comprehensive Monitoring
Vrancken, S. L. A. G., van Heijst, A. F. J., & de Boode, W. P. (2018) · Frontiers in Pediatrics · doi:10.3389/fped.2018.00087 · Cited by 124
Sanders, M. L., Servaas, S., & Slagt, C. (2019) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-019-00330-y · Cited by 101
Hemodynamic Assessment and Monitoring of Premature Infants
EL-Khuffash, A. F., & McNamara, P. J. (2017) · Clinics in Perinatology · doi:10.1016/j.clp.2017.02.001 · Cited by 73
Suehiro, K., Joosten, A., Murphy, L. S.-L., Desebbe, O., Alexander, B. S., Kim, S.-H., & Cannesson, M. P. (2015) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-015-9757-9 · Cited by 73
Advances in Diagnosis and Management of Hemodynamic Instability in Neonatal Shock
Singh, Y., Katheria, A. C., & Vora, F. (2018) · Frontiers in Pediatrics · doi:10.3389/fped.2018.00002 · Cited by 68
Blohm, M. E. G., Obrecht, D., Hartwich, J., Mueller, G. C., Kersten, J. F., Weil, J., & Singer, D. (2014) · Critical Care · doi:10.1186/s13054-014-0603-0 · Cited by 63
Boët, A., Jourdain, G., Demontoux, S., & Luca, D. D. (2016) · Journal of Perinatology · doi:10.1038/jp.2016.65 · Cited by 62
Rauch, R., Welisch, E., Lansdell, N., Burrill, E., Jones, J., Robinson, T., Bock, D. E., Clarson, C. L., Filler, G., & Norozi, K. (2012) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-012-9412-7 · Cited by 59
Hsu, K.-H., Wu, T.-W., Wang, Y., Lim, W.-H., Lee, C.-C., & Lien, R.-I. (2016) · Journal of Perinatology · doi:10.1038/jp.2016.2 · Cited by 57
What Should We Do about Low Blood Pressure in Preterm Infants
Dempsey, E. M. (2017) · Neonatology · doi:10.1159/000460603 · Cited by 55
Non-Invasive Cardiac Output Measurement in Low and Very Low Birth Weight: A Method Comparison
Grollmuss, O., & Gonzalez, P. (2014) · Frontiers in Pediatrics · doi:10.3389/fped.2014.00016 · Cited by 52
Weber, F., & Scoones, G. P. (2019) · Pediatric Anesthesia · doi:10.1111/pan.13726 · Cited by 53
Coté, C. J., Sui, J., Anderson, T. A., Bhattacharya, S. T., Shank, E. S., Tuason, P., August, D. A., Zibaitis, A., Firth, P. G., Fuzaylov, G., et al. (2014) · Pediatric Anesthesia · doi:10.1111/pan.12441 · Cited by 51
Wu, T.-W., Tamrazi, B., Soleymani, S., Seri, I., & Noori, S. (2018) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2018.01.067 · Cited by 49
Self-citationThe Pathophysiology of Low Systemic Blood Flow in the Preterm Infant
Kluckow, M. R. (2018) · Frontiers in Pediatrics · doi:10.3389/fped.2018.00029 · Cited by 48
Prone positioning decreases cardiac output and increases systemic vascular resistance in neonates
Ma, M., Noori, S., Maarek, J.-M. I., Holschneider, D. P., Rubinstein, E. H., & Seri, I. (2015) · Journal of Perinatology · doi:10.1038/jp.2014.230 · Cited by 46
Self-citationSu, B., Lin, S.-P., Huang, F., Tsai, M., & Huang, Y.-T. (2016) · Pediatrics & Neonatology · doi:10.1016/j.pedneo.2016.01.001 · Cited by 46
Vergnaud, E., Vidal, C., Verchère, J., Miatello, J., Meyer, P. G., Carli, P. A., & ORLIAGUET, G. A. (2014) · British Journal of Anaesthesia · doi:10.1093/bja/aeu361 · Cited by 45
Noori, S., & Seri, I. (2015) · Seminars in Fetal and Neonatal Medicine · doi:10.1016/j.siny.2015.03.005 · Cited by 41
Self-citationHaemodynamic Transition after Birth: A New Tool for Non-Invasive Cardiac Output Monitoring
Freidl, T., Baik‐Schneditz, N., Pichler, G., Schwaberger, B., Zingerle, B., Avian, A., & Urlesberger, B. (2016) · Neonatology · doi:10.1159/000446468 · Cited by 41
Torigoe, T., Sato, S., Nagayama, Y., Sato, T., & Yamazaki, H. (2015) · Journal of Perinatology · doi:10.1038/jp.2014.245 · Cited by 41
Narula, J., Chauhan, S., Ramakrishnan, S., & Gupta, S. K. (2017) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1053/j.jvca.2016.12.009 · Cited by 39
Non-invasive Cardiac Output Monitoring in Neonates
O'Neill, R., Dempsey, E. M., Garvey, A. A., & Schwarz, C. E. (2021) · Frontiers in Pediatrics · doi:10.3389/fped.2020.614585 · Cited by 38
Neurocritical care of high‐risk infants during inter‐hospital transport
Gupta, N., Shipley, L., Goel, N., Carmo, K. B., Leslie, A., & Sharkey, D. (2019) · Acta Paediatrica · doi:10.1111/apa.14940 · Cited by 36
Aziz, K. B., Lavilla, O. C., Wynn, J. L., Lure, A. C., Gipson, D. R., & de la Cruz, D. (2021) · Journal of Perinatology · doi:10.1038/s41372-021-01030-9 · Cited by 36
Hsu, K., Wu, T.-W., Wu, I.-H., Lai, M., Hsu, S.-Y., Huang, H.-W., Mok, T., & Lien, R. (2017) · Neonatology · doi:10.1159/000475774 · Cited by 35
Díllí, D., Soylu, H., & Tekın, N. (2019) · Türk Pediatri Arşivi · doi:10.5152/turkpediatriars.2018.01801 · Cited by 35
Azhibekov, T., Noori, S., Soleymani, S., & Seri, I. (2013) · Seminars in Fetal and Neonatal Medicine · doi:10.1016/j.siny.2013.09.009 · Cited by 35
Self-citationAdvanced Hemodynamic Monitoring in the Neonatal Intensive Care Unit
de Boode, W.-P. (2020) · Clinics in Perinatology · doi:10.1016/j.clp.2020.05.001 · Cited by 34
Time Course Study of Blood Pressure in Term and Preterm Infants Immediately after Birth
Pichler, G., Cheung, P., Binder, C., O’Reilly, M., Schwaberger, B., Aziz, K., Urlesberger, B., & Schmölzer, G. M. (2014) · PLoS ONE · doi:10.1371/journal.pone.0114504 · Cited by 34
Variability in the Hemodynamic Response to Fluid Bolus in Pediatric Septic Shock
Ranjit, S., Natraj, R., Kissoon, N. “., Thiagarajan, R. R., Ramakrishnan, B., & García, M. I. M. (2021) · Pediatric Critical Care Medicine · doi:10.1097/pcc.0000000000002714 · Cited by 32
Katheria, A. C., Wozniak, M., Harari, D., Arnell, K., Petruzzelli, D. L., & Finer, N. N. (2015) · Maternal Health Neonatology and Perinatology · doi:10.1186/s40748-015-0016-3 · Cited by 31
Sehgal, A., Steenhorst, J. J., McLennan, D. I., Merkus, D., Ivy, D. D., & McNamara, P. J. (2020) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2020.06.031 · Cited by 29
Lien, R., Hsu, K., Chu, J., & Chang, Y. (2014) · European Journal of Pediatrics · doi:10.1007/s00431-014-2437-9 · Cited by 28
EL-Khuffash, A. F., McNamara, P. J., Breatnach, C. R., Bussmann, N., Smith, A. M., Feeney, O., Tully, E. C., Griffin, J., de Boode, W. P., Cleary, B., et al. (2018) · Maternal Health Neonatology and Perinatology · doi:10.1186/s40748-018-0093-1 · Cited by 27
Hemodynamic monitoring of the critically ill neonate: An eye on the future
Azhibekov, T., Soleymani, S., Lee, B. H., Noori, S., & Seri, I. (2015) · Seminars in Fetal and Neonatal Medicine · doi:10.1016/j.siny.2015.03.003 · Cited by 27
Self-citationThoracic fluid content by electric bioimpedance correlates with respiratory distress in newborns
Paviotti, G., Cunto, A. D., Moressa, V., Bettiol, C. S., & Demarini, S. (2017) · Journal of Perinatology · doi:10.1038/jp.2017.100 · Cited by 27
Wu, T.-W., Lien, R., Seri, I., & Noori, S. (2017) · Archives of Disease in Childhood Fetal & Neonatal · doi:10.1136/archdischild-2016-311769 · Cited by 24
Self-citationAssessment of neonatal perfusion
Gupta, S., & Donn, S. M. (2020) · Seminars in Fetal and Neonatal Medicine · doi:10.1016/j.siny.2020.101144 · Cited by 24
Neonatal hemodynamics: Monitoring, data acquisition and analysis
Soleymani, S., Borzage, M., Noori, S., & Seri, I. (2012) · Expert Review of Medical Devices · doi:10.1586/erd.12.32 · Cited by 23
Self-citationNederend, I., ten Harkel, A. T., Blom, N., Berntson, G., & de Geus, E. D. (2017) · International Journal of Psychophysiology · doi:10.1016/j.ijpsycho.2017.07.015 · Cited by 20
Cardiac Output Monitoring in Preterm Infants
McGovern, M., & Miletín, J. (2018) · Frontiers in Pediatrics · doi:10.3389/fped.2018.00084 · Cited by 23
Noninvasive hemodynamic monitoring of septic shock in children
Fathi, E. M., Narchı, H., & Chedid, F. D. (2018) · World Journal of Methodology · doi:10.5662/wjm.v8.i1.1 · Cited by 22
Eriksen, V. R., Trautner, S., Hahn, G. H., & Greisen, G. O. (2019) · PLoS ONE · doi:10.1371/journal.pone.0213537 · Cited by 22
Pharmacokinetics and Pharmacology of Drugs Used in Children
Anderson, B. J., Lerman, J., & Coté, C. J. (2019) · Elsevier eBooks · doi:10.1016/b978-0-323-42974-0.00007-0 · Cited by 22
Mansfield, R. C., Kaza, N., Charalambous, A., Milne, A. C., Sathiyamurthy, S., & Banerjee, J. (2021) · Critical Care Medicine · doi:10.1097/ccm.0000000000005144 · Cited by 20
Non-invasive cardiac output monitor validation study in pediatric cardiac surgery patients
Tirotta, C. F., Lagueruela, R. G., MADRIL, D. R., Velis, E., Ojito, J. W., Monroe, D., Aguero, D., Irizarry, M., McBride, J., Hannan, R. L., et al. (2017) · Journal of Clinical Anesthesia · doi:10.1016/j.jclinane.2017.02.001 · Cited by 20
van Wyk, L., Gupta, S., Lawrenson, J., & de Boode, W. P. (2022) · Frontiers in Pediatrics · doi:10.3389/fped.2022.851850 · Cited by 19
Bernstein, D. P., Henry, I., Lemmens, H. J. M., Chaltas, J. L., DeMaria, A. N., Moon, J., & Kahn, A. M. (2015) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-015-9668-9 · Cited by 18
Intervention and Outcome for Neonatal Hypotension
Barrington, K. J., EL-Khuffash, A. F., & Dempsey, E. M. (2020) · Clinics in Perinatology · doi:10.1016/j.clp.2020.05.011 · Cited by 17
Martini, S., Frabboni, G., Rucci, P., Czosnyka, M., Smielewski, P. S., Galletti, S., Cimatti, A. G., Faldella, G., Corvaglia, L., & Austin, T. (2020) · The Journal of Physiology · doi:10.1113/jp279730 · Cited by 17
Cerebral oxygenation in preterm infants receiving transfusion
Jain, D., D’Ugard, C., Bancalari, E., & Claure, N. (2018) · Pediatric Research · doi:10.1038/s41390-018-0266-7 · Cited by 17
Boët, A., Jourdain, G., Demontoux, S., Hascoët, S., Tissières, P., Rücker‐Martin, C., & Luca, D. D. (2017) · Pediatric Critical Care Medicine · doi:10.1097/pcc.0000000000001298 · Cited by 16
Martin, E. I., Anyikam, A. L., Ballas, J., Buono, K., Mantell, K., Huynh‐Covey, T., & Archer, T. L. (2015) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-015-9771-y · Cited by 15
Hemodynamics and Perfusion in Premature Infants During Transfusion
Weaver, B., Guerreso, K., Conner, E. A., Russell, K., Vogel, R. L., & Rodriguez, M. (2018) · AACN Advanced Critical Care · doi:10.4037/aacnacc2018402 · Cited by 15
Martini, S., Gatelli, I. F., Vitelli, O., Galletti, S., Camela, F., Rienzo, F. D., Martinelli, S., & Corvaglia, L. (2022) · European Journal of Pediatrics · doi:10.1007/s00431-022-04764-6 · Cited by 14
de la Blanca, A. R. S., Luna, M. S., González‐Pacheco, N., Arriaga-Redondo, M., & Patiño, N. N. (2017) · European Journal of Pediatrics · doi:10.1007/s00431-017-3063-0 · Cited by 14
Blohm, M. E. G., Hartwich, J., Obrecht, D., Kersten, J. F., & Singer, D. (2016) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-016-9878-9 · Cited by 14
Liu, Y. H., Dhakal, B. P., Keesakul, C., Kacmarek, R. M., Lewis, G. D., & Jiang, Y. (2016) · British Journal of Anaesthesia · doi:10.1093/bja/aew156 · Cited by 13
Patra, A. P., Thakkar, P. S., Makhoul, M., & Bada, H. S. (2021) · Frontiers in Pediatrics · doi:10.3389/fped.2021.648584 · Cited by 13
Wyllie, J. P. (2015) · Seminars in Fetal and Neonatal Medicine · doi:10.1016/j.siny.2015.03.009 · Cited by 13
Sex related difference in cardiac output during neonatal transition in term neonates
Baik‐Schneditz, N., Schwaberger, B., Mileder, L. P., Höller, N., Avian, A., Köestenberger, M., Urlesberger, B., Martensen, J., & Pichler, G. (2021) · Cardiovascular Diagnosis and Therapy · doi:10.21037/cdt-20-844 · Cited by 12
Monitorages physiopathologiques en réanimation néonatale
Luca, D. D., Romain, O., Yousef, N., Andriamanamirija, D., Shankar‐Aguilera, S., Walls, E., Sgaggero, B., Aubé, N., & Tissières, P. (2015) · Journal de Pédiatrie et de Puériculture · doi:10.1016/j.jpp.2015.06.004 · Cited by 12
Baik‐Schneditz, N., Urlesberger, B., Schwaberger, B., Schmölzer, G. M., Köstenberger, M., Avian, A., & Pichler, G. (2016) · Early Human Development · doi:10.1016/j.earlhumdev.2016.08.009 · Cited by 12
van Wyk, L., Smith, J., Lawrenson, J., & de Boode, W. P. (2020) · Neonatology · doi:10.1159/000506203 · Cited by 12
Xu, S., Zhang, J., Zhang, Y., Zhang, P., & Cheng, G. (2021) · Translational Pediatrics · doi:10.21037/tp-21-20 · Cited by 11
A recommendation for the use of electrical biosensing technology in neonatology
van Wyk, L., Austin, T., Barzilay, B., Bravo, M. C., Breindahl, M., Czernik, C., Dempsey, E. M., de Boode, W. P., de Vries, W. B., Eriksen, B. H., et al. (2024) · Pediatric Research · doi:10.1038/s41390-024-03369-z · Cited by 11
A Randomized Controlled Trial of a 30- versus a 120-Second Delay in Cord Clamping after Term Birth
Soliman, R. M., Elgendy, M. M., Said, R. N., Shaarawy, B. I., Helal, O. M., & Aly, H. Z. (2022) · American Journal of Perinatology · doi:10.1055/a-1772-4543 · Cited by 11
Noninvasive cardiac output measurement using bioreactance in postoperative pediatric patients
Vergnaud, E., Vidal, C., Montmayeur, J., Taright, H., Meyer, P. G., Carli, P. A., & ORLIAGUET, G. A. (2014) · Pediatric Anesthesia · doi:10.1111/pan.12412 · Cited by 11
King, M. R., Anderson, T. A., Sui, J., He, G., Poon, K. Y. T., & Coté, C. J. (2016) · Journal of Clinical Anesthesia · doi:10.1016/j.jclinane.2016.02.026 · Cited by 11
Paviotti, G., Todero, S., & Demarini, S. (2017) · Journal of Perinatology · doi:10.1038/jp.2016.251 · Cited by 11
McNamara, P. J., & EL-Khuffash, A. F. (2016) · Elsevier eBooks · doi:10.1016/b978-0-323-35214-7.00071-8 · Cited by 11
Altamirano‐Diaz, L., Welisch, E., Dempsey, A. A., Park, T. S., Grattan, M., & Norozi, K. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aac02b · Cited by 11
Hsu, K., Wu, T.-W., Wu, I.-H., Lai, M., Hsu, S.-Y., Huang, H.-W., Mok, T., Lee, C., & Lien, R. (2019) · BMC Pediatrics · doi:10.1186/s12887-019-1560-1 · Cited by 11
EL-Fattah, N. M. A., Elmahdy, H. S., Hamisa, M. F., & Ibrahim, A. M. (2024) · European Journal of Pediatrics · doi:10.1007/s00431-024-05507-5 · Cited by 10
McCarthy, K. N., Pavel, A., Garvey, A. A., Hawke, A.-L., Levins, C., Livingstone, V., & Dempsey, E. M. (2020) · Archives of Disease in Childhood Fetal & Neonatal · doi:10.1136/archdischild-2019-318244 · Cited by 10
Hassan, M. A., Bryant, M. B., & Hummler, H. (2021) · Neonatology · doi:10.1159/000519713 · Cited by 10
Sun, H., Stockbridge, N., Ariagno, R. L., Murphy, D., Nelson, R. M., & Rodriguez, W. J. (2016) · Journal of Perinatology · doi:10.1038/jp.2016.103 · Cited by 10
Awadhare, P., Patel, R., McCallin, T., Mainali, K., Jackson, K., Starke, H., & Bhalala, U. (2022) · Frontiers in Pediatrics · doi:10.3389/fped.2022.857106 · Cited by 9
Cardiac output monitoring in children: a review
Cave, D. G. W., Bautista, M., Mustafa, K., & Bentham, J. R. (2023) · Archives of Disease in Childhood · doi:10.1136/archdischild-2022-325030 · Cited by 9
Slagt, C., Servaas, S., Ketelaars, R., van Geffen, G., Tacken, M. C. T., Verrips, C. A., Baggen, L. A. M., Scheffer, G. J., & van Eijk, L. T. (2021) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-021-00657-5 · Cited by 9
Gatelli, I. F., Vitelli, O., Fossati, M., Rienzo, F. D., Chiesa, G. B., & Martinelli, S. (2021) · American Journal of Perinatology · doi:10.1055/s-0041-1726123 · Cited by 9
Sumbel, L., Annamalai, M., Wats, A., Salameh, M., Agarwal, A., & Bhalala, U. S. (2020) · Journal of Pediatric Intensive Care · doi:10.1055/s-0040-1718867 · Cited by 8
Balog, V., Vatai, B., Kovács, K. D., Szabo, A. J., Szabó, M., & Jermendy, Á. (2023) · Frontiers in Pediatrics · doi:10.3389/fped.2023.1112959 · Cited by 8
Pedgaonkar, R. A., Singh, N., Dhananjaya, M., Nagaraja, P., Nagesh, K., & Prabhakar, V. (2023) · Annals of Cardiac Anaesthesia · doi:10.4103/aca.aca_9_23 · Cited by 9
Defining information needs in neonatal resuscitation with work domain analysis
Zestic, J., Sanderson, P. M., Dawson, J. A., & Liley, H. G. (2020) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-020-00526-7 · Cited by 8
Noninvasive Cardiac Output Monitoring in Newborn with Hypoplastic Left Heart Syndrome
Gatelli, I. F., Vitelli, O., Chiesa, G. B., Rienzo, F. D., & Martinelli, S. (2020) · American Journal of Perinatology · doi:10.1055/s-0040-1713603 · Cited by 8
Effect of hemodialysis on impedance cardiography (electrical velocimetry) parameters in children
Wilken, M., Oh, J., Pinnschmidt, H. O., Singer, D., & Blohm, M. E. (2020) · Pediatric Nephrology · doi:10.1007/s00467-019-04409-1 · Cited by 8
Elgebaly, A. S., Anwar, A. G., Fathy, S., Sallam, A. A., & Elbarbary, Y. (2020) · Annals of Cardiac Anaesthesia · doi:10.4103/aca.aca_196_18 · Cited by 7
The Use of Epinephrine in Caudal Anesthesia Increases Stroke Volume and Cardiac Output in Children
Liu, C. A., Sui, J., Coté, C. J., & Anderson, T. A. (2016) · Regional Anesthesia & Pain Medicine · doi:10.1097/aap.0000000000000498 · Cited by 7
Objective cardiovascular assessment in the neonatal intensive care unit
Dempsey, E. M., & EL-Khuffash, A. F. (2017) · Archives of Disease in Childhood Fetal & Neonatal · doi:10.1136/archdischild-2017-313837 · Cited by 7
Bioreactance Cardiac Output Trending Ability in Preterm Infants: A Single Centre, Longitudinal Study
van Wyk, L., Smith, J., Lawrenson, J., Lombard, C. J., & de Boode, W. P. (2021) · Neonatology · doi:10.1159/000518656 · Cited by 6
Assessment of Cardiac Output in Neonates
Osypka, M. J., Soleymani, S., Seri, I., & Noori, S. (2012) · Elsevier eBooks · doi:10.1016/b978-1-4377-2763-0.00006-8 · Cited by 6
Self-citationIn-Silico Evaluation of Anthropomorphic Measurement Variations on Electrical Cardiometry in Neonates
Healy, D., Dempsey, E. M., O’Toole, J. M., & Schwarz, C. E. (2021) · Children · doi:10.3390/children8100936 · Cited by 5
Falciola, V., Donath, S. M., Roden, E., Davidson, A. J., & Vutskits, L. (2022) · British Journal of Anaesthesia · doi:10.1016/j.bja.2021.12.032 · Cited by 5
Az ICON® elektromos kardiometrián alapuló nem invazív hemodinamikai monitor használata a klinikumban
Zakariás, D., Marics, G., Kovács, K. D., Jermendy, Á., Vatai, B., Schuster, G., Tóth‐Heyn, P., Attila, S. J., & Lódi, C. (2018) · Orvosi Hetilap · doi:10.1556/650.2018.31225 · Cited by 5
Schwarz, C. E., O’Toole, J. M., Healy, D., Panaviene, J., Livingstone, V., & Dempsey, E. M. (2024) · European Journal of Pediatrics · doi:10.1007/s00431-023-05387-1 · Cited by 4
Parladori, R., Austin, T., Smielewski, P. S., Czosnyka, M., Paoletti, V., Vitali, F., Corvaglia, L., & Martini, S. (2024) · Pediatric Research · doi:10.1038/s41390-024-03194-4 · Cited by 4
Hasija, S., Chauhan, S., Makhija, N., Singh, S. P., Kumar, S., Choudhury, A., Talwar, S., & Kiran, U. (2014) · World Journal of Cardiovascular Surgery · doi:10.4236/wjcs.2014.410024 · Cited by 4
Consistency between Impedance Technique and Echocardiogram Hemodynamic Measurements in Neonates
Wu, W., Lin, S., Xie, C.-X., Li, J., Lie, J. T., & Qiu, S. (2020) · American Journal of Perinatology · doi:10.1055/s-0040-1710030 · Cited by 4
Metadata
- Authors
- Noori, S., Drabu, B., Soleymani, S., & Seri, I.
- Affiliations
- University of Southern California (US)
- Children's Hospital of Oklahoma (US)
- 1Division of Neonatology and the Center for Fetal and Neonatal Medicine, Department of Pediatrics, Children's Hospital Los Angeles and the LAC+USC Medical Center, Keck School of Medicine, University of Southern California, Los Angeles, CA, USA
- Corresponding author
- Noori, Shahab
- ORCID
- Noori, Shahab
- Venue
- Archives of Disease in Childhood - Fetal and Neonatal Edition
- Publisher
- BMJ
- Published
- 2012-08-29
- Volume
- 97
- Issue
- 5
- Pages
- F340-F343
- ISSN
- 1359-2998 (ISSN-L)
- 1468-2052
- OpenAlex
- W2114998295
- Cited by
- 146
- References
- 16
- Citations by year
- 2026: 4 · 2025: 6 · 2024: 10 · 2023: 8 · 2022: 8 · 2021: 14 · 2020: 12 · 2019: 15 · 2018: 13 · 2017: 13 · 2016: 17 · 2015: 11 · 2014: 9 · 2013: 1 · 2012: 5
- FWCI
- 8.43
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Archives of Disease in Childhood Fetal & Neonatal (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrical and Bioimpedance Tomography
- Keywords
- electrical velocimetry, cardiac output monitoring, neonates, impedance cardiography, echocardiography, left ventricular output, measurement agreement, patent ductus arteriosus, noninvasive hemodynamic monitoring
- MeSH terms
- Cardiac Output, Echocardiography, Female, Humans, Infant, Newborn, Infant, Premature, Male, Monitoring, Physiologic, Prospective Studies, Rheology, Ventricular Function, Left, Critical Illness
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review