Multinational Study of Subcutaneous Model-Predictive Closed-Loop Control in Type 1 Diabetes Mellitus: Summary of the Results
Open accessThe DOI matches its Crossref record; 25 of the 25 references it cites were found in a registry or an index.
Abstract
Background: In 2008–2009, the first multinational study was completed comparing closed-loop control (artificial pancreas) to state-of-the-art open-loop therapy in adults with type 1 diabetes mellitus (T1DM). Methods: The design of the control algorithm was done entirely in silico, i.e., using computer simulation experiments with N = 300 synthetic “subjects” with T1DM instead of traditional animal trials. The clinical experiments recruited 20 adults with T1DM at the Universities of Virginia (11); Padova, Italy (6); and Montpellier, France (3). Open-loop and closed-loop admission was scheduled 3–4 weeks apart, continued for 22 h (14.5 h of which were in closed loop), and used a continuous glucose monitor and an insulin pump. The only difference between the two sessions was that insulin dosing was performed by the patient under a physician's supervision during open loop, whereas insulin dosing was performed by a control algorithm during closed loop. Results: In silico design resulted in rapid (less than 6 months compared to years of animal trials) and cost-effective system development, testing, and regulatory approvals in the United States, Italy, and France. In the clinic, compared to open-loop, closed-loop control reduced nocturnal hypoglycemia (blood glucose below 3.9 mmol/liter) from 23 to 5 episodes (p < .01) and increased the amount of time spent overnight within the target range (3.9 to 7.8 mmol/liter) from 64% to 78% (p = .03). Conclusions: In silico experiments can be used as viable alternatives to animal trials for the preclinical testing of insulin treatment strategies. Compared to open-loop treatment under identical conditions, closed-loop control improves the overnight regulation of diabetes.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Only three months later, in April 2008, the first human trials began at the Universities of Virginia, Montpellier (France), and Padova (Italy), using a control system designed entirely in silico.69”
From Developing the UVA/Padova Type 1 Diabetes Simulator: Modeling, Validation, Refinements, and Utility (2023) · cites this paper as [69]
Citations
Cites
Continuous glucose monitoring and intensive treatment of type 1 diabetes
(2008) · New England Journal of Medicine · doi:10.1056/nejmoa0805017 · Cited by 1,559
Nonlinear model predictive control of glucose concentration in subjects with type 1 diabetes
Hovorka, R., Canonico, V., Chassin, L. J., Haueter, U., Massi-Benedetti, M., Federici, M. O., Pieber, T. R., Schaller, H. C., Schaupp, L., Vering, T., et al. (2004) · Physiological Measurement · doi:10.1088/0967-3334/25/4/010 · Cited by 1,274
Meal Simulation Model of the Glucose-Insulin System
Dalla Man, C., Rizza, R. A., & Cobelli, C. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.893506 · Cited by 955
Self-citationIn Silico Preclinical Trials: A Proof of Concept in Closed-Loop Control of Type 1 Diabetes
Kovatchev, B. P., Breton, M., Dalla Man, C., & Cobelli, C. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300106 · Cited by 760
Self-citationWeinzimer, S. A., Steil, G. M., Swan, K. L., Dziura, J., Kurtz, N., & Tamborlane, W. V. (2008) · Diabetes Care · doi:10.2337/dc07-1967 · Cited by 567
A model-based algorithm for blood glucose control in Type I diabetic patients
Parker, R. S., Doyle, F. J., & Peppas, N. A. (1999) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.740877 · Cited by 500 · cited 4×
“, by a series of computer simulations performed on virtual subjects; (2) clinical experiments performed in three independent centers in the United States, Italy, and France, which enhances the external validity of the results; and (3) automated (as opposed to manual) transfer of CGM data to the control system, which b…”
BackgroundInfluentialFeasibility of Automating Insulin Delivery for the Treatment of Type 1 Diabetes
Steil, G. M., Rebrin, K., Darwin, C., Hariri, F., & Saad, M. F. (2006) · Diabetes · doi:10.2337/db06-0419 · Cited by 473
Hovorka, R., Allen, J. M., Elleri, D., Chassin, L. J., Harris, J., Xing, D., Kollman, C., Hovorka, T., Larsen, A. M. F., Nodale, M., et al. (2010) · The Lancet · doi:10.1016/s0140-6736(09)61998-x · Cited by 450
Continuous glucose monitoring and closed‐loop systems
Hovorka, R. (2006) · Diabetic Medicine · doi:10.1111/j.1464-5491.2005.01672.x · Cited by 443 · cited 4×
“The clinical experiments recruited 20 adults with T1DM at the Universities of Virginia (11); Padova, Italy (6); and Montpellier, France (3).”
BackgroundInfluentialA Bihormonal Closed-Loop Artificial Pancreas for Type 1 Diabetes
El-Khatib, F. H., Russell, S. J., Nathan, D. M., Sutherlin, R. G., & Damiano, E. R. (2010) · Science Translational Medicine · doi:10.1126/scitranslmed.3000619 · Cited by 403
Cited by
Items 101–200 of 201
Patient Perception and Satisfaction With Insulin Pump System: Pilot User Experience Survey
Grando, A., Bayuk, M., Karway, G. K., Corrette, K., Groat, D., Cook, C. B., & Thompson, B. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819843146 · Cited by 21
“For those who used the pump as part of the hybrid closed loop, the views were mostly positive: very satisfied (7), satisfied (6), dissatisfied (1), and very dissatisfied (3).”
BackgroundClosed loop control for type 1 diabetes
Kovatchev, B. P. (2011) · BMJ · doi:10.1136/bmj.d1911 · Cited by 22
Self-citationSrinivasan, A., Lee, J. B., Dassau, E., & Doyle, F. J. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814543660 · Cited by 20
Abbes, I. B., Richard, P.-Y., Lefebvre, M.-A., Guilhem, I., & Poirier, J.-Y. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700315 · Cited by 22
Design and in silico evaluation of an intraperitoneal–subcutaneous (IP–SC) artificial pancreas
Lee, J. J., Dassau, E., Zisser, H., & Doyle, F. J. (2014) · Computers & Chemical Engineering · doi:10.1016/j.compchemeng.2014.02.024 · Cited by 19
Facchinetti, A., Sparacino, G., & Cobelli, C. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700522 · Cited by 20
Self-citationMessori, M., Cobelli, C., & Magni, L. (2015) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2015.09.148 · Cited by 20
Self-citationA fundamental control performance limit for a class of positive nonlinear systems
Goodwin, G. C., Carrasco, D. S., Seron, M. M., & Medioli, A. M. (2018) · Automatica · doi:10.1016/j.automatica.2018.05.008 · Cited by 18
Bally, L., Thabit, H., Tauschmann, M., Allen, J. M., Hartnell, S., Wilinska, M. E., Exall, J., Huegel, V., Sibayan, J., Borgman, S. K., et al. (2017) · BMJ Open · doi:10.1136/bmjopen-2017-016738 · Cited by 17
Artificial Pancreas: First Clinical Trials in Argentina
Sánchez‐Peña, R., Colmegna, P. H., Grosembacher, L., Breton, M. D., Battista, H. D., Garelli, F., Belloso, W. H., Campos-Náñez, E., Simonovich, V., Beruto, V., et al. (2017) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2017.08.1151 · Cited by 20
Self-citationCombining data augmentation, EDAs and grammatical evolution for blood glucose forecasting
Velasco, J. M., Garnica, Ó., Lanchares, J., Botella, M., & Hidalgo, J. I. (2018) · Memetic Computing · doi:10.1007/s12293-018-0265-6 · Cited by 20
Lee, J. J., Dassau, E., Zisser, H., Harvey, R. A., Jovanovic, L. G., & Doyle, F. J. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700127 · Cited by 18
Facchinetti, A., Favero, S. D., Sparacino, G., & Cobelli, C. (2011) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2011.6091226 · Cited by 16
Self-citationAn Artificial Pancreas for Automated Blood Glucose Control in Patients With Type 1 Diabetes
Schmidt, S., Boiroux, D., Ranjan, A. G., J�rgensen, J., Madsen, H., & Nørgaard, K. (2015) · Therapeutic Delivery · doi:10.4155/tde.15.12 · Cited by 18
Resalat, N., Youssef, J. E., Reddy, R., & Jacobs, P. G. (2017) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2017.08.2270 · Cited by 15
Faggionato, E., Schiavon, M., & Dalla Man, C. (2021) · Metabolites · doi:10.3390/metabo11040235 · Cited by 15
“Abbreviations The following abbreviations are used in this manuscript: BH Body height BICc Bayesian information criterion corrected for NLME models BMI Body mass index BQL Below the quantification limit BSA Body surface area BSV Between subject variability BW Body weight FDA Food and Drug Administration IWRES Individu…”
MethodInfluentialGoodwin, G. C., & Seron, M. M. (2019) · Control Engineering Practice · doi:10.1016/j.conengprac.2019.06.001 · Cited by 15
Closed-loop artificial pancreas: current studies and promise for the future
WEINZIMER, S. A. (2012) · Current Opinion in Endocrinology Diabetes and Obesity · doi:10.1097/med.0b013e3283514e6b · Cited by 16
Challenges and progress in the development of a closed-loop artificial pancreas
Bequette, B. W. (2012) · Proceedings of the ... American Control Conference/Proceedings of the American Control Conference · doi:10.1109/acc.2012.6315593 · Cited by 15
Closed-loop insulin delivery systems in children and adolescents with type 1 diabetes
Dovč, K., & Battelino, T. (2020) · Expert Opinion on Drug Delivery · doi:10.1080/17425247.2020.1713747 · Cited by 15
A Constrained Model Predictive Controller for an Artificial Pancreas
Messori, M., Fornasiero, E., Toffanin, C., Cobelli, C., & Magni, L. (2014) · IFAC Proceedings Volumes · doi:10.3182/20140824-6-za-1003.01880 · Cited by 15
Self-citationFeichtner, F., Mader, J. K., Schaller, R., Schaupp, L. A., Ellmerer, M., Korsatko, S., Kondepati, V. R., Heise, H. M., Wilinska, M. E., Hovorka, R., et al. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500412 · Cited by 13
Leelarathna, L., Dellweg, S., Mader, J. K., Barnard, K. D., Benesch, C., Ellmerer, M., Heinemann, L., Kojzar, H., Thabit, H., Wilinska, M. E., et al. (2014) · BMJ Open · doi:10.1136/bmjopen-2014-006075 · Cited by 14
Multiple-Input Subject-Specific Modeling of Plasma Glucose Concentration for Feedforward Control
Kotz, K., Çınar, A., Mei, Y., Roggendorf, A., Littlejohn, E., Quinn, L., & Rollins, D. K. (2014) · Industrial & Engineering Chemistry Research · doi:10.1021/ie404119b · Cited by 12
Synthetic biomedical data generation in support of In Silico Clinical Trials
Simalatsar, A. (2023) · Frontiers in Big Data · doi:10.3389/fdata.2023.1085571 · Cited by 12
“This simulator was approved by the FDA as a substitute for preclinical trials for insulin injection pumps driven by closedloop algorithms and was used in several closed-loop algorithm simulation studies (Lee et al., 2009; Kovatchev et al., 2010; Toffanin et al., 2013) that helped many developers to receive rapid FDA a…”
MethodHidalgo, J. I., Colmenar, J. M., Risco‐Martín, J. L., Maqueda, E., Botella, M., Rubio, J. A., Cuesta‐Infante, A., Garnica, Ó., & Lanchares, J. (2014) · Genetic and Evolutionary Computation Conference Companion (GECCO Companion) · doi:10.1145/2598394.2609856 · Cited by 15
“Other personalized control approaches: [12][6][13] are proposals following the clinical practice.”
BackgroundFavero, S. D., Pillonetto, G., Cobelli, C., & Nicolao, G. D. (2011) · IFAC Proceedings Volumes · doi:10.3182/20110828-6-it-1002.01929 · Cited by 12
“For further details on the experiment we refer the reader to (Kovatchev et al., 2010 Nov 1).”
MethodSelf-citationArtificial pancreas systems: experiences from concept to commercialisation
Rodríguez-Sarmiento, D. L., Vargas, F. M. L., & Maira, G. (2022) · Expert Review of Medical Devices · doi:10.1080/17434440.2022.2150546 · Cited by 11
“This system implements a control-torange strategy [78] along with a safety module to reduce the risk of hypoglycemia [79].”
MethodMonitoring and fault detection of continuous glucose sensor measurements
Turksoy, K., Quinn, L., Littlejohn, E., & Çınar, A. (2015) · Proceedings of the ... American Control Conference/Proceedings of the American Control Conference · doi:10.1109/acc.2015.7172133 · Cited by 11
Recent Advances in the Prevention of Hypoglycemia in Type 1 Diabetes
Realsen, J., & Chase, H. P. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0150 · Cited by 10
Constrained Versus Unconstrained Model Predictive Control for Artificial Pancreas
Toffanin, C., & Magni, L. (2023) · IEEE Transactions on Control Systems Technology · doi:10.1109/tcst.2023.3291564 · Cited by 11
Self-citationTowards an artificial pancreas at home
Renard, É., Farret, A., Place, J., Wojtusciszyn, A., & Bringer, J. (2011) · Diabetes & Metabolism · doi:10.1016/s1262-3636(11)70973-9 · Cited by 10
Self-citationAng, K. H., & Sherr, J. L. (2017) · Expert Opinion on Drug Delivery · doi:10.1080/17425247.2017.1360862 · Cited by 11
Enhanced Wearable Medical Systems for Effective Blood Glucose Control
Gao, J., Yi, P., Chi, Z., & Zhu, T. (2016) · IEEE/ACM Conference on Connected Health: Applications, Systems and Engineering Technologies · doi:10.1109/chase.2016.11 · Cited by 9
“Unlike open-loop control mode, feedback results are added to this algorithm to guarantee accurate operation [7].”
Deployment of modular MPC for type 1 diabetes control: the Italian experience 2008–2016
Favero, S. D., Toffanin, C., Magni, L., & Cobelli, C. (2019) · Elsevier eBooks · doi:10.1016/b978-0-12-815655-1.00016-8 · Cited by 11
Self-citationGlucose forecasting using genetic programming and latent glucose variability features
Contador, S., Velasco, J. M., Garnica, Ó., & Hidalgo, J. I. (2021) · Applied Soft Computing · doi:10.1016/j.asoc.2021.107609 · Cited by 9
A novel model-based estimator for real-time prediction of insulin-on-board
Aiello, E. M., Wolkowicz, K. L., Pinsker, J. E., Dassau, E., & Doyle, F. J. (2022) · Chemical Engineering Science · doi:10.1016/j.ces.2022.118321 · Cited by 9
The impact of insulin pharmacokinetics and pharmacodynamics on the closed-loop artificial pancreas
Lee, J. J., Dassau, E., Zisser, H., Tamborlane, W. V., WEINZIMER, S. A., & Doyle, F. J. (2013) · IEEE Conference on Decision and Control (CDC) · doi:10.1109/cdc.2013.6759870 · Cited by 11
“, peak at 50-60 min after injection and clearance time of 6-8 h) result in a classical actuation delay that needs to be addressed as a part of the controller design [6, 10-17].”
BackgroundMethodA smart medical system for dynamic closed-loop blood glucose-insulin control
Gao, J., Yi, P., Chi, Z., & Zhu, T. (2017) · Smart Health · doi:10.1016/j.smhl.2017.04.001 · Cited by 9
“Unlike open-loop control mode, feedback results are added to this algorithm to guarantee accurate operation (Kovatchev et al., 2010).”
BackgroundVelasco, J. M., Garnica, Ó., Contador, S., Botella, M., Lanchares, J., & Hidalgo, J. I. (2017) · Genetic and Evolutionary Computation Conference Companion (GECCO Companion) · doi:10.1145/3067695.3082493 · Cited by 9
“Other personalized control approaches were presented by the main research groups on AP [4, 9, 12, 15].”
MethodA Critique of Observers Used in the Context of Feedback Control
Goodwin, G. C. (2018) · Lecture notes in computer science · doi:10.1007/978-3-319-97586-3_1 · Cited by 11
Sharma, A., Singh, H. P., & Nilam (2022) · International Journal of Biomathematics · doi:10.1142/s1793524522500516 · Cited by 8
Evaluating the Performance of a Novel Embedded Closed-loop System
Leelarathna, L., Thabit, H., Allen, J. M., Nodale, M., Wilinska, M. E., Powell, K., Lane, S., Evans, M. L., & Hovorka, R. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296813519882 · Cited by 10
Physiological models for artificial pancreas development
VISENTIN, R., Schiavon, M., Basu, R., Basu, A., Man, C. D., & Cobelli, C. (2019) · Elsevier eBooks · doi:10.1016/b978-0-12-815655-1.00015-6 · Cited by 11
Self-citationIn silico Models of Alcohol Dependence and Treatment
Kovatchev, B. P., Breton, M. D., & Johnson, B. A. (2012) · Frontiers in Psychiatry · doi:10.3389/fpsyt.2012.00004 · Cited by 7
Self-citationProfiled glucose forecasting using genetic programming and clustering
Contador, S., Velasco, J. M., Garnica, Ó., & Hidalgo, J. I. (2020) · ACM Symposium on Applied Computing · doi:10.1145/3341105.3374003 · Cited by 7
“Leading research groups on artificial pancreas have presented other personalized control approaches [10, 14, 19, 22] that follow the clinical practice.”
BackgroundRecent Trends in Blood Glucose Control Studies
Furutani, E. (2015) · Automatic Control of Physiological State and Function · doi:10.4172/2090-5092.1000106 · Cited by 7
“[30] applied a system (Figure 2a) to T1D patients that adjusts insulin infusion rate using MPC based on blood glucose measured by CGM, and showed that it can reduce the frequency of hypoglycemia and maintain blood glucose levels longer within a desirable range during the night.”
MethodCan clustering improve glucose forecasting with genetic programming models?
Contador, S., Hidalgo, J. I., Garnica, Ó., Velasco, J. M., & Lanchares, J. (2019) · Genetic and Evolutionary Computation Conference Companion (GECCO Companion) · doi:10.1145/3319619.3326809 · Cited by 7
“Leading research groups on artificial pancreas have presented other personalized control approaches [9, 12, 17, 20] that follow the clinical practice.”
BackgroundClosed-loop insulin delivery in adults with type 1 diabetes
Kumareswaran, K. (2012) · Apollo (University of Cambridge) · doi:10.17863/cam.16644 · Cited by 7
“Adults (n = 20) 15 hours fully closed-loop versus open loop (control) [105]”
MethodInfluentialModified control algorithms for patients with type 1 diabetes mellitus undergoing exercise
Liu, S.-W., Huang, H., Lin, C.-H., & Chien, I. (2014) · Journal of the Taiwan Institute of Chemical Engineers · doi:10.1016/j.jtice.2014.06.005 · Cited by 6
Model Predictive Control of Type 1 Diabetes
Magni, L. (2012) · IFAC Proceedings Volumes · doi:10.3182/20120823-5-nl-3013.00071 · Cited by 7
Self-citationWho needs an artificial pancreas? (怎样的患者需要一个人工胰腺?)
Winikoff, J., & Drexler, A. (2013) · Journal of Diabetes · doi:10.1111/1753-0407.12052 · Cited by 6
“mathematical modeling and has been addressed by two main algorithms.(2,3) The initial approach, known as proportional integral derivative control, based the insulin to be delivered on the glucose level and the rate of change of blood glucose at any given time.”
MethodA Portable Low-Power Platform for Ambulatory Closed Loop Control of Blood Glucose in Type 1 Diabetes
El-Sharkawy, M., Daniels, J., Pesl, P., Reddy, M., Oliver, N. S., Herrero, P., & Georgiou, P. (2018) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2018.8351695 · Cited by 6
“Another widely used technique is Model Predictive Control (MPC) which is the most popular control strategy in artificial pancreas systems [16]–[23].”
MethodModel Predictive Control of Type 1 Diabetes added to Conventional Therapy
Magni, L., Toffanin, C., Man, C. D., Kovatchev, B. P., Cobelli, C., & Nicolao, G. D. (2011) · IFAC Proceedings Volumes · doi:10.3182/20110828-6-it-1002.03027 · Cited by 5
“This paper builds on the experience obtained with an MPC strategy presented in Magni et al. (2007) and applied in a clinical study at the University of Virginia, Padova and Montpellier covering night and breakfast on 20 patients (Kovatchev et al. (2010)).”
BackgroundMethodSelf-citationEvaluation of Stochastic Adjustment for Glucose Sensor Bias During Closed-Loop Insulin Delivery
Kollman, C., Calhoun, P. M., Lum, J. W., Sauer, W., & Beck, R. W. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0133 · Cited by 5
Rahmanian, F., Asemani, M. H., Dehghani, M., & Mobayen, S. (2021) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2021.103088 · Cited by 5
Technology and Type 1 Diabetes: Closed-Loop Therapies
Ly, T. T., & Buckingham, B. A. (2015) · Current Pediatrics Reports · doi:10.1007/s40124-015-0083-y · Cited by 5
Closing the Loop: Another Step Forward
Weinstock, R. S. (2011) · Diabetes Care · doi:10.2337/dc11-1217 · Cited by 5
“Food and Drug Administration (FDA) of computer-based simulation studies (in silico testing) (instead of animal trials) has helped to more rapidly advance control algorithm development over the past several years (4,5).”
BackgroundAn LMI-Based Robust Fuzzy Controller for Blood Glucose Regulation in Type 1 Diabetes
Ganji, M., & Pourgholi, M. (2024) · doi:10.1109/icee63041.2024.10668193 · Cited by 4
Use of intraperitoneal insulin delivery for artificial pancreas
Renard, É., Farret, A., & Place, J. (2019) · Elsevier eBooks · doi:10.1016/b978-0-12-815655-1.00014-4 · Cited by 5
Self-citationMealtime correction insulin advisor for CGM-informed insulin pen therapy
Vereshchetin, P., Breton, M. D., & Patek, S. D. (2013) · Proceedings of the ... American Control Conference/Proceedings of the American Control Conference · doi:10.1109/acc.2013.6580277 · Cited by 4
Self-citationModel-Based Sensor-Augmented Pump Therapy
Grosman, B., Voskanyan, G. R., Loutseiko, M., Roy, A., Mehta, A., Kurtz, N., Parikh, N., KAUFMAN, F. R., Mastrototaro, J. J., & Keenan, D. B. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700224 · Cited by 5
An Android Phone Based Artificial Pancreas System for Type- 1 Diabetic Patients
Chandrasekhar, A., Saini, D., & Padhi, R. (2021) · IEEE Region Humanitarian Technology Conference · doi:10.1109/r10-htc53172.2021.9641715 · Cited by 4
Overcoming control challenges in the Artificial Pancreas
Hachimi, M. E., Ballouk, A., & Lebbar, H. (2016) · International Conference on Intelligent Systems: Theories and Applications · doi:10.1109/sita.2016.7772321 · Cited by 3
“C. Population results We consider Control-Variability Grid Analysis (CVGA) [22] of the entire cohort in the University of Padova/Virginia metabolic simulator, consisting of 10-subject and 100-subject simulator cohorts.”
BackgroundInfluentialComputational tools in the assistance of personalized healthcare
Papathanasiou, M. M., Onel, M., Naşcu, I., & Pistikopoulos, E. N. (2018) · Computer-aided chemical engineering/Computer aided chemical engineering · doi:10.1016/b978-0-444-63964-6.00006-4 · Cited by 4
Probabilistic Fitting of Glucose Models with Real-Coded Genetic Algorithms
Cervigón, C., Velasco, J. M., Burgos, C., Villanueva, R. J., & Hidalgo, J. I. (2021) · IEEE Congress on Evolutionary Computation (CEC) · doi:10.1109/cec45853.2021.9504836 · Cited by 3
Blood Glucose Prediction Using a Two Phase TSK Fuzzy Rule Based System
Alvarado, J., Velasco, J. M., Chávez, F., Hidalgo, J. I., & de Vega, F. F. (2021) · IEEE Congress on Evolutionary Computation (CEC) · doi:10.1109/cec45853.2021.9504992 · Cited by 3
Davis, G. M., Forlenza, G. P., Gehr, B., Pemberton, J. S., Penfornis, A., Do, D., Huyett, L. M., & Ly, T. T. (2026) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2026.113450 · Cited by 2
Glycated hemoglobin, serum proteins, and other markers as tools for monitoring
Gerber, S. M., Stickle, D. F., Ahmed, I., & Jabbour, S. A. (2015) · Wiley eBooks · doi:10.1002/9781118387658.ch59 · Cited by 5
Closed‐loop blood glucose control for type 1 diabetes
Furutani, E. (2019) · Electronics and Communications in Japan · doi:10.1002/ecj.12179 · Cited by 2
Artificial Pancreas: A Review of Fundamentals and Inpatient and Outpatient Studies
Favero, S. D., Bruttomesso, D., & Cobelli, C. (2014) · Frontiers in diabetes · doi:10.1159/000363512 · Cited by 3
Self-citationSzalay, P., Szilágyi, L. S., Benyó, Z., & Kovács, L. (2014) · Lecture notes in computer science · doi:10.1007/978-3-319-12640-1_54 · Cited by 2
“There are various methods to choose from: classical PID [1]; run-to-run control [2]; exact linearization based nonlinear control [3]; H∞ control [4,5]; Model Predictive Control (MPC) [6,7]; Linear Parameter Varying (LPV) control [8,9], among others.”
MethodUncertainties and Modeling Errors of Type 1 Diabetes Models
Kovács, L., & Szalay, P. (2015) · Lecture notes in bioengineering · doi:10.1007/978-3-319-25913-0_11 · Cited by 3
Pathophysiology and Management of Type 1 Diabetes
Georga, E. I., Fotiadis, D. I., & Tigas, S. K. (2017) · Elsevier eBooks · doi:10.1016/b978-0-12-804831-3.00002-9 · Cited by 4
Siket, M., Toth, R., Szász, L., Novák, K., Eigner, G., & Kovács, L. (2023) · IEEE World Symposium on Applied Machine Intelligence and Informatics (SAMI) · doi:10.1109/sami58000.2023.10044485 · Cited by 3
Automated optimal glycaemic control using a physiology based pharmacokinetic, pharmacodynamic model
Schaller, S., Schuppert, A., & Mitsos, A. (2015) · RWTH Publications (RWTH Aachen) · Cited by 6
Future Drug Treatments for Type 1 Diabetes
Sherr, J. L., Cengiz, E., Name, M. A. V., WEINZIMER, S. A., & Tamborlane, W. V. (2016) · Wiley eBooks · doi:10.1002/9781118924853.ch66 · Cited by 1
Technology for Hypoglycaemia: CSII and CGM
Schopman, J. E., & DeVries, J. H. (2013) · Wiley eBooks · doi:10.1002/9781118695432.ch9 · Cited by 1
Embedded implementation of modular closed-loop control of diabetes and in silico validation
Gentili, M., Caltabiano, D., Sannino, R., Toffanin, C., Palma, F. D., Magni, L., & Lane, S. (2013) · IEEE International Conference on e-Health Networking, Applications and Services (Healthcom) · doi:10.1109/healthcom.2013.6720730 · Cited by 2
Self-citationInsulinothérapie en boucle fermée : vers le pancréas artificiel ambulatoire
Renard, É., Place, J., Pelletier, M.-É., & Farret, A. (2012) · Médecine des Maladies Métaboliques · doi:10.1016/s1957-2557(12)70468-6 · Cited by 1
Self-citationDevelopment of physiological metabolic models in diabetes based on data mining techniques
Georga, E. I. (2019) · University of Ioannina · doi:10.26268/heal.uoi.1420 · Cited by 1
“all studies consistently concluding that artificial pancreas approaches can reduce nocturnal hypoglycaemia and improve overall nocturnal glycaemic control both in an inpatient and an outpatient setting [56, 61-66].”
BackgroundL'insulinothérapie au xxie siècle : une symphonie de plus en plus harmonieuse
Benhamou, P., Hanaire, H. P., & Renard, É. (2021) · Médecine des Maladies Métaboliques · doi:10.1016/s1957-2557(21)00179-6 · Cited by 1
Fuzzy Model Predictive Controller for Artificial Pancreas
Hachimi, M. E., Ballouk, A., & Baghdad, A. (2018) · International Journal of Power Electronics and Drive Systems (IJPEDS) · doi:10.11591/ijpeds.v9.i3.pp1178-1185 · Cited by 1
“The simulator was approved by the U.S. Food and Drug Administration (FDA) [9] for verification of control algorithms for AP before doing clinical trials on T1DM subjects, this trials are equivalent to animal test.”
MethodModeling and Control of Physiological Systems
Kovács, L. (2015) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-981-287-573-0_2 · Cited by 1
Automatic Control of Physiological State and Function
Furutani, E. (2017) · Automatic Control of Physiological State and Function · doi:10.4172/2090-5092 · Cited by 1
Canciani, G. N., Stewart, Z. A., & Murphy, H. R. (2017) · Wiley eBooks · doi:10.1002/9781119043805.ch16 · Cited by 1
Medical Cyber-Physical Systems: Challenges and Future Directions
Lee, I., Sokolsky, O. V., & Weimer, J. E. (2021) · Springer eBooks · doi:10.1007/978-3-030-44184-5_100115 · Cited by 1
Robust strategies for glucose control in type 1 diabetes
Tomás, A. R. (2015) · RiuNet (Universitat Politècnica de València) · doi:10.4995/thesis/10251/56001
Formal verification of Artificial Pancreas
Kumar, S. A. (2016) · CU Scholar (University of Colorado Boulder)
Part A: Type 1 Diabetes Mellitus
Zavitsanou, S., Mantalaris, A., Georgiadis, M. C., & Pistikopoulos, E. N. (2017) · Wiley eBooks · doi:10.1002/9781118965580.ch5
Risks reduction control techniques analysis in type-1 diabetes
Arif, H., Shuja, S., Imtiaz, S., Khaqan, A., Hasan, Q. U., Malik, S. A., Ahmed, J., & Riaz, R. A. (2018) · Biomedical Research - India · doi:10.4066/biomedicalresearch.29-17-3740
“Strategies like Model Predictive Control (MPC), Proportional Integral Derivative (PID) and adaptive control have been broadly implemented both in silico and in clinical trials [11-17].”
MethodRapid Model Predictive Controller for Artificial Pancreas
Hachimi, M., Ballouk, A., & Baghdad, A. (2019) · Advances in Science, Technology & Innovation/Advances in science, technology & innovation · doi:10.1007/978-3-030-05276-8_8
Renard, É. (2019) · Springer eBooks · doi:10.1007/978-3-319-98119-2_6
Closed-loop Blood Glucose Control for Type 1 Diabetes
Furutani, E. (2019) · IEEJ Transactions on Electronics Information and Systems · doi:10.1541/ieejeiss.139.260
The dawn of automated insulin delivery: from promise to product
Nally, L. M., & Sherr, J. L. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816714-4.00016-8
Diagnosis and Management of Pediatric Diseases
Sergi, C. M. (2019) · MDPI eBooks · doi:10.3390/books978-3-03921-967-4
Medical Cyber-Physical Systems: Challenges and Future Directions
Lee, I., Sokolsky, O. V., & Weimer, J. E. (2020) · Springer eBooks · doi:10.1007/978-1-4471-5102-9_100115-1
Session Three: Innovating Beyond Glucose Control in Diabetes Care
Michel, L. (2017) · EMJ Diabetes · doi:10.33590/emjdiabet/10314488
Tube-based Model Predictive Control of Blood Glucose for Multiple Meal Scenarios
Iimura, Y., & Wakaiki, M. (2023) · Transactions of the Society of Instrument and Control Engineers · doi:10.9746/sicetr.59.529 · Cited by 1
Learning of Input Penalty Function for Zone Model Predictive Blood Glucose Control
Iimura, Y., & Wakaiki, M. (2024) · Transactions of the Society of Instrument and Control Engineers · doi:10.9746/sicetr.60.464
Metadata
- Authors
- Kovatchev, B., Cobelli, C., Renard, E., Anderson, S., Breton, M., Patek, S., Clarke, W., Bruttomesso, D., Maran, A., Costa, S., Avogaro, A., Man, C. D., Facchinetti, A., Magni, L., De Nicolao, G., Place, J., & Farret, A.
- Affiliations
- University of Virginia (US)
- University of Padua (IT)
- Department of Endocrinology and UMR CNRS, CHU and University of Montpellier, Montpellier, France
- University of Pavia (IT)
- Department of System and Informatics, University of Pavia, Pavia, Italy;
- Centre National de la Recherche Scientifique (FR)
- Corresponding author
- Kovatchev, Boris
- ORCID
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2010-11
- Volume
- 4
- Issue
- 6
- Pages
- 1374-1381
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC3005047
- OpenAlex
- W2134017961
- Semantic Scholar
- faf6284087dadbb4011a4ee2564dc9ed8bad111a
- Cited by
- 200
- References
- 25
- Citations by year
- 2026: 2 · 2024: 2 · 2023: 6 · 2022: 5 · 2021: 9 · 2020: 4 · 2019: 15 · 2018: 11 · 2017: 14 · 2016: 13 · 2015: 19 · 2014: 29 · 2013: 32 · 2012: 24
- FWCI
- 17.19
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Padua Research Archive (University of Padova) (submittedVersion)
- Funders
- National Institutes of Health (5r01dk051562-23)
- National Institute of Diabetes and Digestive and Kidney Diseases (R01DK 51562)
- National Center for Research Resources (M01RR000847, M01-RR 00847)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Keywords
- type 1 diabetes, feedback control, model predictive control, artificial pancreas, in silico design, computer simulation, nocturnal hypoglycemia, continuous glucose monitoring, insulin pump
- MeSH terms
- Adolescent, Adult, Aged, Algorithms, Blood Glucose, Child, Computer Simulation, Diabetes Mellitus, Type 1, Female, France, Glycated Hemoglobin, Humans, Hypoglycemia, Hypoglycemic Agents, Insulin, Insulin Infusion Systems, Italy, Male, Middle Aged, Models, Biological, Monitoring, Physiologic, Pilot Projects, Time Factors, Virginia, Treatment Outcome, Pancreas, Artificial, Drug Dosage Calculations, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.