Multinational Study of Subcutaneous Model-Predictive Closed-Loop Control in Type 1 Diabetes Mellitus: Summary of the Results
Open accessThe DOI matches its Crossref record; 25 of the 25 references it cites were found in a registry or an index.
Abstract
Background: In 2008–2009, the first multinational study was completed comparing closed-loop control (artificial pancreas) to state-of-the-art open-loop therapy in adults with type 1 diabetes mellitus (T1DM). Methods: The design of the control algorithm was done entirely in silico, i.e., using computer simulation experiments with N = 300 synthetic “subjects” with T1DM instead of traditional animal trials. The clinical experiments recruited 20 adults with T1DM at the Universities of Virginia (11); Padova, Italy (6); and Montpellier, France (3). Open-loop and closed-loop admission was scheduled 3–4 weeks apart, continued for 22 h (14.5 h of which were in closed loop), and used a continuous glucose monitor and an insulin pump. The only difference between the two sessions was that insulin dosing was performed by the patient under a physician's supervision during open loop, whereas insulin dosing was performed by a control algorithm during closed loop. Results: In silico design resulted in rapid (less than 6 months compared to years of animal trials) and cost-effective system development, testing, and regulatory approvals in the United States, Italy, and France. In the clinic, compared to open-loop, closed-loop control reduced nocturnal hypoglycemia (blood glucose below 3.9 mmol/liter) from 23 to 5 episodes (p < .01) and increased the amount of time spent overnight within the target range (3.9 to 7.8 mmol/liter) from 64% to 78% (p = .03). Conclusions: In silico experiments can be used as viable alternatives to animal trials for the preclinical testing of insulin treatment strategies. Compared to open-loop treatment under identical conditions, closed-loop control improves the overnight regulation of diabetes.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Only three months later, in April 2008, the first human trials began at the Universities of Virginia, Montpellier (France), and Padova (Italy), using a control system designed entirely in silico.69”
From Developing the UVA/Padova Type 1 Diabetes Simulator: Modeling, Validation, Refinements, and Utility (2023) · cites this paper as [69]
Citations
Cites
Continuous glucose monitoring and intensive treatment of type 1 diabetes
(2008) · New England Journal of Medicine · doi:10.1056/nejmoa0805017 · Cited by 1,559
Nonlinear model predictive control of glucose concentration in subjects with type 1 diabetes
Hovorka, R., Canonico, V., Chassin, L. J., Haueter, U., Massi-Benedetti, M., Federici, M. O., Pieber, T. R., Schaller, H. C., Schaupp, L., Vering, T., et al. (2004) · Physiological Measurement · doi:10.1088/0967-3334/25/4/010 · Cited by 1,274
Meal Simulation Model of the Glucose-Insulin System
Dalla Man, C., Rizza, R. A., & Cobelli, C. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.893506 · Cited by 955
Self-citationIn Silico Preclinical Trials: A Proof of Concept in Closed-Loop Control of Type 1 Diabetes
Kovatchev, B. P., Breton, M., Dalla Man, C., & Cobelli, C. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300106 · Cited by 760
Self-citationWeinzimer, S. A., Steil, G. M., Swan, K. L., Dziura, J., Kurtz, N., & Tamborlane, W. V. (2008) · Diabetes Care · doi:10.2337/dc07-1967 · Cited by 567
A model-based algorithm for blood glucose control in Type I diabetic patients
Parker, R. S., Doyle, F. J., & Peppas, N. A. (1999) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.740877 · Cited by 500 · cited 4×
“, by a series of computer simulations performed on virtual subjects; (2) clinical experiments performed in three independent centers in the United States, Italy, and France, which enhances the external validity of the results; and (3) automated (as opposed to manual) transfer of CGM data to the control system, which b…”
BackgroundInfluentialFeasibility of Automating Insulin Delivery for the Treatment of Type 1 Diabetes
Steil, G. M., Rebrin, K., Darwin, C., Hariri, F., & Saad, M. F. (2006) · Diabetes · doi:10.2337/db06-0419 · Cited by 473
Hovorka, R., Allen, J. M., Elleri, D., Chassin, L. J., Harris, J., Xing, D., Kollman, C., Hovorka, T., Larsen, A. M. F., Nodale, M., et al. (2010) · The Lancet · doi:10.1016/s0140-6736(09)61998-x · Cited by 450
Continuous glucose monitoring and closed‐loop systems
Hovorka, R. (2006) · Diabetic Medicine · doi:10.1111/j.1464-5491.2005.01672.x · Cited by 443 · cited 4×
“The clinical experiments recruited 20 adults with T1DM at the Universities of Virginia (11); Padova, Italy (6); and Montpellier, France (3).”
BackgroundInfluentialA Bihormonal Closed-Loop Artificial Pancreas for Type 1 Diabetes
El-Khatib, F. H., Russell, S. J., Nathan, D. M., Sutherlin, R. G., & Damiano, E. R. (2010) · Science Translational Medicine · doi:10.1126/scitranslmed.3000619 · Cited by 403
Cited by
Artificial Pancreas: Past, Present, Future
Cobelli, C., Renard, E., & Kovatchev, B. (2011) · Diabetes · doi:10.2337/db11-0654 · Cited by 600
“(31) reported a study in which the aggressiveness of the controller was individualized using patient parameters such as body weight, carbohydrate ratio, and insulin basal rate.”
BackgroundSelf-citationNocturnal Glucose Control with an Artificial Pancreas at a Diabetes Camp
Phillip, M., Battelino, T., Atlas, E., Kordonouri, O., Bratina, N., Miller, S., Biester, T., Avbelj Stefanija, M., Muller, I., Nimri, R., et al. (2013) · New England Journal of Medicine · doi:10.1056/nejmoa1206881 · Cited by 435
Closed-Loop Artificial Pancreas Systems: Engineering the Algorithms
Doyle, F. J., Huyett, L. M., Lee, J. B., Zisser, H. C., & Dassau, E. (2014) · Diabetes Care · doi:10.2337/dc13-2108 · Cited by 394
“Fully automated closed-loop control showed a comparatively lower time in range, although these systems require considerably less patient effort (9,17,24,26,27,31,34,35,41,42).”
BackgroundClosed-loop insulin delivery from bench to clinical practice
Hovorka, R. (2011) · Nature Reviews Endocrinology · doi:10.1038/nrendo.2011.32 · Cited by 342
In silico clinical trials: how computer simulation will transform the biomedical industry
Viceconti, M., Henney, A., & Morley-Fletcher, E. (2016) · International Journal of Clinical Trials · doi:10.18203/2349-3259.ijct20161408 · Cited by 289
Fully Integrated Artificial Pancreas in Type 1 Diabetes
Breton, M., Farret, A., Bruttomesso, D., Anderson, S., Magni, L., Patek, S., Dalla Man, C., Place, J., Demartini, S., Del Favero, S., et al. (2012) · Diabetes · doi:10.2337/db11-1445 · Cited by 315
Self-citationHovorka, R., Kumareswaran, K., Harris, J., Allen, J. M., Elleri, D., Xing, D., Kollman, C., Nodale, M., Murphy, H. R., Dunger, D. B., et al. (2011) · BMJ · doi:10.1136/bmj.d1855 · Cited by 244
Hovorka, R., Elleri, D., Thabit, H., Allen, J. M., Leelarathna, L., El-Khairi, R., Kumareswaran, K., Caldwell, K., Calhoun, P., Kollman, C., et al. (2014) · Diabetes Care · doi:10.2337/dc13-2644 · Cited by 214
ISPAD Clinical Practice Consensus Guidelines 2018: diabetes technologies
Sherr, J. L., Tauschmann, M., Battelino, T., de Bock, M., Forlenza, G., Roman, R., Hood, K. K., & Maahs, D. M. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12731 · Cited by 197
Challenges and recent progress in the development of a closed-loop artificial pancreas
Bequette, B. W. (2012) · Annual Reviews in Control · doi:10.1016/j.arcontrol.2012.09.007 · Cited by 183
Modular Closed-Loop Control of Diabetes
Patek, S. D., Magni, L., Dassau, E., Hughes-Karvetski, C., Toffanin, C., De Nicolao, G., Del Favero, S., Breton, M., Man, C. D., Renard, E., et al. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2192930 · Cited by 180
“The drawbacks highlighted in [29] can be circumvented by a control scheme that combines feedforward (FF) action, in-”
BackgroundMethodSelf-citationFeasibility of Outpatient Fully Integrated Closed-Loop Control
Kovatchev, B. P., Renard, E., Cobelli, C., Zisser, H. C., Keith-Hynes, P., Anderson, S. M., Brown, S. A., Chernavvsky, D. R., Breton, M. D., Farret, A., et al. (2013) · Diabetes Care · doi:10.2337/dc12-1965 · Cited by 177
Self-citationTurksoy, K., Bayrak, E. S., Quinn, L., Littlejohn, E., & Cinar, A. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0283 · Cited by 161
“In contrast to MPC studies, 12,13,15 there is no need to provide meal and exercise information as known disturbances.”
Haidar, A., Legault, L., Dallaire, M., Alkhateeb, A., Coriati, A., Messier, V., Cheng, P., Millette, M., Boulet, B., & Rabasa-Lhoret, R. (2013) · Canadian Medical Association Journal · doi:10.1503/cmaj.121265 · Cited by 167
“Instead of a single glucose sensor, infusions were based on venous glucose(14) or multiple sensors,(13) and were sometimes overridden by an attending physician.(13,17) Moreover, the glucose sensor was calibrated using referencequality venous plasma glucose(10,17,19) instead of capillary finger-stick measurements, pote…”
BackgroundControl-Relevant Models for Glucose Control Using A Priori Patient Characteristics
van Heusden, K., Dassau, E., Zisser, H. C., Seborg, D. E., & Doyle, F. J. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2176939 · Cited by 160
Closed-Loop Insulin Delivery Systems: Past, Present, and Future Directions
Templer, S. (2022) · Frontiers in Endocrinology · doi:10.3389/fendo.2022.919942 · Cited by 182
Weinzimer, S. A., Sherr, J. L., Cengiz, E., Kim, G., Ruiz, J. L., Carria, L., Voskanyan, G., Roy, A., & Tamborlane, W. V. (2012) · Diabetes Care · doi:10.2337/dc12-0330 · Cited by 134
Ly, T. T., Roy, A., Grosman, B., Shin, J., Campbell, A., Monirabbasi, S., Liang, B., von Eyben, R., Shanmugham, S., Clinton, P., et al. (2015) · Diabetes Care · doi:10.2337/dc14-3073 · Cited by 132
Leelarathna, L., Dellweg, S., Mader, J. K., Allen, J. M., Benesch, C., Doll, W., Ellmerer, M., Hartnell, S., Heinemann, L., Kojzar, H., et al. (2014) · Diabetes Care · doi:10.2337/dc13-2911 · Cited by 130
“These include evaluations using a randomized design by our group in youths (11,12), adults (13), and pregnant women (14) and by others using the model predictive control algorithm (15,16), the proportional-integral-derivative approach (17,18), and the fuzzy logic con-”
MethodSherr, J. L., Cengiz, E., Palerm, C. C., Clark, B., Kurtz, N., Roy, A., Carria, L., Cantwell, M., Tamborlane, W. V., & Weinzimer, S. A. (2013) · Diabetes Care · doi:10.2337/dc13-0010 · Cited by 125
Ly, T. T., Breton, M. D., Keith-Hynes, P., De Salvo, D., Clinton, P., Benassi, K., Mize, B., Chernavvsky, D., Place, J., Wilson, D. M., et al. (2014) · Diabetes Care · doi:10.2337/dc14-0147 · Cited by 121
“More glucose measurements were obtained during OCL compared with control nights, with a median of 3 (3, 4) versus”
Self-citationClinical Evaluation of a Personalized Artificial Pancreas
Dassau, E., Zisser, H., Harvey, R. A., Percival, M. W., Grosman, B., Bevier, W., Atlas, E., Miller, S., Nimri, R., Jovanovič, L., et al. (2013) · Diabetes Care · doi:10.2337/dc12-0948 · Cited by 120
“Either a manual glucose reading from the CGM, manual administration of insulin by the study personnel, or both was used (4,6,9,14,15), or a three-pump system with manual venous blood glucose input was implemented (6).”
MethodMPC based Artificial Pancreas: Strategies for individualization and meal compensation
Soru, P., Nicolao, G. D., Toffanin, C., Man, C. D., Cobelli, C., & Magni, L. (2012) · Annual Reviews in Control · doi:10.1016/j.arcontrol.2012.03.009 · Cited by 130
“…(Steil, Rebrin, Darwin, Hariri, & Saad, 2006), with an acceleration caused by the launch of research projects funded by the JDRF, the European Commission and the NIH (El-Khatib, Russell, Nathan, Sutherlin, & Damiano, 2010; Hovorka et al., 2004, 2010; Kovatchev et al., 2010; Weinzimer et al., 2008).”
BackgroundResultSelf-citationPilot Studies of Wearable Outpatient Artificial Pancreas in Type 1 Diabetes
Cobelli, C., Renard, E., Kovatchev, B. P., Keith-Hynes, P., Ben Brahim, N., Place, J., Del Favero, S., Breton, M., Farret, A., Bruttomesso, D., et al. (2012) · Diabetes Care · doi:10.2337/dc12-0660 · Cited by 120
Self-citationA Review of Emerging Technologies for the Management of Diabetes Mellitus
Zarkogianni, K., Litsa, E., Mitsis, K., Wu, P.-Y., Kaddi, C. D., Cheng, C.-W., Wang, M. D., & Nikita, K. S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2470521 · Cited by 112
“Overnight closed-loop experiments using different MPC controllers have demonstrated the superiority of the closed-loop control over the conventional pump treatment [101], [104].”
Algorithms for a Closed-Loop Artificial Pancreas: The Case for Model Predictive Control
Bequette, B. W. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700624 · Cited by 113
Messori, M., Incremona, G. P., Cobelli, C., & Magni, L. (2018) · IEEE Control Systems · doi:10.1109/mcs.2017.2766314 · Cited by 127
“Starting from 2008, several clinical trials have been performed by relying on a model predictive control (MPC) algorithm [16]–[20] in a hospital setting [21]–[26].”
MethodSelf-citationCircadian Variability of Insulin Sensitivity: Physiological Input for In Silico Artificial Pancreas
Visentin, R., Dalla Man, C., Kudva, Y. C., Basu, A., & Cobelli, C. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0192 · Cited by 113
Self-citationUse of a “Fuzzy Logic” Controller in a Closed-Loop Artificial Pancreas
Mauseth, R., Hirsch, I. B., Bollyky, J., Kircher, R., Matheson, D., Sanda, S., & Greenbaum, C. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0036 · Cited by 112
Nimri, R., Muller, I., Atlas, E., Miller, S., Kordonouri, O., Bratina, N., Tsioli, C., Stefanija, M. A., Danne, T., Battelino, T., et al. (2014) · Pediatric Diabetes · doi:10.1111/pedi.12071 · Cited by 103
Artificial Pancreas: Model Predictive Control Design from Clinical Experience
Toffanin, C., Messori, M., Di Palma, F., De Nicolao, G., Cobelli, C., & Magni, L. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700607 · Cited by 106
“J Diabetes Sci Technol 2013;7(6):1470–1483 SYMPOSIUM”
Self-citationVisentin, R., Dalla Man, C., Kovatchev, B., & Cobelli, C. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0377 · Cited by 103
“To do this, simulations have been compared with a database of 24 T1DM adult subjects studied in two 22-h admissions, in open- and closed-loop, respectively.(8) The comparison consisted, first, in determining if, for each T1DM subject, a virtual subject exists who shows a similar metabolic response and lies in the same…”
MethodSelf-citationClosed-loop insulin delivery for treatment of type 1 diabetes
Elleri, D., Dunger, D. B., & Hovorka, R. (2011) · BMC Medicine · doi:10.1186/1741-7015-9-120 · Cited by 115
“Meal announcement Laboratory studies Feasibility documented in children, adults, and pregnant women using various control algorithms; preferred option by most investigators [26,44,46,58,63]”
BackgroundSherr, J. L., Patel, N. S., Michaud, C. I., Palau-Collazo, M. M., Van Name, M. A., Tamborlane, W. V., Cengiz, E., Carria, L. R., Tichy, E. M., & Weinzimer, S. A. (2016) · Diabetes Care · doi:10.2337/dc16-0089 · Cited by 99
Effect of Insulin Feedback on Closed-Loop Glucose Control: A Crossover Study
Ruiz, J. L., Sherr, J. L., Cengiz, E., Carria, L., Roy, A., Voskanyan, G., Tamborlane, W. V., & Weinzimer, S. A. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600517 · Cited by 97
Day and Night Closed-Loop Control in Adults With Type 1 Diabetes
Luijf, Y. M., DeVries, J. H., Zwinderman, K., Leelarathna, L., Nodale, M., Caldwell, K., Kumareswaran, K., Elleri, D., Allen, J. M., Wilinska, M. E., et al. (2013) · Diabetes Care · doi:10.2337/dc12-1956 · Cited by 95
Self-citationSusceptibility of Interstitial Continuous Glucose Monitor Performance to Sleeping Position
Mensh, B. D., Wisniewski, N. A., Neil, B. M., & Burnett, D. R. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700408 · Cited by 97
The Artificial Pancreas in 2016: A Digital Treatment Ecosystem for Diabetes
Kovatchev, B., Tamborlane, W. V., Cefalu, W. T., & Cobelli, C. (2016) · Diabetes Care · doi:10.2337/dc16-0824 · Cited by 95
Self-citationSharifi, A., De Bock, M. I., Jayawardene, D., Loh, M. M., Horsburgh, J. C., Berthold, C. L., Paramalingam, N., Bach, L. A., Colman, P. G., Davis, E. A., et al. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0288 · Cited by 91
“…at 02:00 h revealed median [IQR] for AUC > 144 mg/dL during Android-HCLS to be higher for all participants on night 4 than on nights 1–3 (73.8 [27.6, 148.0] min $ mg/dL vs. 6.3 [0, 49.0] min $ mg/dL; p = 0.004), for adults (48.1 [20.1, 174.1] min $ mg/dL vs. 6.3 [0, 80.1] min $ mg/dL; p =…”
InfluentialDel Favero, S., Place, J., Kropff, J., Messori, M., Keith‐Hynes, P., Visentin, R., Monaro, M., Galasso, S., Boscari, F., Toffanin, C., et al. (2015) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.12440 · Cited by 90
Self-citationThe Hybrid Closed-Loop System: Evolution and Practical Applications
Weaver, K. W., & Hirsch, I. B. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0091 · Cited by 84
DiAs User Interface: A Patient-Centric Interface for Mobile Artificial Pancreas Systems
Keith-Hynes, P., Guerlain, S., Mize, B., Hughes-Karvetski, C., Khan, M., McElwee-Malloy, M., & Kovatchev, B. P. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700602 · Cited by 85
Self-citationJacobs, P. G., El Youssef, J., Castle, J., Bakhtiani, P., Branigan, D., Breen, M., Bauer, D., Preiser, N., Leonard, G., Stonex, T., et al. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2323248 · Cited by 82
Switched LPV Glucose Control in Type 1 Diabetes
Colmegna, P. H., Sánchez‐Peña, R., Gondhalekar, R., Dassau, E., & Doyle, F. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2487043 · Cited by 84
van Bon, A. C., Luijf, Y. M., Koebrugge, R., Koops, R., Hoekstra, J. B. L., & DeVries, J. H. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0166 · Cited by 76
Closed-Loop Insulin Delivery in Type 1 Diabetes
Thabit, H., & Hovorka, R. (2012) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2011.12.003 · Cited by 74
The Diabetes Assistant: A Smartphone-Based System for Real-Time Control of Blood Glucose
Keith-Hynes, P., Mize, B., Robert, A., & Place, J. (2014) · Electronics · doi:10.3390/electronics3040609 · Cited by 77
Self-citationReport from IPITA-TTS Opinion Leaders Meeting on the Future of β-Cell Replacement
Bartlett, S. T., Markmann, J. F., Johnson, P., Korsgren, O., Hering, B. J., Scharp, D., Kay, T. W. H., Bromberg, J., Odorico, J. S., Weir, G. C., et al. (2016) · Transplantation · doi:10.1097/tp.0000000000001055 · Cited by 75
Self-citationNimri, R., Atlas, E., Ajzensztejn, M., Miller, S., Oron, T., & Phillip, M. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0004 · Cited by 71
Current Diabetes Technology: Striving for the Artificial Pancreas
Allen, N., & Gupta, A. (2019) · Diagnostics · doi:10.3390/diagnostics9010031 · Cited by 65
“In a 2010 study, OmniPod and Freestyle Navigator CGM were used to compare an open- versus closed-loop system showing an improvement from 64% to 78% time within range for users [57].”
MethodFeasibility of Closed-Loop Insulin Delivery in Type 2 Diabetes: A Randomized Controlled Study
Kumareswaran, K., Thabit, H., Leelarathna, L., Caldwell, K., Elleri, D., Allen, J. M., Nodale, M., Wilinska, M. E., Evans, M. L., & Hovorka, R. (2014) · Diabetes Care · doi:10.2337/dc13-1030 · Cited by 66
“Such a system has been shown to be safe and efficacious in controlled overnight studies in adults with type 1 diabetes (4,5).”
BackgroundModel individualization for artificial pancreas
Messori, M., Toffanin, C., Favero, S. D., Nicolao, G. D., Cobelli, C., & Magni, L. (2016) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2016.06.006 · Cited by 75
Self-citationMultinight “Bedside” Closed-Loop Control for Patients with Type 1 Diabetes
Brown, S. A., Kovatchev, B. P., Breton, M. D., Anderson, S. M., Keith-Hynes, P., Patek, S. D., Jiang, B., Ben Brahim, N., Vereshchetin, P., Bruttomesso, D., et al. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0259 · Cited by 64
Self-citationAdaptive Control of Artificial Pancreas Systems ‐ A Review
Turksoy, K., & Çınar, A. (2014) · Journal of Healthcare Engineering · doi:10.1260/2040-2295.5.1.1 · Cited by 70
“Four different types of control strategies have been proposed for AP: Proportionalintegral-derivative (PID) control [2–7], model-based and model-predictive control (MPC) [8–12], fuzzy logic control [13,14] and adaptive control [15–18].”
BackgroundMOSQUERA-LOPEZ, C. M., Wilson, L. M., Youssef, J. E., Hilts, W. W., LEITSCHUH, J. A., Branigan, D. L., Gabo, V. B., Eom, J. H., Castle, J. R., & Jacobs, P. G. (2023) · npj Digital Medicine · doi:10.1038/s41746-023-00783-1 · Cited by 73
Nimri, R., Danne, T., Kordonouri, O., Atlas, E., Bratina, N., Biester, T., Avbelj, M., Miller, S., Muller, I., Phillip, M., et al. (2013) · Pediatric Diabetes · doi:10.1111/pedi.12025 · Cited by 62
Viceconti, M., Cobelli, C., Haddad, T., Himes, A., Kovatchev, B., & Palmer, M. (2017) · Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine · doi:10.1177/0954411917702931 · Cited by 64
“This IDE was issued solely on the basis of in silico testing of the safety and efficacy of AP control algorithm, an event that set a precedent for future clinical studies.(84) In brief, to test the validity of the computer simulation environment independently from the data used for its development, a number of experim…”
MethodSelf-citationFacchinetti, A., Favero, S., Sparacino, G., & Cobelli, C. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2227256 · Cited by 62
“Datasets of three real T1DM subjects were extracted from a larger database of open-loop and closed-loop experimental trials performed at the Department of Clinical and Experimental Medicine (University of Padova, Padova, Italy) [17].”
BackgroundSelf-citationHarvey, R. A., Dassau, E., Bevier, W. C., Seborg, D. E., Jovanovič, L., Doyle, F. J., & Zisser, H. C. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0231 · Cited by 62
Place, J., Robert, A., Ben Brahim, N., Keith-Hynes, P., Farret, A., Pelletier, M.-J., Buckingham, B., Breton, M., Kovatchev, B., & Renard, E. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700603 · Cited by 59
Self-citationCameron, F., Niemeyer, G., Wilson, D. M., Bequette, B. W., Benassi, K. S., Clinton, P., & Buckingham, B. A. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0093 · Cited by 57
Renukuntla, V. S., Ramchandani, N., Trast, J., Cantwell, M., & Heptulla, R. A. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814542153 · Cited by 56
Hovorka, R., Nodale, M., Haidar, A., & Wilinska, M. E. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0185 · Cited by 61
Elleri, D., Allen, J. M., Biagioni, M., Kumareswaran, K., Leelarathna, L., Caldwell, K., Nodale, M., Wilinska, M. E., Acerini, C. L., Dunger, D. B., et al. (2012) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2012.00903.x · Cited by 53
“Learning” Can Improve the Blood Glucose Control Performance for Type 1 Diabetes Mellitus
Wang, Y., Zhang, J., Zeng, F., Wang, N., Chen, X., Zhang, B., Zhao, D. B., Yang, W., & Cobelli, C. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0328 · Cited by 56
Self-citationA Glucose-Specific Metric to Assess Predictors and Identify Models
Favero, S. D., Facchinetti, A., & Cobelli, C. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2185234 · Cited by 51
Self-citationReducing Risks in Type 1 Diabetes Using ${\cal H}_\infty$ Control
Colmegna, P. H., Sánchez‐Peña, R., Gondhalekar, R., Dassau, E., & Doyle, F. J. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2336772 · Cited by 64
Ly, T. T., Keenan, D. B., Roy, A., Han, J., Grosman, B., Cantwell, M., Kurtz, N., von Eyben, R., Clinton, P., Wilson, D. M., et al. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0431 · Cited by 51
van Bon, A. C., Jonker, L., Koebrugge, R., Koops, R., Hoekstra, J. B. L., & DeVries, J. H. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600516 · Cited by 59
Model-based falsification of an artificial pancreas control system
Sankaranarayanan, S., Kumar, S. A., Cameron, F. M., Bequette, B. W., Fainekos, G., & Maahs, D. M. (2017) · ACM SIGBED Review · doi:10.1145/3076125.3076128 · Cited by 55
“Minimizer [4, 46, 29] #2 + additional insulin when glucose above threshold 4 Hybrid Closed Loop [33, 32, 31] Closed loop insulin delivery with manual bolus 5 Fully Automated Closed Loop [9, 10, 12, 18, 40, 36, 21] #4 with no manual boluses 6 Multi-hormone Closed Loop [24, 23] Use glucagon and insulin”
MethodAutomated insulin delivery systems: from early research to routine care of type 1 diabetes
Renard, É. (2022) · Acta Diabetologica · doi:10.1007/s00592-022-01929-5 · Cited by 58
“in adult T1D patients who were investigated for night-time control using another combination of SC glucose monitoring system, SC insulin pump and MPC algorithm [22].”
MethodZisser, H., Renard, E., Kovatchev, B., Cobelli, C., Avogaro, A., Nimri, R., Magni, L., Buckingham, B. A., Chase, H. P., Doyle, F. J., et al. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0066 · Cited by 48
“The control algorithm was of control-to-range class, 22 and its design was based on modular architecture 1,8,11,23 including two interacting modules: the Range Correction Module, 24 developed at the University of Pavia, and the Safety Super-vision Module, 25 developed at the University of Virginia.”
Self-citationZecchin, C., Facchinetti, A., Sparacino, G., & Cobelli, C. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0208 · Cited by 46
Self-citationWearable systems and mobile applications for diabetes disease management
Georga, E. I., Protopappas, V. C., Bellos, C. V., & Fotiadis, D. I. (2014) · Health and Technology · doi:10.1007/s12553-014-0082-y · Cited by 53
Developing the UVA/Padova Type 1 Diabetes Simulator: Modeling, Validation, Refinements, and Utility
Cobelli, C., & Kovatchev, B. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231195081 · Cited by 46
Self-citationBenhamou, P., Catargi, B., Delenne, B., Guerci, B., Hanaire, H. P., Jeandidier, N., Leroy, R., Meyer, L., Penfornis, A., Radermecker, R. P., et al. (2012) · Diabetes & Metabolism · doi:10.1016/s1262-3636(12)71538-0 · Cited by 52
“These models have already proven in a hospital milieu their safety and effectiveness in the prevention of nocturnal hypoglycaemic events in adults [58] and in children and adolescents [59].”
BackgroundEvaluating the Accuracy and Large Inaccuracy of Two Continuous Glucose Monitoring Systems
Leelarathna, L., Nodale, M., Allen, J. M., Elleri, D., Kumareswaran, K., Haidar, A., Caldwell, K., Wilinska, M. E., Acerini, C. L., Evans, M. L., et al. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0245 · Cited by 43
Multicenter Closed-Loop/Hybrid Meal Bolus Insulin Delivery with Type 1 Diabetes
Chase, H. P., Doyle, F. J., Zisser, H., Renard, É., Nimri, R., Cobelli, C., Buckingham, B. A., Maahs, D. M., Anderson, S. M., Magni, L., et al. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0050 · Cited by 47
Self-citationHanaire, H. P., Franc, S., Borot, S., Penfornis, A., Benhamou, P., Schaepelynck, P., Renard, É., Guerci, B., Jeandidier, N., Simon, C., et al. (2019) · Diabetes Obesity and Metabolism · doi:10.1111/dom.13898 · Cited by 46
“Most studies on CL systems have been conducted in current life conditions with standardized meals.(20-22) However, it appears that CL systems perform markedly better at night because daytime glycaemia”
BackgroundMethodModeling glycemia in humans by means of Grammatical Evolution
Hidalgo, J. I., Colmenar, J. M., Risco‐Martín, J. L., Cuesta‐Infante, A., Maqueda, E., Botella, M., & Rubio, J. A. (2013) · Applied Soft Computing · doi:10.1016/j.asoc.2013.11.006 · Cited by 49
“Hence, we can also find some personalized control approaches [10][11][12][13] corresponding to clinical practice.”
BackgroundNimri, R., Bratina, N., Kordonouri, O., Avbelj Stefanija, M., Fath, M., Biester, T., Muller, I., Atlas, E., Miller, S., Fogel, A., et al. (2017) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.12852 · Cited by 42
Abitbol, A., Rabasa-Lhoret, R., Messier, V., Legault, L., Smaoui, M., Cohen, N., & Haidar, A. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0353 · Cited by 39
Continuous-time interval model identification of blood glucose dynamics for type 1 diabetes
Kirchsteiger, H., Johansson, R., Renard, É., & del Re, L. (2014) · International Journal of Control · doi:10.1080/00207179.2014.897004 · Cited by 44
“In spite of intensive research (see, e.g., Harvey et al., 2010) no control system proposed in the literature so far was able to handle this task in presence of meals with reasonable accuracy, although first results presented in Boiroux et al. (2012) and Kovatchev et al. (2010) look promising.”
ResultAP@home: A Novel European Approach to Bring the Artificial Pancreas Home
Heinemann, L., Benesch, C., & DeVries, J. H. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500607 · Cited by 41
Closed loop developments to improve glucose control at home
Renard, É., Cobelli, C., & Kovatchev, B. P. (2013) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2013.09.009 · Cited by 40
Self-citationModel-Based Closed-Loop Glucose Control in Type 1 Diabetes: The DiaCon Experience
Schmidt, S., Boiroux, D., Duun-Henriksen, A. K., Frøssing, L., Skyggebjerg, O., Jørgensen, J. B., Poulsen, N. K., Madsen, H., Madsbad, S., & Nørgaard, K. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700515 · Cited by 35
Real-Time Model-Based Fault Detection of Continuous Glucose Sensor Measurements
Turksoy, K., Roy, A., & Çınar, A. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2535412 · Cited by 38
Wearable and implantable pancreas substitutes
Ricotti, L., Assaf, T., Dario, P., & Menciassi, A. (2012) · Journal of Artificial Organs · doi:10.1007/s10047-012-0660-6 · Cited by 39
Sherr, J. L., Palau Collazo, M., Cengiz, E., Michaud, C., Carria, L., Steffen, A. T., Weyman, K., Zgorski, M., Tichy, E., Tamborlane, W. V., et al. (2014) · Diabetes Care · doi:10.2337/dc13-1608 · Cited by 35
Kovatchev, B. P. (2012) · Scientifica · doi:10.6064/2012/283821 · Cited by 39
“Only three months later, in April 2008, the �rst human trials began at the University of �irginia (USA), Montpellier (France), and Padova (Italy), using a control system designed entirely in silico [151].”
BackgroundSelf-citationLinear Matrix Inequality-based Robust Controller design for Type-1 Diabetes Model
Szalay, P., Eigner, G., & Kovács, L. (2014) · IFAC Proceedings Volumes · doi:10.3182/20140824-6-za-1003.02451 · Cited by 32
“Considering the control algorithm in the AP there were various different approaches, including classical PID controllers Palerm (2011), run-to-run control Zisser et al. (2009), fuzzy logic based control Phillip et al. (2013), model-free soft computing-based control Zarkogianni et al. (2011), exact linearization based…”
BackgroundInfluentialFeasibility Study of a Bio-inspired Artificial Pancreas in Adults with Type 1 Diabetes
Reddy, M., Herrero, P., El-Sharkawy, M., Pesl, P., Jugnee, N., Thomson, H., Pavitt, D. V., Toumazou, C., Johnston, D. G., Georgiou, P., et al. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0009 · Cited by 35
“Twenty adult subjects with T1DM (55% male; mean [SD] age, 44 [10] years; duration of diabetes, 22 [12] years; duration of insulin pump therapy, 3.4 [4] years; HbA1c, 7.4% [0.7%]; body mass index, 25 [4] kg/m 2 ) participated in the 6-h fasting closed-loop study.”
InfluentialTowards a Verified Artificial Pancreas: Challenges and Solutions for Runtime Verification
Cameron, F., Fainekos, G., Maahs, D. M., & Sankaranarayanan, S. (2015) · Lecture notes in computer science · doi:10.1007/978-3-319-23820-3_1 · Cited by 39
“4 Hybrid Closed Loop [35,34,33] Closed loop insulin delivery with manual bolus 5 Fully Autoamted Closed Loop [7,8,9,15,44,38,19] #4 with all manual meal boluses eliminated”
MethodKumareswaran, K., Elleri, D., Allen, J. M., Harris, J., Xing, D., Kollman, C., Nodale, M., Murphy, H. R., Amiel, S. A., Heller, S. R., et al. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500606 · Cited by 31
Fabrication of a Flexible Amperometric Glucose Sensor Using Additive Processes
Du, X., Durgan, C. J., Matthews, D. J., Motley, J. R., Tan, X., Pholsena, K., Árnadóttir, L., Castle, J. R., Jacobs, P. G., Cargill, R. S., et al. (2015) · ECS Journal of Solid State Science and Technology · doi:10.1149/2.0101504jss · Cited by 32
Improved postprandial glucose control with a customized Model Predictive Controller
Messori, M., Ellis, M. J., Cobelli, C., Christofides, P. D., & Magni, L. (2015) · Proceedings of the ... American Control Conference/Proceedings of the American Control Conference · doi:10.1109/acc.2015.7172136 · Cited by 29
Self-citationBastyr, E. J., Zhang, S., Mou, J., Hackett, A. P., Raymond, S. A., & Chang, A. M. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0407 · Cited by 25
Hidalgo, J. I., Botella, M., Velasco, J. M., Garnica, Ó., Cervigón, C., Martínez, R. R., Aramendi, A., Maqueda, E., & Lanchares, J. (2019) · Applied Soft Computing · doi:10.1016/j.asoc.2019.105923 · Cited by 26
Turksoy, K., Bayrak, E. S., Quinn, L., Littlejohn, E., & Çınar, A. (2013) · Proceedings of the ... American Control Conference/Proceedings of the American Control Conference · doi:10.1109/acc.2013.6580275 · Cited by 26
Cobelli, C., & Man, C. D. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211015268 · Cited by 24
“131 In brief, to test the validity of the computer simulation environment independently from the data used for its development, a number of experiments were conducted, aiming to assess the model capability to reflect the variety of clinical situations as closely as possible.”
MethodSelf-citation
Metadata
- Authors
- Kovatchev, B., Cobelli, C., Renard, E., Anderson, S., Breton, M., Patek, S., Clarke, W., Bruttomesso, D., Maran, A., Costa, S., Avogaro, A., Man, C. D., Facchinetti, A., Magni, L., De Nicolao, G., Place, J., & Farret, A.
- Affiliations
- University of Virginia (US)
- University of Padua (IT)
- Department of Endocrinology and UMR CNRS, CHU and University of Montpellier, Montpellier, France
- University of Pavia (IT)
- Department of System and Informatics, University of Pavia, Pavia, Italy;
- Centre National de la Recherche Scientifique (FR)
- Corresponding author
- Kovatchev, Boris
- ORCID
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2010-11
- Volume
- 4
- Issue
- 6
- Pages
- 1374-1381
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC3005047
- OpenAlex
- W2134017961
- Semantic Scholar
- faf6284087dadbb4011a4ee2564dc9ed8bad111a
- Cited by
- 200
- References
- 25
- Citations by year
- 2026: 2 · 2024: 2 · 2023: 6 · 2022: 5 · 2021: 9 · 2020: 4 · 2019: 15 · 2018: 11 · 2017: 14 · 2016: 13 · 2015: 19 · 2014: 29 · 2013: 32 · 2012: 24
- FWCI
- 17.19
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Padua Research Archive (University of Padova) (submittedVersion)
- Funders
- National Institutes of Health (5r01dk051562-23)
- National Institute of Diabetes and Digestive and Kidney Diseases (R01DK 51562)
- National Center for Research Resources (M01RR000847, M01-RR 00847)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Keywords
- type 1 diabetes, feedback control, model predictive control, artificial pancreas, in silico design, computer simulation, nocturnal hypoglycemia, continuous glucose monitoring, insulin pump
- MeSH terms
- Adolescent, Adult, Aged, Algorithms, Blood Glucose, Child, Computer Simulation, Diabetes Mellitus, Type 1, Female, France, Glycated Hemoglobin, Humans, Hypoglycemia, Hypoglycemic Agents, Insulin, Insulin Infusion Systems, Italy, Male, Middle Aged, Models, Biological, Monitoring, Physiologic, Pilot Projects, Time Factors, Virginia, Treatment Outcome, Pancreas, Artificial, Drug Dosage Calculations, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.