Safe Glycemic Management during Closed-Loop Treatment of Type 1 Diabetes: The Role of Glucagon, Use of Multiple Sensors, and Compensation for Stress Hyperglycemia
Open accessAbstract
Patients with type 1 diabetes mellitus (T1DM) must make frequent decisions and lifestyle adjustments in order to manage their disorder. Automated treatment would reduce the need for these self-management decisions and reduce the risk for long-term complications. Investigators in the field of closed-loop glycemic control systems are now moving from inpatient to outpatient testing of such systems. As outpatient systems are developed, the element of safety increases in importance. One such concern is the risk for hypoglycemia, due in part to the delayed onset and prolonged action duration of currently available subcutaneous insulin preparations. We found that, as compared to an insulin-only closed-loop system, a system that also delivers glucagon when needed led to substantially less hypoglycemia. Though the capability of glucagon delivery would mandate the need for a second hormone chamber, glucagon in small doses is tolerated very well. People with T1DM often develop hyperglycemia from emotional stress or medical stress. Automated closed-loop systems should be able to detect such changes in insulin sensitivity and adapt insulin delivery accordingly. We recently verified the adaptability of a model-based closed-loop system in which the gain factors that govern a proportional-integral-derivative-like system are adjusted according to frequently measured insulin sensitivity. Automated systems can be tested by physical exercise to increase glucose uptake and insulin sensitivity or by administering corticosteroids to reduce insulin sensitivity. Another source of risk in closed-loop systems is suboptimal performance of amperometric glucose sensors. Inaccuracy can result from calibration error, biofouling, and current drift. We found that concurrent use of more than one sensor typically leads to better sensor accuracy than use of a single sensor. For example, using the average of two sensors substantially reduces the proportion of large sensor errors. The use of more than two allows the use of voting algorithms, which can temporarily exclude a sensor whose signal is outlying. Elements such as the use of glucagon to minimize hypoglycemia, adaptation to changes in insulin sensitivity, and sensor redundancy will likely increase safety during outpatient use of closed-loop glycemic control systems.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Cites
Bergman, R. N., Phillips, L. S., & Cobelli, C. (1981) · Journal of Clinical Investigation · doi:10.1172/jci110398 · Cited by 1,640 · see citing works
Nonlinear model predictive control of glucose concentration in subjects with type 1 diabetes
Hovorka, R., Canonico, V., Chassin, L. J., Haueter, U., Massi-Benedetti, M., Federici, M. O., Pieber, T. R., Schaller, H. C., Schaupp, L., Vering, T., et al. (2004) · Physiological Measurement · doi:10.1088/0967-3334/25/4/010 · Cited by 1,274 · see citing works
[Lys(B28), Pro(B29)]-Human Insulin: A Rapidly Absorbed Analogue of Human Insulin
Howey, D. C., Bowsher, R. R., Brunelle, R. L., & Woodworth, J. R. (1994) · Diabetes · doi:10.2337/diab.43.3.396 · Cited by 471 · see citing works
A Bihormonal Closed-Loop Artificial Pancreas for Type 1 Diabetes
El-Khatib, F. H., Russell, S. J., Nathan, D. M., Sutherlin, R. G., & Damiano, E. R. (2010) · Science Translational Medicine · doi:10.1126/scitranslmed.3000619 · Cited by 403 · see citing works
Novel Use of Glucagon in a Closed-Loop System for Prevention of Hypoglycemia in Type 1 Diabetes
Castle, J. R., Engle, J. M., Youssef, J. E., Massoud, R. G., Yuen, K. C. J., Kagan, R., & Ward, W. K. (2010) · Diabetes Care · doi:10.2337/dc09-2254 · Cited by 232 · see citing works
Self-citationFederiuk, I. F., Casey, H. M., Quinn, M. J., et al. (2004) · PubMed · Cited by 153 · see citing works
Fear of hypoglycaemia in adults with type 1 diabetes
Anderbro, T., Amsberg, S., Adamson, U., Bolinder, J., Lins, P. ‐., Wredling, R., Moberg, E., Lisspers, J., & Johansson, U. ‐. (2010) · Diabetic Medicine · doi:10.1111/j.1464-5491.2010.03078.x · Cited by 128 · see citing works
Mini-Dose Glucagon Rescue for Hypoglycemia in Children With Type 1 Diabetes
Haymond, M. W., & Schreiner, B. (2001) · Diabetes Care · doi:10.2337/diacare.24.4.643 · Cited by 114 · see citing works
Taborsky, G. J. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400607 · Cited by 96 · see citing works
(1963) · PubMed · Cited by 95 · see citing works
Shulman, G. I., Rossetti, L., Rothman, D. L., et al. (1987) · Journal of Clinical Investigation · doi:10.1172/jci113084 · Cited by 92 · see citing works
Pharmacokinetics of insulin aspart in obesity, renal impairment, or hepatic impairment
Holmes, G. B., Galitz, L., Hu, P., & Lyness, W. H. (2005) · British Journal of Clinical Pharmacology · doi:10.1111/j.1365-2125.2005.02476.x · Cited by 86 · see citing works
Insulin, glucagon, and catecholamines in prevention of hypoglycemia during fasting
Boyle, P. J., SHAH, S. D., & Cryer, P. E. (1989) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.1989.256.5.e651 · Cited by 83 · see citing works
A Review of Closed-Loop Algorithms for Glycemic Control in the Treatment of Type 1 Diabetes
Youssef, J. E., Castle, J. R., & Ward, W. K. (2009) · Algorithms · doi:10.3390/a2010518 · Cited by 69 · see citing works
Self-citationOptimization of the Native Glucagon Sequence for Medicinal Purposes
Chabenne, J. R., DiMarchi, M. A., Gelfanov, V. M., & DiMarchi, R. D. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400605 · Cited by 65 · see citing works
Nixon, R., & Pickup, J. C. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0192 · Cited by 51 · see citing works
Factors Influencing the Effectiveness of Glucagon for Preventing Hypoglycemia
Castle, J. R., Engle, J. M., Youssef, J. E., Massoud, R. G., & Ward, W. K. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400603 · Cited by 47 · see citing works
Self-citationInsulin-induced hypoglycemia increases hepatic sensitivity to glucagon in dogs
Rivera, N., Ramnanan, C. J., An, Z., Farmer, T. E., Smith, M. S., Farmer, B., Irimia, J. M., Snead, W. L., Lautz, M., Roach, P. J., et al. (2010) · Journal of Clinical Investigation · doi:10.1172/jci40919 · Cited by 46 · see citing works
Assessment of Chronically Implanted Subcutaneous Glucose Sensors in Dogs
Ward, W. K., & Troupe, J. E. (1999) · ASAIO Journal · doi:10.1097/00002480-199911000-00009 · Cited by 43 · see citing works
Self-citationYoussef, J. E., Castle, J. R., Branigan, D. L., Massoud, R. G., Breen, M. E., Jacobs, P. G., Bequette, B. W., & Ward, W. K. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500602 · Cited by 42 · see citing works
Self-citationGopakumaran, B., Duman, H. M., Overholser, D. P., Federiuk, I. F., Quinn, M. J., Wood, M. D., & Ward, W. K. (2005) · Artificial Organs · doi:10.1111/j.1525-1594.2005.29096.x · Cited by 41 · see citing works
Self-citationEfficacy Determinants of Subcutaneous Microdose Glucagon during Closed-Loop Control
Russell, S. J., El-Khatib, F. H., Nathan, D. M., & Damiano, E. R. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400602 · Cited by 39 · see citing works
Stabilized Glucagon Formulation for Bihormonal Pump Use
Steiner, S. S., Li, M., Hauser, R. G., et al. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400606 · Cited by 39 · see citing works
El-Khatib, F. H., Jiang, J., Gerrity, R. G., & Damiano, E. R. (2007) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.0006 · Cited by 29 · see citing works
Ward, W. K., Engle, J. M., Duman, H. M., Bergstrom, C. P., Kim, S. F., & Federiuk, I. F. (2008) · IEEE Sensors Journal · doi:10.1109/jsen.2007.912912 · Cited by 24 · see citing works
Self-citationWard, W. K., Massoud, R. G., Szybala, C., et al. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400604 · Cited by 21 · see citing works
Self-citationWard, W. K., Wood, M. D., & Troupe, J. E. (2000) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(00)00051-8 · Cited by 18 · see citing works
Self-citationAn Implantable Subcutaneous Glucose Sensor Array in Ketosis‐prone Rats: Closed Loop Glycemic Control
Ward, W. K., Wood, M. D., Casey, H. M., Quinn, M. J., & Federiuk, I. F. (2005) · Artificial Organs · doi:10.1111/j.1525-1594.2005.29024.x · Cited by 16 · see citing works
Self-citationWard, W. K., Wood, M. D., & Troupe, J. E. (2000) · ASAIO Journal · doi:10.1097/00002480-200009000-00006 · Cited by 13 · see citing works
Self-citationFarhy, L. S., & McCall, A. L. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400608 · Cited by 13 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Continuous glucose monitoring: a review of recent studies demonstrating improved glycemic outcomes
Rodbard, D. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0035 · Cited by 458 · see citing works
An Integrated Multivariable Artificial Pancreas Control System
Turksoy, K., Quinn, L. T., Littlejohn, E., & Cinar, A. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814524862 · Cited by 77 · see citing works
Pharmacokinetics Modeling of Exogenous Glucagon in Type 1 Diabetes Mellitus Patients
Lv, D., Breton, M. D., & Farhy, L. S. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0150 · Cited by 17 · see citing works
- Authors
- Ward, W. K., Castle, J. R., & Youssef, J. E.
- Affiliations
- Oregon Health & Science University (US)
- Corresponding author
- Ward, W. Kenneth
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2011-11
- Volume
- 5
- Issue
- 6
- Pages
- 1373-1380
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC3262703
- OpenAlex
- W2015800577
- Cited by
- 36
- References
- 30
- Citations by year
- 2025: 1 · 2022: 2 · 2021: 1 · 2020: 1 · 2018: 4 · 2017: 2 · 2016: 2 · 2015: 8 · 2014: 7 · 2013: 6 · 2012: 2
- FWCI
- 0.67
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Institutes of Health (5k23dk090133-03)
- National Institute of Diabetes and Digestive and Kidney Diseases (K23-DK-090133)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Medicine › Surgery › Pancreatic function and diabetes
- Keywords
- type 1 diabetes, hypoglycemia prevention, stress hyperglycemia, insulin sensitivity, sensor redundancy, voting algorithm, sensor error
- MeSH terms
- Algorithms, Blood Glucose, Diabetes Mellitus, Type 1, Glucagon, Humans, Hyperglycemia, Hypoglycemic Agents, Insulin, Stress, Psychological, Pancreas, Artificial
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review