Performance Variability of Seven Commonly Used Self-Monitoring of Blood Glucose Systems: Clinical Considerations for Patients and Providers
Open accessThe DOI matches its Crossref record; 15 of the 15 references it cites were found in a registry or an index.
Abstract
Background: Blood glucose data are frequently used in clinical decision making, thus it is critical that self-monitoring of blood glucose (SMBG) systems consistently provide accurate results. Concerns about SMBG accuracy have prompted the development of newly proposed International Organization for Standardization (ISO) standards: ≥95% of individual glucose results shall fall within ±15 mg/dl of the results of the manufacturer's reference procedure at glucose concentrations <100 mg/dl and within ±15% for values >100 mg/dl. We evaluated seven marketed systems against the current and proposed ISO criteria (criterion A). Method: Capillary blood samples were collected from 100 subjects and tested on seven systems: Accu-Chek Aviva Plus, Advocate Redi-Code, Element, Embrace, Prodigy Voice, TRUEbalance, and WaveSense Presto. Results were compared with manufacturer's documented reference system, YSI or perchloric acid hexokinase; three different strip lots from each system were tested on each subject, in duplicate. Results: Compared against current ISO criteria (≥95% within ±15 mg/dl for values <75 mg/dl and ±20% for values ≥75 mg/dl) the Accu-Chek Aviva Plus, Element, and WaveSense Presto systems met accuracy criteria. However, only the Accu-Chek Aviva Plus met the proposed ISO criteria (criterion A) in all three lots. The other six systems failed to meet the criteria in at least two of the three lots, showing lot-to-lot variability, high/low bias, and variations due to hematocrit. Conclusions: Inaccurate SMBG readings can potentially adversely impact clinical decision making and outcomes. Clinicians can reduce controllable variables by prescribing accurate SMBG systems. Adherence to the proposed ISO criteria should enhance patient safety by improving the accuracy of SMBG systems.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“There is substantial variation in the accuracy of widely used BGM systems (17,18).”
From 7. Diabetes Technology: Standards of Care in Diabetes—2025 (2025) · cites this paper as [18]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
International Organization for Standardization
(2011) · SpringerReference · doi:10.1007/springerreference_16847 · Cited by 12,534
Besarab, A., Bolton, W. K., Browne, J. K., Egrie, J. C., Nissenson, A. R., Okamoto, D. M., Schwab, S. J., & Goodkin, D. A. (1998) · New England Journal of Medicine · doi:10.1056/nejm199808273390903 · Cited by 1,930
Polonsky, W. H., Fisher, L., Schikman, C. H., Hinnen, D. A., Parkin, C. G., Jelsovsky, Z., Petersen, B., Schweitzer, M., & Wagner, R. S. (2011) · Diabetes Care · doi:10.2337/dc10-1732 · Cited by 429
Self-citationTang, Z., Lee, J. H., Louie, R. F., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-1135-eodhlo · Cited by 293
System Accuracy Evaluation of 27 Blood Glucose Monitoring Systems According to DIN EN ISO 15197
Freckmann, G., Baumstark, A., Jendrike, N., Zschornack, E., Kocher, S., Tshiananga, J., Heister, F., & Haug, C. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0128 · Cited by 195 · cited 1×
“, Taipei, Taiwan); (6) TRUEbalance (Nipro Diagnostics, Ft.”
Counterregulatory hormone responses to hypoglycemia in the elderly patient with diabetes
Meneilly, G. S., Cheung, E., & Tuokko, H. (1994) · Diabetes · doi:10.2337/diab.43.3.403 · Cited by 167
Counterregulatory hormone responses to hypoglycemia in the elderly patient with diabetes
Meneilly, G. S., Cheung, E., & Tuokko, H. (1994) · Diabetes · doi:10.2337/diabetes.43.3.403 · Cited by 167
Meneilly, G. S., & Tessier, D. (1995) · Diabetic Medicine · doi:10.1111/j.1464-5491.1995.tb00405.x · Cited by 132
Cited by
7. Diabetes Technology: Standards of Medical Care in Diabetes—2022
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s007 · Cited by 358
7. Diabetes Technology: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s007 · Cited by 348
7. Diabetes Technology: Standards of Care in Diabetes — 2023
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s007 · Cited by 301
7. Diabetes Technology: Standards of Care in Diabetes—2024
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Hilliard, M. E., Johnson, E. L., Khunti, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s007 · Cited by 281
7. Diabetes Technology: Standards of Medical Care in Diabetes—2021
American Diabetes Association (2021) · Diabetes Care · doi:10.2337/dc21-s007 · Cited by 202
“There is substantial variation in the accuracy of widely used blood glucose monitoring systems (2,3).”
MethodWadwa, R. P., Laffel, L. M., Shah, V. N., & Garg, S. K. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0150 · Cited by 198
Clinical accuracy of a continuous glucose monitoring system with an advanced algorithm
Bailey, T. S., Chang, A., & Christiansen, M. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814559746 · Cited by 168
Comparative accuracy of 17 point-of-care glucose meters
Ekhlaspour, L., Mondesir, D., Lautsch, N., Balliro, C., Hillard, M., Magyar, K., Radocchia, L. G., Esmaeili, A., Sinha, M., & Russell, S. J. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816672237 · Cited by 127
Continuous Glucose Monitoring: A Brief Review for Primary Care Practitioners
Ajjan, R., Slattery, D., & Wright, E. (2019) · Advances in Therapy · doi:10.1007/s12325-019-0870-x · Cited by 120
Bailey, T. S., Grunberger, G., Bode, B. W., Handelsman, Y., Hirsch, I. B., Jovanovič, L., Roberts, V. L., Rodbard, D., Tamborlane, W. V., & Walsh, J. S. A. (2016) · Endocrine Practice · doi:10.4158/ep151124.cs · Cited by 151
Klatman, E. L., Jenkins, A. J., Ahmedani, M. Y., & Ogle, G. D. (2019) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(18)30074-3 · Cited by 100
Rajendran, R., & Rayman, G. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814538940 · Cited by 84
“Of greater concern was once approved (by Communauté Européenne [CE] in Europe or FDA in the United States) many of the existing meters (both hospital and outpatient use) failed to maintain the original standard.(60-65) Most clinicians and patients are unaware of this.”
BackgroundFreckmann, G., Baumstark, A., Schmid, C., Pleus, S., Link, M., & Haug, C. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0208 · Cited by 77
“5% of the measurement results within the required accuracy limits; comparable results for some of the investigated systems have been reported in other studies.(6,8,17) Considering the recently published revision ISO 15197:2013,(5) the systems’ measurement results within the required limits varied between 54.”
Freckmann, G., Schmid, C., Baumstark, A., Rutschmann, M., Haug, C., & Heinemann, L. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815580160 · Cited by 70
“Unsurprisingly, a remarkable number of systems failed to meet the more stringent accuracy criteria of the new version of this standard.(14,16,23,44) However, accuracy data of such postmarketing evaluations do not allow for a general conclusion regarding the measurement performance of a given SMBG system as this evalua…”
BackgroundPerformance of Cleared Blood Glucose Monitors
Klonoff, D. C., & Prahalad, P. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815584797 · Cited by 59
Feasibility of Factory Calibration for Subcutaneous Glucose Sensors in Subjects With Diabetes
Hoss, U., Budiman, E. S., Liu, H., & Christiansen, M. P. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296813511747 · Cited by 53
“BG test strips themselves have a certain error,(9) and the overall accuracy may vary from one lot to another.(10) More important, the corresponding tissue glucose value may have a time lag and its ratio to BG may be different when measured under steady state conditions.”
BackgroundFreckmann, G., Link, M., Schmid, C., Pleus, S., Baumstark, A., & Haug, C. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0085 · Cited by 46
“17 Although for some of the evaluated systems similar accuracy results have been published, 8,12–14,18–21 the comparability of accuracy data obtained ).”
Continuous Glucose Monitoring in the Subcutaneous Tissue over a 14-Day Sensor Wear Period
Hoss, U., Budiman, E. S., Liu, H., & Christiansen, M. P. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700511 · Cited by 46
A review of blood glucose monitor accuracy
King, F., Ahn, D., Hsiao, V., Porco, T., & Klonoff, D. C. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0232 · Cited by 37
Boettcher, C., Dost, A., Wudy, S. A., Flechtner-Mors, M., Borkenstein, M., Schiel, R., Weitzel, D., Bechtold-Dalla Pozza, S., Wolf, J., Holl, R. W., et al. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0262 · Cited by 35
Klonoff, D. C., & Reyes, J. S. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700432 · Cited by 29
Accuracy Evaluation of CONTOUR®PLUS Compared With Four Blood Glucose Monitoring Systems
Dunne, N., Viggiani, M. T., Pardo, S. A., Robinson, C., & Parkes, J. L. (2015) · Diabetes Therapy · doi:10.1007/s13300-015-0121-3 · Cited by 31
Glurich, I., Bartkowiak, B., Berg, R. L., & Acharya, A. (2018) · International Dental Journal · doi:10.1111/idj.12405 · Cited by 26
Puckrein, G. A., Nunlee-Bland, G., Zangeneh, F., Davidson, J. A., Vigersky, R. A., Xu, L., Parkin, C. G., & Marrero, D. G. (2016) · Diabetes Care · doi:10.2337/dc15-1264 · Cited by 23
“Additionally, because the program may havemade it difficult formany beneficiaries to continue using the SMBG systems theywere familiar with, it is possible that these beneficiaries reduced or discontinued their SMBG due to lack of training and/or lack of confidence in the accuracy of their new SMBG system (20).”
BackgroundSelf-citationGrunberger, G., Bailey, T. S., Camacho, P. M., EINHORN, D. F., Garber, A. J., Handelsman, Y., Harrell, R. M., Lando, H. M., Law, B., Leffert, J. D., et al. (2015) · Endocrine Practice · doi:10.4158/ep15653.cs · Cited by 22
“However, at least 6 studies evaluating the performance of blood GM devices published between 2010 and 2014 have demonstrated that only 40 to 83% met ISO 2003 standards, and only 14 to 67% met ISO 2013 standards (19-24).”
The financial impact of inaccurate blood glucose monitoring systems
Fortwaengler, K., Campos-Náñez, E., Parkin, C. G., & Breton, M. D. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817731423 · Cited by 17
“In their 2013 study, Brazg and colleagues assessed the accuracy of 7 BGM systems, testing three test lots for each system.(27) The investigators observed that although three of the seven systems met ISO 15197:2003 criteria, only one system met the current ISO 15197:2013 criteria.”
BackgroundSelf-citationYu-Fei, W., Wei-Ping, J., Ming-Hsun, W., Miao-O, C., Ming-Chang, H., Chi-Pin, W., & Ming-Shih, L. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817705143 · Cited by 15
Price, D. A., & Walker, T. C. (2016) · touchREVIEWS in Endocrinology · doi:10.17925/ee.2016.12.01.24 · Cited by 20
“Inaccuracy may stem from inadequate manufacturing practices and controls.(33,34) Recent studies have demonstrated that >45% of marketed SMBG systems do not meet the minimum International Organization for Standardization (ISO) accuracy standards.”
BackgroundHowell, A. P., Parrett, J. L., & Malcom, D. R. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819889638 · Cited by 15
Electrochemical Glucose Biosensors for Diabetes Care
Ocvirk, G., Buck, H., & DuVall, S. H. (2016) · Bioanalytical reviews · doi:10.1007/11663_2016_3 · Cited by 16
Heinemann, L., Freckmann, G., & Koschinsky, T. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700230 · Cited by 14
New Generation Blood Glucose Monitoring System Exceeds International Accuracy Standards
Brazg, R., Klaff, L. J., & Sussman, A. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816652902 · Cited by 12
“The system was developed on a foundation of electrochemical technology that provides notably improved accuracy over previous Accu-Chek product generations.(3) Our study demonstrated that the system met and exceeded current ISO accuracy requirements.”
MethodSelf-citationJoseph, J. I. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700519 · Cited by 11
If SMBG Accuracy Is Critical to Patient Safety, Why Are Inaccurate Meters Still on the Market?
Parkin, C. G. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817693256 · Cited by 11
Self-citationChen, H., Yao, Q., Dong, Y., Tang, Z., Li, R., Cai, B., Wang, R., & Chen, Q. (2019) · Primary care diabetes · doi:10.1016/j.pcd.2018.12.010 · Cited by 10
Diabetes Technology and Therapy in the Pediatric Age Group
Maahs, D. M., & Shalitin, S. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.2509 · Cited by 8
“Over the last years several publications have pointed out that not all previously approved glucose-meter devices conform to the current and/or planned ISO standards (12,13), which may have an impact on clinical decision making and glucose control.”
Elabd, S. S., & Poole, R. (2023) · Practical Diabetes · doi:10.1002/pdi.2479 · Cited by 8
The problem with total error models in establishing performance specifications and a simple remedy
Krouwer, J. S. (2016) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2015-1175 · Cited by 7
Link, M., Pleus, S., Schmid, C., Freckmann, G., Baumstark, A., Stolberg, E., & Haug, C. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296813518859 · Cited by 6
Blood Glucose Meter Market: This World is Undergoing Drastic Changes
Heinemann, L., & Klonoff, D. C. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700301 · Cited by 6
Freckmann, G., Pleus, S., Link, M., & Haug, C. (2015) · Lecture notes in bioengineering · doi:10.1007/978-3-319-25913-0_2 · Cited by 5
Impact of the 2013 National Rollout of CMS Competitive Bidding Program: The Disruption Continues
Puckrein, G. A., Hirsch, I. B., Parkin, C. G., Taylor, B. T., Xu, L., & Marrero, D. G. (2017) · Diabetes Care · doi:10.2337/dc17-0960 · Cited by 5
Self-citationJeanny, G., & Hope, P. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815608871 · Cited by 5
“This is a 1-time process, and postmarket release evaluations or independent evaluations are not mandatory, despite studies repeatedly showing test strip lots of available systems not adhering with ISO 15197 accuracy criteria.(3-6) In 2 studies, performed in April and May 2013 and in October 2014, the SMBG systems Gluc…”
Method10th Annual Symposium on Self-Monitoring of Blood Glucose, April 27–29, 2017, Warsaw, Poland
Parkin, C. G., Homberg, A., & Hinzmann, R. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0356 · Cited by 6
“have shown the Accu-Chek Guide system achieving tighter 10/10 accuracy.(48) Further calculations of the study data have shown that 95% of 10TH ANNUAL SMBG SYMPOSIUM 83”
MethodSelf-citationThe Danger of Using Total Error Models to Compare Glucose Meter Performance
Krouwer, J. S. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296813518673 · Cited by 6
“Thus, some glucose meters suffer from hematocrit interference and others not.(7) The CLSI standard EP21-A takes a different approach to estimating total error by directly computing the differences between the candidate assay and reference.”
BackgroundLuppa, P. B., Zeller, M., Pieper, M., Kaiser, P., Weiss, N., Vierbaum, L., & Freckmann, G. (2024) · Frontiers in Molecular Biosciences · doi:10.3389/fmolb.2024.1371426 · Cited by 4
Biases in clinical trials performed for regulatory approval
Krouwer, J. S. (2015) · Accreditation and Quality Assurance · doi:10.1007/s00769-015-1145-0 · Cited by 4
“not meet the performance standards [1] that allowed the meters to be approved [2, 3].”
BackgroundAlzaid, A., Schlaeger, C., & Hinzmann, R. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0260 · Cited by 4
“When accuracy results from the four- and three-lot assessment study were scrutinized further using the current ISO 15197 standard, only one meter system passed the full test: the Accu-Chek Aviva system.(30,31) Hattemer concluded his talk by stating that only the AccuChek Aviva system had successfully passed all three…”
MethodKonsequenzen der Genauigkeit von Blutglucosemesssystemen: Klinische und ökonomische Aspekte
Hessel, F. P., Freckmann, G., & Resch, A. (2014) · Der Diabetologe · doi:10.1007/s11428-013-1124-1 · Cited by 3
The Chronic Injury Glucose Error Grid
Krouwer, J. S., & Cembrowski, G. S. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814554415 · Cited by 3
“For example, some glucose meters are affected by hematocrit with biases of 20% at the extremes of hematocrit and zero bias at the average value of hematocrit.(9) Large patient interferences can be observed visually from a graph.”
BackgroundЕвгеньевна, Р. Г. (2021) · Meditsinskiy sovet = Medical Council · doi:10.21518/2079-701x-2021-12-286-292 · Cited by 3
7 th Annual Symposium on Self-Monitoring of Blood Glucose (SMBG), May 8–10, 2014, Helsinki, Finland
Parkin, C. G., Mlinac, A., & Hinzmann, R. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0239 · Cited by 2
Self-citationOkorie-Kanu, C. O., Okorie-Kanu, O. J., Akwuobu, C. A., Tizhe, E. V., & Antia, R. E. (2021) · Nigerian Veterinary Journal · doi:10.4314/nvj.v41i3.2 · Cited by 2
Medicare Competitive Bidding Program
Parkin, C. G. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816662049 · Cited by 2
“According to former Congresswoman Nancy Johnson, a new round of bidding, which will slash prices by another 20%, occurred in January of this year.(4) After a 6-month phase-in period, the new prices will be applied nationally, starting July 1.”
BackgroundSelf-citationPfützner, A. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700118 · Cited by 1
“It highlights several important topics: (1) the need for more stringent post-marketing requirements for blood glucose meters after launch and (2) low-cost meters use older technologies and their manufacturers do not usually seriously invest in new technology or constant quality assurance efforts.”
The Standard of Analytical Quality of Glucometers
Friedecky, B., Springer, D., & Kratochvíla, J. (2014) · Klinická biochemie a metabolismus · doi:10.61568/kbm.2014.015 · Cited by 2
Modeling the SMBG measurement error
Vettoretti, M. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816714-4.00005-3 · Cited by 1
On the Bus or Under the Bus? The Current Healthcare System is Endangering Elderly Diabetes Patients
Parkin, C. G. (2015) · Journal of Diabetes Metabolic Disorders & Control · doi:10.15406/jdmdc.2015.02.00028
Self-citationGrunberger, G., Abelseth, J. M., Bailey, T. S., Bode, B. W., Handelsman, Y., Hellman, R., Lane, W. S., Raskin, P., Tamborlane, W. V., & Rothermel, C. (2014)
Parkin, C. G. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700420
Self-citationEffects of Postmeal Walking on Postprandial Glucose Control and Oxidative Stress
Knurick, J. R., Johnston, C. S., & Gaesser, G. A. (2016) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000486618.68805.05
Consequences of SMBG systems inaccuracy
Campos-Náñez, E., & Fortwaengler, K. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816714-4.00004-1
Barnabas, R., Shivane, V. K., Tamoli, S., Mahadik, S., Memon, S. S., Bandgar, T., & Mehta, B. (2024) · Medical Research Archives · doi:10.18103/mra.v12i12.6048
Recommendations for the use, selection and kontrol of glucometers
Friedecky, B., Springer, D., Kratochvíla, J., Škrha, J., & Zima, T. (2014) · Klinická biochemie a metabolismus · doi:10.61568/kbm.2014.026 · Cited by 4
Quality, control and validation of glucometers and CGM systems. Status overview
Friedecky, B., Kratochvíla, J., & Springer, D. (2016) · Klinická biochemie a metabolismus · doi:10.61568/kbm.2016.013 · Cited by 2
Friedecky, B. (2016) · Klinická biochemie a metabolismus · doi:10.61568/kbm.2016.028
Sia, G. F., & Yao, C. S. (2017) · Philippine Journal of Internal Medicine · doi:10.65564/pjim.138aeec745
A-/B-Blutzuckermesssysteme: Kommentar der Arbeitsgemeinschaft Diabetologische Technologie
Heinemann, L., & Freckmann, G. (2014) · Der Diabetologe · doi:10.1007/s11428-013-1166-4
Metadata
- Authors
- Brazg, R. L., Klaff, L. J., & Parkin, C. G.
- Affiliations
- Rainier Clinical Research Center (US)
- CGParkin Communications Inc., Boulder City, Nevada
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2013-01
- Volume
- 7
- Issue
- 1
- Pages
- 144-152
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC3692226
- OpenAlex
- W2165915429
- Cited by
- 68
- References
- 15
- Citations by year
- 2024: 2 · 2023: 2 · 2022: 1 · 2021: 3 · 2020: 3 · 2019: 3 · 2018: 4 · 2017: 6 · 2016: 11 · 2015: 10 · 2014: 14 · 2013: 9
- FWCI
- 11.54
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Keywords
- self-monitoring of blood glucose, blood glucose meter accuracy, lot-to-lot variation, measurement bias, hematocrit interference, capillary blood glucose, diabetes management, patient safety
- MeSH terms
- Blood Glucose, Diabetes Mellitus, Humans, Blood Glucose Self-Monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.