Multiplexed Echo Planar Imaging for Sub-Second Whole Brain FMRI and Fast Diffusion Imaging
Open accessThe DOI matches its Crossref record; 59 of the 59 references it cites were found in a registry or an index.
Abstract
Echo planar imaging (EPI) is an MRI technique of particular value to neuroscience, with its use for virtually all functional MRI (fMRI) and diffusion imaging of fiber connections in the human brain. EPI generates a single 2D image in a fraction of a second; however, it requires 2-3 seconds to acquire multi-slice whole brain coverage for fMRI and even longer for diffusion imaging. Here we report on a large reduction in EPI whole brain scan time at 3 and 7 Tesla, without significantly sacrificing spatial resolution, and while gaining functional sensitivity. The multiplexed-EPI (M-EPI) pulse sequence combines two forms of multiplexing: temporal multiplexing (m) utilizing simultaneous echo refocused (SIR) EPI and spatial multiplexing (n) with multibanded RF pulses (MB) to achieve m×n images in an EPI echo train instead of the normal single image. This resulted in an unprecedented reduction in EPI scan time for whole brain fMRI performed at 3 Tesla, permitting TRs of 400 ms and 800 ms compared to a more conventional 2.5 sec TR, and 2-4 times reductions in scan time for HARDI imaging of neuronal fibertracks. The simultaneous SE refocusing of SIR imaging at 7 Tesla advantageously reduced SAR by using fewer RF refocusing pulses and by shifting fat signal out of the image plane so that fat suppression pulses were not required. In preliminary studies of resting state functional networks identified through independent component analysis, the 6-fold higher sampling rate increased the peak functional sensitivity by 60%. The novel M-EPI pulse sequence resulted in a significantly increased temporal resolution for whole brain fMRI, and as such, this new methodology can be used for studying non-stationarity in networks and generally for expanding and enriching the functional information.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures






Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“and also increase the information content of the data (e.g., in terms of reflecting the richness of the neural dynamics) [41, 42]”
From Functional connectomics from resting-state fMRI (2013) · cites this paper as [41]
In the citing paper
“This efficiency increase can lead to substantially improved functional SNR (Feinberg et al., 2010; (Smith et al., 2011), the ability to acquire more diffusion data points (Sotiropoulos et al., 2013c), and/or increases in spatial resolution for fMRI or dMRI (Ugurbil et al. 2013).”
From The WU-Minn Human Connectome Project: An overview (2013) · cites this paper as Feinberg et al., 2010
In the citing paper
“The most significant pulse sequence development for the HCP was the implementation and optimization of slice-accelerated multiband (MB) acquisitions for fMRI and dMRI (Larkman et al., 2001; Moeller et al., 2008; Feinberg et al., 2010; Moeller et al., 2010; Setsompop et al., 2012; Ugurbil et al., 2013).”
From The WU-Minn Human Connectome Project: An overview (2013) · cites this paper as Feinberg et al., 2010
In the citing paper
“Higher temporal resolution of the fMRI data can improve overall statistical sensitivity”
From Functional connectomics from resting-state fMRI (2013) · cites this paper as [41]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Advances in functional and structural MR image analysis and implementation as FSL
Smith, S. M., Jenkinson, M., Woolrich, M. W., Beckmann, C. F., Behrens, T. E. J., Johansen-Berg, H., Bannister, P. R., De Luca, M., Drobnjak, I., Flitney, D. E., et al. (2004) · NeuroImage · doi:10.1016/j.neuroimage.2004.07.051 · Cited by 13,630 · cited 2×
“Analysis was carried out using FSL (FMRIB’s Software Library) [51,52].”
MethodSelf-citationFast robust automated brain extraction
Smith, S. M. (2002) · Human Brain Mapping · doi:10.1002/hbm.10062 · Cited by 10,496 · cited 1×
“The mean (over time) image was brain-extracted using BET [54].”
MethodSelf-citationFunctional connectivity in the motor cortex of resting human brain using echo‐planar mri
Biswal, B., Zerrin Yetkin, F., Haughton, V. M., & Hyde, J. S. (1995) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910340409 · Cited by 9,812
Independent component analysis, A new concept?
Comon, P. (1994) · Signal Processing · doi:10.1016/0165-1684(94)90029-9 · Cited by 8,205 · cited 2×
“Multi-acquisition RSN analysis RSNs and structured artifacts in the data were identified using MELODIC (Multivariate Exploratory Linear Optimized Decomposition into Independent Components [55], FSL’s implementation of ICA (Independent Component Analysis [56]).”
A global optimisation method for robust affine registration of brain images
Jenkinson, M., & Smith, S. (2001) · Medical Image Analysis · doi:10.1016/s1361-8415(01)00036-6 · Cited by 6,872
Self-citationSENSE: Sensitivity encoding for fast MRI
Pruessmann, K. P., Weiger, M., Scheidegger, M. B., & Boesiger, P. (1999) · Magnetic Resonance in Medicine · doi:10.1002/(sici)1522-2594(199911)42:5<952::aid-mrm16>3.0.co;2-s · Cited by 6,251 · cited 5×
“Echo train shortening achieved by reducing the phase encoding steps by factor R using partial parallel acquisition techniques such as SENSE [11] or GRAPPA [10] and their derivatives, which is referred to as acceleration, or the use of partial Fourier sampling [9] can also be incorporated in this sequence and were empl…”
BackgroundMethodInfluentialGeneralized autocalibrating partially parallel acquisitions (GRAPPA)
Griswold, M. A., Jakob, P. M., Heidemann, R. M., Nittka, M., Jellus, V., Wang, J., Kiefer, B., & Haase, A. (2002) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10171 · Cited by 5,303 · cited 5×
“Echo train shortening achieved by reducing the phase encoding steps by factor R using partial parallel acquisition techniques such as SENSE [11] or GRAPPA [10] and their derivatives, which is referred to as acceleration, or the use of partial Fourier sampling [9] can also be incorporated in this sequence and were empl…”
BackgroundMethodInfluentialCorrespondence of the brain's functional architecture during activation and rest
Smith, S. M., Fox, P. T., Miller, K. L., Glahn, D. C., Fox, P. M., Mackay, C. E., Filippini, N., Watkins, K. E., Toro, R., Laird, A. R., et al. (2009) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.0905267106 · Cited by 5,150 · cited 1×
“In addition to carrying out the high-dimensional analyses, we also used a set of 10 spatial maps from another study ([33]; 10 RSNs from 36 subjects) as the dual regression input maps.”
MethodSelf-citationDynamic magnetic resonance imaging of human brain activity during primary sensory stimulation.
Kwong, K. K., Belliveau, J. W., Chesler, D. A., Goldberg, I. E., Weisskoff, R. M., Poncelet, B. P., Kennedy, D. N., Hoppel, B. E., Cohen, M. S., & Turner, R. (1992) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.89.12.5675 · Cited by 4,212 · cited 1×
“[1,2] and diffusion weighted imaging for neuronal fiber tracto-”
BackgroundOgawa, S., Tank, D. W., Menon, R., Ellermann, J. M., Kim, S. G., Merkle, H., & Ugurbil, K. (1992) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.89.13.5951 · Cited by 3,681 · cited 1×
“[1,2] and diffusion weighted imaging for neuronal fiber tracto-”
Background
Cited by
Items 801–900 of 1,000
Lim, P. C., Ward, E. J., Vickery, T. J., & Johnson, M. R. (2019) · Journal of Cognitive Neuroscience · doi:10.1162/jocn_a_01427 · Cited by 13
“Functional scans used a multiband EPI sequence (Feinberg et al., 2010; Moeller et al., 2010) with repetition time (TR) = 1000 msec, echo time = 30 msec, 100 × 100 in-plane resolution, 60 axial slices with a thickness of 2.”
MethodYun, S. D., Weidner, R., Weiss, P. H., & Shah, N. J. (2019) · Scientific Reports · doi:10.1038/s41598-019-47341-y · Cited by 15
Kark, S. M., Slotnick, S. D., & Kensinger, E. A. (2019) · Neuropsychologia · doi:10.1016/j.neuropsychologia.2019.107277 · Cited by 12
Heads in the Cloud: A Primer on Neuroimaging Applications of High Performance Computing
Shatil, A. S., Younas, S., Pourreza, H., & Figley, C. R. (2015) · Magnetic Resonance Insights · doi:10.4137/mri.s23558 · Cited by 14
“However, with the advent of multi-band (aka, simultaneous multi-slice ) acquisition techniques such as simultaneous echo-refocused EPI 12 or controlled aliasing techniques such us CAIPIRINHA 13,14 , it is now possible to acquire whole-brain fMRI data with temporal resolutions below 1 s/volume—ie,…”
Saleem, K. S., Avram, A. V., Yen, C. C., Magdoom, K. N., Schram, V., & Basser, P. J. (2023) · NeuroImage · doi:10.1016/j.neuroimage.2023.120311 · Cited by 10
Kumar, S., Shovon, A. R., & Deshpande, G. (2023) · Human Brain Mapping · doi:10.1002/hbm.26403 · Cited by 11
“Resting-state fMRI data were acquired using multiband echo-planar imaging (EPI; Feinberg et al., 2010) from each subject using three different acquisition protocols with different parameters as follows: (1) 3.0 mm isotropic voxels with 40 slices, TR = 645 ms, TE = 30 ms, FOV = 222 222 mm, number of…”
MethodInfluentialThe Subcortical Atlas of the Marmoset (“SAM”) monkey based on high-resolution MRI and histology
Saleem, K. S., Avram, A. V., Glen, D., Schram, V., & Basser, P. J. (2024) · Cerebral Cortex · doi:10.1093/cercor/bhae120 · Cited by 13
Inhibition of the inferior parietal lobe triggers state-dependent network adaptations
Williams, K. A., Numssen, O., Guerra, J. D., Kopál, J., Bzdok, D., & Hartwigsen, G. (2024) · Heliyon · doi:10.1016/j.heliyon.2024.e39735 · Cited by 10
Functional spectroscopy to no-gradient fMRI
Hennig, J. (2011) · NeuroImage · doi:10.1016/j.neuroimage.2011.09.060 · Cited by 10
“An example is David Feinberg's multiplexed SIR-EPI, which achieves a 100 ms temporal resolution based on multiband excitation of the whole multislice package in a single scan (Feinberg et al., 2010).”
MethodMultiple‐frequency excitation wideband MRI (ME‐WMRI)
Wu, E. L., Chiueh, T., & Chen, J. (2014) · Medical Physics · doi:10.1118/1.4893502 · Cited by 10
Multiband echo‐shifted echo planar imaging
Boyacıoğlu, R., Schulz, J., & Norris, D. G. (2016) · Magnetic Resonance in Medicine · doi:10.1002/mrm.26289 · Cited by 10
“The increases in z scores obtained with MESH compared with MB are in line with those previously reported for MB compared with standard EPI (2).”
BackgroundResultImproving the signal detection accuracy of functional Magnetic Resonance Imaging
Janssen, N., Hernandez-Cabrera, J. A., & Ezama, L. (2018) · NeuroImage · doi:10.1016/j.neuroimage.2018.01.076 · Cited by 9
Altered neural responses to social fairness in bipolar disorder
Lois, G., Schneider, E. E., Kaurin, A., & Wessa, M. (2020) · NeuroImage Clinical · doi:10.1016/j.nicl.2020.102487 · Cited by 11
“5 mm3, 60 slices, MB acceleration factor=4) was used for blood oxygen-level dependent (BOLD) fMRI to achieve a fine-grained temporal resolution (29).”
BackgroundMethodThe neural substrate of noun morphological inflection: A rapid event-related fMRI study in Italian
Russo, A. G., Esposito, F., Laudanna, A., Mancuso, A., Salle, F. D., Elia, A., & Martino, M. D. (2020) · Neuropsychologia · doi:10.1016/j.neuropsychologia.2020.107699 · Cited by 10
In vivo Estimation of Dispersion Anisotropy of Neurites Using Diffusion MRI
Tariq, M., Schneider, T., Alexander, D. C., Wheeler‐Kingshott, C. A. M. G., & Zhang, H. G. (2014) · Lecture notes in computer science · doi:10.1007/978-3-319-10443-0_31 · Cited by 11
“This is not yet clinically feasible with the existing imaging sequences, but emerging technologies, such as multi-band imaging [17], will make it possible to acquire substantially more data per unit time.”
BackgroundFord, J. H., Kim, S. Y., Kark, S. M., Daley, R. T., Payne, J. D., & Kensinger, E. A. (2022) · European Journal of Neuroscience · doi:10.1111/ejn.15777 · Cited by 12
“Resonance Research at University of Minnesota (Feinberg et al., 2010; Moeller et al., 2010; Xu et al., 2013), and were acquired in an interleaved fashion covering the whole brain, with the slices aligned 25 above the anterior commissure–posterior commissure line in the coronal direction.”
MethodEvaluation of fiber tracking from subsampled q-space data in diffusion spectrum imaging
Tefera, G. B., Zhou, Y., Juneja, V., & Narayana, P. A. (2013) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2013.02.006 · Cited by 10
“It is possible to reduce the scan time by multiplexed echo sampling [12, 13], but this method is not commonly available.”
BackgroundImproved autoregressive model for correction of noise serial correlation in fast fMRI
Luo, Q., Misaki, M., Mulyana, B. E., Wong, C. K., & Bodurka, J. (2020) · Magnetic Resonance in Medicine · doi:10.1002/mrm.28203 · Cited by 13
Integrated and segregated frequency architecture of the human brain network
Ma, J., Lin, Y., Hu, C., Zhang, J., Yi, Y., & Dai, Z. (2021) · Brain Structure and Function · doi:10.1007/s00429-020-02174-8 · Cited by 11
Emmenegger, T. M., Dávid, G., Ashtarayeh, M., Fritz, F. J., Ellerbrock, I., Helms, G., Balteau, E., Freund, P., & Mohammadi, S. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.674719 · Cited by 10
“To accelerate the data acquisition, GRAPPA (inplane acceleration with factor two) and simultaneous multi-slice acquisitions (“multiband,” slice acceleration factor two) (Feinberg et al., 2010; Moeller et al., 2010; Xu et al., 2013) were used as described in Setsompop et al. (2012).”
MethodSimultaneous multislice magnetic resonance fingerprinting with low-rank and subspace modeling
Zhao, B., Bilgiç, B., Adalsteinsson, E., Griswold, M. A., Wald, L. L., & Setsompop, K. (2017) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2017.8037553 · Cited by 13
“Simultaneous multislice (SMS) imaging [3–6] is a powerful data acquisition methodology, which enables rapid volumetric imaging through acquiring data from multiple imaging slices simultaneously rather than sequentially.”
BackgroundKohler, P. J., Cottereau, B. R., & Norcia, A. M. (2019) · Scientific Reports · doi:10.1038/s41598-019-45036-y · Cited by 12
How pushing the spatiotemporal resolution of fMRI can advance neuroscience
Vizioli, L., Yacoub, E. S., & Lewis, L. D. (2021) · Progress in Neurobiology · doi:10.1016/j.pneurobio.2021.102184 · Cited by 11
Self-citationA systematically optimized awake mouse fMRI paradigmPreprint
Xu, W., Pei, M., Zhang, K., Tong, C., Bo, B., Feng, J., Zhang, X., & Liang, Z. (2022) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2022.11.16.516376 · Cited by 12
“Moreover, test statistics and temporal signal-to-noise ratio (tSNR) may benefit from increase of the effective sample size (Feinberg et al., 2010; Jahanian et al., 2019; Todd et al., 2016), leading to more significant activation or functional connectivity in task (Boyacioğlu et al., 2015; Chen and…”
BackgroundClark, I. A., & Maguire, E. A. (2023) · Scientific Data · doi:10.1038/s41597-023-02449-9 · Cited by 10
HCP Multi-Pipeline: a derived dataset to investigate analytical variability in fMRI
Germani, É., Fromont, É., Maurel, P., & Maumet, C. (2025) · Scientific Data · doi:10.1038/s41597-025-05247-7 · Cited by 12
Gage, A. T., Stone, J. R., Wilde, E. A., McCauley, S. R., Welsh, R. C., Mugler, J. P., Tustison, N., Avants, B., Whitlow, C. T., Lancashire, L., et al. (2024) · Journal of Neurotrauma · doi:10.1089/neu.2024.0128 · Cited by 9
Highly accelerated EPI with wave encoding and multi‐shot simultaneous multislice imaging
Cho, J., Liao, C., Tian, Q., Zhang, Z., Xu, J., Lo, W., Poser, B. A., Stenger, V. A., Stockmann, J., Setsompop, K., et al. (2022) · Magnetic Resonance in Medicine · doi:10.1002/mrm.29291 · Cited by 9
Geometric deep learning for diffusion MRI signal reconstruction with continuous samplings (DISCUS)
Ewert, C., Kügler, D., Stirnberg, R., Koch, A., Yendiki, A., & Reuter, M. (2024) · Imaging Neuroscience · doi:10.1162/imag_a_00121 · Cited by 13
“…weightings ( b - values) of 1,000, 2,000, and 3,000 s/mm 2 with 90 DWIs per b - value and participant on a modified 3 Tesla Siemens MAGNETOM Skyra scanner ( Feinberg et al., 2010; Moeller et al., 2010; Setsompop et al., 2012; Sotiropoulos et al., 2013; van Essen et al., 2012; Xu et al., 2012).”
MethodNarsude, M., van der Zwaag, W., Kober, T., Rolf Gruetter, A., & Marques, J. P. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24975 · Cited by 9
“The improvement in detection of RSNs with faster imaging by means of increased temporal resolution and tSNR was demonstrated in the multiplexed 2D EPI (23) at 3T.”
MethodTan, F. M., Caballero‐Gaudes, C., Mullinger, K. J., Cho, S., Zhang, Y., Dryden, I. L., Francis, S., & Gowland, P. A. (2017) · Human Brain Mapping · doi:10.1002/hbm.23767 · Cited by 10
“the temporal blurring of the hemodynamic response, we predict that increased temporal resolution, for instance, using simultaneous multislice imaging [Feinberg et al., 2010; Moeller et al., 2010], would help to differentiate between individual movements within blocks, in addition to enhancing the performance of the SP…”
MethodFalco, D., Chowdury, A., Rosenberg, D. R., Bressler, S. L., & Diwadkar, V. A. (2020) · Brain Structure and Function · doi:10.1007/s00429-020-02061-2 · Cited by 10
Talk to us! Communication is a key factor in improving the comfort of MRI research participants
Dewey, R. S., Ward, C., Junor, A., & Horobin, A. J. (2021) · Health Expectations · doi:10.1111/hex.13217 · Cited by 12
Anderson, R. J., Poser, B. A., & Stenger, V. A. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25050 · Cited by 10
Beyond Functional Localization
Sutherland, M. T., Liang, X., Yang, Y., & Stein, E. A. (2015) · Wiley eBooks · doi:10.1002/9781118472415.ch20 · Cited by 10
A Rotational Cylindrical fMRI Phantom for Image Quality Control
Tovar, D. A., Zhan, W., & Rajan, S. S. (2015) · PLoS ONE · doi:10.1371/journal.pone.0143172 · Cited by 9
“For example, multiband-EPI pulse sequence which allows higher temporal or spatial resolution has been recently reported [36].”
BackgroundAn M-estimator for reduced-rank system identification
Chen, S., Liu, K., Yang, Y., Xu, Y., Lee, S., Lindquist, M. A., Caffo, B. S., & Vogelstein, J. T. (2016) · Pattern Recognition Letters · doi:10.1016/j.patrec.2016.12.012 · Cited by 9
“The Human Connectome Project (HCP) is a systematic effort to map macroscopic human brain circuits and their relationship to behavior in a large population of healthy adults [9,18,26].”
BackgroundBilateral Multiband 4D Flow MRI of the Carotid Arteries at 7T
Schmitter, S., Adriany, G., Waks, M., Moeller, S., Aristova, M., Vali, A., Auerbach, E. J., de Moortele, P. V., Uǧurbil, K., & Schnell, S. (2020) · Magnetic Resonance in Medicine · doi:10.1002/mrm.28256 · Cited by 9
Self-citationNeural Correlates of Food Cue Exposure Intervention for Obesity: A Case-Series Approach
Franssen, S., Jansen, A. T. M., Schyns, G., van den Akker, K., & Roefs, A. (2020) · Frontiers in Behavioral Neuroscience · doi:10.3389/fnbeh.2020.00046 · Cited by 9
Chauvin, R. J., Buitelaar, J. K., Sprooten, E., Oldehinkel, M., Franke, B., Hartman, C. A., Heslenfeld, D. J., Hoekstra, P. J., Oosterlaan, J., Beckmann, C. F., et al. (2021) · Translational Psychiatry · doi:10.1038/s41398-021-01284-z · Cited by 13
Reducing Susceptibility Distortion Related Image Blurring in Diffusion MRI EPI Data
Clark, I. A., Callaghan, M. F., Weiskopf, N., Maguire, E. A., & Mohammadi, S. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.706473 · Cited by 9
“The protocol used was the Multi-Band Accelerated EPI Pulse Sequence developed by the Centre for Magnetic Resonance Research at the University of Minnesota (R012a-c, R013a on VB17, https://www.cmrr.umn.edu/multiband/; Feinberg et al., 2010; Xu et al., 2013).”
MethodShou, Q., Shao, X., & Wang, D. J. J. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.737525 · Cited by 9
Zhang, Y., Lemarchand, R., Asyraff, A., & Hoffman, P. (2022) · NeuroImage · doi:10.1016/j.neuroimage.2022.119450 · Cited by 9
MRI Methods for In-Vivo Cortical Parcellation
Turner, R. (2013) · Springer eBooks · doi:10.1007/978-3-642-37824-9_7 · Cited by 13
Teruel, J. R., Kuperman, J. M., Dale, A. M., & White, N. S. (2018) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.25953 · Cited by 9
Meyer, K., Garzón, B., Lövdén, M., & Hildebrandt, A. (2019) · Royal Society Open Science · doi:10.1098/rsos.180857 · Cited by 10
Shearrer, G. E., Sadler, J. R., Papantoni, A., & Burger, K. S. (2020) · Journal of Neuroendocrinology · doi:10.1111/jne.12891 · Cited by 10
Noninvasive Functional and Anatomical Imaging of the Human Medial Temporal Lobe
Brown, T. I., Staresina, B. P., & Wagner, A. D. (2015) · Cold Spring Harbor Perspectives in Biology · doi:10.1101/cshperspect.a021840 · Cited by 11
Guell, X., Schmahmann, J. D., & Gabrieli, J. D. E. (2018) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/424176 · Cited by 10
“EPI data acquired by the WU-Minn HCP used multi-band pulse sequences (Feinberg et al. 2010, Moeller et al. 2010, Setsompop et al. 2012, Xu et al. 2013).”
MethodValue shapes the structure of schematic representations in the medial prefrontal cortexPreprint
Paulus, P. C., Charest, I., & Benoit, R. G. (2020) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2020.08.21.260950 · Cited by 9
“For each of the two functional runs of the simulation task, we acquired 469 volumes of blood-oxygen-level-dependent (BOLD) data with a T2*weighted echo-planar imaging (EPI) pulse sequence (89, 90).”
MethodA Transdiagnostic fMRI Dataset With 300+ Deeply Phenotyped Subjects Across Resting and Task StatesPreprint
Samardzija, A., Shen, X., Greene, A. S., Mehta, S., Iannone, S., Simon, A. J., Parra, F., Luo, W., Arora, J., Tokoglu, F., et al. (2025) · medRxiv · doi:10.1101/2025.07.15.25331595 · Cited by 8
Himmelberg, M. M., Kwak, Y., Carrasco, M., & Winawer, J. A. (2025) · PLoS Computational Biology · doi:10.1371/journal.pcbi.1013599 · Cited by 12
“The slice-time corrected functional data were then resampled to the T1w anatomical space via a one-shot interpolation consisting of all the pertinent transformations (i.e., head motion transform matrices, susceptibility distortion correction).”
Reduced gray matter volume in the default-mode network associated with insulin resistance
Asano, S., Ogawa, A., Osada, T., Oka, S., Nakajima, K., Oshima, Y., Tanaka, S., Kaga, H., Tamura, Y., Watada, H., et al. (2023) · Cerebral Cortex · doi:10.1093/cercor/bhad358 · Cited by 10
Ramage, A. E., Cote, K., Thorson, J. C., Lerner, K., Reidel, M. C., & Laird, A. R. (2024) · Imaging Neuroscience · doi:10.1162/imag_a_00355 · Cited by 8
“Parameters were structural: 0.7-mm isotropic resolution; rsfMRI: TR = 720 ms, TE = 33.1 ms, in-plane FOV = 208 x 180 mm, 72 slices, 2.0 mm isotropic voxels, and multiband acceleration factor = 8 ( Feinberg et al., 2010).”
InfluentialBudak, M., Fausto, B. A., Osiecka, Z., Sheikh, M., Perna, R., Ashton, N. J., Blennow, K., Zetterberg, H., Fitzgerald-Bocarsly, P. A., & Gluck, M. A. (2024) · Alzheimer s Research & Therapy · doi:10.1186/s13195-024-01619-0 · Cited by 9
Sönmez, Ö., Holstein, E. R., Puschmann, S., Schmitt, T., Witt, K., & Thiel, C. M. (2025) · Psychophysiology · doi:10.1111/psyp.14739 · Cited by 8
“Resting-state and task-based data were acquired using a 2D multiband T 2 *- weighted gradient echo planar imaging EPI sequence with BOLD contrast, from the Center for Magnetic Resonance Research (CMRR), University of Minnesota (Feinberg et al., 2010; Moeller et al., 2010; Xu et al., 2013) (TR = 850 ms, TE = 30 ms, fli…”
InfluentialCatalogna, M., Somerville, Y., Saporta, N., Nathansohn-Levi, B., Shelly, S., Edry, L., Zagoory‐Sharon, O., Feldman, R., & Amedi, A. (2025) · Scientific Reports · doi:10.1038/s41598-025-91457-3 · Cited by 9
High‐density MRI coil arrays with integrated field monitoring systems for human connectome mapping
Mahmutović, M., Shrestha, M., Ramos‐Llordén, G., Sung, D., Edwards, L., Chu, Y., Dubovan, P. I., Müller, A., Hansen, S. J. D., Ghotra, A., et al. (2025) · Magnetic Resonance in Medicine · doi:10.1002/mrm.30606 · Cited by 8
“17 This distribution of encoding responsibilities allows for highly accelerated in-plane 18,19 and multi-slice 20–22 parallel imaging acquisition techniques, including advanced volumetric thin-slab acquisition methods such as gSlider-SMS, 23 with minimal noise amplification during image…”
RF peak power reduction in CAIPIRINHA excitation by interslice phase optimization
Sbrizzi, A., Poser, B. A., Tse, D. H. Y., Hoogduin, H. M., Luijten, P. R., & van den Berg, C. A. T. (2015) · NMR in Biomedicine · doi:10.1002/nbm.3360 · Cited by 8
“This improvement is related to the smooth form of the objective function and constraints in Equations A9–A11, derived by introducing the auxiliary variable t, which makes it possible to cast the problem as a linear program with quadratic constraints.”
BackgroundMethodInfluentialDuff, E. P., Makin, T. R., Madugula, S. S., Smith, S. M., & Woolrich, M. W. (2013) · International Workshop on Pattern Recognition in NeuroImaging · doi:10.1109/prni.2013.24 · Cited by 8
“3s) multiband (factor 6) whole-brain acquisition [5], [6].”
Self-citationHughes, M. E., Nkyekyer, J., Innes-Brown, H., Rossell, S. L., Sly, D., Bhar, S. S., Pipingas, A., Hennessy, A., & Meyer, D. (2018) · JMIR Research Protocols · doi:10.2196/resprot.9916 · Cited by 8
“In addition, multiband acquisition sequences will be derived from the Human Connectome Project [58].”
MethodJang, W., Song, J. S., Kwak, H. S., Hwang, S. B., & Paek, M. Y. (2019) · Abdominal Radiology · doi:10.1007/s00261-019-01924-5 · Cited by 10
Effective connectivity modulations related to win and loss outcomes
de Steen, F. V., Krebs, R. M., Colenbier, N., Almgren, H., & Marinazzo, D. (2019) · NeuroImage · doi:10.1016/j.neuroimage.2019.116369 · Cited by 8
Semantic fMRI neurofeedback: a multi-subject study at 3 tesla
Ciarlo, A., Russo, A. G., Ponticorvo, S., Salle, F. D., Lührs, M., Goebel, R., & Esposito, F. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac6f81 · Cited by 8
“…256, voxel size = 1.0 × 1.0 × 1.0 mm 3 ) and a BOLD-fMRI multi-band accelerated echo-planar imaging (EPI) sequence (TR = 1000 ms, TE = 30 ms, number of axial slices = 48, matrix size = 96 × 96, FOV = 288 mm, slice thickness = 3 mm, interslice gap = 0 mm, multi-band accelerator factor = 4) [37–39].”
InfluentialIntrinsic Brain Activity and Resting State Networks
Snyder, A. Z. (2022) · Springer eBooks · doi:10.1007/978-3-030-88832-9_133 · Cited by 8
Jackson, R. L., Humphreys, G. F., Rice, G. E., Binney, R. J., & Ralph, M. A. L. (2023) · Cortex · doi:10.1016/j.cortex.2022.12.016 · Cited by 8
Brain-wide mapping of resting-state networks in mice using high-frame rate functional ultrasound
Hikishima, K., Tsurugizawa, T., Kasahara, K., Takagi, R., Yoshinaka, K., & Nitta, N. (2023) · NeuroImage · doi:10.1016/j.neuroimage.2023.120297 · Cited by 9
Spatially localized fMRI metrics as predictive and highly distinct state-independent fingerprints
Larabi, D. I., Gell, M., Amico, E., Eickhoff, S. B., & Patil, K. R. (2026) · Imaging Neuroscience · doi:10.1162/imag.a.1444 · Cited by 10
Motion correction for diffusion weighted SMS imaging
Herbst, M., Poser, B. A., Singh, A., Deng, W., Knowles, B. R., Zaitsev, M., Stenger, V. A., & Ernst, T. (2016) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2016.12.013 · Cited by 8
“In contrast, simultaneous multislice imaging (SMS) makes it possible to acquire several slices simultaneously [16]–[19] and can considerably shorten volume acquisition times for DTI protocols.”
A consistent organizational structure across multiple functional subnetworks of the human brain
Stillman, P. E., Wilson, J. D., Denny, M. J., Desmarais, B. A., Cranmer, S., & Lu, Z. (2019) · NeuroImage · doi:10.1016/j.neuroimage.2019.03.036 · Cited by 8
fMRIflows: A Consortium of Fully Automatic Univariate and Multivariate fMRI Processing Pipelines
Notter, M. P., Herholz, P., Costa, S. D., Gülban, Ö. F., Isik, A. I., Gaglianese, A., & Murray, M. M. (2022) · Brain Topography · doi:10.1007/s10548-022-00935-8 · Cited by 8
Measuring cerebral hemodynamics with a modified magnetoencephalography system
Burke, B. A., & Diamond, S. G. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/12/2079 · Cited by 9
Demonstration and validation of a new pressure-based MRI-safe pain tolerance device
Davis, M. T., Daniel, T. A., Witte, T. K., Beyers, R. J., Willis, J. Z., Wang, Y., Denney, T. S., Katz, J. S., Salibi, N., & Deshpande, G. (2016) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2016.07.001 · Cited by 8
Peechatka, A. L., & Janes, A. C. (2016) · Nicotine & Tobacco Research · doi:10.1093/ntr/ntw252 · Cited by 8
Caldeira, L., Yun, S. D., da Silva, N. A., Filß, C., & Shah, N. J. (2019) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.26639 · Cited by 8
“One of the possible choices for improving the resolution along the slice direction is to use the multiband scheme.(28,29) This method essentially reduces scanning time for one volume by exciting multiple slices with the use of sinusoidal modulation for a single-slice selective RF pulse.”
MethodInteraction Between Cerebellum and Cerebral Cortex, Evidence from Dynamic Causal Modeling
Bukhari, Q., Ruf, S. F., Guell, X., Whitfield, S., & Anteraper, S. A. (2021) · The Cerebellum · doi:10.1007/s12311-021-01284-1 · Cited by 10
Shashidhara, S., & Erez, Y. (2021) · Neuropsychologia · doi:10.1016/j.neuropsychologia.2021.107981 · Cited by 11
Matyi, M. A., Cioabă, S. M., Banich, M. T., & Spielberg, J. M. (2021) · NeuroImage · doi:10.1016/j.neuroimage.2021.118614 · Cited by 8
Skipper, J. I., & Hasson, U. (2017) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/139550 · Cited by 9
“Gradient-echo EPI was used to collect 1200 images per run, acquired using multiband accelerated imaging (D. A. Feinberg et al., 2010; Moeller, Yacoub, & Olman, 2010; Setsompop et al., 2012; Xu, Moeller, Strupp, Auerbach, & Chen, 2012)(TR = 720 ms; TE = 33.1 ms; Flip angle = 52°; FOV 208 × 180 mm;…”
InfluentialTemporal transitions of spontaneous brain activityPreprint
Ma, Z., & Zhang, N. (2017) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/166512 · Cited by 8
Catalogna, M., Saporta, N., Nathansohn-Levi, B., Tamir, T., Shahaf, A., Molcho, S., Erlich, S., Shelly, S., & Amedi, A. (2025) · npj Digital Medicine · doi:10.1038/s41746-025-01765-1 · Cited by 10
“The MRI protocol included the following sequences: Two runs of 300 volumes (9:28 min) resting state fMRI scans were acquired using a multi-band echo planar imaging sequence, CMRR EPI 2D 114,115 .”
Laminar functional magnetic resonance imaging in vision research
Demirayak, P., Deshpande, G., & Visscher, K. (2022) · Frontiers in Neuroscience · doi:10.3389/fnins.2022.910443 · Cited by 8
“Slicetiming correction is often critical for multi-band acquisitions that are used to increase temporal resolution or spatial coverage (Feinberg et al., 2010; Feinberg and Setsompop, 2013; Barth et al., 2016).”
BackgroundParallel processing relies on a distributed, low-dimensional cortico-cerebellar architecture
Müller, E. J., Hou, K. Y., Palesi, F., Tan, J. B., Close, T., Wheeler‐Kingshott, C. A. M. G., D'Angelo, E. U., Calamante, F., & Shine, J. M. (2023) · Network Neuroscience · doi:10.1162/netn_a_00308 · Cited by 8
“For functional scans, a T2*-weighted multiband gradient echo-planar imaging sequence was used (TR = 700 msec, TE = 30 msec, flip angle = 55°, 48 axial slices, slice thickness = 3 mm, no gap, in-plane resolution 3 × 3 mm) (Feinberg et al., 2010).”
MethodSudo, F. K., Pinto, T. P., Barros‐Aragão, F. G. Q., Bramati, I. E., Marins, T. F., Monteiro, M., Meireles, F., Soares, R., Erthal, P., Calil, V., et al. (2023) · Brain Behavior and Immunity · doi:10.1016/j.bbi.2023.10.020 · Cited by 10
Fei, W., Sun, Y.-N., Zhang, B.-T., Yang, Q., He, A.-D., Xu, W., Liu, J., Liu, M.-X., Li, X., Yu, Y., et al. (2024) · BMC Medical Imaging · doi:10.1186/s12880-024-01368-4 · Cited by 8
“SMS, predominantly during single-shot echo planar imaging DWI (SSEPI-DWI), has been utilized in assessments of the brain [21–23], abdomen [24–26], and breast glands [27–30].”
MethodGuerrero, S. L., Mesite, L. M., & Luk, G. (2024) · Scientific Reports · doi:10.1038/s41598-024-69575-1 · Cited by 8
Hou, T., Ding, W., Huang, J., Jiang, S., Yao, H., Zhou, T., & Ju, H. (2024) · Applied Soft Computing · doi:10.1016/j.asoc.2024.112538 · Cited by 9
Gann, M. A., Paparella, I., Zich, C., Grigoras, I., Huertas‐Penen, S., Rieger, S. W., Thielscher, A., Sharott, A., Stagg, C. J., & Schwab, B. C. (2025) · Brain stimulation · doi:10.1016/j.brs.2025.08.011 · Cited by 10
Comparing the effect of multi-gradient echo and multi-band fMRI during a semantic task
Halai, A. D., Henson, R. N., Finoia, P., & Correia, M. (2025) · Imaging Neuroscience · doi:10.1162/imag.a.1043 · Cited by 8
“Multi-band imaging has been shown to reduce temporal aliasing of high frequency noise sources, increase statistical power, and improve resting-state network estimation/reliability ( Feinberg et al., 2010; Griffanti et al., 2014; Smith et al., 2013).”
Middlebrooks, E. H., Yağmurlu, K., Bennett, J., & Bidari, S. (2014) · Surgical and Radiologic Anatomy · doi:10.1007/s00276-014-1387-2 · Cited by 7
“Diffusion data utilized a spin-echo EPI sequence with multiband EPI [5, 12, 15, 24].”
MethodChang, W., Setsompop, K., Ahveninen, J., Belliveau, J. W., Witzel, T., & Lin, F. (2014) · NeuroImage · doi:10.1016/j.neuroimage.2013.12.037 · Cited by 7
“In addition, higher temporal resolution of fMRI has been suggested to be beneficial for improving the power of detecting neuronally related timing information and connectivity among brain areas (Deshpande et al., 2010; Feinberg et al., 2010; Kayser et al., 2009; Lee et al., 2013; Lin et al., 2013; Roebroeck et al., 20…”
Deconvolution filtering: Temporal smoothing revisited
Bush, K., & Cisler, J. M. (2014) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2014.03.002 · Cited by 7
“technology [28] or alterations to the scan sequence.”
BackgroundZhang, J., Liu, S., Dai, E., Ye, X., Shi, D., Wu, Y.-H., Lu, J., & Guo, H. (2021) · Magnetic Resonance in Medicine · doi:10.1002/mrm.29047 · Cited by 9
The Role of the Left Inferior Frontal Gyrus in Introspection during Verbal Communication
Yoshioka, A., Tanabe, H. C., Nakagawa, E., Sumiya, M., Koike, T., & Sadato, N. (2023) · Brain Sciences · doi:10.3390/brainsci13010111 · Cited by 9
Neural representation of gestalt grouping and attention effect in human visual cortex
Wu, H., Zuo, Z., Yuan, Z., Zhou, T., Zhuo, Y., Zheng, N., & Chen, B. (2023) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2023.109980 · Cited by 7
Cheng, H., & Puce, A. (2014) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2014.02.007 · Cited by 8
Impaired Visual Cortical Processing of Affective Facial Information in Schizophrenia
Maher, S. M., Ekstrom, T., & Chen, Y. (2015) · Clinical Psychological Science · doi:10.1177/2167702615609595 · Cited by 7
Functional MRI and Sensory Perception of Food
Bonny, J.-M., Sinding, C., & Thomas‐Danguin, T. (2017) · Springer eBooks · doi:10.1007/978-3-319-28275-6_132-1 · Cited by 10
Goldhacker, M., Tomé, A. M., Greenlee, M. W., & Lang, E. W. (2018) · Frontiers in Human Neuroscience · doi:10.3389/fnhum.2018.00253 · Cited by 8
Coil-combined split slice-GRAPPA for simultaneous multi-slice diffusion MRI
HashemizadehKolowri, S., Chen, R., Adluru, G., Ying, L. L., & DiBella, E. (2019) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2019.11.017 · Cited by 7
Metadata
- Authors
- Feinberg, D. A., Moeller, S., Smith, S. M., Auerbach, E., Ramanna, S., Glasser, M. F., Miller, K. L., Ugurbil, K., & Yacoub, E.
- Affiliations
- University of California, San Francisco (US)
- University of Minnesota (US)
- John Radcliffe Hospital (GB)
- Advanced MRI Technologies (United States) (US)
- Washington University in St. Louis (US)
- Corresponding author
- Feinberg, David A.
- Venue
- PLoS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2010-12-20
- Volume
- 5
- Issue
- 12
- Pages
- e15710
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC3004955
- OpenAlex
- W2114809469
- Cited by
- 1,296
- References
- 57
- Citations by year
- 2026: 51 · 2025: 73 · 2024: 53 · 2023: 63 · 2022: 88 · 2021: 104 · 2020: 94 · 2019: 94 · 2018: 107 · 2017: 117 · 2016: 98 · 2015: 102 · 2014: 75 · 2013: 99 · 2012: 48
- FWCI
- 44.96
- Type
- journal article
- Language
- en
- Update notices
- correction (2011-09-28)
- Open access
- gold
- doi.org
- Locations
- PLoS ONE (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Digital Commons@Becker (Washington University School of Medicine) (submittedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Figshare (submittedVersion, cc-by)
- Funders
- National Science Foundation (P30 NS057091, 1U54MH091657, 9907842)
- W. M. Keck Foundation
- National Institutes of Health (5u54mh091657-05, S10 RR1395, 1u54mh091657-01, 1U54‐MH‐091657, RR08079)
- National Institute of Mental Health (1U54-MH091657-01, U54-MH091657)
- National Institute of Neurological Disorders and Stroke (R44NS063537, R44NS073417, P30NS057091)
- National Institute of Biomedical Imaging and Bioengineering (R01EB000331)
- National Center for Research Resources (S10-RR1395, P41RR001395, P41RR008079, P41RR08079)
- NIH Blueprint for Neuroscience Research (1U54-MH-091657)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced Neuroimaging Techniques and Applications
- Life Sciences › Neuroscience › Cognitive Neuroscience › Functional Brain Connectivity Studies
- Keywords
- echo planar imaging, whole-brain fMRI, scan time reduction, spatial multiplexing, time-division multiplexing, high angular resolution diffusion imaging, diffusion MRI, fat suppression, signal-to-noise ratio, resting-state networks, independent component analysis, temporal resolution, non-stationarity
- MeSH terms
- Algorithms, Brain, Brain Mapping, Diffusion, Humans, Image Processing, Computer-Assisted, Magnetic Resonance Imaging, Models, Biological, Neurons, Regression Analysis, Echo-Planar Imaging, Phantoms, Imaging
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.