Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients
Open accessThe DOI matches its Crossref record; 23 of the 23 references it cites were found in a registry or an index.
Abstract
PURPOSE: Physiologic monitors are plagued with alarms that create a cacophony of sounds and visual alerts causing "alarm fatigue" which creates an unsafe patient environment because a life-threatening event may be missed in this milieu of sensory overload. Using a state-of-the-art technology acquisition infrastructure, all monitor data including 7 ECG leads, all pressure, SpO(2), and respiration waveforms as well as user settings and alarms were stored on 461 adults treated in intensive care units. Using a well-defined alarm annotation protocol, nurse scientists with 95% inter-rater reliability annotated 12,671 arrhythmia alarms. RESULTS: A total of 2,558,760 unique alarms occurred in the 31-day study period: arrhythmia, 1,154,201; parameter, 612,927; technical, 791,632. There were 381,560 audible alarms for an audible alarm burden of 187/bed/day. 88.8% of the 12,671 annotated arrhythmia alarms were false positives. Conditions causing excessive alarms included inappropriate alarm settings, persistent atrial fibrillation, and non-actionable events such as PVC's and brief spikes in ST segments. Low amplitude QRS complexes in some, but not all available ECG leads caused undercounting and false arrhythmia alarms. Wide QRS complexes due to bundle branch block or ventricular pacemaker rhythm caused false alarms. 93% of the 168 true ventricular tachycardia alarms were not sustained long enough to warrant treatment. DISCUSSION: The excessive number of physiologic monitor alarms is a complex interplay of inappropriate user settings, patient conditions, and algorithm deficiencies. Device solutions should focus on use of all available ECG leads to identify non-artifact leads and leads with adequate QRS amplitude. Devices should provide prompts to aide in more appropriate tailoring of alarm settings to individual patients. Atrial fibrillation alarms should be limited to new onset and termination of the arrhythmia and delays for ST-segment and other parameter alarms should be configurable. Because computer devices are more reliable than humans, an opportunity exists to improve physiologic monitoring and reduce alarm fatigue.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures











![Figure 12 True accelerated ventricular rhythm alarm showing why this arrhythmia is not considered an “actionable” alarm condition. Accelerated ventricular rhythm at a rate of 56 for the first 5 beats followed by 2 fusion beats; the last 2 beats are normal sinus rhythm. The invasive arterial pressure waveform (bottom tracing) shows no change between normal rhythm and accelerated ventricular rhythm which confirms the rationale for the published guidelines stating no treatment is indicated for this arrhythmia in hospital settings [5].](https://figures.cochat.ai/paper-figures/440dfa9937e3409b986af268dec77e1a/insights-into-the-problem-of-alarm-fatigue-with-physiologic-monitor-devices-a-figure-12.jpg)




Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“One study conducted a comprehensive observational analysis over 31 days across five ICUs, involving 461 adults and monitoring over 2.5 million alarm instances. Of 1,154,201 unique alarms, 88.8% were for arrhythmias with a majority being false positive98.”
From Bias recognition and mitigation strategies in artificial intelligence healthcare applications (2025) · cites this paper as [98]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The Cardiac Arrhythmia Suppression Trial (CAST) Investigators (1989) · New England Journal of Medicine · doi:10.1056/nejm198908103210629 · Cited by 3,300
Third Universal Definition of Myocardial Infarction
Thygesen, K., Alpert, J. S., Jaffe, A. S., Simoons, M. L., Chaitman, B. R., & White, H. D. (2012) · Circulation · doi:10.1161/cir.0b013e31826e1058 · Cited by 2,985
Third universal definition of myocardial infarction
the Writing Group on behalf of the Joint ESC/ACCF/AHA/WHF Task Force for the Universal Definition of Myocardial Infarction, Thygesen, K., Alpert, J. S., Jaffe, A. S., Simoons, M. L., Chaitman, B. R., & White, H. D. (2012) · Nature Reviews Cardiology · doi:10.1038/nrcardio.2012.122 · Cited by 2,053
Zipes, D. P., Camm, A. J., Borggrefe, M., Buxton, A. E., Chaitman, B., Fromer, M., Gregoratos, G., Klein, G., Myerburg, R. J., Quinones, M. A., et al. (2006) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2006.07.010 · Cited by 1,973
Weir, M. (1990) · Cardiopulmonary Physical Therapy Journal · doi:10.1097/01823246-199001020-00009 · Cited by 1,058
The association for the advancement of medical instrumentation
(1974) · Ultrasound in Medicine & Biology · doi:10.1016/0301-5629(74)90035-0 · Cited by 1,031
Drew, B. J., Ackerman, M. J., Funk, M., Gibler, W. B., Kligfield, P., Menon, V., Philippides, G. J., Roden, D. M., & Zareba, W. (2010) · Circulation · doi:10.1161/circulationaha.109.192704 · Cited by 821
Self-citationAuthors/Task Force Members (2006) · European Heart Journal · doi:10.1093/eurheartj/ehl199 · Cited by 523
ACC/AHA guidelines for ambulatory electrocardiography
Crawford, M. H., Bernstein, S. J., Deedwania, P. C., DiMarco, J. P., Ferrick, K. J., Garson, A., Green, L. A., Greene, H. L., Silka, M. J., Stone, P. H., et al. (1999) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(99)00354-x · Cited by 309
Improving Alarm Performance in the Medical Intensive Care Unit Using Delays and Clinical Context
Görges, M., Markewitz, B. A., & Westenskow, D. R. (2009) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e31819bdfbb · Cited by 211
Cited by
Items 301–400 of 457
A review of infant apnea monitor design
Nwaneri, S. C., Ezenwa, B. N., Osuntoki, A. A., Ezeaka, C., & Ogunsola, F. T. (2024) · Journal of Clinical Sciences · doi:10.4103/jcls.jcls_48_23 · Cited by 2
Pelter, M. M., Prasad, P. A., MORTARA, D. W., & Badilini, F. (2024) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2024.153777 · Cited by 2
Chromik, J., Flint, A. R., Prendke, M., Bert Arnrich, A., & Poncette, A.-S. (2024) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2024.109244 · Cited by 2
False Crisis Alarms in Cardiopulmonary Monitoring:
Wung, S., Ogoo, M., Chow, J. L., & Muzzy, A. (2025) · Critical Care Nursing Clinics of North America · doi:10.1016/j.cnc.2025.02.001 · Cited by 3
Acute Coronary Syndrome ST-Segment Monitoring
Pelter, M. M., Kozik, T. M., Al‐Zaiti, S. S., & Carey, M. G. (2014) · American Journal of Critical Care · doi:10.4037/ajcc2014847 · Cited by 2
Context-Sensitive Clinical Alert Packages Written in Arden Syntax
Zeckl, J., Adlassnig, K., Fossler, R., Blacky, A., de Bruin, J. S., Koller, W., Rappelsberger, A., & Adlassnig, K.-P. (2017) · Studies in health technology and informatics · doi:10.3233/978-1-61499-830-3-1190 · Cited by 2
“Various studies on alerting suggest that by far the large majority of alerts are ignored or overridden [16, 17].”
BackgroundBig Data Revolution in Health Care Sector
Gagana, S., & Paramesha, K. (2018) · doi:10.21467/proceedings.1.44 · Cited by 2
Big Data in Electroencephalography Analysis
Yedurkar, D. P., & Metkar, S. P. (2019) · Studies in big data · doi:10.1007/978-3-030-31672-3_8 · Cited by 2
Au-Yeung, W.-T. M., Sevakula, R. K., Singh, J. P., Heist, E. K., Isselbacher, E. M., & Armoundas, A. A. (2021) · Springer eBooks · doi:10.1007/978-3-030-63355-4_24 · Cited by 2
Putri, S. D., Tahlil, T., & Hasballah, K. (2021) · STRADA JURNAL ILMIAH KESEHATAN · doi:10.30994/sjik.v10i1.698 · Cited by 2
Hravnak, M., Mazzoccoli, A., Bose, E. L., & Pinsky, M. R. (2017) · Springer eBooks · doi:10.1007/978-3-319-39391-9_10 · Cited by 5
Accurate Dysrhythmia Monitoring in Adults
Bourgault, A. M., Seckel, M. A., & Kramlich, D. L. (2016) · Critical Care Nurse · doi:10.4037/ccn2016767 · Cited by 2
Computers in White Coats: How to Devise Useful Clinical Decision Support Software*
Moorman, J. R., Lake, D. E., & Moss, T. J. (2016) · Critical Care Medicine · doi:10.1097/ccm.0000000000001781 · Cited by 2
“Stress Ulcer Prophylaxis?* T he use of stress ulcer prophylaxis is common in the ICU. damage Although it is commonly accepted that a subset of critically ill patients is at risk for stress-related mucosal (1), the prevalence of clinically important bleeding in the ICU is quite low (1–4).”
BackgroundResultMonitor Watchers and Alarm Fatigue: Cautious Optimism
Ruppel, H., & Funk, M. (2017) · Journal of Hospital Medicine · doi:10.12788/jhm.2758 · Cited by 2
Hundal, J., Veettil, A. A. V., Chung, C., & Hosseini, M. (2026) · American Society of Clinical Oncology Educational Book · doi:10.1200/edbk-26-517280 · Cited by 1
Chappel, E. (2023) · Global Journal of Engineering and Technology Advances · doi:10.30574/gjeta.2023.16.3.0181 · Cited by 1
“Research showed, notably in physiologic monitor devices like ECG, that 80 to more than 90% of alarms are false positives [20].”
BackgroundOptimisation of Electrode Placement for New Ambulatory ECG Monitoring Devices
Kennedy, A., Finlay, D., Guldenring, D., Bond, R., McLaughlin, J. A., & Moran, K. (2016) · Computing in cardiology · doi:10.22489/cinc.2016.032-526 · Cited by 1
Mitra, N., & Bélteki, G. (2020) · Archives of Disease in Childhood Education & Practice · doi:10.1136/archdischild-2019-318242 · Cited by 1
Improving the Efficacy of Deep-Learning Models for Heart Beat Detection on Heterogeneous Datasets
Bizzego, A., Gabrieli, G., Neoh, M. J. Y., & Esposito, G. (2021) · Bioengineering · doi:10.3390/bioengineering8120193 · Cited by 6
“Beat identification errors in medical settings cause false alarms that can make the identification of underlying conditions more difficult [30,31], and can affect both patient [32] and healthcare professionals wellbeing [33].”
BackgroundPhysioTag: An Open-Source Platform for Collaborative Annotation of Physiological Waveforms
McCullum, L. B., Saeed, H., Moody, B., Perry, D., Gottlieb, E. R., Pollard, T., Borrat, X., Li, Q., Clifford, G. D., Mark, R. G., et al. (2022) · Computing in cardiology · doi:10.22489/cinc.2022.335 · Cited by 1
“Despite considerable effort from algorithm developers, false arrhythmia alarms in ICUs are still a major problem [10].”
BackgroundNurses perception and practices regarding clinical alarm in critical care units
Timilsina, S. S. (2025) · Civil Medical Journal · doi:10.59338/cmj.24 · Cited by 1
Selvam, I. J., Srinivasan, C., & Saravanakumar, C. (2025) · The Journal of Supercomputing · doi:10.1007/s11227-025-07543-w · Cited by 1
“However, a study [8] found that in some ICU patients, ECG-derived respiratory signals were inaccurate despite continuous breathing, raising concerns.”
Pelter, M. M., & Carey, M. G. (2025) · European Journal of Cardiovascular Nursing · doi:10.1093/eurjcn/zvaf120 · Cited by 1
The Nurse Work Environment and Staffing in Critical Care
Connell, K. A., Lin, B., Figueroa, H., & Bethel, C. (2025) · Critical Care Nursing Clinics of North America · doi:10.1016/j.cnc.2025.07.002 · Cited by 2
Jamal, H., LaLoggia, M., Boggarapu, S., Moazamian, R., Azzo, J. D., Nyah, W., & McMillan, T. (2025) · JACC Case Reports · doi:10.1016/j.jaccas.2025.105717 · Cited by 1
Мультимодальная объяснимость для ОРИТ-сигналов (VTaC): метрические и асимптотические результаты
Trofimov, Y. V., Аверкин, А. Н., Кузнецов, Е. М., Еремеев, А. П., & Нечаевский, А. В. (2025) · Herald of Tver State University Series Applied Mathematics · doi:10.26456/vtpmk757 · Cited by 1
Management of medical device alarms in intensive care units: A scoping review
Leigh, V., Grech, C., Roderick, A., Kelly, M., & Pelentsov, L. J. (2026) · Australian Critical Care · doi:10.1016/j.aucc.2025.101511 · Cited by 1
Rens, N. E., Rajotte, J. J., Deng, H., Meng, J., Mueller, A. L., Houle, T. T., Wiener-Kronish, J. P., Safavi, K. C., & Ruscic, K. J. (2026) · The Joint Commission Journal on Quality and Patient Safety · doi:10.1016/j.jcjq.2026.02.001 · Cited by 1
Severity‑aware multimodal network for chest x‑ray triage and disease captioning
Ghulam, Z., Mittal, R., & Ukwatta, E. (2026) · SPIE Medical Imaging: Computer-Aided Diagnosis · doi:10.1117/12.3082906 · Cited by 1
Ruppel, H., Hinz, E., Frey, N., Wunderlich, M. M., & Poncette, A.-S. (2026) · The Joint Commission Journal on Quality and Patient Safety · doi:10.1016/j.jcjq.2026.02.004 · Cited by 1
Medical device errors in intensive care units and patient safety: a systematic review
Ziaudeen, Z., Selvam, J., & Dakshinamoorthi, A. (2026) · Expert Review of Medical Devices · doi:10.1080/17434440.2026.2649555 · Cited by 1
“For example, in one observational study, 88.8% of arrhythmia alarms were identified as false positives [82], highlighting system-level alarm management challenges.”
InfluentialKhan, A., & Collins, B. J. (2026) · Critical Care Medicine · doi:10.1097/ccm.0000000000007197 · Cited by 1
Lin, W.-L., Fang, S.-S., & Chen, S.-H. (2026) · International Urology and Nephrology · doi:10.1007/s11255-026-05241-x · Cited by 1
“Alert fatigue constitutes one of the most extensively documented patient safety phenomena in contemporary hospital medicine [12, 29].”
Huang, X., Xie, L., Zhu, X., Cai, J., Zhang, W., & Xia, Y. (2026) · BMJ Open · doi:10.1136/bmjopen-2026-117928 · Cited by 1
Assessing Effect of Beat Detector on Detection Dependent Signal Quality Indices
Daluwatte, C., Johannesen, L., Galeotti, L., Vicente, J., Strauss, D. G., & Scully, C. G. (2016) · Computing in cardiology · doi:10.22489/cinc.2016.267-273 · Cited by 2
“8% of arrhythmia alarms in the ICU are false [2].”
BackgroundElectrocardiographic Alarms in the Acute Care Setting
Jing Jing Zhang, M., & Weathers, N. (2018) · GSTF Journal of Nursing and Health Care · doi:10.5176/2345-7198_5.1.49 · Cited by 1
Detection of Junctional Ectopic Tachycardia by Central Venous Pressure
Tan, X. T., Dai, Y., Humayun, A. I., Chen, H., Allen, G. I., & Jain, P. (2021) · Lecture notes in computer science · doi:10.1007/978-3-030-77211-6_29 · Cited by 1
“based on electrocardiogram (ECG) waveforms and result in a staggering number of false alarms (∼72-99% of clinical alarms are false [9]).”
BackgroundO’Dowd, D., Harte, A., & O’Sullivan, C. (2021) · Physiotherapy Practice and Research · doi:10.3233/ppr-200468 · Cited by 1
Stehling, H., Büscher, A., Euteneier, A., Gräsner, J., Wölfl, C., Seidling, H. M., Stützle, M., Haefeli, W. E., Gastmeier, P., Jürgensen, J. S., et al. (2015) · Springer eBooks · doi:10.1007/978-3-662-45150-2_31 · Cited by 1
An Overview of Applied Smart Health Care Informatics in the Context of Computational Intelligence
De, S., & Das, R. K. K. (2022) · Wiley eBooks · doi:10.1002/9781119743187.ch1 · Cited by 1
Howard, J. (2018) · Springer eBooks · doi:10.1007/978-3-319-93224-8_28 · Cited by 1
Debus, E. S., Grundmann, R. T., & Heilberger, J. (2017) · Springer eBooks · doi:10.1007/978-3-662-54298-9_4 · Cited by 1
Functional and sensible: Patient monitoring alarm tones designed with those who hear them
Sen, A., Sen, Y., Barile, M., Palmedo, S., Walden, A., & Antunes, V. (2022) · Proceedings of DRS · doi:10.21606/drs.2022.684 · Cited by 1
Predictive Alarm Prevention by Forecasting Threshold Alarms at the Intensive Care Unit
Chromik, J., Pfitzner, B., Ihde, N., Michaelis, M., Schmidt, D., Klopfenstein, S. A. I., Poncette, A.-S., Balzer, F., & Bert Arnrich, A. (2023) · Communications in computer and information science · doi:10.1007/978-3-031-38854-5_12 · Cited by 1
Making it easier to listen to clinical alarms
Edworthy, J. R. (2022) · Convention of the European Acoustics Association Forum Acusticum · doi:10.61782/fa.2023.0190 · Cited by 1
Curating Naturally Adversarial Datasets for Learning-Enabled Medical Cyber-Physical Systems
Pugh, S., Ruchkin, I., Weimer, J. E., & Lee, I. (2024) · ACM/IEEE International Conference on Cyber-Physical Systems (ICCPS) · doi:10.1109/iccps61052.2024.00026 · Cited by 1
“Unfortunately, such systems are known to overwhelm caregivers with many in-actionable or non-informative alarms [5], [6].”
BackgroundThatoi, P., Choudhary, R., Selvarathinam, A. L. X. R. A., David Samuel Azariya, S., & Hasan, M. K. (2024) · Artificial Intelligence and Cloud Computing Conference (AICCC) · doi:10.1145/3719384.3719467 · Cited by 1
“Additionally, they use temporal information ineffectively, thus missing more subtle but significant trends that might indicate a slow decline in the baby’s health [1].”
Schatz, M. R. (2016) · UA Campus Repository (The University of Arizona) · Cited by 2
Cruz, S. N. (2018) · doi:10.28971/532018cs118 · Cited by 2
“These alarms totaled 12,671 of which 11,251 or 88.8% were false alarms (Drew et al., 2016).”
BackgroundInfluentialChow, J. L. (2021) · UA Campus Repository (The University of Arizona) · Cited by 1
Hobson, C. (2019) · Faculty Opinions – Post-Publication Peer Review of the Biomedical Literature · doi:10.3410/f.733181042.793560090 · Cited by 1
Mishra, A. K. (2021) · MOspace Institutional Repository (University of Missouri) · doi:10.32469/10355/90073 · Cited by 1
Maharjan, J., Thapa, R., Calvert, J. S., Attwood, M. M., Shokouhi, S., Chetty, S. C., Iqbal, Z., Singh, N. P., Arnold, R., Hoffman, J. L., et al. (2022) · SSRN Electronic Journal · doi:10.2139/ssrn.4130480 · Cited by 1
Veritas-RPM: Provenance-Guided Multi-Agent False Positive Suppression for Remote Patient MonitoringPreprint
Misro, A., Sharma, V., & Gowda, S. N. (2026) · arXiv (Cornell University) · doi:10.48550/arxiv.2604.16081 · Cited by 1
“Published evidence consistently shows that a substantial proportion of physiological monitor alarms are non-actionable, with reported rates ranging from 74% to 99% depending on monitoring context and device configuration (Cvach, 2012; Drew et al., 2014).”
Benchmarked Yet Not Measured -- Generative AI Should be Evaluated Against Real-World UtilityPreprint
Mondal, I., & Bhardwaj, S. (2026) · arXiv (Cornell University) · doi:10.48550/arxiv.2605.06856 · Cited by 1
Kinghorn, K. A. (2015) · USF Scholarship Repository (University of San Francisco)
Kaski, M., Vanhatalo, J., Treacy, S. M., & Viertiö‐Oja, H. (2015) · Critical Care · doi:10.1186/cc14235
Bai, Y. (2016) · eScholarship (California Digital Library)
Self-citationMedicine Enters the Computer Age
Santhosh, L., & Khanna, R. R. (2017) · Springer eBooks · doi:10.1007/978-3-319-48707-6_2
Survey and Analysis of Current State of Ventilator Alarms in the Intensive Care Unit
Lin, Z., & Zheng, K. (2018) · Global Clinical Engineering Journal · doi:10.31354/globalce.v0i1.25
Optimisation of Patient Monitor Alarm Settings Using Annotated Hospital Data
Fontana, M. (2018) · Aaltodoc (Aalto University)
Blinded by the height: A case of telemetry T wave overcounting
Groh, C. A., & Goldschlager, N. (2019) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2019.03.013
Noise fatigue in the ICU: platform for sound data collection and visualization: Cacophony Mapper
Lee, Y. (2019) · Research Repository (Delft University of Technology) · Cited by 1
Kulkarni, A. (2019) · Medical Journal of Clinical Trials & Case Studies · doi:10.23880/mjccs-16000193
Pediatric Meaningful Alarm Management Approach
Gilmore, J. (2017) · Carolina Digital Repository (University of North Carolina at Chapel Hill) · doi:10.17615/tbrs-cg79
Zoller, R., & Nositschka, J. (2019) · intensiv · doi:10.1055/a-0970-0520
“Das kontinu ierliche Messen der Vitalzeichen ist ein hilfreiches Instru ment, um die Patientensicherheit zu gewährleisten [1].”
BackgroundInfluentialSex differences in intensive care unit electrocardiographic alarms
Pelter, M. M., & Hu, X. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-817728-0.00053-x
Self-citationFernandes, C. O., & de Lucena, C. J. P. (2020) · doi:10.5753/sbcas.2020.11556
“…falsos alarmes podem levar a uma sobrecarga de alarmes que pode comprometer a atenção dos cuidadores de saúde (Cvach, 2012; Drew et al., 2014; Jones, 2014; Tanner, 2013), podendo levá-los a perder alarmes relevantes que anunciam eventos significativos ou que representam risco de vida aos pacientes.”
BackgroundLEVERAGING TEMPORAL SUBSEQUENCES FOR TIME-SERIES CLASSIFICATION
Roychoudhury, S. (2020) · TUScholarShare (Temple University)
Italian Journal of Prevention, Diagnostic and Therapeutic Medicine
Zamberlan, C., & Pelati, C. (2018) · doi:10.30459/2018-19
Heidl, C. M., Hofmann, J. M., Prenzel, R., Fiedler, C., & Zerth, J. (2021) · Perspektiven Sozialwirtschaft und Sozialmanagement · doi:10.1007/978-3-658-32556-5_20
Des progrès technologiques et humains en réanimation
Fazilleau, C. (2021) · Soins · doi:10.1016/j.soin.2021.11.009
Care of the patient with cardiac arrhythmias
Lee, G., Fålun, N., Angus, N. J., Hendriks, J., Norekvål, T. M., Berg, S. K., & Fitzsimons, D. (2021) · Oxford University Press eBooks · doi:10.1093/med/9780198849315.003.0007
L’alarme sonore des scopes, un outil au service des malades et des soignants
de Raucourt, I. P. (2022) · La Revue de l Infirmière · doi:10.1016/j.revinf.2022.01.013
Verspoor, K. M. (2019) · Faculty Opinions – Post-Publication Peer Review of the Biomedical Literature · doi:10.3410/f.733181042.793560978
Intelligent alarms using principal component analysis on hemodynamic monitoring (Preprint)Preprint
Fuglerud, S. S., Wedeld, M. D., Martens, H. A., & Skjærvold, N. K. (2019) · doi:10.2196/preprints.16997
Fernandes, C. O., Miles, S., & de Lucena, C. J. P. (2020) · doi:10.2196/preprints.19272
Research Progress on the Impact and Intervention of Noise and Alarms on Intensive Care Units
王, 晗. (2023) · Advances in Clinical Medicine · doi:10.12677/acm.2023.1351171
Ensuring Specificity as a Strategy for Increasing Alarm Safety
Gül, G., & İntepeler, Ş. S. (2023) · Journal of Basic and Clinical Health Sciences · doi:10.30621/jbachs.1251111
Zimpelman, L. T. (2023) · MOspace Institutional Repository (University of Missouri) · doi:10.32469/10355/97012
Sujimongkol, C., Daochai, S., Wichakhrueang, S., & Daochai, C. (2023) · Biomedical Engineering International Conference · doi:10.1109/bmeicon60347.2023.10322051 · Cited by 2
“…have revealed that false alarms occur as frequently as every 90 seconds to two minutes, highlighting the severity of the problem in ICU settings [1, 2] He et al. (2021) reported an incident rate of 142 patient monitor alarms per patient-day in the ICU, indicating that each patient experienced…”
BackgroundHighlights of 2023-Print Issue
Authority, P. S. (2023) · Patient Safety · doi:10.33940/001c.92361
A Software Tool for Anomaly Detection and Labeling of Ventilator Waveforms
Chen, C., Wang, Z., Chen, C., Wang, X., & Liu, S. (2024) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-031-51485-2_29
Jain, D., Gupta, A., Pandey, A. K., & Vats, P. (2024) · Wiley eBooks · doi:10.1002/9781119905202.ch26
Euteneier, A., Gräsner, J., & Wölfl, C. (2024) · Erfolgskonzepte Praxis- & Krankenhaus-Management · doi:10.1007/978-3-662-67565-6_37
Gallagher, C. P., Ziniel, S. I., Newman, J., Wathen, B., Marusich, P., Poppy, C., Carpenter, T. C., & Gist, K. M. (2024) · Cardiology in the Young · doi:10.1017/s1047951124025034
Subbe, C. P., Champunot, R., & Jones, D. A. (2024) · Springer eBooks · doi:10.1007/978-3-031-67951-3_32
Keim‐Malpass, J., Clark, M., Ratcliffe, S. J., Krahn, K. N., Jones, M., Sullivan, B. A., Moorman, L. P., Moorman, J. R., & Bourque, J. M. (2025) · doi:10.2196/preprints.80206
Enhancing Personal Protection in Dealing With Nurses’ Alarm Fatigue: A Prescriptive Model
Movahedi, A., Sadooghiasl, A., Ahmadi, F., & Vaismoradi, M. (2025) · Nursing Forum · doi:10.1155/nuf/6937479
“When critical alarms are neglected, the chance of patient harm is enhanced [7].”
Alarm Fatigue Among Healthcare Workers Working in Intensive Care Unit: A Cross-sectional Study
Pant, C., Shrestha, B. K., & Rajesh, K. C. (2024) · Medical Journal of Shree Birendra Hospital · doi:10.3126/mjsbh.v23i1.75750
Gomez, J. P. V., Pelter, M. M., Tison, G. H., MORTARA, D. W., Badilini, F., Abe‐Jones, Y., Oreper, S., Fang, M. C., & Prasad, P. A. (2025) · Heart Rhythm O2 · doi:10.1016/j.hroo.2025.08.034
Rodriguez, H. H., García, H., & Fajardo, C. A. (2025) · Symposium on Image, Signal Processing and Artificial Vision (STSIVA) · doi:10.1109/stsiva66383.2025.11156764
“Arrhythmia alarms are configured to be highly sensitive to avoid missing critical events; however, this compromises their specificity [2], leading to a false alarm rate that can reach up to 88.8% [3].”
InfluentialDo Re Mi Fa So Pass the Tool: Using Melodic Prediction to Improve Human-Robot Fluency
Rogel, A., Yang, Q., Hayley, J., & Weinberg, G. (2025) · IEEE RO-MAN · doi:10.1109/ro-man63969.2025.11217907 · Cited by 3
“The drone’s design, which avoided standing tones and incorporated subtle variations, appears to have successfully addressed issues of auditory fatigue for long term audio [37], [57].”
Chiziwa, C., Kamndaya, M. S., Phepa, P., IMPALA Study Team, Vweza, A. O., Calis, J. C. J., & Bierling, B. (2025) · Bioengineering · doi:10.3390/bioengineering12111210
“These threshold alarms help caregivers to either attend to emergencies or to flag deterioration [8,9].”
Areia, C., Biggs, C. A., Santos, M., Thurley, N., Gerry, S., Tarassenko, L., Watkinson, P., & Vollam, S. (2021) · Figshare · doi:10.6084/m9.figshare.16696588
Areia, C., Biggs, C. A., Santos, M., Thurley, N., Gerry, S., Tarassenko, L., Watkinson, P., & Vollam, S. (2021) · Figshare · doi:10.6084/m9.figshare.16696603
Areia, C., Biggs, C. A., Santos, M., Thurley, N., Gerry, S., Tarassenko, L., Watkinson, P., & Vollam, S. (2021) · Figshare · doi:10.6084/m9.figshare.16696591
Areia, C., Biggs, C. A., Santos, M., Thurley, N., Gerry, S., Tarassenko, L., Watkinson, P., & Vollam, S. (2021) · Figshare · doi:10.6084/m9.figshare.16696609
Metadata
- Authors
- Drew, B. J., Harris, P., Zègre-Hemsey, J. K., Mammone, T., Schindler, D., Salas-Boni, R., Bai, Y., Tinoco, A., Ding, Q., & Hu, X.
- Affiliations
- University of California, San Francisco (US)
- University of California San Francisco Medical Center (US)
- Corresponding author
- Drew, Barbara J.
- Venue
- PLoS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2014-10-22
- Volume
- 9
- Issue
- 10
- Pages
- e110274
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC4206416
- OpenAlex
- W2001741247
- Cited by
- 438
- References
- 23
- Citations by year
- 2026: 51 · 2025: 31 · 2024: 25 · 2023: 37 · 2022: 26 · 2021: 58 · 2020: 29 · 2019: 38 · 2018: 42 · 2017: 36 · 2016: 43 · 2015: 21 · 2014: 1
- FWCI
- 33.15
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- PLoS ONE (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Carolina Digital Repository (University of North Carolina at Chapel Hill) (submittedVersion)
- Dominican Scholar (Dominican University of California) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- eScholarship (California Digital Library) (submittedVersion)
- UNC Libraries
- Funders
- University of California, San Francisco
- Topics
- Health Sciences › Medicine › Surgery › Healthcare Technology and Patient Monitoring
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Patient Safety and Medication Errors
- Health Sciences › Health Professions › Health Information Management › Electronic Health Records Systems
- Keywords
- alarm fatigue, intensive care unit patients, false positive alerts, alarm settings, ECG leads, QRS amplitude, atrial fibrillation, ventricular tachycardia, bundle branch block, persistent atrial fibrillation, patient safety
- MeSH terms
- Adult, Aged, Arrhythmias, Cardiac, Atrial Fibrillation, Diagnostic Errors, Electrocardiography, Equipment Failure, False Positive Reactions, Female, Humans, Intensive Care Units, Male, Middle Aged, Monitoring, Physiologic, Critical Illness, Clinical Alarms
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.