Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients
Open accessThe DOI matches its Crossref record; 23 of the 23 references it cites were found in a registry or an index.
Abstract
PURPOSE: Physiologic monitors are plagued with alarms that create a cacophony of sounds and visual alerts causing "alarm fatigue" which creates an unsafe patient environment because a life-threatening event may be missed in this milieu of sensory overload. Using a state-of-the-art technology acquisition infrastructure, all monitor data including 7 ECG leads, all pressure, SpO(2), and respiration waveforms as well as user settings and alarms were stored on 461 adults treated in intensive care units. Using a well-defined alarm annotation protocol, nurse scientists with 95% inter-rater reliability annotated 12,671 arrhythmia alarms. RESULTS: A total of 2,558,760 unique alarms occurred in the 31-day study period: arrhythmia, 1,154,201; parameter, 612,927; technical, 791,632. There were 381,560 audible alarms for an audible alarm burden of 187/bed/day. 88.8% of the 12,671 annotated arrhythmia alarms were false positives. Conditions causing excessive alarms included inappropriate alarm settings, persistent atrial fibrillation, and non-actionable events such as PVC's and brief spikes in ST segments. Low amplitude QRS complexes in some, but not all available ECG leads caused undercounting and false arrhythmia alarms. Wide QRS complexes due to bundle branch block or ventricular pacemaker rhythm caused false alarms. 93% of the 168 true ventricular tachycardia alarms were not sustained long enough to warrant treatment. DISCUSSION: The excessive number of physiologic monitor alarms is a complex interplay of inappropriate user settings, patient conditions, and algorithm deficiencies. Device solutions should focus on use of all available ECG leads to identify non-artifact leads and leads with adequate QRS amplitude. Devices should provide prompts to aide in more appropriate tailoring of alarm settings to individual patients. Atrial fibrillation alarms should be limited to new onset and termination of the arrhythmia and delays for ST-segment and other parameter alarms should be configurable. Because computer devices are more reliable than humans, an opportunity exists to improve physiologic monitoring and reduce alarm fatigue.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures











![Figure 12 True accelerated ventricular rhythm alarm showing why this arrhythmia is not considered an “actionable” alarm condition. Accelerated ventricular rhythm at a rate of 56 for the first 5 beats followed by 2 fusion beats; the last 2 beats are normal sinus rhythm. The invasive arterial pressure waveform (bottom tracing) shows no change between normal rhythm and accelerated ventricular rhythm which confirms the rationale for the published guidelines stating no treatment is indicated for this arrhythmia in hospital settings [5].](https://figures.cochat.ai/paper-figures/440dfa9937e3409b986af268dec77e1a/insights-into-the-problem-of-alarm-fatigue-with-physiologic-monitor-devices-a-figure-12.jpg)




Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“One study conducted a comprehensive observational analysis over 31 days across five ICUs, involving 461 adults and monitoring over 2.5 million alarm instances. Of 1,154,201 unique alarms, 88.8% were for arrhythmias with a majority being false positive98.”
From Bias recognition and mitigation strategies in artificial intelligence healthcare applications (2025) · cites this paper as [98]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The Cardiac Arrhythmia Suppression Trial (CAST) Investigators (1989) · New England Journal of Medicine · doi:10.1056/nejm198908103210629 · Cited by 3,300
Third Universal Definition of Myocardial Infarction
Thygesen, K., Alpert, J. S., Jaffe, A. S., Simoons, M. L., Chaitman, B. R., & White, H. D. (2012) · Circulation · doi:10.1161/cir.0b013e31826e1058 · Cited by 2,985
Third universal definition of myocardial infarction
the Writing Group on behalf of the Joint ESC/ACCF/AHA/WHF Task Force for the Universal Definition of Myocardial Infarction, Thygesen, K., Alpert, J. S., Jaffe, A. S., Simoons, M. L., Chaitman, B. R., & White, H. D. (2012) · Nature Reviews Cardiology · doi:10.1038/nrcardio.2012.122 · Cited by 2,053
Zipes, D. P., Camm, A. J., Borggrefe, M., Buxton, A. E., Chaitman, B., Fromer, M., Gregoratos, G., Klein, G., Myerburg, R. J., Quinones, M. A., et al. (2006) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2006.07.010 · Cited by 1,973
Weir, M. (1990) · Cardiopulmonary Physical Therapy Journal · doi:10.1097/01823246-199001020-00009 · Cited by 1,058
The association for the advancement of medical instrumentation
(1974) · Ultrasound in Medicine & Biology · doi:10.1016/0301-5629(74)90035-0 · Cited by 1,031
Drew, B. J., Ackerman, M. J., Funk, M., Gibler, W. B., Kligfield, P., Menon, V., Philippides, G. J., Roden, D. M., & Zareba, W. (2010) · Circulation · doi:10.1161/circulationaha.109.192704 · Cited by 821
Self-citationAuthors/Task Force Members (2006) · European Heart Journal · doi:10.1093/eurheartj/ehl199 · Cited by 523
ACC/AHA guidelines for ambulatory electrocardiography
Crawford, M. H., Bernstein, S. J., Deedwania, P. C., DiMarco, J. P., Ferrick, K. J., Garson, A., Green, L. A., Greene, H. L., Silka, M. J., Stone, P. H., et al. (1999) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(99)00354-x · Cited by 309
Improving Alarm Performance in the Medical Intensive Care Unit Using Delays and Clinical Context
Görges, M., Markewitz, B. A., & Westenskow, D. R. (2009) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e31819bdfbb · Cited by 211
Cited by
Items 401–457 of 457
Areia, C., Biggs, C. A., Santos, M., Thurley, N., Gerry, S., Tarassenko, L., Watkinson, P., & Vollam, S. (2021) · Figshare · doi:10.6084/m9.figshare.16696606
Areia, C., Biggs, C. A., Santos, M., Thurley, N., Gerry, S., Tarassenko, L., Watkinson, P., & Vollam, S. (2021) · Figshare · doi:10.6084/m9.figshare.16696600
Non-invasive detection of hypovolemia by multimodal body sensor network
Tigges, T. (2023) · DepositOnce · doi:10.14279/depositonce-18833
Areia, C., Biggs, C. A., Santos, M., Thurley, N., Gerry, S., Tarassenko, L., Watkinson, P., & Vollam, S. (2021) · Figshare · doi:10.6084/m9.figshare.16696594
Areia, C., Biggs, C. A., Santos, M., Thurley, N., Gerry, S., Tarassenko, L., Watkinson, P., & Vollam, S. (2021) · Figshare · doi:10.6084/m9.figshare.16696597
Notkun hjartarafsjáa við vöktun sjúklinga : vettvangsrannsókn á hjartadeild
Freysdóttir, G. R. (2017) · Skemman
Raaijmakers, G. (. (2024) · Research Repository (Delft University of Technology)
Meten, weten en de werkelijkheid!
Bouwman, R. A. (2023) · TU/e Research Portal
Sjuksköterskors upplevelse av larmtrötthet och övervakningsutrustning - En litteraturstudie
Lind, J., & Brostedt, L. (2025) · Uppsala University Publications (Uppsala University)
Technological interventions to alleviate alarm fatigue in intensive care medicine
Chromik, J. (2025) · publish.UP (University of Potsdam) · doi:10.25932/publishup-68728
von Essen, C., & Eriksson, L. (2024) · Diva portal (Dalarna University Library)
Time-series ECG Imputation Using a Pattern-Based Masking FrameworkPreprint
Suba, S., Novak, A., Xia, X., Al‐Zaiti, S. S., & Pelter, M. M. (2026) · medRxiv · doi:10.64898/2026.01.14.26344164
Khanolkar, U. K., Yadav, A., Mann, A., Pappada, S. M., Deshmukh, A., & Maan, A. (2026) · Journal of Cardiovascular Electrophysiology · doi:10.1111/jce.70271
Keeping Watch: Reviewing Recent Telemetry-Related Events in Pennsylvania Facilities
Sanchez, C. E. (2026) · Patient Safety · doi:10.33940/001c.154644
Khairat, S., Morelli, J. G., Yang, S., Li, Y., Herasevich, V., Mohan, D., Handzel, R. M., Ratwani, R. M., Apathy, N. C., Boynton, M. H., et al. (2026) · Journal of the American Medical Informatics Association · doi:10.1093/jamia/ocag017
Detecting False Arrhythmias Alarms in the ICU Using a Deep Learning Approach
Jaimes, G. D. M., & Garnica, K. M. A. (2024) · Universidad Industrial de Santander
Contextualized clinical anomaly detection with explainable AI and patient modeling
Elketroussi, A., Djebbar, B., & Bekkouche, I. (2026) · Indonesian Journal of Electrical Engineering and Computer Science · doi:10.11591/ijeecs.v41.i2.pp614-623
Karboub, K. (2022) · theses.fr (ABES) · doi:10.70675/c8b638daz5fbcz4524za62dz27eee11ca094 · Cited by 1
Barriers to Electrocardiographic Monitoring Practice Among Nurses: A Scoping Review
Giusti, Â. C. B. D. S., de Medeiros, K. S., Chiavone, F. B. T., da Silva, W. P., Cornélio, M. E., & Gallani, M. C. B. J. (2026) · Cardiology in Review · doi:10.1097/crd.0000000000001261
Designing Novel Visual Physiologic Monitor Displays for Patient Care
Cummins, M. L., Sicard, C., Yang, S., Schlesinger, J. J., & Shive, J. (2026) · Human Factors The Journal of the Human Factors and Ergonomics Society · doi:10.1177/00187208261445900
“Auditory alarms are used in parallel to signal out-of-range physiological values but are not designed to convey detailed or integrated patient information (Drew et al., 2014).”
Impact of Noise Levels of the Dialysis Machine Alarm on Patient Anxiety
Bera, B., & Hemanth, P. (2026) · Hemodialysis International · doi:10.1111/hdi.70080
Hassouneh, D. A., D’emeh, W. M., & Yacoub, M. I. (2026) · Journal of Neonatal Nursing · doi:10.1016/j.jnn.2026.101864
The promise and pragmatics of wearable continuous monitoring on hospital wards
Connell, K. A., & Weissman, G. E. (2026) · Intensive and Critical Care Nursing · doi:10.1016/j.iccn.2026.104447
At the Crossroads of Safety: The Anesthesiologist’s Leadership in Preventing Harm
Methangkool, E. (2026) · Springer eBooks · doi:10.1007/978-3-032-18058-2_23
Wagner, F., Thöle, M., Menz, M., Hoermandinger, C., Kuschnerus, K., Mader, M., Ferentzi, H., Berger, F., Miera, O., Köhler, F., et al. (2026) · Cardiology in the Young · doi:10.1017/s1047951126113018
Yıldız, O., & Subaşı, A. (2026) · medRxiv · doi:10.64898/2026.05.17.26353436
Kaneko, H., & Inoue, S. (2026) · Sensors · doi:10.3390/s26113561
“In practice, comparative evaluations including early warning scores (EWS) report low positive predictive value (PPV) [2,4,5,24]. https://doi.org/10.3390/s26113561 Care facilities also maintain care records that document residents’ conditions and care activities, although research remains limited…”
InfluentialKG-Sepsis: Knowledge-Guided Temporal Graph Neural Networks for Early Sepsis Prediction in ICU
Chowdhury, S. H., Tasneem, Z., Tuhin, G. S. A., Sayam, A. L., Hossain, M. A., & Ifty, R. A. (2026) · IEEE International Conference on Quantum Photonics, Artificial Intelligence & Networking (QPAIN) · doi:10.1109/qpain69676.2026.11546025 · Cited by 1
“The significant jump in AUPRC suggests that our model generates far fewer false alarms, a critical factor for preventing alarm fatigue in ICUs where alert overload affects 72-99% of notifications [22].”
A Cascaded Quantized Spiking Neural Network for Real-Time ECG Arrhythmia Detection on Edge Hardware
Banjo, O. O., & Ghoraani, B. (2026) · Sensors · doi:10.3390/s26123723 · Cited by 1
Sohail, M. A., Heyntsen, B., Safdar, S., Foo, S. Y., & Altaf, M. A. B. (2026) · IEEE Open Journal of Circuits and Systems · doi:10.1109/ojcas.2026.3707506
“In practice, a large fraction of physiologic monitor alarms are false [28], underscoring the need for on-chip, real-time suppression to ensure actionable notifications.”
When stable performance misleads: A formal model of compensation-dependent systems
Loepke, A. W. (2026) · Safety Science · doi:10.1016/j.ssci.2026.107443
A quality improvement study on noise reduction in a level 3 neonatal intensive care unit
Forestier, M., Caeymaex, L., Decobert, F., Durrmeyer, X., Dassieu, G., & Tauzin, M. (2026) · European Journal of Pediatrics · doi:10.1007/s00431-026-07139-3
“Driven by precautionary mandates and liability mitigation, manufacturers often prioritize alarm sensitivity over acoustic clarity, leading to a surplus of non-action-able alerts that contribute to a saturated NICU environment [29].”
Alkubati, S. A., Salameh, B. S., Halboup, A. M., Loutfy, A., Pasay-an, E. A., Zoromba, M. A., El‐Gazar, H. E., El‐Monshed, A. H., Tlili, M. A., & Elsayed, S. M. (2026) · SAGE Open Nursing · doi:10.1177/23779608261470481
“Because computers are more reliable than humans, there is a potential to enhance hemodynamic monitoring as well as reduce alarm fatigue using computers (Drew et al., 2014).”
InfluentialDie Mensch-Achse: Input, Kognition und Handlung
Elsharkawy, A. E. (2026) · Essentials · doi:10.1007/978-3-662-74123-8_4
Jang, J., Choi, Y. J., Sim, T. Y., Lee, K.-B., Kim, J., Cho, E. Y., Choi, Y., Hong, S., Jung, B. M., KIM, S., et al. (2026) · Journal of Clinical Medicine · doi:10.3390/jcm15145753
A Dual-Path Optical Clip for Bubble-Aware Continuous Monitoring in Surgical Drainage Tubing
Chiu, C.-W., Kuo, P.-C., Chang, M.-W., Liao, Y.-T., & Wu, K. (2026) · IEEE Sensors Letters · doi:10.1109/lsens.2026.3721660
Kuznetsov, E. M., Averkin, A. N., Trofimov, Y. V., Lopatin, M. A., & Kondrashova, E. S. (2026) · International Conference on Neural Networks and Neurotechnologies (NeuroNT) · doi:10.1109/neuront71829.2026.11650778
Development and Validation of a Prediction Model for Alarm-Triggered Interventions in Pediatrics
Akahoshi, S., Nagasawa, S., Sakatani, M., Minegishi, H., Kawata, N., Iio, K., Fukuda, K., Kinumaki, A., Hataya, H., Sato, F., et al. (2026) · Hospital Pediatrics · doi:10.1542/hpeds.2026-009243
Patient characteristics and monitor alarms in a cardiac ICU: A retrospective cross-sectional study
Jämsä, J. O., Uutela, K., Lääperi, M., Tapper, A., & Lehtonen, L. (2026) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-026-01486-0
Certified AI Triage of ICU AlarmsPreprint
Syed, M. S., & Yasaei, R. (2026) · arXiv (Cornell University) · doi:10.48550/arxiv.2609.12365
“Bedside monitors in intensive care units generate arrhythmia alarms faster than clinical attention can absorb, and most are false (Drew et al., 2014; Sendel-bach and Funk, 2013; Cvach, 2012).”
management degli allarmi in terapia intensiva: analisi e prevenzione degli incidenti
Zamberlan, C., & Pelati, C. (2019) · Italian Journal of Prevention Diagnostic and Therapeutic Medicine
Automated Detection of Ventricular Flutter and Ventricular Hypertrophy From ECG Using Deep Learning
Pathak, A., Sharma, R., & Sharma, S. D. (2026) · Advances in computational intelligence and robotics book series · doi:10.4018/979-8-2600-4599-2.ch009
Pathak, A., Sharma, R., & Sharma, S. D. (2026) · Advances in computational intelligence and robotics book series · doi:10.4018/979-8-2600-4599-2.ch010
Bazroun, M. I. A., Hakeem, A. A. A., Alhashem, N. M., Alnasser, H. M., Alabadi, Z. M., Alfadhel, R. A., ALSAGER, N. H. H., Jaffar, A. H. A., Alrabie, Z. A. Y., Allabad, Z. T., et al. (2026) · Acta Innovations · doi:10.62441/actainnovations.v62i1s.1185
Palamim, C. V. C., & Marson, F. A. L. (2026) · Intensive and Critical Care Nursing · doi:10.1016/j.iccn.2026.104589
Connectedness, Cognitive Load, and Human-AI Oversight in Cyber OperationsPreprint
Conklin, N., Gao, P., & North, C. L. (2026) · arXiv (Cornell University) · doi:10.48550/arxiv.2610.02384
Tariq, A., Sohn, W., Lee, Y. J., & Kim, Y. (2026) · Arteriosclerosis Thrombosis and Vascular Biology · doi:10.1161/atvbaha.126.324766
Linhares, S. M., Schultz, K. S., Mathew, P. J., D’Aquila, M., Richburg, C. E., Buljubasic, A. M., Schroeder, K. J., Puklin, L. S., & Leeds, I. L. (2026) · Surgery · doi:10.1016/j.surg.2026.110743
Filter-induced ST elevation at the bedside after cardiac surgery
Özdemir, O., Kuzucu, Ö., & Hazır, K. E. (2026) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2026.154466
Pei, Y., Wu, Y., Hou, R., & Wan, K. (2026) · Allergy medicine. · doi:10.1016/j.allmed.2026.100142
Lazzaro, F. D., Barletta, G., Magnanimi, E., Cirillo, B., Frati, P., Marinelli, E., Pugliese, F. R., Paola, L. D., & Giordano, G. G. (2026) · Healthcare · doi:10.3390/healthcare14203382
“False alarms in the ICU, like false critical ECG arrhythmia or arterial blood pressure waveforms, generate a noisy environment and cause alarm fatigue among caregivers [44].”
Sang, M., Wang, S., Weng, F., Ya'nan, L., Qiao, W., Sheng, G., Ren, Y., & Wei, J. (2026) · International Journal of Nursing Sciences · doi:10.1016/j.ijnss.2026.10.006
李冰玉, Yue, L., Peng, Q., & Chen, H. (2026) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2026.105741
Travmaya Bağlı Maksiller Santral Kesici Eksikliğinin Ortodontik Yönetimi: Olgu Sunumu
Cengiz, S. M., Baki, Ö. F., & Cengiz, A. F. (2026) · Mersin Üniversitesi Açık Erişim Sistemi/Mersin University Open Access System
Bråberg, F., Vesterlund, J., Roxberg, Å., & Emilsson, M. (2026) · Critical Care Research and Practice · doi:10.1155/ccrp/3865004
Gim, U., Lim, L., Cho, K.-J., Cho, W. H., & Kim, J. (2026) · Diagnostics · doi:10.3390/diagnostics16213671
Metadata
- Authors
- Drew, B. J., Harris, P., Zègre-Hemsey, J. K., Mammone, T., Schindler, D., Salas-Boni, R., Bai, Y., Tinoco, A., Ding, Q., & Hu, X.
- Affiliations
- University of California, San Francisco (US)
- University of California San Francisco Medical Center (US)
- Corresponding author
- Drew, Barbara J.
- Venue
- PLoS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2014-10-22
- Volume
- 9
- Issue
- 10
- Pages
- e110274
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC4206416
- OpenAlex
- W2001741247
- Cited by
- 438
- References
- 23
- Citations by year
- 2026: 51 · 2025: 31 · 2024: 25 · 2023: 37 · 2022: 26 · 2021: 58 · 2020: 29 · 2019: 38 · 2018: 42 · 2017: 36 · 2016: 43 · 2015: 21 · 2014: 1
- FWCI
- 33.15
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- PLoS ONE (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Carolina Digital Repository (University of North Carolina at Chapel Hill) (submittedVersion)
- Dominican Scholar (Dominican University of California) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- eScholarship (California Digital Library) (submittedVersion)
- UNC Libraries
- Funders
- University of California, San Francisco
- Topics
- Health Sciences › Medicine › Surgery › Healthcare Technology and Patient Monitoring
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Patient Safety and Medication Errors
- Health Sciences › Health Professions › Health Information Management › Electronic Health Records Systems
- Keywords
- alarm fatigue, intensive care unit patients, false positive alerts, alarm settings, ECG leads, QRS amplitude, atrial fibrillation, ventricular tachycardia, bundle branch block, persistent atrial fibrillation, patient safety
- MeSH terms
- Adult, Aged, Arrhythmias, Cardiac, Atrial Fibrillation, Diagnostic Errors, Electrocardiography, Equipment Failure, False Positive Reactions, Female, Humans, Intensive Care Units, Male, Middle Aged, Monitoring, Physiologic, Critical Illness, Clinical Alarms
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.