Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients
Open accessThe DOI matches its Crossref record; 23 of the 23 references it cites were found in a registry or an index.
Abstract
PURPOSE: Physiologic monitors are plagued with alarms that create a cacophony of sounds and visual alerts causing "alarm fatigue" which creates an unsafe patient environment because a life-threatening event may be missed in this milieu of sensory overload. Using a state-of-the-art technology acquisition infrastructure, all monitor data including 7 ECG leads, all pressure, SpO(2), and respiration waveforms as well as user settings and alarms were stored on 461 adults treated in intensive care units. Using a well-defined alarm annotation protocol, nurse scientists with 95% inter-rater reliability annotated 12,671 arrhythmia alarms. RESULTS: A total of 2,558,760 unique alarms occurred in the 31-day study period: arrhythmia, 1,154,201; parameter, 612,927; technical, 791,632. There were 381,560 audible alarms for an audible alarm burden of 187/bed/day. 88.8% of the 12,671 annotated arrhythmia alarms were false positives. Conditions causing excessive alarms included inappropriate alarm settings, persistent atrial fibrillation, and non-actionable events such as PVC's and brief spikes in ST segments. Low amplitude QRS complexes in some, but not all available ECG leads caused undercounting and false arrhythmia alarms. Wide QRS complexes due to bundle branch block or ventricular pacemaker rhythm caused false alarms. 93% of the 168 true ventricular tachycardia alarms were not sustained long enough to warrant treatment. DISCUSSION: The excessive number of physiologic monitor alarms is a complex interplay of inappropriate user settings, patient conditions, and algorithm deficiencies. Device solutions should focus on use of all available ECG leads to identify non-artifact leads and leads with adequate QRS amplitude. Devices should provide prompts to aide in more appropriate tailoring of alarm settings to individual patients. Atrial fibrillation alarms should be limited to new onset and termination of the arrhythmia and delays for ST-segment and other parameter alarms should be configurable. Because computer devices are more reliable than humans, an opportunity exists to improve physiologic monitoring and reduce alarm fatigue.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures











![Figure 12 True accelerated ventricular rhythm alarm showing why this arrhythmia is not considered an “actionable” alarm condition. Accelerated ventricular rhythm at a rate of 56 for the first 5 beats followed by 2 fusion beats; the last 2 beats are normal sinus rhythm. The invasive arterial pressure waveform (bottom tracing) shows no change between normal rhythm and accelerated ventricular rhythm which confirms the rationale for the published guidelines stating no treatment is indicated for this arrhythmia in hospital settings [5].](https://figures.cochat.ai/paper-figures/440dfa9937e3409b986af268dec77e1a/insights-into-the-problem-of-alarm-fatigue-with-physiologic-monitor-devices-a-figure-12.jpg)




Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“One study conducted a comprehensive observational analysis over 31 days across five ICUs, involving 461 adults and monitoring over 2.5 million alarm instances. Of 1,154,201 unique alarms, 88.8% were for arrhythmias with a majority being false positive98.”
From Bias recognition and mitigation strategies in artificial intelligence healthcare applications (2025) · cites this paper as [98]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The Cardiac Arrhythmia Suppression Trial (CAST) Investigators (1989) · New England Journal of Medicine · doi:10.1056/nejm198908103210629 · Cited by 3,300
Third Universal Definition of Myocardial Infarction
Thygesen, K., Alpert, J. S., Jaffe, A. S., Simoons, M. L., Chaitman, B. R., & White, H. D. (2012) · Circulation · doi:10.1161/cir.0b013e31826e1058 · Cited by 2,985
Third universal definition of myocardial infarction
the Writing Group on behalf of the Joint ESC/ACCF/AHA/WHF Task Force for the Universal Definition of Myocardial Infarction, Thygesen, K., Alpert, J. S., Jaffe, A. S., Simoons, M. L., Chaitman, B. R., & White, H. D. (2012) · Nature Reviews Cardiology · doi:10.1038/nrcardio.2012.122 · Cited by 2,053
Zipes, D. P., Camm, A. J., Borggrefe, M., Buxton, A. E., Chaitman, B., Fromer, M., Gregoratos, G., Klein, G., Myerburg, R. J., Quinones, M. A., et al. (2006) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2006.07.010 · Cited by 1,973
Weir, M. (1990) · Cardiopulmonary Physical Therapy Journal · doi:10.1097/01823246-199001020-00009 · Cited by 1,058
The association for the advancement of medical instrumentation
(1974) · Ultrasound in Medicine & Biology · doi:10.1016/0301-5629(74)90035-0 · Cited by 1,031
Drew, B. J., Ackerman, M. J., Funk, M., Gibler, W. B., Kligfield, P., Menon, V., Philippides, G. J., Roden, D. M., & Zareba, W. (2010) · Circulation · doi:10.1161/circulationaha.109.192704 · Cited by 821
Self-citationAuthors/Task Force Members (2006) · European Heart Journal · doi:10.1093/eurheartj/ehl199 · Cited by 523
ACC/AHA guidelines for ambulatory electrocardiography
Crawford, M. H., Bernstein, S. J., Deedwania, P. C., DiMarco, J. P., Ferrick, K. J., Garson, A., Green, L. A., Greene, H. L., Silka, M. J., Stone, P. H., et al. (1999) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(99)00354-x · Cited by 309
Improving Alarm Performance in the Medical Intensive Care Unit Using Delays and Clinical Context
Görges, M., Markewitz, B. A., & Westenskow, D. R. (2009) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e31819bdfbb · Cited by 211
Cited by
Items 201–300 of 457
Invasive and non-invasive monitoring in the ICU
McEachron, K. R., & Costantini, T. W. (2025) · Trauma Surgery & Acute Care Open · doi:10.1136/tsaco-2025-001780 · Cited by 12
Nahid, N., Ahad, M. A. R., & Inoue, S. (2025) · Sensors · doi:10.3390/s25216560 · Cited by 8
Yang, S., Galvagno, S. M., Badjatia, N., Stein, D., Teeter, W. A., Scalea, T. M., Shackelford, S. A., Fang, R., Miller, C. H. T., Hu, P., et al. (2023) · JMIR Human Factors · doi:10.2196/46030 · Cited by 9
“Moreover, auditory alarms often cause “alarm fatigue” instead of increasing situational awareness [3].”
BackgroundToften, S., Kjellstadli, J. T., Kværness, J., Pedersen, L., Laugsand, L. E., & Thu, O. K. F. (2024) · Frontiers in Physiology · doi:10.3389/fphys.2024.1502413 · Cited by 11
Generalizability of SuperAlarm via Cross-Institutional Performance Evaluation
Xiao, R., Do, D. H., Ding, C., Meisel, K. M., Lee, R. J., & Hu, X. (2020) · IEEE Access · doi:10.1109/access.2020.3009667 · Cited by 12
“A variety of factors individually or collectively contribute to the excessive false alarms including inappropriate system settings, suboptimal signal quality and deficiencies in proprietary algorithms [7].”
BackgroundSelf-citationThe Effect of Alarm Fatigue Nursing Management Protocol on Critical Care Nurses' Experience
Sliman, A. M. A., El-Aziz, W. W. A., & Mansour, H. E. (2020) · Egyptian Journal of Health Care · doi:10.21608/ejhc.2020.125756 · Cited by 8
“…organizations have scheduled alarm fatigue the sensory load and desensitization that nurses experience when exposed to an extreme amount of alarms as one of the headmost ten technology risks in acute care settings s (the Advancement of Medical Instrumentation Foundation, 2011, Drew et al., 2014).”
BackgroundThoracic Impedance Pneumography–Derived Respiratory Alarms and Associated Patient Characteristics
Bawua, L. K., Miaskowski, C. A., Suba, S., Badilini, F., Rodway, G. W., Hu, X., & Pelter, M. M. (2022) · American Journal of Critical Care · doi:10.4037/ajcc2022295 · Cited by 8
Self-citationHerráiz, Á. H., Martínez‐Rodrigo, A., Carneiro, D. R., Bertomeu‐González, V., Rieta, J. J., & Alcaraz, R. (2023) · IEEE Access · doi:10.1109/access.2023.3317793 · Cited by 9
“Indeed, these nui-sance interferences have been identified as the cause of more than 70% of false alarms of AF in conventional continuous ECG monitoring, both in intensive care units via bedside monitors [17] and in free-living conditions via insertable Holters [18].”
BackgroundDeveloping new predictive alarms based on ECG metrics for bradyasystolic cardiac arrest
Ding, Q., Bai, Y., Tinoco, A., MORTARA, D. W., Do, D. H., Boyle, N. G., Pelter, M. M., & Hu, X. (2015) · Physiological Measurement · doi:10.1088/0967-3334/36/12/2405 · Cited by 8
“It is estimated that the number of alarms per patient per day can reach several hundred (The Joint Commission, 2013), 80% - 99% of which are false positives and/or clinically insignificant and do not require clinical intervention (Lawless, 1994, Chambrin et al., 1999, Drew et al., 2014).”
BackgroundSelf-citationInpatient-Derived Vital Sign Parameters Implementation: An Initiative to Decrease Alarm Burden
Kipps, A. K., Poole, S. F., Slaney, C., Feehan, S., Longhurst, C., Sharek, P. J., & Goel, V. V. (2017) · PEDIATRICS · doi:10.1542/peds.2016-2458 · Cited by 8
Evaluating Alarm Classifiers with High-confidence Data Programming
Pugh, S., Ruchkin, I., Bonafide, C. P., DeMauro, S. B., Sokolsky, O. V., Lee, I., & Weimer, J. E. (2022) · ACM Transactions on Computing for Healthcare · doi:10.1145/3549942 · Cited by 8
“but nonzero probability of representing a critical patient need [5].”
BackgroundPelter, M. M., Xu, Y., Fidler, R., Xiao, R., MORTARA, D. W., & Hu, X. (2017) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2017.10.005 · Cited by 8
Self-citationJabin, M. S. R. (2026) · JMIR Human Factors · doi:10.2196/92560 · Cited by 9
“Centralized electronic surveillance systems are widely used in intensive and cardiac care units to support continuous physiological monitoring, early detection of deterioration, and retrospective review of patient data [1,2].”
Suliman, A., Farahmand, M., Galeotti, L., & Scully, C. G. (2024) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2024.3386493 · Cited by 7
“…of noise in ECGs can lead to various performance degradations for analysis algorithms such as false or incorrect detection of events or arrhythmias [9], [13], inaccurate heart rate and heart-rate variability estimations [14], and other adverse outcomes that can impact the diagnosis of relevant…”
Ventricular tachycardia and in-hospital mortality in the intensive care unit
Prasad, P. A., Isaksen, J. L., Abe‐Jones, Y., Zègre‐Hemsey, J. K., Sommargren, C. E., Al‐Zaiti, S. S., Carey, M. G., Badilini, F., MORTARA, D. W., Kanters, J. K., et al. (2023) · Heart Rhythm O2 · doi:10.1016/j.hroo.2023.09.008 · Cited by 7
“2,10,12,15,25 Potential consequences of alarm fatigue include the assimilation of alarm noise into a nurses ’ work fl ow with the potential for missing alarms, delayed response to alarms, and/or unsafe alarm adjustments (ie, lowering the vol-umeand/orcompletelysilencingalarms).”
BackgroundInfluentialSelf-citationPatient Safety Implications of Electronic Alerts and Alarms of Maternal–Fetal Status During Labor
Simpson, K. R., Lyndon, A., & Davidson, L. A. (2016) · Nursing for Women s Health · doi:10.1016/j.nwh.2016.07.004 · Cited by 7
Pelter, M. M., Loranger, D. L., Kozik, T. M., Fidler, R., Hu, X., & Carey, M. G. (2016) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2016.08.010 · Cited by 7
Self-citationAlarm Strategies and Surveillance for Mechanical Ventilation
Walsh, B. K., & Waugh, J. B. (2020) · Respiratory Care · doi:10.4187/respcare.07546 · Cited by 7
Suba, S., Hoffmann, T. J., Fleischmann, K. E., Schell-Chaple, H. M., Marcus, G. M., Prasad, P. A., Hu, X., Badilini, F., & Pelter, M. M. (2023) · Research in Nursing & Health · doi:10.1002/nur.22314 · Cited by 7
Self-citationTinoco, A., MORTARA, D. W., Hu, X., Sandoval, C. P., & Pelter, M. M. (2018) · Heart & Lung · doi:10.1016/j.hrtlng.2018.09.003 · Cited by 9
“A sophisticated research infrastructure was used to capture continuous ECG data for off-line analysis.(35) Briefly, continuous ECG data were captured, and then securely transferred using a specially developed research system (CARESCAPE Gateway system, GE Healthcare, Milwaukee, WI) to an external server for off-line an…”
MethodSelf-citationLessons Learned from Efforts to Reduce Overuse of Cardiac Telemetry Monitoring
Chakravarthy, R., Goggins, K. M., Leverenz, D. L., Trumbo, S. P., Kripalani, S. B., & Limper, H. M. (2020) · The Joint Commission Journal on Quality and Patient Safety · doi:10.1016/j.jcjq.2020.05.001 · Cited by 8
A descriptive investigation of alarm activation in a critical care setting
Nguyen, V. N. B., MacDonald, B., Cignarella, A., & Miller, C. (2025) · Nursing in Critical Care · doi:10.1111/nicc.13302 · Cited by 8
“In our study, ST-segment monitoring alarms were only activated in the acute cardiac unit, similar to the study by Drew et al. 38 which included five adult ICUs where only the cardiac ICU utilized ST-segment monitoring.”
InfluentialPrescription Use and Spending After the Introduction of a Real-Time Prescription Benefit Tool
Zink, A., Wehrly, D. J., Bozzi, D., Sutherland, A. J., Boudreau, E. C., McWilliams, J. M., & Chernew, M. E. (2025) · JAMA Network Open · doi:10.1001/jamanetworkopen.2025.19038 · Cited by 7
Choosing the Appropriate QRS Detector
Eilers, J., Chromik, J., & Bert Arnrich, A. (2021) · International Joint Conference on Biomedical Engineering Systems and Technologies (BIOSTEC) · doi:10.5220/0010234600002865 · Cited by 6
“If too many false alarms are raised, this causes nurses and doctors not responding to them anymore (Drew et al., 2014).”
BackgroundDeep learning classification of capnography waveforms: secondary analysis of the PRODIGY study
Conway, A., Rad, M. G., Zhou, W., Parotto, M., & Jungquist, C. R. (2023) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-023-01028-y · Cited by 6
Theelke, L., Metzler, F.-L., Kreimeier, J., Hauer, C., Binder, J., & Roth, D. (2023) · Proceedings - International Symposium on Mixed and Augmented Reality, ISMAR · doi:10.1109/ismar59233.2023.00034 · Cited by 7
“Furthermore, a majority of these alarms are false positive and do not require any clinical intervention [18].”
BackgroundGaleotti, L., & Scully, C. G. (2018) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2018.08.023 · Cited by 6
“When recorded signals from these devices are fed into automatic processing and classification algorithms, the presence of noise may result in degraded performance of the automatic algorithm leading to false alarms (3).”
BackgroundZègre‐Hemsey, J. K. (2020) · Journal of Emergency Nursing · doi:10.1016/j.jen.2020.01.007 · Cited by 7
“Alarm fatigue is a national patient safety hazard issue and occurs when nurses are desensitized by numerous alarms, which can lead potentially to patient injury or death.(21) In this issue of JEN, we learn about strategies to manage alarm fatigue in the emergency department.”
BackgroundSelf-citationNurses’ use of auditory alarms and alerts in high dependency units: A field study
Deschamps, M.-L., & Sanderson, P. M. (2021) · Applied Ergonomics · doi:10.1016/j.apergo.2021.103475 · Cited by 7
Intensive Care Unit False Alarm Identification Based on Convolution Neural Network
Yu, Q., Wang, C., Xi, J., Chen, Y., Li, W., Ge, Y., & Huang, X. (2021) · IEEE Access · doi:10.1109/access.2021.3086862 · Cited by 6
“According to previous research, nearly 90% of alarms in ICU are false due to artifacts and low signal-to-noise ratio [1]–[3].”
BackgroundNeed for a new paradigm in the design of alarms for patient monitors and medical devices
Webster, C. S., & Sanderson, P. M. (2021) · British Journal of Anaesthesia · doi:10.1016/j.bja.2021.08.001 · Cited by 6
Bawua, L. K., Miaskowski, C. A., Suba, S., Badilini, F., MORTARA, D. W., Hu, X., Rodway, G. W., Hoffmann, T. J., & Pelter, M. M. (2021) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2021.12.006 · Cited by 8
Self-citationSuba, S., Hoffmann, T. J., Fleischmann, K. E., Schell-Chaple, H. M., Prasad, P. A., Marcus, G. M., Badilini, F., Hu, X., & Pelter, M. M. (2022) · Journal of Clinical Nursing · doi:10.1111/jocn.16408 · Cited by 6
Self-citationKopitar, L., Fister, I., & Štiglic, G. (2024) · Information · doi:10.3390/info15030162 · Cited by 11
“Additionally, the F1-score, which balances precision and recall, is crucial in medical scenarios where both false positives [28] and false negatives [29] carry significant clinical implications.”
BackgroundMayer, C. M., Mauney, J. A., Barnes, J., Addison, A. J., Dicus, A. K., Hanlin, R., Hsieh, K. B., Soares-Sardinha, S., & White, B. L. (2016) · Proceedings of the International Symposium on Human Factors and Ergonomics in Health Care · doi:10.1177/2327857916051028 · Cited by 6
“Approaches to improving alarm safety in health care environments in the current literature include education (Phillips and Barnsteiner, 2005), equipment redesign recommendations (Drew et al, 2014), and optimizing alarm settings (Creighton and Cvach, 2010).”
BackgroundContinuous ECG monitoring in hospital: part 2, practical issues
Sampson, M. G. (2018) · British Journal of Cardiac Nursing · doi:10.12968/bjca.2018.13.3.128 · Cited by 7
Reducing Redundant Alarms in the Pediatric ICU
Dewan, M. L., Cipriani, L., Boyer, J., Stark, J. E., Seger, B., & Tegtmeyer, K. (2019) · Multimodal Technologies and Interaction · doi:10.3390/mti3010011 · Cited by 6
“Second, these approaches are often time consuming to complete on a large scale without substantial resources [19,21,22].”
BackgroundUncommon music making: The functional roles of music in design for healthcare
Özcan, E., Frankel, L., & Stewart, J. (2019) · Music and Medicine · doi:10.47513/mmd.v11i4.707 · Cited by 6
Suba, S., & Pelter, M. M. (2019) · Systematic Reviews · doi:10.1186/s13643-019-1168-4 · Cited by 11
The perils of modern telemetry: A human factors perspective
Korentsides, J. M., Miller, Z. N., Lazzara, E. H., Fernandez, R., & Keebler, J. R. (2025) · Human Factors in Healthcare · doi:10.1016/j.hfh.2025.100102 · Cited by 5
Role of Artificial Intelligence in Critical Care Medicine: A Literature Review
ABDELBAKY, A., Elmasry, W. G., Awad, A. H., & Khan, S. (2025) · Cureus · doi:10.7759/cureus.90149 · Cited by 14
“For example, one report found 88.8% of arrhythmia alarms were false positives [31].”
InfluentialThe impact of alarm sounds on patient sleep in the intensive care unit: A scoping review
Hiemstra, F. W., van Kekem, M. J., Arbous, M. S., Schoones, J. W., van Westerloo, D. J., & Kervezee, L. (2026) · Journal of Critical Care · doi:10.1016/j.jcrc.2026.155612 · Cited by 5
Clinical Alarms in a Gynaecological Surgical Unit: A Retrospective Data Analysis
Jämsä, J. O., Uutela, K., Tapper, A., & Lehtonen, L. (2023) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph20054193 · Cited by 5
“It is hypothesised that even inaudible technical alarms are relevant to alarm fatigue [13].”
BackgroundThe association between alarm burden and nurse burnout in U.S. hospitals
Ruppel, H., Dougherty, M., Kodavati, M., & Lasater, K. B. (2024) · Nursing Outlook · doi:10.1016/j.outlook.2024.102288 · Cited by 9
Xiao, R. (2024) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2024.153861 · Cited by 5
Cumulative Time Series Representation for Code Blue prediction in the Intensive Care Unit.
Salas-Boni, R., Bai, Y., & Hu, X. (2015) · Europe PMC (PubMed Central) · Cited by 5
Self-citationClinician-in-the-Loop Annotation of ICU Bedside Alarm Data
Roederer, A., Dimartino, J., Gutsche, J. T., Mullen‐Fortino, M., Shah, S. S., Hanson, C. W., & Lee, I. (2016) · IEEE/ACM Conference on Connected Health: Applications, Systems and Engineering Technologies · doi:10.1109/chase.2016.10 · Cited by 5
“The large number of erroneous alarms produced by medical devices in the ICU have been shown to cause clinicians to suffer from alarm fatigue , a desensitization to the presence of these alarms that causes clinicians to lose trust in the systems and begin ignoring them [9], [10].”
Sawhney, R., Joshi, H., Nobles, A. L., & Shah, R. R. (2021) · Proceedings of the International AAAI Conference on Web and Social Media · doi:10.1609/icwsm.v15i1.18088 · Cited by 10
Tweet Classification to Assist Human Moderation for Suicide Prevention.
Sawhney, R., Joshi, H., Nobles, A. L., & Shah, R. R. (2021) · PubMed · Cited by 9
Predict In-Hospital Code Blue Events using Monitor Alarms through Deep Learning Approach
Xiao, R., King, J., Villaroman, A., Do, D. H., Boyle, N. G., & Hu, X. (2018) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2018.8513269 · Cited by 7
“However, current bedside monitors are known to produce excessive false alarms that can desensitize caregivers, causing alarm fatigue [1-2].”
Self-citationPelter, M. M., MORTARA, D. W., & Badilini, F. (2021) · Hearts · doi:10.3390/hearts2040036 · Cited by 5
“We found that many of these patients often have ECG features that contribute to false arrhythmia alarms, including: bundle branch block (BBB), ventricular paced rhythms, and low amplitude QRS complexes [3,23,24].”
BackgroundMethodResultInfluentialModlin, I. M., Kidd, M. S., Drozdov, I. A., & Bodei, L. (2019) · Neuroendocrinology · doi:10.1159/000504605 · Cited by 6
Mendoza, A., Tume, S., Puri, K., Acosta, S., & Cavallaro, J. R. (2025) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2024.3510217 · Cited by 4
Klopfenstein, S. A. I., Flint, A. R., Heeren, P., Prendke, M., Chaoui, A., Ocker, T., Chromik, J., Arnrich, B., Balzer, F., & Poncette, A.-S. (2025) · Journal of Medical Internet Research · doi:10.2196/65961 · Cited by 6
Exploring intensive care unit nurses’ decision-making in alarm management: A qualitative study
Haskell, N., Banner, D., Bouchal, S. R., & King‐Shier, K. M. (2025) · Heart & Lung · doi:10.1016/j.hrtlng.2025.09.013 · Cited by 6
Boulanger, P. (2026) · Sensors · doi:10.3390/s26041135 · Cited by 4
“Studies of intensive care unit (ICU) alarms reveal that 85–99% of arrhythmia alerts are false positives, primarily caused by patient movement, electrode displacement, and electromyographic interference [3,4].”
Continuous Cardiac Arrest Prediction in ICU using PPG Foundation Model
Kataria, S., Xiao, R., Ruchti, T. L., Clark, M. P. A., Lu, J., Lee, R. J., Grunwell, J. R., & Hu, X. (2025) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc58623.2025.11254649 · Cited by 4
“However, this causes issues such as alarm fatigue [18], [19].”
Self-citationZhang, X., & Ding, Q. (2016) · Asilomar Conference on Signals, Systems and Computers (ACSSC) · doi:10.1109/acssc.2016.7869656 · Cited by 4
“repeatable [6] and the monitored respiratory waveform often appears flat in ICU patients who are breathing adequately [7].”
BackgroundSelf-citationClinical decision support during electronic prescribing for elderly patients
Poeran, J., Gorbenko, K. O., Babby, J., Mazumdar, M., & Leipzig, R. M. (2017) · American Journal of Health-System Pharmacy · doi:10.2146/ajhp160942 · Cited by 4
Hospital-Based Electrocardiographic Monitoring: The Good, the Not So Good, and Untapped Potential
Pelter, M. M. (2024) · American Journal of Critical Care · doi:10.4037/ajcc2024781 · Cited by 4
High-Confidence Data Programming for Evaluating Suppression of Physiological Alarms
Pugh, S., Ruchkin, I., Bonafide, C. P., DeMauro, S. B., Sokolsky, O. V., Lee, I., & Weimer, J. E. (2021) · IEEE/ACM Conference on Connected Health: Applications, Systems and Engineering Technologies · doi:10.1109/chase52844.2021.00015 · Cited by 4
“Alarm fatigue is a serious and well-known problem of physiological monitors [1], [2].”
BackgroundAI-Based Internet of Things (AIoT): Applications of AI With IoT
Singampalli, D. T., & Pise, A. A. (2023) · Advances in computational intelligence and robotics book series · doi:10.4018/978-1-6684-6519-6.ch007 · Cited by 6
Ismail, A., Imam, A., Pingue-Raguini, M., Hassan, D., Ali, A., & Alkhotani, A. (2023) · Cureus · doi:10.7759/cureus.50307 · Cited by 5
“However, 90% of monitor alarms are false positives [18].”
BackgroundFalse alarm suppression in early prediction of cardiac arrhythmia
Roychoudhury, S., Ghalwash, M. F., & Zoran Obradovic, A. (2015) · IEEE International Conference on Bioinformatics and Bioengineering (BIBE) · doi:10.1109/bibe.2015.7367628 · Cited by 4
“lead to lower standards of care to patients and also result in fatal consequences [7].”
BackgroundPhysiologic Monitoring: Improving Safety or Increasing Risk?
Durbin, C. G. (2016) · Respiratory Care · doi:10.4187/respcare.04931 · Cited by 4
How to Annotate Patient Monitoring Alarms in Intensive Care Medicine for Machine LearningPreprint
Klopfenstein, S. A. I., Flint, A. R., Heeren, P., Prendke, M., Chaoui, A., Ocker, T., Chromik, J., Arnrich, B., Balzer, F., & Poncette, A.-S. (2023) · Research Square · doi:10.21203/rs.3.rs-2514969/v1 · Cited by 4
Educational Frameworks for Diagnostic Decision-Making in AI-Enhanced Head and Neck Pathology
Tavares, T., Sangalli, L., Menon, R. S., Miller, C. S., & Hochman, G. (2026) · Head and Neck Pathology · doi:10.1007/s12105-026-01909-z · Cited by 5
“[ 42 , 48] Decision Audits and Reflection revised after consulting AI suggestions, learners can identify tendencies toward automation bias, algorithm aversion (loss of confidence, distrust, or negative bias towards AI outputs), or opacity of models [ 61 , 67 , 68 ].”
Smith, D. H. (2017) · Cited by 3
Hardy, M., Dallaire, C., Bouchlaghem, M. A., & Hajji, I. (2023) · Nursing Open · doi:10.1002/nop2.1835 · Cited by 3
Fang, J., Xiong, C., Xia, S., & Ding, P. (2025) · Intensive and Critical Care Nursing · doi:10.1016/j.iccn.2025.104192 · Cited by 3
Navigating Communication: Crafting Guidelines for Epic Secure Chat in an Academic Medical Center
Howard-Williams, E. L., Dreesen, E. B., Downs, J., Schiff, L. D., Dangerfield, C., & Mock, C. (2025) · Quality Management in Health Care · doi:10.1097/qmh.0000000000000547 · Cited by 3
Research in Review: Driving Critical Care Practice Change
Bridges, E. J., McNeill, M. M., & Munro, N. (2016) · American Journal of Critical Care · doi:10.4037/ajcc2016564 · Cited by 3
“The analysis identified 6 major themes: (1) dissonance and desensitization, with respondents referring to a “cry wolf” situation where the nurse no longer trusts the alarm; (2) pollution, panic, and pathology, which reflects the effect of the noise on the patient; (3) calling for accountability at all levels of the pr…”
BackgroundStaggers, N., Elias, B. L., Makar, E. V., Hunt, J., & Alexander, G. L. (2016) · Lecture notes in computer science · doi:10.1007/978-3-319-40250-5_20 · Cited by 3
Asadi, F., Mollakazemi, M. J., Ghiasi, S., & Sadati, S. H. (2016) · Computing in cardiology · doi:10.22489/cinc.2016.019-449 · Cited by 3
“balancing information are used by clinicians to recognize these alarms [4].”
BackgroundDrawing Up a New Game Plan to Reduce Alarm Fatigue
Venella, J. (2017) · Biomedical Instrumentation & Technology · doi:10.2345/0899-8205-51.s2.71 · Cited by 3
Block, F. E. (2017) · Journal of Clinical Engineering · doi:10.1097/jce.0000000000000248 · Cited by 3
Types and Frequency of Infusion Pump Alarms: Protocol for a Retrospective Data Analysis
Glover, K. R., Vitoux, R. R., Schuster, C., & Curtin, C. R. (2018) · JMIR Research Protocols · doi:10.2196/10446 · Cited by 4
“Research has primarily focused on electrocardiograms (ECGs), physiologic monitors, and pulse oximetry in the telemetry and ICUs, where the total alarm incidences of these devices is thought to be highest [1,5,12].”
BackgroundReducing Infusion Pump Alarms Through Structured Interventions
Matocha, D. (2018) · Journal of the Association for Vascular Access · doi:10.1016/j.java.2018.03.002 · Cited by 3
A Comprehensive Medical Equipment Management Software System for Increased Patient Safety
Neslisah, A. K. A. R., Ülgen, Y., & Öztürk-Işık, E. (2019) · IEEE International Symposium on Medical Measurements and Applications (MeMeA) · doi:10.1109/memea.2019.8802156 · Cited by 3
“From literature, the false or unactionable alarms have the highest ratio of 88-90% in the total clinical alarms [23].”
BackgroundFitzgerald, M. C., Rayo, M. F., Reynolds, M. E., Abdel‐Rasoul, M., & Moffatt, S. D. (2019) · Proceedings of the International Symposium on Human Factors and Ergonomics in Health Care · doi:10.1177/2327857919081015 · Cited by 4
Fayyaz, B., & Rehman, H. J. (2020) · Journal of Community Hospital Internal Medicine Perspectives · doi:10.1080/20009666.2020.1759763 · Cited by 4
“Furthermore, this trend of overutilizing TM was associated with harmful effects such as alarm fatigue and patient distress [3] while overburdening an already strained healthcare system and wasting manpower [4].”
BackgroundFalse Alarm Rejection for ICU ECG Monitoring
Dai, J., Sun, Z., & He, X. (2020) · Springer eBooks · doi:10.1007/978-981-15-3824-7_12 · Cited by 3
Cabrera-Quirós, L., Varisco, G., Zhan, Z. Z., Long, X., Andriessen, P., Cottaar, E. J. E., & van Pul, C. (2021) · Companion Publication of the International Conference on Multimodal Interaction (ICMI Companion) · doi:10.1145/3461615.3485435 · Cited by 5
“Other reasons behind the lack of response for some alarms include self-patient stabilization, a large number of false or irrelevant alarms (up to 70% of total number of alarms delivered to caregivers), or work overload in the medical team [7, 13, 17].”
BackgroundResultInfluentialAn Introduction to Big Data Analytics Techniques in Healthcare
Pise, A. A. (2022) · Wiley eBooks · doi:10.1002/9781119792376.ch1 · Cited by 3
Ergın, E., Yaz, Ş. B., & Atay, A. (2023) · Turkiye Klinikleri Journal of Nursing Sciences · doi:10.5336/nurses.2022-94490 · Cited by 3
Cosentino, N., Zhang, X., Farrar, E. J., Yapici, H. O., Coffeng, R., Vaananen, H., & Beard, J. W. (2024) · Cardiovascular Digital Health Journal · doi:10.1016/j.cvdhj.2024.02.002 · Cited by 3
“2,6 Consequently, health care personnel are exposed to a high volume of physiological monitor alarms throughout their shifts, resulting in an excessively noisy environment that is perceived as the normal working atmosphere of a hospital.”
BackgroundMinimizing Alarm Fatigue: Pediatric Perspective
Williams, L. (2018) · AACN Advanced Critical Care · doi:10.4037/aacnacc2018946 · Cited by 3
Liu, X., Gadhoumi, K., Xiao, R., Tran, N., Smielewski, P. S., Czosnyka, M., Hetts, S. W., Ko, N., & Hu, X. (2019) · Physiological Measurement · doi:10.1088/1361-6579/aafab1 · Cited by 3
“All data were recorded through BedMasterEx software (Excel Medical Electronics, Inc., Jupiter, FL) at a sampling rate of 240 Hz (Drew et al 2014).”
MethodSelf-citationRole of nurses' electrocardiography competency in emergent situations
Iacobescu, R.-A., Corneanu, L. E., Dimache, A.-M., Florescu, B.-D., & Coman, A. E. (2023) · Romanian Journal of Medical Practice · doi:10.37897/rjmp.2023.3.5 · Cited by 2
“There are concerns about alarm fatigue and de-sensitization of nurses related to prolonged monitoring of patients which leads to failure in recognition of real life-threatening conditions [17].”
BackgroundThomas, O. D., Linnér, R., & Dardashti, A. (2025) · European Heart Journal - Digital Health · doi:10.1093/ehjdh/ztaf059 · Cited by 2
Chiziwa, C., Kamndaya, M. S., Phepa, P., IMPALA Study Team, Vweza, A. O., Calis, J. C. J., & Bierling, B. (2025) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-025-01311-0 · Cited by 4
“Most of these alarms are triggered when physiological variables go over a predetermined threshold [11].”
Keenan, L., Younger, R. M., Race, P. R., & Bawn, M. (2025) · Emerging Topics in Life Sciences · doi:10.1042/etls20253019 · Cited by 2
“However, their low positive rate was a direct consequence of the limited representation of these individuals in the original study upon which the AI model was based [12,57].”
Kraevsky-Phillips, K., Aqtash, S., Teh, F. C.-E., Pinsky, M. R., Clermont, G., Hravnak, M., & Al‐Zaiti, S. S. (2026) · Scientific Reports · doi:10.1038/s41598-026-43028-3 · Cited by 2
“Our findings for the ICUs were higher than those previously reported by Poncette 10 and Drew 13 , who determined there were 152 and 187 alarms per bed per day, respectively, which could be due to variability in hospital and unit practices, as well as more comprehensive nature of longitudinal data…”
InfluentialQin, C., Lin, K., & Huang, C. (2026) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2026.110127 · Cited by 2
Vrettou, C. S., Koliotsis, P.-T. A., Golemati, S., Mastorakis, G., Karaviti, V., Mavromati, S., Tagara, M., Papadopoulou, M. P., & Dimopoulou, I. (2026) · Critical Care Research and Practice · doi:10.1155/ccrp/4782724 · Cited by 2
Boyd, E., Lynch, M., COATES, V. E., Vanharen, Y., & Chamos, A. J. (2026) · Nurse Education in Practice · doi:10.1016/j.nepr.2026.104856 · Cited by 4
Ferreira, V. C., Carrajola, N., & Catalão, M. J. (2026) · Healthcare · doi:10.3390/healthcare14172720 · Cited by 3
“Optimization of monitoring devices and sensors is an important component of alarm management because technical artifacts and poor signal acquisition contribute to nonac-tionable alarms [6,22,24,34].”
InfluentialDeveloping a Clinical Alarms Management Committee at an Academic Medical Center
Pelter, M. M., Stotts, J. R., Spolini, K., Nguyen, J., Sin, E., & Hu, X. (2017) · Biomedical Instrumentation & Technology · doi:10.2345/0899-8205-51.s2.21 · Cited by 2
Self-citationDaluwatte, C., Vicente, J., Galeotti, L., Johannesen, L., Strauss, D. G., & Scully, C. G. (2017) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2017.08.030 · Cited by 2
“However, heart rate limit alarms related to ECG beat detection have been shown to account for a notable number of false patient monitoring alarms [1,2].”
BackgroundGüder, G., von Rein, E., Flohr, T., Weismann, D., Schmitt, D. R., Störk, S., Frantz, S., Kratzer, V., & Kendi, C. (2023) · Applied Sciences · doi:10.3390/app13169319 · Cited by 2
“, the reaction time of the responsible nurse decreases [5,6].”
Background
Metadata
- Authors
- Drew, B. J., Harris, P., Zègre-Hemsey, J. K., Mammone, T., Schindler, D., Salas-Boni, R., Bai, Y., Tinoco, A., Ding, Q., & Hu, X.
- Affiliations
- University of California, San Francisco (US)
- University of California San Francisco Medical Center (US)
- Corresponding author
- Drew, Barbara J.
- Venue
- PLoS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2014-10-22
- Volume
- 9
- Issue
- 10
- Pages
- e110274
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC4206416
- OpenAlex
- W2001741247
- Cited by
- 438
- References
- 23
- Citations by year
- 2026: 51 · 2025: 31 · 2024: 25 · 2023: 37 · 2022: 26 · 2021: 58 · 2020: 29 · 2019: 38 · 2018: 42 · 2017: 36 · 2016: 43 · 2015: 21 · 2014: 1
- FWCI
- 33.15
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- PLoS ONE (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Carolina Digital Repository (University of North Carolina at Chapel Hill) (submittedVersion)
- Dominican Scholar (Dominican University of California) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- eScholarship (California Digital Library) (submittedVersion)
- UNC Libraries
- Funders
- University of California, San Francisco
- Topics
- Health Sciences › Medicine › Surgery › Healthcare Technology and Patient Monitoring
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Patient Safety and Medication Errors
- Health Sciences › Health Professions › Health Information Management › Electronic Health Records Systems
- Keywords
- alarm fatigue, intensive care unit patients, false positive alerts, alarm settings, ECG leads, QRS amplitude, atrial fibrillation, ventricular tachycardia, bundle branch block, persistent atrial fibrillation, patient safety
- MeSH terms
- Adult, Aged, Arrhythmias, Cardiac, Atrial Fibrillation, Diagnostic Errors, Electrocardiography, Equipment Failure, False Positive Reactions, Female, Humans, Intensive Care Units, Male, Middle Aged, Monitoring, Physiologic, Critical Illness, Clinical Alarms
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.