Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients
Open accessThe DOI matches its Crossref record; 23 of the 23 references it cites were found in a registry or an index.
Abstract
PURPOSE: Physiologic monitors are plagued with alarms that create a cacophony of sounds and visual alerts causing "alarm fatigue" which creates an unsafe patient environment because a life-threatening event may be missed in this milieu of sensory overload. Using a state-of-the-art technology acquisition infrastructure, all monitor data including 7 ECG leads, all pressure, SpO(2), and respiration waveforms as well as user settings and alarms were stored on 461 adults treated in intensive care units. Using a well-defined alarm annotation protocol, nurse scientists with 95% inter-rater reliability annotated 12,671 arrhythmia alarms. RESULTS: A total of 2,558,760 unique alarms occurred in the 31-day study period: arrhythmia, 1,154,201; parameter, 612,927; technical, 791,632. There were 381,560 audible alarms for an audible alarm burden of 187/bed/day. 88.8% of the 12,671 annotated arrhythmia alarms were false positives. Conditions causing excessive alarms included inappropriate alarm settings, persistent atrial fibrillation, and non-actionable events such as PVC's and brief spikes in ST segments. Low amplitude QRS complexes in some, but not all available ECG leads caused undercounting and false arrhythmia alarms. Wide QRS complexes due to bundle branch block or ventricular pacemaker rhythm caused false alarms. 93% of the 168 true ventricular tachycardia alarms were not sustained long enough to warrant treatment. DISCUSSION: The excessive number of physiologic monitor alarms is a complex interplay of inappropriate user settings, patient conditions, and algorithm deficiencies. Device solutions should focus on use of all available ECG leads to identify non-artifact leads and leads with adequate QRS amplitude. Devices should provide prompts to aide in more appropriate tailoring of alarm settings to individual patients. Atrial fibrillation alarms should be limited to new onset and termination of the arrhythmia and delays for ST-segment and other parameter alarms should be configurable. Because computer devices are more reliable than humans, an opportunity exists to improve physiologic monitoring and reduce alarm fatigue.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures











![Figure 12 True accelerated ventricular rhythm alarm showing why this arrhythmia is not considered an “actionable” alarm condition. Accelerated ventricular rhythm at a rate of 56 for the first 5 beats followed by 2 fusion beats; the last 2 beats are normal sinus rhythm. The invasive arterial pressure waveform (bottom tracing) shows no change between normal rhythm and accelerated ventricular rhythm which confirms the rationale for the published guidelines stating no treatment is indicated for this arrhythmia in hospital settings [5].](https://figures.cochat.ai/paper-figures/440dfa9937e3409b986af268dec77e1a/insights-into-the-problem-of-alarm-fatigue-with-physiologic-monitor-devices-a-figure-12.jpg)




Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“One study conducted a comprehensive observational analysis over 31 days across five ICUs, involving 461 adults and monitoring over 2.5 million alarm instances. Of 1,154,201 unique alarms, 88.8% were for arrhythmias with a majority being false positive98.”
From Bias recognition and mitigation strategies in artificial intelligence healthcare applications (2025) · cites this paper as [98]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The Cardiac Arrhythmia Suppression Trial (CAST) Investigators (1989) · New England Journal of Medicine · doi:10.1056/nejm198908103210629 · Cited by 3,300
Third Universal Definition of Myocardial Infarction
Thygesen, K., Alpert, J. S., Jaffe, A. S., Simoons, M. L., Chaitman, B. R., & White, H. D. (2012) · Circulation · doi:10.1161/cir.0b013e31826e1058 · Cited by 2,985
Third universal definition of myocardial infarction
the Writing Group on behalf of the Joint ESC/ACCF/AHA/WHF Task Force for the Universal Definition of Myocardial Infarction, Thygesen, K., Alpert, J. S., Jaffe, A. S., Simoons, M. L., Chaitman, B. R., & White, H. D. (2012) · Nature Reviews Cardiology · doi:10.1038/nrcardio.2012.122 · Cited by 2,053
Zipes, D. P., Camm, A. J., Borggrefe, M., Buxton, A. E., Chaitman, B., Fromer, M., Gregoratos, G., Klein, G., Myerburg, R. J., Quinones, M. A., et al. (2006) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2006.07.010 · Cited by 1,973
Weir, M. (1990) · Cardiopulmonary Physical Therapy Journal · doi:10.1097/01823246-199001020-00009 · Cited by 1,058
The association for the advancement of medical instrumentation
(1974) · Ultrasound in Medicine & Biology · doi:10.1016/0301-5629(74)90035-0 · Cited by 1,031
Drew, B. J., Ackerman, M. J., Funk, M., Gibler, W. B., Kligfield, P., Menon, V., Philippides, G. J., Roden, D. M., & Zareba, W. (2010) · Circulation · doi:10.1161/circulationaha.109.192704 · Cited by 821
Self-citationAuthors/Task Force Members (2006) · European Heart Journal · doi:10.1093/eurheartj/ehl199 · Cited by 523
ACC/AHA guidelines for ambulatory electrocardiography
Crawford, M. H., Bernstein, S. J., Deedwania, P. C., DiMarco, J. P., Ferrick, K. J., Garson, A., Green, L. A., Greene, H. L., Silka, M. J., Stone, P. H., et al. (1999) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(99)00354-x · Cited by 309
Improving Alarm Performance in the Medical Intensive Care Unit Using Delays and Clinical Context
Görges, M., Markewitz, B. A., & Westenskow, D. R. (2009) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e31819bdfbb · Cited by 211
Cited by
Items 101–200 of 457
Safety analysis of proposed data‐driven physiologic alarm parameters for hospitalized children
Goel, V. V., Poole, S. F., Longhurst, C., Platchek, T. S., Pageler, N. M., Sharek, P. J., & Palma, J. P. (2016) · Journal of Hospital Medicine · doi:10.1002/jhm.2635 · Cited by 25
Boussi Rahmouni, H., Hassine, N. B. E. H., Chouchen, M., Ceylan, H. İ., Muntean, R. I., Bragazzi, N. L., & Dergaa, I. (2025) · Healthcare · doi:10.3390/healthcare13202553 · Cited by 25
“It substantially reduces clinician responsiveness when false positive rates exceed acceptable thresholds [16], while integration disruptions can compromise established care processes.”
An Evidence-Based Initiative to Reduce Alarm Fatigue in a Burn Intensive Care Unit
Gorisek, R., Mayer, C. M., Hicks, W. B., & Barnes, J. (2021) · Critical Care Nurse · doi:10.4037/ccn2021166 · Cited by 25
Customizing Alarm Limits Based on Specific Needs of Patients
Cvach, M. M., Kitchens, M., Smith, K. N., Harris, P. R. E., & Flack, M. N. (2017) · Biomedical Instrumentation & Technology · doi:10.2345/0899-8205-51.3.227 · Cited by 28
“Where possible, adding a short delay before an alarm is enunciated may be desirable.(4) For example, an ST alarm may signal a lack of oxygen to the heart.”
Self-citationMathis, M. R., Engoren, M. C., Williams, A. M., Biesterveld, B. E., Croteau, A. J., Cai, L., Kim, R. B., Liu, G., Ward, K. R., Najarian, K., et al. (2022) · Anesthesiology · doi:10.1097/aln.0000000000004345 · Cited by 24
Au-Yeung, W.-T. M., Sevakula, R. K., Sahani, A. K., Kassab, M. B., Boyer, R. B., Isselbacher, E. M., & Armoundas, A. A. (2021) · European Heart Journal - Digital Health · doi:10.1093/ehjdh/ztab058 · Cited by 25
Joseph, A. L., Borycki, E. M., Kushniruk, A. W., & |, |. (2021) · Knowledge Management & E-Learning An International Journal · doi:10.34105/j.kmel.2021.13.027 · Cited by 35
Zhang, X., & Ding, Q. (2017) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2017.02.003 · Cited by 24
Self-citationA contrastive learning approach for ICU false arrhythmia alarm reduction
Zhou, Y., Zhao, G., Li, J., Sun, G., Qian, X., Moody, B., Mark, R. G., & Lehman, L.-W. H. (2022) · Scientific Reports · doi:10.1038/s41598-022-07761-9 · Cited by 23
“(2) FCN We use a fully-connected convolutional network as the feature extractor of the input waveform.”
MethodRespiratory Monitoring During Mechanical Ventilation: The Present and the Future
Roshdy, A. (2023) · Journal of Intensive Care Medicine · doi:10.1177/08850666231153371 · Cited by 25
“Unfortunately, this culminated to what is called “alarm fatigue” which refers to the reduced attention and inability to recognize true from false alarms risking patients’ harm.(48) Algorithms were then introduced to allow automatic adaptation of MV through a closed loop feedback serving also the goal of an intelligent…”
BackgroundDo, D. H., Kuo, A., Lee, E. S., MORTARA, D. W., Elashoff, D. A., Hu, X., & Boyle, N. G. (2019) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2019.06.032 · Cited by 26
Self-citationDarbyshire, J. L., & Young, D. (2021) · Journal of the Intensive Care Society · doi:10.1177/17511437211022127 · Cited by 22
Frequency, duration and cause of ventilator alarms on a neonatal intensive care unit
Bélteki, G., & Morley, C. J. (2017) · Archives of Disease in Childhood Fetal & Neonatal · doi:10.1136/archdischild-2017-313493 · Cited by 22
Challenges of customizing electrocardiography alarms in intensive care units: A mixed methods study
Ruppel, H., Funk, M., Kennedy, H. P., Bonafide, C. P., Wung, S., & Whittemore, R. (2018) · Heart & Lung · doi:10.1016/j.hrtlng.2018.06.010 · Cited by 24
Understanding heart rate alarm adjustment in the intensive care units through an analytical approach
Fidler, R., Pelter, M. M., Drew, B. J., Arroyo-Palacios, J., Bai, Y., Stannard, D., Aldrich, J. M., & Hu, X. (2017) · PLoS ONE · doi:10.1371/journal.pone.0187855 · Cited by 22
“This study is a secondary, quantitative, descriptive analysis of HR alarms using the UCSF Alarm Study data from five adult ICUs at a quaternary care hospital [6].”
BackgroundMethodSelf-citationCritical Care Nurses’ Cognitive Ergonomics Related to Medical Device Alarms
Wung, S., & Schatz, M. R. (2018) · Critical Care Nursing Clinics of North America · doi:10.1016/j.cnc.2018.02.002 · Cited by 23
Voscopoulos, C. J., Theos, K., Hein, H. A. T., & George, E. (2016) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-016-9841-9 · Cited by 21
Galvagno, S. M., Duke, P. G., Eversole, D., & George, E. (2016) · The Journal of Trauma: Injury, Infection, and Critical Care · doi:10.1097/ta.0000000000001152 · Cited by 20
Acoustic Biotopes, Listeners and Sound-Induced Action: A Case Study of Operating Rooms
Özcan, E., Broekmeulen, C. L. H., Luck, Z. A., van Velzen, M., Stappers, P. J., & Edworthy, J. R. (2022) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph192416674 · Cited by 29
Classifying Alarms: Seeking Durability, Credibility, Consistency, and Simplicity
Edworthy, J. R., Schlesinger, J. J., McNeer, R. R., Kristensen, M. S., & Bennett, C. (2017) · Biomedical Instrumentation & Technology · doi:10.2345/0899-8205-51.s2.50 · Cited by 20
Alarm fatigue in the Intensive Care Unit: Relevance and response time
Méndez, B. A., Torres, D. O. A., & Gómez-Tovar, L. O. (2020) · Enfermería Intensiva (English ed ) · doi:10.1016/j.enfie.2019.11.001 · Cited by 20
Chowdhury, M. H., Shuzan, M. N. I., Chowdhury, M. E. H., Reaz, M. B. I., Mahmud, S., Emadi, N. A., Ayari, M. A., Ali, S. H. M., Bakar, A. A. A., Rahman, S. M., et al. (2022) · Bioengineering · doi:10.3390/bioengineering9100558 · Cited by 20
“However, in a study [28], it has been observed that, in some ICU patients, the respiratory signals extracted from ECG were not accurate, despite the patients still breathing.”
BackgroundGaleotti, L., Scully, C. G., Vicente, J., Johannesen, L., & Strauss, D. G. (2015) · Physiological Measurement · doi:10.1088/0967-3334/36/8/1705 · Cited by 20
Cardiovascular nurses' adherence to practice standards in in‐hospital telemetry monitoring
Fålun, N., Oterhals, K., Pettersen, T. R., Brørs, G., Olsen, S. S., Norekvål, T. M., & On behalf of the TELMON‐NOR investigators (2019) · Nursing in Critical Care · doi:10.1111/nicc.12425 · Cited by 22
An algorithm strategy for precise patient monitoring in a connected healthcare enterprise
Hu, X. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0107-z · Cited by 23
Self-citationFujita, L. Y., & Choi, S. Y. (2019) · Journal of Emergency Nursing · doi:10.1016/j.jen.2019.10.017 · Cited by 18
Cost Sensitive Time-Series Classification
Roychoudhury, S., Ghalwash, M. F., & Obradovic, Z. (2017) · Lecture notes in computer science · doi:10.1007/978-3-319-71246-8_30 · Cited by 22
“rates cause desensitization among care providers, thus risking patients’ lives [7].”
BackgroundA Holistic and Collaborative Approach to Audible Alarm Design
Özcan, E., Birdja, D., & Edworthy, J. R. (2018) · Biomedical Instrumentation & Technology · doi:10.2345/0899-8205-52.6.422 · Cited by 22
Double Trouble: Patients With Both True and False Arrhythmia Alarms
Nguyen, S. C., Suba, S., Hu, X., & Pelter, M. M. (2020) · Critical Care Nurse · doi:10.4037/ccn2020363 · Cited by 17
Self-citationChen, S., Palchaudhuri, S., Johnson, A. E., Trost, J., Ponor, I. L., & Zakaria, S. (2017) · Journal of Evaluation in Clinical Practice · doi:10.1111/jep.12708 · Cited by 17
Alsaad, A. A., Alman, C. R., Thompson, K. M., Park, S. H., Monteau, R. E., & Maniaci, M. J. (2017) · Postgraduate Medical Journal · doi:10.1136/postgradmedj-2016-134764 · Cited by 18
Alarm in the ICU! Envisioning Patient Monitoring and Alarm Management in Future Intensive Care Units
Özcan, E., Birdja, D., Simonse, L. W. L., & Struijs, A. (2018) · Springer eBooks · doi:10.1007/978-3-030-00749-2_24 · Cited by 17
Li, D., Vaidya, J., Wang, M. C., Bush, B., Lu, C., Kollef, M. H., & Bailey, T. C. (2020) · ACM Transactions on Computing for Healthcare · doi:10.1145/3344256 · Cited by 21
“• In clinical practice, excessive false alarms generated by a clinical warning system can cause nurses to ignore important alarms [14].”
BackgroundSinno, Z.-C., Shay, D., Kruppa, J., Klopfenstein, S. A. I., Giesa, N., Flint, A. R., Herren, P., Scheibe, F., Spies, C., Hinrichs, C., et al. (2022) · Scientific Reports · doi:10.1038/s41598-022-26261-4 · Cited by 18
When will less monitoring and diagnostic testing benefit the patient more?
Zampieri, F. G., & Einav, S. (2019) · Intensive Care Medicine · doi:10.1007/s00134-019-05715-w · Cited by 16
Team-Based Intervention to Reduce the Impact of Nonactionable Alarms in an Adult Intensive Care Unit
Yeh, J., Wilson, R., Young, L., Pahl, L., Whitney, S., Dellsperger, K. C., & Schafer, P. E. (2019) · Journal of Nursing Care Quality · doi:10.1097/ncq.0000000000000436 · Cited by 17
“The most frequent alarms considered nonactionable (eg, PVCs, Spo2, and heart rate) remained consistent, both internally in this study and with previous data gathered from the intensive care setting.(27) Therefore, applying the alarm setting changes used in this study to ICUs at other similar institutions would likely…”
ResultSowan, A. K., Staggers, N., Berndt, A., Austin, T., Reed, C. C., Malshe, A. V., Kilger, M., Fonseca, E., Vera, A., & Chen, Q. (2021) · JMIR Nursing · doi:10.2196/20584 · Cited by 16
“5%) from physiologic monitors leads to a phenomenon called alarm fatigue [2-8], which further results in nurses having to respond to an average of 150-400 alarms per patient per day in intensive care units (ICUs) [9] and, more startlingly, ignoring alarms or inappropriately turning off alarms [10].”
BackgroundNatural Brain-Inspired Intelligence for Screening in Healthcare Applications
Naghshvarianjahromi, M., Majumder, S., Kumar, S., Naghshvarianjahromi, N., & Deen, M. J. (2021) · IEEE Access · doi:10.1109/access.2021.3077529 · Cited by 17
“These diagnosis errors are achieved with acceptable false alarm rates [19] of 20.1%, 25%, 28.4%, and 54.7%, respectively.”
BackgroundInfluentialThe Influence of Audible Alarm Loudness and Type on Clinical Multitasking
Bruder, A. L., Rothwell, C. D., Fuhr, L. I., Shotwell, M. S., Edworthy, J. R., & Schlesinger, J. J. (2021) · Journal of Medical Systems · doi:10.1007/s10916-021-01794-9 · Cited by 17
“Very few alarms signal immediately life-threatening situations [48, 52], therefore strongly attention-capturing and attention-focusing alarms may be largely unnecessary in clinical practice.”
BackgroundA Machine Learning Methodology for Identification and Triage of Heart Failure Exacerbations
Morrill, J., Qirko, K., Kelly, J. P., Ambrosy, A. P., Toro, B., Smith, T., Wysham, N. G., Fudim, M., & Swaminathan, S. (2021) · Journal of Cardiovascular Translational Research · doi:10.1007/s12265-021-10151-7 · Cited by 15
Interdisciplinary collaboration in critical care alarm research: A bibliometric analysis
Agha-Mir-Salim, L., McCullum, L. B., Dähnert, E., Scheel, Y.-D., Wilson, A., Carpio, M., Chan, C., Lo, C., Maher, L., Dressler, C., et al. (2023) · International Journal of Medical Informatics · doi:10.1016/j.ijmedinf.2023.105285 · Cited by 15
Karway, G. K., Koyner, J. L., Caskey, J., Spicer, A. B., Carey, K. A., Gilbert, E. R., Dligach, D., Mayampurath, A. M., Afshar, M., & Churpek, M. M. (2023) · JAMIA Open · doi:10.1093/jamiaopen/ooad109 · Cited by 17
“51 Furthermore, our models are developed in 1 setting and validated in another, enabling investigation of clinical outcomes across different settings and generalization to diverse patient populations.”
MethodActionable Ventricular Tachycardia During In-Hospital ECG Monitoring and Its Impact on Alarm Fatigue
Pelter, M. M., Suba, S., Sandoval, C. P., Zègre‐Hemsey, J. K., Berger, S., Larsen, A. M., Badilini, F., & Hu, X. (2020) · Critical Pathways in Cardiology A Journal of Evidence-Based Medicine · doi:10.1097/hpc.0000000000000216 · Cited by 18
“Purpose: This study was designed to determine (1) the proportion of clinically actionable true V-tach events; (2) whether true actionable versus nonactionable V-tach differs in terms of heart rate and/or duration (seconds); and (3) if actionable V-tach is associated with adverse outcomes.”
BackgroundInfluentialSelf-citationResearch : Association of Low-Amplitude QRSs with False-Positive Asystole Alarms
Pelter, M. M., Fidler, R., & Hu, X. (2016) · Biomedical Instrumentation & Technology · doi:10.2345/0899-8205-50.5.329 · Cited by 18
Self-citationAlarm fatigue in the ward: An acoustical problem?
Kristensen, M. S., Edworthy, J. R., & Özcan, E. (2016) · SoundEffects - An Interdisciplinary Journal of Sound and Sound Experience · doi:10.7146/se.v6i1.24915 · Cited by 25
“In a recent fi eld study Drew et al. (2014) treated the total number of auditory and visual alarms as one uni fi ed variable without any further discussion of the effect of the distribution between auditory and visual alarms, let alone the attributes of the different kinds of auditory and visual…”
BackgroundAlarm-Fatigue – wieviel Alarm verträgt der Mensch?
Lorenz, B., Peters, J., & Frey, U. H. (2017) · AINS - Anästhesiologie · Intensivmedizin · Notfallmedizin · Schmerztherapie · doi:10.1055/s-0042-118618 · Cited by 15
“[18] Drew BJ, Harris P, Zègre-Hemsey JK et al.”
BackgroundAysha, Z. M., & Ahmed, S. E. S. (2019) · American journal of nursing research · doi:10.12691/ajnr-7-5-16 · Cited by 23
“[5,6] Excessive clinical alarms can cause an alarm hazard, that leads to inappropriate responses by clinicians to these alarms, such as ignoring, and disabling alarms without any action that may result in unsafe nursing practices and effects on patient's recovery and increase length of intensive care unit stay [7].”
BackgroundNot to Cry Wolf: Distantly Supervised Multitask Learning in Critical CarePreprint
Schwab, P., Keller, E., Muroi, C., Mack, D. J., Strässle, C., & Karlen, W. (2018) · arXiv (Cornell University) · doi:10.48550/arxiv.1802.05027 · Cited by 18
“False alarms are an enormous mental burden for clinical staff and are extremely dangerous to patients, as alarm fatigue and desensitisation may lead to clinically important alarms being missed (Drew et al., 2014).”
BackgroundGálvez‐Barrón, C., Pérez‐López, C., Álvarez, F. V., Ribas, J., Formiga, F., Chivite, D., Boixeda, R., Iborra, C., & Rodríguez‐Molinero, A. (2023) · Scientific Reports · doi:10.1038/s41598-023-39329-6 · Cited by 14
Bai, Y., Do, D. H., Ding, Q., Arroyo-Palacios, J., Shahriari, Y., Pelter, M. M., Boyle, N., Fidler, R., & Hu, X. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2586443 · Cited by 14
“Unfortunately, the ability to effectively and precisely detect and anticipate patient status changes using current patient monitors remains unsatisfactory as evidenced by the widespread alarm fatigue problems in hospitals [1], [2].”
Self-citationZong, W., Nielsen, L., Gross, B. D., Brea, J., & Frassica, J. J. (2016) · Physiological Measurement · doi:10.1088/0967-3334/37/8/1355 · Cited by 14
Searching for meaning in sound: Learning and interpreting alarm signals in visual environments.
McDougall, S. J. P., Edworthy, J. R., Sinimeri, D., Goodliffe, J., Bradley, D., & Foster, J. (2019) · Journal of Experimental Psychology Applied · doi:10.1037/xap0000238 · Cited by 14
McFarlane, D. C., Doig, A. K., Agutter, J. A., Mercurio, J. L., Mittu, R., Brewer, L., & Syroid, N. D. (2016) · The Journal of Defense Modeling and Simulation Applications Methodology Technology · doi:10.1177/1548512916667246 · Cited by 15
Principles of Automation for Patient Safety in Intensive Care: Learning From Aviation
Dominiczak, J., & Khansa, L. (2018) · The Joint Commission Journal on Quality and Patient Safety · doi:10.1016/j.jcjq.2017.11.008 · Cited by 14
Pelter, M. M., Loranger, D. L., Kozik, T. M., Kedia, A., Ganchan, R. P., Ganchan, D., Hu, X., & Carey, M. G. (2015) · The Journal of Cardiovascular Nursing · doi:10.1097/jcn.0000000000000310 · Cited by 15
“enzyme inhibitor 135 (46) 63 (51) 72 (43) .”
Self-citationPsychophysiological models of hypovigilance detection: A scoping review
Marois, A., Kopf, M., Fortin, M., Huot‐Lavoie, M., Martel, A., Boyd, J. G., Gagnon, J., & Archambault, P. M. (2023) · Psychophysiology · doi:10.1111/psyp.14370 · Cited by 15
Lee, S., Zheng, G., Koh, J., Li, H., Xu, Z., Cho, S. P., Im, S. I., Braverman, V., & Jeong, I. C. (2025) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2025.108898 · Cited by 13
Mehta, J. H., Cattano, D., Brayanov, J. B., & George, E. (2017) · BMC Anesthesiology · doi:10.1186/s12871-017-0352-0 · Cited by 13
“Moreover, both SpO2 and EtCO2 monitoring systems are plagued by false alarms in the post-operative setting [11, 12].”
BackgroundSuba, S., Fleischmann, K. E., Schell-Chaple, H. M., Prasad, P. A., Marcus, G. M., Hu, X., & Pelter, M. M. (2021) · PLoS ONE · doi:10.1371/journal.pone.0261712 · Cited by 14
“5 million total alarms or 18 PVCs/monitoring hour per patient [7].”
BackgroundSelf-citationPerrin, K., Ernst, N., Nelson, T. S., Sawyer, M. D., Pfoh, E. R., & Cvach, M. M. (2016) · Journal of Nursing Care Quality · doi:10.1097/ncq.0000000000000230 · Cited by 13
Contribution of Electrocardiographic Accelerated Ventricular Rhythm Alarms to Alarm Fatigue
Suba, S., Sandoval, C. P., Zègre‐Hemsey, J. K., Hu, X., & Pelter, M. M. (2019) · American Journal of Critical Care · doi:10.4037/ajcc2019314 · Cited by 16
Self-citationCertainty in QRS detection with artificial neural networks
Chromik, J., Pirl, L., Beilharz, J., Bert Arnrich, A., & Polze, A. (2021) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2021.102628 · Cited by 13
Probabilistic model-based approach for heart beat detection
Chen, H., Erol, Y., Shen, E., & Russell, S. (2016) · Physiological Measurement · doi:10.1088/0967-3334/37/9/1404 · Cited by 14
Ashrafi, S., Mehri, S. N., & Nehrir, B. (2017) · Critical Care Nursing · doi:10.5812/ccn.11647 · Cited by 14
“Physiologic monitors generate fatigue in nurses with annoying alarms and visual warnings and this threatens the patients’ lives (8).”
BackgroundFrequency and Duration of Infusion Pump Alarms: Establishing National Benchmarks
Vitoux, R. R., Schuster, C., Glover, K. R., & Dekker, M. (2018) · Biomedical Instrumentation & Technology · doi:10.2345/0899-8205-52.6.433 · Cited by 13
Gajowniczek, K., Grzegorczyk, I., & Ząbkowski, T. (2019) · Sensors · doi:10.3390/s19071588 · Cited by 12
“[2], the rates of false alarms in ICUs might be as high as almost 90%.”
BackgroundDing, C., Guo, Z., Rudin, C. D., Xiao, R., Shah, A. J., Do, D. H., Lee, R. J., Clifford, G. D., Nahab, F. B., & Hu, X. (2024) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2024.3360952 · Cited by 15
“We could not find studies that quantify the accuracy of these alarms, and it is reasonable to assume that these alarms are not accurate based on previous studies of other more lethal cardiac arrhythmia alarms [24], [25].”
ResultSelf-citationAlarm fatigue in the perception of medical soundscapes
Kristensen, Edworthy, J. R., Özcan, E., & Denham, S. L. (2015) · Cited by 12
Pelter, M. M., Carey, M. G., Al‐Zaiti, S. S., Zègre‐Hemsey, J. K., Sommargren, C. E., Isola, L., Prasad, P. A., MORTARA, D. W., & Badilini, F. (2023) · Annals of Noninvasive Electrocardiology · doi:10.1111/anec.13054 · Cited by 12
“These confounders have been identified as a source of false VT in other studies (Aboukhalil et al., 2008; Drew et al., 2014; Harris et al., 2017; Nguyen et al., 2020; Pelter et al., 2020; Watanakeeree et al., 2021).”
BackgroundMethodInfluentialSelf-citationIndividualized Dynamic Patient Monitoring Under Alarm Fatigue
Piri, H., Huh, W. T., Shechter, S. M., & Hudson, D. (2022) · Operations Research · doi:10.1287/opre.2022.2300 · Cited by 13
Herráiz, Á. H., Martínez‐Rodrigo, A., Bertomeu‐González, V., Martín, Ó. A., Rieta, J. J., & Alcaraz, R. (2024) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2023.105920 · Cited by 14
Mumuni, A. N., Hasford, F., Udeme, N. I., Dada, M. O., & Awojoyogbe, B. O. (2022) · Physical Sciences Reviews · doi:10.1515/psr-2022-0121 · Cited by 18
Dai, L., Yang, X., Li, H., Zhao, X., Lin, L., Jiang, Y., Wang, Y., Li, Z., & Shen, H. (2024) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2024.102772 · Cited by 11
ECG Monitoring during End of Life Care: Implications on Alarm Fatigue
Suba, S., Sandoval, C. P., Hu, X., & Pelter, M. M. (2019) · Multimodal Technologies and Interaction · doi:10.3390/mti3010018 · Cited by 11
Self-citationExplaining an increase in predicted risk for clinical alerts
Hardt, M., Rajkomar, A., Flores, G., Dai, A. M., Howell, M., Corrado, G. S., Cui, C. M., & Hardt, M. (2020) · ACM Conference on Health, Inference, and Learning (CHIL) · doi:10.1145/3368555.3384460 · Cited by 16
“Alerts should have high precision – too many false alerts can lead to fatigue [4].”
BackgroundDawson, D., Barham, R. G., Hamilton, M. A., & Philips, B. J. (2021) · Applied Acoustics · doi:10.1016/j.apacoust.2021.108485 · Cited by 11
Intensivist-to-patient ratios in ICUs: is there a number?
Ward, N. S., & Howell, M. D. (2015) · Current Opinion in Anaesthesiology · doi:10.1097/aco.0000000000000170 · Cited by 12
“For example, in one study of just 461 patients over a 31day period, there were 2 558 760 unique alarms [71].”
BackgroundCan Big Data and Machine Learning Improve Our Understanding of Acute Respiratory Distress Syndrome?
Bhattarai, S., Gupta, A., Ali, E. A., Ali, M., Riad, M., Adhikari, P., & Mostafa, J. A. (2021) · Cureus · doi:10.7759/cureus.13529 · Cited by 15
“Clinicians also fear the possibility of “alarm fatigue” which creates an unsafe patient environment because a life-2021 threatening event may be missed due to sensory overload created by alerts, especially in the ICU setting [40].”
BackgroundBergstedt, A., Hilliard, B., Alabsi, S. M., Usher, M. G., Peters, M., Grace, J., Melton, G. B., Beebe, T. J., & Pestka, D. L. (2024) · Journal of the American Heart Association · doi:10.1161/jaha.123.031523 · Cited by 10
“5,6 Therefore, many health care clinical staff and providers may become desensitized and experience alarm fatigue, 7 which has been shown to directly affect patient outcomes.”
Informing Healthcare Alarm Design and Use: A Human Factors Cross-Industry Perspective
Pruitt, Z. M., Bocknek, L. S., Busog, D.-N. C., Spaar, P. A., Milicia, A. P., Howe, J. L., Franklin, E. S., Krevat, S. A., Jones, R., & Ratwani, R. M. (2023) · Patient Safety · doi:10.33940/med/2023.3.1 · Cited by 12
Knowledge of patient observation among critical care nurses
Alastalo, M., Salminen, L., Vahlberg, T. J., & Leino‐Kilpi, H. (2020) · Nursing in Critical Care · doi:10.1111/nicc.12573 · Cited by 10
“This involves identifying the factors that cause observational bias, such as technical errors, artefacts in monitoring and incorrect physical assessment [5], [13].”
BackgroundImproving appropriate imaging for non-specific low back pain
Al-Hihi, E. M., Gibson, C., Lee, J., Mount, R. R., Irani, N., & McGowan, C. (2022) · BMJ Open Quality · doi:10.1136/bmjoq-2021-001539 · Cited by 10
Block, F. E., & Block, F. E. (2015) · Biomedical Instrumentation & Technology · doi:10.2345/0899-8205-49.6.423 · Cited by 10
Heartbeat fusion algorithm to reduce false alarms for arrhythmias
Daluwatte, C., Johannesen, L., Vicente, J., Scully, C. G., Galeotti, L., & Strauss, D. G. (2015) · Computing in cardiology · doi:10.1109/cic.2015.7411018 · Cited by 10
“However, physiological signals from different monitors contain complementary information that is used by clinicians to recognize false alarms.”
BackgroundInfluentialAcute Coronary Syndrome and ST Segment Monitoring
Carey, M. G. (2016) · Critical Care Nursing Clinics of North America · doi:10.1016/j.cnc.2016.04.006 · Cited by 10
Debatable issues in automated ECG reporting
Macfarlane, P. W., Mason, J. W., Kligfield, P., Sommargren, C. E., Drew, B. J., van Dam, P. M., Abächerli, R., Albert, D. J., & Hodges, M. (2017) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2017.08.027 · Cited by 10
“in one study, 91% of 6196 ST alarms were considered to be false positive[16].”
BackgroundSelf-citationECG alarms during left ventricular assist device (LVAD) therapy in the ICU
Watanakeeree, K., Suba, S., Mackin, L. A., Badilini, F., & Pelter, M. M. (2021) · Heart & Lung · doi:10.1016/j.hrtlng.2021.03.080 · Cited by 13
A Risk-Averse Mechanism for Suicidality Assessment on Social Media
Sawhney, R., Neerkaje, A. T., & Gaur, M. (2022) · Proceedings of the conference - Association for Computational Linguistics. Meeting · doi:10.18653/v1/2022.acl-short.70 · Cited by 16
Rehman, A., & Mun, S. (2026) · Sensors · doi:10.3390/s26030863 · Cited by 11
“Studies have documented that 72–99% of physiological monitor alarms are false or clinically irrelevant, leading clinicians to disable alarms, delay responses, or ignore alerts entirely [37].”
Nurses’ use of monitors in patient surveillance: an ethnographic study on a coronary care unit
Freysdóttir, G. R., Björnsdóttir, K., & Svavarsdóttir, M. H. (2018) · European Journal of Cardiovascular Nursing · doi:10.1177/1474515118816930 · Cited by 9
Artificial intelligence changing the future of healthcare diagnostics
Mishra, V. H., Ugemuge, S., & Tiwade, Y. R. (2023) · Journal of Cellular Biotechnology · doi:10.3233/jcb-230118 · Cited by 11
“Human supervision and surveillance are still necessary to ensure the best possible care [45].”
ARTEMIS: An alarm threshold and policy mining system for the intensive care unit
Chromik, J., Flint, A. R., & Arnrich, B. (2024) · International Journal of Medical Informatics · doi:10.1016/j.ijmedinf.2024.105349 · Cited by 10
Yahyaei, S., Khoddam, H., Alinaghimaddah, S., & Modanloo, M. (2023) · Payām-i naw. · doi:10.61186/jgbfnm.20.1.20 · Cited by 9
Sowan, A. K., Reed, C. C., & Staggers, N. (2016) · JMIR Human Factors · doi:10.2196/humanfactors.6427 · Cited by 9
“Previous research presented the number and types of alarms, limits of parameters, and changes in parameters’ limits [10-13,15-17,20-23,25-27].”
BackgroundJalilian, L., & Khairat, S. (2022) · PubMed · Cited by 16
Derivation and Validation of an Algorithm to Detect Stroke Using Arm Accelerometry Data
Messe, S. R., Kasner, S. E., Cucchiara, B. L., McGarvey, M. L., Cummings, S., Acker, M. A., Desai, N. D., Atluri, P., Wang, G. J., Jackson, B. M., et al. (2023) · Journal of the American Heart Association · doi:10.1161/jaha.122.028819 · Cited by 9
Assessment of Alarm Fatigue Risk Among Oncology Nurses in Jordan
Al-Quraan, H. A., Eid, A., & Alloubani, A. (2023) · SAGE Open Nursing · doi:10.1177/23779608231170730 · Cited by 11
“The authors concluded that adjusting the setting may decrease the number of alarms, and decreasing the number leads to decreased alarm fatigue (Drew et al., 2014).”
BackgroundResultInfluentialEvaluating performance of early warning indices to predict physiological instabilities
Scully, C. G., & Daluwatte, C. (2017) · Journal of Biomedical Informatics · doi:10.1016/j.jbi.2017.09.008 · Cited by 9
Medical Cyber-Physical Systems: IoMT Applications and Challenges
Watson, A., Park, J., Pugh, S., Sokolsky, O. V., Weimer, J. E., & Lee, I. (2022) · Asilomar Conference on Signals, Systems and Computers (ACSSC) · doi:10.1109/ieeeconf56349.2022.10052004 · Cited by 9
“In these environments, clinicians can experience alarm fatigue, desensitization to alarms that can cause missed critical alarms or delayed responses [18], [19].”
BackgroundLiu, N. T., Holcomb, J. B., Wade, C. E., Darrah, M. I., & Salinas, J. (2015) · Journal of Medical Engineering & Technology · doi:10.3109/03091902.2015.1054524 · Cited by 9
“In principle it would seem that more data are better, but, as recent work has shown with bedside alarms [17], that may not necessarily be the case.”
Background
Metadata
- Authors
- Drew, B. J., Harris, P., Zègre-Hemsey, J. K., Mammone, T., Schindler, D., Salas-Boni, R., Bai, Y., Tinoco, A., Ding, Q., & Hu, X.
- Affiliations
- University of California, San Francisco (US)
- University of California San Francisco Medical Center (US)
- Corresponding author
- Drew, Barbara J.
- Venue
- PLoS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2014-10-22
- Volume
- 9
- Issue
- 10
- Pages
- e110274
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC4206416
- OpenAlex
- W2001741247
- Cited by
- 438
- References
- 23
- Citations by year
- 2026: 51 · 2025: 31 · 2024: 25 · 2023: 37 · 2022: 26 · 2021: 58 · 2020: 29 · 2019: 38 · 2018: 42 · 2017: 36 · 2016: 43 · 2015: 21 · 2014: 1
- FWCI
- 33.15
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- PLoS ONE (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Carolina Digital Repository (University of North Carolina at Chapel Hill) (submittedVersion)
- Dominican Scholar (Dominican University of California) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- eScholarship (California Digital Library) (submittedVersion)
- UNC Libraries
- Funders
- University of California, San Francisco
- Topics
- Health Sciences › Medicine › Surgery › Healthcare Technology and Patient Monitoring
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Patient Safety and Medication Errors
- Health Sciences › Health Professions › Health Information Management › Electronic Health Records Systems
- Keywords
- alarm fatigue, intensive care unit patients, false positive alerts, alarm settings, ECG leads, QRS amplitude, atrial fibrillation, ventricular tachycardia, bundle branch block, persistent atrial fibrillation, patient safety
- MeSH terms
- Adult, Aged, Arrhythmias, Cardiac, Atrial Fibrillation, Diagnostic Errors, Electrocardiography, Equipment Failure, False Positive Reactions, Female, Humans, Intensive Care Units, Male, Middle Aged, Monitoring, Physiologic, Critical Illness, Clinical Alarms
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.